Acta Cryst. (2009). E65, o538 [ doi:10.1107/S1600536809005005 ]
Online 18 February 2009
No syntax errors found
Alert level A PLAT027_ALERT_3_A _diffrn_reflns_theta_full (too) Low ............ 61.20 Deg.
| Author Response: see PLAT023 |
Alert level B THETM01_ALERT_3_B The value of sine(theta_max)/wavelength is less than 0.575 Calculated sin(theta_max)/wavelength = 0.5694
| Author Response: see PLAT023 |
PLAT023_ALERT_3_B Resolution (too) Low [sin(theta)/Lambda < 0.6].. 61.40 Deg.
| Author Response: The diffractometer Gemini used for the data collection has two X-ray tubes, Mo and Cu. The collimator of Mo tube limits the maximal theta angle of CCD detector and the required resolution cannot be easily achieved with Cu radiation. In order to achieve resolution 0.83 instead of 0.89 almost 2000 frames would be added to the experiment or we would need to measure with removed Mo collimator; the first way is not efficient, the second one is laborius ans unsafe. Because the required resolution is not very far of the one actually reached we are using the normal data collection mode. |
Alert level C Value of measurement temperature given = 295.000 Value of melting point given = 0.000 GOODF01_ALERT_2_C The least squares goodness of fit parameter lies outside the range 0.80 <> 2.00 Goodness of fit given = 2.320
| Author Response: see PLAT087 |
PLAT087_ALERT_2_C Unsatisfactory S value (Too High) .............. 2.32
| Author Response: Program Jana has a different weighting scheme of SHELX, see _refine_special_details, and th eresulting g.o.f is usually higher compared with SHELX. Another reason for higher g.o.f. was high exposition time used for data collection of this compound and merohedric twinning with non-equal twin fractions. |
1 ALERT level A = In general: serious problem 2 ALERT level B = Potentially serious problem 2 ALERT level C = Check and explain 0 ALERT level G = General alerts; check 0 ALERT type 1 CIF construction/syntax error, inconsistent or missing data 2 ALERT type 2 Indicator that the structure model may be wrong or deficient 3 ALERT type 3 Indicator that the structure quality may be low 0 ALERT type 4 Improvement, methodology, query or suggestion 0 ALERT type 5 Informative message, check