J. Appl. Cryst. (1999). 32, 65-70 [doi:10.1107/S0021889898010474]
Abstract: An improved method for intensity-uniformity calibration of diffraction data collected on spiral-readout image-plate (IP) systems is described. This technique is applicable to all types of spiral-readout IP detectors. The procedure utilizes an attenuated direct-beam scan of the IP to generate a radial-sensitivity calibration table. Exposure and scanning of the calibration frame are done on the same time scale as typical data collections, and require no additional equipment. Specific examples are presented for use with Mac Science DIP2000 systems. The new radial calibration is shown to reduce significantly structure-based R factors. The improved radial calibration is also shown to lower Rmerge when the IP is offset from the beam center. In addition to improving data quality and statistics, this method provides a quick and simple diagnostic tool to monitor changes in the sensitivity of the IP detector as a function of age.
Online 1 February 1999
![]() ![]() PDF file (1170.1 kbytes) | |
To open or display or play some files, you may need to set your browser up to use the appropriate software. See the full list of file types for an explanation of the different file types and their related mime types and, where available links to sites from where the appropriate software may be obtained.