J. Appl. Cryst. (2008). 41, 373-376 [ doi:10.1107/S0021889808004652 ]
Abstract: To ensure concentricity of the
-2
rotation stages, but with a very limited 2
range of rotation, an alignment technique using two dial gauges mounted at right angles to each other (X and Y axes) is described. The deflections of the orthogonal dial gauges are measured as a function of 2
, with their probes impinging on a cylindrical pin centred on the
axis. The X and Y corrections required to bring the two rotation stages into concentricity can then be found using a least-squares fitting via a freestanding program or via the `solver' optimization within the Microsoft Excel spreadsheet software. The same dial gauges provide guidance for applying the required positional corrections. An Excel spreadsheet for determining the X and Y corrections via least squares has been deposited.
Keywords: neutron powder diffraction; off-line alignment; diffractometry; concentricity; circles of rotation.
![]() ![]() excel file (19.0 kbytes) | |
![]() ![]() OpenDocument spreadsheet file (22.0 kbytes) | |
To open or display or play some files, you may need to set your browser up to use the appropriate software. See the full list of file types for an explanation of the different file types and their related mime types and, where available links to sites from where the appropriate software may be obtained.