Figure 5
The scattered intensities for two samples of latex particles (R ≃ 54 nm) determined over a broad Q range using different experimental setups at the V16/VSANS instrument (black line) and the V4 instrument (red line). The data from V4 comprise five individual measurements at three different sample-to-detector distances (1, 4 and 16 m) and three different wavelengths (4.6, 9.1 and 18.2 Å). The data set from the V16/VSANS instrument consists of three individual measurements at two sample-to-detector distances (11.3 and 2 m) and three chopper configurations (compare Fig. 4). |