data validation criteria - PLATON tests


PLAT020 Type_3 Check Rint
 The value of Rint (i.e. _diffrn_reflns_av_R_equivalents) should normally be
 considerably less than 0.12 and in the order of magnitude of the reported
 R-values. Rint may be relatively meaningless when based on a very limited
 number of averaged data. Higher values should be accompanied by a suitable
 explanation in the _publ_section_exptl_refinement section. However,
 authors should first ensure that there are not overlooked problems
 associated with the data or the space-group. Elevated values for 
 _diffrn_reflns_av_R_equivalents may be indicative of a need to recollect 
 the data from a crystal of higher quality or that there is a problem with
 the data treatment. Consider the following:
 (a) The absorption corrections are inadequate or inappropriate.
 (b) The overall quality of the data may be poor due to the crystal quality.
 (c) The crystal is very weakly diffracting, so that a large proportion of
     essentially "unobserved" reflections are being used in the refinement.
     You should consider using a better crystal or a data collection at
     low temperature and/or, if the compound is organic, using Cu radiation.
 (d) You are working in the wrong crystal system or Laue group.
 (e) You have only a very small number of equivalent reflections, which
     may lead to artificially high values of _diffrn_reflns_av_R_equivalents 
 Note that if _diffrn_reflns_av_sigmaI/netI is also large, the quality
 of the data should be considered to be suspect. 

PLAT021 Type_1 Check Expected number of Reflections (Max = 1 Centro, 2 - non-centro)
 The expected number of reflections corresponds to that in the asymmetric
 unit of the Laue group. Expected ratio: less-or-equal 1 for centro symmetric
 structures and less than 2 for non-centrosymmetric structures.
 Reasons to exceed those numbers can be:
  1 - Systematic extinctions not omitted from the obsd data count.
  2 - Refinement with redundant (i.e. not merged/unique) data set.

PLAT022 Type_3 Check Expected number of Reflections
 Test for data completeness. The ratio of the reported number of unique
 reflections and expected number of reflections for the resolution given is
 reported. The ratio can be low due to a missing cusp of data when collected
 with a 2D-detector. Alternatively, the wrong asymmetric part of reciprocal
 space was collected on a serial detector system.

PLAT023 Type_3 Check Theta-Max
 Check resolution of the data set. Alert is issued when sin(theta)/lambda < 0.6

PLAT024 Type_4 Check for required Friedel pair averaging Z<=Si
 Averaging of Friedel pairs is strongly advised when indicated by a
 large su on the Flack parameter. Large su values indicate that the
 anomalous scattering power is too small in combination with the data
 quality at hand to merit refinement with a non-averaged dataset. This
 will generally be the case with MoKa datasets for structures containing
 atoms not heavier than Si. The value of the Flack parameter will be 
 largely meaningless anyway for large su values. 
 
 Use MERG 3 or MERG 4 in case of refinement with the SHELXL97 program.
 Non-compliance with the above for valid scientific reasons should be
 discussed in detail in the experimental section of the paper. 

PLAT025 Type_1 Check for Hmin..Lmax
 Check reported h,k,l - range with calculated range based on reported theta-max.

PLAT026 Type_3 Check for weak data
 Check whether a sufficient fraction of the unique data is indeed above the
 2 sigma level.

PLAT027 Type_3 Check _diffrn_reflns_theta_full
 Ideally (and a requirement for publication in Acta Crystallographica),
 the dataset should be essentially complete, as defined by 
 -diffrn-measured-fraction-theta-full (close to 1.0), up to 
 sin(theta)/lambda = 0.6 (i.e. 25.24 degrees MoKa).       
 The three major causes of incomplete data sets are:
 1 - A missing cusp of data due to data collection by rotation around
     the spindle axis only (standard on some image-plate systems).
     Cure: collect an additional data set after remounting the crystal.
 2 - The DENZO image processing package has problems with certain strong
     reflections. They are often excluded from the data set.
     Cure: Add an additional scan at lower power setting in order to
     include strong low order reflections.
 3 - Incomplete scans.
 

PLAT028 Type_3 Check _diffrn_measured_fraction_theta_max
 Ideally, the reported '_diffrn_measured_fraction_theta_max' value,
 corresponding to theta-max, should be close to 1.0.   

PLAT029 Type_3 Check _diffrn_measured_fraction_theta_full
 Ideally (and a requirement for publication in Acta Crystallographica),
 this fraction should be close to 1.0 for theta-full greater or equal
 to sin(theta/lambda) = 0.6 (i.e. 25.24 degrees for MoKa and 67.7 degrees
 for CuKa  radiation).
 The three major causes of incomplete data sets are:
 1 - A missing cusp of data due to data collection by rotation around
     the spindle axis only (standard on some image-plate systems).
     Cure: collect an additional data set after remounting the crystal.
 2 - The DENZO imageprocessing package has problems with certain strong
     reflections. They are often excluded from the data set.
     Cure: Add an additional scan at lower power setting in order to
     include strong low order reflections.
 3 - Incomplete scans, possibly based on erroneously assumed higher than
     actual symmetry.
 Note: The default value of _diffrn_measured_fraction_theta_full that
 is automatically calculated and inserted in the CIF by SHELXL-97 might
 generate A-level ALERTS when significant numbers of reflections are 
 missing at higher theta values. In order to avoid such an ALERT,
 substitute the values calculated with the SHELXL instruction 'ACTA 50'
 for _diffrn_reflns_theta_full and _diffrn_measured_fraction_theta_full
 respectively. For Mo-radiation, corresponding values of 25 degees (or
 higher) and 0.99 (or higher) are expected. 
 PLATON may be used to analyse the case at hand (by invoking either the
 'FCF-VALIDATION' mode or the 'ASYM-VIEW' mode).

PLAT030 Type_1 Check _diffrn_reflns_number > reflns_number_total
 The number of measured reflections should be equal or greater than the
 number of unique reflections. 

PLAT031 Type_4 Check need for Extinction Correction Parameter
 This test checks whether a refined extinction parameter is meaningful
 i.e. whether its value is significantly larger than its corresponding s.u.
 If not, this parameter should be removed from the model and the structure
 refined without this meaningless additional parameter.
 The current default gives a warning when its value is within 3.33 s.u.
 SHELXL97-2 will not allow negative values leading to ill-convergence and
 non-zero maximum shift/error values: remove extinction parameter from the
 refinement.

PLAT032 Type_4 Check su Flack Parameter
 Check the validity of the absolute structure determination.
 A high su indicates that the experimental data do not support the
 determination of the absolute structure. This will generally be the case
 with light atom MoKa data where f" is nearly zero.
 Note: Use the TWIN & BASF 0.0 instructions in SHELXL97. The default 
 FLACK parameter is not always reliable, in particular when strongly 
 correlated with the position of the origin (e.g. along y in space-group P21). 
 Please refer to Flack,H.D. & Bernardinelli, G. (1999) Acta Cryst. A55,
 908-915 and (2000) J. Appl. Cryst., 33, 1143-1148.

PLAT033 Type_4 Check Flack Parameter value
 Check the relevance/validity of the absolute structure determination.
 Please refer to Flack,H.D. & Bernardinelli, G. (1999) Acta Cryst. A55,
 908-915 and (2000) J. Appl. Cryst., 33, 1143-1148.
 A value of the Flack parameter that deviates significantly from zero
 (taking into account the associated s.u.) might indicate that the
 absolute structure should be inverted in case of a value closer to 1.0
 than to zero. A value close to 0.5 may be indicative of an inversion
 twin or a missed centre of inversion. 
 For valid absolute structure assignments, abs(x) should be less thans 2 * su,
 with su < 0.04. For enantiopure compounds, su should be less than 0.1.
 

PLAT034 Type_1 Check for Flack parameter value specified Z>Si, non-centro
 No Flack parameter value given for non-centrosymmetric structure with
 heaviest atom Z > Si. This might be intentional.

PLAT035 Type_1 Check for _chemical_absolute_configuration
 Options are 'rm', 'ad', 'rmad', 'syn', 'unk' or '.'
                      
 rm   : absolute configuration established by the structure determination
        of a compound containing a chiral reference molecule of known
        absolute configuration.
 ad   : absolute configuration established by anomalous dispersion effects
        in diffraction measurements on the crystal.
 rmad : absolute configuration established by the structure determination
        of a compound containing a chiral reference molecule of known
        absolute configuration and confirmed by anomalous dispersion
        effects in diffraction measurements on the crystal.
 syn  : absolute configuration has not been established by anomalous
        dispersion effects in diffraction measurements on the crystal.
        The enantiomer has been assigned by reference to an unchanging
        chiral centre in the synthetic procedure.
 unk  : absolute configuration is unknown, there being no firm chemical
        evidence for its assignment to hand and it having not been 
        established by anomalous dispersion effects in diffraction 
        measurements on the crystal. An arbitrary choice of enantiomer
        has been made.
 .    : inapplicable.
   

PLAT036 Type_1 Check for missing Flack Parameter su
 No standard uncertainty found for the Flack parameter.
 When the structure refinement was done with SHELXL97-2, the likely 
 reason for this is a missing BASF instruction. This applies in
 particular when the associated Flack parameter has the value 0.000.
 No valid conclusions on the absolute structure can be drawn.

PLAT037 Type_1 Check _diffrn_reflns_theta_full
 No information is given about the theta value for which the dataset is
 complete, subject to the percentage given with the dataname 
 _diffrn_measured_fraction_theta_full. 
 

PLAT038 Type_1 Check _diffrn_measured_fraction_theta_max
 This fraction should be specified in combination with the theta value
 given with the dataname _diffrn_reflns_theta_full.

PLAT039 Type_1 Check _diffrn_measured_fraction_theta_full
 This fraction should be specified in combination with the value for theta-max.

PLAT040 Type_1 Test for H-atoms [0,1]
 Alert for 'no H-atoms' in CIF. This is unusual for carbon containing compounds,
 but may be correct.

PLAT041 Type_1 Test SumFormula
 In the ideal case, both SumFormula strings (reported and calculated) should
 be identical. If not, the reason for the difference should be clear. Examples
 are cases where populations do not add up to integer numbers, or when solvent
 molecules have been SQUEEZED.
 Note: SHELXL97 reports population parameters in the CIF with two decimals 
 only. This may lead to non-integer atom counts in cases of disorder
 due to rounding.
 Note: Alerts _041, _042 & _045 can probably be ignored when the relevant
 values differ by the same factor.

PLAT042 Type_1 Test MoietyFormula
 In the ideal case, the MoietyFormula string as reported should be identical
 to the MoietyFormula string calculated from the data in the CIF. If not, the
 reason should be clear. Examples are cases where there is no separating space
 between two element names or cases where populations do not add up to 
 integer numbers or when moieties are separated by '.' instead of ','.
 Example: NO3 should be given as N O3  
 Note: Alerts _041, _042 & _045 can probably be ignored when the relevant
 values differ by the same factor.

PLAT043 Type_1 Test for MolWeight
 Note: atomic weights used in the calculation of the molecular weight are
 taken from Inorg. Chim. Acta 217 (1994) 217-218 which deviate in a few cases
 slightly from the older values used in SHELXL97-2.

PLAT044 Type_1 Check Reported with calculated density
 In the ideal case, both data items should be the same within a small 
 tolerance. If not, the reason should be clear.
 

PLAT045 Type_1 Check Reported and Calculated Z
 In the ideal case, both data items (Z(calc) & Z(reported)) should be
 the same. If not, the reason for the difference should be clear. An example
 is the situation where PLATON gives Z = 1 when the program cannot work out
 a proper Z.
 Note: Alerts _041, _042 & _045 can probably be ignored when the relevant
 values differ by the same factor.
 

PLAT046 Type_1 Check Reported Density with calculated density from Z*MW
 D(calc) as calculated from the reported Z and MW is compared for consistency 
 with the reported d(calc).

PLAT047 Type_1 Test SumFormula Given
 The Sumformula, corresponding with the Moietyformula, should be given.

PLAT048 Type_1 Test MoietyFormula Given
 The Moiety formula (i.e. the specification of the various species in the
 structure) should be given in the CIF.
 Example: '(Cd 2+)3, (C6 N6Cr 3-)2, 2(H2 O)'

PLAT049 Type_1 Check Calculated Density .GT. 1.0
 The calculated density will with a few exceptions be larger than 1.0.
 A smaller value may indicate either an incomplete model or incorrect
 symmetry. (e.g. a missing 'bar' in P-1 etc.)

PLAT050 Type_1 Test for mu given [0,1]
 The linear absorption coefficient corresponding to the Sumformula should be
 given.

PLAT051 Type_1 Test for difference mu(cif) with mu(calc) [%]
 In the ideal case, both data items should be the same within a small
 tolerance. If not, the reason should be clear.
 

PLAT052 Type_1 Test for specification absorption correction method [0,1]
 The treatment/method of absorption(correction) should be given explicitly.
 Set _exptl_absorpt_correction_type to 'none when no correction is done.
 Other recognized values are 'psi-scan', 'empirical', 'multi-scan',
 'refdelf', 'analytical', 'numerical', 'gaussian'.
 

PLAT053 Type_1 Test for specification xtal_dimension_min [0,1]
 The smallest crystal dimension should be supplied in the CIF.
 The expected value should be a real number (i.e. not 0.35mm)

PLAT054 Type_1 Test for specification xtal_dimension_mid [0,1]
 The medium crystal dimension should be supplied in the CIF.
 The expected value should be a real number (i.e. not 0.35mm)

PLAT055 Type_1 Test for specification xtal_dimension_max [0,1]
 The largest crystal dimension should be supplied in the CIF.
 The expected value should be a real number (i.e. not 0.35mm)

PLAT056 Type_1 Test for specification xtal_radius [0,1]

PLAT057 Type_3 Test for correction for absorption needed
 You have indicated that an absorption correction has not been applied.
 (_exptl_absorpt_correction_type 'none'). However, the predicted values of
 Tmin & Tmax, based on the crystal dimensions given in the CIF, are 
 sufficiently unequal that absorption effects appear to be significant.
 Therefore, the application of a suitable absorption correction would
 appear to be required. Also check that the crystal dimensions given in the
 CIF do represent the actual crystal dimensions as closely as possible.
 Inaccuracies here can lead to a poor prediction of Tmin & Tmax and give
 rise to these alerts. It should normally be possible to estimate the 
 crystal dimensions to 2 decimal places. Rough estimates to only 1 decimal
 place may be too inaccurate to provide reliable estimates of Tmin & Tmax.
 

PLAT058 Type_1 Test for specification Tmax [0,1]
 The Maximum transmission factor should be specified in the case a correction 
 for absorption was done. This is NOT the value that is calculated 
 automatically with SHELXL when a SIZE instruction is given in the SHELXL
 instruction file. The values reported by SHELXL represent the EXPECTED
 correction range. Some correction packages (e.g. SADABS) will provide
 only one 'relative- correction-factor'. In such cases, Tmax should be 
 given as Tmax-expected and Tmin = relative-correction-factor * Tmax.

PLAT059 Type_1 Test for specification Tmin [0,1]
 The Minimum transmission factor should be specified in case a correction 
 for absorption was done. This is NOT the value that is calculated 
 automatically with SHELXL when a SIZE instruction is given in the SHELXL
 instruction file. The values reported by SHELXL represent the EXPECTED
 correction range. Some correction packages (e.g. SADABS) will provide
 only one 'relative-correction-factor'. In such cases, Tmax should be 
 given as Tmax-expected and Tmin = relative-correction-factor * Tmax.

PLAT060 Type_4 RR Test
 see IUCR WEB-Pages

PLAT061 Type_4 RR' Test
 see IUCR WEB-Pages

PLAT062 Type_4 Rescale Tmin & Tmax
 Some (empirical) correction packages (e.g. SADABS) will provide only one
 'relative-correction-factor'. In such cases, Tmax should be given as 
 Tmax-expected (as calculated from the crystal dimensions) and
 Tmin = relative-correction-factor * Tmax.
 

PLAT063 Type_4 Test for Crystal Size
 Alert for crystals with at least one dimension probably too large for the
 homogeneous part of the X-ray beam when used for datacollection using 
 crystal monochromated radiation. An exception will be datacollection using 
 a beta-filter and a sufficiently large collimator.
 See also: C.H.Gorbitz (1999), Acta Cryst. B55, 1090-1098.

PLAT064 Type_1 Test for T(max) .GE. T(min)
 Check that the values entered under _exptl_correction_T_min and
 _exptl_correction_T_max have not been reversed or if there is a 
 typographical error for one of these two items.

PLAT065 Type_3 Test for applicability of (semi-)empirical abs.corr. [0,1]
 For high mu * r values, numerical absorption correction procedures are
 recommended.  

PLAT066 Type_1 Test whether Predicted and Reported Transmission Ranges are Identical
 The predicted and reported transmission ranges are found to be identical
 which is not to be expected. CIF's generated with SHELXL97 report 
 transmission ranges based on the crystal dimensions supplied on the
 SIZE card. Those values have nothing to do with the actual corrections
 for absorption as applied to the data: they just report the EXPECTED range.
 Some correction packages (e.g. SADABS) will provide only one
 'relative-correction-factor'. In such cases, Tmax should be given as 
 Tmax-expected and Tmin = relative-correction-factor * Tmax.

PLAT067 Type_1 Insure that minimum dimension less max dimension
 Minimum an Maximum dimensions are likely exchanged in the CIF.

PLAT068 Type_1 Test for F000 Calc/Reported difference
 In the ideal case, both data items should have the same value. If not, 
 the reason should be clear. A reason might be the output by SHELXL-97
 of population parameters to the CIF with only two decimals. 
 Note: SHELXL counts the number of electrons in the unit cell. The result 
 will in general be an integer. This is also the number checked for here.
 The official definition calls for 'The effective number of electrons in the 
 crystal unit cell contributing to F(000)'. It may contain dispersion
 contributions and is calculated as:
 F(000) = [ (sum f~r~)^2^ + (sum f~i~)^2^ ]^1/2^
 f~r~ = real part of the scattering factors at theta = 0
 f~i~ = imaginary part of the scattering factors at theta = 0
 

PLAT070 Type_1 Test for duplicate labels
 The CIF contains duplicate labels posing interpretation problems for
 PLATON/CHECK. Derived geometry ALERTS may have their origin in this
 problem.

PLAT071 Type_1 Test for uninterpretable labels
 The CIF contains labels posing problems for PLATON/CHECK.
 Example: label HN1 with no scattering type information supplied.
 Validation is aborted.

PLAT072 Type_2 Test for extreme first weighting parameter (SHELXL)
 The first parameter on the SHELXL weighting line has an exceptionally
 large value. This may indicate either improper reflection s.u.'s or
 an unresolved problem such as missed twinning.

PLAT073 Type_1 Test for inconsistency 'constr' versus 'H-Atoms refined'
 The structure contains refined hydrogen atoms. However the data item 
 _refine_ls_hydrogen_treatment has the value 'constr'. 
 The value 'mixed' is more appropriate.

PLAT074 Type_1 Test for Occupancy equal 0.0
 The CIF contains an atom with occupacy less than 0.0001 

PLAT075 Type_1 Test for Occupancy greater than 1.0
 The CIF contains an atom with Occupancy greater than 1.0.

PLAT076 Type_1 Test for Occupancy less than 1.0 for atom on special position
 The CIF contains an atom sitting on a special position with occupancy
 specified as less than 1.0. This is often an error and the result of the
 confusion of the notions 'occupancy' and 'population parameter'. The first
 should be 1.0 for a fully occupied site. The latter multiplies the
 site-symmetry with the occupancy. Thus, for a fully occupied site on a
 mirror plane the site-symmetry will be 0.5 * 1.0 = 0.5.
 
 Note: a wrong occupancy number will lead to an incorrect expected
 chemical formula.

PLAT077 Type_4 Test for Non-Integral # of atoms in Unit Cell
 The unit-cell contains a non-integer number of atoms of a given atom type.
 Valid reasons include partially occupied (solvent) sites and substitutional
 disorder.

PLAT078 Type_1 Test for inconsistency 'geom' versus 'no H-Atoms'
 The structure contains no hydrogen atoms. However the data item 
 _atom_sites_solution_hydrogen had the value 'geom'. This value is
 likely the SHELXL default and should be replaced by '.'.
 

PLAT079 Type_1 Test for inconsistency 'mixed' versus 'no H-Atoms'
 The structure contains no hydrogen atoms. However the data item 
 _refine_ls_hydrogen_treatment has the value 'mixed'. This value is
 likely the SHELXL default and should be replaced by '.'.

PLAT080 Type_2 Test maximum shift/error
 Convergence of the refinement is proven with a close to zero shift/error
 value for all refined parameters. Such a convergence is easily achieved
 with a few additional refinement cycles at little cost.
 Note: Some SHELXL-97 versions do not allow for negative Flack parameter
 values. Convergence in such a case may be never reached because the 
 Flack parameter value is reset to zero. 

PLAT082 Type_2 Test for reasonable R1
 A higher than usual R1 indicates either an insufficient model or poor
 quality data.

PLAT083 Type_2 Test for extreme second weighting parameter (SHELXL)
 The second parameter on the SHELXL weighting line has an exceptionally
 large value. This may indicate either improper reflection s.u.'s or
 an unresolved problem such as missed twinning.

PLAT084 Type_2 Test for reasonable wR2
 wR2 will in general have a value twice of that of R1 with refinement on F**2.
 Significantly larger values usually indicate a poor refinement model.
 Also check for unaccounted for twinning.

PLAT085 Type_2 Test for default SHELXL weighting scheme
 The weighting scheme is found to be left at the SHELXL default.

PLAT086 Type_2 Test for reasonable S (Too Low)
 S should in general be close to 1 at the end of a refinement with a proper
 weighting scheme. If not, there might be significant unresolved problems
 with the model.

PLAT087 Type_2 Test for reasonable S (Too High)
 S should in general be close to 1 at the end of a refinement with a proper
 weighting scheme. If not, there might be significant unresolved problems
 with the model.

PLAT088 Type_3 Test for reasonable Data / parameter ratio (centro)
 The data/parameter ratio should in general be higher than 10 for a quality
 structure determination. This ratio can be improved by not refining 
 C-H parameters other than riding on their carrier atom.

PLAT089 Type_3 Test for reasonable Data / parameter ratio (non-centro) (Zmax < 18)
 The data/parameter ratio should in general be higher than 8 for a quality
 determination of a structure containing atoms with Z less than 18.
 This ratio can be improved by not refining C-H parameters other than riding
 on their carrier atom.
 Note that with light atom non-centrosymmetric structures where anomalous
 dispersion effects are insignificant, it is unwise to attempt to use
 unmerged Friedel-related reflections simply to boost the r/p ratio. 

PLAT090 Type_3 Test for reasonable Data / parameter ratio (non-centro) (ZMAX > 18)
 The data/parameter ratio should in general be higher than 10 for a quality
 determination for a structure containing heavy atoms with ZMAX greater
 than 17. This ratio can be improved by not refining C-H parameters other
 than riding on their carrier atom.

PLAT091 Type_1 Test for 'No-wavelength given'
 No Wavelength specification found in the CIF.

PLAT092 Type_4 Test for wavelength type [Cu,Mo,Ag]
 Warning: specified wavelength is not Cu,Mo or Ag Ka radiation.
 Valid exceptions are Neutron and Synchrotron data.

PLAT093 Type_1 Test for inconsistency 'mixed' versus 'no refined H'
 The 'mixed' type Hydrogen atom refinement is reported (SHELXL-97 default).
 However, no Hydrogen atoms with freely refined positions are found in the
 CIF. Likely, the value 'constr' for '_refine_ls_hydrogen_treatment' will
 be more appropriate (e.g. when all Hydrogen atoms have been refined in
 the riding mode on their carrier atom). 
  

PLAT094 Type_2 Test for maximum/minimum residual density ratio
 The ratio of the maximum and minimum residual density excursions is 
 unusual. This might indicate unaccounted for twinning or missing atoms
 (e.g. associated with disordered solvent).
 

PLAT095 Type_1 Test for residual density maximum given [0,1]
 No residual electron density maximum given in CIF.

PLAT096 Type_1 Test for residual density minimum given [0,1]
 No residual electron density minimum given in CIF.

PLAT097 Type_2 Test maximum residual density
 Residual density maximum larger than expected. This might be caused by
 residual absorption artefacts, unaccounted for twinning, wrongly assigned 
 atom types and other model errors. 
 

PLAT098 Type_2 Test for minimum residual density
 Residual density minimum larger than expected. This might be caused by
 residual absorption artefacts, wrongly assigned atom types and other
 model errors.
 

PLAT099 Type_1 Test for minimum residual density less zero [0, 1]
 Likely interchanged maximum and minimum values.
 Alternatively, the minimum residual density has the (unlikely) value 
 zero.
  

PLAT110 Type_2 Test for additional translational symmetry [0, 1]
 Tests for missed symmetry are done with ADDSYM, an expanded MISSYM (C) clone.
 These tests warn for missed or possible higher (pseudo) symmetry in the 
 structural model (i.e. based on the coordinate data).
 Close examination of the situation at hand is indicated in order to
 prove/disprove the issue (usually in combination with the reflection data).
 Report on potential (pseudo/real) lattice centering or cell halving.
 Note: H-atoms and disordered atoms are not taken into account in the tests.

PLAT111 Type_2 Test for additional centre of symmetry [0, 100]
 Tests for missed symmetry are done with ADDSYM, an expanded MISSYM (C) clone.
 These tests warn for missed or possible higher (pseudo) symmetry in the 
 structural model (i.e. based on the coordinate data). Close examination of
 the situation at hand is indicated in order to prove/disprove the issue
 (usually in combination with the reflection data).
 This ALERT reports on a potential additional (pseudo/real) inversion centre.
 A pseudo-centre may be incompatible with existing symmetry elements.
 Chiral molecules are incompatible with an inversion centre.
 Note: H-atoms and disordered atoms are not taken into account in the test.

PLAT112 Type_2 Test for additional symmetry [0, 1]
 Tests for missed symmetry are done with ADDSYM, an expanded MISSYM (C) clone.
 These tests warn for missed or possible higher (pseudo) symmetry in the 
 structural model (i.e. based on the coordinate data). Close examination of
 the situation at hand is indicated in order to prove/disprove the issue 
 (usually in combination with the reflection data).
 This ALERT reports on potential additional (pseudo/real) rotation axes and
 mirrors.
 
 In addition, (pseudo/real) lattice centering/translations are reported as 
 A, B, C, I, X, Y, Z, S. (Here S stands for special and not covered by the
 preceding types). Full details on the situation at hand should be gleaned
 from an actual PLATON/ADDSYM run.
 Chiral molecules are incompatible with an inversion centre or (glide)planes.
 Note: H-atoms and disordered atoms are not taken into account in the tests.

PLAT113 Type_2 Report New space-group suggested by ADDSYM
 Tests for missed symmetry are done with ADDSYM, an extended MISSYM (C) clone.
 These tests warn for missed or possible higher (pseudo) symmetry in the 
 structural model (i.e. based on the coordinate data). Close examination of
 the situation at hand is indicated in order to prove/disprove the issue
 (usually in combination with the reflection data). 
 Chiral molecules are incompatible with an inversion centre or (glide)planes.
 For an example of reported pseudo-symmetry see I.A.Guzei et al, (2002).
 Acta Cryst. C58, m141-m143.
 Note: H-atoms and disordered atoms (i.e. atoms with population less
 than 1.0) are not taken into account in the tests. This may artificially
 lead to a symmetry higher than the actual one.
 Note: Atoms are treated as having the same atom type in order to catch
 certain types of disorder or incorrect atom type assignment.
      

PLAT114 Type_2 Report on ADDSYM problem
 ADDSYM has problems to reconstruct a space-group from the symmetry 
 operation found in the symmetry expanded coordinate set. The reason being
 either intricate additionally detected pseudo-symmetry or serious errors
 in the data set.

PLAT115 Type_5 Test for non-crystallographic centre of symmetry [0, 100]
 Tests for missed symmetry are done with ADDSYM, an expanded MISSYM (C) clone.
 This ALERT reports on local inversion symmetry, not compatible with the
 reported space-group symmetry.
 Note: H-atoms and disordered atoms are not taken into account in the test.

PLAT119 Type_1 Report Problem with symmetry operator syntax
 A symmetry operation should be specified in the CIF either without
 spaces or between quotes.

PLAT120 Type_1 Test for consistent _symmetry_space_group_name_H-M and Symm Opp
 Space-group symmetry should be provided in the CIF both explicitly with a
 _symmetry_equiv_pos_as_xyz loop and implicitly with
 _symmetry_space_group_name-H-M.
 An unusual (non-standard) choice of origin may also raise this ALERT.
 Please check and Explain.

PLAT121 Type_1 Test for valid _symmetry_space_group_name_H-M
 Symmetry in the CIF should be provided both explicitly with a
 _symmetry_equiv_pos_as_xyz loop and implicitly with
 _symmetry_space_group_name_H-M.
 Test for valid _symmetry_space_group_name_H-M symbol.

PLAT122 Type_1 Test for ? _symmetry_space_group_name_H-M
 Symmetry in the CIF should be provided both explicitly with a
 _symmetry_equiv_pos_as_xyz loop and implicitly with
 _symmetry_space_group_name_H-M.
 Test for missing (i.e. ?) _symmetry_space_group_name_H-M symbol.

PLAT123 Type_1 Test for Interpretable Space-Group Symmetry
 Symmetry in the CIF should be provided in the CIF both explicitly with a
 _symmetry_equiv_pos_as_xyz loop and implicitly with
 _symmetry_space_group_name_H-M.
 Test for uninterpretable or inconsistent Space-group information.

PLAT124 Type_1 Test for _symmetry_equiv_pos_as_xyz present
 Symmetry in the CIF should be provided in the CIF both explicitly with a
 _symmetry_equiv_pos_as_xyz loop and implicitly with
 _symmetry_space_group_name_H-M.
 Test for uninterpretable or absent explicit symmetry records

PLAT125 Type_4 Test for ? _symmetry_space_group_name_Hall
 Optionally specify the Hall symbol. The Hall symbol provides an
 unambiguous definition of the space-group symmetry where the Hermann-
 Mauguin symbol leaves room for alternative choices of the origin.
 E.g. for space-group P21, the screw axis is in general taken to coincide 
 with the b-axis. However, sometimes it is chosen to be shifted by 1/4
 in the c-axis direction to bring out the relation with P21/c. The Hall
 symbols will be 'P 2yb' and 'P 2ybc' respectively. 
 Refer to: S.R.Hall, Space-Group Notation with an Explicit Origin;
           Acta Cryst. (1981), A37, 517-525.
       or: http://www.kristall.ethz.ch/LFK/software/sginfo/hall_symbols.html

PLAT126 Type_1 Test for _symmetry_space_group_name_Hall error
 The reported Hall-symbol is found to be in error or uninterpretable. 
 Refer to: S.R.Hall, Space-Group Notation with an Explicit Origin;
           Acta Cryst. (1981), A37, 517-525.
       or: http://www.kristall.ethz.ch/LFK/software/sginfo/hall_symbols.html

PLAT127 Type_1 Test for _symmetry_space_group_name_Hall consistency
 The reported Hall-symbol is found to be inconsistent with the one derived
 from the explicit symmetry operations.
 Alternatively, no Hall-symbol could be derived by PLATON for the explicit 
 set of symmetry operations. This may be the case when an unusual origin
 is chosen.
 Refer to: S.R.Hall, Space-Group Notation with an Explicit Origin;
           Acta Cryst. (1981), A37, 517-525.
       or: http://www.kristall.ethz.ch/LFK/software/sginfo/hall_symbols.html

PLAT128 Type_4 Test for non-standard monoclinic space-group setting
 The reported monoclinic space-group is in a non-standard setting. 
 Transformation to the conventional setting is indicated unless there
 is a good (scientific) reason not to do so.

PLAT129 Type_4 Test for unusual non-standard Space-group name
 The reported space-group name is unusual. 

PLAT130 Type_1 Test for Cubic: a = b = c
 Symmetry constraints on cell dimensions are checked.

PLAT131 Type_1 Test for Cubic: alpha = beta = gamma = 90
 Symmetry constraints on cell dimensions are checked.

PLAT132 Type_1 Test for Trigonal/Hexagonal : a = b
 Symmetry constraints on cell dimensions are checked.

PLAT133 Type_1 Test for Trigonal/Hexagonal : alpha = beta = 90
 Symmetry constraints on cell dimensions are checked.

PLAT134 Type_1 Test for Trigonal/Hexagonal : gamma = 120
 Symmetry constraints on cell dimensions are checked.

PLAT135 Type_1 Test for Tetragonal: a = b
 Symmetry constraints on cell dimensions are checked.

PLAT136 Type_1 Test for Tetragonal: alpha = beta = gamma = 90
 Symmetry constraints on cell dimensions are checked.

PLAT137 Type_1 Test for Orthorhombic: alpha = beta = gamma = 90
 Symmetry constraints on cell dimensions are checked.

PLAT138 Type_1 Test for Monoclinic more than 1 angle off 90 degrees
 Symmetry constraints on cell dimensions are checked.

PLAT139 Type_1 Test for Rhombohedral a = b = c
 Symmetry constraints on cell dimensions are checked.

PLAT140 Type_1 Test for Rhombohedral alpha = beta = gamma
 Symmetry constraints on cell dimensions are checked.

PLAT141 Type_4 su on a - axis small or missing
 The su on the a-axis is small or missing. The presence of su's (where
 required) and value are checked. Su's as given by the diffractometer
 software are often much smaller then realistic.

PLAT142 Type_4 su on b - axis small or missing
 The su on the b-axis is small or missing. The presence of su's (where
 required) and value are checked. Su's as given by the diffractometer 
 software are often much smaller then realistic.

PLAT143 Type_4 su on c - axis small or missing
 The su on the c-axis is small or missing. The presence of su's (where
 required) and value are checked. Su's as given by the diffractometer
 software are often much smaller then realistic.

PLAT144 Type_4 su on alpha small or missing
 The su on alpha is small or missing. The presence of su's (where required)
 and value are checked. Su's as given by the diffractometer software are
 often much smaller then realistic.

PLAT145 Type_4 su on beta small or missing
 The su on beta is small or missing. The presence of su's (where required)
 and value are checked. Su's as given by the diffractometer software are
 often much smaller then realistic.

PLAT146 Type_4 su on gamma small or missing
 The su on gamma is small or missing. The presence of su's (where required)
 and value are checked. Su's as given by the diffractometer software are
 often much smaller then realistic.

PLAT147 Type_1 su on symmetry restricted cell angle
 There should be no s.u. on symmetry constrained cell angles.
 Example: No su on alpha, beta and gamma for orthorhombic symmetry.

PLAT148 Type_3 su on a,b, or c - axis too large
 The su on the reported -axis is unexpectedly large. 

PLAT149 Type_3 su on alpha, beta or gamma too large
 The su on the reported angle is too large.

PLAT150 Type_1 Check Volume
 An ALERT is issued when the reported unit cell volume differs significantly 
 from the volume calculated on the basis of the supplied cell dimensions.

PLAT151 Type_1 Check for s.u. on Volume
 Missing s.u. on cell volume.

PLAT152 Type_1 Check for consistency of su on Volume and cell parameters
 Some software packages calculate Volume su's incorrectly. The correct formula
 for triclinic, monoclinic and orthorhombic systems
 (based on the propagation of error expression) may be found in:
 M. Nardelli, Computer & Chemistry, (1983), 7, 95-98.
 or
 C. Giacovazzo ed. in 'Fundamentals of Crystallography', Second Edition,
 Oxford University Press, 2003,  p135.

PLAT153 Type_1 test for equal axial su's
 The reported cell axes su's are reported equal. Please check whether
 this is correct or a software default value.

PLAT154 Type_1 test for equal cell angle su's
 The reported cell angle su's are reported equal. Please check whether
 this is correct or a software default value.

PLAT155 Type_4 Check for reduced cell aP
 Unless for special reasons related to the structure/content, a unit-cell and
 structure is best reported with reference to the Niggli Reduced Cell.
 This ALERT may originate also from a failure to order the axes from
 small to large.

PLAT156 Type_4 Check for non-standard axial order
 The axial order should be from small to large in the triclinic cell.

PLAT157 Type_4 Check for non-standard monoclinic beta angle less 90 Degrees.
 By convention, the Monoclinic beta angle is always chosen to be larger than
 90.0 Degrees. A trivial transformation (1 0 0/0 -1 0/0 0 -1) should be
 applied to the data.

PLAT158 Type_4 Check for standard reduced cell
 Unless for special reasons related to the structure/content, a unit-cell and
 structure is best reported with reference to the Niggli Reduced Cell.

PLAT161 Type_4 Missing x-coordinate su
 Missing or Zero su (esd) on x-coordinate.
 Positional parameters for all non-hydrogen atoms in general positions are
 checked for the presence of a non-zero s.u. on them. This includes
 parameters fixed to fix the origin in polar space-groups which is no longer
 necessary when refinement is done with modern programs (e.g. SHELXL, XTAL).

PLAT162 Type_4 Missing y-coordinate su
 Missing or Zero su (esd) on y-coordinate.
 Positional parameters for all non-hydrogen atoms in general positions are
 checked for the presence of a non-zero s.u. on them. This includes
 parameters fixed to fix the origin in polar space-groups (e.g. P21) which 
 is no longer necessary when refinement is done with modern programs
 (e.g. SHELXL, XTAL).

PLAT163 Type_4 Missing z-coordinate su
 Missing or Zero su (esd) on z-coordinate.
 Positional parameters for all non-hydrogen atoms in general positions are
 checked for the presence of a non-zero s.u. on them. This includes
 parameters fixed to fix the origin in polar space-groups (e.g. P41) which 
 is no longer necessary when refinement is done with modern programs 
 (e.g. SHELXL, XTAL).

PLAT164 Type_4 Check for Refined C-H H-Atoms
 Warning: Refined C-H H-atoms in heavy-atom structure (i.e. containing an
 element beyond element #18). Such H-atoms are in general better refined
 at calculated positions riding on the atoms they are attached to. A
 better data over parameter ratio will be achieved.

PLAT165 Type_3 Check for R-flagged Non-H Atoms
 Report on restrained (riding) Non-Hydrogen atoms. Note: This may lead to
 non meaningfull bond and angle su's (ALERTS _751, _752). 
 R-flagged atoms may arise unintentional being caused by an "AFIX 0" line 
 being missing in a shelxl.ins file (SHELXL-97 refinement).
 Alternatively, the number of refined parameters may have been limited
 deliberately (e.g. by refinement of C-F with fixed known geometry,
 similar to C-H) in order to keep the data/parameter ratio acceptable. 

PLAT166 Type_4 Check for calc flagged atoms with s.u.s on coordinates
 Calc-flagged atoms are not supposed to carry s.u.'s on their coordinates.
   

PLAT170 Type_4 Check for sufficient data in in Atom data loop
 Insufficient data encountered in coordinate loop.
 A possible cause might be the use of a SHELX style '=' continuation line. 
   

PLAT180 Type_4 Check Rounding of Cell Axes and Angles
 It is unusual that more cell parameters end with a zero and the su is 10.
 This problem might be caused by the specification of insufficient 
 'meaningful' digits as compared to the reported su. 
 
 see also: W.Clegg, Acta Cryst. (2003) E59, e2-e5.

PLAT193 Type_1 Test for consistency of cell & diffraction temperatures
 The reported _cell_measurement_temperature deviates from the reported 
 _diffrn_ambient_temperature values.

PLAT194 Type_1 Check for missing _cell_measurement_reflns_used datum
 Please supply value for _cell_measurement_reflns_used
  

PLAT195 Type_1 Check for missing _cell_measurement_theta_max datum
 Please supply value for _cell_measurement_theta_max  

PLAT196 Type_1 Check for missing _cell_measurement_theta_min datum
 Please supply value for _cell_measurement_theta_min
  

PLAT197 Type_1 Test for specification of unitcell measurement temperature
 Please specify the temperature (Kelvin) at which the unit-cell was
 determined.

PLAT198 Type_1 Test for specification of Datacollection temperature
 Please specify the temperature (Kelvin) at which the intensity data were 
 collected.

PLAT199 Type_1 Test for SHELXL Roomtemperature Default (Cell)
 The cell determination temperature is set in the CIF by default by SHELXL to
 293 K if the TEMP instruction is not used. The actual temperature is likely 
 either slightly or significantly (for a low temperature data collection)
 different.

PLAT200 Type_1 Test for SHELXL Roomtemperature Default (Datacollection)
 The data collection temperature is set in the CIF by default by SHELXL to
 293 K if the TEMP instruction is not used. The actual temperature is likely
 either slightly or significantly (for a low temperature data collection) 
 different.

PLAT201 Type_2 Test for isotropic non-H atoms in main residue(s)
 This test reports on non-hydrogen atoms that were refined with isotropic
 displacement parameters only in the main residue. Such a practice is unusual
 by modern standards and only needed for minor disorder modelling.

PLAT202 Type_3 Test for isotropic non-H atoms in anion ? or solvent ?
 This test reports on isotropically refined atoms in small moieties (usually
 anions or solvent).

PLAT210 Type_3 Test for 'all-isotropic adp(s)
 No anisotropically refined atoms in CIF ?

PLAT211 Type_2 Test for NPD ADP's (1.0) in main residue(s)
 This test reports on non-positive definite (i.e. complex and unrealistic) 
 anisotropic displacement parameters in the main residue.

PLAT212 Type_2 Test for NPD ADP's in anion? & solvent ? [0, 1]
 This test reports on non-positive definite (i.e. complex and
 unrealistic) anisotropic displacement parameters in an anion or solvent
 residue.

PLAT213 Type_2 Test ratio adp max/min in main residue(s)
 The maximum and minimum main axis ADP ratio (Angstrom Units) is tested for 
 the main residue. Large values may indicate unresolved disorder.

PLAT214 Type_2 Test ratio adp max/min in anion ? or solvent ?
 The maximum and minimum main axis ADP ratio (Angstrom Units) is tested for
 the minor residue(s). Large values may indicate unresolved disorder.

PLAT215 Type_3 Test for unusual disordered atom ADP in main residue
 The maximum and minimum main axis ADP ratio (Angstrom Units) is tested for
 the main residue. Large values may indicate unresolved disorder.

PLAT216 Type_3 Test for unusual disordered atom ADP in minor residue
 The maximum and minimum main axis ADP ratio (Angstrom Units) is tested for
 the minor residue(s). Large values may indicate unresolved disorder.

PLAT217 Type_1 Test for Incomplete UIJ data
 Check & Correct U(aniso) data for completeness etc.
 Do not use SHELX style '=' continuation line
 

PLAT220 Type_2 Test Ueq(max)/Ueq(Min) range for non-H atoms in non-solvent
 This test reports on a larger than usual U(eq) range for the specified
 element type in the non-solvent/anion part of the structure.
 Too high or too low Ueq's may be an indication for incorrectly
 identified atomic species (i.e. O versus N).

PLAT221 Type_2 Test Ueq(max)/Ueq(Min) range for non-H atoms in solvent
 This test reports on a larger than usual U(eq) range for the non-hydrogen
 atoms solvent/anion. Too high or too low Ueq's may be an indication for
 incorrectly identified atomic species (i.e. Br versus Ag).

PLAT222 Type_3 Test Uiso(max)/Uiso(Min) range for H atoms in non-solvent
 
 This test reports on a larger than usual range of U(eq) values for hydrogen
 atoms in the non-solvent/anion part of the structure.
 Possible causes are:
 1 - disorder, e.g. in t-butyl moieties.
 2 - poor data, not adequate for the refinement of individual displacement
     parameters.
 3 - Misplaced hydrogen atoms (i.e. there is no density at the position
     where one of the H-atoms is positioned).

PLAT223 Type_4 Test Ueq(max)/Ueq(Min) range for H atoms in solvent
 This test reports on large ranges in displacement parameters for hydrogen
 atoms in the solvent/anion part of the structure.

PLAT224 Type_1 Test for difference in implicit and explicit U(eq)
 This test reports on a large difference between Ueq in the CIF and the Ueq
 calculated from the 6 reported Uij values.                            .

PLAT230 Type_2 Hirshfeld Rigid-Bond Test (Acta Cryst (1976),A32,239-244
 The components of the anisotropic displacement parameters along chemical
 bonds are assumed to be equal in magnitude. Large differences supposedly
 indicate contamination of these parameters with other (unresolved) effects 
 such as (substitutional) disorder, model or data errors and/or 
 over refinement. Atomic sites assigned the wrong scattering type (e.g. Ag 
 versus Br) should generate 'problem signals' with this test. Data sets
 corrected for absorption effects with DELREF techniques (e.g. DIFABS, SHELXA,
 XABS2) often show large DELU values for bonds involving the heaviest atom.
 Note: The original 'Hirshfeld-test' was defined in absolute terms (see
 F.L.Hirshfeld, Acta Cryst. (1976). A32, 239-244). The current test is with
 reference to the associated standard uncertainty.

PLAT231 Type_4 Hirshfeld Rigid-Bond Test (Acta Cryst (1976),A32,239-244
 The components of the anisotropic displacement parameters along chemical
 bonds are assumed to be equal in magnitude. Large differences supposedly
 indicate contamination of these parameters with other (unresolved) effects 
 such as (substitutional) disorder, model or data errors and/or 
 over refinement. Atomic sites assigned the wrong scattering type (e.g. Ag 
 versus Br) should generate 'problem signals' with this test. Data sets
 corrected for absorption effects with DELREF techniques (e.g. DIFABS, SHELXA,
 XABS2) often show large DELU values for bonds involving the heaviest atom.
 Note: The original 'Hirshfeld-test' was defined in absolute terms (see
 F.L.Hirshfeld, Acta Cryst. (1976). A32, 239-244). The current test is with
 reference to the associated standard uncertainty.

PLAT232 Type_2 Hirshfeld Rigid-Bond Test (Metal-X) (Acta Cryst (1976),A32,239-244
 The components of the anisotropic displacement parameters along chemical
 bonds are assumed to be equal in magnitude. Large differences supposedly
 indicate contamination of these parameters with other (unresolved) effects 
 such as (substitutional) disorder, model or data errors and/or 
 over-refinement. Atomic sites assigned the wrong scattering type (e.g. Ag 
 versus Br) should generate 'problem signals' with this test. Data sets
 corrected for absorption effects with DELREF techniques (e.g. DIFABS, SHELXA,
 XABS2) often show large DELU values for bonds involving the heaviest atom.
 A special case are M-C=O type of systems that generally show significant
 differences for the M-C bond. See D.Braga & T.F. Koetzle (1988), Acta Cryst.
 B44, 151-155). 
 
 Note: The original 'Hirshfeld-test' was defined in absolute terms (see
 F.L.Hirshfeld, Acta Cryst. (1976). A32, 239-244). The current test is with
 reference to the associated standard uncertainty.

PLAT233 Type_4 Hirshfeld Rigid-Bond Test (Metal-X) (Acta Cryst (1976),A32,239-244
 The components of the anisotropic displacement parameters along chemical
 bonds are assumed to be equal in magnitude. Large differences supposedly
 indicate contamination of these parameters with other (unresolved) effects 
 such as (substitutional) disorder, model or data errors and/or 
 over refinement. Atomic sites assigned the wrong scattering type (e.g. Ag 
 versus Br) should generate 'problem signals' with this test. Data sets
 corrected for absorption effects with DELREF techniques (e.g. DIFABS, SHELXA,
 XABS2) often show large DELU values for bonds involving the heaviest atom.
 Note: The original 'Hirshfeld-test' was defined in absolute terms (see
 F.L.Hirshfeld, Acta Cryst. (1976). A32, 239-244). The current test is with
 reference to the associated standard uncertainty.

PLAT234 Type_4 Hirshfeld Rigid-Bond Test (Acta Cryst (1976),A32,239-244
 The components of the anisotropic displacement parameters along chemical
 bonds are assumed to be equal in magnitude. Large differences supposedly
 indicate contamination of these parameters with other (unresolved) effects 
 such as (substitutional) disorder, model or data errors and/or 
 over refinement. Atomic sites assigned the wrong scattering type (e.g. Ag 
 versus Br) should generate 'problem signals' with this test. Data sets
 corrected for absorption effects with DELREF techniques (e.g. DIFABS, SHELXA,
 XABS2) often show large DELU values for bonds involving the heaviest atom.
 Note: The original 'Hirshfeld-test' was defined in absolute terms (see
 F.L. Hirshfeld, Acta Cryst. (1976). A32, 239-244).

PLAT241 Type_2 Test for unusually high U(eq) as compared with bonded neighbors
 The U(eq) value of an atom is compared with the average U(eq) for to
 non-hydrogen atoms bonded to it. Large differences may indicate that the
 wrong atom type was assigned (e.g. N instead of O).

PLAT242 Type_2 Test for Unusually Low U(eq) as compared with bonded neighbors
 The U(eq) value of an atom is compared with the average U(eq) for 
 non-hydrogen atoms bonded to it. Large differences may indicate that the
 wrong atom type was assigned (e.g. N instead of O). False alarms may
 occur for terminal groups such as the t-butyl moiety.

PLAT243 Type_4 Test for unusually high solvent U(eq) as compared with bonded neighbors
 The U(eq) value of an atom in the solvent or ion is compared with the average 
 U(eq) for non-hydrogen atoms bonded to it. Large differences may indicate 
 that the wrong atom type was assigned (e.g. N instead of O).

PLAT244 Type_4 Test for unusually low solvent U(eq) as compared with bonded neighbors
 The U(eq) value of an atom in the solvent or ion is compared with the average
 U(eq) for non-hydrogen atoms bonded to it. Large differences may indicate 
 that the wrong atom type was assigned (e.g. N instead of O). False alarms may
 occur for terminal groups such as the t-butyl moiety.

PLAT245 Type_2 Test for unusually low H-U(eq) as compared with bonded atom
 U(iso) of a hydrogen atom is generally expected to be greater than the
 U(eq) of the non-hydrogen atom it is attached to.
           

PLAT250 Type_2 Test for unusual anisotropic average UIJ
 An average value of the U(i,j) tensor of the asymmetric unit of a residue
 is calculated. An ALERT is generated when the corresponding U3/U1 ratio
 deviates significantly from 1.0. Large values of this ratio should be taken
 as an indication of possible systematic errors in the data or errors in 
 the model. Visual inspection of an ORTEP plot will show that many 
 displacement ellipsoids have their major axis pointing in the same direction.

PLAT301 Type_3 Test for main residue(s) disorder %
 Atom sites that are not fully occupied are counted. A large fraction of
 disordered atoms may be both a signal for serious structure analysis 
 problems or less reliable/interesting results. 

PLAT302 Type_4 Test for (anion/solvent) disorder %
 Atom sites that are not fully occupied are counted. A large fraction of

PLAT303 Type_2 Test for more than 1 connection to Hydrogen Atoms
 Hydrogen atoms are generally connected to only one other atom. A hydrogen
 atom between two oxygen atoms is a special case. Investigate whether this
 hydrogen atom is better described with a disorder model with two partially
 occupied sites. A difference map might show a double-well density.        

PLAT305 Type_2 Test for isolated Hydrogen Atoms
 This test reports on hydrogen atoms that are not on bonding distance to any
 atom. This ALERT may indicate that the hydrogen atom refined to a
 non-bonding position or needs a symmetry operation to bring it to bonding
 distance. It also may indicate a problem with incompatible population 
 parameters (e.g. C - H with population 0.8 and 0.9 respectively).

PLAT306 Type_2 Test for isolated Oxygen Atoms
 This test reports on oxygen atoms that are not within bonding distance to
 any other atom in the structure. A common reason may be that no hydrogen
 atoms are given for a water molecule. Attempts should be made to locate
 those hydrogen atoms from a difference map.

PLAT307 Type_2 Test for isolated Metal Atoms
 This test reports on metal atoms that are not bonded or at coordination
 distance of other atoms. Isolated ions are very unusual (or non-existent ?)

PLAT308 Type_2 Test for single bonded Metal Atoms
 This test reports on single bonded (coordinated) metal atoms/ions. This
 represents a very unusual situation. There are literature examples where such
 a 'single bonded metal' was shown to be a halogen.

PLAT309 Type_2 Test for single bonded Oxygen Atoms
 Single bonded Oxygen with C-O .GT. 1.3 Angstrom. Missing H-Atom ?

PLAT310 Type_2 Test for 'too close' (symmetry related) full weight atoms
 This test identifies (very) short contacts between atoms that only becomes
 apparent after the application of symmetry on the primary coordinate set.

PLAT311 Type_2 Test for isolated Disordered Oxygen Atoms
 This test reports on oxygen atoms (not full weight) that are not within 
 bonding distance to any other atom in the structure.
 A common reason may be that no hydrogen atoms are given for a water molecule.

PLAT312 Type_2 Test for C=O-H
 Strange C-O-H geometry with C-O .LT. 1.25 Angstrom detected. Misplaced
 H-Atom ?

PLAT313 Type_2 Test for O with three covalent bonds
 Oxygen atom with three covalent bonds detected. Check for correct atom type
 assignment (e.g. N rather than O)
 Note: Exceptions are H3O+ (Oximium or Hydroxonium) and
                      H5O2+ (Hydronium or aqua-hydroxonium) species.

PLAT314 Type_2 Test for Metal-O-H angle of H2O
 A water molecule coordinated to a metal is detected with an unusually
 small value of the Metal-Oxygen-Hydrogen Angle. 

PLAT316 Type_2 Check for too many H's on C in C=N bond in main residue(s)
 An sp3 hybridized C was detected as part of a C=N moiety. Only one 
 attached H atom in sp2 configuration is expected and not two.
 In SHELXL terms this corresponds with an erroneous AFIX 23 rather than an
 AFIX 43 type of H atom position generation and refinement.

PLAT317 Type_2 Check for too many H's on C in C=N bond in Solvent/Ion (s)
 An sp3 hybridized C was detected as part of a C=N moiety. Only one 
 attached H atom in sp2 configuration is expected and not two.
 In SHELXL terms this corresponds with an erroneous AFIX 23 rather than an
 AFIX 43 type of H atom position generation and refinement.

PLAT318 Type_2 Hybridisation Problem on N in main residue(s)
 The test attempts to assign one of three hybridisations to N atoms in main
 residue: sp, sp2 or sp3 on the basis of the angles around N. 
 This ALERT may indicate a mis-assigned H atom position (e.g. an atom
 placed in a sp2 position instead of sp3).

PLAT319 Type_2 Hybridisation Problem on N in solvent/ion
 The test attempts to assign one of three hybridisations to N atoms in main
 residue: sp, sp2 or sp3 on the basis of the angles around N. 
 This ALERT may indicate a mis-assigned H atom position (e.g. an atom
 placed in a sp2 position instead of sp3).

PLAT320 Type_2 Hybridisation Problem on C in main residue(s)
 The test attempts to assign one of three hybridisations to C atoms in main
 residue: sp, sp2 or sp3 on the basis of the angles around C. In this way,
 missing H atoms or too many H-atoms on a carbon atom should be detected.

PLAT321 Type_2 Hybridisation Problem on C in solvent/ion
 The test attempts to assign one of three hybridisations to C atoms in 
 solven/anion: sp, sp2 or sp3 on the basis of the angles around C. In this 
 way missing H atoms or too many H-atoms on a carbon atom should be detected.

PLAT322 Type_2 Hybridisation Problem on non-C in main residue(s)
 The test attempts to assign one of three hybridisations to a non-C atom in
 the main residue: sp, sp2 or sp3 on the basis of the angles around the
 non-C atom. In this way, missing H atoms or too many H-atoms should be 
 detected.

PLAT323 Type_2 Hybridisation Problem on non-C in Solvent/Ion
 The test attempts to assign one of three hybridisations to a non-C atom in
 the solvent/anion: sp, sp2 or sp3 on the basis of the angles around the
 non-C atom. In this way, missing H atoms or too many H-atoms should be 
 detected.

PLAT324 Type_2 Check for possibly missing H on coordinating X-N-X in main residue
 Check for possibly missing Hydrogen atom on Nitrogen coordinating to
 a metal in the main residue.

PLAT325 Type_2 Check for possibly missing H on coordinating X-N-X in solvent/anion
 Check for possibly missing Hydrogen atom on Nitrogen coordinating to
 a metal in the solvent/anion.

PLAT326 Type_2 Check for possibly missing H on potentially sp3 Carbon
 Check for possibly missing Hydrogen atom on Carbon with sp3-like geometry 
 in the main residue.

PLAT327 Type_2
 Check for possibly missing Hydrogen atom on Carbon with sp3-like geometry
 in the solvent/anion.

PLAT328 Type_4 Check for a possibly missing H on potentially sp3 Phosphorus
 Check for a possibly missing Hydrogen atom on Phosphorus with sp3-like 
 geometry. 

PLAT330 Type_2 Check Average Phenyl C-C
 The standard average C-C bond distance in a phenyl ring is 1.395 Angstrom.
 The actual average ring distance may be larger than expected due to 
 systematic errors in the cell dimensions (e.g. use of incorrect
 wavelength value for the determination of the cell parameters).
 

PLAT331 Type_2 Check Average Phenyl C-C
 The standard average C-C bond distance in a phenyl ring is 1.395 Angstrom.
 The average ring distance may be smaller due to large thermal motion or
 incorrect cell dimensions.

PLAT332 Type_2 Check Phenyl C-C Range
 The standard average C-C in a phenyl ring is 1.395 Angstrom. Bond distances 
 in the ring are expected to vary only slightly due to thermal motion or
 substituent effects. Large deviations are likely due to data or model errors.
  

PLAT333 Type_2 Check Average in Multiple Substituted Benzene Type C-C
 The standard average C-C bond distance in a flat six carbon atom containing
 aromatic ring is 1.395 Angstrom.
 The actual average ring distance may be larger than expected due to 
 substituents such as =O, single bonds or systematic errors in the cell
 dimensions (E.g. when the wrong wavelength is used in the derivation of
 the cell parameters).
 

PLAT334 Type_2 Check Average in Multiple Substituted Benzene Type C-C
 The standard average C-C bond distance in a benzene ring is 1.395 Angstrom.
 The average ring distance may be smaller due to large thermal motion,
 substituents such as =O or incorrect cell dimensions.

PLAT335 Type_2 Check Multiple Substituted Benzene Type C-C Range
 The standard average C-C bond distance in a benzene ring is 1.395 Angstrom.
 Bond distances in the ring are expected to vary only slightly when due to 
 substituent effects (exceptions include =O substituents). Large deviations
 may indicate data or model errors.
  

PLAT338 Type_4 Check Average Torsion Angle in cyclo-hexane ring
 Cyclohexane moieties should have be significantly puckered as measured by
 the average torsion angle tau. Unresolved disorder generally results in 
 flattened rings and elongated displacement ellipsoids. A disorder model
 should be included in the calculations.
  

PLAT340 Type_3 Check Bond Precision for C-C in Light Atom Structures (Z(max) < 20)
 The average su for X-Y bonds is tested (named bond-precision). X-Y will 
 generally be C-C bonds, unless there are none. In the last case the su's of
 the lowest element numbers are considered (excluding hydrogen). There are
 three test ranges: one for structures with the largest element Z < 20, one 
 for the largest Z in the range 20 to 39 and one for structures with Z(max) 40
 or higher (_340, _341 and _342 respectively)

PLAT341 Type_3 Check Bond Precision for C-C in Structures (19 < Z(max) < 40)
 The average su for X-Y bonds is tested (named bond-precision). X-Y will 
 generally be C-C bonds, unless there are none. In the last case the su's of
 the lowest element numbers are considered (excluding hydrogen). There are
 three test ranges: one for structures with the largest element Z < 20, one 
 for the largest Z in the range 20 to 39 and one for structures with Z(max) 40
 or higher (_340, _341 and _342 respectively)

PLAT342 Type_3 Check Bond Precision for C-C in Structures (Z(max) > 39)
 The average su for X-Y bonds is tested (named bond-precision). X-Y will 
 generally be C-C bonds, unless there are none. In the last case the su's of
 the lowest element numbers are considered (excluding hydrogen). There are
 three test ranges: one for structures with the largest element Z < 20, one 
 for the largest Z in the range 20 to 39 and one for structures with Z(max) 40
 or higher (_340, _341 and _342 respectively)

PLAT343 Type_2 Hybridisation Problem on C in main residue(s)
 The angle range is larger than usual for the tentatively assigned
 hybridisation of the reported atom in the main residue. 

PLAT344 Type_2 Hybridisation Problem on C in solvent/ion
 The angle range is larger than usual for the tentatively assigned
 hybridisation of the reported atom in the solven/anion. 

PLAT350 Type_3 Test for short C - H (Angstrom Difference) XRAY: 0.96
 Large deviations from generally accepted values may indicate model problems,
 over refinement etc.  Default C-H = 0.96 Ang. (X-Ray) value from SHELXL. 

PLAT351 Type_3 Test for long C - H (Angstrom Difference) XRAY: 0.96
 Large deviations from generally accepted values may indicate model problems,
 over refinement etc.  Default C-H = 0.96 Ang. (X-Ray) value from SHELXL. 

PLAT352 Type_3 Test for short N - H (Angstrom Difference) XRAY: 0.87
 Large deviations from generally accepted values may indicate model problems,
 over refinement etc.  Default N-H = 0.87 Ang. (X-Ray) value from SHELXL. 

PLAT353 Type_3 Test for long N - H (Angstrom Difference) XRAY: 0.87
 Large deviations from generally accepted values may indicate model problems,
 over refinement etc.  Default N-H = 0.87 Ang. (X-Ray) value from SHELXL. 

PLAT354 Type_3 Test for short O - H (Angstrom Difference) XRAY: 0.82
 Large deviations from generally accepted values may indicate model problems,
 over refinement etc.  Default O-H = 0.82 Ang. (X-Ray) value from SHELXL.

PLAT355 Type_3 Test for long O - H (Angstrom Difference) XRAY: 0.82
 Large deviations from generally accepted values may indicate model problems,
 over refinement etc.  Default O-H = 0.82 Ang. (X-Ray) value from SHELXL. 

PLAT360 Type_2 Test for short C4 - C4 (Angstrom Difference) XRAY: 1.54
 Large deviations from generally accepted values may indicate model problems,
 over refinement etc. Default C4-C4 = 1.54 Ang. (X-Ray) value from
 Ladd & Palmer, Structure Determination by Xray Crystallography (1985).
 Note: 
  - C4-C4 indicates a bond between atoms with 4 bonds each.

PLAT361 Type_2 Test for long C4 - C4 (Angstrom Difference) XRAY: 1.54
 Large deviations from generally accepted values may indicate model problems,
 over refinement etc. Default C4-C4 = 1.54 Ang. (X-Ray) value from
 Ladd & Palmer, Structure Determination by Xray Crystallography (1985).
 Note: 
  - C4-C4 indicates a bond between atoms with 4 bonds each.

PLAT362 Type_2 Test for short C4 - C3 (Angstrom Difference) XRAY: 1.52
 Large deviations from generally accepted values may indicate model problems,
 over refinement etc. Default C4-C3 = 1.52 Ang. (X-Ray) value from
 Ladd & Palmer, Structure Determination by Xray Crystallography (1985).
 Note: 
  - C4-C3 indicates a bond between an atom with 4 bonds and one with 3 bonds.
  - Conjugated systems may cause some 'false alarm' messages.

PLAT363 Type_2 Test for long C4 - C3 (Angstrom Difference) XRAY: 1.52
 Large deviations from generally accepted values may indicate model problems,
 over refinement etc. Default C4-C3 = 1.52 Ang. (X-Ray) value from
 Ladd & Palmer, Structure Determination by Xray Crystallography (1985).
 Note: 
  - C4-C3 indicates a bond between an atom with 4 bonds and one with 3 bonds.
  - Conjugated systems may cause some 'false alarm' messages.

PLAT364 Type_2 Test for short C4 - C2 (Angstrom Difference) XRAY: 1.46
 Large deviations from generally accepted values may indicate model problems,
 over refinement etc. Default C4-C2 = 1.46 Ang. (X-Ray) value from
 Ladd & Palmer, Structure Determination by Xray Crystallography (1985).
 Note: 
  - C4-C2 indicates a bond between an atom with 4 bonds and one with 2 bonds.
  - Conjugated systems may cause some 'false alarm' messages.

PLAT365 Type_2 Test for long C4 - C2 (Angstrom Difference) XRAY: 1.46
 Large deviations from generally accepted values may indicate model problems,
 over refinement etc. Default C4-C2 = 1.46 Ang. (X-Ray) value from
 Ladd & Palmer, Structure Determination by Xray Crystallography (1985).
 Note: 
  - C4-C2 indicates a bond between an atom with 4 bonds and one with 2 bonds.
  - Conjugated systems may cause some 'false alarm' messages.

PLAT366 Type_2 Test for short C? - C? (Angstrom Difference) XRAY: 1.50
 Large deviations from generally accepted values may indicate model problems,
 over refinement etc. Default C?-C? = 1.50 Ang. (X-Ray) value from
 Ladd & Palmer, Structure Determination by Xray Crystallography (1985).
 This ALERT may also arise when the hybridisation at least one atom is
 not recognized.

PLAT367 Type_2 Test for long C? - C? (Angstrom Difference) XRAY: 1.50
 Large deviations from generally accepted values may indicate model problems,
 over refinement etc. Default C?-C? = 1.50 Ang. (X-Ray) value from
 Ladd & Palmer, Structure Determination by Xray Crystallography (1985).

PLAT368 Type_2 Test for short C3 - C3 (Angstrom Difference) XRAY: 1.34
 Large deviations from generally accepted values may indicate model problems,
 over refinement etc. Default C3-C3 = 1.34 Ang. (X-Ray) value from
 Ladd & Palmer, Structure Determination by Xray Crystallography (1985).
 Note: 
  - C3-C3 indicates a bond between atoms with 3 bonds each.
  - Conjugated systems may cause some 'false alarm' messages.

PLAT369 Type_2 Test for long C3 - C3 (Angstrom Difference) XRAY: 1.34
 Large deviations from generally accepted values may indicate model problems,
 over refinement etc.  Default C3-C3 = 1.34 Ang. (X-Ray) value from
 Ladd & Palmer, Structure Determination by Xray Crystallography (1985).
 Note: 
  - C3-C3 indicates a bond between atoms with 4 with 3 bonds each.
  - Conjugated systems may cause some 'false alarm' messages.
  - A notable exception is the C-C bond in -C(=O)-C(=O)- systems with an 
    observed mean value of 1.54 Angstrom.

PLAT370 Type_2 Test for short C3 - C2 (Angstrom Difference) XRAY: 1.31
 Large deviations from generally accepted values may indicate model problems,
 over refinement etc. Default C3-C2 = 1.31 Ang. (X-Ray) value from
 Ladd & Palmer, Structure Determination by Xray Crystallography (1985).
 Note: 
  - C3-C2 indicates a bond between an atom with 3 bonds and one with 2 bonds.
  - Conjugated systems may cause some 'false alarm' messages.

PLAT371 Type_2 Test for long C3 - C2 (Angstrom Difference) XRAY: 1.31
 Large deviations from generally accepted values may indicate model problems,
 over refinement etc. Default C3-C2 = 1.31 Ang. (X-Ray) value from
 Ladd & Palmer, Structure Determination by Xray Crystallography (1985).
 Note: 
  - C3-C2 indicates a bond between an atom with 3 bonds and one with 2 bonds.
  - Conjugated systems may cause some 'false alarm' messages.

PLAT372 Type_2 Test for short C2 - C2 (Angstrom Difference) XRAY: 1.25
 Large deviations from generally accepted values may indicate model problems,
 over refinement etc. Default C2-C2 = 1.25 Ang. (X-Ray) value from
 Ladd & Palmer, Structure Determination by Xray Crystallography (1985).
 Note: 
  - C2-C2 indicates a bond between atoms with 2 bonds each.
  - Conjugated systems may cause some 'false alarm' messages.

PLAT373 Type_2 Test for long C2 - C2 (Angstrom Difference) XRAY: 1.25
 Large deviations from generally accepted values may indicate model problems,
 over refinement etc. Default C2-C2 = 1.25 Ang. (X-Ray) value from
 Ladd & Palmer, Structure Determination by Xray Crystallography (1985).
 Note: 
  - C2-C2 indicates a bond between atoms with 2 bonds each.
  - Conjugated systems may cause some 'false alarm' messages.

PLAT374 Type_2 Test for long N - N Bond (> 1.45 Angstrom)
 Large deviations from generally observed bond distances may indicate model
 problems, over-refinement etc. Check for wrong atom-type assignments.
 For an example see: Acta Cryst. (2003) E59, m710-m712.

PLAT380 Type_4 Test for incorrectly Oriented Methyl Moiety
 This test alerts for possible incorrectly oriented CH3 moieties.
 (E.g. AFIX 33 instead of AFIX 137 etc. within the SHELXL realm)

PLAT390 Type_3 Test Methyl Moiety X-C-H Bond Angle
 Unusual Methyl Moiety X-C-H Angle (Ideally 109 Degrees for 4-bonded C).

PLAT391 Type_3 Test Methyl Moiety H-C-H Bond Angle
 Unusual Methyl Moiety H-C-H Angle (ideally 109 Degrees).

PLAT395 Type_2 Test X-O-Y Angle
 The X-O-Y angle is significantly larger than the expected 120.0 Degrees.

PLAT396 Type_2 Test Si-O-Si Angle
 The Si-O-Si angle is significantly larger than the expected 150.0 Degrees.

PLAT410 Type_2 Test for short non-bonding intra H..H contacts
 Short intramolecular contacts may arise when H-atoms are in (false)
 calculated positions. Short intramolecular contacts may also be a sign for a
 false structure with the molecule sitting on a site with improper site 
 symmetry (e.g. '2' instead of '-1') which may happen when a lattice
 translation is missed. Short contacts are defined using a van der Waals
 radius of 1.2 Angstrom. For intramolecular contacts alerts are generated
 for contacts less than 2.0 Angstrom.

PLAT411 Type_2 Test for short non-bonding inter H..H contacts
 Short intermolecular H..H contacts may indicate incorrectly determined 
 structures (i.e. wrong symmetry, missed translation symmetry, wrong position 
 with reference to the symmetry elements, hydrogen atoms on atoms where there
 should not be any  etc..)
 Short intermolecular contacts may be indicative for inconsistent symmetry
 data (e.g. coordinates for space-group P43 and symmetry specified as P41 or
 P21/n & P21/c confusions). Short contacts are defined using a van der Waals
 radius of 1.2 Angstrom. For intermolecular contacts, an alert is generated
 for contacts less than 2.4 Angstrom. 

PLAT412 Type_2 Test for short non-bonding intra H..H contacts (involving XH3)
 Short intramolecular contacts may arise when H-atoms are in (false)
 calculated positions. Short intramolecular contacts may also be a sign for a
 false structure with the molecule sitting on a site with improper site
 symmetry (e.g. '2' instead of '-1') which may happen when a lattice
 translation is missed. Short contacts are defined using a van der Waals
 radius of 1.2 Angstrom. Short H .. H contact involving CH3 H-atoms are often
 hampered by the fact that they involve H atoms at not optimal calculated
 positions.

PLAT413 Type_2 Test for short non-bonding inter H..H contacts (involving XH3)
 Short intermolecular H..H contacts may indicate incorrectly determined 
 structures (i.e. wrong symmetry, missed translation symmetry, wrong position 
 with reference to the symmetry elements, hydrogen atoms on atoms where there
 should not be any  etc..). Short intermolecular contacts may be indicative
 for inconsistent symmetry data (e.g. coordinates for space-group P43 and
 symmetry specified as P41 or P21/n & P21/c confusions). Short contacts are
 defined using a van der Waals radius of 1.2 Angstrom. Short H .. H contact
 involving CH3 H-atoms are often hampered by the fact that they involve
 H atoms at not optimal calculated positions.

PLAT414 Type_2 Test for short non-bonding intra D-H..H-X contacts
 Short non-bonding intra D-H..H-X contact.

PLAT415 Type_2 Test for short non-bonding inter D-H..H-X contacts
 Short non-bonding inter D-H..H-X contact.

PLAT416 Type_2 Test for short non-bonding intra D-H..H-D contacts
 Short non-bonding intra D-H..H-D contacts may be related to disordered or
 misplaced H-atoms.

PLAT417 Type_2 Test for short non-bonding inter D-H..H-D contacts
 Short non-bonding inter D-H..H-D contacts may be related to disordered or
 misplaced H-atoms.
 Experience has shown that any intermolecular H...H separation of less than
 1.8 Angstroms between unit-occupancy H atoms is a clear indicator that one
 or both of these H atoms may be wrongly placed.

PLAT420 Type_2 Test for D-H without acceptor
 Potential hydrogen bond donors are checked for the presence of suitable 
 acceptors using commonly used (Jeffrey) H-bond criteria. As a general rule
 there should be an acceptor for each donor. Exceptions are very rare for O-H
 and more common for -NH and -NH2. A common error is an -OH on a calculated
 position pointing in the wrong direction.

PLAT430 Type_2 Test for short non-bonding inter D...A contacts
 This test alerts for possibly missed Hydrogen bonds as indicated by short
 (i.e. shorter than sum of the van der Waals radii - 0.2 Angstrom)
 Donor - Acceptor distances.
 Note: Short C=O .. O=C are observed sometimes when part of three-centre
 O-H, N-H or C-H O..O bridging.  

PLAT431 Type_2 Test for short non-bonding inter HL...A contacts
 This test reports on short intermolecular Halogen .. Donor/Acceptor
 atom-type distances.

PLAT432 Type_2 Test for short non-bonding inter X...Y contacts
 This test raised an ALERT for short intermolecular contacts.
 In general, intermolecular contact distances should be not much smaller than
 the sum of the associated van der Waals Radii. More often than not, such  
 short contacts can be a warning sign for errors. All short contacts should 
 therefore be examined in some detail. Interesting exceptions are carbonyl-
 carbonyl interactions that often feature short O...C contacts (see Allen 
 et al. (1998) B54, 320-329, short NO2 O...O interactions and BF4(-) to
 (aromatic) carbon contacts.

PLAT433 Type_4 Test for short non-bonding minor..minor inter X...Y contacts
 This test raised an ALERT for short intermolecular contacts between minor 
 disorder components. In general, intermolecular contact distances should be 
 not much smaller than the sum of the associated van der Waals Radii. More
 often than not, such short contacts can be a warning sign for errors. All
 short contacts should therefore examined in some detail. Interesting 
 exceptions are carbonyl-carbonyl interactions that often feature short O...C 
 contacts (see Allen et al. (1998) B54, 320-329.

PLAT480 Type_4 Test for too large H...A
 Check this (unrealistically) long reported H..A contact.
 Jeffrey criterium: Contact .LT. vdWR(H) + vdWR(A) - 0.12 Angstrom. 

PLAT481 Type_4 Test for too large D...A
 Check this (unrealistically) long reported D..A contact.
 Jeffrey criterium: Contact .LT. vdWR(D) + vdWR(A) + 0.50 Angstrom.

PLAT482 Type_4 Test for too small D-H...A Angle
 Check this unrealistically small reported D-H..A Angle.
 Jeffrey criterium: D-H..A Angle .GT. 100 degrees.

PLAT601 Type_2 Test for (Unreported) solvent accessible voids
 Crystal structures in general do not contain large solvent accessible voids
 in the lattice. Most structures lose their long-range ordering when solvent
 molecules leave the crystal. Only when the remaining network is strongly
 bonded (e.g. zeolites and some hydrogen bonded networks) the crystal 
 structure may survive.
 Residual voids in a structure may indicate the omission of (disordered) 
 density from the model. Disordered density may go undetected when smeared
 since peak search programs are not designed to locate maxima on density
 ridges. The presence or absence of residual density in the void may be
 verified on a printed/plotted difference Fourier map or with PLATON/SQUEEZE.
 Voids of 40 Ang**3 may accommodate H2O. Small molecules such as 
 Tetrahydrofuran have typical volumes in the 100 to 200 Ang**3 range.
 This test reports the volume of the largest solvent accessible void in the
 structure.
 A paper reporting a crystal structure with a significant solvent accessible 
 void should at the least discuss the issue.

PLAT602 Type_2 Test for TOO LARGE (Unreported) solvent accessible voids
 This test reports on a solvent accessible void in the structure, too large 
 or too time consuming for the current PLATON version for a more detailed
 analysis as part of the validation run. Use the SOLV option for more details.
 Such a warning might also indicate that the symmetry is incomplete
 e.g. should have been specified as P-1 and not P1, leaving out half of
 the unit cell content.

PLAT603 Type_4 Test for TOO LARGE Unit Cell for VOID search
 No search for solvent accessible VOIDS done as part of VALIDATION in 
 view of large unit-cell. 

PLAT604 Type_4 Test for TOO Many VOIDS
 Too many solvent accessible VOIDS. 

PLAT605 Type_4 Test for (Reported) solvent accessible voids
 Crystal structures in general do not contain large solvent accessible voids
 in the lattice. Most structures lose their long-range ordering when solvent
 molecules leave the crystal. Only when the remaining network is strongly
 bonded (e.g. zeolites and some hydrogen bonded networks) the crystal 
 structure may survive.
 Residual voids in a structure may indicate the omission of (disordered) 
 density from the model. Disordered density may go undetected when smeared
 since peak search programs are not designed to locate maxima on density
 ridges. The presence or absence of residual density in the void may be
 verified on a printed/plotted difference Fourier map or with PLATON/SQUEEZE.
 Voids of 40 Ang**3 may accommodate H2O. Small molecules such as 
 Tetrahydrofuran have typical volumes in the 100 to 200 Ang**3 range.
 This test reports the volume of the largest solvent accessible void in the
 structure.
 A paper reporting a crystal structure with a significant solvent accessible 
 void should at the least discuss the issue.
 Note: The use of PLATON/SQUEEZE was reported in the CIF

PLAT606 Type_4 Test for TOO LARGE (Reported) solvent accessible voids
 This test reports on a solvent accessible void in the structure, too large 
 or too time consuming for the current PLATON version for a more detailed
 analysis as part of the validation run. Use the SOLV option for more details.
 Such a warning might also indicate that the symmetry is incomplete
 e.g. should have been specified as P-1 and not P1, leaving out half of
 the unit cell content.

PLAT701 Type_1 Test for consistency of Bonds and Coordinates in CIF
 Bond distances given in the CIF are cross-checked with corresponding
 values calculated from the coordinates.  Alerts are set at 1,2 and 3 sigma
 deviation levels.
 Note: Default s.u.'s are used where no su given (e.g. for C-H)
 In general, all differences should be within the associated s.u.
 Small differences may arise from rounding.
 Very large deviation (or zero distance) normally indicate incorrectly 
 specified symmetry operations on the associated atoms, or 'cut-and-pasting'
 of incompatible CIF's.

PLAT702 Type_1 Test for consistency of Angles and Coordinates in CIF
 Bond Angles given in the CIF are cross-checked with corresponding values
 calculated from the coordinates.  Alerts are set at 1,2 and 3 sigma 
 deviation levels.
 In general, all differences should be within the associated s.u.
 Small differences may arise from rounding. Very large deviations normally
 indicate incorrectly specified symmetry operations on the associated atoms,
 or 'cut-and-pasting' of incompatible CIF's.

PLAT703 Type_1 Test for consistency of Torsions and Coordinates in CIF
 Torsion angles given in the CIF are cross-checked with corresponding values
 calculated from the coordinates.  Alerts are set at 1,2 and 3 sigma 
 deviation levels.
 In general, all differences should be within the associated s.u.
 Small differences may arise from rounding. Very large deviations normally
 indicate incorrectly specified symmetry operations on the associated atoms,
 or 'cut-and-pasting' of incompatible CIF's.

PLAT704 Type_1 Test for consistency of Contact Distances and Coordinates in CIF
 Intermolecular contacts listed in the CIF are checked against the
 coordinates in the CIF.
 Alerts are set at 1,2 and 3 sigma deviation levels.

PLAT705 Type_1 Test for consistency of H-Bond D-H distances and Coordinates in CIF
 Hydrogen-Bond D-H listed in the CIF is checked.  Alerts are set at
 1,2 and 3 sigma deviation levels.

PLAT706 Type_1 Test for consistency of H-Bond H..A Distances and Coordinates in CIF
 Hydrogen-Bond H..A listed in the CIF is checked. Alerts are set at 1,2 and
 3 sigma deviation levels.
 This ALERT is generally related to incorrect symmetry codes. The symmetry 
 number s in the symmetry code s_pqr should correspond to the expression 
 for s in the CIF. Those expressions can be different for different 
 software packages. E.g. pasting H-bond table data generated with PLATON
 into a CIF generated with SHELXL may raise this ALERT. Manual correction 
 of the symmetry code should be trivial.

PLAT707 Type_1 Test for consistency of H-Bond D..A Distances and Coordinates in CIF
 Hydrogen-Bond D..A listed in the CIF is checked. Alerts are set at 1,2 and
 3 sigma deviation levels.
 This ALERT is generally related to incorrect symmetry codes. The symmetry 
 number s in the symmetry code s_pqr should correspond to the expression 
 for s in the CIF. Those expressions can be different for different 
 software packages. E.g. pasting H-bond table data generated with PLATON
 into a CIF generated with SHELXL may raise this ALERT. Manual correction 
 of the symmetry code should be trivial.

PLAT708 Type_1 Test for consistency of H-Bond D-H..A Angles and Coordinates in CIF
 Hydrogen-Bond Angle D-H..A listed in the CIF is checked. Alerts are set at
 1,2 and 3 sigma deviation levels.
 This ALERT is generally related to incorrect symmetry codes. The symmetry 
 number s in the symmetry code s_pqr should correspond to the expression 
 for s in the CIF. Those expressions can be different for different 
 software packages. E.g. pasting H-bond table data generated with PLATON
 into a CIF generated with SHELXL may raise this ALERT. Manual correction 
 of the symmetry code should be trivial.

PLAT710 Type_4 Test for Linear Torsions in CIF
 Torsion angles specified in the CIF are checked for the
 'linear variety' where one or both of the 1-2-3 and 2-3-4 bond
 angles are close to 180 Deg. SHELXL97 will generate those 'torsions' for
 molecules containing linear moieties (E.g. Metal-C=O).

PLAT711 Type_1 Test for label problems for Bonds in CIF
 When labels are found on geometry items (bonds, angles etc.) that are not in
 the coordinate list, and alert _71n is issued, related to alert _70n.

PLAT712 Type_1 Test for label problem for Angles in CIF
 When labels are found on geometry items (bonds, angles etc.) that are not in
 the coordinate list, and alert _71n is issued, related to alert _70n.

PLAT713 Type_1 Test for label problem for Torsions in CIF
 When labels are found on geometry items (bonds, angles etc.) that are not in
 the coordinate list, and alert _71n is issued, related to alert _70n.

PLAT714 Type_1 Test for label problem for Contact Distances in CIF
 When labels are found on geometry items (bonds, angles etc.) that are not in
 the coordinate list, and alert _71n is issued, related to alert _70n.

PLAT715 Type_1 Test for label problem for H-Bond D-H distances in CIF
 When labels are found on geometry items (bonds, angles etc.) that are not in
 the coordinate list, and alert _71n is issued, related to alert _70n.

PLAT716 Type_1 Test for label problem for H-Bond H..A Distances in CIF
 When labels are found on geometry items (bonds, angles etc.) that are not in
 the coordinate list, and alert _71n is issued, related to alert _70n.

PLAT717 Type_1 Test for label problem for H-Bond D..A Distances in CIF
 When labels are found on geometry items (bonds, angles etc.) that are not in
 the coordinate list, and alert _71n is issued, related to alert _70n.

PLAT718 Type_1 Test for label problem for H-Bond D-H..A Angles in CIF
 When labels are found on geometry items (bonds, angles etc.) that are not in
 the coordinate list, and alert _71n is issued, related to alert _70n.

PLAT720 Type_4 Test for Unusual Labels
 Up to 4 Character Labels of the type C11, H101, N10A, i.e. chemical symbol +
 number + optional letter are to be preferred.

PLAT721 Type_1 Test for consistency of Bonds and Coordinates in CIF
 Same as 701 but for distance without su (esd). Difference is tested in terms
 of Angstroms.

PLAT722 Type_1 Test for consistency of Angles and Coordinates in CIF
 Same as 702 but for angle without su (esd). Difference is tested in terms of
 Degrees.

PLAT723 Type_1 Test for consistency of Torsions and Coordinates in CIF
 Same as 703 but for torsion without su (esd). Difference is tested in terms 
 of Degrees.

PLAT724 Type_2 Test for consistency of Contact Distances and Coordinates in CIF
 Same as 704, but for distance without su (esd). Difference is tested in 
 terms of Angstroms.

PLAT725 Type_2 Test for consistency of H-Bond D-H distances and Coordinates in CIF
 Same as 705 but for distance without s.u. (esd). Differences are tested in
 terms of Angstrom.

PLAT726 Type_2 Test for consistency of H-Bond H..A Distances and Coordinates in CIF
 Same as 706 but for distance without s.u. (esd). Differences are tested in 
 terms of Angstrom.

PLAT727 Type_1 Test for consistency of H-Bond D..A Distances and Coordinates in CIF
 Same as 707 but for distance without s.u. (esd). Differences are tested in
 terms of Angstrom.

PLAT728 Type_1 Test for consistency of H-Bond D-H..A Angles and Coordinates in CIF
 Same as 708 but for angle without s.u. (esd). Differences are tested in 
 terms of Degrees.

PLAT731 Type_1 Test for consistency of Bond su's and Coordinate su's in CIF
 A large ratio of the reported and calculated bond s.u.'s is found. 
 The use of a DFIX instruction might cause such a warning since
 calculated s.u.'s are based on reported variances only.
 Note_1: su's on the unit-cell dimensions are taken into account in the
 calculation of expected su's. This may result in large differences 
 between expected and reported su's when this contribution is not included
 in the reported su's, in particular for inaccurate unit-cells.
 Note_2: Another source for the discrepancy between calculated and 
 reported su's can be that the validation software has access only to
 the variances of the refined parameters as opposed to the full co-variance
 matrix used by e.g. SHELXL for the calculation of derived parameters
 with associated su's. Constrained/restrained refinement may cause largei
 co-variances.  
 

PLAT732 Type_1 Test for consistency of Angles and Coordinates in CIF s.u.'s
 A large ratio of the reported and calculated bond angle s.u.'s is found. 
 This check should warn for erroneous rounding: E.g. 105.5(19) to 105.5(2) or 
 105.0(5) to 105(5) etc. Note: Large differences are possible when certain
 constraints/restraints were applied in the refinement (e.g. the FLAT option 
 in SHELXL97).
 Note: su's on the unit-cell dimensions are taken into account in the
 calculation of expected su's. This may result in large differences 
 between expected and reported su's when this contribution is not included
 in the reported su's, in particular for inaccurate unit-cells.
 Note_2: Another source for the discrepancy between calculated and 
 reported su's can be that the validation software has access only to
 the variances of the refined parameters as opposed to the full co-variance
 matrix used by e.g. SHELXL for the calculation of derived parameters
 with associated su's. Constrained/restrained refinement may cause large
 co-variances.  

PLAT733 Type_1 Test for consistency of Torsions and Coordinates in CIF s.u's
 A large ratio of the reported and calculated torsion angle s.u.'s is found. 
 This check should warn for erroneous rounding: E.g. 105.5(19) to 105.5(2) 
 or 105.0(5) to 105(5) etc.
 Note: su's on the unit-cell dimensions are taken into account in the
 calculation of expected su's. This may result in large differences 
 between expected and reported su's when this contribution is not included
 in the reported su's, in particular for inaccurate unit-cells.
 Note_2: Another source for the discrepancy between calculated and 
 reported su's can be that the validation software has access only to
 the variances of the refined parameters as opposed to the full co-variance
 matrix used by e.g. SHELXL for the calculation of derived parameters
 with associated su's. Constrained/restrained refinement may cause large
 co-variances.  

PLAT734 Type_1 Test for consistency of Contact Distance s.u. and Coordinate s.u. in CIF
 A large ratio of the reported and calculated contact distance s.u.'s is found. 
 Note: su's on the unit-cell dimensions are taken into account in the
 calculation of expected su's. This may result in large differences 
 between expected and reported su's when this contribution is not included
 in the reported su's, in particular for inaccurate unit-cells.

PLAT735 Type_1 Test for consistency of H-Bond D-H distance s.u. and Coordinate s.u in CIF
 A large ratio of the reported and calculated H-bond D-H distance s.u.'s is
 found. The use of a DFIX instruction might cause such a warning since
 calculated s.u.'s are based on reported variances only.
 Note: su's on the unit-cell dimensions are taken into account in the
 calculation of expected su's. This may result in large differences 
 between expected and reported su's when this contribution is not included
 in the reported su's, in particular for inaccurate unit-cells.

PLAT736 Type_1 Test for consistency of H-Bond H..A Distance s.u. and Coordinates in CIF
 A large ratio of the reported and calculated H-bond H..A distance s.u.'s 
 is found.
 Note: su's on the unit-cell dimensions are taken into account in the
 calculation of expected su's. This may result in large differences 
 between expected and reported su's when this contribution is not included
 in the reported su's, in particular for inaccurate unit-cells.

PLAT737 Type_1 Test for consistency of H-Bond D..A Distance s.u. and Coordinates in CIF
 A large ratio of the reported and calculated H-Bond D...A distance s.u.'s is
 found. 

PLAT738 Type_1 Test for consistency of H-Bond D-H..A Angle and Coordinates in CIF s.u.
 A large ratio of the reported and calculated H-Bond D-H..A angle s.u.'s 
 is found. 
 Note: su's on the unit-cell dimensions are taken into account in the
 calculation of expected su's. This may result in large differences 
 between expected and reported su's when this contribution is not included
 in the reported su's, in particular for inaccurate unit-cells.

PLAT741 Type_1 Test for missing Bond su in CIF
 Likely missing s.u. on Bond in CIF.

PLAT742 Type_1 Test for missing Angle s.u. in CIF
 Likely missing s.u. on Bond angle in CIF.

PLAT743 Type_1 Test for missing Torsion s.u. in CIF
 Likely missing s.u. on Torsion angle in CIF.

PLAT744 Type_1 Test for missing Contact Distance s.u. in CIF
 Likely missing s.u. on contact Distance in CIF.

PLAT745 Type_1 Test for missing H-Bond D-H distance s.u. in CIF
 Likely missing s.u. on H-Bond D-H distance in CIF.

PLAT746 Type_1 Test for missing H-Bond H..A Distance s.u. in CIF
 
 Likely missing s.u. on H-Bond H...A distance in CIF.

PLAT747 Type_1 Test for missing H-Bond D..A Distance s.u. in CIF
 Likely missing s.u. on H-Bond D...A distance in CIF.

PLAT748 Type_1 Test for missing H-Bond D-H..A Angle s.u. in CIF
 Likely missing s.u. on H-Bond D-H..A angle in CIF.

PLAT751 Type_4 Test for senseless Bond s.u. in CIF
 An s.u. should not be given in the CIF for constrained distances.
 Please check for proper refinement status flags (e.g. R)

PLAT752 Type_4 Test for senseless Angle s.u. in CIF
 An s.u. should not be given in the CIF for constrained angles.
 Please check for proper refinement status flags (e.g. R)

PLAT753 Type_4 Test for senseless Torsion s.u. in CIF
 An s.u. should not be given in the CIF for constrained torsion angles.
 Please check for proper refinement status flags (e.g. R)

PLAT754 Type_4 Test for senseless Contact Distance s.u. in CIF
 An s.u. should not be given in the CIF for constrained contact distances.
 Please check for proper refinement status flags (e.g. R)

PLAT755 Type_4 Test for senseless H-Bond D-H distance s.u. in CIF
 An s.u. should not be given in the CIF for constrained distances.
 Please check for proper refinement status flags (e.g. R)

PLAT756 Type_4 Test for senseless H-Bond H..A Distance s.u. in CIF
 An s.u. should not be given in the CIF for constrained distances.
 Please check for proper refinement status flags (e.g. R)

PLAT757 Type_4 Test for senseless H-Bond D..A Distance s.u. in CIF
 An s.u. should not be given in the CIF for constrained distances.
 Please check for proper refinement status flags (e.g. R)

PLAT758 Type_4 Test for Senseless H-Bond D-H..A Angle s.u. in CIF
 An s.u. should not be given in the CIF for constrained angles.
 Please check for proper refinement status flags (e.g. R)

PLAT761 Type_1 Test for the presence of at least one X-H in the CIF
 The CIF contains no X-H Bonds. This might be caused by not using the
 SHELXL instruction BOND $H
 Inclusion is required by Acta Cryst. but not necessarily so by other journals.
 

PLAT762 Type_1 Test for at least one X-Y-H or H-Y-H entry in the CIF
 The CIF contains no X-Y-H or H-Y-H bond angles. This might be caused by not
 using the SHELXL instruction BOND $H. Those data should also be supplied
 when H-atoms are introduced on calculated positions and/or refined riding
 on their carrier atom.  
 Inclusion is required by Acta Cryst. but not necessarily so by other journals.

PLAT763 Type_1 Test for missing bonds in CIF
 Bond list in CIF likely incomplete.

PLAT764 Type_4 Test for overcomplete bonds in CIF
 The CIF contains more bonds than the unique set, indicating redundancy. 
 An example is redundancy due to the inclusion of symmetry related bonds.

PLAT770 Type_2 Test for suspect C-H bonds in CIF (Not caught otherwise)
 Report on unusual C-H bonds not caught in other tests. 

PLAT771 Type_2 Test for suspect N-H bonds in CIF (Not caught otherwise)
 Report on unusual N-H bonds not caught in other tests.

PLAT772 Type_2 Test for suspect O-H bonds in CIF (Not caught otherwise)
 Report on unusual O-H bonds not caught in other tests.
 Note: Exceptions can be H-atoms in acid O..H..O bridges or in H5O2+
       (Hydronium) species.

PLAT773 Type_2 Test for suspect C-C bonds in CIF (Not caught otherwise)
 Report on unusual C-C bonds, possibly  not caught in other tests.
 Exceptions include C-C distances of around 1.75 Ang. in e.g.
 1,2-dicarba-closo-dodecaborane.

PLAT774 Type_1 Test for too large / erroneous bond distance
 Likely Erroneous Bond Entry  

PLAT775 Type_1 Test for too large / erroneous Contact distance
 Likely Erroneous Contact Entry  

PLAT776 Type_1 Test for too large / erroneous H-Bond D-H distance
 Likely Erroneous D-H Entry  

PLAT779 Type_4 Test for suspect Angle in CIF (Not caught otherwise)
 Possibly erroneous (Bond)angle less than 45 degree. The angle might be
 considered for elimination from the CIF when irrelevant. This ALERT can
 also be triggered when the assigned occupancy factors are incorrect.

PLAT780 Type_1 Test whether coordinates form a connected set
 Atoms given in a CIF should form a 'connected set', i.e. no symmetry 
 operations are needed to get atoms in a bonding position. A connected set of
 atoms is not needed for the least squares refinement (unless hydrogen atoms
 are to be added at calculated positions). Geometry listings (bonds, angles,
 torsions & H-bonds) become unwieldy for non-connected atom sets.

PLAT781 Type_1 Test for Flack x value for Centrosymmetric space-group
 A Flack parameter value is erroneously given for a structure reported
 in a centrosymmetric space-group.

PLAT782 Type_2 Test for Unusual C-NO2 an C-CO2 moiety geometry
 The geometry of the reported moiety appears to be unusual/inconsistent.
 The C-O bond distances in C-CO2 are expected to add up to about 2.5
 The N-O bond distances in C-NO2 are expected to add up to about 2.4

PLAT790 Type_4 Test Whether C.G. Residue in Unitcell Box
 Unless for a good reason, molecular species should be transformed
 (by symmetry and/or translation) so that their centres of gravity
 are close to or within the unit-cell bounds. This is a strict rule for the
 main species. Deviations from this general rule are for smaller additional
 species when relevant for intermolecular interactions with the main species.

PLAT791 Type_4 Check the absolute configuration of chiral atom in 'chiral' spgr
 This test addresses the consistency of the absolute configuration 
 assignment in non-centrosymmetric structures with proper symmetry 
 operations (i.e. all matrices with determinant = 1) only.
 Verify the (R/S) absolute configuration assignment of this atom and
 the consistency of the absolute configuration implicit in the CIF-data 
 with that in the 'ORTEP' illustration. Torsion angles should have the
 correct sign.
 The absolute structure assignment should also be consistent with the 
 lowest value of the Flack parameter and/or know absolute configuration.

PLAT792 Type_1 Check the absolute configuration of chiral atom in 'non-chiral' spgr
 This test addresses the consistency of the absolute structure assignment
 (i.e. polarity etc.) in non-centrosymmetric structures in space groups
 that include improper symmetry operations (e.g. mirror planes).
 
 Check the (R/S) absolute configuration assignment of this atom and
 the consistency of the absolute configuration implicit in the CIF-data 
 with that in the 'ORTEP' illustration.

PLAT793 Type_4 Check the absolute configuration of chiral atom in centrosymmetric spgr
 This test addresses the consistency of the absolute configuration assignment
 of molecules in the reported asymmetric unit among coordinates,
 molecular presentations and chemical diagrams.
 Check the (R/S) absolute configuration assignment of this atom and
 the consistency of the absolute configuration implicit in the CIF-data 
 with that in the 'ORTEP' illustration.
 

PLAT794 Type_5 Report the calculated 'Valence Bond' valency for metals
 This test reports the valency of an atom as predicted by the Valence
 Bond Model. See: 
 N.E. Brese & M. O'Keeffe (1991) Acta Cryst. B47, 192-197.
 I.D. Brown (2002). The Chemical Bond in Inorganic Chemistry: 
      The Bond Valence Model. Oxford University Press.
 More explicit info on the calculations can be obtained by running the
 calculations explicitly with the PLATON option BondValence.
 Note: The underlying theory is empirical and might not apply to the
       case at hand (e.g. charged species).

PLAT798 Type_4 Test for Alphanumeric Label on coordinate record
 Atom labels are generally not a number (i.e. starting with one or two
 characters indicating the atom type). Labels can be erroneously numeric
 due to typing errors (e.g. 'Oxygen' typed as 'zero').
 

PLAT799 Type_4 Test for Alphanumeric Label on displacement par. record
 Atom labels are generally not a number (i.e. starting with one or two
 characters indicating the atom type). Labels can be erroneously numeric
 due to typing errors (e.g. 'Oxygen' typed as 'zero').
 

PLAT801 Type_4 Test for missing, incomplete or out-of-order Cell data
 
 PLATON has a problem with the Cell data. The possible reason can be that
 the cell data are missing, incomplete or out-of-sequence. PLATON/CHECK wishes
 to see the cell and symmetry data before any coordinates are given.
 PLATON expects the values of all six cell parameters.

PLAT802 Type_4 Test for Input lines longer than 80 Characters
 The CIF contains records longer than 80 characters. Not all software will  
 read beyond column 80.
 The CIF-1.1 definition specifies a maximum of 2048 character per record.

PLAT803 Type_1 Test for Loop problem in CIF-Read
 Fatal Problem: Check loop data names and data for errors. There are 
 likely too many or to few data in the loop.

PLAT804 Type_5 Test for ARU-Pack Problem in PLATON
 Problem: ARU representations turn out to be needed outside the ORTEP
 style -5:5 unit-cell translation range. The Analysis might be incomplete.
 The problem often occurs for structures with aliphatic chains stretching
 over many unitcells or network structures. Transformation of the unitcell
 content to a symmetry related position might solve the problem. 

PLAT805 Type_1 Test for insufficient 'coordinate data'
 Check Coordinate Data Loop.
 

PLAT806 Type_4 Test for insufficient 'UIJ data'
 Check UIJ Data Loop.
 

PLAT807 Type_5 Test for Maximum number of ATOMS Exceeded Problem
 PLATON can handle up to 7000 atoms in the (expanded) ATOM list. 
 This might happen with disordered or network structures in high
 symmetry space groups. Deletion of the symmetry information might solve
 part of the problem and provide a partial validation. 
 Alternatively, clicking on 'NOSYMM' on the PLATON menu before invoking
 validation might address the problem.
 

PLAT808 Type_5 Test for Parseable SHELXL style Weighting scheme
 The software did not succeed in finding/analyzing a parsable weighting 
 scheme.      
 SHELXL style weight parameters are expected to be given in the format: 
 _refine_ls_weighting_details
 'calc w=1/[\s^2^(Fo^2^)+(0.1000P)^2^+0.0000P] where P=(Fo^2^+2Fc^2^)/3'
 JANA style weight is expected to be given in the format:
 _refine_ls_weighting_details 'w=1/(\s^2^(I)+0.0016I^2^)'  
  Do not edit this string or make it into a text block between ';'. 
 

PLAT809 Type_1 Test the SHELXL style Weighting scheme
 The software did not succeed in Parsing SHELXL style weighting 
 scheme. The string might have been edited.
 

PLAT810 Type_5 Test for Out-of-Memory Problem
 Analysis for missing reflections may be incomplete due to an out-of-memory
 problem.
 

PLAT811 Type_5 Test for No ADDSYM Analysis
 The ADDSYM test for missed symmetry is not executed for structures with too
 many disordered atoms. 
 

PLAT820 Type_5 Report Read Problem in PLATON/PLA230
 Internal PLATON Problem. Please refer problem to author at a.l.spek@uu.nl

PLAT850 Type_4 Test for BASF/TWIN Problem in SHELXL
 This G_ALERT can be ignored in the case that the so-called 'on-the-cheap' 
 Flack parameter is reported as determined with SHELXL. 
 Exactly zero values are possible but may also be a software artefact.
 The following should be checked:
 
 Problem #1: Some SHELXL97 versions do not allow negative values of the
 Flack parameter when determined using the BASF/TWIN instructions.
 Negative values are set to 0.00001. Refinement may not converge completely. 
 Problem #2: Some SHELXL97 versions put meaningless values in the CIF for the
 Flack parameter when 'TWIN -1 0 0 0 -1 0 0 0 -1 2 / BASF' instructions (i.e.
 an explicit matrix is specified on the TWIN instruction) are used.
 Please check the value of BASF (in the list output)  against the Flack
 parameter in the CIF. 	
 

PLAT860 Type_3 Test for restraints used in refinement
 The use of restraints used in the refinement should be explained in
 the write-up of a structure analysis.
 Note: An exception are restraints for floating origins (e.g. in P21).

PLAT870 Type_4 Report the Suppression of some Twinning Related ALERTS
 ALERTS related to twinning effects that can not (yet) be accounted
 for as part of the VALIDATION algorithms have been suppressed.

PLAT871 Type_4 Report the Suppression of some non-Mo,Cu,Ag Radiation ALERTS
 ALERTS related to the use of radiation other than AgKa, MoKa, CuKa
 have been suppressed.

PLAT900 Type_1 Test for 'No-matching reflection file'
 Likely cause: Dataset names on CIF and FCF differ.
 Note: FCF Validation is Skipped for this Entry.

PLAT901 Type_1 Test for CIF & FCF CELL Not Matching
 Likely causes: Wrong Dataset, CIF or FCF Parameters Edited inconsistently.
 Note: FCF Validation is Skipped for this Entry.

PLAT902 Type_1 Test for non-zero number of recognised reflections on FCF
 Either no reflections in FCF or uninterpretable due to unknown
 format or editing.
 Note: FCF Validation is Skipped for this Entry.

PLAT908 Type_2 Report on Max Observed data in any Resolution Shell
 A low maximum percentage of reflections with I .gt. 2*s(I) may     
 indicate:
 1 - Missed translation symmetry. E.g. all reflections hkl weak
     for l = 2n +1
 2 - Pseudo-merohedral twinning, index .gt. 1. (e.g. non-spacegroup
     extinctions.
 3 - Very weak observed data.

PLAT909 Type_3 Report of Observed data at Theta Cutoff
 This ALERT Reports on whether there is still a significant level of
 observed data beyond the Theta cutoff of the Dataset. There should
 be a good reason for a cutoff below sin(theta)/lambda = 0.6. 
 

PLAT910 Type_3 Test for missing reflections below Theta-Min
 Possible causes: Beamstop theta-min limit set too high, large unit-cell etc.
 A possible technical solution on CCD based equipment involves the collection 
 of additional images with the detector at a larger distance from the 
 crystal with the beamstop setting changed accordingly..

PLAT911 Type_3 Test for missing reflections between Theta-Min and sinth/lambda=0.6
 Possible causes: Missing cusp of data (due to rotation about one axis),
 deleted (overflow) reflections or improper strategy (orthorhombic for
 monoclinic crystal etc.) 

PLAT912 Type_4 Test for missing reflections above Sin(TH)/Lambda = 0.6
 Possible causes: Missing cusp of data (due to rotation about one axis),
 deleted (overflow) reflections or improper strategy (orthorhombic for
 monoclinic crystal etc.) 

PLAT913 Type_3 Test for missing strong reflections
 This ALERT reports the number of missing reflections with Fc**2 values
 greater than the largest Fc**2 value in the FCF. 
 Possible causes: Missing cusp of data (due to rotation about one axis),
 deleted (overflow) reflections or improper strategy (orthorhombic for
 monoclinic crystal etc.) or behind the beamstop. 

PLAT914 Type_3 Test for absence of Bijvoet Pairs in non-centro structure
 This ALERT reflects the notion that a dataset should contain a sufficient
 number of Bijvoet (Friedel) pairs for the reliable determination of the
 absolute structure of a non-centrosymmetric crystal structure. 
 This test is invoked when a Flack parameter value is specified.
 Note: SHELXL97 will calculate/report a Flack parameter value even for
 refinement against Friedel merged data. Remove the Flack entry from the 
 CIF.

PLAT915 Type_3 Test for low Friedel Pair Coverage in non-centro structure
 This ALERT reflects the notion that a dataset should contain a sufficient
 number of Bijvoet (Friedel) pairs for the reliable determination of the 
 absolute structure of a non-centrosymmetric crystal structure. 
 A Friedel coverage that deviates significantly from 100 percent may
 bias/invalidate the value of the Flack parameter.

PLAT916 Type_2 Test for differing Flack x and Hooft y Parameter values
 The Hooft y Parameter is calculated independently from the Bijvoet 
 differences and should have a value similar (observing the s.u.'s) to that of
 the Flack x Parameter. 
 See:
 Hooft, R.W.W, Straver, L.H. & Spek,A.L. (2008). J. Appl, Cryst. 41, 96-103.
 Thompson,A.L. & Watkin, D.J. (2009). Tetrahedron: Asymmetry,
   doi:10.1016/j.tetasy.2009.02.025
 Large differences may arise in cases where the Flack parameter was not
 done with BASF/TWIN or with essentially centrosymmetric data.
 See:
 Flack, H.D., Bernardinelli, G, Clemente, D.A., Linden, A. Spek, A.L. (2006)
 Acta Cryst. B62, 695-701.

PLAT917 Type_2 Test/Report whether FCF is based on a BASF/TWIN refinement
 The contribution of F(-h,-k,-l) to F(h,k,l) is likely not included in the
 FCF file. This usually indicates that the Flack parameter was NOT determined
 with a BASF/TWIN type of refinement. 

PLAT918 Type_3 Test for reflections with I(obs) << I(calc)
 This ALERT reports on the number of reflections with  
 (Fo**2 - Fc**2) / Sigma(Fo**2) < - 100.0. Those reflections are better
 removed from the final refinement since they are in systematic error.
 Of course, a valid reason for this problem should be found.

PLAT919 Type_3 Test for reflections effected by the beamstop
 This ALERT reports the number of reflections with intensities seriously 
 effected by the beamstop. Reflections are counted for which theta < 3 
 Degrees and (Fo**2 - Fc**2) / sqrt(weight) < - 10.0. Those reflections 
 are better removed from the final refinement since they are in systematic
 error.

PLAT920 Type_1 Test for TH(Max) Consistency between CIF & FCF
 Check reflection statistics of the data in the FCF for consistency with 
 the data reported in the CIF.     
 A difference usually indicates an edited CIF or an FCF file that was
 not created in the same SHELXL run where the CIF was created.

PLAT921 Type_1 Test for R1 Consistency between CIF & FCF(Reported)
 Please check whether the supplied FCF corresponds with the CIF produced in 
 the same least squares refinement job.      
 The test is based on the observed and calculated F**2 in the FCF and de
 weight parameters taken from the CIF.

PLAT922 Type_1 Test for wR2 Consistency between CIF & FCF(Reported)
 Please check whether the supplied FCF corresponds with the CIF produced in 
 the same least squares refinement job.     
 The test is based on the observed and calculated F**2 in the FCF and de
 weight parameters taken from the CIF.

PLAT923 Type_1 Test for S Consistency between CIF & FCF(Reported)
 Please check whether the supplied FCF corresponds with the CIF produced in 
 the same least squares refinement job.     
 The test is based on the observed and calculated F**2 in the FCF and de
 weight parameters taken from the CIF.

PLAT924 Type_1 Test for Consistency of the Reported & Calculated Rho(min)
 Check & Explain why the Reported Rho(min) differs significantly from 
 the value calculated on the basis of the reported structure.     
 Note: The Reported and Calculated values may differ slightly due to
 a differing peak interpolation algorithm.

PLAT925 Type_1 Test for Consistency of the Reported & Calculated Rho(max)
 Check & Explain why the Reported Rho(max) differs significantly from 
 the value calculated on the basis of the reported structure.     
 Note: The Reported and Calculated values may differ slightly due to
 a differing peak interpolation algorithm.

PLAT926 Type_1 Test for R1 Consistency between CIF & FCF(Calculated)
 Please check whether the R1 value that is reported in the CIF corresponds
 with the R1 value calculated from the parameters supplied in the CIF.
 This test is based on the observed reflection data in the FCF and 
 reflection data that are calculated with the parameters (i.e. coordinates,
 displacement and weight parameters) taken from the CIF.

PLAT927 Type_1 Test for wR2 Consistency between CIF & FCF(Calculated)
 Please check whether the wR2 value that is reported in the CIF corresponds
 with the wR2 value calculated from the parameters supplied in the CIF. 
 This test is based on the observed reflection data in the FCF and 
 reflection data that are calculated with the parameters (i.e. coordinates,
 displacement and weight parameters) taken from the CIF.

PLAT928 Type_1 Test for S Consistency between CIF & FCF(Calculated)
 Please check whether the S value that is reported in the CIF corresponds
 with the S value calculated from the parameters supplied in the CIF. 
 This test is based on the observed reflection data in the FCF and 
 reflection data that are calculated with the parameters (i.e. coordinates,
 displacement and weight parameters) taken from the CIF.

PLAT929 Type_5 Test for interpretable weight parameters for R1,wR2 & S Comparison
 SHELXL weight parameters are expected to be given in the format below: 
 _refine_ls_weighting_details
 'calc w=1/[\s^2^(Fo^2^)+(0.1000P)^2^+0.0000P] where P=(Fo^2^+2Fc^2^)/3'
 JANA style weight is expected to be given in the format:
 _refine_ls_weighting_details 'w=1/(\s^2^(I)+0.0016I^2^)'  
  Do not edit this string or make it into a text block between ';'. 

PLAT930 Type_2 Test for Missed Twinning from FCF data
 Check the proposed Twin Law. The entry in () represents the proposed 
 rotation axis in reciprocal space and the one in [] the corresponding
 rotation is direct space. 
 The relevant Twin Matrix can be found in the file '.ckf'.
 Note: This analysis is based on Fo/Fc differences with Fc data given
 in the .fcf file (i.e. Fobs, Fcalc listing). 
 
 ALERT-930 is expected to generate an related ALERT-931 as well.

PLAT931 Type_5 Test for Missed Twinning from FCF/CIF data
 Check the proposed Twin Law. The entry in () represents the proposed 
 rotation axis in reciprocal space and the one in [] the corresponding
 rotation is direct space. 
 The relevant Twin Matrix can be found in the file '.fck'.
 Note: This test is based on F(calc) values calculated with the data
 in the CIF. This ALERT can be ignored when twinning has been addressed
 in the refinement (As indicated by the Absence of ALERT 930).
 Please check whether twinning is mentioned in the write-up of the
 paper.

PLAT934 Type_3 Report number of outliers
 This ALERT reports on the number of reflections for which I(obs) and I(calc)
 differ more that 10 times SigmaW. (The latter being the square root of the
 weight for that reflection in the L.S. refinement). The reason for those 
 deviations should be investigated. When shown to be systematic errors,
 those reflections are best removed from the refinement and their omission
 from the refinement reported in the experimental section of an associated
 paper.

PLAT935 Type_2 Pseudo Extinction Parameter Test
  
 Both significantly positive and significantly negative values should
 invoke a search for a likely cause and a corrective action.
  

PLAT940 Type_3 Test for wR2 refinement with all data
 Apparently, observed data with I > n * sigma(I) were used in the F**2
 least squares refinement, rather than all observed data. 

PLAT950 Type_1 Test for Reported and Calculated Hmax Difference
 Reported and Calculated Max(Hmax,-Hmin) values differ by more than one
 unit. Check the consistency of wavelength and reported resolution data.

PLAT951 Type_1 Test for Reported and Calculated Kmax Difference
 Reported and Calculated Max(Kmax,-Kmin) values differ by more than one
 unit. Check the consistency of wavelength and reported resolution data.

PLAT952 Type_1 Test for Reported and Calculated Lmax Difference
 Reported and Calculated Max(Lmax,-Lmin) values differ by more than one
 unit. Check the consistency of wavelength and reported resolution data.

PLAT953 Type_1 Test for Reported and Actual Hmax Difference in the FCF File
 Reported (in the CIF) and Actual (in the FCF)  Max(Hmax,-Hmin) values
 differ by more than one unit. Check for data set truncation.

PLAT954 Type_1 Test for Reported and Actual Kmax Difference in the FCF File
 Reported (in the CIF) and Actual (in the FCF) Max(Kmax,-Kmin) values
 differ by more than one unit. Check for data set truncation.

PLAT955 Type_1 Test for Reported and Actual Lmax Difference in the FCF file
 Reported (in the CIF) and Actual (in the FCF) Max(Lmax,-Lmin) values
 differ by more than one unit. Check for data set truncation.

PLAT960 Type_3 Test for reflections with I < - 2 sigma
 Multiple strongly negative intensities may be indicative for poor 
 integration of the  diffraction images. 
 Too many negative intensities may result in higher than usual wR2 values.

PLAT971 Type_2 Test maximum calculated residual density
 Larger than expected residual density maximum outside heavy atom locations.
 This might be caused by unaccounted for twinning, wrongly assigned 
 atom types and other model errors. 
 

PLAT972 Type_2 Test minimum calculated residual density
 Larger than expected residual density minimum outside heavy atom locations. 
 This might be caused by unaccounted for twinning, wrongly assigned 
 atom types and other model errors.

PLAT973 Type_2 Test for large positive density on metal atom
 Larger than expected residual density maximum on heavy atom location. 
 This might be caused by unaccounted for twinning, wrongly assigned 
 atom types and other model errors.
 Another cause may be a SHELXL 'DAMP 0 0' instruction for a non-converged
 refinement.

PLAT974 Type_2 Test for large negative density close to metal atom
 Larger than expected residual density minimum on heavy atom location. 
 This might be caused by unaccounted for twinning, wrongly assigned 
 atom types and other model errors.
 Another cause may be a SHELXL 'DAMP 0 0' instruction for a non-converged
 refinement.

PLAT980 Type_1 Test for non-zero number of anomalous scattering factors
 Check for missing anomalous scattering factors.

PLAT981 Type_1 Test for non-zero f" anomalous scattering factor values
 Check for non-zero f" anomalous scattering factor values in the CIF.
 Note: Zero values are correct for SHELXL MERG 4 refinements.

PLAT982 Type_1 Test the anomalous scattering factor f' values against IT
 Check the anomalous scattering factor values.

PLAT983 Type_1 Test the anomalous scattering factor f" values against IT
 Check the anomalous scattering factor values.