Figure 16
Rcp versus dose (image number) for Au derivatives of proteinase K, under the assumption that the dose per image is constant across all crystals. (a) Data from the first beamline visit display signs of radiation damage after around 25–30 images. (b) Data from the second beamline visit display no obvious signs of radiation damage in the plot of Rcp versus dose (image number). (c) Combined data from both beamline visits. A plot of Rcp versus dose (image number) indicates possible radiation damage after around 25 images. (d) Combined data from both beamline visits, using only the first 25 images from each data set. The plot of Rcp versus dose (image number) displays no obvious sign of radiation damage.