research papers
A synchrotron-radiation-based circular-dichroism end-station has been implemented at beamline BL04B at the National Synchrotron Radiation Research Center (NSRRC) in Taiwan for biological research. The design and performance of this compact end-station for measuring circular-dichroism spectra in the vacuum-ultraviolet region are described. The linearly polarized light from the beamline is converted to modulated circularly polarized light with a LiF photoelastic modulator to provide a usable wavelength region of 130-330 nm. The light spot at the sample position is 5 mm × 5 mm at a slit width of 300 µm and provides a flux greater than 1 × 1011 photons s-1 (0.1% bandwidth)-1. A vacuum-compatible cell made of two CaF2 windows has a variable path length from 1.3 µm to 1 mm and a temperature range of 253-363 K. Measured CD spectra of (1S)-(+)-10-camphorsulfonic acid and proteins demonstrated the ability of this system to extend the wavelength down to 172 nm in aqueous solution and 153 nm in hexafluoro-2-propanol.