laboratory notes
The surge of macromolecular crystallography is intimately linked to the advent of methods for cryoprotecting macromolecular crystals. Only if crystals are kept cold during data collection can they withstand the effects of radiation damage during a diffraction experiment, especially at third-generation synchrotron sources. While a number of different cryoprotective agents and procedures have been described in the literature over the past three decades, it is still a time- and crystal-consuming process to establish and optimize a good cryo-condition for a specific crystal. In this study, trimethylamine N-oxide (TMAO) has been identified as a very versatile cryoprotectant for macromolecular crystals. In a few test cases it was shown that diffraction data collected from crystals treated with TMAO are of very good quality.