research papers
Serial crystallography on in vivo grown microcrystals using synchrotron radiation
aCenter for Free-Electron Laser Science (CFEL), Deutsches Elektronensynchrotron (DESY), Notkestrasse 85, 22607 Hamburg, Germany, bEuropean Molecular Biology Laboratory (EMBL), Hamburg Outstation, Notkestrasse 85, 22607 Hamburg, Germany, cJoint Laboratory for Structural Biology of Infection and Inflammation, Institute of Biochemistry and Molecular Biology, University of Hamburg, and Institute of Biochemistry, University of Lübeck, Notkestrasse 85, 22607 Hamburg, Germany, dInstitute of Biochemistry and Molecular Biology, University of Hamburg, Notkestrasse 85, 22607 Hamburg, Germany, eInterfaculty Institute of Biochemistry, University of Tübingen, Hoppe-Seyler-Strasse 4, 72076 Tübingen, Germany, and fInstitute of Experimental Physics, University of Hamburg, Luruper Chaussee 149, 22761 Hamburg, Germany
*Correspondence e-mail: thomas.schneider@embl-hamburg.de, henry.chapman@desy.de, redecke@biochem.uni-luebeck.de
Trypanosoma brucei procathepsin B crystals with an average volume of 9 µm3, a complete data set to 3.0 Å resolution has been assembled. The data allowed the of a structural model that is consistent with that previously obtained using free-electron laser radiation, providing mutual validation. Further improvements of the serial synchrotron crystallography technique and its combination with serial femtosecond crystallography are discussed that may allow the determination of high-resolution structures of micrometre-sized crystals.
determinations of biological macromolecules are limited by the availability of sufficiently sized crystals and by the fact that crystal quality deteriorates during data collection owing to radiation damage. Exploiting a micrometre-sized X-ray beam, high-precision diffractometry and shutterless data acquisition with a pixel-array detector, a strategy for collecting data from many micrometre-sized crystals presented to an X-ray beam in a vitrified suspension is demonstrated. By combining diffraction data from 803D view: 4n4z
PDB reference: cathepsin B, 4n4z
1. Introduction
Macromolecular crystallography (MX) is a powerful method for obtaining structural information about biological macromolecules and their assemblies. Since the 1990s, advanced third-generation synchrotrons have been used to produce micrometre-sized high-flux X-ray beams whose focus size matches the size of small crystals (Cusack et al., 1998; Riekel et al., 2005; Evans et al., 2011; Smith et al., 2012). X-ray beams with dimensions of less than 10 µm are now in routine use at many synchrotron-radiation facilities (Evans et al., 2011) and enable the determination of crystal structures from crystals with volumes of less than 1000 µm3 (Cusack et al., 1998). Using these microbeams, the structures of the cypovirus polyhedra protein (Coulibaly et al., 2007), amyloid-like fibres (Nelson et al., 2005) and a number of complexes addressing the structure and function of G-protein-coupled receptors (GPCRs; Cherezov et al., 2007; Rasmussen et al., 2007, 2011) have been determined from micrometre-sized crystals.
Inelastic and absorption events deposit energy in the crystal, triggering a variety of chemical reactions that lead to a repositioning of atoms (during the experiment), ultimately reducing the crystalline order and corresponding to increasing levels of `radiation damage'. As a consequence, even under cryogenic conditions, the quality of the diffraction data, in particular those at high resolution, quickly deteriorates as a function of the X-ray dose applied to the crystal. Since the earliest days of macromolecular crystallography, this problem has been addressed by combining partial data sets collected from multiple crystals into a full data set to high resolution. In the 1990s cryogenic cooling became routine in macromolecular crystallography (Henderson, 1990; Garman & Schneider, 1997), extending the lifetime of individual crystals, but with the implementation of undulator beamlines on third-generation synchrotrons the tolerable X-ray dose could be deposited in a short time even for cryocooled crystals (Ravelli & Garman, 2006). Data-collection schemes employing multiple positions on a single crystal (Riekel et al., 2005), to distribute the delivered dose over the entire crystal, or partial data sets from multiple crystals (Brodersen et al., 2003) were developed. The main complication with multicrystal methods is that systematic differences between data measured on different crystals, or at different positions on a single crystal, can prevent the derivation of a consistent assembled data set. Nevertheless, Hendrickson and coworkers (Liu et al., 2012) have recently shown that by removing sources of experimental error and by carefully selecting consistent partial data sets for merging, data sets of sufficient accuracy for phasing using the anomalous signal from S atoms in native proteins can in fact be assembled. Systematic errors in data collections from multiple positions in the same crystal can, amongst other techniques, be reduced by `helical scan' procedures (Flot et al., 2010), in which a single crystal, usually needle-shaped, is continuously moved during data collection to expose fresh parts of the same crystal.
When working with small crystals, it is often not possible to locate the crystals accurately using optical microscopy owing to their small size, optical distortions caused by materials surrounding the crystal, or the crystal being embedded in an opaque matrix. All of these problems are present during the data collection of GPCR crystals grown from lipidic cubic phase (LCP). Cherezov et al. (2009) described a method in which the mounted LCP containing crystals is first systematically rastered using a small beam to test each position for diffraction. Data are then collected in a second step using the rotation method (Arndt & Wonacott, 1977) from positions where diffraction was actually detected. The application of this method has been crucial in the of GPCRs, where the crystals are small (of the order of 25 × 6 × 4 µm3) and invisible to optical light because the lipidic in which the crystals grow turns opaque upon cryogenic cooling. Recently, an alternative procedure was proposed in which radiographs are taken of the entire sample mount to localize crystals (Warren et al., 2013).
Extremely short pulses of X-rays, each of tens of femtoseconds duration or even shorter, generated by a free-electron laser (FEL), have recently been shown to overcome the dose limitations mentioned above. Referred to as `diffraction before destruction', the inertia of the atoms prevents significant atomic displacement during the short exposing pulse (Neutze et al., 2000; Boutet et al., 2012; Barty et al., 2012), even though the initial photoabsorption eventually leads to plasma formation and the complete vaporization of the sample. A typical serial femtosecond crystallography (SFX) experiment collects tens or hundreds of thousands of single-crystal X-ray diffraction `snapshots' from a liquid suspension of protein microcrystals that flows across the focused X-ray beam at room temperature (Chapman et al., 2011). Since each FEL pulse destroys the sample, only one diffraction pattern can be collected per crystal. New data-processing tools, such as Cheetah, CrystFEL (White et al., 2012, 2013; Kirian et al., 2011) and cctbx.xfel (Kern et al., 2013), have been developed to process these large data volumes and have been successfully applied to several examples (Chapman et al., 2011; Boutet et al., 2012; Redecke et al., 2013; Kern et al., 2013).
The Trypanosoma brucei procathepsin B (TbCatB) in complex with its native propeptide represents the first novel bioinformation obtained by applying the SFX approach at a free-electron laser (Redecke et al., 2013). This enzyme is of scientific and medical interest, since the knockout of its encoding gene has been shown to be lethal to the parasite that causes sleeping sickness in Africa (Abdullah et al., 2008; Bryant et al., 2009), which turns TbCatB into an urgently required potential new drug target (Fairlamb, 2003). TbCatB crystals grew spontaneously within living baculovirus-infected insect cells during protein overexpression to a size of 10–15 µm in the longest dimension (Koopmann et al., 2012). In a recent experiment (Redecke et al., 2013), these crystals were used to obtain 178 875 single-crystal diffraction patterns by SFX at LCLS, which enabled of the enzyme to 2.1 Å resolution.
ofInspired by the SFX methodology and the new capabilities to process large data sets, as well as the successful T. brucei procathepsin B using a suspension of in vivo grown microcrystals mounted in a standard nylon loop for crystallographic data collection on a microfocus synchrotron beamline. Owing to the presence of cell remnants in the in vivo crystal suspension, individual TbCatB crystals are difficult to detect in the cryocooled sample. Our procedure combined elements from SFX with a helical line-scan approach used in microcrystallography. As with the SFX approach, an initially unknown subset of the recorded detector frames contain diffraction signals, which are selected for further processing into a crystallographic data set. Unlike the snapshots recorded at an FEL, the sample is rotated during exposure and multiple exposures from the same crystal can be obtained and subsequently processed in a consistent manner. In the following, we describe the diffraction experiment, data processing and The independent determination of the of pro-cathepsin B from T. brucei via two different methods using synchrotron and FEL radiation sources provided a unique opportunity to validate the results obtained against each other, and we present a comparison of the crystallographic models.
of TbCatB in the pro-form using the SFX approach, we embarked on an experiment to determine the of2. Results
2.1. Sample preparation and data collection
Spontaneous crystallization of TbCatB was obtained in baculovirus-infected Sf9 insect cells following our previously established protocol (Fig. 1; Koopmann et al., 2012). The in vivo crystals were isolated and purified by cell lysis and stepwise centrifugation based on their high mechanical and chemical stability. The purity of the crystal preparation was verified by (SEM). Prior to diffraction data collection, the suspension of needle-shaped crystals containing approximately 5 × 108 crystals per millilitre with average dimensions of 0.9 × 0.9 × 11 µm (approximately 9 µm3 in volume) was supplemented with 40%(v/v) glycerol as a cryoprotectant. A small volume of approximately 13 nl of the crystalline suspension was mounted after settling in a standard 20 µm thick nylon loop (Hampton Research, USA; 0.7 mm diameter) containing approximately 5000 crystals.
Diffraction experiments were conducted on the P14 microfocus beamline at the PETRA III storage ring (DESY, Hamburg) with a 4 × 5 µm (FWHM) microfocus beam, a total 12 photons s−1 at the sample position and a photon energy of 10.00 keV. The nylon loop was mounted on an MK3 mini-kappa goniometer head attached to an MD3 microdiffractometer (ARINAX, Moirans, France) and kept at 110 K using a gaseous nitrogen stream (Cryojet XL, Oxford Instruments, England). Diffraction data were recorded on a PILATUS 6M-F detector (DECTRIS Ltd, Baden, Switzerland).
of 1.2 × 10In a set of initial experiments, the diffraction properties of individual TbCatB crystals were characterized. After optically centring selected crystals with respect to the X-ray beam, single diffraction images collected with an oscillation range of 1° at an exposure time of 2 s showed diffraction spots extending to a resolution higher than 3 Å (Supporting Fig. S2). Using a series of short rotation exposures, by visual inspection of diffraction images and using the ability to index them as a criterion (see Supporting Information), we empirically determined a crystal lifetime of 0.5–1 s under the conditions at hand. This observed crystal lifetime corresponded well to the lifetime of 0.88 s calculated via RADDOSE (Paithankar & Garman, 2010) at an estimated maximum dose rate of 34 MGy s−1.
To collect a complete data set, we employed a data-collection strategy in which a region of interest of 600 × 600 µm was raster scanned with rotation exposures (Fig. 2). 120 parallel helical scans were performed, spaced 5 µm apart. During each helical scan, the goniostat was rotated from Ω = −45° to Ω = +45° (where Ω = 0° corresponds to the orientation of the loop surface perpendicular to the incoming beam) and translated by 600 µm. Taking 240 exposures of 1 s duration over the course of a each helical scan, each individual frame recorded on the detector corresponded to a rotation of 0.375° and a translation of 2.5 µm of the sample. Under these conditions, every crystal within the region of interest received a dose of between 50 and 60 MGy. This high total (integrated over multiple exposures) dose was chosen to collect the highest possible resolution data from each crystal. Using the above strategy, 28 800 detector frames were acquired in a period of 8 h.
2.2. Data processing and structure determination
Adopting the recently established methods for the processing of diffraction patterns from SFX experiments, the CrystFEL software suite (White et al., 2012) was used as a first step to identify and index single-crystal diffraction patterns within the large set of detector frames. Frames containing diffraction patterns that were recorded consecutively during the same helical scan were considered as originating from the same crystals and were assembled into 595 groups containing between two and ten consecutive frames (Fig. 2d and Supporting Fig. S2). These groups were further treated as regular rotation data for re-indexing and integration applying XDS (Kabsch, 2010). Horizontally adjacent groups of diffraction images, potentially containing diffraction patterns from the same crystal, were treated independently. In a standard three-dimensional profile-fitting procedure, both fully and partially recorded reflections were integrated. Processing was successful for 130 groups containing a total of 557 frames.
After iterative merging and scaling, 109 661 reflection intensities in the resolution range from 88 to 3.0 Å were merged into a final data set consisting of 8881 merged reflection intensities with an overall completeness of 99.8%. This final set of reflection intensities included data from 426 diffraction patterns collected from 80 individual TbCatB crystals in 120 groups. The distribution of the size of the groups (Supporting Fig. S2) apparently reflects the variation in the crystal size in TbCatB preparations (see Fig. S1 of Redecke et al., 2013). Most of the data were derived from groups of three to five consecutive frames corresponding to a total rotation range of 1.125–1.875° of one crystal. A small fraction of data originated from groups containing eight to ten frames, while the majority of groups contained three to five consecutive frames.
The quality and internal consistency of the data were judged on the basis of standard 〈I/σ(I)〉 statistics and on the basis of the CC* criteria recently advocated as a single statistically valid guide for deciding the resolution cutoff of the obtained data (McCoy et al., 2007; Karplus & Diederichs, 2012; Evans, 2012). The CC* calculated in resolution shells for the TbCatB data set (Supporting Fig. S1) indicated the presence of statistically significant data to a resolution of 3.0 Å and below.
Following the same strategy as for the previous determination of the T. brucei pro-cathespsin B via SFX (Redecke et al., 2013), initial phases were obtained by with Phaser (McCoy et al., 2007) using the structure of the nonglycosylated and in vitro crystallized TbCatB (PDB entry 3mor ; Koopmann et al., 2012) that lacks the propeptide and the carbohydrate chains as a search model. During stepwise model building and 62 propeptide residues and five carbohydrate residues were manually placed in difference electron-density maps.
The refined TbCatB structure (R factor = 22.3%, Rfree = 26.4%) shares the papain-like fold which is characteristic of cathepsin B enzymes, including the propeptide residues 27–72 and 79–85 without defined electron density in between, as well as a carbohydrate chain consisting of two N-acetylglucosamine (NAG) monomers N-linked to Asn58 (in the propeptide) and another carbohydrate chain consisting of two NAG monomers and one β-mannose (BMA) molecule N-linked to Asn216 of the enzyme domain. Overall, the 3.0 Å resolution electron-density map is well defined by the TbCatB model. No electron density is observed for nine flexible amino-acid side chains mainly located within a loop region spanning residues His195–Asn209 or for ten atoms of the carbohydrate structures. In particular, as for the SFX the expected features of the electron-density map that were not part of the search model are well defined by the propeptide and two carbohydrate chains after manual model building and (Fig. 3).
2.3. Comparison of the structural TbCatB models
For detailed comparison of the T. brucei procathepsin B structure solved in this study at 110 K using synchrotron radiation with that previously obtained at room temperature using the FEL-based SFX technique (PDB entry 4hwy ; Redecke et al., 2013), electron-density maps were generated using the SFX data truncated at 3.0 Å resolution. Applying an identical protocol that omits solvent atoms resulted in an R factor of 17.0% (Rfree = 19.6%). A slight shrinking of the unit-cell parameters of the TbCatB in vivo crystals observed for the synchrotron data set (Table 1) can be attributed to the cryogenic data-collection conditions. At room temperature, unit-cell parameters of a = b = 125.5, c = 54.6 Å were previously obtained by SFX. The superposition of the peptide backbone atoms of both structures revealed a high degree of consistency, resulting in an average r.m.s.d. value of 0.35 ± 0.19 Å, which is comparable to the overall coordinate error of 0.32 Å estimated based on by REFMAC5.5 (Murshudov et al., 2011). No significant structural differences are present, including no major features related to radiation damage (Supporting Fig. S3). Main-chain deviations of more than 0.8 Å are limited to nine residues located at the N-terminus and C-terminus, in flexible loop regions and at positions flanking the disordered part of the propeptide region that results from an increased flexibility of the residues after proteolytic cleavage between Ser78 and Ile79 (Redecke et al., 2013). Even the two carbohydrate chains are clearly defined and largely superimposable between the two models (Figs. 3c–3f). Slight differences were only observed for the second N-acetylglucosamine residue of the propeptide carbohydrate, which represents the most flexible carbohydrate within the model. This is further reflected by the almost identical number of amino-acid side chains/carbohydrate atoms not defined by electron density in both TbCatB structures (nine side chains and ten carbohydrate atoms in this structure versus 11 side chains and eight carbohydrate atoms in the SFX structure).
|
Despite the overall similarity in atomic coordinates, systematic differences were observed in the relative heights of the electron-density peaks at the 12 Cys SG atoms involved in disulfide bridges. Considering refined Debye–Waller factors as an (anticorrelated) measure of the height of electron-density maxima, we note that in the synchrotron structure the average Debye–Waller factor (〈BSG〉 = 57 Å2) is higher than that averaged over all atoms (〈BAll〉 = 41 Å2). In the SFX TbCatB structure refined using an identical protocol at 3.0 Å resolution, 〈BSG〉 (38 Å2) is lower than 〈BAll〉 (45 Å2). This observation is consistent with a significant diffraction contribution from reduced disulfide bonds in the synchrotron data but not in the SFX data.
3. Discussions and conclusions
Particularly for crystals with dimensions in the low micometre range, the determination of macromolecular crystal structures is inherently limited by radiation damage. In most cases, when an X-ray et al., 2011; Siu et al., 2013; Li et al., 2013), despite the difficulties arising from systematic errors when data from multiple crystals are merged. The recent introduction of SFX exploiting X-rays from an FEL to collect single diffraction images to the maximum resolution from large numbers of crystals in the `diffraction-before-destruction' regime has realised an extreme approach to overcoming the radiation-damage problem.
sufficient to measure Bragg reflections to the highest resolution as defined by the degree of crystalline order in a given crystal is used, the crystal will be severely damaged before complete diffraction data can be collected. As a consequence, in practice a compromise is sought balancing the resolution and the completeness of the data to be measured on a single crystal. In recent years, combining data from multiple crystals has enabled the determination of a number of important structures (RasmussenHere, we have demonstrated a strategy for the collection of complete diffraction data close to the diffraction limit from micrometre-sized crystals using synchrotron radiation. This method is based on the serial illumination of subvolumes of a sample consisting of a cryogenically vitrified suspension of microcrystals mounted in a standard nylon loop. During exposure, the sample is rotated, resulting in `classical' rotation frames (Arndt & Wonacott, 1977) for subvolumes presenting crystalline material to the X-ray beam. The X-ray dose received by an exposed subvolume is chosen to fully exploit the crystal lifetime during the exposure time used, a requirement to achieve maximum resolution under these conditions. The combination of microcrystals assuming quasi-random orientations in the suspension and continuous rotation of the loop during each helical scan will present a variety of different crystal orientations to the beam, effectively covering the rotation space and thus providing complete diffraction data in the end.
The crystals used in this study are smaller in volume (∼107 unit cells in a crystal of volume 9 µm3) than those of cypovirus polyhedra (CPV) studied by Metcalf and coworkers (Coulibaly et al., 2007; ∼108 unit cells in 125 µm3). In addition, the TbCatB crystals are embedded in a matrix giving rise to a high scattering background, similar to the situation of GPCR crystals mounted in LCP, where the crystals are commonly of larger volume (∼109 unit cells in 600 µm3; Cherezov et al., 2007). While systematic strategies collecting diffraction data on individually pre-centred crystals were employed for both the CPV and GPCR cases, the serial strategy suggested here has delivered data of comparable quality in terms of signal to noise of the measured diffraction intensities without the need to identify and centre micrometre-sized crystals before data collection (see Supporting Information).
The serial synchrotron-radiation diffraction data allowed us to phase and refine the TbCatB structure using standard technologies. With respect to the data set collected from TbCatB using FEL radiation, the data collected at the synchrotron extend to lower resolution, reflecting the difference between measuring radiation-damage-free `infinitesimally still' data with femtosecond laser pulses and a macroscopic rotation of the crystal over 0.375°, with a similar dose in both cases. Additionally, different levels of background influence the data quality. While a liquid jet of approximately 4 µm diameter delivered the sample for SFX, a 20 µm nylon loop suspending a film of 20 µm thickness was used to mount the crystals in this study. The structural model built against the synchrotron data is consistent with the model obtained with FEL radiation, providing mutual validation.
The present implementation of the serial synchrotron strategy can be improved in many ways. The embedding of the TbCatB crystals with a maximum dimension of approximately 10 µm in a vitrified matrix of approximately 20 µm thickness consisting of buffer and cell debris causes significant background scatter that could be minimized by employing different mounting technologies, such as adaptations of cryo-EM techniques (Nederlof et al., 2013) or graphene supports (Wierman et al., 2013). Reducing the beam size to 1 × 1 µm will reduce the background scattering 20-fold. Increasing the of the X-ray beam will reduce the data-collection time, which is currently several hours for a single loop-mounted drop of crystalline suspension. This can be accomplished using improved X-ray optics and the implementation of wide-bandpass monochromators. Merging data collected on multiple loops would increase the signal-to-noise ratio, while decreasing the rotation increment per X-ray dose unit could increase the resolution to which diffraction data can be recorded closer to the limit posed by the degree of order in the crystal.
The method presented here is conceptually simple and could be implemented at many microfocus synchrotron-radiation beamlines employing existing helical scan schemes. It lends itself to data collection on small crystals in suspension, such as those obtained from in vivo preparations, as it avoids the centring of hardly visible (or invisible) crystals. Parameters can be tuned to maximize accuracy (e.g. applying a larger rotation range per exposed subvolume may allow more accurate integration and scaling) or to maximize resolution (by using a smaller rotation range during the application of the tolerable X-ray dose).
In addition to the promising application as a standalone approach, the combination of serial synchrotron and SFX data collected for a given crystallized protein further offers a new strategy for scaling and phasing of SFX data. In comparison to SFX data collection, serial synchrotron crystallography allows the extraction of accurate diffraction data, albeit to lower resolution owing to the finite rotation range and the onset of radiation damage during the exposure, from a small number of microcrystals by the systematic acquisition of structure-factor amplitudes followed by the application of well defined scaling models modelling a finely controlled experimental process. In contrast, at present, a three orders of magnitude larger number of microcrystals is required for the convergence of Monte Carlo intensity integration when arithmetic means of partially recorded intensities are used without scaling (Kirian et al., 2011). The use of complete and accurate low-resolution data sets obtained using synchrotron radiation for bootstrapping scaling procedures for SFX data could improve the convergence behaviour of these procedures. If diffraction data can be collected on the same system using X-rays from both synchrotron and free-electron laser sources, the combined use of these data therefore has the potential to provide more accurate crystallographic data than those originating from only one of the two methods, ultimately resulting in higher quality macromolecular structures from micrometre-sized crystals.
Supporting information
3D view: 4n4z
PDB reference: cathepsin B, 4n4z
Supporting information. Includes Table S1 as well as Figs. S1, S2 and S3, which provide details of the quality measures of the diffraction data set, an exemplary diffraction pattern, the size distribution in the groups of adjacent indexed frames and details of the structural comparison of the T. brucei procathepsin B structure determined by SFX using free-electron laser radiation and by serial synchrotron crystallography in this study. DOI: 10.1107/S2052252513033939/jt5002sup1.pdf
Footnotes
‡These authors contributed equally to this study.
Acknowledgements
The X-ray diffraction experiments were carried out at beamline P14 of the PETRA III synchrotron source operated by the European Molecular Biology Laboratory (EMBL) at the German Electron Synchrotron (DESY) in Hamburg, Germany in May and June 2013. We gratefully acknowledge discussions with Ilme Schlichting, and we thank Thomas White for help with using CrystFEL. CG acknowledges support from the PIER Graduate School, Helmholtz Association Scholarship. LR, MK, DR and CB thank the German Federal Ministry for Education and Research (BMBF) for funding (grants 01KX0806 and 01KX0807). BPS, DO, LR, MD, CB and HNC acknowledge support from the BMBF in the context of the Röntgen-Angström-Cluster (grant 05K12GU3). Support from the Hamburg Ministry of Science and Research and Joachim Herz Stiftung as part of the Hamburg Initiative for Excellence in Research (LEXI) and the Hamburg School for Structure and Dynamics (SDI) as well as the DFG Cluster of Excellence `Inflammation at Interfaces' (EXC 306) is gratefully acknowledged.
References
Abdullah, M. H., O'Brien, T., Mackey, Z. B., Sajid, M., Grab, D. J. & McKerrow, J. H. (2008). PLoS Negl. Trop. Dis. 2, 298–304. Google Scholar
Arndt, U. M. & Wonacott, A. J. (1977). The Rotation Method in Crystallography. Amsterdam: North Holland. Google Scholar
Barty, A. et al. (2012). Nature Photonics, 6, 35–40. Web of Science CrossRef CAS PubMed Google Scholar
Boutet, S. et al. (2012). Science, 20, 362–364. CrossRef Google Scholar
Brodersen, D. E., Clemons, W. M., Carter, A. P., Wimberly, B. T. & Ramakrishnan, V. (2003). Acta Cryst. D59, 2044–2050. Web of Science CrossRef CAS IUCr Journals Google Scholar
Bryant, C., Kerr, I. D., Debnath, M., Ang, K. K. H., Ratnam, J., Ferreira, R. S., Jaishankar, P., Zhao, D., Arkin, M. R., McKerrow, J. H., Brinen, L. S. & Renslo, A. R. (2009). Bioorg. Med. Chem. Lett. 19, 6218–6221. Web of Science CrossRef PubMed CAS Google Scholar
Chapman, H. N. et al. (2011). Nature (London), 470, 73–77. Web of Science CrossRef CAS PubMed Google Scholar
Cherezov, V., Hanson, M. A., Griffith, M. T., Hilgart, M. C., Sanishvili, R., Nagarajan, V., Stepanov, S., Fischetti, R. F., Kuhn, P. & Stevens, R. C. (2009). J. R. Soc. Interface, 6, S587–S597. Web of Science CrossRef PubMed CAS Google Scholar
Cherezov, V., Rosenbaum, D. M., Hanson, M. A., Rasmussen, S. G. F., Thian, F. S., Kobilka, T. S., Choi, H.-J., Kuhn, P., Weis, W. I., Kobilka, B. K. & Stevens, R. C. (2007). Science, 318, 1258–1265. Web of Science CrossRef PubMed CAS Google Scholar
Coulibaly, F., Chiu, E., Ikeda, K., Gutmann, S., Haebel, P. W., Schulze-Briese, C., Mori, H. & Metcalf, P. (2007). Nature (London), 446, 97–101. Web of Science CrossRef PubMed CAS Google Scholar
Cusack, S., Belrhali, H., Bram, A., Burghammer, M., Perrakis, A. & Riekel, C. (1998). Nature Struct. Biol. 5, 634–637. Web of Science CrossRef CAS PubMed Google Scholar
Evans, G., Axford, D., Waterman, D. & Owen, R. L. (2011). Crystallogr. Rev. 17, 105–142. Web of Science CrossRef Google Scholar
Evans, P. (2012). Science, 336, 986–987. Web of Science CrossRef CAS PubMed Google Scholar
Fairlamb, A. H. (2003). Trends Parasitol. 19, 488–494. Web of Science CrossRef PubMed CAS Google Scholar
Flot, D., Mairs, T., Giraud, T., Guijarro, M., Lesourd, M., Rey, V., van Brussel, D., Morawe, C., Borel, C., Hignette, O., Chavanne, J., Nurizzo, D., McSweeney, S. & Mitchell, E. (2010). J. Synchrotron Rad. 17, 107–118. Web of Science CrossRef CAS IUCr Journals Google Scholar
Garman, E. F. & Schneider, T. R. (1997). J. Appl. Cryst. 30, 211–237. CrossRef Web of Science IUCr Journals Google Scholar
Henderson, R. (1990). Proc. R. Soc. London Ser. B, 241, 6–8. CrossRef CAS Web of Science Google Scholar
Kabsch, W. (2010). Acta Cryst. D66, 125–132. Web of Science CrossRef CAS IUCr Journals Google Scholar
Karplus, P. A. & Diederichs, K. (2012). Science, 25, 1030–1033. Web of Science CrossRef Google Scholar
Kern, J. et al. (2013). Science, 26, 491–495. Web of Science CrossRef Google Scholar
Kirian, R. A., White, T. A., Holton, J. M., Chapman, H. N., Fromme, P., Barty, A., Lomb, L., Aquila, A., Maia, F. R. N. C., Martin, A. V., Fromme, R., Wang, X., Hunter, M. S., Schmidt, K. E. & Spence, J. C. H. (2011). Acta Cryst. A67, 131–140. Web of Science CrossRef CAS IUCr Journals Google Scholar
Koopmann, R. et al. (2012). Nature Methods, 9, 259–262. Web of Science CrossRef CAS PubMed Google Scholar
Li, D., Lyons, J. A., Pye, V. E., Vogeley, L., Aragão, D., Kenyon, C. P., Shah, S. T., Doherty, C., Aherne, M. & Caffrey, M. (2013). Nature (London), 497, 521–524. Web of Science CrossRef CAS PubMed Google Scholar
Liu, Q., Dahmane, T., Zhang, Z., Assur, Z., Brasch, J., Shapiro, L., Mancia, F. & Hendrickson, W. A. (2012). Science, 336, 1033–1037. Web of Science CrossRef CAS PubMed Google Scholar
McCoy, A. J., Grosse-Kunstleve, R. W., Adams, P. D., Winn, M. D., Storoni, L. C. & Read, R. J. (2007). J. Appl. Cryst. 40, 658–674. Web of Science CrossRef CAS IUCr Journals Google Scholar
Murshudov, G. N., Skubák, P., Lebedev, A. A., Pannu, N. S., Steiner, R. A., Nicholls, R. A., Winn, M. D., Long, F. & Vagin, A. A. (2011). Acta Cryst. D67, 355–367. Web of Science CrossRef CAS IUCr Journals Google Scholar
Nederlof, I., Li, Y. W., van Heel, M. & Abrahams, J. P. (2013). Acta Cryst. D69, 852–859. Web of Science CrossRef CAS IUCr Journals Google Scholar
Nelson, R., Sawaya, M. R., Balbirnie, M., Madsen, A. Ø., Riekel, C., Grothe, R. & Eisenberg, D. (2005). Nature (London), 435, 773–778. Web of Science CrossRef PubMed CAS Google Scholar
Neutze, R., Wouts, R., van der Spoel, D., Weckert, E. & Hajdu, J. (2000). Nature (London), 406, 752–757. Web of Science CrossRef PubMed CAS Google Scholar
Paithankar, K. S. & Garman, E. F. (2010). Acta Cryst. D66, 381–388. Web of Science CrossRef CAS IUCr Journals Google Scholar
Rasmussen, S. G. F., Choi, H.-J., Rosenbaum, D. M., Kobilka, T. S., Thian, F. S., Edwards, P. C., Burghammer, M., Ratnala, V. R., Sanishvili, R., Fischetti, R. F., Schertler, G. F., Weis, W. I. & Kobilka, B. K. (2007). Nature (London), 450, 383–387. Web of Science CrossRef PubMed CAS Google Scholar
Rasmussen, S. G. F. et al. (2011). Nature (London), 477, 549–555. Web of Science CrossRef CAS PubMed Google Scholar
Ravelli, R. B. G. & Garman, E. F. (2006). Curr. Opin. Struct. Biol. 16, 624–629. Web of Science CrossRef PubMed CAS Google Scholar
Redecke, L. et al. (2013). Science, 339, 227–230. Web of Science CrossRef CAS PubMed Google Scholar
Riekel, C., Burghammer, M. & Schertler, G. (2005). Curr. Opin. Struct. Biol. 15, 556–562. Web of Science CrossRef PubMed CAS Google Scholar
Siu, F. Y. et al. (2013). Nature (London), 499, 444–449. Web of Science CrossRef CAS PubMed Google Scholar
Smith, J. L., Fischetti, R. F. & Yamamoto, M. (2012). Curr. Opin. Struct. Biol. 22, 602–612. Web of Science CrossRef CAS PubMed Google Scholar
Warren, A. J., Armour, W., Axford, D., Basham, M., Connolley, T., Hall, D. R., Horrell, S., McAuley, K. E., Mykhaylyk, V., Wagner, A. & Evans, G. (2013). Acta Cryst. D69, 1252–1259. Web of Science CrossRef CAS IUCr Journals Google Scholar
White, T. A., Barty, A., Stellato, F., Holton, J. M., Kirian, R. A., Zatsepin, N. A. & Chapman, H. N. (2013). Acta Cryst. D69, 1231–1240. Web of Science CrossRef CAS IUCr Journals Google Scholar
White, T. A., Kirian, R. A., Martin, A. V., Aquila, A., Nass, K., Barty, A. & Chapman, H. N. (2012). J. Appl. Cryst. 45, 335–341. Web of Science CrossRef CAS IUCr Journals Google Scholar
Wierman, J. L., Alden, J. S., Kim, C. U., McEuen, P. L. & Gruner, S. M. (2013). J. Appl. Cryst. 46, 1501–1507. Web of Science CrossRef CAS IUCr Journals Google Scholar
This is an open-access article distributed under the terms of the Creative Commons Attribution (CC-BY) Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.