Figure 7
Effect of adding one valence electron on the structure-factor intensities. Starting with the published PDB structure (1m2a), and using the anomalous scattering factors of Fig. 1 , the structure factors (including bulk solvent) are calculated at 7122 eV for the oxidized and reduced forms of ferredoxin, for Miller indices in the 2.1–2.5 Å resolution range. The plot shows the change upon reduction of the structure-factor intensity |Ftrue(λ = 7122 eV)|2 normalized by the average intensity in that range. The r.m.s. difference is 1.7%, sizable enough to permit the modeling of anomalous scattering factors demonstrated in Table 3 . A number of intensities (173 of the total 8 234) change more than 5%. |
Open
access
access

journal menu![[Figure 7]](tz5101fig7.jpg)
![[link]](../../../../../../logos/arrows/d_arr.gif)