Scherrer grain-size analysis adapted to grazing-incidence scattering with area detectors. Erratum

CROSSMARK_Color_square_no_text.svg

aCornell High Energy Synchrotron Source (CHESS), Cornell University, Ithaca, NY 14853, USA
*Correspondence e-mail: dms79@cornell.edu

(Received 17 November 2012; accepted 6 December 2012; online 17 January 2013)

A poor divergence value in the paper by Smilgies [J. Appl. Cryst. (2009), 42, 1030–1034 ] is corrected.

It was brought to my attention that the vertical beam divergence as determined by energy analysis using an Si(111) Bragg crystal [see the paragraph immediately above equation (9) of Smilgies (2009[Smilgies, D.-M. (2009). J. Appl. Cryst. 42, 1030-1034.])] appeared to be too high, as the experimental vertical spot widths were clearly much smaller than predicted. The problem seems to be related to the standard energy analysis program not taking the high mismatch of multilayer and analyser Bragg angles properly into account (see Kazimirov et al., 2006[Kazimirov, A., Smilgies, D.-M., Shen, Q., Xiao, X., Hao, Q., Fontes, E., Bilderback, D. H., Gruner, S. M., Platonov, Y. & Martynov, V. V. (2006). J. Synchrotron Rad. 13, 204-210.]), and thus overestimating the beam divergence.

I devised a simple independent scheme to measure beam size and divergence using the sample holder and the beam stop for knife-edge scans. If the beam widths at the sample [{\sigma _{\rm S}}] and at the beamstop [{\sigma _{\rm B}}] are measured, then the beam divergence [\sigma ^\prime] can be determined by

[\sigma ^\prime = {{\left( {\sigma _{\rm B}^2 - \sigma _{\rm S}^2}\right)^{1/2} } / {{L_{\rm SB}}}}, \eqno(1)]

assuming Gaussian line shapes, where LSB is the sample-to-beamstop distance. Measurements can be made this way for both the horizontal and the vertical beamwidths and divergences. Fig. 1[link] shows the beam profiles for the GISAXS-I geometry with LSB = 1800 mm.

[Figure 1]
Figure 1
Horizontal and vertical beam profiles at the sample (samx, samz) and the beamstop (bsx, bsz) positions, as obtained by knife-edge scans. Solid lines show the measured profiles and dashed lines the corresponding Gaussian fits.

The vertical beam divergence determined this way is 0.14 mrad, which is significantly lower than the originally reported value of 1.6 mrad obtained from analysing the beamwidths of the Si(111) and Si(333) reflections using existing software. The measured divergence value is yet significantly larger than the intrinsic source divergence of 0.08 mrad, as given by source size and distance (Smilgies, 2009[Smilgies, D.-M. (2009). J. Appl. Cryst. 42, 1030-1034.]). The horizontal beam divergence based on the knife-edge scan is 0.18 mrad, which compares quite well to the reported value of 0.16 mrad, as estimated from source size and source-to-sample distance in the original paper. On the basis of the new vertical divergence value, Fig. 2[link] [to be compared with the right panel of Fig. 3(a) of Smilgies (2009[Smilgies, D.-M. (2009). J. Appl. Cryst. 42, 1030-1034.])] shows the revised resolution estimates for the GISAXS-I geometry, which were most affected by the original poor vertical divergence value.

[Figure 2]
Figure 2
Revised instrumental resolution Δqres for the GISAXS-I geometry (solid line) using the vertical beam divergence as determined by knife-edge scans. Also shown are the partial contributions due to geometric (dashed line), band width (dotted line) and divergence smearing (dash–dotted line).

Acknowledgements

I thank Christine Papadakis and Zhenyu Di at the Technical University of Munich, Germany, for pointing out the problem.

References

First citationKazimirov, A., Smilgies, D.-M., Shen, Q., Xiao, X., Hao, Q., Fontes, E., Bilderback, D. H., Gruner, S. M., Platonov, Y. & Martynov, V. V. (2006). J. Synchrotron Rad. 13, 204–210.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationSmilgies, D.-M. (2009). J. Appl. Cryst. 42, 1030–1034.  Web of Science CrossRef CAS IUCr Journals Google Scholar

© International Union of Crystallography. Prior permission is not required to reproduce short quotations, tables and figures from this article, provided the original authors and source are cited. For more information, click here.

Journal logoJOURNAL OF
APPLIED
CRYSTALLOGRAPHY
ISSN: 1600-5767
Follow J. Appl. Cryst.
Sign up for e-alerts
Follow J. Appl. Cryst. on Twitter
Follow us on facebook
Sign up for RSS feeds