Figure 5
Scattering intensity on absolute scale for samples L5 and L5I as a function of the magnitude of the scattering vector recorded using the SAXS instrument in configuration 4. The (colored) solid lines are model form factors, obtained by least-squares minimization, with smearing included using direct integration of the trapezoidal representation of the beam profile in Fig. 3. The black solid lines are the corresponding ideal intensities with no instrumental resolution effects included. The results have been shifted vertically for clarity as labeled. |