research papers
The fluence–resolution relationship in holographic and coherent diffractive imaging1
aInstitut für Röntgenphysik, Friedrich-Hund-Platz 1, 37077 Göttingen, Germany
*Correspondence e-mail: jhagema@gwdg.de
This work presents a numerical study of the fluence–resolution behaviour for two coherent lensless X-ray imaging techniques. To this end the fluence–resolution relationship of inline near-field holography and far-field coherent diffractive imaging are compared in numerical experiments. To achieve this, the phase reconstruction is carried out using iterative phase-retrieval algorithms on simulated noisy data. Using the incident
on the specimen as the control parameter, the achievable resolution for two example phantoms (cell and bitmap) is studied. The results indicate the superior performance of holography compared with coherent diffractive imaging, for the same fluence and phase-reconstruction procedure.1. Introduction
The simple question `What resolution do I get for the invested photon fluence?' is extremely important for X-ray imaging of radiation-sensitive specimens, such as biological cells and tissues. Structure analysis by diffraction is in general based on et al., 2004; Howells et al., 2009) and have found a steep power law, , relating dose D and resolution d for the case of imaging three-dimensional structures at isotropic resolution. Note that this corresponds to equivalent imaging of two-dimensional slices of a width which is scaled down with d and hence loses contrast. Conversely, for constant width, increasing only the two-dimensional resolution yields for diffraction as for absorption (see e.g. Kirz et al., 1978). Further work has studied the effect of having a certain feature of interest embedded in other structures (matrix) (Schropp & Schroer, 2010), showing that the reconstruction quality in CDI is nearly independent of the surroundings (for a given dose). Before addressing the case of (coherent) diffraction, which became important after the advent of CDI (Miao et al., 1999), earlier work had already compared X-rays, neutrons and electrons as microscopy probes, but had exclusively considered image formation by absorption (Henderson, 1995). This is understandable for the simple reason that X-ray microscopy started in absorption contrast, and was only later extended to (phase contrast) diffractive imaging. A comparison between X-ray microscopy in absorption contrast (water window spectral range) and by numerical simulation was provided by Huang et al. (2009), showing that isolated low-Z materials such as biological cells can be imaged with fewer photons by CDI.
of photons and hence on the Thompson scattering which is much smaller than the for photon absorption. This results in significant energy uptake within the sample and hence causes radiation damage. However, for coherent imaging the dose issue is accentuated, since the information is collected from a single copy of the imaged structure, rather than from a large ensemble of identical constituents over which the dose is distributed. Pioneering studies have addressed this topic within the framework of kinematic scattering theory for far-field coherent diffractive imaging (CDI) (ShenThe literature cited above already illustrates the large range of perspectives which one can take to address the dose and resolution issues, at least in a broad sense. One can compare different probes (X-rays versus other probes), different types of contrast (absorption versus phase contrast), different experimental parameters (notably wavelength) or different types of imaging (e.g. lens-based X-ray microscopy versus lensless diffractive imaging). To this list, we here add the optical regime of a coherent diffractive imaging experiment.
Notably, we want to compare direct reconstruction of lensless coherent imaging data in the near-field and far-field regimes. While the previous studies addressing CDI mentioned above were concerned with far-field diffraction, the numerical simulations used in this work are carried out in the optical near-field regime. Fig. 1 shows a sketch for (a) a near-field inline holographic (NFH) imaging experiment and (b) a CDI experiment. Fig. 1(c) shows both cases tranferred to the parallel-beam setting, as used in the numerical experiments of this study. Our goal is to provide a quantitative comparison between NFH (Kellström, 1932; Gabor, 1948; Howells et al., 1986; Snigirev et al., 1995) and CDI (Miao et al., 1999; Robinson & Miao, 2004). The main difference between this work and the above-mentioned studies is thus the imaging regime in use.
Further, a recent numerical study (Villanueva-Perez et al., 2016) also assessed NFH and CDI and formulated a signal-to-noise criterion. Rather than resolution as in the present study, Villanueva-Perez and co-workers focused on the sensitivity with respect to the phase shift of a given feature and its size at a constant fluence.
The motivations for this study are the experimental indications for the high dose effectiveness of NFH imaging (Bartels et al., 2015; Wilke et al., 2014; Jones et al., 2014). In the work of Bartels et al. (2015) for example, NFH images of bacteria were recorded in the multi-keV regime, where a single bacterium is essentially a pure phase-contrast object. Reconstructions were obtained at a dose which was orders of magnitude smaller than reconstructions of similar resolution obtained previously for the same bacteria by (far-field) ptychography (Wilke et al., 2012). Since experimental work can always be influenced by a number of additional parameters which can, for example, easily render the data inconsistent, a higher or lower dose required for a particular experiment is not conclusive per se. In the light of the limited evidence, we therefore turn to numerical analysis, comparing CDI with NFH for simulated noisy data on the same phantoms. To this end, we first used the (ML) approach introduced by Elser & Eisebitt (2011). Accordingly, a critical fluence μc can be defined, above which the correct phantom (random bitmap) out of a selection of random bitmaps could be identified with a chosen tolerance (error) level ∊ and for given photon shot noise (Jahn et al., 2017). In this way, one can test the information content in noisy two-dimensional diffraction patterns and investigate the dependence of μc on object contrast levels, the accepted error level and the bitmap size. For the experimentally relevant case of weak phase contrast, applicable to most biological samples, NFH required a lower dose than CDI for the optimum propagation distance (Fresnel number) (Jahn et al., 2017). However, apart from small oscillations in μc as a function of distance, as expected based on the contrast transfer function, the results in the near field were almost identical to the far-field results (Jahn et al., 2017). Hence, as far as the encoding of information is concerned, which can be tested by the ML approach, far-field CDI and near-field NFH seem in principle to be roughly equal in dose efficiency. What the ML approach cannot address, however, is whether an unknown object can actually be reconstructed from the noisy data, rather than just comparing likelihoods between the true object and some alternatives (bitmaps with randomly switched bits).
In this work, we fill this gap and actually test the real process of reconstruction from noisy diffraction patterns and not just an ML reconstructability criterion. The main control parameter in this numerical work is again the fluence μ, i.e. the average number of photons per pixel in the plane of the object. Using μ we are able to tune our numerical experiment from the case `barely reconstructing' to `best object reconstruction'. According to this parameter we generate test data of two phantoms: (i) a cell, and (ii) a bitmap object [as done by Jahn et al. (2017)] (see Fig. 2). Following the generation of these noisy diffraction patterns, we run phase-retrieval algorithms on the data and determine the resolution by Fourier ring correlation (FRC) (Harauz & van Heel, 1986; van Heel & Schatz, 2005). Appendix A presents a benchmark of the NFH/CDI propagator without the need of a phase reconstruction. §2 gives details of the data generation and reconstruction scheme. §3 presents the results of the comparison of NFH and CDI. The paper closes in §4 with a summary and outlook.
Of course, implementing both NFH and CDI on the same sample can be experimentally challenging owing to limitations of the setup (coherence, beam size, sampling constraints). These considerations are beyond the scope of this work. Moreover, we consider only coherent scattering (elastic Thomson scattering) and no further interactions of the radiation with matter. Our main focus is the optical regime and the decoding of (phase) information. Thus all simulations are carried out in a dimensionless setting (pixel units and Fresnel number), as detailed below.
2. Numerical setup
Fig. 2 introduces the concept of the numerical study. In essence the two optical setups – far-field CDI and near-field holography (NFH) – are simulated for two different phantoms, namely a phantom of two adhering biological cells [as used by Giewekemeyer et al. (2011)] and a random binary bitmap; see Figs. 2(a) and 2(d), respectively. Both phantoms are pure phase-contrast objects, with phases in the range [−1, 0] rad (cell) and {0, −1} rad (bitmap). Note that binary bitmaps with no correlations between pixels are to some extent amenable to analytical treatments and have been used before, for example by Elser & Eisebitt (2011) (one-dimensional bitmap) and by Jahn et al. (2017) (two-dimensional bitmap).
Both images have a size of 512 × 512 pixels embedded in 1024 × 1024 (Nx × Ny) pixels. This embedding ensures that the simulated numerical aperture (NA) is sufficently large to recover details down to the pixel level. The NA in a vacuum is given by the opening angle α of the detector,
where paraxiality is assumed and is the pixel size. The resolution limit dmin due to the NA is
with wavelength λ. Inserting equation (1) and using the definition of the Fresnel number Fr = yields
Setting thus yields Fr 1/N as a requirement for the NA.
In both CDI and NFH, we assume perfect illumination by a point source or in equivalent geometry by a plane wave (cf. Fig. 1c), such that the exit wave Ψ is given by the phantom Ψ = . The measurements for NFH were generated by applying the Fresnel propagator given by
where kx = 2nx/Nx and ky = 2ny/Ny are spatial frequencies in Fourier space with nx,y and a Fresnel number of Fr = 10−3. The measurements for CDI were generated by discrete Fourier transformation of the corresponding exit wave. Next, the generated intensity patterns (far-field and near-field, respectively) in the detection plane were subjected to Poissonian noise using the routine imnoise (Matlab Inc.), with the average μ (in photons per pixel) in the object plane as the only parameter. Figs. 2(b) and 2(c), and 2(e) and 2(f), show in each case the ideal noiseless simulated data or `measurements' (left-hand side) and a random realization of the noisy measurement for a fluence of 200 photons per pixel (in the exit plane; right-hand side). The example NFH and CDI measurements simulated for the phantom nicely illustrate the completely different nature of the signals. In the case of NFH, the signal varies around one (normalized primary beam) by self-interference of the primary beam with the diffracted beam behind the object, and is best represented on a linear scale with a narrow range of intensities. In contrast, the CDI data cover many orders of magnitude from the central pixel to the edge of the detector, where most pixels have zero photon counts. Note that in this idealized simulation we take the full diffraction pattern into account, i.e. the numerical aperture is sufficiently large, and we assume that the detector does not need any kind of beamstop, which would result in a loss of information.
Thus, in summary, the noisy measurements were generated using the following recipe:
(i) Propagate the field Ψ from the sample plane to the measurement plane (detection plane) using the respective propagator ( or ).
(ii) Calculate the intensities of the field, yielding the measurement M = . Normalize M so that M′ = .
(iii) Multiply M′ by μNxNy and use the result as input for a Poisson random number generator. This yields the noisy measurement used in the phase reconstruction.
The reconstructions from the noisy data were obtained using the relaxed averaged alternating reflections (RAAR) algorithm (Luke, 2005). The iterations are given by
where = − Ψ denotes a (mirror) reflection by a given constraint set and n the iteration index. The parameter βn controls the relaxation. It follows the function
where β0 denotes the starting value, βmax the final value of βn and βs the iteration number when the relaxation is switched. This relaxation strategy follows equation (37) of Luke (2005). The parameters were set for β0 = 0.99, βmax = 0.75 and βs = 150 iterations for all reconstructions. The projection on the measurements PM is the standard magnitude projector
where is either or for far-field or near-field propagation, respectively. In equation (74) of Luke et al. (2002) an alternative version of PM is given which should handle numerical inconsistencies such as noise. In our case, experiments using this version did not show any improvement in the resolution.
The operator PS is used to enforce the support S, which is assumed to be perfectly known, and the pure phase constraint in the object plane, i.e.
Details on the implementation can be found in the supporting information.
3. Results
Before addressing the fluence–resolution relationship, we present an example reconstruction to illustrate the steps that are necessary to obtain the reconstruction data, which are then analysed by massively parallel batch processing. Fig. 3 shows an example reconstruction for μ = 200 photons per pixel for both phantoms (Figs. 3a and 3c), using the noisy measurements shown in Figs. 2(b), 2(c), 2(e) and 2(f).
The left- and right-hand sides of Figs. 3(a) and 3(c) show the phases of the NFH and CDI reconstructions, respectively, again for both (a) the cell and (c) the bitmap phantom. The reconstructions are based on the same set of parameters and constraints as far as possible. There are two differences: (i) the propagation operator ( or ) and (ii) the starting guess. For the holographic reconstruction, an array uniformly initialized with amplitude 1 and phase 0 was used, whereas the CDI reconstruction used an initialization consisting of uniform amplitude 1 but randomly chosen phases from the range , both in the object plane. Changing the initial guess typically had only a small effect on the results. Inspection of the holographic reconstructions in Figs. 3(a) and 3(c) shows some high-frequency noise in the background of the reconstructions, but the fine structures of the cells (small black dots) are still clearly visible and the edges of the bitmap are still sharp. The CDI reconstruction of the cell clearly shows a loss of detail, but the background is less noisy. For the bitmap we see washed out edges and some structured background which matches the length scale of the bits. Note that the reconstructions show the object after the measurement projection (before the support is enforced).
For a quantitative comparison of the resolution we used FRC. To this end, the phases of the reconstructions were correlated with the corresponding phantom phases (Fig. 2), and the decrease in correlation was plotted as a function of spatial frequency. The resolution is determined by the first intersection of fr, the resolved spatial frequency, with the so-called half-bit threshold, indicating the degree of correlation at which sufficient signal has been acquired (van Heel & Schatz, 2005). The results are shown in Fig. 3 for (b) the cell and (d) the bitmap. The FRC curve in Fig. 3(b) for the CDI reconstruction decays much faster (fr = 0.17) than for the NFH reconstruction (fr = 0.48), in agreement with visual inspection. Interestingly, the FRC curves for the bitmap phantom show an oscillatory behaviour, but again the CDI curve decays faster (fr = 0.2) than the NFH curve (fr = 0.4), i.e. it shows lower resolution.
Next, we turn to the fluence–resolution relationships, which were computed by performing the automated reconstruction and FRC analysis for measurements of systematically varied fluence μ. For each μ covering the range from 1 to 20 000 photons per pixel (phantom plane), 30 realizations were generated and reconstructed, each with the same parameters. Fig. 4 shows the results. Comparing the results for (a) the cell and (b) the bitmap, we notice that in both cases the NFH reconstruction reaches the maximum achievable resolution at a significantly smaller fluence. Note that the spatial frequency of 0.5 periods per pixel corresponds to the maximum (half-period) resolution of a pixel. However, reaching the full resolution does not necessarily equate to having a perfect reconstruction. For example, the l2-norm of the difference image (reconstruction phantom) can be non-zero, while the FRC has already saturated. Comparing Figs. 4(a) and 4(b), we notice that NFH reaches a maximum resolution for both objects at the same fluence of around 300 photons per pixel. At the same time, the error bars of the bitmap results are larger than those for the cell. In contrast, CDI needs 11 000 photons to reach full resolution for the cell, and 3000 photons for the bitmap. Furthermore, we analysed the error of the reconstruction by the l2-norm [see Fig. 4(c) for the cell and Fig. 4(d) for the bitmap]. To this end, the l2-norm Δ of the phase difference,
was computed for all pixels within the support. The Δ curves in Figs. 4(c) and 4(d) are normalized by the number of pixels in S and show an unexpected behaviour. At low fluences the error in CDI is smaller than that in NFH, but there is a crossover at μ = 140 (cell) and μ = 4 (bitmap), where NFH becomes superior in terms of Δ. On closer inspection of the reconstruction result, however, it becomes clear that the smaller Δ at low fluence is misleading. CDI yields an unstructured reconstruction with no representation of structural details [see insets in Fig. 4(c)]. The reconstructions are much worse than the NFH results for the same fluence, but exhibit a smaller Δ by way of averaging the signal deviations. We must conclude that Δ is not a well suited error metric at low fluence.
Thus, it becomes clear that in all cases tested, NFH yields superior results to CDI. Note that the absolute Δ values also depend on the number of iterations (200 in both cases). Running the algorithm for more iterations, e.g. 800, led to further reductions in Δ of about 30% in the case of NFH and 10% for CDI (cell phantom). Furthermore, the introduction of additional constraints can of course also change the error value. For example, using the prior knowledge that the binary bitmap must have discrete phase values 0, −1 suggests the use of a thresholding constraint (binary value projector)
Fig. 5 shows the results using this projection in addition to the support and pure phase constraint for the bitmap phantom. For these results we used a bitmap with 1:1 pixel correspondence of bitmap to object plane pixel. Thus the entire object has a size of only 10 × 10 pixels. Fig. 5(a) shows the Δ/N error [as in Fig. 4(d), but now after the threshold constraint], corresponding to the fraction of wrong pixels. Here we see the expected behaviour in that NFH reconstructs better at low fluence than CDI. For comparison, we also plot the theoretical function μc(Δ/N) (solid line) based on the ML analysis, describing the critical fluence to identify the correct bitmap from the noisy diffraction pattern (CDI) out of a set of neighbouring bitmaps (Jahn et al., 2017). Fig. 5(b) shows the fraction of successful reconstructions, i.e. the fraction of successful reconstructions (Δ = 0) from an ensemble of 100 runs as a function of fluence. Comparing these results, we see that CDI reconstructions require substantially more at any error level. Furthermore, the functional form of the curve is smoother for NFH, while the transition from non-reconstructible to reconstructible is extremely sharp for CDI, similar to a phase transition.
4. Summary and outlook
In this work, we have investigated the fluence efficiency of variants of lensless X-ray imaging techniques, notably coherent diffractive imaging operating in the optical far field, and inline holography operating in the optical near field. Despite the entirely different nature of the signals and the modes of image formation, which can be classified as heterodyne and homodyne, i.e. with and without adding a reference wave, it is commonly assumed that the information contents in the diffraction pattern for a given should be equal. The analytical work and simulations of Jahn et al. (2017) have already pointed out that this can never be exactly true, since the oscillatory nature of the contrast transfer functions in NFH results in a dependence on the Fresnel number. Therefore, absorption and phase contrast have to be distinguished, and also the regime of weak or strong contrast. However, the approach of Jahn et al. (2017) addresses the information content of the noisy pattern and not the reconstruction quality, which can actually be obtained by standard methods of iterative algorithms. As we have shown here, the latter case is characterized by substantial differences between NFH and CDI. In other words, while the information content may be similar, the ability of the algorithms to decode the diffraction pattern deviates significantly. These conclusions have been substantiated both by the error metric of the l2-norm Δ and by Fourier ring correlation. For example, Fig. 5(b) shows that both NFH and CDI reach full reconstructability within one decade of but for CDI the fluence curve was shifted up by two decades. Furthermore, both NFH and CDI reconstructions required substantially higher fluence, as predicted by the ML approach. In conclusion, our findings point to an important advantage of NFH, in addition to its large tolerance for partial coherence, its compatibility with extended specimens and its flexibility in reconstruction constraints, e.g. the pure phase constraint is often sufficient to reconstruct at least a coarse image of the object. Some of these advantages may also apply to ptychography, for example when mixed states are taken into account (Thibault & Menzel, 2013). It goes without saying that these conclusions await further validation by other reconstruction codes, as well as by careful experimental testing. If the evidence is substantiated, more imaging experiments of radiation-sensitive specimens such as biological objects should be carried out in the holographic regime, for which dedicated synchrotron beamlines are now available.
APPENDIX A
Direct back propagation
Following the suggestion of a reviewer, we have investigated to what extent the results obtained in this work depend on the reconstruction (which always requires a specific choice of constraints and reconstruction algorithm) and we have also performed simulations in the simplest possible setting, based on direct back propagation. To this end, the exit wave in the sample plane was first propagated to the detector plane to generate diffraction data as before. Then Poisson noise was added to the intensity (amplitude), again as before. Finally, the wave with noisy amplitude and ideal phase was propagated back to the sample plane. The idea of this procedure was to eliminate the role of the reconstruction algorithm used. The ideal complex-valued data M is given as
where is either or for NFH or CDI, respectively, and Ψ is the exit wave in the sample plane. Using the noising procedure from the main text gives the noisy intensities , which are then combined with the phases to yield ,
i.e. the noisy complex amplitudes. The are generated for different fluences μ ranging from 1 to 1000 photons per pixel and are then used as input for the inverse propagator for NFH and CDI. Fig. 6 shows the results of this numerical experiment. We compare three different phantoms: (i) the cell phantom, (ii) the bitmap phantom with one bitmap pixel represented by 10 × 10 image pixels (oversampled information) and (iii) one bitmap pixel represented by one image pixel.
NFH again shows superior performance for the cell phantom and the 1 × 1 bitmap, but not to the same degree as before. Furthermore, CDI reaches a higher resolution at a lower dose for the 10 × 10 bitmap. This can be understood based on the fact that in this numerical experiment the phases are given, and the main advantage of NFH in encoding phase information by interference does not play a role.
In addition, we have varied the contrast of the object. The results illustrate the same trend for both CDI and NFH: as the phase shift of the phantom increases, fewer photons are needed to reach maximum resolution, as expected. This is in agreement with the findings of Villanueva-Perez et al. (2016).
Supporting information
Simulation details and implementation. DOI: https://doi.org/10.1107/S1600576717003065/vc5008sup1.pdf
Scripts. DOI: https://doi.org/10.1107/S1600576717003065/vc5008sup2.zip
Footnotes
1This article will form part of a virtual special issue of the journal, presenting some highlights of the 13th Biennial Conference on High-Resolution X-ray Diffraction and Imaging (XTOP2016).
Acknowledgements
We thank Jan-David Nicolas, Malte Vassholz and Simon Maretzke for fruitful discussions, and Chris Jacobsen for inspiring comments. Funding by BMBF project No. 05K16MGB WeKokUbS, `Wellenfront- und Kohrenzkontrolle an Undulatorbasierten Strahlquellen zur Optimierung bildgebender Verfahren', and DFG grant SFB755, `Nanoscale Photonic Imaging', is gratefully acknowledged.
References
Bartels, M., Krenkel, M., Haber, J., Wilke, R. N. & Salditt, T. (2015). Phys. Rev. Lett. 114, 048103. Web of Science CrossRef PubMed Google Scholar
Elser, V. & Eisebitt, S. (2011). New J. Phys. 13, 023001. Web of Science CrossRef Google Scholar
Gabor, D. (1948). Nature, 161, 777–778. CrossRef PubMed CAS Web of Science Google Scholar
Giewekemeyer, K., Krüger, S. P., Kalbfleisch, S., Bartels, M., Beta, C. & Salditt, T. (2011). Phys. Rev. A, 83, 023804. Web of Science CrossRef Google Scholar
Harauz, G. & van Heel, M. (1986). Proceedings of the IEEE Computer Vision and Pattern Recognition Conference, Vol. 73, pp. 146–156. Google Scholar
Heel, M. van & Schatz, M. (2005). J. Struct. Biol. 151, 250–262. Web of Science PubMed Google Scholar
Henderson, R. (1995). Q. Rev. Biophys. 28, 171–193. CrossRef CAS PubMed Web of Science Google Scholar
Howells, M., Beetz, T., Chapman, H., Cui, C., Holton, J., Jacobsen, C., Kirz, J., Lima, E., Marchesini, S., Miao, H., Sayre, D., Shapiro, D., Spence, J. & Starodub, D. (2009). J. Electron Spectrosc. Relat. Phenom. 170, 4–12. Web of Science CrossRef CAS Google Scholar
Howells, M. R., Kirz, J. & Iarocci, M. A. (1986). J. Opt. Soc. Am. A, 3, 2171–2178. CrossRef CAS Google Scholar
Huang, X., Miao, H., Steinbrener, J., Nelson, J., Shapiro, D., Stewart, A., Turner, J. & Jacobsen, C. (2009). Opt. Express, 17, 13541–13553. Web of Science CrossRef PubMed Google Scholar
Jahn, T., Wilke, R. N., Chushkin, Y. & Salditt, T. (2017). Acta Cryst. A73, 19–29. CrossRef IUCr Journals Google Scholar
Jones, M. W., Dearnley, M. K., van Riessen, G. A., Abbey, B., Putkunz, C. T., Junker, M. D., Vine, D. J., McNulty, I., Nugent, K. A., Peele, A. G. & Tilley, L. (2014). Ultramicroscopy, 143, 88–92. Web of Science CrossRef CAS PubMed Google Scholar
Kellström, G. (1932). PhD thesis, Nova Acta Regiae Societatis Scientiarum Upsaliensis. Google Scholar
Kirz, J., Sayre, D. & Dilger, J. (1978). Ann. N. Y. Acad. Sci. 306, 291–305. CrossRef CAS Google Scholar
Luke, D. R. (2005). Inverse Probl. 21, 37–50. Web of Science CrossRef Google Scholar
Luke, D. R., Burke, J. V. & Lyon, R. G. (2002). SIAM Rev. 44, 169–224. Web of Science CrossRef Google Scholar
Miao, J., Charalambous, P., Kirz, J. & Sayre, D. (1999). Nature, 400, 342–344. Web of Science CrossRef CAS Google Scholar
Paganin, D. M. (2006). Coherent X-ray Optics. New York: Oxford University Press. Google Scholar
Robinson, I. & Miao, J. (2004). MRS Bull. 29, 177–181. CrossRef CAS Google Scholar
Salditt, T., Osterhoff, M., Krenkel, M., Wilke, R. N., Priebe, M., Bartels, M., Kalbfleisch, S. & Sprung, M. (2015). J. Synchrotron Rad. 22, 867–878. Web of Science CrossRef CAS IUCr Journals Google Scholar
Schropp, A. & Schroer, C. G. (2010). New J. Phys. 12, 035016. Web of Science CrossRef Google Scholar
Shen, Q., Bazarov, I. & Thibault, P. (2004). J. Synchrotron Rad. 11, 432–438. Web of Science CrossRef IUCr Journals Google Scholar
Snigirev, A., Snigireva, I., Kohn, V., Kuznetsov, S. & Schelokov, I. (1995). Rev. Sci. Instrum. 66, 5486–5492. CrossRef CAS Web of Science Google Scholar
Thibault, P. & Menzel, A. (2013). Nature, 494, 68–71. Web of Science CrossRef CAS PubMed Google Scholar
Villanueva-Perez, P., Pedrini, B., Mokso, R., Guizar-Sicairos, M., Arcadu, F. & Stampanoni, M. (2016). Opt. Express, 24, 3189–3201. PubMed Google Scholar
Wilke, R. N., Priebe, M., Bartels, M., Giewekemeyer, K., Diaz, A., Karvinen, P. & Salditt, T. (2012). Opt. Express, 20, 19232–19254. Web of Science CrossRef CAS PubMed Google Scholar
Wilke, R. N., Wallentin, J., Osterhoff, M., Pennicard, D., Zozulya, A., Sprung, M. & Salditt, T. (2014). Acta Cryst. A70, 552–562. Web of Science CrossRef IUCr Journals Google Scholar
This is an open-access article distributed under the terms of the Creative Commons Attribution (CC-BY) Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.