Estimation of the average errors in orientation and CoM position of subgrains as a function of misalignment of the instrument. The errors are plotted as a function of misalignment in (a) detector tilt z and (b) detector z centre. The errors represent the difference between grains in the misaligned case and matching grains for the ideal alignment. The presented CoM errors are calculated after subtracting the rigid-body translations due to the considered misalignments.