research papers
An inclined detector geometry for improved X-ray total scattering measurements
aDepartment of Physics, University of Guelph, Guelph, Ontario, Canada, and bCanadian Light Source, Saskatoon, Saskatchewan, Canada
*Correspondence e-mail: skycia@uoguelph.ca
X-ray total scattering measurements are implemented using a digital flat-panel area detector in an inclined geometry and compared with the traditional geometry. The traditional geometry is defined here by the incident X-ray beam impinging on and normal to the center-most pixel of a detector. The inclined geometry is defined here by a detector at a pitch angle α, set to 15° in this case, bisected by the vertical The detector is positioned such that the incident X-ray beam strikes the pixels along the bottom edge and 90° scattered X-rays impinge on the pixels along the top edge. The geometric attributes of the inclined geometry translate into multiple benefits, such as an extension of the measurable scattering range to 90°, a 47% increase in the accessible magnitudes of the reciprocal-space vector Q and a leveling of the in the measured total scattering pattern. As a result, a sixfold improvement in signal-to-noise ratios is observed at higher scattering angles, enabling up to a 36-fold reduction in acquisition time. Additionally, the extent of applied modification functions is reduced, decreasing the magnitude of termination ripples and improving the real-space resolution of the pair distribution function G(r). Taken all together, these factors indicate that the inclined geometry produces higher quality data than the traditional geometry, usable for simultaneous and total scattering studies.
Keywords: synchrotron radiation; X-ray diffraction; total scattering; instrumentation; area detectors.
1. Introduction
The reduced pair G(r) and radial R(r) distribution functions have been instrumental in characterizing condensed matter systems by means of X-ray total scattering experiments (Billinge, 2019; Petkov, 2012). The distribution function methodology has been successful in analyzing large subsets of materials such as melts and amorphous and polycrystalline systems (Laaziri et al., 1999a,b; Petkov et al., 1999; Skinner et al., 2013; Tomberli et al., 2015). In order to produce a high real-space resolution G(r) by means of a Fourier transform, the measurement of the X-ray total scattering pattern's coherent component should be taken out to large values in Q, the magnitude of the reciprocal-space vector. Here, Q is defined in equation (1), where λ is the wavelength of the incident X-ray beam and 2θ is the scattering angle,
Increasingly, effort is being dedicated to reducing image acquisition times without compromising real-space resolution in G(r). The push is fueled by challenging experiments in thin films, systems of low and time-resolved in situ measurements (Jensen et al., 2015; Li et al., 2021; Wiaderek et al., 2013; Chapman et al., 2015).
In order to achieve the desired counting statistics and fast acquisition times, two-dimensional digital flat-panel area detectors are commonly used when performing X-ray total scattering measurements (Marlton et al., 2019). Traditional detector geometries have the incident X-ray beam normal to the detector surface, as shown in Fig. 1(a). In these geometries, the experimenter must find a balance between the X-ray beam energy, the sample-to-detector distance (SDD) and the quality of the acquired measurement. Increasing the X-ray energy extends the maximum reachable value of Q, while conversely decreasing the measured scattered intensity by reducing the incident X-ray beam and reducing the interaction cross sections for both the sample and detector (Waseda, 2003; Hubbell & Seltzer, 1995). Decreasing the SDD extends the measurable scattering range but shifts the of the measured X-ray total scattering pattern by means of the solid angle, which becomes larger at lower scattering angles and smaller at higher scattering angles. This results in possible detector saturation at lower scattering angles and poor signal-to-noise ratios at higher scattering angles. In addition, decreasing the SDD in the traditional geometry leads to compromises in the quality of the measured lower scattering angle region, as the resolution of Q is decreased and the beam-stop shadow blocks more of the lower angle scattering pattern.
In order to address many of these issues, we explore here the performance of an inclined detector geometry. The inclined geometry is defined by a detector at a pitch angle α, set to 15° in this case, bisected by the vertical The detector is positioned such that the incident X-ray beam strikes the pixels along the bottom edge and 90° scattered X-rays impinge on the pixels along the top edge, as shown in Fig. 1(b). Lower pitch angles maximize the favorable attributes of the geometry. However, collisions between the detector and the sample rotation stage prevent lower pitch angles from being obtained. As a result, the minimum physically allowable pitch angle angle of 15° was chosen for this example. A series of synchrotron-based measurements were conducted in each configuration which prove the advantages of the inclined geometry over the traditional geometry. This was done by measuring a powdered silicon laboratory standard to calibrate the detector and a powdered nickel sample to serve as an unknown for comparison between the geometries. It was shown that, with careful two-dimensional image analysis, parallax corrections and addressing of intrinsic detector curvature, the inclined geometry is superior to the traditional geometry, producing higher quality data usable for simultaneous and total scattering studies.
2. Methods
X-ray total scattering patterns were collected for a powdered laboratory standard of silicon (NIST SRM 640f) and for powdered nickel (Sigma–Aldrich). All samples were measured at a temperature of 100 K for a summed total of 512 1 s exposures. The measurements were conducted on the high energy wiggler X-ray diffraction and scattering beamline in the Brockhouse sector of the Canadian Light Source (Gomez et al., 2018). All measurements were performed with 60 keV X-rays. A PerkinElmer XRD 1621 CN3-EHS (PE-1621) flat-panel area detector containing 2048 × 2048 pixels with a pitch of 200 µm was used to measure two-dimensional X-ray total scattering patterns.
The flat-panel area detector was placed in two different geometries. In the first configuration, represented in Fig. 1(a), the detector was placed in the traditional geometry, defined with the incident X-ray beam impinging on and normal to the center-most pixel of the detector. In the second configuration represented in Fig. 1(b), the detector was placed in an inclined geometry, defined with the detector at a pitch angle α of 15°, bisected by the vertical The detector is positioned such that the incident X-ray beam strikes the pixels along the bottom edge and 90° scattered X-rays impinge on the pixels along the top edge. To avoid the loss of coherent intensity at higher scattering angles as a result of 98% horizontally polarized X-rays, the detector is strategically positioned to be bisected by the vertical scattering plane.
The PE-1621 is a commonly used detector on many beamlines for its large active area and relatively low cost. However, the PE-1621 introduces significant artifacts and distortions into the collected X-ray scattering patterns as a result of its electronic and physical design. Due to the electronic design, the detector exhibits a characteristic intensity distortion in the measured X-ray total scattering pattern. The intensity distortion is produced from a variation in read-out gain between pixels in the detector (Skinner et al., 2012). The gain of each pixel is calculated through a flood field measurement, producing a correction for this PE-1621 detector which reduces the intensity distortion to an unobservable level (Burns, 2022). The physical design of the detector introduces parallax and topography distortions in the measured position of an X-ray beam incident on the flat-panel X-ray detector (Weiß et al., 2012; Lüthi et al., 2019; Wu et al., 2002; Zaleski et al., 1998). Using only the measured X-ray total scattering pattern of a powdered silicon laboratory standard at 100 K, these distortions were quantified and corrected. For detailed information on the parallax and topography distortion corrections in this work, see the supporting information. The corrections calculated from the silicon were applied to all measured X-ray total scattering patterns and in all geometries used in these experiments.
The selected scattering areas used for generation of the distribution functions cover a different range in each geometry, as displayed in Fig. 2. In the traditional geometry, the selected scattering area covers a full circular 360° in the azimuthal angle χ, where the beam-stop shadow is masked and removed from any integrations. The maximum Q reached is 29 Å−1. In the inclined geometry, the selected scattering area covers a 125° wedge in χ bisected by the vertical and a maximum Q of 42.5 Å−1 is reached. In the inclined geometry, the choice of a scattering area covering a smaller range in χ about the bisecting vertical reduces peak broadening due to the scattering footprint of a line beam on the sample compared with the full circular χ range used in the traditional geometry (Sulyanov et al., 1994; He, 2018). The same χ range could be chosen for the traditional geometry to make this effect equivalent in both cases but this would result in reduced signal-to-noise ratios.
3. Results and discussion
Implementation of the inclined geometry extends the accessible maximum scattering range to 90° and thus results in a 47% larger Q range. Additionally, the of the measured X-ray total scattering pattern is leveled. The measured total scattering intensity of the lower scattering angle region is reduced, while the intensity of the higher scattering angle region is increased. Reducing the counts per pixel in the lower scattering angle region allows for the incident beam to be increased when measuring samples which saturate the detector in these lower regions. In this way, the scattered intensity of the entire pattern is increased, which results in an improved signal-to-noise ratio. In each experiment, the incident beam is adjusted such that the most intense pixels of the first significant Debye–Scherrer ring are measured at 45 000 counts. This is below the 65 000 count saturation limit of the detector. The inclined detector geometry is compared with the traditional by integrating along χ to a one-dimensional plot in Q, within the scattering areas as defined in Fig. 2. Interpolation and integration from the collected two-dimensional scattering patterns to one-dimensional patterns in Q are performed using custom Python code which allows for simultaneous correction of the parallax and topography distortions. For detailed information on the parallax and topography distortion corrections used in this work, see the supporting information. Fig. 3 shows the integrated mean total scattering intensities after background subtractions μTotal for powdered nickel in both geometries.
The solid angle subtended by a pixel ΔΩ(γ) is quantified using equation (2), where Px,y is the pixel pitch, γ is the incidence angle and SDD is the sample-to-detector distance (Grillo, 2008),
The most significant geometric effect introduced by implementation of the inclined detector geometry is the change to the distribution of ΔΩ(γ) across the detector surface. The point of normal incidence (PONI) is the point of maximum ΔΩ(γ) for flat detectors of uniform pixel size. The PONI in the traditional geometry is positioned at a scattering angle 2θ = 0°. In the inclined geometry, the PONI is repositioned away from the beam center to the scattering angle 2θ = 90° − α along the vertical Shifting the PONI has the effect of decreasing ΔΩ(γ) in the lower scattering angle region, reducing the measured counts per pixel, and increasing ΔΩ(γ) in the higher scattering angle region, resulting in more measured counts per pixel. Figs. 4(a) and 4(b) show ΔΩ(γ) in the traditional and inclined geometries, respectively.
The solid angle indicates both the change in pixel intensity and the change in angular resolution of the pixel. However, when considering powder diffraction, the resolution between the pixels along the azimuthal direction is not typically relevant, as total scattering measurements are integrated along this dimension. In order to investigate only the resolution along the direction parallel to the scattering angle, the magnitude of the reciprocal-space vector gradient |∇Q(xP, yP)| is calculated using equation (3),
Here, xP and yP are the two pixel coordinates defined along the detector's surface. |∇Q(xP, yP)| quantifies the amount of Q covered per pixel along the direction parallel to the scattering angle after multiplying by the pixel size, which is simply one in pixel space. The smaller the value of |∇Q(xP, yP)|, the more pixels will fit within the width of a measured Debye–Scherrer ring. Comparing the inclined geometry and the traditional as shown in Figs. 4(c) and 4(d), |∇Q(xP, yP)| is reduced in the lower scattering angle region but increased in the higher scattering angle region.
The geometric effect of changing |∇Q(xP, yP)| is highlighted in Fig. 5, by comparing the centerline traces as defined in Fig. 2, for the first significant Debye–Scherrer rings in both the traditional and inclined geometries. The centerline trace is plotted as intensity per pixel and in the structure function S(Q) as defined in equation (4), where I(Q) is the normalized isolated coherent intensity, N is the number of atoms illuminated by the incident X-ray beam and f(Q) is the coherent form factor for a monoatomic system (Peterson et al., 2021),
The benefits of a reduced |∇Q(xP, yP)| at lower scattering angles are also highlighted in Fig. 5. In Fig. 5(a), the line shape of the measured Debye–Scherrer ring in the inclined geometry appears broader in pixel space relative to the traditional geometry. However, it is important to understand that this broadening in pixel space does not translate to broadening in Q space, as shown in Fig. 5(b) for the first significant peak in S(Q). Here, the Q-space resolution is instead improved, as shown by a sharper and more highly sampled peak. The opposite is true for the traditional geometry. The inclined geometry case is preferred for X-ray total scattering measurements and analysis. Specifically, high Q-space resolution for intense low scattering angle Debye–Scherrer rings improves the real-space resolution envelope for generated G(r) functions as higher frequency signals may be captured. Additionally, increased sampling of the lower scattering angle region benefits traditional analysis methods for polycrystalline materials such as which focuses on these lower scattering angle regions.
The theoretical incoherent and photoelectric intensities were subtracted out of μTotal leaving only the integrated mean unnormalized coherent intensities μCoherent (Burns, 2022). We observed an increased μCoherent in the high scattering angle regions for the inclined geometry compared with the traditional geometry. A higher coherent intensity largely contributes to improved signal-to-noise ratios for the inclined geometry over the traditional geometry. A comparison of μCoherent is plotted in Fig. 6 for powdered nickel.
A comparison of the integrated mean unnormalized coherent intensity's signal-to-noise ratio SNRCoherent in the inclined and traditional geometries is calculated using experimentally determined Poisson-like statistics for the PE-1621 detector. The standard deviation of an ensemble of binned snapshots and pixels for total scattering is given by equation (5),
Here, NCounts is the number of measured counts, NSnapshots is the number of integrated snapshots, NPixels is the number of integrated pixels in χ, σBackground is the standard deviation of the background after the subtraction of all of its signals and β is the conversion factor from visible photons to measured counts (Michel et al., 2006; Hughes & Hase, 2010; Lyons, 2004). For detailed information on how the experimentally determined Poisson-like statistics were calculated for the PE-1621 detector, see the supporting information. SNRCoherent can then be calculated from the ratio of μCoherent and as defined in equation (6) (Lyons, 2004),
μCoherent is calculated by subtraction of the theoretically calculated incoherent and photoelectric signals from μTotal. As a result, the standard deviation of μCoherent remains as as it does not contain statistical noise and is assumed to be unchanged by the subtraction of the theoretical signal.
Ratio comparisons of μCoherent and between the inclined and traditional geometries are shown in Fig. 7(a), calculated from the theoretical nickel gas intensities defined in Figs. 3 and 6. Experimentally determined values of β = 2.83 and σBackground = 53 are used to calculate . Additionally, the ratio comparison of SNRCoherent is shown in Fig. 7(b), from which it is observed that the SNRCoherent of the inclined geometry is reduced in the lower scattering angle regions and increased in the higher scattering angle regions compared with that of the traditional geometry. This is an ideal compromise between the two geometries, as the signal-to-noise ratios are typically higher in the low scattering angle regions and lower in the high scattering angle regions. Effectively a trade-off in signal-to-noise ratio has been made, reducing SNRcoherent where it is not needed and increasing it where it is needed most. The ratio comparison of SNRCoherent indicates that a measurement of the same quality up to 29 Å−1 could be acquired measuring 36 times fewer snapshots, greatly reducing the acquisition time.
The inclined geometry provides access to a broader range in Q. The additionally probed helps to preserve the real-space resolution by allowing the reduced total scattering structure function F(Q) defined in equation (7) to damp out naturally due to the Debye–Waller factor,
Modified Lorch functions with different Lorch power factors ζ were multiplied by the F(Q) functions in each geometry (Lorch, 1969). The modified F(Q) functions produce radial distribution functions R(r) containing minimal termination ripples of equivalent proportionality. The generated R(r) functions are compared in Fig. 8. The traditional geometry required a ζ value of 0.6, while the inclined geometry only required a ζ value of 0.25 in the case of the powdered nickel sample. For detailed information on the use of the modified Lorch function, see the supporting information.
Similar to R(r), the inclined geometry also produces high resolution G(r), as shown in Fig. 9. We observed higher resolution G(r) correlations extending out past the traditional geometry, even after the correlations of the traditional geometry had decayed to near zero. The extended long-range correlations are resolved as a result of the inclined geometry, firstly by increasing the maximum value of Q and secondly as a result of the higher Q-space resolution in the low scattering angle regions, as implied by |∇Q(xP, yP)| in Figs. 4(c) and 4(d), allowing for the capture of higher frequency signals in F(Q) when Fourier transforming to G(r). Additionally, |∇Q(xP, yP)| in the inclined geometry decays in a more natural manner, beginning as higher Q resolution at lower scattering angles and decaying to lower Q resolution at higher scattering angles. The opposite trend is true for the traditional geometry. Capturing the high frequency correlations in the low scattering angle regions where the signal is most intense is of greater benefit to establishing the long-range order.
The physical design of a flat-panel area detector can vary greatly by manufacturer and model. Specifically, the thickness of the scintillator and the mounting design can have large effects on the magnitude of the parallax and topography distortions introduced by a detector. It has been shown by many published reports that it is necessary to correct these distortions in the traditional geometry in order to measure and analyze X-ray scattering patterns properly (Weiß et al., 2012; Lüthi et al., 2019; Marlton et al., 2019; Wu et al., 2002; Zaleski et al., 1998). Equivalently, it was observed that implementing the PE-1621 in the inclined geometry similarly necessitated distortion corrections for proper measurement and analysis.
Flat-panel area detectors are often manufactured for medical applications and commonly repurposed for X-ray total scattering measurements. In these cases, the mounting design of the scintillator is intended for the detector to be in the traditional geometry. In the inclined geometry an increase in the magnitude of the topography distortion is observed due to insufficient scintillator anchoring points, causing nonuniform detector sag (Lüthi et al., 2019; Burns, 2022). Additionally, the large angles of incidence introduced by the detector tilt increase the magnitude of the parallax distortions in the measured pattern (Marlton et al., 2019; Wu et al., 2002; Zaleski et al., 1998; Burns, 2022).
In the inclined geometry, distortions from parallax and topography lead to more noticeable errors in peak positions and amplitudes in the calculated G(r) that need to be addressed. Using the X-ray total scattering pattern of a powdered silicon laboratory standard, the parallax and topography distortions were quantified and corrected. For detailed information on the parallax and topography distortion corrections used here, see the supporting information. Shown in Fig. 10 are the theoretical G(r) models generated by the PDFgui software package (Farrow et al., 2007) for a powdered nickel sample (a) before and (b) after parallax and topography corrections.
The corrections are applied and any distortions in the pair distribution functions G(r) are greatly reduced, improving the peak positions and amplitudes. The quality of the fit is dramatically improved and a significant reduction is observed in the R factor RW from 20.5 to 4.5%. The same distortions due to parallax and topography produce incorrect peak positions in the X-ray total scattering pattern. Fig. 11 shows the calculated generated by the GSAS-II software package (Toby & Von Dreele, 2013) for the powdered nickel after parallax and topography corrections. The produced an acceptable RW of 6.3% over the entire selected scattering region out to 2θ = 90°. As shown in Figs. 10 and 11, with appropriate care taken to correct the parallax and topography distortions, the X-ray total scattering patterns in the inclined geometry can be corrected to levels adequate for both high resolution G(r) analysis and Rietveld refinements.
4. Conclusions
The inclined geometry produces distribution functions of improved quality compared with traditional geometries through two main mechanisms. First, the measurable scattering range is extended to 90°, increasing the accessible magnitudes of the reciprocal-space vector Q by 47%. Consequently, the generated reduced total scattering structure function F(Q) at an increased maximum Q decays to a lower intensity by means of the Debye–Waller factor. This reduces the magnitude of modification functions typically applied to F(Q) to satisfy the requirement of for the Fourier transform to G(r). In concert, by extending the measurable range in Q and reducing the extent of the applied modification functions to F(Q), the real-space resolution in the generated G(r) is preserved with minimal termination ripples. Secondly, the inclined detector geometry shifts the maximum solid angle ΔΩ(γ) and the magnitude of the reciprocal-space vector gradient |∇Q(xP, yP)| away from the beam center to the scattering angle 2θ = 90° − α along the vertical The shift results in the measured counts per pixel at the lower scattering angle regions being reduced, while increasing the measured counts per pixel at higher scattering angle regions, leveling the of the measured total scattering pattern. Reducing the intensity of potentially saturating signals in the lower scattering angle region allows for the incident beam to be increased. In concert, these effects lead to a sixfold increase in coherent scattering signal-to-noise ratio compared with the traditional geometry, implying that a measurement of the same quality at the traditional geometry's maximum value of Q could be acquired measuring 36 times fewer snapshots, greatly reducing acquisition time.
While distortions and artifacts are accentuated in the inclined geometry, these effects are well known in the traditional geometry (Weiß et al., 2012; Lüthi et al., 2019; Marlton et al., 2019; Wu et al., 2002; Zaleski et al., 1998). Through careful distortion and gain corrections, the errors can be corrected to an unobserved level.
Comparing the two detector configurations, using an inclined geometry provides higher quality reciprocal-space information with better statistics. The inclined geometry enables the acquisition of high quality data in less time. It extends the measurable scattering range to 90°, increasing the accessible magnitudes of the reciprocal-space vector Q. The inclined geometry enables acquisition of higher quality data for simultaneous and total scattering studies, yielding structural information on the short-, medium- and long-range orders from one single measurement.
5. Related literature
For further literature related to the supporting information, see Kieffer et al. (2020), O'Donnell et al. (2018), Shah (1971) and Soper & Barney (2012).
Supporting information
Mathematical background. DOI: https://doi.org/10.1107/S1600576723001747/xx5017sup1.pdf
Acknowledgements
We thank Adam Leontowich, Graham King and Narayan Appathurai for helping set up multiple detector configurations and for many insightful discussions on the beamline. The experiments described in this paper were performed at the Canadian Light Source, a national research facility of the University of Saskatchewan, which is supported by the Canada Foundation for Innovation (CFI), the Natural Sciences and Engineering Research Council (NSERC), the National Research Council (NRC), the Canadian Institutes of Health Research (CIHR), the Government of Saskatchewan and the University of Saskatchewan.
Funding information
The following funding is acknowledged: Natural Sciences and Engineering Research Council of Canada (grant No. CRDPJ-2018-531166 to Stefan Kycia and John Dutcher); Canada Foundation for Innovation (grant to Canadian Light Source); National Research Council Canada (grant to Canadian Light Source); Canada Foundation for Innovation (grant to Stefan Kycia).
References
Billinge, S. J. (2019). Phil. Trans. R. Soc. A. 377, 20180413. CrossRef PubMed Google Scholar
Burns, N. (2022). Master's thesis, University of Guelph, Canada. Google Scholar
Chapman, K. W., Lapidus, S. H. & Chupas, P. J. (2015). J. Appl. Cryst. 48, 1619–1626. Web of Science CrossRef CAS IUCr Journals Google Scholar
Farrow, C., Juhas, P., Liu, J., Bryndin, D., Božin, E., Bloch, J., Proffen, T. & Billinge, S. (2007). J. Phys. Condens. Matter, 19, 335219. Web of Science CrossRef PubMed Google Scholar
Gomez, A., Dina, G. & Kycia, S. (2018). Rev. Sci. Instrum. 89, 063301. Web of Science CrossRef PubMed Google Scholar
Grillo, I. (2008). Soft Matter Characterization, edited by R. Barsali & R. Pecora, ch. 13.4.3, p. 739. Heidelberg: Springer. Google Scholar
He, B. B. (2018). Two-Dimensional X-ray Diffraction, ch. 3, pp. 70–71. Chichester: John Wiley & Sons. Google Scholar
Hubbell, J. H. & Seltzer, S. M. (1995). Tables of X-ray Mass Attenuation Coefficients and Mass Energy-Absorption Coefficients 1 keV to 20 MeV for Elements Z = 1 to 92 and 48 Additional Substances of Dosimetric Interest. Technical report, NIST, Gaithersburg, Maryland, USA. Google Scholar
Hughes, I. & Hase, T. (2010). Measurements and Their Uncertainties: A Practical Guide to Modern Error Analysis, ch. 7.4, p. 95. Oxford University Press. Google Scholar
Jensen, K. M. Ø., Blichfeld, A. B., Bauers, S. R., Wood, S. R., Dooryhée, E., Johnson, D. C., Iversen, B. B. & Billinge, S. J. L. (2015). IUCrJ, 2, 481–489. Web of Science CrossRef CAS PubMed IUCr Journals Google Scholar
Kieffer, J., Valls, V., Blanc, N. & Hennig, C. (2020). J. Synchrotron Rad. 27, 558–566. Web of Science CrossRef CAS IUCr Journals Google Scholar
Laaziri, K., Kycia, S., Roorda, S., Chicoine, M., Robertson, J., Wang, J. & Moss, S. (1999a). Phys. Rev. B, 60, 13520–13533. CrossRef CAS Google Scholar
Laaziri, K., Kycia, S., Roorda, S., Chicoine, M., Robertson, J., Wang, J. & Moss, S. (1999b). Phys. Rev. Lett. 82, 3460–3463. CrossRef CAS Google Scholar
Li, H., Karina, A., Ladd-Parada, M., Späh, A., Perakis, F., Benmore, C. & Amann-Winkel, K. (2021). J. Phys. Chem. B, 125, 13320–13328. CrossRef CAS PubMed Google Scholar
Lorch, E. (1969). J. Phys. C Solid State Phys. 2, 229–237. CrossRef Web of Science Google Scholar
Lüthi, M., Bircher, B. A., Meli, F., Küng, A. & Thalmann, R. (2019). Meas. Sci. Technol. 31, 035002. Google Scholar
Lyons, R. (2004). Understanding Digital Signal Processing, ch. 8, pp. 319–327. Upper Saddle River: Prentice Hall/PTR. Google Scholar
Marlton, F., Ivashko, O., Zimmerman, M., Gutowski, O., Dippel, A.-C. & Jørgensen, M. R. V. (2019). J. Appl. Cryst. 52, 1072–1076. Web of Science CrossRef CAS IUCr Journals Google Scholar
Michel, T., Anton, G., Böhnel, M., Durst, J., Firsching, M., Korn, A., Kreisler, B., Loehr, A., Nachtrab, F., Niederlöhner, D., Sukowski, F. & Takoukam Talla, P. (2006). Nucl. Instrum. Methods Phys. Res. A, 568, 799–802. CrossRef CAS Google Scholar
O'Donnell, J. H., Von Dreele, R. B., Chan, M. K. Y. & Toby, B. H. (2018). J. Appl. Cryst. 51, 1244–1250. CrossRef CAS IUCr Journals Google Scholar
Peterson, P. F., Olds, D., McDonnell, M. T. & Page, K. (2021). J. Appl. Cryst. 54, 317–332. Web of Science CrossRef CAS IUCr Journals Google Scholar
Petkov, V. (2012). Characterization of Materials, edited by E. N. Kaufmann, pp. 1361–1372. Chichester: John Wiley & Sons. Google Scholar
Petkov, V., Jeong, I.-K., Chung, J., Thorpe, M., Kycia, S. & Billinge, S. J. (1999). Phys. Rev. Lett. 83, 4089–4092. CrossRef CAS Google Scholar
Shah, J. S. (1971). PhD dissertation, University of Missouri–Rolla, USA. Google Scholar
Skinner, L. B., Benmore, C. J. & Parise, J. B. (2012). Nucl. Instrum. Methods Phys. Res. A, 662, 61–70. Web of Science CrossRef CAS Google Scholar
Skinner, L. B., Huang, C., Schlesinger, D., Pettersson, L. G., Nilsson, A. & Benmore, C. J. (2013). J. Chem. Phys. 138, 074506. Web of Science CrossRef PubMed Google Scholar
Soper, A. K. & Barney, E. R. (2012). J. Appl. Cryst. 45, 1314–1317. Web of Science CrossRef CAS IUCr Journals Google Scholar
Sulyanov, S. N., Popov, A. N. & Kheiker, D. M. (1994). J. Appl. Cryst. 27, 934–942. CrossRef CAS Web of Science IUCr Journals Google Scholar
Toby, B. H. & Von Dreele, R. B. (2013). J. Appl. Cryst. 46, 544–549. Web of Science CrossRef CAS IUCr Journals Google Scholar
Tomberli, B., Rahemtulla, A., Kim, E., Roorda, S. & Kycia, S. (2015). Phys. Rev. B, 92, 064204. Web of Science CrossRef Google Scholar
Waseda, Y. (2003). Anomalous X-ray Scattering for Materials Characterization: Atomic-Scale Structure Determination, Springer Tracts in Modern Physics, Vol. 179. Heidelberg: Springer. Google Scholar
Weiß, D., Lonardoni, R., Deffner, A. & Kuhn, C. (2012). 4th Conference on Industrial Computed Tomography (iCT), 19–21 September 2012, Wels, Austria, pp. 175–181. Aachen: Shaker Verlag. Google Scholar
Wiaderek, K. M., Borkiewicz, O. J., Castillo-Martínez, E., Robert, R., Pereira, N., Amatucci, G. G., Grey, C. P., Chupas, P. J. & Chapman, K. W. (2013). J. Am. Chem. Soc. 135, 4070–4078. Web of Science CrossRef CAS PubMed Google Scholar
Wu, G., Rodrigues, B. L. & Coppens, P. (2002). J. Appl. Cryst. 35, 356–359. Web of Science CrossRef CAS IUCr Journals Google Scholar
Zaleski, J., Wu, G. & Coppens, P. (1998). J. Appl. Cryst. 31, 302–304. Web of Science CrossRef CAS IUCr Journals Google Scholar
This is an open-access article distributed under the terms of the Creative Commons Attribution (CC-BY) Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.