view article

Figure 4
Instrumental Qdamp and SDD obtained from ten repetitions on LaB6 NIST 660b. The gray shaded area highlights the standard deviation for both Qdamp and SDD, and the dashed line corresponds to the average value. The least-squares error of the SDD is not provided by the pyFAI algorithm and therefore cannot be reported.