IUCrJ
Volume 7| Part 3| May 2020| Pages 445-452
ISSN: 2052-2525

High-resolution cryo-EM reconstructions in the presence of substantial aberrations

CROSSMARK_Color_square_no_text.svg

aDepartment of Biophysics, The University of Texas Southwestern Medical Center, Dallas, TX 75390, USA
*Correspondence e-mail: dominika.borek@utsouthwestern.edu, zbyszek.otwinowski@utsouthwestern.edu

Edited by F. Sun, Chinese Academy of Sciences, China (Received 24 December 2019; accepted 20 February 2020; online 26 March 2020)

Here, an analysis is performed of how uncorrected antisymmetric aberrations, such as coma and trefoil, affect cryo-EM single-particle reconstruction (SPR) results, and an analytical formula quantifying information loss owing to their presence is inferred that explains why Fourier-shell coefficient-based statistics may report significantly overestimated resolution if these aberrations are not fully corrected. The analysis is validated with reference-based aberration refinement for two cryo-EM SPR data sets acquired with a 200 kV microscope in the presence of coma exceeding 40 µm, and 2.3 and 2.7 Å reconstructions for 144 and 173 kDa particles, respectively, were obtained. The results provide a description of an efficient approach for assessing information loss in cryo-EM SPR data acquired in the presence of higher order aberrations, and address inconsistent guidelines regarding the level of aberrations that is acceptable in cryo-EM SPR experiments.

1. Introduction

In a typical cryo-EM single-particle reconstruction (SPR) experiment, some aberrations such as defocus are introduced intentionally, while others such as spherical aberration are unavoidable for a given setup (Cheng et al., 2015[Cheng, Y., Grigorieff, N., Penczek, P. A. & Walz, T. (2015). Cell, 161, 438-449.]; Wade, 1992[Wade, R. H. (1992). Ultramicroscopy, 46, 145-156.]; Scherzer, 1936[Scherzer, O. (1936). Z. Phys. 101, 593-603.]). The significance of the remaining aberrations is evaluated on a case-by-case basis (Glaeser et al., 2011[Glaeser, R. M., Typke, D., Tiemeijer, P. C., Pulokas, J. & Cheng, A. (2011). J. Struct. Biol. 174, 1-10.]; Uhlemann & Haider, 1998[Uhlemann, S. & Haider, M. (1998). Ultramicroscopy, 72, 109-119.]) or, more recently, by correcting for small aberrations using reference-based refinement (Zivanov et al., 2020[Zivanov, J., Nakane, T. & Scheres, S. H. W. (2020). IUCrJ, 7, 253-267.]).

In the analysis of aberrations in cryo-EM SPR, where in one image we record only a small part of the focal plane, we can use an isoplanatic approximation in which aberrations are represented by convolutions, and in Fourier space they depend only on the angle of the scattered electrons. In phase-contrast illumination mode, as used in cryo-EM SPR, axial aberrations can be divided into two distinct categories depending on the symmetry properties of the image phase shift as a function of a scattering vector. If the phase shift is centrosymmetric, then aberrations will result in modulations of the image power spectrum. If the phase shift is antisymmetric then the power spectrum will not be modulated, because aberrations will not affect the amplitude of the image but only its phase. The lowest order of antisymmetric phase shift is a translation, which owing to the lack of an absolute coordinate system can be set to zero. Antisymmetric aberrations of the next, third order are called axial coma and trefoil, and these are important for cryo-EM SPR data quality in practice (Cheng et al., 2018[Cheng, A., Eng, E. T., Alink, L., Rice, W. J., Jordan, K. D., Kim, L. Y., Potter, C. S. & Carragher, B. (2018). J. Struct. Biol. 204, 270-275.]; Glaeser et al., 2011[Glaeser, R. M., Typke, D., Tiemeijer, P. C., Pulokas, J. & Cheng, A. (2011). J. Struct. Biol. 174, 1-10.]; Uhlemann & Haider, 1998[Uhlemann, S. & Haider, M. (1998). Ultramicroscopy, 72, 109-119.]). The alignment procedures used in cryo-EM SPR minimize coma and trefoil indirectly, for instance by analyzing changes in the image power spectrum owing to interactions between beam tilt and spherical aberration (Uhlemann & Haider, 1998[Uhlemann, S. & Haider, M. (1998). Ultramicroscopy, 72, 109-119.]). The success of this approach and similar approaches requires co-alignment of the optical axes for multiple lenses, and if the alignment procedures are not properly executed, coma and trefoil may be present and affect the quality of the SPR results, and yet only manifest in specialized analyses (Uhlemann & Haider, 1998[Uhlemann, S. & Haider, M. (1998). Ultramicroscopy, 72, 109-119.]; Mastronarde, 2005[Mastronarde, D. N. (2005). J. Struct. Biol. 152, 36-51.]; Suloway et al., 2005[Suloway, C., Pulokas, J., Fellmann, D., Cheng, A., Guerra, F., Quispe, J., Stagg, S., Potter, C. S. & Carragher, B. (2005). J. Struct. Biol. 151, 41-60.]; Cheng et al., 2018[Cheng, A., Eng, E. T., Alink, L., Rice, W. J., Jordan, K. D., Kim, L. Y., Potter, C. S. & Carragher, B. (2018). J. Struct. Biol. 204, 270-275.]; Zivanov et al., 2020[Zivanov, J., Nakane, T. & Scheres, S. H. W. (2020). IUCrJ, 7, 253-267.]). Furthermore, these and additional, higher order antisymmetric aberrations are induced when data are acquired using the beam-image shift method (Mastronarde, 2005[Mastronarde, D. N. (2005). J. Struct. Biol. 152, 36-51.]; Suloway et al., 2005[Suloway, C., Pulokas, J., Fellmann, D., Cheng, A., Guerra, F., Quispe, J., Stagg, S., Potter, C. S. & Carragher, B. (2005). J. Struct. Biol. 151, 41-60.]; Cheng et al., 2018[Cheng, A., Eng, E. T., Alink, L., Rice, W. J., Jordan, K. D., Kim, L. Y., Potter, C. S. & Carragher, B. (2018). J. Struct. Biol. 204, 270-275.]; Wu et al., 2019[Wu, C., Huang, X., Cheng, J., Zhu, D. & Zhang, X. (2019). J. Struct. Biol. 208, https://doi.org/10.1016/j.jsb.2019.09.013.]), in which coordinated electronic shifts of an illuminating beam and an image are used to navigate away from the optical axis. We restrict our discussion here to axial aberrations, but with the understanding that these aberrations will have different values in different positions of an optical system for data acquired with the beam-image shift method, except for coma, which can be compensated for by the beam tilt (Wu et al., 2019[Wu, C., Huang, X., Cheng, J., Zhu, D. & Zhang, X. (2019). J. Struct. Biol. 208, https://doi.org/10.1016/j.jsb.2019.09.013.]; Glaeser et al., 2011[Glaeser, R. M., Typke, D., Tiemeijer, P. C., Pulokas, J. & Cheng, A. (2011). J. Struct. Biol. 174, 1-10.]).

The acceleration in cryo-EM data acquisition enabled by the beam-image shift method (Mastronarde, 2005[Mastronarde, D. N. (2005). J. Struct. Biol. 152, 36-51.]; Suloway et al., 2005[Suloway, C., Pulokas, J., Fellmann, D., Cheng, A., Guerra, F., Quispe, J., Stagg, S., Potter, C. S. & Carragher, B. (2005). J. Struct. Biol. 151, 41-60.]; Cheng et al., 2016[Cheng, A., Tan, Y. Z., Dandey, V. P., Potter, C. S. & Carragher, B. (2016). Methods Enzymol. 579, 87-102.], 2018[Cheng, A., Eng, E. T., Alink, L., Rice, W. J., Jordan, K. D., Kim, L. Y., Potter, C. S. & Carragher, B. (2018). J. Struct. Biol. 204, 270-275.]) has resulted in discussions of the best experimental strategies for data collection, approaches to compensate for or to correct for optical aberrations and methods to assess modulation and the loss of signal owing to the presence of uncorrected optical aberrations (Glaeser et al., 2011[Glaeser, R. M., Typke, D., Tiemeijer, P. C., Pulokas, J. & Cheng, A. (2011). J. Struct. Biol. 174, 1-10.]; Herzik et al., 2017[Herzik, M. A. Jr, Wu, M. & Lander, G. C. (2017). Nat. Methods, 14, 1075-1078.]; Cheng et al., 2018[Cheng, A., Eng, E. T., Alink, L., Rice, W. J., Jordan, K. D., Kim, L. Y., Potter, C. S. & Carragher, B. (2018). J. Struct. Biol. 204, 270-275.]; Li et al., 2019[Li, K., Sun, C., Klose, T., Irimia-Dominguez, J., Vago, F. S., Vidal, R. & Jiang, W. (2019). J. Struct. Biol. 206, 225-232.]; Wu et al., 2019[Wu, C., Huang, X., Cheng, J., Zhu, D. & Zhang, X. (2019). J. Struct. Biol. 208, https://doi.org/10.1016/j.jsb.2019.09.013.]). Axial coma can be corrected by applying compensating beam tilt during data collection; however, beam tilting does not correct other aberrations (Glaeser et al., 2011[Glaeser, R. M., Typke, D., Tiemeijer, P. C., Pulokas, J. & Cheng, A. (2011). J. Struct. Biol. 174, 1-10.]), and thus the extent to which one can apply beam-image shift without compromising data quality remains open. There are, however, strong indications that higher order aberrations generate a significant loss of information (Wu et al., 2019[Wu, C., Huang, X., Cheng, J., Zhu, D. & Zhang, X. (2019). J. Struct. Biol. 208, https://doi.org/10.1016/j.jsb.2019.09.013.]). Wu and coworkers showed that in the case of 6 × 6 hole data collection with compensating beam tilt, the loss of resolution for the final reconstructions was up to 0.45 Å in comparison to data collection in which the beam was on the optical axis. In their experiments, the coma was compensated and corrected, and thus the loss of resolution in the final reconstruction was probably caused by uncorrected higher order aberrations.

We found that large values of axial aberrations can be precisely estimated and accurately corrected for, leading to large, case-specific improvements in SPR results. We provide formulae for assessing how the levels of uncorrected coma and trefoil affect the resolution of SPR, and discuss their impact on the validation statistics.

2. Materials and methods

2.1. Protein expression, purification and grid preparation

Glucose isomerase (GI), also called xylose isomerase, from Streptomyces rubiginosus was purchased from Hampton Research. Protein slurry was dialyzed three times against an excess of distilled H2O and concentrated to ∼40 mg ml−1 with an Amicon filter (Borek et al., 2018[Borek, D., Bromberg, R., Hattne, J. & Otwinowski, Z. (2018). J. Synchrotron Rad. 25, 451-467.]).

The HemQ protein was a structural genomics target (Midwest Center for Structural Genomics APC35880). We have solved its X-ray crystallographic structure (PDB entry 1t0t) and recently others have determined its function (Celis et al., 2017[Celis, A. I., Gauss, G. H., Streit, B. R., Shisler, K., Moraski, G. C., Rodgers, K. R., Lukat-Rodgers, G. S., Peters, J. W. & DuBois, J. L. (2017). J. Am. Chem. Soc. 139, 1900-1911.]). The expression and purification of the GYMC52_3505 plasmid encoding HemQ in pMCSG7 vector with a Tobacco etch virus (TEV)-cleavable N-terminal His6 tag (Stols et al., 2002[Stols, L., Gu, M., Dieckman, L., Raffen, R., Collart, F. R. & Donnelly, M. I. (2002). Protein Expr. Purif. 25, 8-15.]) followed a previously established protocol (Kim et al., 2011[Kim, Y., Babnigg, G., Jedrzejczak, R., Eschenfeldt, W. H., Li, H., Maltseva, N., Hatzos-Skintges, C., Gu, M., Makowska-Grzyska, M., Wu, R., An, H., Chhor, G. & Joachimiak, A. (2011). Methods, 55, 12-28.]). After purification and tag cleavage, the protein was extensively dialyzed against 20 mM HEPES pH 7.5 and used at a concentration of ∼28 mg ml−1 for grid preparation. The plasmid GYMC52-3505 is available from the DNASU Plasmid Repository (https://dnasu.org).

Cryo-EM grids for both proteins were prepared with an FEI Vitrobot Mark IV. In each case, 3 µl protein solution was applied to the grid at 4°C and 100% humidity followed by 6 s blotting with blot force 20 before the grids were plunged into liquid ethane cooled with liquid nitrogen.

2.2. Data acquisition and analysis

The cryo-EM data set for GI was collected using a 200 kV Talos Arctica microscope equipped with a K2 Gatan camera, with a physical pixel size of 0.91 Å. A phase plate was not used and the objective aperture was not inserted. A total of 202 movies with an exposure time of 100 s per movie were collected. Each movie contains 200 frames with an exposure time of 0.5 s per frame and an electron dose of 120 e Å−2 per movie (Table 1[link]).

Table 1
Data collection and processing

  GI HemQ-57K HemQ-45K EMPIAR 10185 EMPIAR 10186 EMPIAR 10204
Instrument Talos Arctica 200 kV Cryo-Arm 200 kV
Phase plate No No No No No No
Energy filter No No No No No No
Objective aperture No No No Yes Yes Not known
Frames per movie 200 100 100 68 68 49
Electron dose (e Å−2 per frame) 0.7 0.9 0.9 0.99 1.0 1.38
Exposure time (s per frame) 0.5 0.4 0.4 0.25 0.25 Not known
K2 super-resolution mode No Yes Yes Yes Yes No
Detector pixel size (Å) 0.91 0.72 0.91 0.91 0.91 0.885
Data pixel size (Å) N/A 0.36 0.455 0.455 0.455 N/A
Movies collected/deposited 202 268 257 315 260 2161
Movies used for processing 149 258 173 315 260 415
Molecular weight (kDa) 173 144 144 659 659 465
Particle symmetry D2 C5 C5 D7 D7 D2
Total picked particles 114522 156210 236091 109695 91186 157513
Particles after 2D averaging 85527 145966 174776 No 2D classification No 2D classification 82721
Particles used in refinement 61909 81302 129446 85847 78689 52340
†Count-based estimation.

Both the HemQ-57K and HemQ-45K data sets were also collected in the same alignment conditions on a 200 kV Talos Arctica microscope with a K2 Gatan camera run in super-resolution mode, with physical pixel sizes of 0.72 Å for HemQ-57K and 0.91 Å for HemQ-45K. For HemQ-57K, 268 movies were collected with an exposure time of 40 s per movie. Each movie contains 100 frames with an exposure time of 0.4 s per frame and an electron dose of 90 e Å−2 per movie. For HemQ-45 K, 257 movies were collected with an exposure time of 40 s per movie. Each movie contains 100 frames with an exposure time of 0.4 s per frame and an electron dose of 90 e Å−2 per movie.

Complete data sets for EMPIAR depositions 10204, 10185 and 10186 were processed as examples of data sets collected at 200 kV. EMPIAR 10185 and EMPIAR 10186 were collected consecutively on the same instrument, with EMPIAR 10185 collected with a traditional setup by moving only the stage and EMPIAR 10186 collected with the beam-image shift method. We performed image-specific correction for coma in both data sets, producing a material improvement in resolution (Table 2[link]). EMPIAR 10263, data set III, served as an example of a data set collected at 300 kV with a large coma value that was partially corrected by alternative methods.

Table 2
The resolution without and with correction for coma for the analyzed data sets

  GI HemQ-45K HemQ-57K EMPIAR 10204 EMPIAR 10185 EMPIAR 10186
Reported resolution (Å) NA NA NA NA 3.1 3.3
FSC0.143-based resolution before correction (Å) 4.1 3.8 4.3 2.5 3.1 3.2
FSC0.143-based resolution after correction (Å) 2.7 2.6 2.3 2.5 2.5 2.4
Coma (µm)/beam tilt (mrad) 42.7/5.3 56.9/7.0 56.2/6.9 0.36–5.22/0.13–1.93 0.89–3.57/0.33–1.32 1.41–4.61/0.52–1.70
Trefoil (µm) 0.62 0.79 0.49 0.01–0.56 0.76–0.98 0.93–1.07
Resolution at the first oscillation of J0 with 40% compensation 5.2 5.7 5.7 ND ND ND
†EMPIAR 10185 and 10186 were collected consecutively and share the same stable value of trefoil. EMPIAR 10185 was collected with stage shift and has similar coma variation as EMPIAR 10204. EMPIAR 10186 was collected with beam-image shift that induced additional coma variation.

We processed all data sets with cisTEM (Grant et al., 2018[Grant, T., Rohou, A. & Grigorieff, N. (2018). eLife, 7, e35383.]). We modified the cisTEM pipeline by adding reference-based refinement of aberrations, including coma and trefoil, as in JSPR and RELION (Li et al., 2019[Li, K., Sun, C., Klose, T., Irimia-Dominguez, J., Vago, F. S., Vidal, R. & Jiang, W. (2019). J. Struct. Biol. 206, 225-232.]; Zivanov et al., 2018[Zivanov, J., Nakane, T., Forsberg, B. O., Kimanius, D., Hagen, W. J. H., Lindahl, E. & Scheres, S. H. W. (2018). eLife, 7, e42166.]). As discussed in Sections 3[link] and 4[link], we performed multiple cycles that included aberration refinement, orientation refinement and creation of a new reference (Grant et al., 2018[Grant, T., Rohou, A. & Grigorieff, N. (2018). eLife, 7, e35383.]). In these cycles, the resolution limit of the data used for the orientation refinement was selected based on manual assessment of the signal to noise (SNR) estimate in cisTEM exceeding a threshold value, typically around 4. The data-collection and analysis statistics are summarized in Table 1[link]. The cryo-EM movies, maps and models used in the data analysis of GI and HemQ proteins have been deposited in the PDB, the EMDB and EMPIAR under codes 6vrs, EMD-21371 and EMPIAR-10360 for GI; 6vsa, EMPIAR-10363 and EMD-21373 for HemQ-57K; and 6vsc, EMPIAR-10362 and EMD-21376 for HemQ-45K. We solved all of the structures using MOLREP (Vagin & Teplyakov, 2010[Vagin, A. & Teplyakov, A. (2010). Acta Cryst. D66, 22-25.]) and refined them with REFMAC (Murshudov et al., 2011[Murshudov, G. N., Skubák, P., Lebedev, A. A., Pannu, N. S., Steiner, R. A., Nicholls, R. A., Winn, M. D., Long, F. & Vagin, A. A. (2011). Acta Cryst. D67, 355-367.]) used within CCPEM (Wood et al., 2015[Wood, C., Burnley, T., Patwardhan, A., Scheres, S., Topf, M., Roseman, A. & Winn, M. (2015). Acta Cryst. D71, 123-126.]; Burnley et al., 2017[Burnley, T., Palmer, C. M. & Winn, M. (2017). Acta Cryst. D73, 469-477.]; Nicholls et al., 2018[Nicholls, R. A., Tykac, M., Kovalevskiy, O. & Murshudov, G. N. (2018). Acta Cryst. D74, 492-505.]) with manual inspection using Coot (Emsley & Cowtan, 2004[Emsley, P. & Cowtan, K. (2004). Acta Cryst. D60, 2126-2132.]; Emsley et al., 2010[Emsley, P., Lohkamp, B., Scott, W. G. & Cowtan, K. (2010). Acta Cryst. D66, 486-501.]).

In the process of modifying cisTEM, we introduced many changes in the underlying data structures, which resulted in the development of a separate software package. The package does not rely on the particle-stack concept which is used in other cryo-EM packages. We plan to release it in the future, and it will include the implementation of aberration refinement described above. The module is, however, not ready yet for standalone release owing to its reliance on unusual data structures.

3. Results

CTF determination by power spectrum analysis is an inherent part of high-resolution cryo-EM SPR and provides estimates of the magnitude of symmetric aberrations (Wade, 1992[Wade, R. H. (1992). Ultramicroscopy, 46, 145-156.]). However, phase shift does not modulate the power spectrum, so the determination of antisymmetric aberrations has to be performed by other methods (Glaeser et al., 2011[Glaeser, R. M., Typke, D., Tiemeijer, P. C., Pulokas, J. & Cheng, A. (2011). J. Struct. Biol. 174, 1-10.]; Zivanov et al., 2020[Zivanov, J., Nakane, T. & Scheres, S. H. W. (2020). IUCrJ, 7, 253-267.]).

In our analysis of antisymmetric aberrations, we utilize an associative property of pure phase shift that also holds when it is combined with image translation (Hopkins, 1984[Hopkins, H. H. (1984). Opt. Acta Int. J. Opt. 31, 345-368.]). Thus, corrections for the presence of antisymmetric aberrations can be split into separate steps and applied in any order, without loss of accuracy in retrieving information. Consequently, for both large and small magnitudes of antisymmetric aberrations, their impact is defined only by the difference between their true and their assumed or refined values. If their estimates are inaccurate, the difference generates a component of the image phase shift that, after averaging over multiple particles, produces the signal modulation that we analyze here.

Antisymmetric aberration is a convolution with signal, so in the frequency domain the convolution is represented as the multiplication of a signal and a Fourier representation of an aberration. The impact of aberrations on 3D cryo-EM SPR has been theoretically analyzed by considering how images are affected (Uhlemann & Haider, 1998[Uhlemann, S. & Haider, M. (1998). Ultramicroscopy, 72, 109-119.]; Zivanov et al., 2018[Zivanov, J., Nakane, T., Forsberg, B. O., Kimanius, D., Hagen, W. J. H., Lindahl, E. & Scheres, S. H. W. (2018). eLife, 7, e42166.]). However, cryo-EM SPR relies on averaging multiple particles to increase the SNR. Thus, we investigated how the impact of an aberration will propagate to averaged representations of particles.

To this end, we assume a large number of particles which are randomly oriented on a grid with respect to rotation around the beam axis but with potential preferred orientation dependence on the other two Eulerian angles. Averaging all such particles removes the dependence of the average signal on the angle of the aberration in the microscope frame (Cheng et al., 2018[Cheng, A., Eng, E. T., Alink, L., Rice, W. J., Jordan, K. D., Kim, L. Y., Potter, C. S. & Carragher, B. (2018). J. Struct. Biol. 204, 270-275.]). Therefore, the final consequence of an aberration is the resolution-dependent modulation of signal amplitude (Cheng et al., 2018[Cheng, A., Eng, E. T., Alink, L., Rice, W. J., Jordan, K. D., Kim, L. Y., Potter, C. S. & Carragher, B. (2018). J. Struct. Biol. 204, 270-275.]). For each unique projection that is conceptually equivalent to a 2D class average, we can define an in-plane rotational angle of particle orientation φ in the microscope and the difference Δφ between φ and the characteristic direction of a particular antisymmetric aberration (Supplementary Fig. S1). Single particles have no force aligning them with this angle, so we can assume that their distribution is uniform with respect to rotation perpendicular to the grid. We then express the maximum phase shift a for a particular aberration and resolution, which for coma and trefoil is

[a_{3,x} = 2\pi {1 \over 3} C_{3,x} {{\lambda ^2} \over {d^3}}, \eqno (1)]

where d represents resolution and λ represents electron wavelength. The first index of the aberration coefficients identifies the power dependence on resolution and the other index defines angular periodicity in the microscope frame of the phase shift, and so for coma x = 1 and for trefoil x = 3 (Barthel, 2007[Barthel, J. (2007). PhD thesis. Rheinisch-Westfälische Technische Hochschule Aachen.]). Coma can also originate from interactions between beam tilt b and spherical aberrations Cs: C3,1 = 3bCs so b = (C3,1/3Cs). This representation may be convenient when one is interested in expressing coma with respect to beam tilt rather than coma values.

In weak phase approximation, third-order aberrations can be expressed using terms resulting from the Taylor expansion of a wavefront: exp[−i2π(a3,1cosΔφ3,1 + a3,3cos3Δφ3,3)]. To obtain a modulation term for the signal at a given resolution, we average the wavefront over the angular distribution of all possible particles, with the result being the Bessel function of order zero, J0 (Fig. 1[link]):

[{{1}\over{2\pi }} \int\limits_{0}^{2\pi}\exp[ia_{3,x}\cos(x\Delta\varphi_{3,x})]\,{\rm d} \Delta {\varphi}_{3,x} = J_0(a_{3,x}). \eqno (2)]

To our knowledge, this analytical result has not been noticed before in assessing aberrations despite having highly important consequences for their analysis. As discussed in detail later, the foremost consequence is that the structure factors of the reconstruction may become anticorrelated from the reality in some resolution shells, an effect which can be missed in the standard, FSC-based half-maps assessment of resolution (Saxton & Baumeister, 1982[Saxton, W. O. & Baumeister, W. (1982). J. Microsc. 127, 127-138.]; van Heel, 1987[Heel, M. van (1987). Ultramicroscopy, 21, 95-99.]; Scheres & Chen, 2012[Scheres, S. H. W. & Chen, S. (2012). Nat. Methods, 9, 853-854.]), implying much higher resolution than that achieved.

[Figure 1]
Figure 1
Resolution dependence of third-order aberrations on single-particle reconstruction. The x axis represents reciprocal-space resolution, scaled to a = 1 for x = 1 (equation 1[link]). The y axis represents reciprocal-space signal modulation of the reconstruction resulting from averaging the aberration. Three values of aberration coefficients are shown here: black is 1, blue is 0.5 and orange is 0.25.

We expect that data acquired with the beam-image shift method will be affected by more than one type of axial aberration. In such a case, aberrations that have the same angular dependence order (second index) but different radial order (first index) will be strongly correlated (Uhlemann & Haider, 1998[Uhlemann, S. & Haider, M. (1998). Ultramicroscopy, 72, 109-119.]) and these correlations have to be considered when values of aberrations are refined. If in refinement the data were to have uniform information content across resolutions, the refinement will be orthogonalized by Zernike polynomials,

[\matrix{^oZ_n^{-m}(\rho, \varphi) \cr ^eZ_n^m(\rho, \varphi)} = R_n^m(\rho)\matrix{\sin \cr \cos} (m\varphi), \eqno (3)]

where Rnm is the radial function

[\eqalign{&R_n^m (\rho) = \cr&\!\cases {\!\sum \limits_{l=0}^{(n-m)/2} \displaystyle\!{{(-1)^{l}(n\!-\!l)!}\over{l!\left[{{1}\over{2}}(n\!+\!m)\!-\!l\right]!\left[{{1}\over{2}}(n\!-\!m)\!-\!l\right]!}}\rho^{n-2l} &for $n-m$ even \cr \!0 &for $n-m$ odd}}]

defined for integers n and m such as that nm ≥ 0; φ is the azimuthal angle and ρ is the radial distance, with 0 ≤ ρ ≤ 1. The argument of Zernike polynomial ρ = 1 corresponds to the limiting resolution (Bhatia & Wolf, 1954[Bhatia, A. B. & Wolf, E. (1954). Math. Proc. Camb. Phil. Soc. 50, 40-48.]).

For coma, the corresponding Zernike polynomial is ρ3 − 2/3ρ, with the first term representing coma and the second representing translation. Consequently, at the resolution limit, two-thirds of the coma-produced phase shifts will be compensated for by image translation in refinement. This translation can be executed on the whole image or at the level of particles by shifting their positions in the image by the same value, with such an effect not being noticeable in a typical refinement. However, the compensation factor is reduced from the value of two-thirds because the signal is much stronger at low resolution than at the resolution limit. We determined for the analyzed high-resolution cryo-EM SPR data sets (Table 1[link]) that translation produced only about a two-fifths compensating contribution, reducing the maximum value of the coma-induced phase shift by a factor of ∼0.6. The compensation owing to resolution dependence explains some of the discrepancies in the literature discussing the acceptable limits of uncorrected coma. For instance, one proposed limit was a π/4 phase shift in the direction of coma distortion, and without translational compensation (Uhlemann & Haider, 1998[Uhlemann, S. & Haider, M. (1998). Ultramicroscopy, 72, 109-119.]; Glaeser et al., 2011[Glaeser, R. M., Typke, D., Tiemeijer, P. C., Pulokas, J. & Cheng, A. (2011). J. Struct. Biol. 174, 1-10.]). Using our formula (equation 2[link]), we found that this will preserve J0(0.6π/4) = ∼0.94 of the original signal, a reduction in the SNR that is barely significant, and so the π/4 limit is too conservative, as was postulated by Cheng et al. (2018[Cheng, A., Eng, E. T., Alink, L., Rice, W. J., Jordan, K. D., Kim, L. Y., Potter, C. S. & Carragher, B. (2018). J. Struct. Biol. 204, 270-275.]). In the analysis of Cheng and coworkers the impact of coma was derived from FSC curves calculated from numerical experiments, in which a particular value of beam tilt (and corresponding coma) was assumed. FSC-based resolution analysis in the presence of uniform large coma can be misleading because the J0 function oscillates (Fig. 1[link]). When the signal modulation term defined by J0 (equation 2[link]) is negative, both halves of the split data are affected, and so the correlation coefficient between the halves is positive even if the resulting reconstruction has a negative correlation with the truth. If the coma causes very strong modulation in the FSC (Li et al., 2019[Li, K., Sun, C., Klose, T., Irimia-Dominguez, J., Vago, F. S., Vidal, R. & Jiang, W. (2019). J. Struct. Biol. 206, 225-232.]), then it is easy to recognize the correct resolution limit corresponding to the first zero of the Bessel function J0 (equations 1[link] and 2[link]). However, oscillations in the FSC curve may be pronounced or smoothed to different degrees (Fig. 2[link]), even when data sets with very similar values of coma (Table 2[link]) were analyzed. The method that we employed from cisTEM (Grant et al., 2018[Grant, T., Rohou, A. & Grigorieff, N. (2018). eLife, 7, e35383.]) uses only a smooth spherical mask to calculate FSC curves, and so the oscillations could not have come from a molecular-mask effect (Penczek, 2010[Penczek, P. A. (2010). Methods Enzymol. 482, 73-100.]). Thus, in the absence of pronounced oscillations, the resolution limit may be significantly overestimated compared with the consideration based on the first zero of the Bessel function. Consequently, simulation-based procedures relying solely on FSC (Cheng et al., 2018[Cheng, A., Eng, E. T., Alink, L., Rice, W. J., Jordan, K. D., Kim, L. Y., Potter, C. S. & Carragher, B. (2018). J. Struct. Biol. 204, 270-275.]) may grossly underestimate the significance of coma.

[Figure 2]
Figure 2
Each row corresponds to one of our three experiments performed with high coma (beam-tilt) values. The left panel shows oscillations of amplitudes caused by uncorrected aberrations calculated with (1)[link] and (2)[link] for each experiment, the middle panel shows FSC plots before (black) and after (red) coma correction and the corresponding final map fragment for three experiments with high coma (beam-tilt) values. The statistics from each experiment are presented in Table 1[link]. The vertical dotted gray line represents the first zero of the modulation function (2[link]) and the solid line represents the first zero of the modulation function with the assumption that the coma impact was compensated by image translation (5[link]). The resolutions corresponding to the first zero values are listed in Table 2[link]. The right panel shows fragments of unsharpened maps corresponding to the reconstructed maps.

To be more precise, in the case of translation compensation, we introduce the translation using the same notation as for aberration coefficients,

[a_{1,1} = 2\pi C_{1,1}/d. \eqno (4)]

We subtract the phase term with translation to emphasize that this compensates for coma. With such notation, the phase for compensating translation and coma is the same. While the terms a3,1 and a1,1 have a different dependence on resolution, the effect is material where the cubic dependence on resolution dominates, so the overall oscillating behavior is similar,

[\eqalignno {{{1}\over{2\pi}} \textstyle\int\limits_{0}^{2\pi}\exp&[ia_{3,1}\cos(\Delta\varphi_{3,1})-ia_{1,1}\cos(\Delta\varphi_{3,1})]\,\,{\rm d}\Delta \varphi_{3,1} \cr & = J_0(a_{3,1}-a_{1,1}). &(5)}]

The curves with translation compensation are shown in Fig. 2[link].

We tested our approach for two small proteins with particle sizes of 144 and 173 kDa (Fig. 2[link], Table 1[link]), and we obtained reconstructions of 2.32 and 2.70 Å resolution, respectively, in the presence of very high coma, using data acquired with a Talos Arctica 200 kV and K2 Gatan camera. An objective aperture, an energy filter and a phase plate were not used, and the numbers of micrographs and particles in data analysis were moderate (Table 1[link]). We have not found any detrimental effects for correcting coma, even in cases where the generated phase shift is very large, of the order of 5 × 2π. We reprocessed data from EMPIAR for 200 kV instruments, applying the same aberration estimation approach to data for the larger molecules of the proteasome (700 kDa, EMPIAR 10185 and 10186; Herzik et al., 2017[Herzik, M. A. Jr, Wu, M. & Lander, G. C. (2017). Nat. Methods, 14, 1075-1078.]) and β-galactosidase (430 kDa, EMPIAR 10204; Iudin et al., 2016[Iudin, A., Korir, P. K., Salavert-Torres, J., Kleywegt, G. J. & Patwardhan, A. (2016). Nat. Methods, 13, 387-388.]) (Table 1[link]). We refined coma and trefoil independently on each micrograph. We noticed that coma can fluctuate far above refinement uncertainty, and we attribute this observation to differences in the stability of the beam-tilt direction between micrographs. In addition, coma refinement has a strong correlation with overall image shift, affecting the accuracy of coma determination. We found that trefoil on the other hand was remarkably stable and has no significant correlation with other parameters of refinement, so variations in its refined value can be used as a good indicator of statistical uncertainty for third-order aberrations (Fig. 3[link]).

[Figure 3]
Figure 3
Heat maps for coma and trefoil values refined separately per image from the HemQ-57K data set. The leftmost panel shows tight clustering of the coma values, with the center panel magnifying the region of clustering. The right panel shows the values of trefoil, which for this data set were very low and very consistent across the entire data collection. Dividing coma expressed in micrometres by a factor of 8.1 converts it to the equivalent mrad beam-tilt units assuming Cs = 2.7 mm.

The problem generated by the presence of large coma has been analyzed in a recent publication (Li et al., 2019[Li, K., Sun, C., Klose, T., Irimia-Dominguez, J., Vago, F. S., Vidal, R. & Jiang, W. (2019). J. Struct. Biol. 206, 225-232.]) which described the refinement of EMPIAR 10263 (data set III) with coma refinement performed by RELION-3.0 (Zivanov et al., 2018[Zivanov, J., Nakane, T., Forsberg, B. O., Kimanius, D., Hagen, W. J. H., Lindahl, E. & Scheres, S. H. W. (2018). eLife, 7, e42166.]) and JSPR (Guo & Jiang, 2014[Guo, F. & Jiang, W. (2014). Methods Mol. Biol. 1117, 401-443.]; Li et al., 2019[Li, K., Sun, C., Klose, T., Irimia-Dominguez, J., Vago, F. S., Vidal, R. & Jiang, W. (2019). J. Struct. Biol. 206, 225-232.]), with the results presented in Fig. 6(b) in Li et al. (2019[Li, K., Sun, C., Klose, T., Irimia-Dominguez, J., Vago, F. S., Vidal, R. & Jiang, W. (2019). J. Struct. Biol. 206, 225-232.]).

We reprocessed this data set and obtained a much tighter clustering of coma values, similar to the clustering observed for our data sets. RELION-3.0 (Zivanov et al., 2018[Zivanov, J., Nakane, T., Forsberg, B. O., Kimanius, D., Hagen, W. J. H., Lindahl, E. & Scheres, S. H. W. (2018). eLife, 7, e42166.]) underestimated coma by a factor of ∼4 compared with our estimate, and the average of the JSPR individual refinements per micrograph had a value that was underestimated by 25% compared with our values. However, we used a similar reference-based refinement procedure as JSPR (Li et al., 2019[Li, K., Sun, C., Klose, T., Irimia-Dominguez, J., Vago, F. S., Vidal, R. & Jiang, W. (2019). J. Struct. Biol. 206, 225-232.]; Guo & Jiang, 2014[Guo, F. & Jiang, W. (2014). Methods Mol. Biol. 1117, 401-443.]) and RELION-3.0 (Zivanov et al., 2018[Zivanov, J., Nakane, T., Forsberg, B. O., Kimanius, D., Hagen, W. J. H., Lindahl, E. & Scheres, S. H. W. (2018). eLife, 7, e42166.], 2020[Zivanov, J., Nakane, T. & Scheres, S. H. W. (2020). IUCrJ, 7, 253-267.]), and so the different outcomes are most probably owing to differing refinement schemes (Li et al., 2019[Li, K., Sun, C., Klose, T., Irimia-Dominguez, J., Vago, F. S., Vidal, R. & Jiang, W. (2019). J. Struct. Biol. 206, 225-232.]). This was confirmed by a discussion with the author of the publication, who stated that coma refinement was not followed by additional cycles of particle global orientation and coma refinement, unlike in our approach. Thus, the results presented in Fig. 6(b) in Li et al. (2019[Li, K., Sun, C., Klose, T., Irimia-Dominguez, J., Vago, F. S., Vidal, R. & Jiang, W. (2019). J. Struct. Biol. 206, 225-232.]) are consistent with the behavior of our procedure when compared with the results from our first cycle.

We observed that only the presence of a large fraction of incorrect particles or particles with grossly incorrect orientation in a micrograph would bias the coma refinement towards the starting point. When starting with coma refined by the previous cycle, re-refining the particle orientation attenuated the bias in the next cycle, so even the presence of a moderate number of bad particles did not affect the convergence of the procedure (Cash et al., 2020[Cash, J. N., Kearns, S., Li, Y. & Cianfrocco, M. A. (2020). bioRxiv, 2020.01.21.914507.]). We achieved additional improvement, in terms of resolution and the spread of coma values, when we changed the null hypothesis regarding coma from a value of zero to the average of the refined coma values. This more appropriate null hypothesis was applied by taking advantage of the associative properties of coma, which allowed us to apply coma phase-shift correction to the images; therefore, all of the steps in the final round of the data analysis, starting from particle picking, were performed on images corrected by the average value of coma. Subsequent coma refinement, representing a difference from the previous average used in image correction, was still performed independently for all micrographs, but the residual bias resulting from the initial coma value being zero was eradicated. Coma or beam-tilt refinement has to use a reference-based target function, irrespective of implementation. We would expect that small corrections would be equally well characterized by all programs. However, local refinements and restraints may have different convergence ranges depending on the implementation, and this is mostly likely to be the explanation for the differences in results between programs.

Uncorrected aberrations create systematic patterns of phase shift, and if such aberration error is constant across a data set then the reconstruction will also be altered in a systematic way, not only by losing amplitude but also by flipping its sign at some resolutions. In this case, the FSC curve may undergo oscillations, with the first minimum being the effective resolution limit of the result. Oscillations in the FSC curve are recognized as a qualitative sign of problems with the quality of cryo-EM SPR results (Penczek, 2010[Penczek, P. A. (2010). Methods Enzymol. 482, 73-100.]). We provide here in (2[link]) an explanation for another possible source of these oscillations. In our glucose isomerase (GI) data acquired with 58 µm coma, corresponding to 7.2 mrad beam tilt, before correcting coma we observed four oscillations in the FSC curve resulting from phase shift (Fig. 2[link]), with the map interpretability being inconsistent with the FSC-based resolution indicator. Therefore, if FSC oscillations are encountered, refining coma is highly recommended to diagnose the problem, with the potential outcome being substantial reconstruction improvement by correcting the aberration.

4. Discussion

Although instruments may be aligned quite accurately before data collection, our analyses and those of others (Li et al., 2019[Li, K., Sun, C., Klose, T., Irimia-Dominguez, J., Vago, F. S., Vidal, R. & Jiang, W. (2019). J. Struct. Biol. 206, 225-232.]; Cheng et al., 2018[Cheng, A., Eng, E. T., Alink, L., Rice, W. J., Jordan, K. D., Kim, L. Y., Potter, C. S. & Carragher, B. (2018). J. Struct. Biol. 204, 270-275.]; Herzik et al., 2017[Herzik, M. A. Jr, Wu, M. & Lander, G. C. (2017). Nat. Methods, 14, 1075-1078.]; Glaeser et al., 2011[Glaeser, R. M., Typke, D., Tiemeijer, P. C., Pulokas, J. & Cheng, A. (2011). J. Struct. Biol. 174, 1-10.]) indicate that coma can vary significantly between images and the extent of the variation is data-set dependent. We do not fully understand why coma varies, but a possible explanation may involve avoiding stage shift for small position changes without beam-tilt adjustment, with the allowed range of such shifts being a user-adjustable parameter. Another possible explanation is small fluctuations in the stability of the objective lens. The reason for this does not necessarily have to be hardware problems, for example temperature fluctuations; it can also arise from the data-collection scheme, as we and others (Wu et al., 2019[Wu, C., Huang, X., Cheng, J., Zhu, D. & Zhang, X. (2019). J. Struct. Biol. 208, https://doi.org/10.1016/j.jsb.2019.09.013.]) have observed that the beam-image shift approach is not only faster but generates different patterns of experimental instabilities. However, until this problem is further resolved, we recommend refinement of not only overall coma but also coma for individual images. Trefoil typically is not pronounced, but for one pair of EMPIAR data sets (EMPIAR 10185 and EMPIAR 10186), and also in our data sets (not shown), it was highly significant. However, even in these cases only the overall value of trefoil for the data set was important, with an insignificant level of variations between individual images. (2)[link] provides quantification of the impact that uncorrected aberrations have on reconstruction in specific cryo-EM SPR experiments and associated computations. The dependence of SNR on resolution in the presence of uncorrected aberrations described by (2)[link] has characteristic modulations (Fig. 1[link]) that allow the separation of the impact of aberrations from other effects in reconstruction, consequently guiding experimental strategy.

Correction for antisymmetric aberrations can be performed during reconstruction (Li et al., 2019[Li, K., Sun, C., Klose, T., Irimia-Dominguez, J., Vago, F. S., Vidal, R. & Jiang, W. (2019). J. Struct. Biol. 206, 225-232.]; Zivanov et al., 2018[Zivanov, J., Nakane, T., Forsberg, B. O., Kimanius, D., Hagen, W. J. H., Lindahl, E. & Scheres, S. H. W. (2018). eLife, 7, e42166.], 2020[Zivanov, J., Nakane, T. & Scheres, S. H. W. (2020). IUCrJ, 7, 253-267.]), but the consequence of the associative property is that it can just as well be performed at the whole-micrograph level before reconstruction commences. In the presence of large coma, correcting coma at the image level reduces the point-spread function of the imaging system so it may improve particle-masking operations. If the value of these aberrations is known from external calibration then the image-based correction may simplify the application of these corrections to the data without modifying downstream analysis programs.

What affects cryo-EM SPRs is the error in the aberration model used and not the magnitude of the aberrations themselves, at least up to the theoretical limit where the image of a point source (for example due to coma) extends outside the detector. Therefore, it is important to calibrate all aberrations and not only those that affect the power spectrum. This can be accomplished prior to data collection or afterwards by reference-based refinement using the structure being solved. Once appropriate coma calibrations and corrections are used, this means that procedures where the beam is intentionally tilted can be used on a larger scale than those applied until now (Wu et al., 2019[Wu, C., Huang, X., Cheng, J., Zhu, D. & Zhang, X. (2019). J. Struct. Biol. 208, https://doi.org/10.1016/j.jsb.2019.09.013.]; Cheng et al., 2018[Cheng, A., Eng, E. T., Alink, L., Rice, W. J., Jordan, K. D., Kim, L. Y., Potter, C. S. & Carragher, B. (2018). J. Struct. Biol. 204, 270-275.]). This can provide limited but additional three-dimensional particle information on top of the projection image without any of the time and precision penalties associated with mechanical rotations.

Footnotes

Current address: Ligo Analytics, Dallas, TX 75206, USA

Acknowledgements

We thank Tabitha Emde for protein purification and grid preparation for the HemQ protein. Author contributions were as follows. R. Bromberg, D. Borek and Z. Otwinowski developed an approach to aberration analysis; R. Bromberg and Z. Otwinowski implemented the approach; D. Borek acquired data; R. Bromberg, Y. Guo and Z. Otwinowski analyzed the data; R. Bromberg, Y. Guo, D. Borek and Z. Otwinowski wrote the manuscript. Conflict of interest statement: R. Bromberg, Y. Guo, D. Borek and Z. Otwinowski are co-founders of Ligo Analytics. Y. Guo serves as the CEO of Ligo Analytics. Z. Otwinowski is a co-founder of HKL Research. R. Bromberg, Z. Otwinowski and D. Borek are co-inventors listed on a provisional patent application that has been filed on work related to the subject of this manuscript. This report was prepared as an account of work sponsored by an agency of the United States Government. Neither the United States Government nor any agency thereof, nor any of their employees, makes any warranty, express or implied, or assumes any legal liability or responsibility for the accuracy, completeness, or usefulness of any information, apparatus, product, or process disclosed, or represents that its use would not infringe privately owned rights. Reference herein to any specific commercial product, process, or service by trade name, trademark, manufacturer, or otherwise does not necessarily constitute or imply its endorsement, recommendation, or favoring by the United States Government or any agency thereof. The views and opinions of authors expressed herein do not necessarily state or reflect those of the United States Government or any agency thereof.

Funding information

This project has been funded in part with federal funds from the National Institute of Allergy and Infectious Diseases, National Institutes of Health, Department of Health and Human Services, under Contract No. HHSN272201700060C. This project was also supported by the National Institutes of Health (R21GM126406 to D. Borek and R01GM117080 and R01GM118619 to Z. Otwinowski) and the Department of Energy (DE-SC0019600 to Y. Guo). We thank the Cryo-Electron Microscopy Facility at UT Southwestern Medical Center, which is supported by grant RP170644 from the Cancer Prevention and Research Institute of Texas (CPRIT), for maintaining the Talos Arctica microscope.

References

First citationBarthel, J. (2007). PhD thesis. Rheinisch-Westfälische Technische Hochschule Aachen.  Google Scholar
First citationBhatia, A. B. & Wolf, E. (1954). Math. Proc. Camb. Phil. Soc. 50, 40–48.  CrossRef Web of Science Google Scholar
First citationBorek, D., Bromberg, R., Hattne, J. & Otwinowski, Z. (2018). J. Synchrotron Rad. 25, 451–467.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationBurnley, T., Palmer, C. M. & Winn, M. (2017). Acta Cryst. D73, 469–477.  Web of Science CrossRef IUCr Journals Google Scholar
First citationCash, J. N., Kearns, S., Li, Y. & Cianfrocco, M. A. (2020). bioRxiv, 2020.01.21.914507.  Google Scholar
First citationCelis, A. I., Gauss, G. H., Streit, B. R., Shisler, K., Moraski, G. C., Rodgers, K. R., Lukat-Rodgers, G. S., Peters, J. W. & DuBois, J. L. (2017). J. Am. Chem. Soc. 139, 1900–1911.  Web of Science CrossRef CAS PubMed Google Scholar
First citationCheng, A., Eng, E. T., Alink, L., Rice, W. J., Jordan, K. D., Kim, L. Y., Potter, C. S. & Carragher, B. (2018). J. Struct. Biol. 204, 270–275.  Web of Science CrossRef CAS PubMed Google Scholar
First citationCheng, A., Tan, Y. Z., Dandey, V. P., Potter, C. S. & Carragher, B. (2016). Methods Enzymol. 579, 87–102.  Web of Science CrossRef CAS PubMed Google Scholar
First citationCheng, Y., Grigorieff, N., Penczek, P. A. & Walz, T. (2015). Cell, 161, 438–449.  Web of Science CrossRef CAS PubMed Google Scholar
First citationEmsley, P. & Cowtan, K. (2004). Acta Cryst. D60, 2126–2132.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationEmsley, P., Lohkamp, B., Scott, W. G. & Cowtan, K. (2010). Acta Cryst. D66, 486–501.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationGlaeser, R. M., Typke, D., Tiemeijer, P. C., Pulokas, J. & Cheng, A. (2011). J. Struct. Biol. 174, 1–10.  Web of Science CrossRef PubMed Google Scholar
First citationGrant, T., Rohou, A. & Grigorieff, N. (2018). eLife, 7, e35383.  Web of Science CrossRef PubMed Google Scholar
First citationGuo, F. & Jiang, W. (2014). Methods Mol. Biol. 1117, 401–443.  CrossRef CAS PubMed Google Scholar
First citationHeel, M. van (1987). Ultramicroscopy, 21, 95–99.  Google Scholar
First citationHerzik, M. A. Jr, Wu, M. & Lander, G. C. (2017). Nat. Methods, 14, 1075–1078.  Web of Science CrossRef CAS PubMed Google Scholar
First citationHopkins, H. H. (1984). Opt. Acta Int. J. Opt. 31, 345–368.  CrossRef Web of Science Google Scholar
First citationIudin, A., Korir, P. K., Salavert-Torres, J., Kleywegt, G. J. & Patwardhan, A. (2016). Nat. Methods, 13, 387–388.  Web of Science CrossRef CAS PubMed Google Scholar
First citationKim, Y., Babnigg, G., Jedrzejczak, R., Eschenfeldt, W. H., Li, H., Maltseva, N., Hatzos-Skintges, C., Gu, M., Makowska-Grzyska, M., Wu, R., An, H., Chhor, G. & Joachimiak, A. (2011). Methods, 55, 12–28.  Web of Science CrossRef CAS PubMed Google Scholar
First citationLi, K., Sun, C., Klose, T., Irimia-Dominguez, J., Vago, F. S., Vidal, R. & Jiang, W. (2019). J. Struct. Biol. 206, 225–232.  Web of Science CrossRef CAS PubMed Google Scholar
First citationMastronarde, D. N. (2005). J. Struct. Biol. 152, 36–51.  Web of Science CrossRef PubMed Google Scholar
First citationMurshudov, G. N., Skubák, P., Lebedev, A. A., Pannu, N. S., Steiner, R. A., Nicholls, R. A., Winn, M. D., Long, F. & Vagin, A. A. (2011). Acta Cryst. D67, 355–367.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationNicholls, R. A., Tykac, M., Kovalevskiy, O. & Murshudov, G. N. (2018). Acta Cryst. D74, 492–505.  Web of Science CrossRef IUCr Journals Google Scholar
First citationPenczek, P. A. (2010). Methods Enzymol. 482, 73–100.  Web of Science CrossRef CAS PubMed Google Scholar
First citationSaxton, W. O. & Baumeister, W. (1982). J. Microsc. 127, 127–138.  CrossRef CAS PubMed Web of Science Google Scholar
First citationScheres, S. H. W. & Chen, S. (2012). Nat. Methods, 9, 853–854.  Web of Science CrossRef CAS PubMed Google Scholar
First citationScherzer, O. (1936). Z. Phys. 101, 593–603.  CrossRef Google Scholar
First citationStols, L., Gu, M., Dieckman, L., Raffen, R., Collart, F. R. & Donnelly, M. I. (2002). Protein Expr. Purif. 25, 8–15.  Web of Science CrossRef PubMed CAS Google Scholar
First citationSuloway, C., Pulokas, J., Fellmann, D., Cheng, A., Guerra, F., Quispe, J., Stagg, S., Potter, C. S. & Carragher, B. (2005). J. Struct. Biol. 151, 41–60.  Web of Science CrossRef PubMed CAS Google Scholar
First citationUhlemann, S. & Haider, M. (1998). Ultramicroscopy, 72, 109–119.  Web of Science CrossRef CAS Google Scholar
First citationVagin, A. & Teplyakov, A. (2010). Acta Cryst. D66, 22–25.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationWade, R. H. (1992). Ultramicroscopy, 46, 145–156.  CrossRef CAS Web of Science Google Scholar
First citationWood, C., Burnley, T., Patwardhan, A., Scheres, S., Topf, M., Roseman, A. & Winn, M. (2015). Acta Cryst. D71, 123–126.  Web of Science CrossRef IUCr Journals Google Scholar
First citationWu, C., Huang, X., Cheng, J., Zhu, D. & Zhang, X. (2019). J. Struct. Biol. 208, https://doi.org/10.1016/j.jsb.2019.09.013Web of Science CrossRef Google Scholar
First citationZivanov, J., Nakane, T., Forsberg, B. O., Kimanius, D., Hagen, W. J. H., Lindahl, E. & Scheres, S. H. W. (2018). eLife, 7, e42166.  Web of Science CrossRef PubMed Google Scholar
First citationZivanov, J., Nakane, T. & Scheres, S. H. W. (2020). IUCrJ, 7, 253–267.  Web of Science CrossRef CAS PubMed IUCr Journals Google Scholar

This is an open-access article distributed under the terms of the Creative Commons Attribution (CC-BY) Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.

IUCrJ
Volume 7| Part 3| May 2020| Pages 445-452
ISSN: 2052-2525