Download citation
Download citation
link to html
Quantitative phase analysis of a system of n phases can be effected without the need for calibration standards provided at least n different mixtures of these phases are available. A series of linear equations relating diffracted X-ray intensities, weight fractions and quantitation factors coupled with mass balance relationships can be solved for the unknown weight fractions and factors. Uncertainties associated with the measured X-ray intensities, owing to counting of random X-ray quanta, are used to estimate the errors in the calculated parameters utilizing a Monte Carlo simulation. The Monte Carlo approach can be generalized and applied to any quantitative X-ray diffraction phase analysis method. Two examples utilizing mixtures of CaCO3, Fe2O3 and CaF2 with an α-SiO2 (quartz) internal standard illustrate the quantitative method and corresponding error analysis. One example is well conditioned; the other is poorly conditioned and, therefore, very sensitive to errors in the measured intensities.
Follow J. Appl. Cryst.
Sign up for e-alerts
Follow J. Appl. Cryst. on Twitter
Follow us on facebook
Sign up for RSS feeds