A bespoke capillary sample holder is described that attaches to the cold head of a commercially manufactured (PheniX) closed-cycle helium cryostat originally intended for flat-plate geometry. The new holder allows high-resolution synchrotron powder diffraction data to be collected from samples in Debye-Scherrer geometry over the temperature range 11-295 K. To demonstrate that high-quality powder data can be obtained using this new sample holder, structural refinement (Rietveld) and thermal expansion results measured from reference samples (Si and Al) are presented.