short communications
Absolute soft X-ray measurements using an ion chamber
aElectrotechnical Laboratory, Umezono, Tsukuba, Ibaraki 305-8568, Japan
*Correspondence e-mail: nsaito@etl.go.jp
Measurements of soft X-ray absolute intensities have been carried out using a double ion chamber and monochromated synchrotron radiation. The chamber is cylindrical and 1.3 m long. The soft X-ray beam enters the chamber at a position off the central axis, and the produced ions are collected with electrodes on the opposite side to the photon entrance. An index constant for rare gas, the γ-value, which is the average number of electrons emitted from an atom having absorbed a photon, was used for obtaining the absolute photon intensity. The obtained intensity ranges from 1 Gphotons s−1 to 25 Gphotons s−1 in the energy range 72–800 eV. The estimated uncertainty is about 5–20% depending on the intensity and the spectral purity of the soft X-rays.
Keywords: soft X-rays; absolute intensity; double ion chamber; γ-values.
1. Introduction
The utilization of soft X-rays is currently making advances into micro-fabrication techniques, characterization of different materials and micro-analyses of biological samples with the efficient use of a number of synchrotron radiation facilities (Saisho & Gohshi, 1996). Much basic research needs a monochromatic beam with high resolution, but some studies require absolute intensity of the photon beam, e.g. quantitative characterization of irradiation effects and impurity analysis of several elements. Synchrotron radiation has a standard-light-source capability in the UV to X-ray region, because the intensity and characteristics of the radiation can be calculated (Krumrey & Tegeler, 1992; Madden et al., 1985; Sakurai et al., 1989; Suzuki & Saito, 1986, 1993). However, this radiation has a wide energy spectrum and thus it is usually necessary to employ some focusing and monochromatizing optical elements. Since these elements have unknown efficiencies on photon transfer, the photon beam at the experimental point is not of a standard whose intensity and energy width are well determined. Therefore it is necessary to develop a technique for measuring the absolute intensity of the soft X-ray beam. In the present study, a double ion chamber has been developed, with which the absolute intensity of the soft X-ray beam is obtained with the aid of an index constant for rare gas on multiple photoionization.
2. Methods for absolute measurement
A ), but they have the serious drawback of limited or of no available data on the W-values in the soft X-ray region (Saito & Suzuki, 1986; Wyckoff, 1979). A double ion chamber including a rare gas is well known to work effectively as the absolute detection device in the vacuum ultraviolet radiation region (Saito, 1994; Samson, 1967). Since singly charged ions of rare-gas atoms are dominantly produced in this energy region, the photon intensity is given by
and an ion chamber with a single electrode have been used for the absolute measurement of X-rays for many years (Caruso & Neupert, 1965where i1 and i2 denote the ion currents collected with the first and second electrodes, respectively, and e indicates the Since multiple takes place significantly in the soft X-ray region (Holland et al., 1979; Saito & Suzuki, 1992, 1997), an index constant, the γ-value (the average number of electrons emitted from the atom having absorbed a single photon), should be multiplied to the denominator of (1). Since available γ-value data were limited to only those near 100 eV, the γ-value had to be measured at the initial stage of the present work.
3. Experimental
Monochromatic soft X-rays were obtained by dispersing synchrotron radiation from the electron storage ring at the Electrotechnical Laboratory. The soft X-ray beam passed through a differential pumping stage and entered a chamber including a time-of-flight (TOF) γ-values of Ne, Ar, Kr and Xe. In the absolute measurement, the photon beam was introduced to the cylindrical-shaped double ion chamber (see Fig. 1). The window, a thin VYNS foil (Manson type `F' filter), for the incident photon is positioned at 15 mm from the central axis, and electrodes for collecting the produced ions are fixed at 22 mm from the axis. A positive potential is applied to the outer cylindrical electrode, of diameter 65 mm, for the collection of ions. The set of other electrodes is composed of six cylinders of diameter 5 mm; their lengths are shown in Fig. 1. Each Au-plated electrode is electrically separated with Teflon insulators. The ion chamber was evacuated with a 280 l s−1 turbo-molecular pump just before measurements, and supplied with a rare gas at about 10−2–103 Pa for obtaining ion currents. The ion current at each electrode was led to a pico-amperemeter and transferred to a personal computer. The gas density was detected with a capacitance monometer (Baratron 690) and controlled with an automatic valve system. The size of the soft X-ray beam was defined using a circular aperture of diameter 1 mm.
for measurements of the4. Results and discussion
Branching ratios for multi-charged ions from rare-gas atoms were measured over the soft X-ray range using a pulse electric field applied to the TOF γ-value was derived from the measured ratios. The γ-value data for Ne, Ar, Kr and Xe are reported elsewhere (Suzuki & Saito, 1993). In selectiing adequate species of atoms, a large and little variation in the γ-value are important factors in the energies of interest.
TheThe performance test of the double ion chamber was carried out using Ne at a photon energy of 72 eV. Fig. 2 shows collected ion currents for the second through to the fifth electrode as a function of the voltage applied to the outer cylinder at a gas pressure of 53.2 Pa. The currents increase linearly with voltage below 0.8 V and approach saturation plateaus near 5 V. Since the electrodes are positioned off the central axis, calculations indicated that a strong electric field is not formed even with high applied voltages.
Apparent photon intensities at photon energies of 72, 150, 600 and 800 eV are plotted as a function of the Ar gas pressure at the applied voltage of 15 V in Fig. 3. Data points denote the experimental results when the ion current for the fourth electrode, of length 500 mm, was used as i1 and that of the fifth electrode was employed as i2 in equation (1). An Au photon-intensity monitor was used during these measurements. Data were normalized with an electron beam current of 100 mA in the storage ring. Real photon intensities have been obtained through extrapolation to zero gas density. The data for 600 eV were seriously affected by higher-order light, which induced a sharp increase in intensity above 500 Pa.
The ratio between the apparent photon intensity at sufficiently high pressure and that at almost zero pressure, N, indicates the effect of through collisions of the electron ejected from the atom with other ambient atoms. The W-value has been obtained as
Table 1 lists measured absolute intensities of the soft X-rays (fluence rate) in the Ar sample gas, as well as the W-values. The uncertainties range from 5% to 20%. The main source of uncertainty originates from the estimation of a crossing point at the ordinate extrapolated from apparent photon intensities. The uncertainty of the γ-value used is about 1% at the present photon energies. The obtained W-values are close to those for low-energy electrons (Combecher, 1980), suggesting that the present ion chamber has been correctly operated. Measurements using other rare gases were tentatively carried out, which yielded similar fluence rates of the soft X-rays to the values listed in Table 1.
|
5. Summary
A trial determination of the fluence rate of soft X-rays in the energy range 72–800 eV has been made using a double ion chamber and the γ-value for Ar. Obtained absolute values of the fluence rate are comparable with those expected from the Au monitor. Uncertainties are not small and thus a decrease in noise currents is necessary for establishing a standard soft X-ray field of high quality.
Acknowledgements
The authors wish to thank the staff of the accelerator group at the Electrotechnical Laboratory for their continued operation of the storage ring.
References
Caruso, A. J. & Neupert, W. M. (1965). Appl. Opt. 4, 247–251. CrossRef Web of Science
Combecher, D. (1980). Rad. Res. 84, 189–218. CrossRef CAS Web of Science
Holland, D. M. P., Codling, K., West, J. B. & Marr, G. V. (1979). J. Phys. B, 12, 2465–2484. CrossRef CAS Web of Science
Krumrey, M. & Tegeler, E. (1992). Rev. Sci. Instrum. 63, 797–801. CrossRef CAS Web of Science
Madden, R. P., Ederer, D. L. & Parr, A. C. (1985). Nucl. Instrum. Methods, B10/11, 289–292. CrossRef Web of Science
Saisho, H. & Gohshi, Y. (1996). Analytical Spectroscopy Library, Vol. 7, Applications of Synchrotron Radiation to Material Analysis. Amsterdam: Elsevier Science.
Saito, N. & Suzuki, I. H. (1986). Chem. Phys. 108, 327–333. CrossRef CAS Web of Science
Saito, N. & Suzuki, I. H. (1992). Int. J. Mass Spectrom. Ion Processes, 115, 157–172. CrossRef CAS Web of Science
Saito, N. & Suzuki, I. H. (1997). J. Phys. Soc. Jpn. 66, 1979. CrossRef Web of Science
Saito, T. (1994). Res. ETL, 960, 39–56. (In Japanese.)
Sakurai, M., Morita, S., Fujita, J., Yonezu, H., Fukui, K., Sakai, K., Nakamura, E., Watanabe, M., Ishiguro, E. & Yamashita, K. (1989). Rev. Sci. Instrum. 60, 2089–2092. CrossRef CAS Web of Science
Samson, J. A. R. (1967). Techniques of Vacuum Ultraviolet Spectroscopy. New York: Wiley Interscience.
Suzuki, I. H. & Saito, N. (1986). Jpn. J. Appl. Phys. 25, 130–136. CrossRef CAS Web of Science
Suzuki, I. H. & Saito, N. (1993). Bull. ETL, 57, 34–51. (In Japanese.)
Wyckoff, H. O. (1979). ICRU Report, Vol. 31, W-Values for Photons and Electrons. Washington: Int. Commis. Radiat. Units. Meas.