short communications
A new UHV angle encoder for high-resolution synchrotron radiation monochromators
aBESSY GmbH, Lentzeallee 100, D-14195 Berlin, Germany, and bDr Johannes HEIDENHAIN GmbH, D-83292 Traunreut, Germany
*Correspondence e-mail: senf@exp.bessy.de
The high precision of 0.1 arcsec required for the positioning of optical elements in new two-axes monochromators at the undulator beamlines at BESSY II has led to the development of UHV-compatible high-precision angle encoders. Mounted directly on the rotation axes, they provide substantial advantages over measuring systems connected outside the vacuum vessel. Making use of a fast closed-loop control system, an accuracy of 0.1 arcsec at a resolution of less than 0.01 arcsec has been experimentally verified.
Keywords: angle encoders; monochromators; ultrahigh vacuums.
1. Introduction
For an optimum transformation of the high et al., 1995) we have pointed out improvements of the wavelength readout of a monochromator by making use of a high-precision incremental angle encoder connected directly to the rotation axis of an optical element. The development of such a measurement system has now been successfully carried out in a cooperative effort by the measuring system manufacturer HEIDENHAIN and by BESSY.
2. Incremental angle encoder
During the past two years a conventional high-precision angle encoder has been modified in order to make it compatible with UHV requirements and particularly with high temperatures. For this purpose certain parts have been replaced or even transferred to an interface card. Fig. 1 illustrates the principle of the incremental angle encoder. When a circular scale with a very fine line grating is moved relative to another grating with the same structure, the lines of the two graduations alternately align. The resulting light–dark modulation is sensed by four photovoltaic cells. The sinusoidal phase-shifted current signals are then amplified, digitized, etc. in the subsequent electronics.
![]() | Figure 1 Principle of photoelectric scanning of radial gratings on glass. |
Fig. 2 shows a simplified schematic through the angle encoder with its integral stator-side coupling and two of the four scanning heads. The fourfold scanning of the grating contributes to its high degree of accuracy. Moreover, a special correction routine can be carried out to determine and store the correction data. The photograph in Fig. 3
presents the UHV-compatible angle encoder RON905-UHV, with its 170 mm-diameter stainless-steel shielding and electrical connector. Additional improvements in the analysis and data treatment yielded a total gain in resolution and accuracy as detailed in Table 1
and described below.
|
![]() | Figure 2 Simplified schematic diagram of an angle encoder with integral stator coupling and two of the four scanning heads. |
![]() | Figure 3 Photograph of the UHV-compatible angle encoder RON905-UHV. |
3. Results
After successful tests of a prototype, further UHV encoders have been manufactured. By now, two of them have been tested for more than six months in the first of the new BESSY II monochromators. The positioning system of the monochromators consists of a conventional drive unit, i.e. stepper motor, spindle and lever arm. One angle encoder has been mounted directly on the grating rotation axis and another one on the mirror rotation axis, both inside the vacuum vessel. For comparison, two precision linear encoder systems have been mounted as already described and as used with former monochromators. For independent positioning control an autocollimator (ELCOMAT 2000, Müller/Wedel, Germany) has been set up, realizing an autocollimation path together with the installed X-ray optics, i.e. plane mirror, plane grating blank and normal retro-reflecting mirror. A comparison between the readout of the angle encoder on the grating axis and that of the autocollimator is shown in Fig. 4. In order to reveal backlash, stick-slip effects etc., the grating drive has been moved with highest resolution starting from its resting point. The new angle encoder unambiguously displays the expected linear relation. (0.1 arcsec at the encoder corresponds to 0.2 arcsec at the autocollimator since the autocollimator reading counts twice the actual rotation of the grating.) The observed broadening of the line is caused by the total noise of the autocollimator. The noise level of the angle encoder is about one order of magnitude lower, as can be also seen from Fig. 4
. The readout of the linear encoder was recorded simultaneously and is displayed in Fig. 5
. In contrast to the data of the angle encoder, the result is not an unambiguous straight line. The graph of the angle positions calculated from the readout of the linear encoder at the lever arm and of the autocollimator readout should give, at least for the small scan range, a line with the same slope as indicated in Figs. 4
and 5
. What is observed at the start of the scan is most likely an elastic deformation of the mechanics between the two different encoders. The scan range displayed is fairly small; however, the comparison between the two curves clearly shows the advantage of the UHV angle encoder being directly mounted on the rotation axis. The measuring position of the linear encoder outside the vacuum vessel is too far away from the relevant optical elements. Because of the unambiguous and precise response of the angle encoders, a closed-loop control system between the angle encoders of the optical elements and the corresponding drive systems has been installed and successfully tested during recent months. This system allows control of the optical elements with sub-0.1 arcsec precision and 0.01 arcsec resolution. As shown, this is impossible by using the readout of linear encoders.
![]() | Figure 4 Comparison between the readout of the RON905-UHV on the grating axis and that of a high-precision autocollimator. |
![]() | Figure 5 Comparison between the readout of the linear encoder which has been attached outside the vacuum vessel and a high-precision autocollimator. The data, calculated in arcsec* for a better comparison, were taken simultaneously to the data displayed in Fig. 4 ![]() |
4. Conclusions
New UHV-compatible high-precision angle encoders have been built. By using these it has been experimentally verified that the mechanical positioning precision required for today's almost perfect plane and spherical optical elements is achievable. The new RON905-UHV encoders, which are now commercially available, are key components of our new set of high-resolution grating monochromators at the undulator beamlines of BESSY II.
References
Senf, F., Lammert, H., Flechsig, U., Zeschke, T. & Peatman, W. B. (1995). Rev. Sci. Instrum. 66, 2154–2156. CrossRef CAS Web of Science
© International Union of Crystallography. Prior permission is not required to reproduce short quotations, tables and figures from this article, provided the original authors and source are cited. For more information, click here.