short communications
A synchrotron radiation beam-position monitor at the Taiwan Light Source
aSynchrotron Radiation Research Center, No.1 R&D Road VI, Hsinchu Science-Based Industrial Park, Hsinchu 30077, Taiwan, and bDepartment of Nuclear Science, National Tsing-Hua University, Hsinchu 30043, Taiwan
*Correspondence e-mail: jrchen@alpha1.srrc.gov.tw
A prototype photon-beam-position monitor has been designed, fabricated and tested at the Taiwan Light Source of the Synchrotron Radiation Research Center. Aluminium was chosen as the material of the blade electrodes due to its low in situ. Results of measurements for bending-magnet light and undulator light with different gaps are described.
and high The resolution of this photon-beam-position monitor was <±1 µm. The sensitivity of the blade electrode has been measuredKeywords: beam-position monitors.
1. Introduction
Photon-beam-position monitors (PBPM) have been widely studied at many synchrotron radiation facilities (Aoyagi, 1996; Elleaume, 1987; Loyer, 1994; Mitsuhashi et al., 1992; Mortazavi et al., 1986; Proceedings of the Workshop on X-ray Position Monitors, 1993; Warwick et al., 1995; Zhang et al., 1995). The blade-type PBPM is the most promising one to meet the stringent performance requirements. However, several problems, such as contamination from the light emitted by bending magnets, different responses due to different undulator gaps, cross-talk effect among blade electrodes, thermal effects etc., have to be considered when a blade-type PBPM is employed in an undulator beamline.
There are several undulators in commission or under construction at the Taiwan Light Source (TLS) (SRRC Annual Report, 1995; Wang et al., 1996). To keep the electron beam orbit constant during the operation of undulators, good BPMs (either electron or photon BPMs) and an effective feedback system are necessary. Due to the limitation of the existing electron BPMs, a prototype photon BPM has been developed. Several tests have been carried out. The experiments and the results are described in the following sections.
2. Experiments
2.1. PBPM design considerations
Fig. 1 shows schematic diagrams of the PBPM structure. Two blades (A), each with a thickness of 0.5 mm, were used as the photoemissive elements. Aluminium was chosen as the blade material due to its low and high These two blades were separated by ∼8 cm and were able to be electrically biased in order to reduce the cross-talk effect or to increase the sensitivity. No deformation or melting was observed on the blades after the irradiation of undulator light.
In order to reduce contamination from the scattered photoelectrons, two mask plates (B and C) were added to form entrance and exit windows of the PBPM. A biased photoelectron-collector (D), located at one side of the PBPM, was used to collect the scattered photoelectrons.
The aluminium blades were fixed to a cooled aluminium block (E) with ceramic insulators (F) for electrical insulation. The aluminium block was mounted on an XY stage driven by stepping motors. The whole PBPM unit was installed in a chamber which could be adjusted in the yaw direction for alignment purposes.
2.2. Experimental system
Three PBPMs were installed in the front-end of the U-10 undulator beamline. These three PBPMs were located at about 7.5 m, 9.0 m and 10.3 m from the middle of the undulator. An aluminium plate with a slit opening of 1 mm (H) × 20 mm (W) was mounted on a motor-driven unit, which was installed in the PBPM1 chamber and located right in front of PBPM1. The slit was used to provide a defined photon beam of height 1 mm for in situ photoemission sensitivity measurements of the blades. In order to have a softer radiation irradiated on the blades during the measurement, the slit was installed about 3 mm above the central plane.
3. Results and discussions
3.1. Calibration
The photon beam position, y, was calculated from the blade currents by using the formula
where Ky is the proportional constant, IU and IL are the photoemission currents of the upper (U) blade and the lower (L) blade, α and β are the photoemission sensitivities of the upper and the lower blades, and γ = β/α. By changing the vertical position of the PBPM, the Ky value could be derived from the relationship of the blade currents (IU and IL) and the PBPM position (y).
The sensitivity test result was obtained by scanning the upper and lower blades of the PBPM across the photon beam (1 mm in height) which penetrated through the slit. Fig. 2 shows a typical result of the sensitivity measurements. The ratio, γ, was defined as the ratio of the area under the two peaks. In Fig. 2 it was about 0.7.
3.2. Effects of bias voltages on Ky and γ
It was observed from the results that the effects of the bias voltages (of the blade and the collector) on Ky and γ were small if suitable bias voltages were chosen. It was also noted that the Ky value with a blade bias at 0 V is almost the same as the `saturation' value as long as the collector bias voltage is large enough (e.g. ≥200 V). In this work, however, −24 V and +200 V were chosen as the bias voltages for the blade and the collector, respectively, due to a higher stability of Ky and γ at these voltages.
3.3. Effects of the horizontal position and the angular misalignment on Ky and γ
An experiment was carried out to study the effects of the misalignment in the yaw direction on the Ky and γ values. The results showed that the variation of both values was ≤1% within an angular (yaw) range of ±0.5°.
Fig. 3 shows the effect of the PBPM horizontal position on the Ky value. It is shown that the Ky value increased as the horizontal position moved outwards from the storage ring. The increase of Ky might be due to the increase of the vertical beam size of the photon beam. The decrease in photon beam (increase beam size) could decrease the conversion sensitivity of the blade currents to the PBPM position (increase the proportional constant Ky). As the horizontal position moved outwards from the storage ring, the contamination from the upstream bending-magnet light increased. This effect could increase the total beam size. The variation of the effect of the horizontal position on the γ factor was about ±8%. Although this value is not large, it is still significant. The variations in the γ value might be due to the variations of the photon energy, the photon spatial distribution or some other factors.
From the experimental results it was observed that both the Ky and γ values were sensitive to the horizontal position of the PBPM. It seemed that there was a systematic factor governing this phenomenon.
3.4. Effect of the undulator gap on Ky and γ
It is shown in Fig. 3 that the Ky value increased as the undulator gap increased except for the case at a full open gap of 220 mm. The increase of Ky might be attributed to the larger beam size or other unknown reasons.
Fig. 4 shows the relationship of γ versus the undulator gap. The γ value increased from 0.67 at 25 mm gap to 0.81 at 220 mm full open gap. The increase of γ might be due to a systematic changing of the photon beam parameters, or some other unknown reasons.
3.5. Index of uncertainty of the measurement
There are many factors contributing to the uncertainty of the PBPM measurement results. In order to study the confidence level and to find out the factors that cause the uncertainty, three sets of PBPMs were used in this work.
Fig. 5 shows the results of measurements with three PBPMs. In Fig. 5(a) there are three curves corresponding to beam positions measured by the related PBPMs. It can be seen that the values of the three curves are in good coincidence in both the trend and the microfluctuation. This implies good reliability of the measurement system. In Fig. 5(b) the three curves show the angles derived from the beam positions of the three PBPMs. Since there was no optical element in the PBPM region, the three measured angles should in principle be exactly the same. In our measurements, although the trends of the three angle curves were similar, the values of these three angle curves were different from each other. The variations in the PBPM structure, the Ky factor, the γ factor, the electron beam quality or other factors might induce the differences in the values of these three angle curves.
Although there were discrepancies among the values of the three angles, the differences were not large (e.g. ≤4 µrad in Fig. 5). Furthermore, it was observed that the drifts in the angle curves were small. A result of ≤±1 µrad h−1 drift was obtained (this value depended on the stability of the electron beam orbit).
The results of this work showed that the position resolution of the PBPM was better than ±1 µm and the drift was better than ±1 µm h−1. The index of uncertainty, evaluated from the deviations of the three angles, was estimated to be <±3 µrad without changing the undulator gap. The index of uncertainty increased to about ±70 µrad as the undulator gap changed from 25 mm to 65 mm. More studies are necessary for further improvements.
4. Summary
A two-blade PBPM was designed and tested at the TLS. Aluminium was chosen as the blade material due to its high
and low atomic number.A slit located upstream of the beamline was installed. With such a design, the photoemission sensitivity of the blades could be measured in situ.
The effects of several factors, such as bias voltage, transverse and angular misalignment, and different undulator gaps on the Ky and γ factors, were studied. No significant effect (<2%) was found due to small bias voltage variation or angular misalignment. However, the change in both horizontal position and undulator gap had a significant effect on the Ky and γ parameters. The deviation was about ±10% in the undulator gaps of interest and the PBPM horizontal positions.
A system with three PBPMs was set up in this work. With such a design, data on three beam positions and three beam angles were obtained. The uncertainties of the measurements were evaluated by comparing these data. A position resolution of ±1 µm and a drift in angle of ±1 µrad h−1 were obtained. The index of uncertainty of the measurements in this work was estimated to be <±3 µrad without changing the undulator gap. In the case of the undulator gap varying from 25 mm to 65 mm, the index of uncertainty degraded to about ±70 µrad.
References
Aoyagi, H. (1996). Third Joint APS-ESRF-Spring-8 Workshop, April 16. Himej, Japan.
Elleaume, P. (1987). ESRF Report, ESRF-Synchrotron Radiation/ID-87–11. ESRF, Grenoble, France.
Loyer, F. (1994). ESRF Report, ESRF/Mach DIAG 94/22. ESRF, Grenoble, France.
Mitsuhashi, T., Ueda, A. & Katsura, T. (1992). Rev. Sci. Instrum. 63, 534–537. CrossRef Web of Science
Mortazavi, P., Woodle, M., Rarback, H., Shu, D. & Howells, M. (1986). Nucl. Instrum. Methods, A246, 389–393. CrossRef CAS Web of Science
Proceedings of the Workshop on X-ray Position Monitors (1993). ESRF, Grenoble, 22–23 November.
SRRC Annual Report (1995). SRRC, Hsinchu, Japan.
Wang, Ch., Chang, L. H., Chang, C. H., Chen, H. H., Fan, T. C., Hsu, K. T., Hsu, J. Y., Hwang, C. S., Lin, M. C. & Pan, K. T. (1996). Proceedings of the European Particle Accelerator Conference, EPAC96, Barcelona, June 10–14. Bristol: Institute of Physics.
Warwick, T., Andresen, N., Portmann, G. & Jackson, A. (1995). Rev. Sci. Instrum. 66, 1984–1986. CrossRef CAS Web of Science
Zhang, X., Sugiyama, H., Ando, M. & Shiwaku, H. (1995). Rev. Sci. Instrum. 66, 1990–1992. CrossRef CAS Web of Science
© International Union of Crystallography. Prior permission is not required to reproduce short quotations, tables and figures from this article, provided the original authors and source are cited. For more information, click here.