A grazing-incidence reflectometer for BL-39XU at SPring-8
aNational Research Institute for Metals, 1-2-1 Sengen, Tsukuba, Ibaraki 305, Japan, bKohzu Seiki Co. Ltd, 2-27-37 Mishuku, Setagaya, Tokyo 154, Japan, and cJASRI, SPring-8, Kamigori, Ako-gun, Hyogo 678-12, Japan
*Correspondence e-mail: sakurai@yuhgiri.nrim.go.jp
A new grazing-incidence reflectometer has been designed for the X-ray undulator beamline BL-39XU at SPring-8. It will be used for analytical applications, especially ultra-trace-level determination by total reflection
(TXRF), nanometer-scale surface topography by specular/non-specular X-ray reflection, and thin-film interface studies by combined measurements of and X-ray reflection. The requirements for the instrument and the solutions in the present design are discussed. A prototype has been developed to test the feasibility of the design. It has been confirmed that the measurements planned at the beamline are achievable with this reflectometer.Keywords: instrumentation; total reflection; grazing incidence; X-ray fluorescence; TXRF; specular reflection; non-specular reflection; diffuse scattering; ultra-trace-level analysis; surface topography; interface study.
1. Introduction
The recent remarkable progress in total reflection ; Stoev & Sakurai, 1998). One of the most widely known TXRF applications is the evaluation of of semiconductor wafers; according to recent reports, trace determination of the order of 109–108 atoms cm−2 is now possible (Laderman et al., 1995; Wobrauschek et al., 1995). The chemical characterization of trace metals in a small sample, such as a drop of liquid, is also important in biological and environmental analysis (Sakurai, Ida & Shintani, 1997). Chemical shifts of absorption edges are detected by measuring the fluorescence intensity as a function of the incident X-ray energy. A mirror is used as a sample support to improve the signal-to-background ratio, and the chemical state of diluted metals (~0.1 mM) in a drop of liquid (~1 µl) can be determined. Both X-ray specular and non-specular (diffuse) reflection are powerful tools in nanometer-scale surface topography (Bowen & Tanner, 1993; de Boer, 1996), and when combined with angle-resolved TXRF measurements they form a unique probe for interface studies by means of the interference effect caused by multiple reflection of X-rays (Sakurai & Iida, 1997).
(TXRF) and grazing-incidence X-ray reflectometry is significant for materials analysis, especially for ultra trace level determination and the characterization of thin films (Schwenke & Knoth, 1993Third-generation synchrotron radiation, especially the extremely high ) at SPring-8 is one of ten public beamlines completed for initial operation in October 1997. This X-ray undulator beamline is equipped with an Si(111) rotated-inclined double-crystal monochromator and a Pt-coated mirror with a vertical axis. We have designed a new X-ray reflectometer for this beamline as part of the activities of the SPring-8 spectrochemical analysis collaborating group. The group is responsible for the development and installation of two different instruments: an spectrometer for micro-imaging (Hayakawa et al., 1997) and this grazing-incidence X-ray reflectometer. In this paper, the detailed design of the reflectometer and some results of the feasibility tests on a prototype are reported.
available from an undulator, is expected to change the basic principles of sample characterization and result in novel techniques and instruments, besides increasing the potential of conventional analytical methods. The BL-39XU beamline (Japan Synchrotron Radiation Research Institute, 19972. Design considerations
The design policy for the reflectometer is based on the scientific programme planned at the beamline. So far, a number of highly sophisticated instruments have been developed for grazing-incidence X-ray experiments, especially for surface diffraction techniques. Some have been equipped with many stages for multi-purpose measurements and others have been designed for use in an ultra-high vacuum to study well defined surfaces. The present reflectometer is a much simpler, but nevertheless carefully designed, instrument for ultra trace level analysis and realistic surface/interface study by combined measurements of TXRF and specular/non-specular X-ray reflection.
Figs. 1 and 2 show the X-ray reflectometer schematically. A high degree of accuracy and precision in angular scanning are essential. In the present design, a single-axis goniometer [(2) in Figs. 1 and 2] with 0.005 arcsec resolution (Ishikawa et al., 1991) is used to rotate the sample. The resolution is chosen for conventional X-ray reflection/scattering experiments and for future grazing-incidence X-ray interference experiments using a partially coherent beam. For detector (6) scanning, an independent translational stage (8) is used instead of the second axis of the goniometer. Since the angle of incidence is shallow (typically of the order of mrad) straight motion gives exact angular change and high resolution is easily achievable because of the rather long distance from the goniometer centre (250–500 mm). The distance is controlled by the stage (9) and can be read by a magnetic scale. A crystal analyser is not employed to enhance the resolution because a distance of 500 mm can provide good enough resolution for our experiments. Precise alignment capability is extremely important for the grazing-incidence X-ray reflectometer; the goniometer rotation centre and sample surface must be set on the exact centre of the incident beam. Two linear stages with 0.25 µm resolution are used for this alignment; one, (3), for positioning the goniometer and the other, (4), for the sample (1). Although the automatic scan range of the goniometer is limited to 6°, it is designed to allow manual 180° rotation. The beam centre can be detected easily, e.g. by scanning a very thin wire, and the rotation centre position can be corrected so that the difference between the 0 and 180° positions becomes zero. For reliable positioning of the sample surface, a direct beam stopper [a knife edge, (7)] is required just in front of the reflection detector. At maximum reflection intensity the sample surface is exactly on the beam centre and the angle of incidence is just below the critical angle.
This reflectometer is also intended for ultra trace level analysis by TXRF. It is important to obtain a good signal-to-scattering-background ratio and also to reduce fluorescent X-rays from the environment (outside the sample). In the present design, a low-pressure chamber (12) is prepared to reduce air scattering and to avoid contamination from airborne particles. An oil-free pump is used for evacuation. A Teflon surface finish is added to every part around the sample and the ), but vertical axis geometry (Fig. 2) is typically advantageous in the analysis of semiconductor wafers, as it makes full use of the linear polarization of undulator radiation to reduce the scattering further. In horizontal axis geometry, the fluorescence detector is usually set at 10° inclination (`side view' geometry) as shown in Fig. 1(c), but a `top view' layout using a direction-convertible detector is also possible. This is sometimes important when collecting fluorescence signals from extremely small particles on the substrate. The upper part of the chamber can be removed so that different chambers or equipment can be connected. Initially, the roof is a transparent plastic plate to allow internal inspection of the chamber. It is also possible to introduce gas (He, dry N2 etc.) for biological and materials research.
detector [(10) or (11)]. The sample holder is fixed on a small goniometer which allows manual orientation adjustment to avoid possible diffraction by a single-crystal sample. The distance between the sample and the detector (10) can be changed from 0 to 345 mm. The detector is equipped with a collimator to limit the viewing area, which is useful in avoiding the influence of scattering at the edges of the sample. It is also essential to be able to vary the sample orientation by switching the rotation axis between horizontal and vertical, as wet, liquid or small samples need horizontal axis geometry (Fig. 1In the present X-ray reflectometer, a high-resolution et al., 1997) of a crystal analyser may be used instead. However, for X-ray reflection a wide is important and an or an APD detector is used. The incident intensity is monitored by an [(5) in Figs. 1 and 2].
detector is needed. Initially, an Si(Li) detector was used, but in the future a superconductor tunnelling junction detector (Kurakado3. The prototype and feasibility tests
The prototype reflectometer was developed at the National Research Institute for Metals, Tsukuba, Japan. No vacuum chamber was prepared because the main aim was to test the mechanical motion of the reflectometer and the precision and feasibility of real measurements. Only the vertical axis geometry (see Fig. 2) was tested with the Si(Li) detector in the `side view' position, but, in principle, the specification of the reflectometer is very close to the design for the beamline discussed above. The alignment procedure was as follows. The reflectometer was connected to a laboratory X-ray source (45 kV, 80 mA, 30 µm × 3 mm focus) and an Si(111) channel-cut monochromator was installed on the tube wall. The size of the entrance and receiving slits was 40 µm (width) × 10 mm (height). The distance between the goniometer centre and the receiving slit was set to 285 mm. A YAP:Ce was used for X-ray reflection measurements.
Fig. 3 shows the specular and non-specular (diffuse) reflection from synthetic quartz, which is commercially available as mirror material (λ/20, λ = 632.8 nm; 2 arcsec). While specular reflectivity is sensitive to changes in the electron density with depth, non-specular reflection provides information in the lateral direction. Every scan was carried out successfully, and it was found that the experimental data obtained give reasonable surface parameters such as roughness, correlation length and fractal dimension (Sakurai & Stoev, 1998). In this case, the surface is expected to be randomly rough, but the step structure and initial or intermediate stages of island growth of thin films constitute other interesting models. Furthermore, when the experiment is extended to thin films, the most interesting and important target is the interface, and different models are required for analysis. The prototype reflectometer worked well in preliminary studies.
The advantage of an undulator radiation beamline for TXRF measurements is clear. By using the interference effects in the X-ray fluorescence intensity, it is possible to enhance the segregated trace impurities localized at a specific interface (Sakurai & Iida, 1997); this will be one of the main future programmes at the beamline. Another interesting possibility is the separation of true roughness (physical shape) and the degree of chemical gradation for each interface. Although it is difficult to distinguish them by only X-ray specular reflection, which detects changes in electron density, the combined measurement of non-specular X-ray reflection and fluorescent X-rays from the interface is promising as non-specular X-ray reflection is observed when the interface is physically rough, and does not depend on chemical gradation (Sinha et al., 1988), while is sensitive to chemical composition, even at the trace level. When measurements were attempted using the limited intensity of a laboratory source (Sakurai & Stoev, 1998), it was confirmed that the present design fits well with the planned experiments.
4. Conclusions
A grazing-incidence X-ray reflectometer has been designed for analytical use at BL-39XU, SPring-8. The feasibility of the design has been successfully confirmed by testing the prototype. We are now considering additional designs for
and transfer. Since ultra trace level element studies are the main objectives of this beamline, possible contamination prior to the measurements can be a big problem. To achieve much better detection limits than with a second-generation synchrotron radiation source, a clean, versatile instrument will be needed. Controlling small quantities of liquid drop samples is also essential and we plan to install an instrument for producing drops inside the chamber. One of the advantages of the present reflectometer is that it is relatively easy to add or change equipment to suit each experiment.Acknowledgements
The authors are grateful to all members of the SPring-8 spectrochemical collaboration group for valuable discussions. Our thanks also go to Dr K. N. Stoev and Dr L. Orteaga at the National Research Institute for Metals for useful comments. Part of the work, including the development of the reflectometer prototype, received financial support from the Science and Technology Agency, Japan, through the nuclear energy research program.
References
Boer, D. K. G. de (1996). Phys. Rev. B, 53, 6048–6064. CrossRef Web of Science
Bowen, D. K. & Tanner, B. K. (1993). Nanotechnology, 4, 175–182. CrossRef CAS
Hayakawa, S., Goto, S., Shoji, T., Yamada, E. & Gohshi, Y. (1997). J. Synchrotron Rad. 5, 1114–1116. Web of Science CrossRef IUCr Journals
Ishikawa, T., Hirano, K. & Kikuta, S. (1991). Nucl. Instrum. Methods, A308, 356–362. CAS
Japan Synchrotron Radiation Research Institute (1997). SPring-8 Beamline Handbook. Version 1.1, p. 20.
Kurakado, M., Ohsawa, D., Katano, R., Ito, S. & Isozumi, Y. (1997). Rev. Sci. Instrum. 68, 3685–3696. CrossRef CAS Web of Science
Laderman, S. S., Fischer-Colbrie, A., Shimazaki, A., Miyazaki, K., Brennan, S., Takaura, N., Pianetta, P. & Kortright, J. B. (1995). Anal. Sci. 11, 515–518. CrossRef CAS Web of Science
Sakurai, K. & Iida, A. (1997). Adv. X-ray Anal. 39, 695–700. CAS
Sakurai, K., Iida, A. & Shintani, H. (1997). J. Phys. IV (France), C2, 713–714.
Sakurai, K. & Stoev, K. (1998). In preparation.
Schwenke, H. & Knoth, J. (1993). Handbook of X-ray Spectrometry, edited by R. E. Van Grieken & A. A. Markowicz, p. 464. New York: Marcel Dekker.
Sinha, S. K., Sirota, E. B., Garoff, S. & Stanley, H. B. (1988). Phys. Rev. B, 38, 2297–2311. CrossRef CAS Web of Science
Stoev, K. N. & Sakurai, K. (1998). Spectrochim. Acta B. In the press.
Wobrauschek, P., Kregsamer, P., Ladisich, W., Streli, Ch., Pahle, S., Fabry, L., Garbe, S., Haller, M., Knöchel, A. & Radtke, M. (1995). Nucl. Instrum. Methods, A363, 619–620. CrossRef Web of Science
© International Union of Crystallography. Prior permission is not required to reproduce short quotations, tables and figures from this article, provided the original authors and source are cited. For more information, click here.