short communications
Development of a compact six-circle goniometer
aElectrotechnical Laboratory, 1-1-4 Umezono, Tsukuba, Ibaraki 305, Japan
*Correspondence e-mail: oyanagi@etl.go.jp
A compact versatile six-axis goniometer has been built which can be used with the ω, 2θ, φ and χ motions and a two-circle goniometer on the 2θ arm for mounting the detector and analyser crystal. The goniometer is controlled via Ethernet, and standard data-acquisition software with a graphical user interface has been developed. A closed-cycle helium cryostat is mounted on the ω axis for temperature-dependent experiments from 330 K to 14 K. Preliminary experiments on a Bi2Sr2CaCu2O8 single crystal demonstrate the performance. peaks due to the one-dimensional spatial modulations in the b axis were recorded as a function of temperature above 20 K using an image plate.
vertical or horizontal for anomalous X-ray diffraction and diffraction X-ray absorption fine structure using a tunable X-ray undulator at SPring-8. The diffractometer consists of a four-circle goniometer for the standardKeywords: goniometers; XRD; DAFS; X-ray undulators; SPring-8.
1. Introduction
Recently, structural studies using synchrotron radiation have often required the combined use of X-ray scattering and X-ray absorption, or the simultaneous use of these two techniques, i.e. diffraction anomalous fine structure (DAFS) (Stragier et al., 1992). For instance, in studies of high-Tc superconductors, where the CuO2 plane shows a local deviation from the crystallographic structure (Toby et al., 1990) or a spatial modulation (Yamamoto et al., 1990), both local and superstructural information are necessary for understanding the lattice effects on high-Tc superconductivity. In a recently proposed model the two components, locally distorted (localized) and undistorted (itinerant) domains, form a spatially modulated two-dimensional which is related to the enhancement of the Tc (Bianconi, Saini, Lanzara et al., 1996; Bianconi, Saini, Rosetti et al., 1996). For single-crystal samples, X-ray absorption and X-ray (Bianconi, Lusignoli et al., 1996) provide the ratio of locally distorted to undistorted components and the modulation period, respectively. Combining these techniques, the width and frequency of the stripe structure are obtained.
It is desirable to measure both X-ray absorption fine structure (XAFS) in a fluorescence mode and 2Sr2CaCu2O8 (Bi2212) at various temperatures from 300 K down to 14 K.
in the same experimental hutch. We have designed and built a compact Huber six-circle goniometer for joint experiments of and X-ray or DAFS so that the two apparatuses are arranged in tandem at the undulator beamline XU-10 of SPring-8. In this report, we describe the fundamental specifications of the compact goniometer and evaluate its performance by observing the incommensurate modulation of Bi2. Diffractometer
2.1. Goniometer
We have built a small six-axis goniometer which can be used with the ω, 2θ, φ and χ motions, and a two-circle goniometer (Huber 408 and 409) on the 2θ arm to mount the detector and analyser crystal. Since we plan to install the goniometer in the same hutch with an apparatus at the beamline XU-10 of SPring-8, efforts have been made to reduce the dimensions as much as possible yet to retain the precision required for a four-circle diffractometer. The inner and outer diameters of the χ circle are 250 mm and 340 mm, respectively, and the length of the 2θ arm is 550 mm. The angular resolution is 0.0001° for the ω, 2θ and ωA axes, 0.001° for the χ axis, and 0.002° for the φ and 2θA axes.
vertical or horizontal for use with a tunable X-ray undulator. The diffractometer consists of a four-circle goniometer (Huber 424 with a 511.1 Eulerian cradle) for the standardFig. 1 shows a schematic illustration of the goniometer in a cradle for use in a vertical scattering mode. The left figure shows the front view of the goniometer looking upstream through the χ circle. The right-hand side shows the side view perpendicular to the incident X-ray beam. In this figure the X-rays enter from the left and pass through an entrance slit assembly and an ionization-chamber i0 monitor before striking the sample. In a horizontal scattering-mode operation for either in-plane surface diffraction or use with a rotating-anode X-ray generator, the cradle is easily removed. The goniometer can be positioned using an air pad on a thin sheet of stainless steel in horizontal-mode operation.
For use at a synchrotron facility with vertical-mode operation, the six-axis system is mounted on a support stand with x, z and Rx (rotation) motions controlled by three stepping motors (the upper support stand in Fig. 1). The Rx motion rotates the goniometer and cradle in a horizontal plane about an axis which passes through the sample position at the centre of the χ circle. This motion is used for aligning the incident angle for grazing-incidence diffraction (GID) experiments (Feidenhans'l, 1989). The x-motion driving mechanism located underneath the Rx stage translates the goniometer in a direction perpendicular to the incident X-ray beam. The z-motion mechanism lies beneath the x stage and controls the height of the entire assembly. The goniometer can be used at the Photon Factory (beam height 1225 mm) and SPring-8 (beam height 1445 mm). For use at the Photon Factory, only the upper support stand is used, while at SPring-8 both upper and lower support stands are required.
2.2. Detector
Fig. 2 shows a schematic diagram of the beamline optics and Huber control system. The incident X-ray beam intensity is monitored by an with a 140 mm-long electrode filled with dry N2. A Bicron NaI(Tl) is routinely used on the 2θ arm. The output of a single-channel analyser (SCA) is counted by one of two programmable processor units (Labo Equipment, NT2400). The first programmable processor unit has two functions: as a stepping-motor controller for six channels and as a dual timer/scaler. A position-sensitive (PSPC) with an effective length of 150 mm can also be used with two preamplifiers (EG&G ORTEC, Model 142A), two delay line amplifiers (EG&G ORTEC, Model 460) and a position analyser (EG&G ORTEC, Model 464). The second programmable processor unit is used as a multi-channel analyser. A Fuji BAS-III 20 × 25 cm image plate (IP) is used as a two-dimensional detector. Various combinations of adjustable slits and soller slits can be mounted on the 2θ arm. For high 2θ-resolution, a Si(111) analyser crystal is used, and for lower resolution but higher an asymmetric (9° miscut) four-bounce Ge(220) analyser crystal (Crystal Scientific) is available. Ray tracing of the undulator optics (rotated inclined double-crystal monochromator and a double flat mirror) shows that a maximum of 1014 photons s−1 is achievable at the sample position.
2.3. Cryostat
For low-temperature diffraction experiments, a lightweight (3 kg) closed-cycle helium refrigerator (Air Products, Model 201M) can be attached to the Eulerian cradle. It has a cylindrical Be window, aluminized Mylar heat shield and a carbon-fibre vacuum shroud. Good thermal contact is achieved by both conduction and exchange with low-pressure helium gas in the internal shroud. With a Conductus LTC-10 temperature controller the cryostat provides a temperature range from 14 K to 330 K with ±0.1 K stability in most orientations. The temperature can be controlled either from the front panel of the controller or from the computer via an RS-232 connection.
2.4. Data-acquisition system
The goniometer is controlled by a custom-made workstation (Labo Equipment, UX-4100) with one Ethernet cable controlling all six axes and collecting data from the i0 monitor, and PSPC (Fig. 2). For controlling six stepping motors, a programmable processor unit (Labo Equipment, NT2400) and six NIM-based stepping-motor drivers (Labo Equipment, SD-2020) are used. The former has the capability of independently controlling six axes and recording two outputs: the incident-beam intensity and the output. For the PSPC, the position is analysed and recorded by a programmable processor unit acting as a multi-channel analyser (Labo Equipment, NT-2400). The workstation runs either on a UNIX (Solaris 2.4) or Windows NT operating system. We have developed the data-acquisition (DAQ) software starting from a commercial four-circle goniometer control system, Gscan (Labo Equipment), using Microsoft Visual C/C++/Windows NT 3.5.1. The advantage of the present system is that all controllers are connected by Ethernet and a single Ethernet cable is introduced into the hutch. The DAQ software package features a user-friendly graphical user interface (GUI) making it easy to perform single-axis or 2θ/ω scans, two-dimensional maps or hkl scans using a slight modification of the standard formulation (Busing & Levy, 1967) for the different symmetry of this goniometer.
3. Performance
This system has recently been tested at the 27-pole wiggler beamline BL13B of the Photon Factory to measure X-ray diffraction from a Bi2Sr2CaCu2O8 single crystal. This crystal has an incommensurate modulation with a vector of qs = βb* + (1/γ)c* with β ≃ 0.21 and γ = 1. The sample was mounted on the cryostat without the cylindrical Be window, and image-plate data were taken at temperatures of 150, 100, 70, 30 and 15 K. Exposures were made at X-ray wavelengths of 1.4086 Å below the Cu K-shell and 1.3788 Å above the edge. Fig. 3 shows image-plate data taken with the sample at 15 K and an X-ray wavelength of 1.4086 Å. Both spots and diffuse spots can be clearly seen. The indices range from l = 12 to 24 and k = −5 to 4, with h = 0. The data were obtained using the swing-scan mode, in which the φ axis (long axis of the cryostat) was continuously oscillated over a 25° range with a period of 12.5 s for a total exposure time of 500 s.
4. Summary
We describe a compact six-circle diffractometer designed for anomalous X-ray diffraction and DAFS using an X-ray undulator. Despite the small dimensions, the Eulerian cradle can support a lightweight closed-cycle helium cryostat with a high angular resolution. Stepping motors, counters and MCA are controlled by a workstation via Ethernet. This replaces intricate wiring into the experimental hutch by a single Ethernet cable. As a preliminary test, the modulated structure of the CuO2 plane in Bi2Sr2CaCu2O8 single crystal has been investigated at various temperatures in the range 300–15 K. The installation of the goniometer requires a small area (1500 × 1000 mm), which is advantageous when sharing the hutch with other apparatus.
Footnotes
‡Present address: Sharp Microelectronics Technology, 5700 North West Pacific Rim Boulevard, Camas, WA 98607, USA.
Acknowledgements
The authors wish to thank N. L. Saini for collaborating in the performance test. They also express their thanks to P. Fons and Y. Tanabe for help throughout the development of the data-acquisition software.
References
Bianconi, A., Lusignoli, M., Saini, N. L., Bordet, P., Kvick, A. & Radaelli, P. G. (1996), Phys. Rev. B, 54, 4310–4314. CrossRef CAS Web of Science
Bianconi, A., Saini, N. L., Lanzara, A., Missori, M., Rossetti, T., Oyanagi, H., Yamaguchi, H., Oka, K. & Ito, T. (1996). Phys. Rev. Lett. 76, 3412–3415. CrossRef PubMed CAS Web of Science
Bianconi, A., Saini, N. L., Rossetti, T., Lanzara, A., Perali, A., Missori, M., Oyanagi, H., Yamaguchi, H. & Nishihara, Y. (1996). Phys. Rev. B, 54, 12018–12021. CrossRef CAS Web of Science
Busing, W. R. & Levy, H. A. (1967). Acta Cryst. 22, 457–464. CrossRef IUCr Journals Web of Science
Feidenhans'l, R. (1989). Surf. Sci. Rep. 10, 105–188. CrossRef CAS Web of Science
Stragier, H., Cross, J. O., Rehr, J. J., Sorensen, L. B., Bouldin, C. E. & Woicik, J. C. (1992). Phys. Rev. Lett. 69, 3064–3067. CrossRef PubMed CAS Web of Science
Toby, B. H., Egami, T. E., Jorgensen, J. D. & Subramanian, M. A. (1990). Phys. Rev. Lett. 64, 2414–2417. CrossRef PubMed CAS Web of Science
Yamamoto, A., Onoda, M., Takayama-Muromachi, E., Izumi, F., Ishigaki, T. & Asano, H. (1990). Phys. Rev. B, 42, 4228–4239. CrossRef CAS Web of Science
© International Union of Crystallography. Prior permission is not required to reproduce short quotations, tables and figures from this article, provided the original authors and source are cited. For more information, click here.