short communications\(\def\hfill{\hskip 5em}\def\hfil{\hskip 3em}\def\eqno#1{\hfil {#1}}\)

Journal logoJOURNAL OF
SYNCHROTRON
RADIATION
ISSN: 1600-5775

Characterization of windows and filters for coherent X-ray beamlines

aAdvanced Research Laboratory, Hitachi Ltd, Hatoyama, Saitama 350-03, Japan, and bPhoton Factory, KEK, Tsukuba, Ibaraki, 305, Japan
*Correspondence e-mail: suzuki@harl.hitachi.co.jp

(Received 4 August 1997; accepted 22 December 1997)

Beryllium windows and graphite heat absorbers used in coherent X-ray beamlines must be optically flat. If the optical path through the window varies by more than λ/4, the X-ray wave front will be distorted after passing through the window, and the divergence of the X-ray beam may be increased. This reduces the beam coherence. Beryllium, graphite and Kapton films have been tested using ultra-small-angle X-ray scattering. Wave-front distortion was also directly observed by means of phase-contrast X-ray microradiography. The measured increase of angular divergence is about 4 µrad. The wave-front distortion is larger than 2π (optical path difference of λ). These are serious problems for utilizing coherent X-ray beams.

1. Introduction

Beryllium foils are usually used as vacuum windows in hard X-ray synchrotron radiation beamlines. Graphite foils are used as heat absorbers in high-flux beamlines. Thin polymer films, e.g. Kapton, are also sometimes used as He/air windows. Important characteristics of these window and filter materials are high X-ray transmittance and mechanical toughness. Relatively little attention has been paid to the quality of the windows from the point of view of coherent optics, except for the evaluation of beryllium windows by means of Gabor-type holography (Snigirev et al., 1996[Snigirev, A., Snigireva, I., Kohn, I. V. & Kuznetsov, S. M. (1996). Nucl. Instrum. Methods, A370, 634-640.]).

Coherent optics is considered to be one of the important applications at the third-generation synchrotron radiation light sources. The angular spread of an X-ray beam (Δθ) observed at an experimental station is expressed as Δθ ≃ A/L, where A is the transverse source size and L is the distance from the source to the observation point. In third-generation synchrotron radiation sources, the typical source size is A ≃ 0.1 mm, and L is around 50 m. Therefore, Δθ is about 2 µrad (∼0.4 arcsec). These highly collimated coherent beams are, for example, well suited to diffraction-limited microfocusing. However, X-ray beam transportation through the beamline is sometimes a very serious problem. All the optical components of the beamline must have sufficient accuracy for conserving the spatial and temporal coherence of the X-rays.

Undesirable wave-front distortions may be caused by passing through windows and filters having uneven surfaces, as shown in Fig. 1[link]. Increase of the angular divergence of the X-ray beam may also occur due to the distortion of the wave-front. From the point of view of wave optics, it is well known that distortion of the wave-front should be smaller than π/2 (= λ/4 in optical path difference) for diffraction-limited focusing (Rayleigh quarter-wavelength rule). From the point of view of geometrical optics, the angular divergence caused by wave-front distortion should be smaller than the angular spread of the source (Δθ) for conserving the transverse coherence of the light source. When the optical components (Be windows, filters, monochromator etc.) do not fulfil the above conditions, the X-ray wave front may be distorted by transportation through the beamline. The distortion of the wave front causes an angular spread of the transmitted beams.

[Figure 1]
Figure 1
(a) Surface figure of typical beryllium foil (type #1) and (b) schematic diagram of wave-front distortion and angular spread of transmitted X-rays due to surface roughness.

When the window material is uniform in density, the relation between wave front distortion and unevenness of the window may be estimated by

[\Delta\varphi=2\pi\delta\,\Delta t/\lambda,]

where Δφ is the phase difference of the wave front, δ (= 1 − n) is the difference of refractive index between the window material and a vacuum, and Δt is the window thickness error (surface roughness). Assuming the free-electron approximation and Z/A ≃ 0.5 (where Z is the atomic number and A is the atomic weight), δ is expressed by the well known formula (Parrat & Hempstead, 1954[Parrat, L. G. & Hempstead, C. F. (1954). Phys. Rev. 94, 1593.])

[\delta\simeq 1.3\times10^{-6}\rho\lambda^2,]

where ρ (g cm−3) is the density of the window material, and λ (Å) is the X-ray wavelength. An example of the measured surface roughness of a beryllium foil is shown in Fig. 1[link]. The roughness of the Be foil is about 9 µm peak-to-valley (p–v). Considering both sides of one foil, Δt is estimated to be about 18 µm. Therefore, assuming λ = 1 Å and ρ = 1.85 (metallic beryllium), Δφ is about 0.8π. This value is larger than the optical path difference of λ/4. The distortion of the wavefront caused by the X-ray windows cannot be neglected.

Characterization of the window materials has already been performed by Henderson (1995[Henderson, S. (1995). J. Appl. Cryst. 28, 820-826.]) for the purpose of small-angle-scattering experiments. However, the measured scattering angle was limited to larger than 1 mrad at 1.54 Å. Small-angle-scattering measurements with higher angular resolution are needed to discuss the conservation of coherence. Cloetens et al. (1996[Cloetens, P., Barrett, R., Baruchel, J., Guigay, J. & Shlenker, M. (1996). J. Phys. D, 29, 133-146.]) have characterized the angular broadening caused by `random phase plates' using the same method as that used in our experiment. However, the angular resolution was about 50 µrad at λ = 0.7 Å. This angular resolution was insufficient to evaluate the beryllium windows and graphite filters.

It is necessary to characterize the X-ray window/filter materials in terms of conservation of coherence. This paper describes quantitative characterization of these materials, using ultra-small-angle X-ray scattering and phase-contrast X-ray microradiography with an X-ray interferometer.

2. Samples and experimental set-up

Distortion of a wave front can be evaluated quantitatively by two methods: one is measurement of the angular distribution of an X-ray beam transmitted through the window, i.e. a kind of small-angle-scattering measurement, and the other is phase-contrast X-ray microradiography.

The experiments were performed at the experimental station BL-8C2 of the 2.5 GeV storage ring in the Photon Factory, KEK. As shown in Fig. 2[link], we use a Bonse–Hart type ultra-small-angle-scattering measurement system (Bonse & Hart, 1966[Bonse, U. & Hart, M. (1966). Z. Phys. 189, 151-162.]) to measure the small-angle scattering of the X-ray beam through the windows/filters. The experimental set-up is basically the same as that of previous work (Uchida & Suzuki, 1992[Uchida, F. & Suzuki, Y. (1992). Proc. SPIE, 1720, 264-271.]). Synchrotron radiation is monochromated and collimated with an Si 440 channel-cut monochromator, and incident to the specimen. The X-ray beam through the specimen is analysed by an Si 440 crystal. The plane of incidence is set in the horizontal direction to use the π component of synchrotron radiation.

[Figure 2]
Figure 2
Experimental set-up for ultra-small-angle-scattering measurements (Bonse–Hart small-angle camera). The cross section of the X-ray beam incident to the sample is 0.2 mm × 0.2 mm.

We also use a phase-contrast X-ray computed-tomography measurement system developed by Momose (Momose, 1995[Momose, A. (1995). Nucl. Instrum. Methods, A352, 622-628.]; Momose et al., 1998[Momose, A., Takeda, T., Itai, Y., Yoneyama, A. & Hirano, K. (1998). J. Synchrotron Rad. 5, 309-314.]) to measure the X-ray phase-contrast image. The instrument (see Fig. 3[link]) consists of a Laue-Laue-Laue interferometer (Bonse & Hart, 1965[Bonse, U. & Hart, M. (1965). Appl. Phys. Lett. 6, 155-156.]) and an interference-fringe-scanning system with rotatable-plate phase shifter for quantitative two-dimensional phase-difference mapping. The phase distribution of the X-ray beam, φ(x), emerging from the window is connected to the angular distribution F(kx) by a Fourier transformation (Mandel & Wolf, 1995[Mandel, L. & Wolf, E. (1995). Optical Coherence and Quantum Optics. Cambridge University Press.]),

[F(k_x)=\textstyle\int\varphi(x)\exp(-k_xx)\,{\rm d}x,]

where x is the coordinate on the sample surface perpendicular to the incident X-ray beam, and kx is the x component of the wavevector of the scattered wave. kx is expressed as,

[k_x=k_0\sin\theta,]

where k0 = 2π/λ, and θ is the scattering angle, which is related to the spatial period Δx by

[\sin\theta=\lambda/\Delta x.]

Therefore, if the angular resolution is infinitely small, we need not measure the phase image. However, in our experiment, the angular resolution is about 4 µrad at λ = 1.54 Å. This value corresponds to the spatial period of 40 µm. On the other hand, the spatial resolution in phase-contrast radiography is estimated to be less than 40 µm, but not better than 10 µm. Therefore, in this experiment, these two methods give complementary information because of the insufficient angular resolution and lack of spatial resolution.

[Figure 3]
Figure 3
Experimental setup for phase-contrast microradiography (Laue-Laue-Laue interferometer with rotatable-plate phase shifter).

Two kinds of Be windows were tested: one was supplied by NGK Corporation, Japan (Be #1), the other was of IF-1-grade Be foil purchased from Brush–Wellman Inc., USA (Be #2). Be #1 is the type widely used at the Photon Factory beamlines as a vacuum window. Be #2 was used in the Tristan main-ring test beamline (Sugiyama et al., 1997[Sugiyama, H., Zhang, X., Higashi, Y., Arakawa, E. & Ando, M. (1997). To be published.]; Suzuki et al., 1997[Suzuki, Y., Kamijo, N., Tamura, S., Handa, K., Takeuchi, A., Yamamoto, S., Sugiyama, H., Ohsumi, K. & Ando, M. (1997). J. Synchrotron Rad. 4, 60-63.]). The graphite foil was the same as that used in the Tristan main-ring test beamline.

The surface roughness is measured with needle contact apparatus. As shown in Table 1[link], Be #1 has a roughness of 8.9 µm p-v. Be #2 is relatively flat (1.1 µm p-v). The Kapton film is optically flat (0.03 µm p-v). The graphite foil seems to have a fairly smooth surface (1.0 µm p-v).

Table 1
List of samples

Material Thickness Surface roughness (peak-to-valley)
Be metal #1 0.2 mm 8.9 µm
Be metal #2 0.2 mm 1.1 µm
Kapton 0.05 mm 0.03 µm
Graphite 0.1 mm 1.0 µm
†Roughness of this Be foil is specified as JIS-3.2S.

3. Results and discussions

Results of the ultra-small-angle-scattering measurements for a typical combination of windows/filters and the instrumental resolution are shown in Fig. 4[link](a). The sample is a combination of two graphite filters, three Be foils (type #2) and four Kapton films, which is the same as that used in the Tristan main-ring test beamline. The measured angular broadening is about 2 arcsec full width at half-maximum (FWHM) at a wavelength of 1.54 Å. This angular broadening is larger than the instrumental resolution (0.9 arcsec in FWHM). However, the instrumental resolution is not negligibly small so as to derive intrinsic broadening. We estimate the intrinsic angular divergence by deconvoluting the measured small-angle scattering with the measured instrumental function. The measured profiles are considered to consist of strong sharp peaks and weak broad scattering tails. Therefore, for this purpose, Gaussian-type broadening is assumed as

[f(\theta)=a_1\exp(-\theta^2/2\sigma_1^2)+a_2\exp (-\theta^2/2\sigma_2^2),]

and a least-squares fitting is adapted. The first term corresponds to the `strong narrow peak' and the second term represents the `broad weak tail'. The result of the least-squares fitting is shown in Fig. 4[link](b). The angular divergence of the X-ray beam through the windows and filters is 0.91 arcsec (∼4.4 µrad) FWHM, and a broad weak tail is also observed. This angular broadening is not negligible for coherent X-ray optics. Although the origin of the angular broadening is not clear yet, we think that the narrow peak is caused by surface undulation or large porosity and its spatial period is around several tens of µm. The broad tail seems to be due to grain boundaries or other fine structures, and the spatial period is estimated to be a few µm.

[Figure 4]
Figure 4
Small-angle scattering of the X-ray beam passing through the window and filter. (a) Dots represent measured data. The sample is a combination of three Be #2 foils, two graphite foils and four Kapton films. The measured instrumental function is shown by the lower line, and the result of the least-squares curve fitting is shown by the other line. (b) Result of deconvolution from the measured instrumental function. The thinner lines represent the narrow and broad components.

Each component of the windows and filters is investigated to evaluate which component gives the significant effect to angular broadening. Results for Be #1, Be #2, Kapton and graphite films are shown in Figs. 5[link](a)–5[link](d). The instrumental broadening is removed from the measured small-angle-scattering profiles in a similar manner. Results of the least-squares curve fitting are listed in Table 2[link]. The results for Kapton are not shown in the table because the measured angular distribution for Kapton film agrees with the instrumental function within experimental error. Therefore, the Kapton foil is considered to be an optically flat window for hard X-rays. On the contrary, the graphite foil shows the most significant angular broadening of 0.85 arcsec.

Table 2
Results of least-squares fitting for angular broadening

Sample a1 σ1 (arcsec) a2 σ2 (arcsec) FWHM (arcsec)
Be (BW) × 3 + Graphite × 2 + Kapton × 4 1.164 0.3600 0.111 2.864 0.91 (4.4 µrad)
Be (NGK) × 3 2.498 0.1661 0.015 1.469 0.40 (1.9 µrad)
Be (BW) × 3 4.926 0.0762 0.042 2.506 0.18 (0.9 µrad)
Kapton × 6
Graphite × 2 1.270 0.3494 0.068 3.079 0.85 (4.1 µrad)
†The measured profile for Kapton films agrees with the instrumental function.
[Figure 5]
Figure 5
Small-angle scattering of the X-ray beam passing through various windows and filters. (a) Three Be foils (NGK), (b) three Be foils (Brush–Wellman), (c) six Kapton films, (d) two graphite foils. Dots are measured data, and fits are shown by solid lines in (a), (b) and (d). The dashed line in (c) is the measured instrumental function.

Results of phase-contrast imaging of the test samples are shown in Fig. 6[link]; one-dimensional phase-difference profiles are also shown in the figure. The phase-contrast images are taken at an X-ray wavelength of 0.92 Å. All the images are displayed as grey scale of 2π. Although no significant non-uniformities are observed in Be #2 and Kapton foils within experimental errors, the graphite foil shows a phase difference of about 2π p-v, and the Be #1 foil shows a phase perturbation of about 0.7π p-v.

[Figure 6]
Figure 6
Two-dimensional phase images measured with an X-ray interferometer. (a) Be #1 foil (NGK), (b) Be #2 foil (Brush–Wellman), (c) Kapton film, (d) graphite film. One-dimensional phase-difference profiles extracted from the two-dimensional mapping are shown on the phase images.

Peak-to-valley values of phase difference for the Be foils and Kapton film can be explained by the surface figure profiles. Therefore, the phase distortion is considered to be caused mainly by the figure errors (surface roughness) of the films. On the contrary, the graphite foil shows significant wave-front distortion of about 2π in spite of its almost flat surface. It is suggested that this large phase distortion and broad angular divergence may be caused by the porosity of the graphite. Therefore, optical components in coherent optics beamlines cannot be checked only by measurement of surface figures; it is important to investigate them by means of the interferometric method, holography or phase-contrast microscopy, and by means of ultra-small-angle-scattering measurements.

Footnotes

Present address: Japan Synchrotron Radiation Research Institute, SPring-8, Kamigori, Hyogo 678-12, Japan.

References

First citationBonse, U. & Hart, M. (1965). Appl. Phys. Lett. 6, 155–156.  CrossRef Web of Science
First citationBonse, U. & Hart, M. (1966). Z. Phys. 189, 151–162.  CrossRef CAS Web of Science
First citationCloetens, P., Barrett, R., Baruchel, J., Guigay, J. & Shlenker, M. (1996). J. Phys. D, 29, 133–146.  CrossRef CAS Web of Science
First citationHenderson, S. (1995). J. Appl. Cryst. 28, 820–826. CrossRef CAS Web of Science IUCr Journals
First citationMandel, L. & Wolf, E. (1995). Optical Coherence and Quantum Optics. Cambridge University Press.
First citationMomose, A. (1995). Nucl. Instrum. Methods, A352, 622–628.  CrossRef Web of Science
First citationMomose, A., Takeda, T., Itai, Y., Yoneyama, A. & Hirano, K. (1998). J. Synchrotron Rad. 5, 309–314. Web of Science CrossRef CAS IUCr Journals
First citationParrat, L. G. & Hempstead, C. F. (1954). Phys. Rev. 94, 1593.
First citationSnigirev, A., Snigireva, I., Kohn, I. V. & Kuznetsov, S. M. (1996). Nucl. Instrum. Methods, A370, 634–640.  CrossRef Web of Science
First citationSugiyama, H., Zhang, X., Higashi, Y., Arakawa, E. & Ando, M. (1997). To be published.
First citationSuzuki, Y., Kamijo, N., Tamura, S., Handa, K., Takeuchi, A., Yamamoto, S., Sugiyama, H., Ohsumi, K. & Ando, M. (1997). J. Synchrotron Rad. 4, 60–63. CrossRef CAS Web of Science IUCr Journals
First citationUchida, F. & Suzuki, Y. (1992). Proc. SPIE, 1720, 264–271.  CrossRef CAS

© International Union of Crystallography. Prior permission is not required to reproduce short quotations, tables and figures from this article, provided the original authors and source are cited. For more information, click here.

Journal logoJOURNAL OF
SYNCHROTRON
RADIATION
ISSN: 1600-5775
Follow J. Synchrotron Rad.
Sign up for e-alerts
Follow J. Synchrotron Rad. on Twitter
Follow us on facebook
Sign up for RSS feeds