A new high-flux chemical and materials crystallography station at the SRS Daresbury. 1. Design, construction and test results. Corrigendum

aCLRC Daresbury Laboratory, Warrington WA4 4AD, UK, bDepartment of Crystallography, Birkbeck College, Malet Street, London WC1E 7HX, UK, cDepartment of Chemistry, University of Newcastle, Newcastle upon Tyne NE1 7RU, UK, dThe Royal Institution of Great Britain, 21 Albemarle Street, London W1X 4BS, UK, and eDepartment of Physics, University of Aberystwyth, Penglais, Aberystwyth, Ceredigion SY23 3BZ, UK
*Correspondence e-mail: r.j.cernik@dl.ac.uk

(Received 31 August 1999; accepted 4 October 1999)

The scattering efficiencies of four test samples given in Table 1[link] of the original report [Cernik et al. (1997[Cernik, R. J., Clegg, W., Catlow, C. R. A., Bushnell-Wye, G., Flaherty, J. V., Greaves, G. N., Burrows, I., Taylor, D. J., Teat, S. J. & Hamichi, M. (1997). J. Synchrotron Rad. 4, 279-286.]). J. Synchrotron Rad. 4, 279–286 ] were calculated incorrectly. Corrected values are provided; these are two to three orders of magnitude lower.

Table 1
Incorrect and corrected scattering efficiencies (e2 Å3)

Example (1) (2) (3) (4)
Chemical formula C67H56Cl2O5P3Ru2 C18H17NO2 C72H68N4O20S2 Mg0.18Al0.82PO4
Original value 6.3 × 1017 7.0 × 1017 5.1 × 1018 9.9 × 1015
Corrected value 5.5 × 1015 2.0 × 1015 3.2 × 1015 2.4 × 1013

We have previously described a new single-crystal diffraction facility on station 9.8 at the SRS Daresbury (Cernik et al., 1997[Cernik, R. J., Clegg, W., Catlow, C. R. A., Bushnell-Wye, G., Flaherty, J. V., Greaves, G. N., Burrows, I., Taylor, D. J., Teat, S. J. & Hamichi, M. (1997). J. Synchrotron Rad. 4, 279-286.]). Test results were summarized for four samples. The data reported included the scattering efficiency of each test crystal, defined as [\textstyle\sum{f}^2V_{\rm crystal}/V_{\rm cell}^2] (Harding, 1996[Harding, M. M. (1996). J. Synchrotron Rad. 3, 250-259.]), for comparison with other studies.

These efficiencies were incorrectly calculated. They are given, together with the corrected values, in Table 1[link]. The correct values are two to three orders of magnitude lower. This means that the conclusions and comments made on the basis of the incorrect figures were too conservative, the performance of the facility being more impressive than indicated by these original figures. In the meantime, samples with even lower scattering efficiencies have been successfully studied by a number of users, and results will be reported in due course.

References

First citationCernik, R. J., Clegg, W., Catlow, C. R. A., Bushnell-Wye, G., Flaherty, J. V., Greaves, G. N., Burrows, I., Taylor, D. J., Teat, S. J. & Hamichi, M. (1997). J. Synchrotron Rad. 4, 279–286.  CrossRef CAS Web of Science IUCr Journals Google Scholar
First citationHarding, M. M. (1996). J. Synchrotron Rad. 3, 250–259.  CrossRef CAS Web of Science IUCr Journals Google Scholar

© International Union of Crystallography. Prior permission is not required to reproduce short quotations, tables and figures from this article, provided the original authors and source are cited. For more information, click here.

Journal logoJOURNAL OF
SYNCHROTRON
RADIATION
ISSN: 1600-5775
Follow J. Synchrotron Rad.
Sign up for e-alerts
Follow J. Synchrotron Rad. on Twitter
Follow us on facebook
Sign up for RSS feeds