research papers
Precision wavelength measurement of the 14.4 keV Mössbauer photon
aPhoton Factory, High Energy Accelerator Research Organization (KEK), Tsukuba, Ibaraki 305-0801, Japan, bJapan Synchrotron Radiation Research Institute (JASRI), SPring-8, Hyogo 679-5198, Japan, and cDepartment of Applied Physics, School of Engineering, University of Tokyo, Hongo, Bunkyo-ku, Tokyo 113, Japan
*Correspondence e-mail: zhang.xiaowei@kek.jp
The wavelength of the 14.4 keV Mössbauer photon has been determined by using undulator radiation and diffraction of FZ silicon crystals. For the wavelength determination a goniometer equipped with a laser rotary encoder and a sine-bar angle optical interferometer was developed; a temperature-monitoring system and X-ray optics related to the experiment were also developed. The mean wavelength was 0.08602557 nm with an uncertainty of 0.6 p.p.m., derived from 16 measurements of three pieces of FZ Si(840) crystals.
1. Introduction
It is well known that the ratio of the line width to the Kα1 wavelength is of the order of 10−4 in the X-ray range, although it is believed that the central wavelength can be determined to better than one part per million (p.p.m.). From this point of view it is clear that the characteristic X-ray line is not a good reference for a high-precision determination. The lattice parameter of Si plays a main role in the scale of the X-ray range (Cohen & Taylor, 1987), though it is well known that the lattice parameter is not an atomic quantity but a macroscopic quantity obtained from the average value of 1016 unit cells.
The γ-ray lines of the order of 0.1 nm would, in principle, be ideal primary wavelength standards in X-rays because of their narrow line-width, symmetric line-shape and the Mössbauer effect in solids. A feasibility study of 14.4 keV wavelength determination was carried out in the 1960s using a 200 mCi 57Co γ-ray source (Bearden, 1965, 1967). Owing to the low of the γ-ray source, the measurement took a long time; the uncertainty of the determination was 25 p.p.m., poorer than that of the characteristic X-ray lines, a few p.p.m. The results of the experiment showed that, to make a γ-ray standard experimentally feasible, the intensity of the radiation source must be increased by at least a factor of 100. Synchrotron radiation from an undulator can produce a strong Mössbauer photon beam, of which the is thousands of times more intense than any isotope sources (Yamamoto et al., 1993). We are now able to determine the 14.4 keV Mössbauer wavelength.
Compared with the great progress in the ). Although a wavelength-dispersive X-ray interferometer measurement was successfully performed (Appel & Bonse, 1991), owing to the optical-path difference, which depends on both the size of the crystal interferometer and the X-ray diffraction angle, we cannot use this method to measure the X-ray wavelength accurately. The only realistic method is still utilizing the traditional crystal technique (Dumond & Hoyt, 1930). In this paper we describe the experimental instruments and adjustment and report on an absolute measurement of the wavelength of the 14.4 keV Mössbauer photon within an accuracy of 1 p.p.m.
of X-ray sources, there has been no remarkable breakthrough in the X-ray wavelength determination, except absolute lattice parameter measurements by X-ray/optical interferometer technology (Desllates & Henins, 19732. Experimental arrangement and adjustments
A schematic view of the experimental arrangement for the wavelength measurement is shown in Fig. 1. From the left, the third-harmonic radiation from the undulator is monochromated by a beamline monochromator on beamline NE3 (Zhang et al., 1992). Through an and a slit, the beam is succeedingly monochromated to a bandwidth of 6.5 meV at 14.4 keV by a high-resolution monochromator. The operation of the Accumulation Ring at KEK is single-bunch with a revolution of 1.2 µs. Nuclear events can be easily separated from those of electronic scattering using an (APD) with the combination of a time-resolved detection technique. The bandwidth was measured by nuclear resonant from a 57Fe foil, and is shown in Fig. 2. Because the design of the high-resolution monochromator was focused on a high throughput, the beam divergence after the beamline monochromator is about 4 arcsec, which is not suitable for a high-precision angular determination. It is necessary to add a beam collimator to the arrangement. Owing to the action of an Si(840) channel-cut collimator, the beam divergence was reduced to 0.4 arcsec, and the Laue and 840 Bragg diffraction were narrowed to sub-arcsec widths, as shown in Fig. 3. The final part in the illustration of the arrangement is a precision goniometer (Kohzu KTG-10) equipped with a laser rotary encoder (Canon X-1M) and an optical sine-bar small-angle measurement interferometer. Two crossed tilting stages and one rotary table are mounted on the rotary axis of the goniometer for adjusting the posture of an Si crystal. Three pieces of FZ Si(840) crystal were cut to chips of 10 mm × 40 mm, where the partial thickness of crystals for X-ray diffraction is 0.5 mm and the thickness of the mount part of the chips is 1.2 mm. The 840 atomic plane is parallel to the crystal surface within 0.05°, as illustrated in Fig. 4.
The principle of our measurement is based on the Bragg formula; the geometric relationship is shown in Fig. 4. We selected two equivalent planes, 840 and , in one FZ Si crystal piece for X-ray diffraction. This method was proposed by Siddons et al. (1989) and was used in our first nuclear Bragg scattering observation experiment (Kikuta et al., 1991). Because the diffraction angle of 840 for the 14.4 keV photon wavelength is 45.1°, we can find both the Bragg and the Laue diffractions within a small angle range of 0.2°. The advantage of this method is that a small angle can be determined with high precision and high accuracy; the disadvantage is that two tilt adjustments for the two diffractions of one crystal are required, which is more difficult than just tuning one.
Let θL be the diffraction angle for the Laue case, and θB be that for the Bragg case; the 2ω angle in Fig. 4 is given by the following equation, which is the angle we should measure,
After consideration, the X-ray refraction correction, θB = θL + δ/sin 2θL (δ is the index of refraction), the diffraction angle is given by
and the Mössbauer wavelength λ is given by
Here, d840 is the lattice parameter of Si(840), and t is the temperature of the crystal piece.
For perfect alignment of the wavelength measurement system, (i) the rotary axis of the goniometer is set perpendicular to the X-ray beam, and (ii) the 001 direction (a common vector perpendicular to 840 and vectors) of the Si crystal is set parallel to the rotary axis. We narrowed the beam width to 0.3 mm, utilized a Polaroid film on a long co-axis 2θ arm of the goniometer, and checked that the first condition was satisfied within an error of 0.13 mrad. We then adjusted the two tilting stages as follows, allowing the condition (ii) to be satisfied. From Fig. 4 we know that the Laue diffraction angle should reach a maximum with the smallest width when the plane is adjusted to be parallel to the rotary axis, and the Bragg diffraction angle should reach a minimum (consider the sign of the in Fig. 4) for the 840 plane. The curves of the peak shift versus the tilting angle of and 840, in units of mrad, are shown in Fig. 5, and are consistent with the solid geometrical theory. When we adjust one tilting angle of the atomic plane, the diffraction angular position of another atomic plane could be moved, because we cannot expect that the crystal is set completely at the position where the and 840 planes are just perpendicular to their respective tilting stages, even with the help of the rotary table on the tilting stages. Therefore, we need an additional `2ω max-maximum condition' for complete alignment of the crystal. We considered that under this condition the common line of two atomic planes will be parallel to the rotary axis and perpendicular to the X-ray beam. After the process of adjusting two tilting stages, shown in Fig. 5, we measured the 2ω angle versus the tilting angles of 840 and , as shown in Fig. 6. Because the configuration of the 840 Bragg diffraction is a (+,+) arrangement to the collimator, where the wavelength is sensitive to the Bragg angular position, the top of the 2ω curve is located at −15.3 mrad, which was slightly different from that of the Bragg peak at −16.3 mrad shown in Fig. 5. For the (+,−) arrangement of Laue diffraction, where the wavelength versus the diffraction angle is not as sensitive as the (+,+) arrangement, the top position of the 2ω curve was consistent with that of the Laue peak curve shown in the graph.
3. Temperature monitor system and angular measurement equipment
The temperatures of the crystals and the experimental environment were monitored by six platinum and three thermistor sensors. The accuracy of a standard platinum 25 Ω resistance thermometer (Chino/R800-2, serial No. RS96Z-3) is 0.01 K (with an AZONIX controller); other sensors were calibrated by the standard sensor at the melting points of ice and gallium. The temperatures of the high-resolution monochromator, atmosphere and holder of the Si(840) crystal were recorded every 10 s during the experiment. A section of the record chart is shown in Fig. 7. The temperature readings from the standard thermometer and a mini-Pt sensor, which were used to monitor the atmosphere of the crystal, are in between that of the two temperatures of the crystal holder. We considered that the true temperature of the crystal should be a value between them. The temperature fluctuation with an amplitude of 0.01 K and a period of 23 min recorded by the atmosphere monitors was due to the air conditioner in the experimental hall.
We assembled the optical sine-bar angular measurement system by mounting two one-inch cube-corners (Edmund Scientific) on a super-invar frame (Marzolf, 1964), and measured the angular displacement by a commercial laser metric system (HP 5529A dynamic calibrator). The bar length was calibrated in the ±3.5° angular range and 281 measurement points by the X-1M rotary encoder equipped on a precision goniometer. One of the calibration measurements, its sin-fitting curve and the residuals are shown in Fig. 8. The average of the bar length is 140.2030 ± 0.0006 mm (5 p.p.m., 1σ), which corresponds to the accuracy of the X-1M encoder, 1 arcsec in 360°. Using the sine-bar interferometer angular measurement system, we can measure 0.1° with an accuracy of 10 p.p.m. (2σ). When we convert this accuracy into a of 45°, the accuracy of the angular determination is better than 0.1 p.p.m.
4. Discussion of experimental results
One pair of measurements of the 840 Bragg and the Laue electronic diffractions are shown in Fig. 9. Because the intensity of the nuclear signal is not sufficiently strong at beamline NE3, the curves in Fig. 9 were not directly measured by the time-delayed nuclear signal. However, this is not essential for the wavelength determination. Because the bandwidth of the X-ray is sub-p.p.m., which is satisfactory for the precision we need, we also confirmed that the centre of the nuclear resonant signal was identical to that of the electronic signal before and after every measurement circle by rocking the collimator as in the plot in Fig. 10.
Sixteen measurement results obtained from three 840 Si pieces are listed in Table 1. The first column is the crystal code, measurement direction and order; the second column is the angle of 2ω in degrees observed by the sine-bar angular-measurement system. The third column is the Laue diffraction angle obtained by equation (2) with δ/sin2θL = 4.672 × 10−6 (Sasaki, 1989). The fourth column is the temperature of the crystal obtained from the average of two readings of the crystal holder and one reading from the atmosphere. The last column is the wavelength obtained using equation (3) with an Si lattice parameter of 0.543101773 nm (1 atm, 295.5 K; Cohen & Taylor, 1986) and the coefficient of α(t) = 2.56 × 10−6 K−1 (298 K; Okada & Tokumaru, 1984). The average of the 16 measurements is 0.08602557 nm and its standard deviation is 9.5 × 10−9 nm, corresponding to a relative deviation of 0.1 p.p.m., which means that the reproducibility of our measurements was very good.
|
The errors in our measurements were estimated to be as follows, and are listed in Table 2. Because Δλ/λ = Δd/dsinθL + cotθLΔθL, the uncertainty in the wavelength determination arises from two parts. The first part relates to the lattice parameter and temperature of the silicon crystal, contributing to 0.5 p.p.m. in the uncertainty, which is five times larger than the measurement reproducibility. The second part arises from a determination of the diffraction angle. Although there are many factors in this term, as shown in Table 2, we estimated that the uncertainty due to this part is only 0.1 p.p.m., smaller than the first part. The total uncertainty was estimated to be 0.6 p.p.m. (2σ).
|
In conclusion, we have plotted in Fig. 11 the previous and present 14.4 keV wavelength and their uncertainties. The present value is 4.5 p.p.m. larger than Bearden's original (Bearden, 1965, 1967) with the old Å* to Å conversion factor given in 1967, which corresponds to 0.8602516 Å. The recommended conversion factor in 1986 was reduced by a factor of 5.7 p.p.m. compared with the old one. Under the new conversion ratio our result is 10 p.p.m. larger than Bearden's. The present value is 7 p.p.m. lower than Siddons' result (Siddons et al., 1988), measured 11 years ago using an Si() crystal with synchrotron radiation, and near to our early experimental result obtained using beamline 14B, a vertical wiggler beamline on the 2.5 GeV ring of the Photon Factory. The uncertainty of the 14.4 keV wavelength is reduced from Bearden's 25 p.p.m. to the present experiment of 0.6 p.p.m. For higher-precision wavelength measurements we should develop X-ray interferometer technology, where the wavelength in the X-ray range can be directly compared with visible light.
References
Appel, A. & Bonse, U. (1991). Phys. Rev. Lett. 67, 1673–1677. CrossRef PubMed CAS Web of Science Google Scholar
Bearden, J. A. (1965). Phys. Rev. 137, B455–461. CrossRef Web of Science Google Scholar
Bearden, J. A. (1967). Rev. Mod. Phys. 39, 78–124. CrossRef CAS Web of Science Google Scholar
Cohen, R. E. & Taylor, B. N. (1987). Rev. Mod. Phys. 59, 1121–1148. CrossRef CAS Google Scholar
Desllates, R. D. & Henins, A. (1973). Phys. Rev. Lett. 31, 972–976. CAS Google Scholar
Dumond, J. W. & Hoyt, A. (1930). Phys. Rev. 36, 1702–1720. CrossRef CAS Google Scholar
Kikuta, S., Yoda, Y., Kudo, Y., Izumi, K., Ishikawa, T., Suzuki, C. K., Ohno, H., Takei, H., Nakayama, K., Zhang, X. W., Matushita, T., Kishimoto, S. & Ando, M. (1991). J. Appl. Phys. 30, L1686–1688. CrossRef CAS Google Scholar
Marzolf, J. G. S. J. (1964). Rev. Sci. Instrum. 35, 1212–1215. CrossRef Google Scholar
Okada, Y. & Tokumaru, Y. (1984). J. Appl. Phys. 56, 314–320. CrossRef CAS Web of Science Google Scholar
Sasaki, S. (1989). Report 88–14. KEK, Tsukuba, Ibaraki 305, Japan. Google Scholar
Siddons, D. P., Hastings, J. B. & Faigel, G. (1988). Nucl. Instrum. Methods, A266, 329–335. CrossRef CAS Google Scholar
Siddons, D. P., Hastings, J. B., Faigel, G., Grover, J. R., Haustein, P. E. & Berman, L. E. (1989). Rev. Sci. Instrum. 60, 1649. CrossRef Google Scholar
Yamamoto, S., Zhang, X. W., Kitamura, H., Shioya, T., Mochizuki, T., Sugiyama, H., Ando, M., Yoda, Y., Kikuta, S. & Takei, J. (1993). J. Appl. Phys. 74, 500–503. CrossRef CAS Web of Science Google Scholar
Zhang, X. W., Mochizuki, T., Sugiyama, H., Yamamoto, S., Kitamura, H., Shioya, T., Ando, M., Yoda, Y., Ishikawa, T., Kikuta, S. & Suzuki, C. K. (1992). Rev. Sci. Instrum. 63, 404–407. CrossRef Web of Science Google Scholar
© International Union of Crystallography. Prior permission is not required to reproduce short quotations, tables and figures from this article, provided the original authors and source are cited. For more information, click here.