view article

Figure 4
Si (533) analyzer for the Co K-edge. Dashed line: energy spread of the incident beam (FWHM = 169 meV); dashed-dotted line: source size contribution (FWHM = 171 meV); dotted line: intrinsic resolution of the bent crystal calculated by the TT method (FWHM = 50 meV); solid line: total calculated resolution from convolution of all contributions (FWHM = 259 meV); solid line with circles: experimental resolution, FWHM = 297 meV. All curves have been normalized to maximum for better visibility.