Feedback system of a liquid-nitrogen-cooled double-crystal monochromator: design and performances

CROSSMARK_Color_square_no_text.svg

aLaboratoire de Géophysique Interne et Tectonophysique, UMR CNRS/Université Joseph Fourier, 1381 rue de la Piscine, Domaine Universitaire, F-38400 Saint-Martin-D'Hères, France, bCEA/Grenoble, DRFMC/SP2M/NRS, 17 avenue des Martyrs, F-38054 Grenoble CEDEX 9, France, and cLaboratoire de Cristallographie, CNRS, 25 avenue des Martyrs, BP 166, F-38042 Grenoble CEDEX 9, France
*Correspondence e-mail: proux@grenoble.cnrs.fr

(Received 8 July 2005; accepted 14 November 2005)

A new set-up is reported of an indirect cryogenic cooling system for a double-crystal monochromator which runs on the BM30b/FAME beamline at the ESRF (Grenoble, France). This device has been conceived to limit the vibrations on the first diffracting crystal and to maintain it at a constant temperature. These points are crucial for maximizing the beamline stability. Moreover, the relative angular position of the second crystal can be dynamically adjusted by a piezoelectric transducer coupled with a feedback system in order to always be at the maximum of the rocking curve during an X-ray absorption spectroscopy scan. The temperature is stabilized to an accuracy of 0.01°, with two principal consequences. The energy resolution is close to the theoretical value [ΔE/E = 5.6 × 10−6 for Si(220)] and the precision of the energy positioning is extremely good even if the power load changes. A feedback mechanism allows a permanent and automatic optimization of the angle between the two crystals of the monochromator. The intensity of the monochromatic beam remains optimized (i) when the intensity of the electron beam decreases in the storage ring and (ii) during an energy scan.

1. Introduction

FAME is the French absorption spectroscopy beamline for material and environmental sciences at the ESRF (France), one of the four French Collaborating Research Group (CRG) beamlines. FAME is dedicated to X-ray absorption spectroscopy (XAS). A wide range of fields such as environmental and earth sciences, electrochemistry, catalysis, biology and solid-state physics require the extraction of electronic and local structure information of very diluted elements. The stability requirements of low-concentration XAS studies and the operational simplicity of the beamline for the users were the main leading factors in the design of the whole beamline, especially the monochromator, designed and realised by our team.

The double-crystal monochromator can be schematically divided into three parts: the axis, the cooled flat first crystal and the sagitally bent second crystal. The axis was completely characterized before its installation on the beamline (Proux et al., 2005[Proux, O., Biquard, X., Lahera, E., Menthonnex, J.-J., Prat, A., Ulrich, O., Soldo, Y., Trévisson, P., Kapoujyan, G., Perroux, G., Taunier, P., Grand, D., Jeantet, P., Deleglize, M., Roux, J.-P. & Hazemann, J.-L. (2005). Phys. Scr. 115, 970-973.]). The bender of the second crystal was developed in collaboration with the ESRF optics group ten years ago and successfully used on numerous beamlines (Hazemann et al., 1995[Hazemann, J. L., Nayouf, K. & de Bergevin, F. (1995). Nucl. Instrum. Methods B, 97, 547-550.]). The support of this second crystal was carefully designed in order to achieve a perfect decoupling between all the second-crystal movements, and between the relative positions of the first and second crystals during Bragg-angle rotation.

During the conception and commissioning phases of the monochromator, we always kept in mind three factors which were crucial for all the optical elements: (i) the quality of the cooling devices and their thermal stability (Bilderback et al., 2000[Bilderback, D. H., Freund, A. K., Knapp, G. S. & Mills, D. M. (2000). J. Synchrotron Rad. 7, 53-60.]), (ii) the limitations of the vibrations due to the pumping and cooling systems (Hettel, 1989[Hettel, R. (1989). Rev. Sci. Instrum. 60, 1501-1506.]) and (iii) the permanent maximization of the beamline transmittance (especially linked to the optimization of the parallelism between the two crystals). In this paper we will show how we managed to build FAME's monochromator in order to meet these requirements. We will describe the first-crystal cooling principle (§3[link]), the second-crystal feedback adjustment (§4[link]) and the achieved performances (§5[link]).

2. State of the art

2.1. Monochromators cooling principles

Silicon crystal monochromators close to liquid-nitrogen (LN2) temperatures are now commonly used at third-generation synchrotron radiation sources (Bilderback et al., 2000[Bilderback, D. H., Freund, A. K., Knapp, G. S. & Mills, D. M. (2000). J. Synchrotron Rad. 7, 53-60.]; Bilderback, 1986[Bilderback, D. H. (1986). Nucl. Instrum. Methods A, 246, 434-436.]), on insertion-device beamlines but also on bending-magnet beamlines. For temperatures lower than or around 120 K, the thermal expansion of silicon is close to zero and its thermal conductivity (600 W m−1 K−1 at 125 K; Carpentier et al., 2001[Carpentier, P., Rossat, M., Charrault, P., Joly, J., Pirocchi, M., Ferrer, J. L., Kaïkati, O. & Roth, M. (2001). Nucl. Instrum. Methods A, 456, 163-176.]) is slightly higher than that of copper at room temperature (around 450 W m−1 K−1). The thermal distortion of the silicon is then minimized: the lack of a thermal bump leads to an optimization of the energy resolution. The cooling process is commonly achieved by directly cooling the diffracting crystal (Knapp et al., 1995[Knapp, G. S., Rogers, C. S., Beno, M. A., Wiley, C. L., Jennings, G. & Cowan, P. L. (1995). Rev. Sci. Instrum. 66, 2138-2140.]; Rowen et al., 2001[Rowen, M., Peck, J. W. & Rabedeau, T. (2001). Nucl. Instrum. Methods A, 467-468, 400-403.]; Dufresne et al., 2002[Dufresne, E. M., Arms, D. A., Dierker, S. B., Clarke, R., Yacoby, Y., Pitney, J., MacHarrie, B. & Pindak, R. (2002). Rev. Sci. Instrum. 73, 1511-1513.]; Zhang et al., 2003[Zhang, L., Lee, W. K., Wulff, M. & Eybert, L. (2003). J. Synchrotron Rad. 10, 313-319.]) or the heat exchangers which clamp this crystal (Carpentier et al., 2001[Carpentier, P., Rossat, M., Charrault, P., Joly, J., Pirocchi, M., Ferrer, J. L., Kaïkati, O. & Roth, M. (2001). Nucl. Instrum. Methods A, 456, 163-176.]; Chumakov et al., 2004[Chumakov, A., Rüffer, R., Leupold, O., Celse, J. P., Martel, K., Rossat, M. & Wah-Keat, L. (2004). J. Synchrotron Rad. 11, 132-141.]) or that are simply in contact with it through a strain-relieving system (Comin, 1995[Comin, F. (1995). Rev. Sci. Instrum. 66, 2082-2084.]). LN2 cooling for the first crystal was then chosen from the beginning of the design of the beamline. With respect to water-cooled crystals, cryogenic-cooled crystals are not limited in their heat-transfer performance by the material properties (Rogers & Assoufid, 1995[Rogers, C. S. & Assoufid, L. (1995). Rev. Sci. Instrum. 66, 2200-2202.]).

However, various parameters pointed out in the literature led us to choose a slightly different system than those commercially found. In most of the cases, LN2 circulation is pressurized (typically between 0.2 and 1 MPa) to prevent the appearance of vapour nitrogen. The refill of the LN2 tank can increase the pressure inside the closed-loop circulation (around 0.01 MPa) leading to a small angular displacement of the crystal (Dufresne et al., 2002[Dufresne, E. M., Arms, D. A., Dierker, S. B., Clarke, R., Yacoby, Y., Pitney, J., MacHarrie, B. & Pindak, R. (2002). Rev. Sci. Instrum. 73, 1511-1513.]; Alkire et al., 2000[Alkire, R. W., Rosenbaum, G. & Evans, G. (2000). J. Synchrotron Rad. 7, 61-68.]). A second factor has to be checked with care, i.e. the vibrations on the diffracting crystal. These vibrations may result from the circulation of LN2 if the flow is not completely laminar and/or by the nitrogen bubbles created during the heat-exchange process. Finally, the diminution of the heating power to the first crystal, during the decay of the intensity of the electron current in the storage ring, can cause a small decrease in the temperature of the diffracting silicon volume, even if the thermal properties are optimized around 120 K.

2.2. Feedback systems for double-crystal monochromators

Various feedback systems have been detailed in the literature which maximize the output from a double-crystal X-ray monochromator or stabilize it at a given offset from its maximum to improve the harmonic rejection. These systems are suitable for stepper motors (Siddons, 1998[Siddons, D. P. (1998). J. Synchrotron Rad. 5, 685-686.]; Jemian, 2002[Jemian, P. R. (2002). Rev. Sci. Instrum. 73, 1533.]), piezoceramic electrostrictive actuators (Shard et al., 1998[Shard, A. G., Dhanak, V. R. & Smith, A. D. (1998). J. Synchrotron Rad. 5, 829-831.]) or piezoelectric transducers (PZT) (Bridges, 1987[Bridges, F. (1987). Nucl. Instrum. Methods A, 257, 447-450.]; Ramanathan et al., 1988[Ramanathan, M., Engbretson, M. & Montano, P. A. (1988). Nucl. Instrum. Methods A, 266, 471-474.]; Jemian, 2002[Jemian, P. R. (2002). Rev. Sci. Instrum. 73, 1533.]). The optimization of the δθ angle between the two crystals is then achieved by maximizing either the signal proportional to the intensity of the monochromatic beam measured after the monochromator (I0) or the intensity of the transmitted beam (or of the fluorescence current) through (or delivered by) the sample.

In the latter case (Bridges, 1987[Bridges, F. (1987). Nucl. Instrum. Methods A, 257, 447-450.]), the sample vibrates at a frequency above 10 Hz, a lock-in detection extracts the signal and dynamically controls the PZT voltage and so δθ. This kind of adjustment is now used to maximize the intensity of the sample diffracting peaks measured during diffraction anomalous fine-structure experiments (Cross et al., 1998[Cross, J. O., Elam, W. T., Harris, V. G., Kirkland, J.-P., Bouldin, C. E. & Sorensen, L. B. (1998). J. Synchrotron Rad. 5, 911-913.]; Renevier et al., 2003[Renevier, H., Grenier, S., Arnaud, S., Berar, J.-F., Caillot, B., Hodeau, J.-L., Letoublon, A., Proietti, M. G. & Ravel, B. (2003). J. Synchrotron Rad. 10, 435-444.]). The direct optimization of the monochromator transmittance can also be achieved by maximizing the I0 value (Shard et al., 1998[Shard, A. G., Dhanak, V. R. & Smith, A. D. (1998). J. Synchrotron Rad. 5, 829-831.]), the I0 value normalized by the intensity of the electron storage ring (Ramanathan et al., 1988[Ramanathan, M., Engbretson, M. & Montano, P. A. (1988). Nucl. Instrum. Methods A, 266, 471-474.]; Siddons, 1998[Siddons, D. P. (1998). J. Synchrotron Rad. 5, 685-686.]) or that normalized by the intensity of the white beam before the monochromator (Jemian, 2002[Jemian, P. R. (2002). Rev. Sci. Instrum. 73, 1533.]).

3. First-crystal monochromator cooling device

The leading ideas for the monochromator cooling device are: no thermal bump, no vibration and optimal thermal stability. We decided to work in two directions in order to limit the vibrations of the first crystal brought by the LN2 circulation. Firstly, the vibrations are damped before the crystal. Secondly, the vibrations of the LN2 circulation are themselves limited at each point of the device. The stabilization of the first-crystal temperature is performed using a system with a large heat inertia, in order to eliminate the temperature fluctuation due to the fast evolution of the heat power on the sample (displacement of the slits during the alignment procedure, for example). Moreover, to achieve this stabilization, the first crystal can be directly heated to keep its temperature constant.

3.1. Internal cooling device

The internal first-crystal cooling device is shown in Figs. 1[link] and 2[link]. The crystal is indirectly cooled at Tcrystal. The media between the heat exchanger and the crystal serves two purposes: it has good heat-transfer capacity and vibration damping properties. The 110 mm-long heat exchanger is composed of 61 cooling fins (of diameter 3 mm) regularly spaced and carved in the copper block. The interaction surface area is maximized at around 630 cm2. The LN2 flow cross section is strongly expanded, from 7 mm2 before the exchanger (internal diameter of the tube, 3 mm) to 4.3 cm2 inside it. The fluid velocity is then reduced to ensure a laminar flowing mode, limiting turbulent flow vibrations in the exchanger.

[Figure 1]
Figure 1
First-crystal cooling device fixed on the monochromator axis via a thermal bridge (a) and Cu heat-exchanger schematic detail (b, transversal cut).
[Figure 2]
Figure 2
Photograph (a) and drawing (b) of the first-crystal complete block. 1: first Si diffracting crystal. 2: Cu braids. 3: Si support piece. 4: extremity of silicon fingers. 5: fixing Cu L-shaped part. 6: thermocouple. 7: heating resistance. 8: Invar screw. 9: weak point hinge. 10: moving stainless steel piece. 11: fixed stainless steel piece. 12: thermal bridge. 13: main holder. 14: weak points system for the manual positioning of the first-crystal surface with respect to the rotation axis. 15: Invar plate.

The copper braids (item 2 in Fig. 2[link], mean diameter ∼100 µm) are fixed to the silicon support piece (3) using three silicon cylindrical pieces, mechanically maintained with screws (4), the thermal contact being ensured by French alloy (Woods metal, Bi50%Pb25%Sn12.5%Cd12.5%, saturated with indium; 12.5 g of Woods for 8.5 g of In). The diffracting silicon crystal (1) is linked to the support piece by gallium.

The thermal bridge (12) which supports the first-crystal device is composed of three stainless steel tubes to ensure an optimal rigidity linked by an Invar (Fe64%Ni36%) plate (15). The choice of Invar has been made because its contraction between 300 and 80 K (0.016%) is close to the contraction of silicon, 0.023% (Marot et al., 1992[Marot, G., Rossat, M., Freund, A., Joksch, S., Kawata, H., Zhang, L., Ziegler, E., Berman, L., Chapman, D., Hastings, J. B. & Iarocci, M. (1992). Rev. Sci. Instrum. 63, 477-480.]). Moreover, its thermal conductivity (around 8.5 W m−1 K−1 at 115 K) is smaller than that of silicon: the Invar plate works as a thermal screen. The fixation to the thermal bridge is indirect, ensured by a block which can accept the thermal contraction of the silicon support piece. The support is fixed with two Invar screws (8) on a stainless steel piece (10) via a gold foil. This moving stainless steel piece is linked to the fixed one (11) by two symmetric hinges (9). The crystal temperature can be adjusted with two symmetric heating resistances on each crystal side (7) and monitored using a K thermocouple (6) linked to a copper L-shaped part (5). The regulation of the temperature is ensured with a Eurotherm system which controls the power supply. Temperature stabilization is then achieved to an a accuracy of 0.01°.

Copper braids are strongly fixed to the 70 mm-long copper heat exchanger with three M8 screws and via a gallium film in order to ensure a good thermal contact (Fig. 1b[link]). The effective section of the 50 mm-long braids is 6 cm2. An estimation of the power (ΔQCu) dissipated by such a copper device is given by the following relation,

[\Delta{Q}_{\rm{Cu}}=S/l\,\,\left[\Lambda\left(T_{\rm{silicon\,support}}\right)-\Lambda\left(T_{\rm{exchanger}}\right)\right],\eqno(1)]

where S and l are the effective section and the length of the copper braids, respectively, and Λ is the integrated thermal conductivity of copper. Tsilicon support is the temperature of the Cu braids close to the first-crystal Si holder interface. Texchanger equals the LN2 temperature, 77 K. At equilibrium, the total dissipated power (ΔQ) equals that received, coming from both the white X-ray beam (QXray) and the thermal radiation owing to the monochromator's vessel and other elements having a temperature of ∼300 K (Qradiations). With the cooling device not entirely being made of bulk copper, its thermal properties are slightly modified compared with those obtained theoretically from equation (1)[link] for pure copper. Then, one can write

[Q_{\rm{Xray}}=\Delta{Q}-Q_{\rm{radiations}}=\alpha\Delta{Q}_{\rm{Cu}}-Q_{\rm{radiations}},\eqno(2)]

where α is characteristic of the modification of the thermal properties of our device compared with bulk copper.

The `crystal temperature' was measured on the silicon holder with or without the white X-ray beam. The obtained values are reported in Fig. 3[link], with the corresponding QXray power brought by the X-ray beam and calculated using the XOP-shadow code (Dejus & del Rio, 1996[Dejus, R. J. & del Rio, M. S. (1996). Rev. Sci. Instrum. 67, 3356.]). The experimental points were adjusted using the semi-empirical equation (2)[link] via the α and Qradiations parameters. Adjustment gives an α value equal to 0.5 and a Qradiations value of 32 W.

[Figure 3]
Figure 3
Evolution of the heating power owing to the incoming X-ray white beam as a function of the first-crystal temperature. Experimental points and semi-empirical values are shown.

Our complete set-up contributes then to a thermal conductivity degradation of 50% with respect to that of bulk copper. This decrease in the thermal properties is the consequence of the indirect cooling of the first crystal. The Cu braids, used in order to reduce the vibrations, lead to the following interfaces, which introduce thermal resistances on the path from the exchanger to the thermocouple linked to the silicon support:

(i) Cu–Cu interface between the exchanger and the braids, via a gallium film (Fig. 1b[link]),

(ii) Cu–Si interface between the braids and the silicon fingers, via French alloy (elements 2 and 4 in Fig. 2[link]),

(iii) Si–Si interface between the fingers and the support piece, via French alloy (element 3 in Fig. 2[link]),

(iv) Si–Cu interface between the silicon support piece and the fixing copper part for the thermocouple, via an Au foil (element 5 in Fig. 2[link]).

However, even with these reduced thermal properties, the cooling device allows one to work in the optimal temperature range for silicon. Under the usual experimental conditions, the X-ray power is in the 50–60 W range, leading to a first-crystal temperature of ∼105–110 K.

3.2. Internal LN2 circulation

The LN2 flow inside the monochromator is sketched in the inset of Fig. 4[link]. The stainless steel tubes and bellows allow the rotation of the heat exchanger without strain. Two thermal bridges on the monochromator vessel stop the vibrations given by the LN2 flow. Moreover, the use of special bellows (thickness of the wall, 15/100 mm) decreases the transmission of the vibrations.

[Figure 4]
Figure 4
Scheme of the LN2 circulation. 1: LN2 flux from ESRF tank. 2: pneumatic valve. 3: LN2 Dewar at atmospheric pressure. 4: heater for the N2 cold vapour produced inside the LN2 tank (especially during the refill of 3). 5 and 6: motorized pneumatic valves. 7: monochromator (cf. inset). 8: exit of the N2 cold vapour. Continuous and dashed arrows: liquid and gaseous circulations, respectively. Internal diameter of the LN2 tubes: 15 mm. Inset: LN2 circulation inside the monochromator. 9: thermal bridges. 10: copper heat exchanger. 11: bellows. 12: monochromator rotation axis.

3.3. External LN2 circulation

The external circulation of LN2 is purely gravitational (Fig. 4[link]). The flow can be adjusted using two valves (5 and 6), just after the Dewar (3) and just before the monochromator (7). This allows a precise flow optimization. Moreover, in this geometry the Dewar works as a phase separator. No gaseous nitrogen is mixed in the LN2 flux, which would strongly decrease the heat exchanger efficiency.

Even if the Dewar is at atmospheric pressure, the position of the beam after the monochromator is slightly perturbed during its refill (valve 2 opened). The control of the motorized valve (5) by a Eurotherm system allows a complete and precise management of the Dewar. The complete stage of this refill is then divided into different steps:

(i) In operation, the motorized valve is opened to ∼10% of its maximum.

(ii) When the LN2 Dewar level reaches 30% of its total capacity (100 l), the refill procedure cycle is started (duration around 5 min, every 5 h).

(iii) Valve 5 is closed, valve 2 is opened. The LN2 flow through the monochromator remains constant, the 4 m-long tubes between valve 5 and the monochromator entrance being then considered as a tank (capacity of the LN2 tubes ∼0.7 l).

(iv) When the LN2 Dewar level reaches the high level (70%), valve 2 is closed and, after 30 seconds, valve 5 is opened to 100% to eliminate the N2 gas from the tubes.

(v) After 1 min, the aperture of valve 5 returns to its initial value.

4. Closed-loop feedback to optimize the X-ray flux through the monochromator

4.1. Temperature differences considerations

The feedback mechanisms are often used to increase the stability over a long period, particularly under the decrease of the heat load on the monochromator encountered in synchrotron radiation operations. In our case, the first-crystal temperature is kept constant and the goal of the feedback device here is to counterbalance the effect of the temperature difference between the first and the second crystals, around 120 and 300 K, respectively. Their dhkl parameters are different (Table 1[link]), leading to a small difference δθ between the Bragg angles (θBragg),

[n\lambda=2d_{hkl}\sin\theta_{\rm{Bragg}},\eqno(3)]

where λ [Å] = 12.3984/E [keV] is the wavelength of the diffracted beam directly linked to its energy E, and n is the order of diffraction (n = 1 in this case).

Table 1
Numerical values obtained for two crystallographic orientations

δθ is the difference between the θBragg values of the first and the second crystals. ωDarwin is the Darwin width of the crystal. The interplanar distances are calculated from the a lattice parameters and the linear expansion of Si (Marot et al., 1992[Marot, G., Rossat, M., Freund, A., Joksch, S., Kawata, H., Zhang, L., Ziegler, E., Berman, L., Chapman, D., Hastings, J. B. & Iarocci, M. (1992). Rev. Sci. Instrum. 63, 477-480.]).

  Si(111) Si(220)
dhkl (Å) at T = 125/300 K 3.13560/3.13633 1.92015/1.92059
Energy range (keV) 4–25 5–40
δθ range (µrad) 132.2–18.4 193.7–19.0
Δδθ range (µrad) 32.1–0.7 47.5–0.5
Δδθ > ωDarwin/10 (keV) E < 16.2 E ≲ 44

Equation (3)[link] in the differential form gives

[\delta\theta=\tan\theta_{\rm{Bragg}}\delta{d}/d_{hkl}.\eqno(4)]

The δθ value evolves during an XAS energy scan, the evolution being larger for Si(220) than for Si(111). This δθ evolution is quantified for a typical 1 keV energy EXAFS scan,

[\Delta\delta\theta(E)=\delta\theta_E-\delta\theta_{E+1}.\eqno(5)]

The obtained Δδθ(E) curves are shown in Fig. 5[link].

[Figure 5]
Figure 5
Δδθ and ωDarwin evolution with respect to energy for two crystallographic orientations.
4.1.1. Consequence on the X-ray beam intensity

The Δδθ(E) absolute difference values are significant, especially by comparison with the Darwin width of the crystals, ωDarwin, their theoretical angular acceptance given by the following equation,

[\omega_{\rm{Darwin}}=\left[2r_{\rm{e}}\lambda^2C|F(hkl)|\exp(-M)\right]/\left[V\pi\sin\left(2\theta_{\rm{Bragg}}\right)\right],\eqno(6)]

where re is the classical electron radius, C is the polarization factor (equal to 1 in this case), F(hkl) is the structure factor and exp(−M) is the Debye–Waller factor (= 1). The ωDarwin/10 curves are shown in Fig. 5[link] for the two crystallographic orientations (111) and (220), assuming that a Δδθ evolution larger than ωDarwin/10 needs a δθ feedback during an EXAFS scan.

For the Fe K-edge (7.112 keV), Δδθ equals 8.8 µrad [Si(111)] and 17.4 µrad [Si(220)], the values of the Darwin width being 39.6 µrad and 26.1 µrad, respectively. Without any δθ adjustment, the beam flux, optimized at the beginning of the scan, would strongly decrease, especially when using Si(220) crystals. A δθ feedback is then necessary in our case over the entire energy range to optimize the X-ray flux [except for energies higher than 16.2 keV when using Si(111) crystals].

4.1.2. Consequence of the vertical X-ray beam position

Another consequence of the temperature difference is the angular change of the monochromatic beam during an energy scan. A Δδθ(E) angular difference between the two crystals on a 1 keV energy scan leads to an angular evolution of the outgoing beam equal to 2Δδθ(E). With the distance between the monochromator and the sample being L = 12.51 m (Proux et al., 2005[Proux, O., Biquard, X., Lahera, E., Menthonnex, J.-J., Prat, A., Ulrich, O., Soldo, Y., Trévisson, P., Kapoujyan, G., Perroux, G., Taunier, P., Grand, D., Jeantet, P., Deleglize, M., Roux, J.-P. & Hazemann, J.-L. (2005). Phys. Scr. 115, 970-973.]), the X-ray vertical position shift on the sample equals 2LΔδθ(E). For the low-energy scans, these beam translations become really large compared with the vertical beam size (in the 100–200 µm range, depending on the second-mirror curvature). For the Fe K-edge, the vertical translations during the scans are 220 µm and 435 µm using Si(111) and Si(220), respectively. This motion is clearly not desirable for the study of inhomogeneous or small samples. To avoid this effect, we choose to move the experimental table during the energy scan with an accuracy of 1 µm. The position of the beam is then kept constant over the entire energy scan range with respect to the sample and all the detectors (for the transmission and fluorescence measurements), both in classical step-by-step and quick continuous EXAFS scan configurations.

4.2. Technical considerations

The implantation of the PZT in the monochromator is shown in Fig. 6[link]. Both a micro-motor and a PZT are used. The stepper micro-motor allows coarse adjustments of the parallelism between the two crystals during the alignment procedures and for the large Δδθ evolutions. The PZT allows fine adjustments and feedback optimization via the lock-in system. The main interest in the PZT with respect to the micro-motor is in its positioning resolution (sub-nanometric), but also the lack of a mechanical backlash and the possibility of providing a modulated excitation.

[Figure 6]
Figure 6
Sketch and photograph of the PZT implantation in the double-crystal monochromator. 1: piezoelectric transducer. 2: stepper micro-motor. 3: hinge.

4.3. Electronic considerations

Our aim was to use only one flux measurement system (I0, Fig. 7[link]) as an input parameter for the feedback system, in order to decorrelate as much as possible the δθ adjustment from the other parameters. The system described by Shard et al. (1998[Shard, A. G., Dhanak, V. R. & Smith, A. D. (1998). J. Synchrotron Rad. 5, 829-831.]) could not be adapted on our beamline owing to the dead-time induced by the process (minimum 1 second per point) and the impossibility of adapting it for a continuous energy scan (as used during quick-EXAFS experiments). We adapted then the system developed by Renevier et al. (2003[Renevier, H., Grenier, S., Arnaud, S., Berar, J.-F., Caillot, B., Hodeau, J.-L., Letoublon, A., Proietti, M. G. & Ravel, B. (2003). J. Synchrotron Rad. 10, 435-444.]) for DAFS experiments.

[Figure 7]
Figure 7
Block diagrams of the control system and the lock-in amplifier.

The idea was to measure a quantity proportional to the derivative of the rocking-curve peak. At the low- and high-angle sides of the peak maximum, the sign of the derivative is changing, whereas the derivative equals zero at the maximum. Fig. 7[link] shows a scheme of the set-up used.

The intensity of the transmitted beam versus the angular difference between first and second crystals (δθ) can be defined as I0 = f(δθ) = f(V), where V is the [0–10 V] input voltage of the PZT high-voltage supply. The V voltage, and so the δθ angle, is modulated with a frequency FOSC = ω/2π and an amplitude Vexc by the lock-in amplifier (EG&G instruments, 7220 DSP),

[V=\langle{V}\rangle+V_{\rm{exc}}\,\,2^{1/2}\cos(\omega{t})=\langle{V}\rangle+\Delta{V}_{\rm{exc}}.\eqno(7)]

For small ΔVexc modulations, a Taylor expansion of f(V) to the first order is expressed by

[f(V)\simeq{f}(\langle{V}\rangle)+\Delta{V}_{\rm{exc}}\left[\,f'(V)\right]_{\langle{V}\rangle}.\eqno(8)]

The transmitted signal I0 = f(V) is fed into the lock-in input. A block diagram of the internal process of the lock-in amplifier is sketched in the lower part of Fig. 7[link]. The Vmult voltage is expressed as

[V_{\rm{mult}}=f(V)\cos\left(\omega{t}+\varphi\right),\eqno(9)]

[\eqalignno{V_{\rm{mult}}\simeq{}&{f}(\langle{V}\rangle)\cos(\omega{t}+\varphi)\cr&+{\textstyle{1\over2}}\,V_{\rm{exc}}\,2^{1/2}\,\left[\,f'(V)\right]_{\langle{V}\rangle}[\cos(2\omega{t}+\varphi)+\cos(\varphi)].&(10)}]

After the low-pass filter, the output voltage is expressed as

[V_{\rm{out}}={\textstyle{1\over2}}\,V_{\rm{exc}}\,2^{1/2}\,\left[\,f'(V)\right]_{\langle{V}\rangle}\cos(\varphi).\eqno(11)]

By varying the V voltage, the derivative of the rocking curve f′(V) can be calculated and minimized with a simple PID (proportional-integral-derivative) software. This single PID software takes the output filter voltage Vout as an input signal. It computes a voltage command and is able to adjust this voltage 〈V〉 to maintain the θ angle at the maximum of the rocking curve.

The main interest in nulling a parameter is that in the case, for example, of electron beam loss in the storage ring or X-ray shutter closing, I0 = f(V) is constant (equal to 0) so f′(V) = 0 for all V values. The PZT voltage then remains constant. When the electron beam is restored, the δθ angle is very close to the optimal value and the optimization process is achieved very quickly.

4.4. Characteristics of the modulation

The optimal parameters are a modulation frequency FOSC equal to 440 Hz and a modulation amplitude equal to 1 mV (high-energy experiment, E > 20 keV) or 2 mV r.m.s. (E < 20 keV). The used PZT is controlled on a 0–10 V range and its maximal elongation is 10 µm. Such a voltage modulation amplitude gives rise to an angular modulation equal to 0.015 or 0.03 µrad, much smaller than the Darwin width of the crystals [in our case, the smallest value is obtained for E = 40 keV using Si(220) crystals, ωDarwin = 2.76 µrad]. This modulation does not induce any noise on the monochromatic X-ray beam intensity. Moreover, the vertical displacement of the beam associated with this modulation on the sample, located at 15 m after the monochromator, is less than 0.5 µm, the minimum vertical size being around 100 µm.

5. Beamline performances

5.1. Thermal stability

The first-crystal temperature stabilization is achieved with a 0.01° accuracy at a value ranging from 103 K to 118 K. This temperature value depends on the white-beam total power on the first crystal. For example, the stabilization temperature was 103 K at the Ti K-edge (K-edge energy Eedge = 4.966 keV), 110 K at the Cu K-edge (Eedge = 8.979 keV) and between 114 and 116 K for the experiments performed at higher energy (Eedge > 22 keV).

The stability of the energy resolution of the beamline is directly linked to the thermal stability of the first crystal. Two sets of experiments have been performed at the Ti K-edge on two amorphous alloys using Si(220) crystals for the monochromator (θBragg = 40.553°). The energy reproducibility for each set of ten spectra (defined unambiguously as the first zero of the derivative curve) is ±0.03 eV.

5.2. Energy resolution

The energy resolution of the beamline can be evaluated by the different XANES features of reference compounds. Fig. 8[link] represents several XANES spectra obtained in the transmission mode for pure metallic references (Ti, V, Cu, Zr and Nb) at their K-edge. Measurements were performed using Si(220) crystals and with the maximum vertical divergences delivered by the bending-magnet source on the first mirror (δθdiverg ranges from 0.23 mrad for Ti to 0.12 mrad for Nb, depending on the 1.15 m-long first-mirror angle). The pre-edge parts of the edges are shown in the inset. These features, which are often called pre-edge fine structures, have been shown to be caused by transitions of the 1s electron to the unoccupied electron states near the conduction band. Such transitions are dipole-forbidden for the free atoms but can become dipole-allowed for the atoms in the matrix, because of dp mixing. To extract these features, the contribution of the edge jump to the pre-edge was modeled by interpolating the background with a spline function, using the data several eV before and after the pre-edge. Experimental FWHM values (ΔEmeas) have then been quantified (Table 2[link]).

Table 2
Comparison between the experimental and theoretical FWHM peak of the XANES pre-edge for metallic references, ΔEelectr, ΔEdiverg and ΔEmeas (eV)

EK: K-edge energy (keV). Γi and Γf: core–hole lifetime broadenings of the initial and final states (eV) (Fuggle & Inglesfield, 1992[Fuggle, J. C. & Inglesfield, J. E. (1992). Editors. Topics in Applied Physics, Vol. 69, pp. 339-350. Berlin: Springer-Verlag.]). δcrystals: intrinsic resolution of the Si(220) crystals (eV). Transitions: 1s to 4p, in the 3d2 (a), 3d3 (b) and 3d10 (c) configurations (Raj et al., 2002[Raj, S., Padhi, H. C., Palit, P., Basa, D. K., Polasik, M. & Pawłowski, F. (2002). Phys. Rev. B, 65, 193105.]; Bazin & Rehr, 2003[Bazin, D. & Rehr, J. J. (2003). J. Phys. Chem. B, 107, 12398-12402.]; Wang et al., 2002[Wang, C. M., Slade Cargill, G., Chan, H. M. & Harmer, M. P. (2002). J. Am. Ceram. Soc. 85, 2492-2498.]; Ruckman et al., 1998[Ruckman, M. W., Reisfeld, G., Jisrawi Najeh, M., Weinert, M., Strongin, M., Wiesmann, H., Croft, M., Sahiner, A., Sills, D. & Ansari, P. (1998). Phys. Rev. B, 57, 3881-3886.]).

  EK Γi Γf δcrystals ΔEelectr ΔEdiverg ΔEmeas
Ti 4.966 0.94 0.1 (a) 0.25 0.98 5.28 1.85
V 5.465 1.01 0.2 (b) 0.28 1.07 6.16 1.5
Cu 8.979 1.55 1.6 (c) 0.46 2.27 10.6 3.4
Zr 17.998 3.84 0.09 0.92 3.95 13.4 4
Nb 18.986 4.14 0.11 0.97 4.25 14.1 4.5
[Figure 8]
Figure 8
K-edge XANES spectra measured in transmission mode for Ti, V, Cu, Zr and Nb metallic references. Insets: edge features after edge subtraction. Crystals: Si(220).

The broadening of these peaks can be theoretically evaluated roughly from the core–hole lifetime broadenings of the initial and final states (Γi and Γf) and from the intrinsic energy resolution of the monochromator (δcrystals). δcrystals values are directly calculated for each edge energy using the ωDarwin values obtained for Si(220) crystals (equation 6[link]),

[\delta\varepsilon_{\rm{crystals}}/E_{\rm{edge}}=\delta\theta/\theta=\omega_{\rm{\,Darwin}}/\theta_{\rm{Bragg}}.\eqno(12)]

Assuming a Gaussian contribution of all these parameters, the total broadening owing to these electronic considerations can then be estimated via the following equation,

[\Delta{E}_{\rm{electr}}=\left(\Gamma_{\rm{i}}^2+\Gamma_{\rm{f}}^2+\delta\varepsilon_{\rm{crystals}}^2\right)^{1/2}.\eqno(13)]

The efficiency of the vertical collimation of the beam by the first mirror can also be evaluated (Klementev, 2001[Klementev, K. V. (2001). J. Phys. D, 34, 2241-2247.]). The vertical divergence of the incident beam, δθdiverg, is larger (from 5× to 14×) than the angular acceptance of the crystals, ωDarwin (equation 6[link], Fig. 5[link]). Without the first mirror, the energy resolution would be

[\Delta{E}_{\rm{diverg}}=\delta\theta_{\rm{diverg}}E_{\rm{edge}}\,{\rm{cotan}}\left(\theta_{\rm{Bragg}}\right).\eqno(14)]

The theoretically calculated and experimentally measured data are collected in Table 2[link]. The ΔEdiverg values are always larger than ΔEmeas. The vertical collimation of the beam is correctly achieved by the first mirror: the vertical divergence of the beam on the first crystal is strongly reduced with respect to the total divergence. At the reverse, the differences between the ΔEmeas and ΔEelectr values are small, between 0.05 eV (Zr case) and 1.13 eV (Cu case). The experimental resolution of the beamline is close to the theoretical resolution.

5.3. Optimization and stability of the beamline

The feedback system allows a dynamical optimization of the I0 intensity during an energy scan. It leads to a significant improvement in the signal-to-noise ratio. Moreover, the device is dead-time free and therefore suitable for classical step-by-step and continuous EXAFS scans (Fig. 9[link]).

[Figure 9]
Figure 9
Monochromatic X-ray beam intensity, I0(E), normalized with the storage-ring current, ISR,mA(E), and with the integration time, for two kinds of acquisition modes: step-by-step (EXAFS) and continuous (QEXAFS) scans, Mo K-edge.

The intensity of the X-ray monochromatic beam I0, measured after the monochromator, decreases of course when the intensity of the current in the storage ring ISR decreases, but the ratio I0/ISR is constant even during a storage-ring refill (i.e. during a strong variation in the heat power on the first crystal). The standard deviation of this ratio over a long period is then a good way to quantify the beamline stability.

Over a short period and without any energy displacement, the optimization of the beamline is achieved with a noise of ∼0.03% r.m.s. for a counting time equal to 1 second per point (Fig. 10[link]). To estimate the quality of the beamline stability over a longer period, we report in Table 3[link] different measurements performed during continuous experiments at various energies and various storage-ring filling modes. During these sets of experiments, no beamline optimization was performed except for the feedback adjustment. The spectra are energy scans on a 1 keV range for a total acquisition time of approximately 30–45 min. For a period of 1 day (including two or three refills), the standard deviation is lower than 1%. It increases to around 2–3% for longer periods (2 and 4 days).

Table 3
Measured standard deviation of the intensity of the monochromatic beam I0 normalized by the current of the electron beam in the storage ring, ISR

Storage-ring mode: 16 bunches – lifetime 9 h, 50 < I (mA) < 90, one refill every 6 h; uniform – lifetime 75 h, 160 < I (mA) < 200, one refill every 12 h.

  Fe K-edge Br K-edge Zn K-edge Sb K-edge
Energy (keV) 7.112 13.474 9.659 30.491
Storage-ring mode 16 bunchs 16 bunchs Uniform Uniform
Number of spectra 88 (including 3 refills) 69 (including 2 refills) 90 (including 7 refills) 54 (including 4 refills)
Standard deviation 0.6% 0.9% 2.6% 2.1%
[Figure 10]
Figure 10
Monochromatic X-ray beam intensity, I0(t), as measured and normalized for a constant storage-ring current, ISR = 176 mA. During the 1000 s acquisition time, ISR decreased from 176.37 to 175.54 mA. Measurement 1 second per point at the As K-edge using Si(220) crystals (θBragg = 15.79°).

6. Conclusion

We have investigated the design and the performances of an indirect cryogenic cooling system for a double-crystal monochromator including a second-crystal angular optimization system. The first-crystal cooling system was adjusted to avoid vibrations and to work at a constant temperature in the 95–115 K range. The automatic angular correction of the angle between the two crystals leads to an optimization of the monochromatic beam intensity, during an energy scan (in the step-by-step or continuous mode) and when the storage-ring current decreases. The performance of the monochromator (thermal stability, energy resolution and optimization over short and long periods) has been checked with care.

The stability of the electron beam position in the ESRF storage ring is extremely good. However, owing to the large distances of the optic elements and the sample with respect to the source (Proux et al., 2005[Proux, O., Biquard, X., Lahera, E., Menthonnex, J.-J., Prat, A., Ulrich, O., Soldo, Y., Trévisson, P., Kapoujyan, G., Perroux, G., Taunier, P., Grand, D., Jeantet, P., Deleglize, M., Roux, J.-P. & Hazemann, J.-L. (2005). Phys. Scr. 115, 970-973.]), even a small perturbation of the electron beam position can induce a loss in the X-ray beam stability. This stability will be improved in the near future with the dynamic control of the evolution of the position of the incident beam owing to the long-range evolution of the position of the electron beam in the storage ring (Fischetti et al., 2002[Fischetti, R. F., Heurich, R., Perry, D., Stepanov, S., Kondrashkina, E. & Rosenbaum, G. (2002). Rev. Sci. Instrum. 73, 1518.]).

Acknowledgements

The authors wish to acknowledge the help of the ESRF technical staff and support groups, and in particular J. P. Vassalli and P. Villermet for the monochromator crystals realisation and H. P. Van Der Kleij for monochromator axis metrology. Thanks is also given to all the users who helped us in the beamline commissioning.

References

First citationAlkire, R. W., Rosenbaum, G. & Evans, G. (2000). J. Synchrotron Rad. 7, 61–68.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationBazin, D. & Rehr, J. J. (2003). J. Phys. Chem. B, 107, 12398–12402.  Web of Science CrossRef CAS Google Scholar
First citationBilderback, D. H. (1986). Nucl. Instrum. Methods A, 246, 434–436.  CrossRef Web of Science Google Scholar
First citationBilderback, D. H., Freund, A. K., Knapp, G. S. & Mills, D. M. (2000). J. Synchrotron Rad. 7, 53–60.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationBridges, F. (1987). Nucl. Instrum. Methods A, 257, 447–450.  CrossRef Web of Science Google Scholar
First citationCarpentier, P., Rossat, M., Charrault, P., Joly, J., Pirocchi, M., Ferrer, J. L., Kaïkati, O. & Roth, M. (2001). Nucl. Instrum. Methods A, 456, 163–176.  Web of Science CrossRef CAS Google Scholar
First citationChumakov, A., Rüffer, R., Leupold, O., Celse, J. P., Martel, K., Rossat, M. & Wah-Keat, L. (2004). J. Synchrotron Rad. 11, 132–141.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationComin, F. (1995). Rev. Sci. Instrum. 66, 2082–2084.  CrossRef CAS Web of Science Google Scholar
First citationCross, J. O., Elam, W. T., Harris, V. G., Kirkland, J.-P., Bouldin, C. E. & Sorensen, L. B. (1998). J. Synchrotron Rad. 5, 911–913.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationDejus, R. J. & del Rio, M. S. (1996). Rev. Sci. Instrum. 67, 3356.  CrossRef Google Scholar
First citationDufresne, E. M., Arms, D. A., Dierker, S. B., Clarke, R., Yacoby, Y., Pitney, J., MacHarrie, B. & Pindak, R. (2002). Rev. Sci. Instrum. 73, 1511–1513.  Web of Science CrossRef CAS Google Scholar
First citationFischetti, R. F., Heurich, R., Perry, D., Stepanov, S., Kondrashkina, E. & Rosenbaum, G. (2002). Rev. Sci. Instrum. 73, 1518.  CrossRef Google Scholar
First citationFuggle, J. C. & Inglesfield, J. E. (1992). Editors. Topics in Applied Physics, Vol. 69, pp. 339–350. Berlin: Springer-Verlag.  Google Scholar
First citationHazemann, J. L., Nayouf, K. & de Bergevin, F. (1995). Nucl. Instrum. Methods B, 97, 547–550.  CrossRef CAS Web of Science Google Scholar
First citationHettel, R. (1989). Rev. Sci. Instrum. 60, 1501–1506.  CrossRef CAS Web of Science Google Scholar
First citationJemian, P. R. (2002). Rev. Sci. Instrum. 73, 1533.  CrossRef Google Scholar
First citationKlementev, K. V. (2001). J. Phys. D, 34, 2241–2247.  Web of Science CrossRef CAS Google Scholar
First citationKnapp, G. S., Rogers, C. S., Beno, M. A., Wiley, C. L., Jennings, G. & Cowan, P. L. (1995). Rev. Sci. Instrum. 66, 2138–2140.  CrossRef CAS Web of Science Google Scholar
First citationMarot, G., Rossat, M., Freund, A., Joksch, S., Kawata, H., Zhang, L., Ziegler, E., Berman, L., Chapman, D., Hastings, J. B. & Iarocci, M. (1992). Rev. Sci. Instrum. 63, 477–480.  CrossRef CAS Web of Science Google Scholar
First citationProux, O., Biquard, X., Lahera, E., Menthonnex, J.-J., Prat, A., Ulrich, O., Soldo, Y., Trévisson, P., Kapoujyan, G., Perroux, G., Taunier, P., Grand, D., Jeantet, P., Deleglize, M., Roux, J.-P. & Hazemann, J.-L. (2005). Phys. Scr. 115, 970–973.  CrossRef Google Scholar
First citationRaj, S., Padhi, H. C., Palit, P., Basa, D. K., Polasik, M. & Pawłowski, F. (2002). Phys. Rev. B, 65, 193105.  CrossRef Google Scholar
First citationRamanathan, M., Engbretson, M. & Montano, P. A. (1988). Nucl. Instrum. Methods A, 266, 471–474.  CrossRef Web of Science Google Scholar
First citationRenevier, H., Grenier, S., Arnaud, S., Berar, J.-F., Caillot, B., Hodeau, J.-L., Letoublon, A., Proietti, M. G. & Ravel, B. (2003). J. Synchrotron Rad. 10, 435–444.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationRogers, C. S. & Assoufid, L. (1995). Rev. Sci. Instrum. 66, 2200–2202.  CrossRef CAS Web of Science Google Scholar
First citationRowen, M., Peck, J. W. & Rabedeau, T. (2001). Nucl. Instrum. Methods A, 467468, 400–403.  CrossRef CAS Google Scholar
First citationRuckman, M. W., Reisfeld, G., Jisrawi Najeh, M., Weinert, M., Strongin, M., Wiesmann, H., Croft, M., Sahiner, A., Sills, D. & Ansari, P. (1998). Phys. Rev. B, 57, 3881–3886.  Web of Science CrossRef CAS Google Scholar
First citationShard, A. G., Dhanak, V. R. & Smith, A. D. (1998). J. Synchrotron Rad. 5, 829–831.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationSiddons, D. P. (1998). J. Synchrotron Rad. 5, 685–686.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationWang, C. M., Slade Cargill, G., Chan, H. M. & Harmer, M. P. (2002). J. Am. Ceram. Soc. 85, 2492–2498.  Web of Science CrossRef CAS Google Scholar
First citationZhang, L., Lee, W. K., Wulff, M. & Eybert, L. (2003). J. Synchrotron Rad. 10, 313–319.  Web of Science CrossRef CAS IUCr Journals Google Scholar