Gas-based detectors for synchrotron radiation

CROSSMARK_Color_square_no_text.svg

aInstrumentation Division, Brookhaven National Laboratory, Upton, NY 11973-5000, USA
*Correspondence e-mail: gsmith@bnl.gov

(Received 21 June 2005; accepted 19 October 2005)

Gas-filled detectors have played an important role in experiments at synchrotron sources for many years. Although other X-ray detection technologies are gaining wider use, particularly with the higher flux of new sources, gas-filled detectors are still well suited for specific experiments. This article describes some fundamental characteristics of gas-filled detectors for X-ray detection, with emphasis on position encoding, position resolution and linearity. Recently developed micropattern structures for achieving electron multiplication are described. Some applications of conventional wire chambers and micro-pattern detectors are given.

1. Introduction

Gas-filled detectors, in a variety of forms and operating with some level of electron gain, have been used as X-ray converters and position-encoding instruments since the very first synchrotron experiments were carried out. Over this approximately 30 year period, they have at various times been the mainstay detector in specific classes of experiment, sometimes to be superseded later by a more appropriate and generally commercially produced device. A classic example is in X-ray protein crystallography, where in the 1980s two-dimensional multi-wire chambers (MWPCs) were found on many of these beamlines (Phizackerley et al., 1986[Phizackerley, R. P., Cork, C. W. & Merritt, E. A. (1986). Nucl. Instrum. Methods, A246, 579-595.]). At that time they competed with TV detectors (Arndt, 1984[Arndt, U. W. (1984). Nucl. Instrum. Methods, 222, 252-255.]), but both were progressively replaced by the image plate (Amemiya et al., 1988[Amemiya, Y., Matsushita, T., Nakagawa, A., Satow, Y., Miyahara, J. & Chikawa, J. (1988). Nucl. Instrum. Methods, A266, 645-653.]) and then the scintillator/CCD (Gruner & Ealick, 1995[Gruner, S. M. & Ealick, S. E. (1995). Structure, 3, 13-15.]) as the de facto protein crystallography detector. The driving force in this class of experiment has always been, primarily, higher count rate and better position resolution. While the MWPC has been unable to match the capabilities of more recent detectors for this application, it's inherent properties, and those of later gas-filled devices based on micro-pattern detectors, provide synchrotron users in a wide range of other experiments with capabilities that are, still, unattainable with other detector types. Wide-angle and small-angle X-ray scattering (WAXS, SAXS) and studies with weak signals in a diffuse background (e.g. speckle) are three such fields where gas-filled detectors are almost indispensable. This article will illustrate some of their key characteristics.

2. Some fundamental properties

2.1. Efficiency

One of the most basic properties is X-ray stopping power. Because of their low operating voltages compared with complex molecules, and their inherent stability, argon, krypton and xenon are the main constituents used in gas-filled detectors, together with a small volume (of order 10%) of quench gas, used to prevent the electron avalanche from becoming self-sustaining. The absorption depth (in which 63% of incoming photons are absorbed) in each of these noble gases is shown in Fig. 1[link], calculated from measured X-ray cross-section data (Storm & Israel, 1970[Storm, E. & Israel, H. I. (1970). Nucl. Data Tables, A7, 565-681.]). Because it has the highest atomic number, xenon has the greatest stopping power. In general, good stopping power can be achieved with argon in the X-ray range 3–8 keV, and with xenon in the range 8–40 keV. Krypton has slightly better stopping power than xenon in the range between their respective K-edges, 14.3–34.6 keV. These properties can be useful in choosing a suitable gas for an experiment if cost is a serious constraint (xenon about $10 per liter, krypton about $3 per liter, and argon is much cheaper than both). Exact conversion efficiencies can be easily evaluated for specific gas mixtures with the use of web tools, for example the Lawrence Berkeley Laboratory on-line calculator for photon transmission/absorption (https://www.cxro.lbl.gov/optical_constants/gastrn2.html).

[Figure 1]
Figure 1
Absorption depth as a function of X-ray energy.

2.2. Position resolution

Position resolution, σdet, is determined by several factors. In non-pixelated detectors, it is generally determined by four main contributors,

[\sigma_{\rm{det}}=\left(\sigma_{\rm{noise}}+\sigma_{\rm{diff}}+\sigma_{\rm{er}}+\sigma_{\rm{aval}}\right)^{1/2},\eqno(1)]

where σnoise is the r.m.s. contribution from electronic noise, σdiff the contribution from electron diffusion, σer the contribution from photoelectron and Auger electron range and σaval the contribution from avalanche centroid fluctuation. The effect of parallax is not considered here, and will be described in a later section. σnoise is dependent upon the type of readout and will not be quantified at this point. σdiff is generally the smallest of the three contributors. σer is the parameter that can be quantified independent of detector type. σaval derives from statistical variations in the electron avalanche centroid, which is also dependent upon detector type, and greatly dependent upon gas gain; in practice, it can be easily the major contributor, and is probably the least considered phenomenon in gas detector operation. Here we will focus on σer, the results based largely on three sets of measurements (Fischer et al., 1986a[Fischer, J., Radeka, V. & Smith, G. C. (1986a). Nucl. Instrum. Methods, A252, 239-245.],b[Fischer, J., Radeka, V. & Smith, G. C. (1986b). IEEE Tran. Nucl. Sci. NS-33, 257-260.]; Smith et al., 1994[Smith, G. C., Cox, D. E. & Jephcoat, A. P. (1994). Nucl. Instrum. Methods, A350, 621-623.]) in which the detector and its mode of operation ensured that all but σer in equation (1)[link] were negligible. A synopsis of these `electron range limited' position-resolution measurements, all at 105 Pa operating pressure, is shown in Fig. 2[link]. Several generalizations can be made. In the X-ray energy range 3.2–6 keV, argon provides the capability of 100 µm resolution. In the X-ray energy range 6–10 keV, xenon provides the capability of 100 µm resolution. In the 10–15 keV region, xenon achieves a resolution that increases from 100 µm to 300 µm. In the 15–20 keV region, gating on the KrK escape peaks can yield resolution ranging from <100 µm to 200 µm. The region 20–34.6 keV is totally dominated by long-range effects of photoelectrons in both Kr and Xe, but there is a dramatic improvement above the XeK edge. In the range 34.6–45 keV, gating on XeK escape peak events provides resolution from 100 µm to 350 µm. Taken in conjunction with efficiency curves in Fig. 1[link], a user requiring a specific position resolution can make an empirical choice of gas mixture to suit the energy requirements of the experiment.

[Figure 2]
Figure 2
Position resolution at 105 Pa (with 10% quench gas).

2.3. Counting rate (limitation owing to space charge)

One of the first models for describing the electric field reduction at an anode wire surface, owing to positive-ion space charge, was given by Sipila et al. (1980[Sipila, H., Vanha-Honko, V. & Bergqvist, J. (1980). Nucl. Instrum. Methods, 176, 381-387.]). This was followed by a more comprehensive description that accounted for second-order effects such as anode wire diameter (Mathieson & Smith, 1992[Mathieson, E. & Smith, G. C. (1992). Nucl. Instrum. Methods, A316, 246-251.]). The presence of any positive ions in the volume between anode and cathode will cause an effective reduction in the electric field at the anode surface, and hence there is a decrease in the gas gain. This model calculates the effective field reduction and excellent agreement was obtained between prediction and measurement. An example of the predicted avalanche size versus count rate for a specific detector geometry is shown in Fig. 3[link]. Since avalanche size (in the absence of major charge saturation) is the product of primary charge and gas gain, higher count rates will be sustained with lower gas gain. Lower gas gain will also prolong the onset of aging effects, in which deposits are formed on the anode electrode and another, independent, reduction in gas gain occurs. A thorough analysis of this `aging' phenomenon was given by Kadyk (1991[Kadyk, J. (1991). Nucl. Instrum. Methods, A300, 436-479.]) and more recently by Va'vra (2003[Va'vra, J. (2003). Nucl. Instrum. Methods, A515, 1-14.]) and Sauli (2003[Sauli, F. (2003). Nucl. Instrum. Methods, A515, 358-363.]).

[Figure 3]
Figure 3
Relative drop in gain versus count rate.

3. Basis of conventional position encoding

In the majority of gas-filled detectors, the basis of both one- and two-dimensional X-ray encoding is determination of the centroid of cathode charge, induced from an anode avalanche. Fig. 4[link] shows that the footprint of charge on the cathodes is significant, compared with the small (<1 mm) physical extent of the anode avalanche. In fact the induced cathode charge has a quasi-Gaussian profile with a FWHM of about 1.6d, where d is the anode–cathode spacing (Gatti et al., 1979[Gatti, E., Longoni, A., Okuno, H. & Semenza, P. (1979) Nucl. Instrum. Methods, 163, 83-92.]; Gordon & Mathieson, 1984[Gordon, J. S. & Mathieson, E. (1984). Nucl. Instrum. Methods, 227, 267-276.]). In practice, readouts for cathodes have been developed that essentially determine the centroid of charge on the strips/wires, a measurement that yields a precision much better than the strip pitch. The most common readout techniques are delay line, or amplifier per strip/wire. In general, delay-line readout achieves 105 to 106 counts s−1, and advanced examples of amplifier per strip can achieve 107 counts s−1 and a little beyond.

[Figure 4]
Figure 4
Cathode-induced charge from an anode avalanche.

4. Delay-line detectors

4.1. General operation

Position readout with delay line represents the most highly developed, and most frequently used, encoding system for X-ray synchrotron experiments (Smith et al., 1992[Smith, G. C., Yu, B., Fischer, J., Radeka, V. & Harder, J. A. (1992). Nucl. Instrum. Methods, A323, 78-85.]). Their robust and excellent performance in terms of resolution and linearity have long been known (Boie et al., 1982[Boie, R. A., Fischer, J., Inagaki, Y., Merritt, F. C., Radeka, V., Rogers, L. C. & Xi, D. M. (1982). Nucl. Instrum. Methods, 201, 93-115.]). With this readout, all cathode strips/wires feed the nodes of a delay line, and timing information of the signals at each end of the line determines the centroid location. Fig. 5[link] shows the principle. The classic calibration measurement with such a system is determination of X-ray position resolution as a function of anode charge, as shown in Fig. 6[link] (Capel et al., 1995[Capel, M. S., Smith, G. C. & Yu, B. (1995). Rev. Sci. Instrum. 66, 2295-2299.]). This illustrates that, as the anode charge is increased, position resolution, at first, improves as the reciprocal of the anode charge. This corresponds to detector operation in a regime where the electronic noise contribution, σnoise, is dominant. As anode charge is increased, the resolution begins to deviate from the 45° line, because other factors, primarily electron range, become significant. However, if the anode signal is increased too much, the resolution begins to deteriorate, a mode of operation to be avoided because the avalanche begins to spread along the wire, contributing to an increasing value of σaval. This diagram also illustrates the effect of changing delay-line transit time. In this example, reducing the transit time from 1 µs to 0.5 µs, the electronic noise (which in terms of number of electrons remains virtually the same) has increased its contribution, σnoise, by a factor of two because the time scale over which position is determined has been halved, while the absolute uncertainty in timing at each end of the delay line remains essentially unchanged.

[Figure 5]
Figure 5
Front-end synopsis of delay-line readout (one-axis only shown).
[Figure 6]
Figure 6
Position resolution (FWHM, µm) versus anode charge.

4.2. Time-to-digital converter

Timing information from the delay line is converted into position by a time-to-digital converter (TDC), which is one of the most crucial elements in a correctly operational instrument. Excellent differential non-linearity is a key desired characteristic, and few really good TDCs have been developed specifically for delay-line detectors. Some examples over the last 15 years are given by Harder (1988[Harder, J. A. (1988). Nucl. Instrum. Methods, A265, 500-510.]), Berry (1993[Berry, A. (1993). Rev. Sci. Instrum. 64, 1222-1228.]), Chemloul & Comparat (1995[Chemloul, M. & Comparat, V. (1995). Nucl. Instrum. Methods, A355, 532-536.]), Epstein & Boulin (1998[Epstein, A. & Boulin, C. (1998). IEEE Trans. Nucl. Sci. NS-45, 1931-1933.]), Hervé & Le Caër (1999[Hervé, C. & Le Caër, T. (1999). Rev. Sci. Instrum. 70, 226-231.]) and Levchanovski et al. (2004[Levchanovski, F. V., Gebauer, B., Litvinenko, E. I., Nikiforov, A. S., Prikhodko, V. I., Schulz, Ch. & Wilpert, Th. (2004). Nucl. Instrum. Methods, A529, 413-416.]). It is probably true to say that the ready availability of a quality TDC containing the proper discriminators would have contributed significantly to efficient use of delay-line detectors in the synchrotron environment. While the examples cited have helped tremendously in this respect, appropriate TDCs are in general not available in great numbers or commercially. Two important discriminator functions were incorporated into one of the first TDC developments (Harder, 1988[Harder, J. A. (1988). Nucl. Instrum. Methods, A265, 500-510.]). The first, and most important, is the sum discriminator. With reference to Fig. 5[link], this discriminator ensures that the sum of the left scaler and right scaler contents, for a single event, is equal to the delay-line transit time, τ. If this condition is not met, all events relating to those timing signals are aborted. The response of this TDC to an electronic input that simulates, accurately, all positions, is shown in Fig. 7(a)[link]. The differential non-linearity is <0.1%, an excellent value. When the TDC is connected to a (one-dimensional) delay-line detector (τ = 1 µs) which is illuminated with X-rays, the uniform irradiation response (UIR) at 20 × 103 counts s−1 is shown in Fig. 7(b)[link]. This illustrates the good linearity of the detector in terms of uniform conversion per unit length and uniformity of delay-line response. If the sum discriminator is not incorporated, then at high rate a peaked response is generated, as shown by the UIR at 106 counts s−1 shown in Fig. 7(c)[link]. This occurs because of a tendency for the left and right scaler to be stopped close together in time, thereby simulating an event toward the center of the detector. An unusual, and at first surprising, second-order artifact occurs when the sum discriminator is applied at high rate, as shown in Fig. 7(d)[link]: there is still a small excess of events at the center of the detector. There is, in fact, a position dependency to the events that are rejected: an event at either end of the delay line will generate one signal that takes almost τ to reach the end of the delay line, while signals for an event in the center will take only τ/2 to reach each end of the line. Therefore, events at the edge of the detector are more likely to be rejected through the occurrence of the next event than those in the center. Application of a `minimum inter-arrival time' discriminator generates a flat spectrum very similar to that in Fig. 7(b)[link], at the expense of throughput. The latest TDC development (Levchanovski et al., 2004[Levchanovski, F. V., Gebauer, B., Litvinenko, E. I., Nikiforov, A. S., Prikhodko, V. I., Schulz, Ch. & Wilpert, Th. (2004). Nucl. Instrum. Methods, A529, 413-416.]) includes logic to try and disentangle signals from more than one event to improve throughput at high rates.

[Figure 7]
Figure 7
Uniform irradiation responses (UIRs) using purpose-designed TDC. (a) Electronic calibration, intrinsic non-linearity <0.1%. (b) 20 × 103 counts s−1, no discriminators. (c) 106 counts s−1, no discriminators. (d) 106 counts s−1, with sum discriminator.

4.3. Linearity

An application that is ideally suited to the low background sensitivity of a gas-filled detector is measurement of speckle. This has been described by Kocsis (1997[Kocsis, M. (1997). Nucl. Instrum. Methods, A392, 28-31.]). At Brookhaven's National Synchrotron Light Source, an improved version of a detector (Smith & Yu, 1995[Smith, G. C. & Yu, B. (1995). IEEE Trans. Nucl. Sci. NS-42, 541-547.]), with wire and electrode spacing less than 1 mm, is under study for speckle measurements (Dierker & Shin, 2004[Dierker, S. & Shin, T. J. (2004). Private communication.]). It possesses extremely good position resolution and linearity in both axes, as shown by the image of a 15 mm-diameter plastic gear wheel in Fig. 8[link], and, with 3 × 105 Pa of xenon, can achieve a resolution of about 50 µm.

[Figure 8]
Figure 8
Example of good linearity in both axes of an MWPC (transmission image of a 1.5 cm-diameter gear wheel).

5. Micro-pattern detectors

A number of detectors in which electron multiplication takes place in high field regions around lithographically produced anodes have been developed over the last 15 years. Among other features, these offer the hope that gas-filled detectors can be fabricated more automatically, with much less of the labor-intensive effort required for MWPCs. While not an exhaustive list, some of these devices are the micro-strip gas chamber, MSGC (Oed, 1988[Oed, A. (1988). Nucl. Instrum. Methods, A263, 351-359.]), the micro-gap chamber, MGC (Bellazzini & Spandre, 1995[Bellazzini, R. & Spandre, G. (1995). Nucl. Instrum. Methods, A368, 259-264.]), the compteur a trous, CAT (Bartol et al., 1996[Bartol, F., Bordessoule, M., Chaplier, G., Lemonnier, M. & Megtert, S. (1996). J. Phys. III Fr. 6, 337-347.]), the gas electron multiplier, GEM (Sauli, 1997[Sauli, F. (1997). Nucl. Instrum. Methods, A386, 531-534.]), the micro-mesh-gaseous structure, Micromegas (Giomataris et al., 1996[Giomataris, Y., Rebourgeard, Ph., Robert, J. P. & Charpak, G. (1996). Nucl. Instrum. Methods, A376, 29-35.]) and micro-pin array, MIPA (Rehak et al., 1997[Rehak, P., Smith, G. C. & Yu, B. (1997). IEEE Trans. Nucl. Sci. NS-44, 651-655.]). Photographs of a part of a GEM structure, and part of a MIPA structure, are shown in Figs. 9[link] and 10[link], respectively.

[Figure 9]
Figure 9
Example of a gas electron multiplier (GEM), etched from Cu-coated kapton. The diameter of the holes is about 75 µm.
[Figure 10]
Figure 10
Micro-pin array (MIPA). Anode pin spacing is 400 µm.

6. Further developments

Developments have taken place at a number of synchrotron sources aimed at improving the count-rate capability of two-dimensional detectors. We will describe three that have achieved significant application in synchrotron experiments.

6.1. Amplifier per wire in a MWPC

A development over the last few years at Daresbury Laboratory has resulted in an instrument known as the RAPID (refined ADC per input detector) system. RAPID1 (Lewis et al., 1997[Lewis, R. A., Helsby, W. I., Jones, A. O., Hall, C. J., Parker, B., Sheldon, J., Clifford, P., Hillem, M., Sumner, I., Fore, N. S., Jones, R. W. M. & Roberts, K. M. (1997). Nucl. Instrum. Methods, A392, 32-41.]) evolved into a 20 cm × 20 cm micro-gap detector with 256 wires/strips on each x, y readout axis, achieved with 128 readout channels per axis. Each readout channel comprises a preamplifier, ADC, with wire/strip digital information used to calculate a center-of-gravity for each event. A schematic of the micro-gap format is shown in Fig. 11[link]. A combination of shorter positive-ion collection time, short shaping time, and a degree of parallelism in the analysis permits this system to sustain counting rates that are in excess of standard delay-line readout. Global count rates of 2 × 107 s−1, local (single diffraction peak) count rates of 106 s−1, and a position resolution of about 300 µm FWHM have been reported. RAPID2 (Berry et al., 2003[Berry, A., Helsby, W. I., Parker, B. T., Hall, C. J., Buksh, P. A., Hill, A., Clague, N., Hillon, M., Corbett, G., Clifford, P., Tidbury, A., Lewis, R. A., Cernik, B. J., Barnes, P. & Derbyshire, G. E. (2003). Nucl. Instrum. Methods, A513, 260-263.]) is a second-generation system consisting of a curved detector and an improved algorithm/faster RAM.

[Figure 11]
Figure 11
Concept of the micro-gap chamber.

6.2. Amplifier per wire in a MSGC

Significant development in MSGCs has taken place in Japan. A two-dimensional detector was produced on a thin substrate, with the second coordinate determined from cross-strips separated from the front surface by a 17 µm polyimide spacer. The principle is shown in Fig. 12[link] (Tanimori et al., 1996[Tanimori, T., Ochi, A., Minami, S. & Nagae, T. (1996). Nucl. Instrum. Methods, A381, 280-288.]). Improved larger versions of this early device have been used at Spring-8 SAXS experiments (Toyokawa et al., 2001[Toyokawa, H., Fujisawa, T., Inoko, Y., Nagayoshi, T., Nishi, Y., Nishikawa, Y., Ochi, A., Suzuki, M. & Tanimori, T. (2001). Nucl. Instrum. Methods, A467-468, 1144-1147.]). Further, advanced micro-pattern detector developments are under way also in Japan (Ochi et al., 2002[Ochi, A., Nagayoshi, T., Koishi, S., Tanimori, T., Nagae, T. & Nakamura, M. (2002). Nucl. Instrum. Methods, A478, 196-199.]; Takahashi et al., 2003[Takahashi, H., Hagai, C., Yano, K., Fukuda, D., Nakazawa, M., Kishimoto, S., Ino, T., Furusaka, M. & Hasegawa, K. (2003). Nucl. Instrum. Methods, A410, 238-258.]).

[Figure 12]
Figure 12
MSGC with two-dimensional readout.

6.3. CAT- and GEM-based detectors

At Synchrotron Trieste, advances have been achieved using the CAT. This features a gas amplifying micro-hole structure in combination with a resistive position-encoding structure, the latter divided into small independent elements (Sarvestani et al., 1998[Sarvestani, A., Besch, H. J., Junk, M., Meißner, W., Sauer, N., Stiehler, R., Walenta, A. H. & Menk, R. H. (1998). Nucl. Instrum. Methods, A410, 238-258.]). A schematic diagram of a small part of the detector structure is shown in Fig. 13[link]. X-rays convert as usual, between the window and MicroCAT electrode, and primary electrons multiply in the small gap between the MicroCAT and the readout structure. As shown at the bottom right, the readout structure is divided into square resistive pads, with a readout node at each corner; measurement of the signal charge at each corner will permit localization of the center-of-gravity of the event to a fraction of the 4 mm side. The method of multiple readout, and recent developments regarding replacement of the MicroCAT with GEMs, are described by Orthen et al. (2004[Orthen, A., Wagner, H., Martoiu, S., Amenitsch, H., Bernstorff, S., Besch, H. J., Menk, R. H., Nurdan, K., Rappolt, M., Walenta, A. H. & Werthenbach, U. (2004). J. Synchrotron Rad. 11, 177-186.]).

[Figure 13]
Figure 13
Principle of the MicroCAT detector.

7. Parallax reduction

7.1. One-dimensional encoding

The conventional approach to eliminating parallax errors in one dimension is to construct the conversion depth of the detector with an appropriate radius of curvature. This is quite impossible in the case of a wire detector, with the wire direction along the sensing axis. Alternative electron multiplying technologies have been adopted to resolve this, such as a blade (Ballon et al., 1983[Ballon, J., Comparat, V. & Pouxe, J. (1983). Nucl. Instrum. Methods, 217, 213-216.]; Yu et al., 1998[Yu, B., Smith, G. C., Siddons, D. P. & Pietraski, P. J. (1998). Nucl. Instrum. Methods, A419, 519-524.]), in which multiplication occurs along the sharp edge of a metal blade or sheet, and which can be fabricated as an arc. A method that permitted off-line correction of parallax by determination of the depth at which X-ray conversion occurred in a planar slab of gas was demonstrated with a multi-step avalanche chamber (Bateman et al., 1985[Bateman, J. E., Connolly, J. F. & Stephenson, R. (1985). Nucl. Instrum. Methods, A239, 251-259.]). None of these approaches have been adopted in a serious way at synchrotron sources, although curved detectors are available commercially (https://www.inel.fr/en/contact/) using a blade operating in the limited streamer mode, with generally restricted rate and resolution capability.

The conventional geometry for operating MWPCs and most micro-pattern detectors in X-ray scattering experiments requires photons to enter the detector in a direction perpendicular to the planes of the electrodes. All X-rays converted at a specific x, y location create primary electrons that drift to the same location for multiplication. This situation leads to the phenomenon of space charge saturation, outlined earlier, in which the local electric field strength is reduced and the gas gain is also reduced.

A method that largely eliminates the space charge problem, and also eliminates parallax errors, is to permit X-rays to enter the detector such that their trajectory is parallel to the electron multiplying plane. Applicable only as a one-dimensional device, this approach is well suited to the MSGC, and Fig. 14[link] shows the principle. This mode of operation has been developed at the European Synchrotron Radiation Facility (Zhukov et al., 1997[Zhukov, V., Udo, F., Marchena, O., Hartjes, F. G., van den Berg, F. D., Bras, W. & Vlieg, E. (1997). Nucl. Instrum. Methods, A392, 83-88.]) and at the Rutherford Appleton Laboratory (Bateman et al., 2002[Bateman, J. E., Connelly, J. F., Derbyshire, G. E., Duxbury, D. M., Lipp, J., Mir, J. A., Simmons, J. E., Spill, E. J., Stephenson, R., Dobson, B. R., Farrow, R. C., Helsby, W. I., Mutikainen, R. & Suni, I. (2002). Nucl. Instrum. Methods, A477, 340-346.]).

[Figure 14]
Figure 14
One-dimensional detector with radial readout, and no parallax error.

7.2. Two-dimensional encoding

Only a few years after the initial concept of the MWPC had been developed, Charpak's group at CERN studied the possibility of adding a spherical drift chamber to the front of a conventional planar MWPC (Charpak et al., 1974[Charpak, G., Hajduk, Z., Jeavons, A., Stubbs, R. J. & Kahn, R. (1974). Nucl. Instrum. Methods, 122, 307.]). This scheme was mechanically challenging, but a working device was developed for the LURE synchrotron source in the 1980s (Kahn et al., 1986[Kahn, R., Fourme, R., Bosshard, R. & Saintagne, V. (1986). Nucl. Instrum. Methods, A246, 596-603.]) as a workhorse for a number of years. More recently, the mechanical difficulties associated with the latter approach were removed by adapting the electric field in an existing planar detector to exhibit a quasi-spherical field (Rehak et al., 1997[Rehak, P., Smith, G. C. & Yu, B. (1997). IEEE Trans. Nucl. Sci. NS-44, 651-655.]). A comparison of the normal field, and modified field, of this arrangement is shown in Fig. 15[link]. The much improved position resolution of this scheme is shown in Fig. 16[link], in which a 5.4 keV beam is incident on a planar detector containing Xe/10%-CO2 at angles from zero to 12°. The improvement in FWHM is very significant. This scheme has not been adopted in a significant way as yet; the Daresbury Laboratory is studying an application (Helsby, 2004[Helsby, W. (2004). Private communication.]) and the Institute Laue Langevin is investigating the technique for application to neutron detectors (Van Esch et al., 2005[Van Esch, P., Clergeau, J. F. & Medjoubi, K. (2005). Nucl. Instrum. Methods, A540, 361-367.]).

[Figure 15]
Figure 15
Principle of parallax reduction in a planar detector.
[Figure 16]
Figure 16
Position-resolution measurements without (thick lines) and with (thin lines) parallax reduction. Collimated 5.4 keV X-rays, Xe/10%-CO2.

8. Conclusion

Gas-filled detectors can provide efficient high-resolution high-rate detection of X-rays. They have extremely low background, and in many applications provide a cost-effective solution that is unattainable with any other device. While they are considered a mature instrument, new developments in operation of wire chambers and innovative methods for electron multiplication in a gas are occurring. It appears likely that these detectors will continue to find useful application in synchrotron experiments in the foreseeable future.

Acknowledgements

The development work and results presented here reflect the efforts of many research groups and collaborators. The author is indebted to colleagues working in the field of gas detectors for many informative and enjoyable discussions. This work was supported by the United States Department of Energy under Contract No. DE-AC02-98CH10886.

References

First citationAmemiya, Y., Matsushita, T., Nakagawa, A., Satow, Y., Miyahara, J. & Chikawa, J. (1988). Nucl. Instrum. Methods, A266, 645–653.  CrossRef Web of Science Google Scholar
First citationArndt, U. W. (1984). Nucl. Instrum. Methods, 222, 252–255.  CrossRef CAS Web of Science Google Scholar
First citationBallon, J., Comparat, V. & Pouxe, J. (1983). Nucl. Instrum. Methods, 217, 213–216.  CrossRef CAS Web of Science Google Scholar
First citationBartol, F., Bordessoule, M., Chaplier, G., Lemonnier, M. & Megtert, S. (1996). J. Phys. III Fr. 6, 337–347.  CrossRef CAS Google Scholar
First citationBateman, J. E., Connelly, J. F., Derbyshire, G. E., Duxbury, D. M., Lipp, J., Mir, J. A., Simmons, J. E., Spill, E. J., Stephenson, R., Dobson, B. R., Farrow, R. C., Helsby, W. I., Mutikainen, R. & Suni, I. (2002). Nucl. Instrum. Methods, A477, 340–346.  CrossRef Google Scholar
First citationBateman, J. E., Connolly, J. F. & Stephenson, R. (1985). Nucl. Instrum. Methods, A239, 251–259.  CrossRef CAS Google Scholar
First citationBellazzini, R. & Spandre, G. (1995). Nucl. Instrum. Methods, A368, 259–264.  CrossRef Google Scholar
First citationBerry, A. (1993). Rev. Sci. Instrum. 64, 1222–1228.  CrossRef CAS Web of Science Google Scholar
First citationBerry, A., Helsby, W. I., Parker, B. T., Hall, C. J., Buksh, P. A., Hill, A., Clague, N., Hillon, M., Corbett, G., Clifford, P., Tidbury, A., Lewis, R. A., Cernik, B. J., Barnes, P. & Derbyshire, G. E. (2003). Nucl. Instrum. Methods, A513, 260–263.  CrossRef Google Scholar
First citationBoie, R. A., Fischer, J., Inagaki, Y., Merritt, F. C., Radeka, V., Rogers, L. C. & Xi, D. M. (1982). Nucl. Instrum. Methods, 201, 93–115.  CrossRef CAS Web of Science Google Scholar
First citationCapel, M. S., Smith, G. C. & Yu, B. (1995). Rev. Sci. Instrum. 66, 2295–2299.  CrossRef CAS Web of Science Google Scholar
First citationCharpak, G., Hajduk, Z., Jeavons, A., Stubbs, R. J. & Kahn, R. (1974). Nucl. Instrum. Methods, 122, 307.  CrossRef Google Scholar
First citationChemloul, M. & Comparat, V. (1995). Nucl. Instrum. Methods, A355, 532–536.  CrossRef Google Scholar
First citationDierker, S. & Shin, T. J. (2004). Private communication.  Google Scholar
First citationEpstein, A. & Boulin, C. (1998). IEEE Trans. Nucl. Sci. NS-45, 1931–1933.  CrossRef Google Scholar
First citationFischer, J., Radeka, V. & Smith, G. C. (1986a). Nucl. Instrum. Methods, A252, 239–245.  CrossRef CAS Google Scholar
First citationFischer, J., Radeka, V. & Smith, G. C. (1986b). IEEE Tran. Nucl. Sci. NS-33, 257–260.  CrossRef CAS Google Scholar
First citationGatti, E., Longoni, A., Okuno, H. & Semenza, P. (1979) Nucl. Instrum. Methods, 163, 83–92.  CrossRef CAS Google Scholar
First citationGiomataris, Y., Rebourgeard, Ph., Robert, J. P. & Charpak, G. (1996). Nucl. Instrum. Methods, A376, 29–35.  CrossRef Google Scholar
First citationGordon, J. S. & Mathieson, E. (1984). Nucl. Instrum. Methods, 227, 267–276.  CrossRef CAS Google Scholar
First citationGruner, S. M. & Ealick, S. E. (1995). Structure, 3, 13–15.  CrossRef CAS PubMed Web of Science Google Scholar
First citationHarder, J. A. (1988). Nucl. Instrum. Methods, A265, 500–510.  CrossRef CAS Google Scholar
First citationHelsby, W. (2004). Private communication.  Google Scholar
First citationHervé, C. & Le Caër, T. (1999). Rev. Sci. Instrum. 70, 226–231.  CrossRef CAS Google Scholar
First citationKadyk, J. (1991). Nucl. Instrum. Methods, A300, 436–479.  CrossRef CAS Google Scholar
First citationKahn, R., Fourme, R., Bosshard, R. & Saintagne, V. (1986). Nucl. Instrum. Methods, A246, 596–603.  CrossRef CAS Web of Science Google Scholar
First citationKocsis, M. (1997). Nucl. Instrum. Methods, A392, 28–31.  CrossRef Google Scholar
First citationLevchanovski, F. V., Gebauer, B., Litvinenko, E. I., Nikiforov, A. S., Prikhodko, V. I., Schulz, Ch. & Wilpert, Th. (2004). Nucl. Instrum. Methods, A529, 413–416.  CrossRef Google Scholar
First citationLewis, R. A., Helsby, W. I., Jones, A. O., Hall, C. J., Parker, B., Sheldon, J., Clifford, P., Hillem, M., Sumner, I., Fore, N. S., Jones, R. W. M. & Roberts, K. M. (1997). Nucl. Instrum. Methods, A392, 32–41.  CrossRef Google Scholar
First citationMathieson, E. & Smith, G. C. (1992). Nucl. Instrum. Methods, A316, 246–251.  CrossRef CAS Google Scholar
First citationOchi, A., Nagayoshi, T., Koishi, S., Tanimori, T., Nagae, T. & Nakamura, M. (2002). Nucl. Instrum. Methods, A478, 196–199.  CrossRef Google Scholar
First citationOed, A. (1988). Nucl. Instrum. Methods, A263, 351–359.  CrossRef CAS Google Scholar
First citationOrthen, A., Wagner, H., Martoiu, S., Amenitsch, H., Bernstorff, S., Besch, H. J., Menk, R. H., Nurdan, K., Rappolt, M., Walenta, A. H. & Werthenbach, U. (2004). J. Synchrotron Rad. 11, 177–186.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationPhizackerley, R. P., Cork, C. W. & Merritt, E. A. (1986). Nucl. Instrum. Methods, A246, 579–595.  CrossRef CAS Google Scholar
First citationRehak, P., Smith, G. C. & Yu, B. (1997). IEEE Trans. Nucl. Sci. NS-44, 651–655.  CrossRef Google Scholar
First citationSarvestani, A., Besch, H. J., Junk, M., Meißner, W., Sauer, N., Stiehler, R., Walenta, A. H. & Menk, R. H. (1998). Nucl. Instrum. Methods, A410, 238–258.  CrossRef Google Scholar
First citationSauli, F. (1997). Nucl. Instrum. Methods, A386, 531–534.  CrossRef Google Scholar
First citationSauli, F. (2003). Nucl. Instrum. Methods, A515, 358–363.  CrossRef Google Scholar
First citationSipila, H., Vanha-Honko, V. & Bergqvist, J. (1980). Nucl. Instrum. Methods, 176, 381–387.  CrossRef CAS Web of Science Google Scholar
First citationSmith, G. C., Cox, D. E. & Jephcoat, A. P. (1994). Nucl. Instrum. Methods, A350, 621–623.  CrossRef Google Scholar
First citationSmith, G. C. & Yu, B. (1995). IEEE Trans. Nucl. Sci. NS-42, 541–547.  CrossRef Google Scholar
First citationSmith, G. C., Yu, B., Fischer, J., Radeka, V. & Harder, J. A. (1992). Nucl. Instrum. Methods, A323, 78–85.  CrossRef CAS Google Scholar
First citationStorm, E. & Israel, H. I. (1970). Nucl. Data Tables, A7, 565–681.  CrossRef Google Scholar
First citationTakahashi, H., Hagai, C., Yano, K., Fukuda, D., Nakazawa, M., Kishimoto, S., Ino, T., Furusaka, M. & Hasegawa, K. (2003). Nucl. Instrum. Methods, A410, 238–258.  Google Scholar
First citationTanimori, T., Ochi, A., Minami, S. & Nagae, T. (1996). Nucl. Instrum. Methods, A381, 280–288.  CrossRef Web of Science Google Scholar
First citationToyokawa, H., Fujisawa, T., Inoko, Y., Nagayoshi, T., Nishi, Y., Nishikawa, Y., Ochi, A., Suzuki, M. & Tanimori, T. (2001). Nucl. Instrum. Methods, A467468, 1144–1147.  CrossRef Google Scholar
First citationVan Esch, P., Clergeau, J. F. & Medjoubi, K. (2005). Nucl. Instrum. Methods, A540, 361–367.  CrossRef Google Scholar
First citationVa'vra, J. (2003). Nucl. Instrum. Methods, A515, 1–14.  Google Scholar
First citationYu, B., Smith, G. C., Siddons, D. P. & Pietraski, P. J. (1998). Nucl. Instrum. Methods, A419, 519–524.  CrossRef Google Scholar
First citationZhukov, V., Udo, F., Marchena, O., Hartjes, F. G., van den Berg, F. D., Bras, W. & Vlieg, E. (1997). Nucl. Instrum. Methods, A392, 83–88.  CrossRef Google Scholar

© International Union of Crystallography. Prior permission is not required to reproduce short quotations, tables and figures from this article, provided the original authors and source are cited. For more information, click here.

Journal logoJOURNAL OF
SYNCHROTRON
RADIATION
ISSN: 1600-5775
Follow J. Synchrotron Rad.
Sign up for e-alerts
Follow J. Synchrotron Rad. on Twitter
Follow us on facebook
Sign up for RSS feeds