research papers
X-ray diffractometer combining synchrotron radiation and pulsed magnetic fields up to 40 T
aISSP, University of Tokyo, Kashiwa, Chiba 277-858, Japan, bRIKEN SPring-8 Center, Harima Institute, Sayo, Hyogo 679-5148, Japan, cKYOKUGEN, Osaka University, Toyonaka, Osaka 560-8531, Japan, dSwiss Light Source, Paul Scherrer Institut, 5232 Villigen, Switzerland, and eSPring-8/JASRI, Sayo, Hyogo 679-5198, Japan
*Correspondence e-mail: narumi@issp.u-tokyo.ac.jp
A synchrotron X-ray diffractometer incorporating a pulsed field magnet for high fields up to 40 T has been developed and a detailed description of this instrument is reported. The pulsed field magnet is composed of two coaxial coils with a gap of 3 mm at the mid-plane for passage of the X-rays. The pixel detector PILATUS 100K is used to store the diffracted X-rays. As a test of this instrument, X-ray diffraction by a powder sample of the antiferromagnet CoO is measured below the Néel temperature. A field-dependent
of CoO due to magnetostriction is observed up to 38 T.Keywords: X-ray diffraction; pulsed high magnetic field; pixel detector; magnetostriction; antiferromagnet.
1. Introduction
The study of materials by X-ray diffraction techniques using third-generation synchrotron sources has been performed extensively in recent years. Studies of magnetic materials are an example of a discipline that has particularly benefited from techniques making use of synchrotron radiation. In studying the properties of magnetic materials, it is useful to change the temperature, T, and/or applied magnetic field, B, in the measurements. Various new phenomena may appear at high fields (for a review, see Berthier et al., 2002) which can be studied by X-ray diffraction techniques. We have developed a high-field and low-temperature facility in conjunction with an X-ray diffractometer at the synchrotron source SPring-8 at Harima, Japan (Katsumata, 2005). At this facility a split-pair superconducting magnet up to 15 T is used in conjunction with a dilution refrigerator. In order to obtain a magnetic field of 15 T at the sample position by this split-pair magnet, each coil should generate a higher magnetic field of ∼20 T, which is close to the limit of modern superconductor technology. The magnetic field could be increased by a few more teslas using new materials such as high-temperature superconductors. In order to achieve much higher fields, a hybrid magnet or a pulsed field magnet is needed (for a review, see Herlach & Miura, 2003). A hybrid magnet, in which a water-cooled resistive magnet is inserted into a superconducting magnet, requires a large space for the installation as well as a facility for cooling. Therefore, a hybrid magnet is not suitable for synchrotron X-ray experiments. On the other hand, a pulsed field magnet can be placed on the goniometer with a relatively small power source. Moreover, the synchrotron X-ray beam is bright enough for time-resolved data collecting, which is inevitable in experiments exploiting pulsed magnetic fields. A synchrotron X-ray diffraction measurement under pulsed fields up to 20 T has been reported and the field dependence of the in Pr1−xCaxMnO3 (x ≃ 0.4) has been measured up to 16 T (Matsuda et al., 2004). In this paper we report technical details of our new X-ray diffractometer with pulsed high magnetic fields up to 40 T.
2. Apparatus
2.1. General considerations
Our diffractometer is installed at the beamline BL19LXU (Yabashi et al., 2001) at SPring-8. This beamline is equipped with a 27 m-long undulator (Hara et al., 2002) providing a of X-rays that is five times stronger than at any other beamline with a 5 m undulator at SPring-8. The X-ray beam of BL19LXU yields sufficient and energy resolution in the ranges 7–19 keV, 23–56 keV and 38–94 keV, utilizing the first, third and fifth harmonics, respectively, for typical experiments. A Si(111) double-crystal monochromator equipped with a cryogenic cooling system for high heat load is located in the optics hutch. Four experimental hutches are arranged in a tandem fashion and our hutch (#4) is located at the end of the beamline at approximately 135 m from the center of the undulator. The incident beam size in the experimental hutch is 1 mm × 2 mm and the estimated is ∼5 × 1013 photons s−1 for an X-ray energy of 30 keV. For a typical Bragg reflection from a single crystal the intensity of the diffracted beam is ∼109 photons s−1 and ∼103 photons µs−1. Consequently, an X-ray diffraction measurement using a pulsed field with a of a few milliseconds becomes feasible at this beamline.
2.2. Diffractometer
Fig. 1 shows schematically the diffractometer installed at experimental hutch #4. Incident X-rays come from the left-hand side of the figure and are reflected by a Pt-coated Si mirror with an angle of 0.1° to eliminate higher harmonic X-rays. An attenuator (ATT), composed of Cu and Al plates of various thicknesses, can be inserted in front of the sample. The reflected X-rays are introduced to the four-circle goniometer manufactured by Huber, Germany, with the 2θ (2th), θ (th), χ (chi) and φ (phi) rotations, in which the χ rotation is limited to ±5°. Here we define the z axis upwards, the x axis perpendicular to the plane of the paper in Fig. 1, and the y axis along the direction of the beam. The vertical and horizontal positions of the entire goniometer are adjusted by the z1 and z2 motors and the x2 motor in Fig. 1, respectively. The 2θ arm (2th) rotates in the horizontal plane (the xy plane). The position of the sample table, which is on the φ circle (phi), is adjusted by two translations, x and y. All components of the goniometer, including the worm gears and ball bearings, are made of non-magnetic materials. In addition, all the motors are more than 70 cm away from the center of the magnet, and their function is not affected by stray fields.
X-rays diffracted by the sample are detected by either a pixel detector or an et al., 2005). The PILATUS 100K single-module detector consists of an array of 487 × 197 pixels, with a pixel size of 0.172 mm × 0.172 mm. Each pixel is operated in single-photon-counting mode, i.e. only X-rays above a certain are counted. Pixels feature 20-bit counters, i.e. up to 106 photons can be counted in a pixel per exposure. The outstanding features of the detector are: (i) no noise contribution owing to or read-out noise leading to an excellent signal-to-noise ratio, (ii) short readout time of ∼5 ms, (iii) electronic gating, i.e. synchronization of the shutter signal with an external trigger is easily performed, and (iv) very simple operation at room temperature. The detector was operated with its set to 12 keV which is significantly below the X-ray energy of 24 keV used in the measurements. The detector was mounted on the 2θ arm of the diffractometer at a distance of 50 cm from the sample. In this geometry the angular resolution corresponding to the pixel size is approximately 0.018° in 2θ.
(APD). In this study we used a novel single-photon-counting pixel detector called PILATUS 100K that was developed at the Swiss Light Source (Hülsen2.3. Pulsed field magnet
The pulsed field magnet used in this study is a split-pair based on the non-destructive solenoid coil design developed by Kindo and collaborators (Kindo, 2001). Fig. 2 shows a schematic of the magnet.
We have made three magnets (#1, #2 and #3) with similar size (the coils have ∼65 mm diameter and ∼150 mm axial length). The #1 magnet has a spacer between the two coils made of a fiber-reinforced plastic, the #2 magnet has a stainless steel spacer, and the #3 magnet has a spacer made of Maraging steel. The magnet has an 18 mm cold bore and four windows with a slit width of 3 mm for passage of the X-ray beam. Each coil consists of ten layers of coaxially wound CuAg wires with a 2 mm × 3 mm . Fig. 4 shows the measured distribution of the magnetic field along the vertical axis of the #2 magnet. The is about 5% within ±5 mm from the center of the coil. A cryostat made of glass can be inserted into the magnet and the sample temperature can be lowered to liquid-helium temperatures.
coated with a Kapton sheet for insulation. This wire, manufactured by Showa Electric Wire & Cable, contains 24 wt.% Ag and has a tensile strength of 900 MPa. The two coils, with a spacer between them, are inserted into a Maraging steel tube for reinforcement. This tube is equipped with four windows that align with the windows in the spacer, and it has an inner diameter of 67 mm, outer diameter of 100 mm and height of 150 mm. Maraging steel has a tensile strength of 2200 MPa. The magnet is immersed in a liquid-nitrogen bath to decrease the resistance and to remove heat generated by the electric discharge. With a capacitor bank of 5 mF and 160 kJ, magnetic fields up to 42 T are obtained in the #2 magnet with a of 5.5 ms, as shown in Fig. 33. Example of a measurement
In order to test the performance of the apparatus, we have carried out Bragg diffraction measurements on CoO, a typical example of an antiferromagnet with cubic structure above the magnetic ordering temperature, TN = 290 K. The ground state of a free Co2+ ion is 4F. In a cubic crystal field Co2+ has a triply degenerate orbital ground state. The orbital angular momentum is not fully quenched, and these properties of CoO are sensitive to Below TN the changes from cubic to tetragonal (Tombs & Rooksby, 1950), with I4/mmm and lattice constants a = 3.015 Å and b = 4.2143 Å (Villars & Calvert, 1985). This phenomenon, known as magnetostriction, occurs to gain the magnetic energy at the cost of lattice energy (for a review see Kanamori, 1963). Nakamichi observed the field-dependent magnetostriction on a single crystal of CoO up to 1.1 T using a strain gauge (Nakamichi, 1965).
We measured X-ray diffraction by a powder sample of CoO under pulsed high magnetic fields. The powder sample was purchased from Sigma-Aldrich and had a purity of 99.99%. Fig. 5 shows the timing chart for the measurement. Before the pulsed field is applied, an X-ray diffraction pattern is recorded in zero field for 10 ms. After waiting for 1 s, a trigger pulse is applied to discharge the capacitor bank through the magnet. An X-ray diffraction image is stored for 1 ms after waiting for 1.7 ms from the trigger. The second exposure is carried out at a time interval where the variation of the magnetic field is at a minimum. During this exposure the magnetic field changes by about 5%. A final exposure of 10 ms after another delay of 1 s is used to confirm the zero field data. The change in the field of ∼5% owing to the chosen time window of 1 ms leads to a scatter in the distribution of planes. This distribution increases the line width of the reflections by a few %, which is below the limit of our detection. Therefore, the exposure time can be chosen such that the collected statistics are optimized compared with the width of the reflection. Fig. 6 shows an image of the powder diffraction rings taken at a magnetic field B = 20 T and a temperature T = 160 K.
To analyze the shift of the diffraction peaks as a function of applied magnetic field, all three data sets from the exposures shown in Fig. 5 were fitted with a Voigt function. The Bragg position under an applied field, including the zero-field calibration pattern, was then subtracted from the average of the two zero-field data sets. Fig. 7 shows the change in the (110) Bragg position with magnetic field measured at T = 160 K. We were able to follow the change of the Bragg reflection up to 38 T. To the best of our knowledge, this is the highest magnetic field achieved to date in an X-ray diffraction measurement. A full description and analysis of the results will be reported elsewhere.
4. Conclusions and future direction
We have developed an X-ray diffractometer equipped with a very high magnetic field facility. This diffractometer will be useful for the study of magnetic materials and superconductors, since many novel phenomena are expected to occur at high fields (for a review, see Berthier et al., 2002 and Herlach & Miura, 2003). A magnet consisting of a single solenoid coil is easier to construct than the split-pair magnet. However, the for such a single-coil magnet is limited and it is not suitable for diffraction measurements. Although the split-pair magnet is not easy to design, it has a great advantage over the solenoid magnet in that a scan over a wide range of space can be performed. In the case of a split-pair magnet, a strong attractive force is exerted between the two coils which limits the magnetic field available with it. We tested the #2 magnet and it was degraded at about 48 T. At the same time the spacer made of stainless steel was severely compressed. In order to avoid this damage, we employed a spacer made of Maraging steel and have been successful as described above.
We are planning to increase the energy of the capacitor bank which will increase both the peak field and the ) in the case of a solenoid coil. Heating of the sample by pulsed fields is not a serious problem in insulating samples. In metallic samples, depending on the conductivity, we use a powder sample or a pillar-shaped single crystal to reduce the heating. In the near future it may become possible to conduct an X-ray diffraction measurement up to ∼60 T at a temperature as low as ∼50 mK.
of the field. Doubling the energy and a new design of a coil will increase the peak field to ∼50 T and the to ∼10 ms. A longer makes the measurement easier of course. It is possible to insert a dilution fridge into the magnet as demonstrated by Kindo (2001When a pixel detector with higher readout time becomes available, time-resolved measurements will be feasible. For example, if we take a two-dimensional image within a few microseconds, then a diffraction pattern could be stored during the field pulse with a resolution of ∼0.04 T. Using this technique it will be possible to study the time dependence of a magnetic
in applied magnetic fields and many other time-dependent phenomena.Acknowledgements
We would like to thank S. W. Lovesey for useful comments. This work was partially supported by a Grant-in-Aid for Scientific Research from the Japan Society for the Promotion of Science.
References
Berthier, C., Lévy, L. P. & Martinez, G. (2002). Editors. High Magnetic Fields, Applications in Condensed Matter Physics and Spectroscopy. Berlin: Springer. Google Scholar
Hara, T., Yabashi, M., Tanaka, T., Bizen, T., Goto, S., Maréchal, X. M., Seike, T., Tamasaku, K., Ishikawa, T. & Kitamura, H. (2002). Rev. Sci. Instrum. 73, 1125–1128. Web of Science CrossRef CAS Google Scholar
Herlach, F. & Miura, N. (2003). High Magnetic Fields, Science and Technology. Singapore: World Scientific. Google Scholar
Hülsen, G., Brönnimann, Ch. & Eikenberry, E. F. (2005). Nucl. Instrum. Methods, A548, 540–554. Google Scholar
Kanamori, J. (1963). Magnetism, Vol. 1, edited by G. T. Rado and H. Suhl, pp. 189–199. New York: Academic Press. Google Scholar
Katsumata, K. (2005). Phys. Scr. 71, CC7–13. CrossRef CAS Google Scholar
Kindo, K. (2001). Physica B, 294–295, 585–590. CrossRef Google Scholar
Matsuda, Y. H., Ueda, Y., Nojiri, H., Takahashi, T., Inami, T., Ohwada, K., Murakami, Y. & Arima, T. (2004). Physica B, 346–347, 519–523. CrossRef CAS Google Scholar
Nakamichi, T. (1965). J. Phys. Soc. Jpn, 20, 720–726. CrossRef CAS Web of Science Google Scholar
Tombs, N. C. & Rooksby, H. P. (1950). Nature (London), 165, 442–443. CrossRef CAS Google Scholar
Villars, P. & Calvert, L. D. (1985). Pearson's Handbook of Crystallographic Data for Intermetallic Phases, Vol. 1, pp. 473. Ohio: American Society for Metals. Google Scholar
Yabashi, M., Mochizuki, T., Yamazaki, H., Goto, S., Ohashi, H., Takeshita, K., Ohata, T., Matsushita, T., Tamasaku, K., Tanaka, Y. & Ishikawa, T. (2001). Nucl. Instrum. Methods, A467–468, 678–681. CrossRef Google Scholar
© International Union of Crystallography. Prior permission is not required to reproduce short quotations, tables and figures from this article, provided the original authors and source are cited. For more information, click here.