Figure 2
Schematic representations of the experimental set-up in beamline X28C. (a) The stand set-up and the KinTek quench-flow apparatus are mounted separately on the precision motorized table along the beam path inside the hutch. The exposure chambers for both set-ups are aligned to maximize the intensity of the beam. The PC interface for the beam alignment, KinTek controller and the shutter controller are situated outside the X28C hutch. (b) The design of the flow cell is shown on the left-hand side. The sample residing inside the 0.8 mm-diameter flow-path is exposed through a 0.2–0.25 mm-thick and 16 mm-wide Vespel window. The right-hand panel schematically represents the position of the flow cell attached to the mounting plate (the entire apparatus is not shown) on the KinTek apparatus, the position of the horizontal slit assembly, the X-ray beam and beam-pipe after a vertical beam alignment. Reproduced with permission from Gupta et al. (2005BB10).