The MX2 macromolecular crystallography beamline: a wiggler X-ray source at the LNLS
aLaboratório Nacional de Luz Síncrotron (LNLS), Campinas, SP, Brazil, and bInstituto de Física de São Carlos, Universidade de São Paulo, Av. Trabalhador São-carlense, 400 São Carlos, SP 13560-970, Brazil
*Correspondence e-mail: beatriz@lnls.br, ipolikarpov@ifsc.usp.br
The Brazilian Synchrotron Light Laboratory [Laboratório Nacional de Luz Síncrotron (LNLS), Campinas, SP, Brazil] is the first commissioned synchrotron light source in the southern hemisphere. The first wiggler macromolecular crystallography beamline (MX2) at the LNLS has been recently constructed and brought into operation. Here the technical design, experimental set-up, parameters of the beamline and the first experimental results obtained at MX2 are described. The beamline operates on a 2.0 T hybrid 30-pole wiggler, and its optical layout includes collimating mirror, Si(111) double-crystal monochromator and toroidal bendable mirror. The measured 11 photons s−1 mm−2 (100 mA)−1. The beamline is equipped with a MarResearch Desktop Beamline Goniostat (MarDTB) and 3 × 3 MarMosaic225 CCD detector, and is controlled by a customized version of the Blu-Ice software. A description of the first X-ray diffraction data sets collected at the MX2 LNLS beamline and used for macromolecular solution is also provided.
at the sample position at 8.7 eV reaches 4.8 × 101. Introduction
The new era of structural molecular biology claims for structural comprehension of molecular phenomena in their cellular contexts, usually requiring the study of macromolecular assemblies, such as protein homo/hetero complexes or protein–nucleic acid complexes. Since X-ray crystallography remains as the method of choice for gathering this structural information, accounting for more than 83% of the macromolecular structures currently deposited in the Protein Data Bank (PDB), pressure lies on having better X-ray data from increasingly poorly diffracting crystals. The inherent experimental difficulties of the macromolecular X-ray diffraction experiments have been overcome in part by the use of increasingly brighter synchrotron sources. The improvements in synchrotron stations dedicated to macromolecular crystallography (MX) data collection include the use of insertion devices (IDs), leading to higher
even in the limiting wavelengths.A number of ID-equipped synchrotron stations are currently available and their outstanding relevance to structural biology can be evaluated in the BIOSYNC database (http://biosync.rcsb.org/BiosyncStat.html ; Jiang & Sweet, 2004). In 2003, 75% of the macromolecular structures solved and deposited in the PDB took advantage of synchrotron sources. In 2007, according to the BIOSYNC database, 6807 structures were solved and deposited in the PDB, with 5504 of these deposited structures using synchrotron sources, corresponding to 81% of the total. Furthermore, comparing the statistics from the National Synchrotron Light Source (NSLS) as an example, the recently set up and ID-equipped X29 beamline (Shi et al., 2006) alone was used in 32% of the total deposited structures solved at the light source and this tendency seems to be followed by the other synchrotron sources.
The Brazilian Synchrotron Light Laboratory [Laboratório Nacional de Luz Síncrotron (LNLS), Campinas, SP, Brazil] is the first commissioned synchrotron light source in the southern hemisphere and has been extensively used by the international scientific community, especially by research groups from Latin America, since 1997 (Rodrigues et al., 1998). The LNLS is a 1.37 GeV source operating at a maximum current of 250 mA with a critical photon energy reaching 2.08 keV. The current LNLS beamline dedicated to macromolecular crystallography (D03B-MX1) (Polikarpov et al., 1998) was commissioned in 1997 and has been widely used since then. According to BIOSYNC statistics, based on deposited file headers, to date 123 structures have been solved using this beamline. Although demand for macromolecular diffraction data collection is increasing rapidly in Brazil and Latin America, the LNLS storage-ring characteristics limit the number of conducted structural studies, partly as a result of the fact that the beam at wavelengths shorter than 1.0 Å is low and the design of the MX1 beamline is inadequate for typical MAD experiments. In addition, the of 4.5 × 1010 photons s−1 mm−2 (100 mA)−1 at 8.7 eV sets limits on the crystal size to be used for X-ray diffraction data collection even for well diffracting crystals.
In this work we describe the set-up, experimental parameters and first results obtained at the W01B-MX2, an ID-equipped beamline dedicated to macromolecular crystallography. The measured 11 photons s−1 mm−2 (100 mA)−1 at 8.7 eV. Some successful MAD/SAD experiments as well as native diffraction data set collection are described.
reaches 4.8 × 102. Hardware set-up
2.1. Beamline set-up
W01B-MX2 is the first LNLS beamline based on an insertion device. Owing to the storage-ring characteristics, a wiggler was the choice to improve the
density of the beamline, since undulators, at the low- and medium-energy synchrotron rings, would not be capable of producing synchrotron radiation in the X-ray region needed for macromolecular diffraction data collection. The 2.0 T hybrid multipolar wiggler was purchased from STI Optronics (Seattle, USA). A minimum gap of 22 mm generates the peak magnetic field and a total of 30 poles accounts for the length of the magnetic assembly (2.70 m). Various changes to the light-source subsystems were implemented prior to the wiggler installation, including an upgrade of the storage-ring RF system. The wiggler vacuum chamber was designed and built at the LNLS and consists of a 3 m-long stainless steel tube shaped into an approximately elliptical form, with pumping ports at its ends. The chamber inner surface is NEG-coated in order to improve the pressure in the long thin chamber. At the beamline front-end a graphite filter absorbs most of the thermal power generated by the wiggler. It consists of a 250 µm HOPG foil clamped between two water-cooled copper plates, assuring the dissipation, approximately, of a power of 1300 W.The MX2 optical layout consists of a collimating mirror that sends parallel synchrotron beam onto the two flat Si(111) monochromator crystals (Fig. 1). The monochromatic radiation after the monochromator is refocused by a toroidal bendable mirror. Similar optical and experimental design have been used on several other synchrotron facilities around the world, including the GCPCC beamline at CAMD (USA; Miller et al., 2001), Bl14 at BESSY (Germany; http://www.mx.bessy.de/userguide/beamlines/bl1/141_software.shtml ), BW6 at DESY (Germany; Buth et al., 1996) and beamline 10 at SRS (UK; Cianci et al., 2005), to name a few. Mirrors were purchased from SESO (France) and the silicon monochromator crystals were cut at the LNLS with dimensions of 8 × 45 × 45 mm (for the first crystal) and 8 × 35 × 110 mm (for the second crystal). The accuracy of the (111) plane orientation is ±0.1°. The beamline optical design was optimized to a maximum divergence of 3 mrad on the sample position with a 2:1 focusing ratio (Fig. 1). The first-mirror (990 mm of rhodium-coated ULE glass) specifications were defined in order to guarantee 90% reflectivity of 15 keV photons for an incident angle of 4.0 mrad. The monochromator crystals are mounted onto a Huber goniometer and vertical translation of the second crystal provides a fixed exit beam. Both the collimating mirror and the monochromator first crystal are water cooled. Cooling systems were designed taking into account the calculated absorbed power, of approximately 25 W and 35 W for the first mirror and monochromator crystal, respectively. The focusing toroidal mirror consists of 1200 mm-long rhodium-coated silicon. A mechanical bender tunes its longitudinal radius. Mirrors and monochromator mechanics are mounted on isolated concrete blocks to assure optical stability.
The LNLS MX2 beamline components are schematically represented in Fig. 2. Fluorescent screens are placed along the beamline for visual inspection of the beam (3, 7, 12 in Fig. 2), and the outcome photocurrent of a 5 µm-thick Al foil placed downstream of the monochromator (9 in Fig. 2) is used to tune the second crystal pitch. A vertically motorized tungsten-wire monitor allows analysis of the beam centroid and profile upstream of the collimating mirror (3 in Fig. 2) and a copper-blades beam-position monitor (BMP) is placed downstream of the mirror (5 in Fig. 2). Finally, to track vertical and horizontal displacement of the beam, a quadrant photodiode-based BPM was placed a few centimeters upstream of the sample position (14 in Fig. 2). The quadrant BPM was designed and built based on the model developed by Alkire and co-workers (Alkire et al., 2000).
Six 125 µm-thick beryllium windows separate the vacuum sections, four of them placed at the extremities of the mirror chambers. Pressure along the beamline is kept at approximately 5 × 10−9 mbar at the front-end and first mirror, 7 × 10−9 mbar at the monochromator and 5 × 10−9 mbar at the second mirror. Downstream of the second mirror chamber a mechanical pump keeps the pressure at approximately 10−3 mbar along the beam path until the experimental station.
2.2. Experimental station
The MX2 beamline is equipped with a MarResearch Desktop Beamline Goniostat (MarDTB) providing a high-precision φ-axis movement and automatic positional and directional alignment and adaptive beam optimization. The desktop motorized slits permit an automated beam shape optimization with individual settings for horizontal and vertical beam width, varying from 0 to 5 mm with 2.5 µm resolution. The manual crystal mounting is accompanied by a CCD microscope placed along the beam path, with an offset of 20° and coupled to a 4′′ LCD monitor; Φ-axis with motor-driven crystal centering allows mouse-click sample centering.
A 3 × 3 MarMosaic225 CCD detector is mounted on the desktop, providing a 50625 mm2 X-ray-sensitive surface. The read noise of 6 e− for 2 s readout allows reliable data collection in a very short time, consistent with the beamline high-photon-flux density. A Cryojet from Oxford Instruments and an Amptek XR100CR fluorescence detector complete the experimental station. The user-accessible experimental hutch is equipped with a key-based security system, which guarantees safety, even to the less experienced user.
3. Control software
At present, most of the LNLS MX2 beamline components are controlled by a customized version of the Blu-Ice software (McPhillips et al., 2002), developed by the Macromolecular Crystallography Group at the Stanford Synchrotron Radiation Laboratory (http://smb.slac.stanford.edu/public/research/developments/blu-ice/ ). The exceptions are the MarDTB goniostat and the MarCCD detector. Thus, the beamline alignment, energy selection and fluorescence scans are controlled via the Blu-Ice GUI, whereas crystal centering and X-ray data collection are carried out via the MarResearch software. The Blu-Ice software makes use of a well designed structure for controlling hardware in a heterogeneous computing environment system. The user interface of Blu-Ice has been developed in Tcl/Tk (Tool Command Language), a multiplatform scripting language. It makes use of a package termed DCS, Distributed Control System, composed of hardware servers (DHS) and a unique DCS server (DCSS). The DHS is responsible for the execution of low-level code. Since DCS uses the TCP/IP protocol, it is independent from the operational system of the machine. The DCSS controls all the messages between the Blu-Ice graphical interface and that of the DHS. It routes an incoming message from the users interface to the proper DHS based on the device it refers to. It also prevents the hardware from being damaged by controlling all the flow of commands. Fig. 3 shows an example of the Blu-Ice GUI frames. The Blu-Ice GUI allows the user to easily perform fluorescence scans and to select the data collection energy. Tuning of the pitch of the monochromator second crystal can also be easily performed when needed. For example, energy change from 7 to 12 keV and pitch optimization can be carried out in a few minutes. Integration of diffraction data collection control into the Blu-Ice software is in progress. DHSs have been written and are currently being tested. Users X-ray data are stored in the hard disks of the beamline linux machines for two weeks. Backups are mainly carried out using external hard disks, but data can also be retrieved by the users via ftp.
4. Experimental parameters
The MX2 beamline operates in the energy range 5 to 15 keV with maximum a) shows the as a function of energy measured at the sample position. Low-energy X-ray photons are absorbed by the beryllium windows, and the collimating mirror energy cut-off limits the at high energy. Beamline energy resolution capacity was determined by measuring the KMnO4 `white line' at low energies (Fig. 4b). An average value of ΔE/E = 2.2 × 10−4 was obtained based on the measured full width at half-maximum (FWHM) and the natural width of the Mn core level. Changing the vertical aperture of pre-monochromator slits produces no significant effect on the beamline resolution, indicating good collimation of the first-mirror outcoming beam. The toroidal mirror focuses the beam to a spot size (FWHM) at the sample position of 0.17 × 0.53 mm (vertical × horizontal) at a synchrotron radiation energy of 8 keV. The experimentally determined size of the focused beam agrees favorably with the theoretical value of 0.097 × 0.461 mm for the beam size at the same energy of X-ray radiation, obtained by simulation using the ray-tracing program XOP/Shadow (Welnak et al., 1992). The focused beam profile at the synchrotron radiation energy of 8 keV is shown in Fig. 5.
at 8 keV. Fig. 4(5. First X-ray crystallographic results
A number of native structures have been solved on the MX2 beamline, including a complex between human interleukin 22 and its receptor R1, solved at 1.9 Å resolution (PDB entry 3DLQ; Bleicher et al., 2008) and a double mutant of the human transthyretin solved at 1.38 Å resolution (PDB entry 3DGD). In the following sections we describe some examples of protein structures solved using SAD and MAD techniques.
5.1. SAD phasing
5.1.1. Sulfur SAD phasing
A number of successful cases of macromolecular phasing utilizing the weak anomalous signal of intrinsic S atoms present in cysteine and methionine residues have been reported (for example, Pal et al., 2008; Wang et al., 2006). This approach requires both a well diffracting crystal and a stable source. To conduct the experiment, we used a single native HEWL crystal for data collection conducted at the MX2 wiggler beamline. A complete and redundant data set was collected (see Table 1) and then integrated and scaled using MOSFLM (Leslie, 1992) and SCALA (Collaborative Computational Project, Number 4, 1994). SHELXD (Schneider & Sheldrick, 2002) was used to find the weak anomalous scatterers truncating the resolution at 1.9 Å. The best solution found after 1000 trials resulted in a CC/CCweak of 32.99/16.84 and clear discrimination after phasing in SHELXE. After 50 cycles of model building in ARP/wARP (Lamzin et al., 2001), 127 out of the 129 residues were built without human interference. The entire process of data reduction, phasing and model building took no more than 4 h. The built model (deposited in the PDB with entry 3EXD), and representative example of both anomalous and 2Fobs − Fcalc electron density maps are shown in Fig. 6(a).
|
5.1.2. Quick cryo-soaking approach
A number of derivatization techniques have been recently developed to solve the structures, which cannot be solved by et al., 2000; Nagem et al., 2001; Banumathi et al., 2003; Girard et al., 2003; Xie et al., 2004). One very successful technique implemented at the LNLS is quick cryo-soaking, which takes advantage of the introduction of a high concentration of anomalous scatterers directly into the cryo-protecting solution prior to flash-cooling of protein crystals for X-ray data collection (Dauter et al., 2000; Nagem et al., 2001, 2003, 2005). This allows heavy atom (e.g. I, Cs, Gd) diffusion into the and their rapid binding to the surface of macromolecules within it, enabling sufficient phasing power to solve novel structures from typical protein crystals (Nagem et al., 2002, 2004; Golubev et al., 2004; Rojas et al., 2004; Krauchenco et al., 2004).
methods. These techniques are particularly useful when MAD experiments cannot be conducted or for crystals that do not diffract at high resolution (DauterDuring commissioning of the MX2 beamline, we performed phasing of test protein crystals using the quick cryo-soaking approach. A single HEWL tetragonal crystal of typical size 100 µm was soaked in a mother liquor solution containing 0.5 M CsCl2 and 15% ethylene glycol for for 1 min. Experimental parameters of the data set collected from the flash-frozen crystal are shown in Table 1.
Two Cs sites were found using SHELXD after 1000 trials, truncating the resolution at 2.2 Å. The best solution yielded a CC/CCweak of 31.88/17.14 and a clear discrimination between the enantiomorphs was achieved as well using SHELXE. The improvements in the anomalous signal by the addition of the anomalous scatterers during the cryo-soaking are readily observed (see supplementary Fig. S11) with the increase in the average anomalous signal, taken as 〈|F+| − |F−|〉/〈F〉, from 0.08 for the native data set to 0.14 in the quick cryo-soaking data set.
5.2. SAD and from selenomethionine-labeled crystals
Selenomethionine-labeled crystals of a putative lipoprotein from Leptospira interrogans were grown in 0.2 M LiSO4, 0.1 M Tris-HCl pH 8.5 and 15% PEG 4000. factors were derived from the emission fluorescence spectrum of the crystal using the program CHOOCH (Evans & Pettifer, 2001). Then, diffraction data were collected at a single wavelength (0.9795 Å) corresponding to the maximum of f ′′. Table 1 summarizes the data collection statistics. All 14 expected selenium atoms were located by the hybrid search method (Grosse-Kunstleve & Adams, 2003) and initial phasing was calculated using Phaser (McCoy et al., 2007). Density modification was carried out using RESOLVE (Terwilliger, 2003) which was also used to build and refine a preliminary model. Further automatic model building was carried out using ARP/wARP (Lamzin et al., 2001). Fig. 6(b) shows the obtained from experimental phases. The is composed of two monomers and the final refined model comprises 814 residues. and analysis of the structure were recently published (Giuseppe et al., 2008) and atomic coordinates were deposited in the PDB (ID 3BWS).
Multiwavelength anomalous diffraction data were collected from selenomethionine-labeled crystals of a soluble Xanthomonas axonopodis protein (Guzzo & Farah, 2009). A MAD data set was collected at three different wavelengths for a single crystal that diffracted to 1.85 Å. These data sets were initially indexed in P6122, though symmetry was reduced to P61 at a later stage during model All expected selenium sites were found for the single monomer in the P6122 using SHELXD (Schneider & Sheldrick, 2002). Se sites were refined using SHARP (La Fortelle & Bricogne, 1997) which produced the first phase estimates up to 1.9 Å. Phases were refined by density modification using the programs SOLOMON (Abrahams & Leslie, 1996) and DM (Cowtan, 1994). Automatic model building was carried out with ARP/wARP (Lamzin et al., 2001). Structural of the atomic model of the two monomers in the P61 was performed using REFMAC (Murshudov et al., 1997), CNS (Brünger et al., 1998) and COOT (Emsley & Cowtan, 2004).
6. Conclusions and perspectives
MX2 is the first dedicated protein crystallography wiggler beamline in Latin America suitable both for native and anomalous data collection. Its technical parameters and characteristics as well as the first X-ray diffraction data sets collected at the beamline assure that it can be very useful for macromolecular
The beamline will provide an important technical support for rapidly growing research in structural molecular biology in Brazil and other Latin American countries. In the future both hardware and software capabilities of the beamline will be extended by installing a higher-flux fluorescence detector with better energy resolution and allowing for better integration of the software for beamline operation, data collection and data processing. Furthermore, the feasibility of sample-changer installation and remote data collection are currently being studied. Experience with the beamline operation and data collection at the first dedicated wiggler protein crystallography beamline will be useful for the development and installation of the ID macromolecular crystallography beamlines to follow. One also could expect that MX2 capabilities for MAD data collection will increase the number of novel structures solved in this part of the world.Supporting information
Native HEWL and HEWL cesium derivative data sets. DOI: 10.1107/S0909049508034870/hi5587sup1.pdf
Acknowledgements
We are very grateful to the Advanced Light Source team for helpful discussions during the assembly and commissioning of the beamline, in particular to James Holton, Alastair MacDowell, Corie Ralston and Thomas Earnest. We are very grateful to Scott McPhilips and the Blu-Ice developers' team for providing the software code and for their huge help with Blu-Ice installation and customization. We are thankful to Robert Sweet for a number of useful suggestions and friendly support. We are also indebted to Shaker Chuck Farah, Lucas Bleicher and Marcelo V. Liberato for their kindness in providing X-ray diffraction data sets collected at the LNLS MX2 beamline.
References
Abrahams, J. P. & Leslie, A. G. W. (1996). Acta Cryst. D52, 30–42. CrossRef CAS Web of Science IUCr Journals Google Scholar
Alkire, R. W., Rosenbaum, G. & Evans, G. (2000). J. Synchrotron Rad. 7, 61–68. Web of Science CrossRef CAS IUCr Journals Google Scholar
Banumathi, S., Dauter, M. & Dauter, Z. (2003). Acta Cryst. D59, 492–498. Web of Science CrossRef CAS IUCr Journals Google Scholar
Bleicher, L., Moura, P. R., Watanabe, L., Colau, D., Dumoutier, L., Reanauld, J.-C. & Polikarpov, I. (2008). FEBS Lett. 582, 2985–2992. Web of Science CrossRef PubMed CAS Google Scholar
Brünger, A. T., Adams, P. D., Clore, G. M., DeLano, W. L., Gros, P., Grosse-Kunstleve, R. W., Jiang, J.-S., Kuszewski, J., Nilges, M., Pannu, N. S., Read, R. J., Rice, L. M., Simonson, T. & Warren, G. L. (1998). Acta Cryst. D54, 905–921. Web of Science CrossRef IUCr Journals Google Scholar
Buth, G., Kölln, I. & Bartunik, H. D. (1996). Rev. Sci. Instrum. 67, 3369. CrossRef Google Scholar
Cianci, M. et al. (2005). J. Synchrotron Rad. 12, 455–466. Web of Science CrossRef CAS IUCr Journals Google Scholar
Collaborative Computational Project, Number 4 (1994). Acta Cryst. D50, 760–763. CrossRef IUCr Journals Google Scholar
Cowtan, K. (1994). Jnt CCP4/ESF-EACBM Newsl. Protein Crystallogr. 31, 34–38. Google Scholar
Dauter, Z., Dauter, M. & Rajashankar, K. R. (2000). Acta Cryst. D56, 232–237. Web of Science CrossRef CAS IUCr Journals Google Scholar
DeLano, W. L. (2002). The PyMOL Molecular Graphics System. DeLano Scientific, San Carlos, CA, USA. (http://www.pymol.org .) Google Scholar
Emsley, P. & Cowtan, K. (2004). Acta Cryst. D60, 2126–2132. Web of Science CrossRef CAS IUCr Journals Google Scholar
Evans, G. & Pettifer, R. F. (2001). J. Appl. Cryst. 34, 82–86. Web of Science CrossRef CAS IUCr Journals Google Scholar
Girard, É., Stelter, M., Anelli, P. L., Vicat, J. & Kahn, R. (2003). Acta Cryst. D59, 118–126. Web of Science CrossRef CAS IUCr Journals Google Scholar
Giuseppe, P. O., Neves, F. O., Nascimento, A. L. T. O. & Guimarães, B. G. (2008). J. Struct. Biol. 163, 53–60. Web of Science CrossRef PubMed CAS Google Scholar
Golubev, A. M., Nagem, R. A. P., Brandão Neto, J. R., Neustroev, K. N., Eneyskaya, E. V., Kulminskaya, A. A., Shabalin, K. A., Savel'ev, A. N. & Polikarpov, I. (2004). J. Mol. Biol. 339, 413–422. Web of Science CrossRef PubMed CAS Google Scholar
Grosse-Kunstleve, R. W. & Adams, P. D. (2003). Acta Cryst. D59, 1966–1973. Web of Science CrossRef CAS IUCr Journals Google Scholar
Guzzo, C. & Farah, C. (2009). In preparation. Google Scholar
Jiang, J. & Sweet, R. M. (2004). J. Synchrotron Rad. 11, 319–327. Web of Science CrossRef CAS IUCr Journals Google Scholar
Krauchenco, S., Nagem, R. A. P., da Silva, J. A., Marangoni, S. & Polikarpov, I. (2004). Biochimie, 86, 167–172. Web of Science CrossRef PubMed CAS Google Scholar
La Fortelle, E. & Bricogne, G. (1997). Methods Enzymol. 276, 472–493. Google Scholar
Lamzin, V. S., Perrakis, A. & Wilson, K. S. (2001). International Tables for Crystallography, Vol. F, pp. 720–722. Dordrecht: Kluwer. Google Scholar
Leslie, A. G. W. (1992). Jnt CCP4/ESF-EAMCB Newsl. Protein Crystallogr. 26. Google Scholar
McCoy, A. J., Grosse-Kunstleve, R. W., Adams, P. D., Winn, M. D., Storoni, L. C. & Read, R. J. (2007). J. Appl. Cryst. 40, 658–674. Web of Science CrossRef CAS IUCr Journals Google Scholar
McPhillips, T. M., McPhillips, S. E., Chiu, H.-J., Cohen, A. E., Deacon, A. M., Ellis, P. J., Garman, E., Gonzalez, A., Sauter, N. K., Phizackerley, R. P., Soltis, S. M. & Kuhn, P. (2002). J. Synchrotron Rad. 9, 401–406. Web of Science CrossRef CAS IUCr Journals Google Scholar
Miller, M. D., Phillips, G. N., White, M. A., Fox, R. O. & Craft, B. C. (2001). AIP Conf. Proc. 576, 734–740. CrossRef Google Scholar
Murshudov, G. N., Vagin, A. A. & Dodson, E. J. (1997). Acta Cryst. D53, 240–255. CrossRef CAS Web of Science IUCr Journals Google Scholar
Nagem, R. A. P., Ambrosio, A. L. B., Rojas, A. L., Navarro, M. V. A. S., Golubev, A. M., Garratt, R. C. & Polikarpov, I. (2005). Acta Cryst. D61, 1022–1030. Web of Science CrossRef CAS IUCr Journals Google Scholar
Nagem, R. A. P., Colau, D., Dumoutier, L., Renauld, J.-C., Ogata, C. & Polikarpov, I. (2002). Structure, 10, 1051–1062. Web of Science CrossRef PubMed CAS Google Scholar
Nagem, R. A. P., Dauter, Z. & Polikarpov, I. (2001). Acta Cryst. D57, 996–1002. Web of Science CrossRef CAS IUCr Journals Google Scholar
Nagem, R. A. P., Polikarpov, I. & Dauter, Z. (2003). Methods Enzymol. 374, 120–137. Web of Science CrossRef PubMed CAS Google Scholar
Nagem, R. A. P., Rojas, A. L., Golubev, A. M., Korneeva, O. S., Eneyskaya, E. V., Kulminskaya, A. A., Neustroev, K. N. & Polikarpov, I. (2004). J. Mol. Biol. 344, 471–480. Web of Science CrossRef PubMed CAS Google Scholar
Pal, A., Debreczeni, J. É., Sevvana, M., Gruene, T., Kahle, B., Zeeck, A. & Sheldrick, G. M. (2008). Acta Cryst. D64, 985–992. Web of Science CrossRef IUCr Journals Google Scholar
Polikarpov, I., Perles, L. A., de Oliveira, R. T., Oliva, G., Castellano, E. E., Garratt, R. C. & Craievich, A. (1998). J. Synchrotron Rad. 5, 72–76. Web of Science CrossRef CAS IUCr Journals Google Scholar
Rodrigues, A. R. D., Craievich, A. F. & Gonçalves da Silva, C. E. T. (1998). J. Synchrotron Rad. 5, 1157–1161. Web of Science CrossRef CAS IUCr Journals Google Scholar
Rojas, A. L., Nagem, R. A. P., Neustroev, K. N., Arand, M., Adamska, M., Eneyskaya, E. V., Kulminskaya, A. A., Garratt, R. C., Golubev, A. M. & Polikarpov, I. (2004). J. Mol. Biol. 343, 1281–1292. Web of Science CrossRef PubMed CAS Google Scholar
Schneider, T. R. & Sheldrick, G. M. (2002). Acta Cryst. D58, 1772–1779. Web of Science CrossRef CAS IUCr Journals Google Scholar
Shi, W. et al. (2006). J. Synchrotron Rad. 13, 365–372. Web of Science CrossRef CAS IUCr Journals Google Scholar
Terwilliger, T. C. (2003). Acta Cryst. D59, 38–44. Web of Science CrossRef CAS IUCr Journals Google Scholar
Wang, J., Dauter, M. & Dauter, Z. (2006). Acta Cryst. D62, 1475–1483. Web of Science CrossRef CAS IUCr Journals Google Scholar
Welnak, C., Anderson, P., Khan, M., Singh, S. & Cerrina, F. (1992). Rev. Sci. Instrum. 63, 865–868. CrossRef Web of Science Google Scholar
Xie, J., Wang, L., Wu, N., Brock, A., Spraggon, G. & Schultz, P. G. (2004). Nat. Biotechnol. 22, 1297–1301. Web of Science CrossRef PubMed CAS Google Scholar
© International Union of Crystallography. Prior permission is not required to reproduce short quotations, tables and figures from this article, provided the original authors and source are cited. For more information, click here.