Performance of single-photon-counting PILATUS detector modules

CROSSMARK_Color_square_no_text.svg

aPaul Scherrer Institut, CH-5232 Villigen PSI, Switzerland, and bDECTRIS Ltd, CH-5400 Baden, Switzerland
*Correspondence e-mail: philipp.kraft@psi.ch

(Received 22 December 2008; accepted 17 March 2009; online 7 April 2009)

PILATUS is a silicon hybrid pixel detector system, operating in single-photon-counting mode, that has been developed at the Paul Scherrer Institut for the needs of macromolecular crystallography at the Swiss Light Source (SLS). A calibrated PILATUS module has been characterized with monochromatic synchrotron radiation. The influence of charge sharing on the count rate and the overall energy resolution of the detector were investigated. The dead-time of the system was determined using the attenuated direct synchrotron beam. A single module detector was also tested in surface diffraction experiments at the SLS, whereby its performance regarding fluorescence suppression and saturation tolerance were evaluated, and have shown to greatly improve the sensitivity, reliability and speed of surface diffraction data acquisition.

1. Introduction

Synchrotron radiation diffraction experiments require detection systems with a large dynamic range, high count-rate capability, high detection efficiency, low or zero dark noise, and a very good point spread function to achieve the best possible signal quality. In modern diffraction and diffuse scattering experiments, the information to be gleaned from weak signals, that often lie very close to much stronger signals, is becoming increasingly important, such as in small-angle X-ray scattering and in protein crystallography (Welberry & Butler, 1994[Welberry, T. R. & Butler, B. D. (1994). J. Appl. Cryst. 27, 205-231.]; Wall et al., 1997[Wall, M. E., Clarage, J. B. & Phillips, G. N. Jr (1997). Structure, 5, 1599-1612.], and references therein). The problem of `blooming', associated with integrating devices such as CCDs, has until now set serious practical limitations to the effective dynamic range (Gruner et al., 2001[Gruner, S. M., Tate, M. W. & Eikenberry, E. F. (2001). International Tables for Crystallography, Vol. F, edited by M. G. Rossmann and E. Arnold, pp. 148-153. Dordrecht: Kluwer.]).

In addition, the need to record large data sets in times short enough to avoid radiation damage (a particular problem for organic systems) means that improved sensitivity and low background noise compared with those offered by established detector systems such as scintillators, image plates and CCD arrays are becoming a prerequisite for furthering the capabilities of diffraction facilities. PILATUS is a single-photon-counting hybrid pixel detector which has been designed to meet these requirements.

The PILATUS detector system is briefly described in the next section. Its characterization using monochromatic X-rays with respect to charge sharing, energy resolution and count rate behaviour is presented in §3[link]. We show some of the unique detector features with surface X-ray diffraction measurements of a NdGaO3 sample in §4[link].

2. System description

PILATUS is a multi-purpose silicon hybrid pixel detector system for X-rays operated in single-photon-counting mode, purpose-built for macromolecular crystallography at the Swiss Light Source (SLS) (Broennimann et al., 2006[Broennimann, Ch., Eikenberry, E. F., Henrich, B., Horisberger, R., Huelsen, G., Pohl, E., Schmitt, B., Schulze-Briese, C., Suzuki, M., Tomizaki, T., Toyokawa, H. & Wagner, A. (2006). J. Synchrotron Rad. 13, 120-130.]; Henrich et al., 2007[Henrich, B., Bergamaschi, A., Broennimann, Ch., Eikenberry, E. F., Kobas, M., Kraft, P. & Schmitt, B. (2007). Proceedings of the 16th International Workshop on Vertex Detectors, 23-28 September 2007, Lake Placid, NY, USA.]). A PILATUS detector module (cf. Fig. 1[link]) consists of a single pixelated silicon sensor bump-bonded1 to an array of 8 × 2 custom-designed and radiation-tolerant CMOS readout chips (ROC). The principal specifications of the PILATUS ROC and module are listed in Table 1[link]. The modularity of the system allows the construction of large area detectors of any array size.

Table 1
Principal properties of the PILATUS detector system

ROC design 0.25 µm CMOS (radiation tolerant layout)
Pixel size 172 µm × 172 µm
Counter depth 1048574 (20 bit counter pixel−1)
Threshold fine adjustment 6-bit DAC pixel−1
Readout clock frequency 66[\,{{2}\over{3}}] MHz
ROC array size 60 × 97 = 5820 pixel
ROC size 17.54 mm × 10.45 mm
Module array size 487 × 195 = 94965 pixel
Module active area 83.78 mm × 33.56 mm (continuous)
Module readout time 2.85 ms
[Figure 1]
Figure 1
PILATUS detector module (top) on a mounting bracket (grey) with module control board (green).

Incident photons are directly transformed into electric charge in the silicon sensor, which is transferred via the bump bond to the input of the ROC pixel. A schematic of the PILATUS readout chip pixel cell is presented in Fig. 2[link]. The analog front-end of a readout pixel consists of a charge-sensitive preamplifier (CSA) and an AC coupled shaper. The gain and shaping time of the CSA are adjusted with a global voltage (Vrf). An analog pulse from the shaper is discriminated against a threshold in the comparator (Comp) after amplification. The comparator threshold of each pixel is set with a global threshold voltage (Vcmp) and is further individually trimmed using an additional in-pixel 6-bit digital-to-analog converter (DAC). If the pulse amplitude exceeds the threshold, a digital signal is produced which increments the 20-bit counter. This detection principle is free of dark current and readout noise effects but requires precise calibration of the pixel threshold for optimum performance.

[Figure 2]
Figure 2
Architecture of the PILATUS ROC pixel cell.

The amplitude and width of the analog pulses after the shaper are correlated such that larger pulses are simultaneously wider. The CSA gain settings can be selected to meet the experimental requirements regarding energy and rate of the radiation. Three different settings of Vrf were defined for the detector characterization presented here: low-gain, medium-gain and high-gain CSA settings. Each setting requires an individual global threshold calibration, threshold trimming and rate correction. The global threshold of the module was calibrated with respect to the X-ray energy, and trim files were created and applied to minimize the pixel-to-pixel threshold dispersion. The PILATUS detector system and the calibration methods have been described in detail by Kraft et al. (2009[Kraft, P., Bergamaschi, A., Broennimann, Ch., Dinapoli, R., Eikenberry, E. F., Graafsma, H., Henrich, B., Johnson, I., Kobas, M., Mozzanica, A., Schlepütz, C. M. & Schmitt, B. (2009). IEEE Trans. Nucl. Sci. In the press.]).

3. Characterization

All the characterizations presented here were carried out at the surface diffraction station of the material science beamline X04SA of the SLS (Patterson et al., 2005[Patterson, B. D. et al. (2005). Nucl. Instrum. Methods Phys. Res. A, 540, 42-67.]), since better results are obtainable using monochromatic X-rays than using the internal calibration signal (CAL, cf. Fig. 2[link]) of the readout chip. Either the direct synchrotron beam in combination with absorbing filters or an elastic scatterer for homogeneous detector illumination was used.

3.1. Threshold scan

Many detector characteristics can be derived from threshold scans. For that purpose the module is homogeneously illuminated with monochromatic X-rays. Images of equal exposure time are taken while the global threshold (Eth) is increased with respect to energy for each frame.

The s-curve method (Dinapoli, 2004[Dinapoli, R. (2004). PhD thesis, Universite Montpellier II, France.]) is used to analyze the threshold scans. In our case the s-curve is extended by a linear factor to take into account the charge-sharing2 contribution. This representation of the s-curve is empirically based on observations and describes the data well. A previous energy calibration of the Vcmp bias controlling the global threshold of the module in order to convert the abscissa from voltage to energy is required,

[n\left(E_{\rm{th}}\right)={{1}\over{2}}\left[1-{\rm{erf}}\left({{E_{\rm{th}}-a_1}\over{2^{1/2}a_2}}\right)\right]\left[a_3+a_4\left(E_{\rm{th}}-a_1\right)\right].\eqno(1)]

The s-curve (1)[link] has a well defined inflection point a1 which is the threshold of the pixel for the incident X-ray energy (Ein) that was used. The parameter a2 is related to the electronic noise, charge sharing and energy spectrum3 of the incident X-rays. The magnitude of a3 is determined by the flux of the source and the exposure time. The slope a4 in the linear term models the charge sharing of the sensor.

Threshold scans of an individual PILATUS pixel for five different beam energies as a function of the global threshold are shown in Fig. 3[link]. For the purpose of comparison, the data of each scan in Fig. 3[link] are normalized to the number of counts recorded with the global threshold set at 50% of Ein.

[Figure 3]
Figure 3
Threshold scans of a pixel with low-gain CSA settings for different incident X-ray energies.

3.2. Charge sharing

Incident photons are converted into charge clouds inside the sensor, which are transported by the applied electrical field to the collection electrodes. Owing to diffusion and Coulomb repulsion, the diameter of the charge cloud increases while it drifts towards the collection electrode. In the case of conversion close to or at the border between pixels, the signal will therefore be shared among adjacent pixels (Ponchut et al., 2002[Ponchut, C., Visschers, J. L., Fornaini, A., Graafsma, G., Maiorino, M., Mettivier, G. & Calvet, D. (2002). Nucl. Instrum. Methods Phys. Res. A, 484, 396-406.]; Bergamaschi et al., 2008[Bergamaschi, A., Broennimann, Ch., Dinapoli, R., Eikenberry, E. F., Gozzo, F., Henrich, B., Kobas, M., Kraft, P., Patterson, B. & Schmitt, B. (2008). Nucl. Instrum. Methods Phys. Res. A, 591, 163-166.]).

Charge sharing in the predecessor of the current PILATUS detector system was characterized using an infrared laser to determine the effective charge collection area (Broennimann et al., 2002[Broennimann, Ch., Baur, R., Eikenberry, E. F., Fischer, P., Florin, S., Horisberger, R., Lindner, M., Schmitt, B. & Schulze, C. (2002). Nucl. Instrum. Methods Phys. Res. A, 447, 531-535.]). Alternatively, the influence of charge sharing on the count rate can be directly derived from threshold scans (Tlustos et al., 2004[Tlustos, L., Campbell, M., Heijne, E. & Llopart, X. (2004). IEEE Trans. Nucl. Sci. 51, 3006-3012.]). However, the method of measuring the charge sharing by means of threshold scans with monochromatic X-rays is advantageous over the laser method with respect to real operation conditions, because the infrared photons convert into charge at the surface of the sensor, whereas X-rays convert throughout the bulk of the silicon.

The parameter a4 in (1)[link] represents the slope of the linear decrease in count rate of the threshold scans occurring for global threshold settings below the incident photon energy. This slope is due to charge sharing among adjacent pixels. Since (1)[link] describes a threshold scan in absolute counts, a4 is normalized becoming independent of exposure time and photon flux, thus thereby being comparable. For the normalization, the count N50 is used, which is registered with the global threshold set to 50% of Ein. At this particular threshold setting, the count rate is independent of charge sharing and represents the number of correctly counted photons. Thus the normalized charge sharing slope is k = a4/N50.

Since charge sharing occurs at the borders of the sensor pixel, its effect depends on the ratio between the perimeter and pixel area. In a simple model, the normalized charge-sharing slope is associated with a corresponding fraction of area of the sensor pixel in which the charge of converting photons is shared, e.g. for k = 4.5% keV−1 at Ein = 12 keV the corresponding area is kEin/2 = 27.2% of the pixel. Regarding this area to be a strip along the pixel border, we can calculate its width. Using the example above, the strip width is 12.6 µm for a normal PILATUS pixel. Assuming the width to be the same for any pixel size, k can be calculated for different pixel sizes from the known k of a particular pixel size.

The normalized charge-sharing slopes of the PILATUS were determined from threshold scans taken with 8, 10, 12, 14 and 16 keV X-rays. The above-mentioned geometrical considerations were applied to calculate the k value for the large sensor pixels spanning the gaps between the readout chips4 (cf. Fig. 4[link]) and compared with the measurements. The averaged k value for normal-sized pixels and the averaged k value for the larger pixels with the calculated k value according to the above model are shown in Fig. 5[link].

[Figure 4]
Figure 4
Light microscope image of the sensor flip side with normal pixels (square) and large pixels (rectangular) spanning the gaps between readout chips.
[Figure 5]
Figure 5
Average normalized charge-sharing slope (k) at different X-ray energies for normal-sized and large pixels including the calculated k for large pixels.

3.3. Energy resolution

A threshold scan curve of a pixel corresponds to the integrated energy spectrum of the X-ray source above the threshold. The electronic noise of the sensor, and the analog and digital front-end, further broaden the spectrum. Hence, the derivative of a threshold scan performed with monochromatic X-rays yields a spectral peak in which the apparent width is a measure of the energy resolution of the pixel. In addition to the resolution of a single pixel, the pixel-to-pixel threshold dispersion has to be included for the overall energy resolution (OER) of the detector system. Since noise in the threshold scan data is enhanced by numeric differentiation, several thousand incident photons per pixel are required to obtain a reliable peak. To achieve better statistics, the threshold scans of a module were averaged before differentiation, i.e. the average count of all pixels was calculated for each threshold value. Thereby, the threshold dispersion is also taken into account.

The derivative of an averaged threshold scan of a trimmed5 module for 12 keV X-rays is shown in Fig. 6[link]. The constant tail towards low threshold energies stems from charge-sharing effects which only marginally affect the peak shoulder towards high energies. A Gaussian is fitted to the data in the region of the right shoulder including the peak. The OER of the system is given by the full width at half-maximum [FWHM = 2(2ln2)1/2σ] of the peak, where σ2 is the variance of the fitted Gaussian.

[Figure 6]
Figure 6
Derivative of an averaged threshold scan for 12 keV X-rays fitted with a Gaussian.

The PILATUS module was trimmed for three different CSA gain settings and for each setting a threshold scan was taken using 8 keV X-rays. The OERs obtained by the described method from the scans are listed in Table 2[link]. The module was also trimmed at 8, 10, 12, 14 and 16 keV for low-gain CSA settings, and a threshold scan was recorded for each energy to study the energy dependence of the OER. The resulting OER as a function of incident X-ray energy is shown in Fig. 7[link].

Table 2
Overall energy resolution of a trimmed PILATUS module for 8 keV X-rays and three different CSA gain settings

  Low gain Medium gain High gain
FWHM (eV) 1082 ± 3 1038 ± 9 863 ± 4
[Figure 7]
Figure 7
Overall energy resolution of a trimmed PILATUS module for different X-ray energies using low-gain CSA settings. Inset: relative overall energy resolution.

Since electronic noise scales with the reciprocal of the CSA gain, and the achieved threshold dispersion after trimming is smaller for high-gain CSA settings (Kraft et al., 2009[Kraft, P., Bergamaschi, A., Broennimann, Ch., Dinapoli, R., Eikenberry, E. F., Graafsma, H., Henrich, B., Johnson, I., Kobas, M., Mozzanica, A., Schlepütz, C. M. & Schmitt, B. (2009). IEEE Trans. Nucl. Sci. In the press.]), the OER improves with an increase in gain.

3.4. Dead-time

The counting behaviour of the PILATUS detector for high-intensity synchrotron beams was investigated at the surface diffraction station of the material science beamline X04SA of the SLS. The beamline features 15 filter sheets of Al, Ti and Mo of different thicknesses to control the attenuation of the beam. To avoid beam hardening owing to absorbers, the monochromator was set to 16 keV, because the higher harmonics (≥32 keV) are greatly suppressed by the lower wiggler flux at high energy, the low mirror reflectivity and the low Si scattering factors (Patterson et al., 2005[Patterson, B. D. et al. (2005). Nucl. Instrum. Methods Phys. Res. A, 540, 42-67.]). The remaining radiation from higher harmonics is further suppressed, since the efficiency of the Si sensor is less than 10% above 30 keV. Therefore the use of filter sheets to create different beam intensities is justified. The transmissions of the sheets were previously calibrated using the PILATUS detector at low intensity (<100000 counts pixel−1 s−1). In doing so, the direct synchrotron beam was defocused on the module such that the spot size was several millimetres in diameter. The same set-up was used to investigate the counting behaviour of the detector. For each filter transmission which was increased in small steps, a frame was recorded. The exposure time was set to 20 ms, preventing the counter overflow at very high intensities. Data were recorded with medium- and low-gain CSA settings at Eth = 6, 8 and 10.7 keV.

The time structure of the SLS during the experiment was a flat-filled electron beam of ton = 780 ns with 390 electron bunches of approximately 50 ps length every 2 ns, followed by a gap of 180 ns containing a single bunch. Regarding the X-ray delivery as being Poisson distributed during ton from the detector point of view and considering the given time structure of the beam, a Monte Carlo (MC) model can be employed to simulate the response of the detector to the incident rate (Bateman, 2000[Bateman, J. E. (2000). J. Synchrotron Rad. 7, 307-312.]).

MC data were generated for 100 different dead-times between 50 and 250 ns. The MC data for each dead-time were parameterized and the parameterizations compared with the experimental data of a pixel (cf. Fig. 8[link]). By means of minimum χ2 between parameterization and experimental data, the corresponding dead-time (τ) was determined. The average τ, obtained in this way, with respect to the relative global threshold setting (Eth/Ein), are presented in Fig. 9[link]. The shorter τ for low-gain CSA settings originates from leaner pulses entering the comparator, hence the time to resolve two successive pulses is less than for medium-gain CSA settings. Common for both CSA settings is a monotone drop in τ for increasing Eth/Ein. Experimental data taken at Eth/Ein = 83% show a significant deviation from the Monte Carlo model at incident rates above 1 × 106 photons s−1, meaning that pile-up effects become dominant and thus the experimental data are insufficiently described by the utilized model (Laundy & Collins, 2003[Laundy, D. & Collins, S. (2003). J. Synchrotron Rad. 10, 214-218.]).

[Figure 8]
Figure 8
Detected count rate of a pixel as a function of the incident rate for 16 keV X-rays using 50% relative global threshold and medium-gain CSA settings. The Monte Carlo model prediction agrees well with the data.
[Figure 9]
Figure 9
Average dead-times (τ) for 12 and 16 keV X-rays with respect to the relative global threshold settings for medium gain (a) and low gain (b) CSA settings. The black error bars represent the statistical errors on the average τ, the red error bars represent the total statistical and systematic errors.

In case of a flat-filled synchrotron beam without gap or an X-ray tube, the photon delivery in time can be considered as uniform, and the loss in counting efficiency at high rates can be calculated using the analytic model (Bateman, 2000[Bateman, J. E. (2000). J. Synchrotron Rad. 7, 307-312.]),

[N_{\rm{obs}}=N_0\exp\left(-N_0\,\tau\right),\eqno(2)]

where Nobs depicts the detected rate and N0 the true incident photon rate. Therefore an offline rate correction in software can be applied. In order to minimize computing time for large series of exposures with constant exposure time (t), the rate correction is accomplished by means of a look-up table. A look-up table mapping recorded counts (Nobst) to the corresponding incident counts (N0t) using (2)[link] is created for a specific dead-time and a specific exposure time, when either of them is changed. After an image is transferred to the data acquisition computer, the counts of each pixel are replaced according to the look-up table. The incident rate per pixel should stay at least below the detectable maximum rate given by N0max = 1/τ, since a higher rate becomes ambiguous in terms of the detected rate and would lead to misinterpretation. The exposure time has to be adapted to the detected rate in order to avoid overflow of the counter because the counter starts to count again from zero after its range is exceeded. Also noteworthy is the loss of statistics owing to the reduced counting efficiency at high rates, meaning that the relative statistical counting error of a pixel is given by the number of detected photons, which is larger than the same error of the number of incident photons. This leads to a decrease of the signal-to-noise ratio in the data for an increase in rate. This simple rate correction will fail if the incident rate changes during an exposure, since the correction method assumes a constant rate per pixel. In the case of the above-mentioned time structure of the SLS, the error between (2)[link] and the Monte Carlo model is approximately 2% for an input rate of 420000 photons s−1 pixel−1 (630000 photons s−1 pixel−1) at 50% relative global threshold for low (medium) gain CSA settings.

4. Surface diffraction with the PILATUS 100K

A PILATUS 100K system consisting of one module has been used for surface diffraction at the material science beamline X04SA of the SLS since 2006. It was demonstrated by surface diffraction experiments that PILATUS provides data of superior quality and that measuring times are shortened owing to the fast readout (Willmott et al., 2007[Willmott, P. R., Pauli, S. A., Herger, R., Schlepütz, C. M., Martoccia, D., Patterson, B. D., Delley, B., Clarke, R., Kumah, D., Cionca, C. & Yacobi, Y. (2007). Phys. Rev. Lett. 99, 155502.]). The diffraction signal on the 20L crystal truncation rod (CTR) of a NdGaO3(110) surface was measured in order to demonstrate the unique detector features of fluorescence suppression and saturation tolerance.

4.1. Fluorescence suppression

Setting the pixel threshold to exactly 50% of the incident photon energy results in maximum detection efficiency, while avoiding double counting owing to charge sharing (Kraft et al., 2009[Kraft, P., Bergamaschi, A., Broennimann, Ch., Dinapoli, R., Eikenberry, E. F., Graafsma, H., Henrich, B., Johnson, I., Kobas, M., Mozzanica, A., Schlepütz, C. M. & Schmitt, B. (2009). IEEE Trans. Nucl. Sci. In the press.]). In some circumstances, however, it can be favourable to choose a higher threshold in order to suppress disturbing the fluorescent background at the expense of a slight decrease in efficiency (Ponchut & Zontone, 2003[Ponchut, C. & Zontone, F. (2003). Nucl. Instrum. Methods Phys. Res. A, 510, 29-34.]). In the case of NdGaO3, the Ga absorbs at 10.367 keV and fluoresces at 9.252 keV (Kα1), 9.225 keV (Kα2) and 10.264 keV (Kβ1), producing a significant background (Thompson et al., 2001[Thompson, A. C., Attwood, D. T., Gullikson, E. M., Howells, M. R., Kortright, J. B., Robinson, A. L., Underwood, J. H., Kim, K., Kirz, J., Lindau, I., Pianetta, P., Winick, H., Williams, G. P. & Scofield, J. H. (2001). X-ray Data Booklet. Lawrence Berkeley National Laboratory, Berkeley, CA, USA.]).

To demonstrate the effect of changing the threshold levels, the incident X-ray energy was set to 15.92 keV and the detector was positioned at the (201) point in reciprocal space, which is the weakest point on the CTR in between the (200) and the (202) Bragg peaks. A series of images was recorded while increasing the global threshold between 8 and 15 keV (cf. Fig. 10[link]). The total intensity of the (201) diffraction signal was obtained by summing all counts inside a region of interest (ROI) comprising those pixels containing the diffraction feature. To determine the average background, the counts of all pixels inside a second ROI in the form of a box below the signal were averaged (cf. inset of Fig. 11[link]). The background below the signal is equal to the average background multiplied by the number of pixels in the signal ROI. Subtracting this background from the integrated counts in the signal ROI yields the true diffraction signal.

[Figure 10]
Figure 10
Detector image (zoom) of the 201 signal of NdGaO3 using 15.92 keV X-rays and a global threshold of (a) 9.5 keV and (b) 12 keV.
[Figure 11]
Figure 11
Integrated signal, integrated background and integrated signal plus background of the 201 signal of NdGaO3 with respect to the global threshold using 15.92 keV X-rays. Inset: the white rectangle marks the ROI where the background was determined; the white shape on top of the rectangle marks the ROI for signal integration.

The integrated signal, integrated background and the total measured intensity inside the signal ROI as a function of the global threshold energy are shown in Fig. 11[link]. The background is significantly reduced when the global threshold exceeds the fluorescence energies of the Ga, as expected. In comparison, the true diffraction signal decreases only slightly over the scanned threshold range, owing to the loss of detection efficiency. As a result, the signal-to-background ratio is strongly enhanced from 2.7 at 8 keV threshold to 44.5 at 12 keV threshold (Fig. 12[link]).

[Figure 12]
Figure 12
Signal-to-background ratio of the 201 signal of NdGaO3 with respect to the global threshold using 15.92 keV X-rays.

4.2. Saturation tolerance

A frequent problem in diffraction experiments is the measurement of weak signals very close to strong signals. Owing to its radiation tolerant design and complete lack of blooming, the PILATUS detector allows exposures with respect to the weak signals to be optimized even if this involves complete saturation of any neighbouring strong signal. To investigate whether this is really true, or whether in fact heavy over-irradiation on one part of the detector affects other areas which are not saturated, the intensity distribution around the very intense (202) Bragg peak was recorded in two reciprocal-space scans taken with nominal filter transmission values of 1 ± 0.01 (high) and (1 ± 0.01) × 10−2 (low), under otherwise identical conditions.

The image taken at the nominal Bragg peak position is shown in Fig. 13[link]. The strong Bragg peak in the centre of the visible feature is heavily saturated when using the high-transmission value. Several contour lines are plotted on top of the image. Contours of 8000, 16000 and 32000 counts for the data taken with high (dark, blue) transmission are overlayed on the corresponding contour lines of the low (light, orange) transmission data (i.e. with 80, 160 and 320 counts).

[Figure 13]
Figure 13
Detector image taken at the nominal Bragg peak position (shaded frame in Fig. 14[link]). The central part is saturated to several orders of magnitude above the maximum detectable count rate. Plotted on top there are several contour lines for high (dark, blue) and low (light, orange) transmission data, respectively.

By stacking of the individual detector frames of each scan, artificial intensity volume data sets can be constructed.6 These volume data can be used to produce intensity maps or line cuts through the sampled reciprocal-space volume, or to visualize isosurfaces of the intensity distributions, as shown in Fig. 14[link]. Here, we compare the isosurfaces for the same nominal intensities of 10000 and 10000/100 = 100 counts for high and low transmission data, respectively. The fact that the two isosurfaces coincide excellently, within the noise of the data, proves that over-irradiation in one part of the images does not affect the performance of those pixels which are not saturated.

[Figure 14]
Figure 14
Intensity isosurfaces for 10000 (dark, blue) and 100 (light, orange) counts from data taken with filter transmissions of 1 (high) and 0.01 (low), respectively. The two isosurfaces coincide perfectly.

An important consequence of this is that weak signals near strong peaks can be measured with an unattenuated beam without the strong signal either affecting the weak signal or damaging the detector. Hence there is no need to use a beam stop to obscure the strong diffraction signal (e.g. a Bragg peak), and simpler measuring geometries can be used which offer more flexibility, reliability and speed. One should bear in mind, however, that very strong signals such as that produced by the direct focused synchrotron beam will still cause radiation damage to the detector in a few seconds (Sobott et al., 2009[Sobott, B. A., Broennimann, Ch., Eikenberry, E. F., Dinapoli, R., Kraft, P., Taylor, G. N., Willmott, P. R. & Rassool, R. P. (2009). J. Synchrotron Rad. Submitted.]). In addition, it should be mentioned that the above measurements were made using a focused beam produced by a wiggler. If focused undulator radiation is used (where the focus spot is usually much smaller compared with a wiggler beam), it is possible that unattenuated Bragg peaks could damage the detector.

5. Conclusions and outlook

PILATUS detector modules were carefully calibrated and characterized with monochromatic X-rays. The effect of charge sharing on the count rate is relatively small and significantly decreases for higher X-ray energies. The measured normalized charge-sharing slope for the larger sensor pixels is well described by the simple geometric charge-sharing model. The presented overall energy resolution is of interest if the threshold needs to be set above a fluorescent background or if position-sensitive spectroscopy is performed by subtracting images of different thresholds.

Offline rate correction is crucial for photon-counting detectors if high rate signals are measured. Correction in case of constant rate during exposures is accomplished with an analytic model and is routinely applied by the PILATUS detector systems. If rate fluctuations owing to a gapped synchrotron beam or changes in an investigated sample occur, this issue becomes more complex and requires special treatment depending on the reason for the fluctuations. Owing to the modularity of the PILATUS system, all of the reported methods and results also apply for large arrays of modules. Detectors with arrays of 60 modules (PILATUS 6M) and of 24 modules (PILATUS 2M) were realised. These detectors were calibrated by the same methods as the single module detector (PILATUS 100K) and are successfully operated at beamlines of the SLS.

It was verified in surface diffraction measurements that fluorescence suppression strongly improves the signal over background ratio and that low intensity data is not affected by a neighbouring saturated signal. Therefore, this detector is the best available instrument for modern diffraction and diffuse scattering measurements (Weber et al., 2008[Weber, T., Deloudi, S., Kobas, M., Yokoyama, Y., Inoue, A. & Steurer, W. (2008). J. Appl. Cryst. 41, 669-674.]).

Developments in the near future are expected to include faster framing, smaller pixels and thicker and high-Z sensors for improved detection efficiency at higher energies.

Footnotes

1Each readout channel is electrically connected to its corresponding detecting element through a microscopic indium ball.

2Sharing of signal charge between neighbouring pixels (cf. §3.2[link]).

3ΔE/E of the incident beam is < 2 × 10−4 [Si(111)] (Patterson et al., 2005[Patterson, B. D. et al. (2005). Nucl. Instrum. Methods Phys. Res. A, 540, 42-67.]).

4Two rows or columns of larger sensor pixels span the gaps between ROCs, thus there is no dead area between chips. The registered counts of two of these larger pixels are redistributed by software into three normal-sized pixels by introducing a virtual pixel in between (Kraft et al., 2009[Kraft, P., Bergamaschi, A., Broennimann, Ch., Dinapoli, R., Eikenberry, E. F., Graafsma, H., Henrich, B., Johnson, I., Kobas, M., Mozzanica, A., Schlepütz, C. M. & Schmitt, B. (2009). IEEE Trans. Nucl. Sci. In the press.]).

5The intrinsic threshold dispersion of a module is significantly reduced by pixel-to-pixel threshold trimming (Eikenberry et al., 2003[Eikenberry, E. F., Broennimann, Ch., Hülsen, G., Toyokawa, H., Horisberger, R., Schmitt, B., Schulze-Briese, C. & Tomizaki, T. (2003). Nucl. Instrum. Methods Phys. Res. A, 501, 260-266.]; Kraft et al., 2009[Kraft, P., Bergamaschi, A., Broennimann, Ch., Dinapoli, R., Eikenberry, E. F., Graafsma, H., Henrich, B., Johnson, I., Kobas, M., Mozzanica, A., Schlepütz, C. M. & Schmitt, B. (2009). IEEE Trans. Nucl. Sci. In the press.]).

6Note that these are distorted with respect to the reciprocal-space coordinates, unless a proper coordinate transformation is carried out, although this is unnecessary for the present purpose.

Acknowledgements

The authors would like to thank H. Rickert and F. Glaus from PSI and M. Naef from DECTRIS for module fabrication. R. Herger, D. Martoccia, S. Pauli, M. Bjoerck and D. Meister of PSI were instrumental in setting up and collecting data at the surface diffraction station of beamline X04SA at the SLS.

References

First citationBateman, J. E. (2000). J. Synchrotron Rad. 7, 307–312.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationBergamaschi, A., Broennimann, Ch., Dinapoli, R., Eikenberry, E. F., Gozzo, F., Henrich, B., Kobas, M., Kraft, P., Patterson, B. & Schmitt, B. (2008). Nucl. Instrum. Methods Phys. Res. A, 591, 163–166.  Web of Science CrossRef CAS Google Scholar
First citationBroennimann, Ch., Baur, R., Eikenberry, E. F., Fischer, P., Florin, S., Horisberger, R., Lindner, M., Schmitt, B. & Schulze, C. (2002). Nucl. Instrum. Methods Phys. Res. A, 447, 531–535.  Google Scholar
First citationBroennimann, Ch., Eikenberry, E. F., Henrich, B., Horisberger, R., Huelsen, G., Pohl, E., Schmitt, B., Schulze-Briese, C., Suzuki, M., Tomizaki, T., Toyokawa, H. & Wagner, A. (2006). J. Synchrotron Rad. 13, 120–130.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationDinapoli, R. (2004). PhD thesis, Universite Montpellier II, France.  Google Scholar
First citationEikenberry, E. F., Broennimann, Ch., Hülsen, G., Toyokawa, H., Horisberger, R., Schmitt, B., Schulze-Briese, C. & Tomizaki, T. (2003). Nucl. Instrum. Methods Phys. Res. A, 501, 260–266.  Web of Science CrossRef CAS Google Scholar
First citationGruner, S. M., Tate, M. W. & Eikenberry, E. F. (2001). International Tables for Crystallography, Vol. F, edited by M. G. Rossmann and E. Arnold, pp. 148–153. Dordrecht: Kluwer.  Google Scholar
First citationHenrich, B., Bergamaschi, A., Broennimann, Ch., Eikenberry, E. F., Kobas, M., Kraft, P. & Schmitt, B. (2007). Proceedings of the 16th International Workshop on Vertex Detectors, 23–28 September 2007, Lake Placid, NY, USA.  Google Scholar
First citationKraft, P., Bergamaschi, A., Broennimann, Ch., Dinapoli, R., Eikenberry, E. F., Graafsma, H., Henrich, B., Johnson, I., Kobas, M., Mozzanica, A., Schlepütz, C. M. & Schmitt, B. (2009). IEEE Trans. Nucl. Sci. In the press.  Google Scholar
First citationLaundy, D. & Collins, S. (2003). J. Synchrotron Rad. 10, 214–218.  Web of Science CrossRef IUCr Journals Google Scholar
First citationPatterson, B. D. et al. (2005). Nucl. Instrum. Methods Phys. Res. A, 540, 42–67.  Web of Science CrossRef CAS Google Scholar
First citationPonchut, C., Visschers, J. L., Fornaini, A., Graafsma, G., Maiorino, M., Mettivier, G. & Calvet, D. (2002). Nucl. Instrum. Methods Phys. Res. A, 484, 396–406.  Web of Science CrossRef CAS Google Scholar
First citationPonchut, C. & Zontone, F. (2003). Nucl. Instrum. Methods Phys. Res. A, 510, 29–34.  Web of Science CrossRef CAS Google Scholar
First citationSobott, B. A., Broennimann, Ch., Eikenberry, E. F., Dinapoli, R., Kraft, P., Taylor, G. N., Willmott, P. R. & Rassool, R. P. (2009). J. Synchrotron Rad. Submitted.  Google Scholar
First citationThompson, A. C., Attwood, D. T., Gullikson, E. M., Howells, M. R., Kortright, J. B., Robinson, A. L., Underwood, J. H., Kim, K., Kirz, J., Lindau, I., Pianetta, P., Winick, H., Williams, G. P. & Scofield, J. H. (2001). X-ray Data Booklet. Lawrence Berkeley National Laboratory, Berkeley, CA, USA.  Google Scholar
First citationTlustos, L., Campbell, M., Heijne, E. & Llopart, X. (2004). IEEE Trans. Nucl. Sci. 51, 3006–3012.  Web of Science CrossRef CAS Google Scholar
First citationWall, M. E., Clarage, J. B. & Phillips, G. N. Jr (1997). Structure, 5, 1599–1612.  Web of Science CrossRef CAS PubMed Google Scholar
First citationWeber, T., Deloudi, S., Kobas, M., Yokoyama, Y., Inoue, A. & Steurer, W. (2008). J. Appl. Cryst. 41, 669–674.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationWelberry, T. R. & Butler, B. D. (1994). J. Appl. Cryst. 27, 205–231.  CrossRef CAS Web of Science IUCr Journals Google Scholar
First citationWillmott, P. R., Pauli, S. A., Herger, R., Schlepütz, C. M., Martoccia, D., Patterson, B. D., Delley, B., Clarke, R., Kumah, D., Cionca, C. & Yacobi, Y. (2007). Phys. Rev. Lett. 99, 155502.  Web of Science CrossRef PubMed Google Scholar

This is an open-access article distributed under the terms of the Creative Commons Attribution (CC-BY) Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.

Journal logoJOURNAL OF
SYNCHROTRON
RADIATION
ISSN: 1600-5775
Volume 16| Part 3| May 2009| Pages 368-375
Follow J. Synchrotron Rad.
Sign up for e-alerts
Follow J. Synchrotron Rad. on Twitter
Follow us on facebook
Sign up for RSS feeds