First experiments on the Australian Synchrotron Imaging and Medical beamline, including investigations of the effective source size in respect of X-ray imaging

CROSSMARK_Color_square_no_text.svg

aCSIRO Materials Science and Engineering, Private Bag 33, Clayton South, Victoria 3169, Australia, bAustralian Synchrotron, 800 Blackburn Road, Clayton, Victoria 3168, Australia, cMonash Centre for Synchrotron Science, Monash University, Melbourne, Victoria 3800, Australia, dBragg Institute, ANSTO, PMB 1, Menai, NSW 2234, Australia, eSchool of Physics, Monash University, Melbourne, Victoria 3800, Australia, and fDepartment of Medicine (RMH/WH), University of Melbourne, Parkville, Victoria 3050, Australia
*Correspondence e-mail: andrew.stevenson@csiro.au

(Received 18 February 2009; accepted 13 October 2009; online 1 December 2009)

The Imaging and Medical beamline at the Australian Synchrotron achieved `first light' in December 2008. Here, the first experiments performed on the beamline are reported, which involved both X-ray imaging and tomography studies for a range of samples. The use of a plastic-edge phantom for quantitative measurements of contrast and resolution proved to be very instructive and helped to confirm certain parameter values such as the effective horizontal source size, detector resolution and average X-ray energy for the polychromatic beam.

1. Introduction

The Imaging and Medical beamline (IMBL) at the Australian Synchrotron [3.0 GeV (1/γ = 170 µrad); 200 mA; circumference 216 m; see Boldeman & Einfeld (2004[Boldeman, J. W. & Einfeld, D. (2004). Nucl. Instrum. Methods Phys. Res. A, 521, 306-317.]) for a description of the physics design] achieved `first light' in December 2008. This paper reports the first experiments by a team from CSIRO, Monash University and the Australian Synchrotron. These experiments involved both X-ray imaging and tomography studies for various samples, including medical/bio­medical and materials-science applications. The focus of this paper is on the imaging results obtained, including some fundamental experiments aimed at estimating the effective X-ray source size.

The experiments were performed in the second hutch (1B), with a minimum source-to-sample distance (R1) of 20 m. Sample-to-detector distances (R2) of up to 3 m were used. The current insertion device is an Advanced Photon Source (APS) type-A permanent-magnet wiggler, which was operated with a gap of 55 mm to protect a temporary Be window from excessive thermal loading. Given that hutch 2B (31.7–40.0 m) and the satellite building (including hutches 3A and 3B; sample position at 136 m) have already been constructed, there is interest in the quality of the X-ray imaging that can be achieved with this interim insertion device, and, ultimately, with the future superconducting multipole wiggler. The RMS electron beam size in the straight sections at the Australian Synchrotron is 320 µm horizontally and 16 µm vertically (1% coupling), with distributed dispersion of 0.1 m. These values correspond to Gaussian FWHM of 754 µm and 38 µm, respectively. The electron-beam deviation caused by the field of the APS wiggler is small in comparison with the electron-beam size and so it is the latter which dictates the X-ray source size. This large horizontal source size and the 20:1 aspect ratio are important factors to be investigated in respect of X-ray imaging on this beamline, especially with reference to spatial coherence for propagation-based phase-contrast imaging. In the case of analyzer-based phase-contrast imaging, similar issues will arise if the plane of diffraction is vertical.

The experiment end-station in the satellite building will not only enable the use of a wide X-ray beam for imaging large objects but also significantly increase the demagnification of the source. Reduced source demagnification will significantly limit the degree of phase contrast achievable in the horizontal direction in hutch 1B.

In conventional treatment of X-ray imaging, and implicit in the above discussion, the X-ray source is regarded as totally incoherent within the range of the source size. In reality, and especially in the case of synchrotron radiation, some degree of coherence across the spatial extent of the source is possible. Such partial coherence of the source may affect image quality, and our aim is to measure the effective source size.

2. Experimental

The synchrotron operated at 200 mA, with a beam decay to about 150 mA in the 12 h between injections. The experiments were performed in hutch 1B, with the beam delivered from the Be window in upstream hutch 1A through a He-filled tube. The He reduced ozone production and protected the 0.5 mm Be window in the aggressive environment caused by the white beam (no monochromators having been installed as yet). The high X-ray flux also necessitated the use of polished Al filters (total thickness 6.0 mm) to protect the CCD detector and minimize damage to the more sensitive samples. The filters were selected so that the X-ray image quality, after processing, was not noticeably compromised. Fig. 1[link] shows a schematic diagram of the experimental configuration used.

[Figure 1]
Figure 1
Schematic diagram of the experimental layout used for these experiments at IMBL.

The CCD used was a 10 MHz 16M FDI-VHR camera, capable of 12-bit and 16-bit operation, supplied by Photonic Science. It was operated at approximately 241 K via the multi-stage Peltier cooling, and with a chilled water supply to remove the heat generated by the Peltier. The CCD has 4872 × 3248 (horizontal × vertical) 7.4 µm pixels, is optically bonded to a straight fibre-optic bundle (no magnification), and employs a (P43) Tb-doped Gadox (gadolinium oxysulphide) input phosphor (5 mg cm−2). It is optimized for X-rays in the 2–30 keV range. The images were pre-processed in conjunction with flat-field and dark-current images, i.e. images without an object and without X-rays, respectively. Correlation of sequences of image frames from a static sample, collected over a period of time, showed no significant drift present, i.e. high correlation coefficients and (translational) movement of less than a single pixel.

The APS wiggler has 28 × 8.5 cm periods with a total length of 2.4 m. At a gap of 55 mm, the field is approximately 0.24 T and the deflection parameter K is 1.9 (see Lai et al., 1993[Lai, B., Khounsary, A. & Gluskin, E. (1993). Technical Bulletin ANL/APS/TB-11. Argonne National Laboratory/Advanced Photon Source, Argonne, IL, USA.]; https://aps.anl.gov/Science/Publications/techbulletins/TB-11.pdf). This quite small K-value means that this insertion device is behaving with considerable undulator character (see Clarke, 2004[Clarke, J. A. (2004). The Science and Technology of Undulators and Wigglers, Oxford Series on Synchrotron Radiation, No. 4. Oxford University Press.], p. 43). The critical energy is 1.4 keV and the first-order harmonic energy is 0.36 keV. The program SPECTRA 8.0 (Tanaka & Kitamura, 2001[Tanaka, T. & Kitamura, H. (2001). J. Synchrotron Rad. 8, 1221-1228.]; https://radiant.harima.riken.go.jp/spectra/index_e.html) was used to calculate the spectral brightness (Mills et al., 2005[Mills, D. M., Helliwell, J. R., Kvick, Å., Ohta, T., Robinson, I. A. & Authier, A. (2005). J. Synchrotron Rad. 12, 385.]) as a function of energy in steps of 0.1 keV, using the storage-ring parameter values from https://radiant.harima.riken.go.jp/spectra/asp.prm. This spectrum was converted into flux density within 1 mm2 at the centre of the beam and 20 m from the source. Allowance was then made for the effect of the various filters, windows and materials present in the beam path, and for the quantum efficiency of the detector. The spectrum calculation was performed for both a wiggler and an undulator; the final results shown in Fig. 2[link] have been smoothed by using a `running average' over 11 data points. The undulator calculation yields, as expected, a (final) spectrum peak with considerable harmonic structure. Table 1[link] shows the parameter values which characterize the (wiggler) spectrum at each stage after inclusion of the various factors (without applying the running average). The final results for the corresponding undulator calculation are also given in Table 1[link].

Table 1
Values of parameters used to characterize the X-ray spectrum after inclusion of the effects of windows, filters and other materials present in the beam path (wiggler-based calculation)

The final row (in italics) corresponds to an undulator-based calculation of the spectrum. The Al filter thickness subsumes a 12 µm Al window in the CCD for excluding light. We also note that the number of visible photons produced in the CCD is a function not only of the number of X-ray photons but also of their energy. Allowance for this energy dependence has little effect in the current context however, e.g. the final value of the weighted-average energy changes to 21.6 keV. NA = not applicable.

Factor Maximum flux density [photons s−1 (0.1% bandwidth)−1 mm−2] Energy (keV) for maximum flux density FWHM (keV) Weighted-average energy (keV)
None 3.4 × 1013 1.9 NA NA
+ He (6.5 m) 2.3 × 1013 2.9 3.6 4.3
+ Be (0.5 mm) 9.4 × 1012 4.5 3.8 5.8
+ Sigradur (0.1 mm) 6.8 × 1012 5.0 3.9 6.2
+ Kapton (0.1 mm) 5.0 × 1012 5.5 4.0 6.7
+ graphite (1 mm) 9.3 × 1011 7.8 4.5 8.9
+ Al (6.0 mm) 9.1 × 106 21.3 6.7 22.2
         
+ Gd2O2S:Tb (5 mg cm−2)§ 1.4 × 106 19.9 6.2 21.2
+ Gd2O2S:Tb (5 mg cm−2)§ 8.3 × 106 22.5 4.0 21.3
†All absorption coefficients are calculated using total cross sections (Zschornack, 2007[Zschornack, G. (2007). Handbook of X-ray Data. Springer: Berlin.]).
‡Glassy carbon CCD window.
§CCD quantum efficiency calculated as 1 − exp(−μmρs), where μm is the energy-dependent mass absorption coefficient and ρs is the `surface density' (or `phosphor concentration'); a more rigorous calculation of the energy-dependent CCD response, to be reported elsewhere, does not have any significant effect in the current context
[Figure 2]
Figure 2
Calculated spectrum used for X-ray imaging and tomography experiments. The short solid line indicates the peak position (19.9 keV), the short dashed line the weighted-average energy (21.2 keV), for the smooth curve (wiggler-based calculation). The jagged curve is for the undulator-based calculation.

3. Results

Fig. 3[link] shows X-ray images collected for (a) a gum leaf (R1 = 20.8 m; R2 = 155 cm) and (b) a coarse-fibre paper sample (R1 = 20.3 m; R2 = 159 cm). The experimental magnification M = (R1 + R2)/R1 is approximately 1.08. There is little X-ray absorption for these samples and the dominant contrast mechanism is phase contrast. The `salt and pepper' noise near the edges of these images is due to the rapid drop in beam flux here, an effect exacerbated by the flat-field correction.

[Figure 3]
Figure 3
X-ray phase-contrast images obtained for (a) a gum leaf and (b) a coarse-fibre paper sample.

In Fig. 4[link] the images of three fixed mouse tibiae are shown, one in each column [see Cornish et al. (2002[Cornish, J., Callon, K. E., Bava, U., Lin, C., Naot, D., Hill, B. L., Broom, N., Myers, D. E., Nicholson, G. C. & Reid, I. R. (2002). J. Endocrinol. 175, 405-415.]) for details of the samples]. The images reveal that the tibiae have quite distinct trabecular microstructures. The first row corresponds to M = 1.00; the second row to M = 1.08; the third row to M = 1.15. Absorption- and phase-contrast effects are present in these images to varying degrees, with the M = 1.00 (essentially `contact') images being dominated by absorption contrast. Whilst the effects are quite subtle, the increase in phase contrast and reduced demagnification of the X-ray source as R2 increases from 158 cm to 305 cm can be detected.

[Figure 4]
Figure 4
X-ray images obtained for three fixed mouse tibiae. The tibiae were obtained from mice treated with control (left column), α-MSH (middle column) and leptin (right column). Details are provided by Cornish et al. (2002[Cornish, J., Callon, K. E., Bava, U., Lin, C., Naot, D., Hill, B. L., Broom, N., Myers, D. E., Nicholson, G. C. & Reid, I. R. (2002). J. Endocrinol. 175, 405-415.]). The rows correspond to, from top to bottom, R1 = 22.5 m and R2 = 8 cm (M = 1.00); R1 = 20.5 m and R2 = 158 cm (M = 1.08); R1 = 20.5 m and R2 = 305 cm (M = 1.15).

The image of the dragon-fly presented in Fig. 5[link] is the result of combining more than 80 individual images, obtained by scanning the sample in 2 mm steps vertically and 6 mm steps horizontally across the beam to create a montage (R1 = 20.8 m; R2 = 155 cm). The images to the right are magnified sections from a wing and the head. Whilst absorption contrast is present, phase contrast is again the dominant mechanism, especially for the delicate wing structure [see Snigirev et al. (1995[Snigirev, A., Snigireva, I., Kohn, V., Kuznetsov, S. & Schelokov, I. (1995). Rev. Sci. Instrum. 66, 5486-5492.]) and Wilkins et al. (1996[Wilkins, S. W., Gureyev, T. E., Gao, D., Pogany, A. & Stevenson, A. W. (1996). Nature (London), 384, 335-338.]) for the fundamentals of propagation-based X-ray phase-contrast imaging].

[Figure 5]
Figure 5
X-ray image of a dragon-fly obtained by combining more than 80 individual images to create a montage. Magnified sections from a wing and the head are shown on the right.

A tomography data set was collected for another mouse tibia. A total of 900 images were collected over a 180° rotation in 0.2° steps with R1 = 21.6 m and R2 = 23 cm (M = 1.01). In addition to the usual pre-processing steps mentioned earlier, a phase-retrieval step was included for the tomography data. The algorithm used was that of Paganin et al. (2002[Paganin, D., Mayo, S. C., Gureyev, T. E., Miller, P. R. & Wilkins, S. W. (2002). J. Microsc. 206, 33-40.]), which assumes a homogeneous object and is based on the transport-of-intensity equation (Teague, 1983[Teague, M. R. (1983). J. Opt. Soc. Am. 73, 1434-1441.]). It requires a value of δ/β (= −2φ/μ) for the object, where the X-ray refractive index is given by n = 1 − δiβ, φ is the phase shift per unit length, and μ is the linear absorption coefficient (all these quantities being energy- or wavelength-dependent). As the results are not particularly sensitive to this parameter, we used a spectrally weighted average value (180) for hydroxylapatite [Ca5(PO4)3OH; ρ = 3.18 g cm−3], the major component of osseous tissue. A cone-beam reconstruction was performed using the conventional FDK algorithm (Feldkamp et al., 1984[Feldkamp, L. A., Davis, L. C. & Kress, J. W. (1984). J. Opt. Soc. Am. A, 1, 612-619.]). These data processing and analysis operations were all performed using the X-TRACT (version 4) software package; https://ts-imaging.net/Services/. Figs. 6(a)–6(d)[link] show reconstructed (xy) slices perpendicular to the direction of the tibia shaft at 1.72, 1.36, 0.99 and 0.62 mm from the top of the tibia. Fig. 6(e)[link] is a reconstructed yz slice and Fig. 6(f)[link] is a reconstructed xz slice, both at the longitudinal mid-line of the tibia.

[Figure 6]
Figure 6
X-ray tomography reconstructed slices for a leptin-treated fixed mouse tibia. (a), (b), (c) and (d) show xy slices perpendicular to the main shaft of the bone at 1.72, 1.36, 0.99 and 0.62 mm, respectively, from the top of the tibia. (e) and (f) are yz and xz slices, respectively, at the longitudinal mid-line of the bone.

The images presented thus far are all remarkable in that the relatively large horizontal source size and 20:1 aspect ratio do not appear to be significant factors for the geometrical conditions being employed. Close examination of the images, especially those for the tibiae in Fig. 4[link], do indicate that the resolution of fine details is superior in the vertical direction than in the horizontal. However, the degree of improvement is not consistent with a simplistic view of the effect of the X-ray source. Given that these observations are qualitative and to a certain extent subjective, it was decided that images should be collected for a plastic-edge phantom.

The plastic-edge phantom was studied previously with a laboratory-based microfocus X-ray source (Gureyev et al., 2008[Gureyev, T. E., Nesterets, Ya. I., Stevenson, A. W., Miller, P. R., Pogany, A. & Wilkins, S. W. (2008). Opt. Express, 16, 3223-3241.]). It provides a means of obtaining quantitative contrast and resolution values, and of simultaneously checking vertical and horizontal directions. This simple phantom consists of two 100 µm-thick polyethylene sheets, partially overlapped to have edges running vertical and horizontal. Given our experimental conditions, in particular the X-ray energy range, this phantom provides images with both absorption and phase contrast, but dominated by the latter, in the form of characteristic single black–white fringes. The images were analysed to provide contrast and resolution values. The observed (experimental) contrast values are obtained from the difference between the maximum and minimum intensity values divided by their sum. The observed resolution values are obtained from the spatial separation of these features, referred to the object plane. The observed data values for vertical and horizontal edges, affected by the horizontal and vertical source dimensions, respectively, are listed in Table 2[link] as a function of R2 (R1 = 21.6 m). The estimated standard deviations (e.s.d.s) for the contrast values are 0.3% and for the resolution values are 2 µm (these values are smaller than the effective pixel size as the positions of the maximum and minimum intensities were extracted from the data by fitting). These e.s.d.s are consistent with the agreement between values of contrast and resolution obtained for a single polyethylene edge and for a single polyethylene edge overlapped by a single polyethylene sheet.

Table 2
Summary of experimental and theoretical (least-squares fitting) contrast and resolution values for a plastic-edge phantom

Errors associated with the experimental quantities are discussed in the text.

R2 (cm) [M] Plastic-edge disposition Contrast C (%) experiment Contrast C (%) calculated Resolution R (µm) experiment Resolution R (µm) calculated
19.8 [1.01] Vertical 1.2 0.82 40 39
19.8 [1.01] Horizontal 1.3 0.84 41 39
69.8 [1.03] Vertical 2.6 2.2 43 44
69.8 [1.03] Horizontal 3.5 3.0 36 38
119.8 [1.06] Vertical 3.0 2.6 55 54
119.8 [1.06] Horizontal 5.2 5.3 37 38
169.8 [1.08] Vertical 2.8 2.4 64 65
169.8 [1.08] Horizontal 7.2 7.7 37 37

It can be shown that the contrast for a Gaussian-blurred edge can be expressed as

[C={{R^{\,\prime}\lambda\left|{\varphi{t}}\right|}\over{2\pi (2\pi{e})^{1/2}\sigma_{\rm{tot}}^2}},\eqno(1)]

and the resolution as

[R=\left(4\sigma_{\rm{tot}}^2+R^{\,\prime}\lambda\right)^{1/2},\eqno(2)]

where

[\sigma_{\rm{tot}}=\left[\sigma_{\rm{b}}^2+(M-1)^2\sigma_{\rm{s}}^2/M^2+\sigma_{\rm{d}}^2/M^2\right]^{1/2}.\eqno(3)]

The derivation of these formulae and the associated validity conditions are detailed by Gureyev et al. (2008[Gureyev, T. E., Nesterets, Ya. I., Stevenson, A. W., Miller, P. R., Pogany, A. & Wilkins, S. W. (2008). Opt. Express, 16, 3223-3241.]). R′ = R1R2/(R1 + R2) = R2/M is the effective propagation or `defocus' distance, λ is the X-ray wavelength, t is the thickness of the polyethylene, and σb, σs and σd are the standard deviations associated with the blurring of the edge, the source emissivity distribution (`source size') and the detector PSF, respectively. The second term on the right-hand side of (2)[link] is associated with diffraction and is generally relatively small, but it may be significant in certain cases. These equations only relate to phase-contrast effects: neglecting absorption is justified in this case, i.e. the polyethylene sheet only absorbs 0.4% at 20 keV.

A non-linear least-squares refinement program [see Gureyev et al. (2008[Gureyev, T. E., Nesterets, Ya. I., Stevenson, A. W., Miller, P. R., Pogany, A. & Wilkins, S. W. (2008). Opt. Express, 16, 3223-3241.]) for further details] was used to fit the experimental data given in Table 2[link] with the above equations. The parameters which can be refined are σb, σs(horiz), σs(vert), σd, t and the effective X-ray energy E. Given the limited amount of experimental data and the existence of significant correlations between certain parameters (t and E for example), the refined parameters were σs(horiz), σd and E. The value of φ was changed dynamically during the least-squares refinement as E was adjusted and was calculated using a density of 0.923 g cm−3. The value of σb was fixed at 4 µm in accord with the findings of Gureyev et al. (2008[Gureyev, T. E., Nesterets, Ya. I., Stevenson, A. W., Miller, P. R., Pogany, A. & Wilkins, S. W. (2008). Opt. Express, 16, 3223-3241.]) for the same plastic edges. Some preliminary refinements showed that σs(vert) was, as expected, quite small and so its contribution in (3)[link], compared with the term involving σd, was insignificant. σs(vert) was therefore fixed at zero in subsequent refinements. The value of t was set to 100 µm. The refined values for the remaining parameters, giving equal weight to all 16 experimental data points, were as follows: σs(horiz) = 361 (9) µm, σd = 18.9 (0.3) µm and E = 18.4 (1.0) keV. The final fit to the experimental data was reasonable, as represented by an R-factor (Hamilton, 1965[Hamilton, W. C. (1965). Acta Cryst. 18, 502-510.]) value of 3.30% (see also the calculated values for contrast and resolution in Table 2[link]), and the maximum correlation coefficient between refined parameters was only 0.40 [between σs(horiz) and σd]. The results obtained are very robust as large changes in the starting values for refined parameters produced no change in the final values. Fig. 7[link] shows the experimental intensity-profile data points corresponding to the M = 1.08 data summarized in Table 2[link]. The solid curves are calculated using the Kirchhoff formula for the X-ray wavefunction including the parameter values detailed above. These theoretical curves include the effect of absorption in the polyethylene sheet and this effect can be seen in the experimental data points.

[Figure 7]
Figure 7
Experimental and calculated (Kirchhoff formula) intensity profiles for the M = 1.08 data summarized in Table 2[link]. The solid data points and associated theoretical curve correspond to the profile across the horizontal plastic edge; the open data points and associated curve correspond to the vertical plastic edge.

A least-squares refinement was also performed using a weighting scheme based on the e.s.d.s mentioned above, a given weight being the reciprocal of the square of the associated e.s.d. This has the realistic effect of giving significantly greater weight to the contrast values relative to the resolution values. The corresponding refined parameter values were σs(horiz) = 346 (14) µm, σd = 19.2 (0.5) µm and E = 18.0 (0.5) keV. Whilst these values have not changed markedly, the (weighted) R-factor is now 5.53%, indicating a poorer fit to the data (the standard goodness-of-fit parameter has a value of 1.13, the optimum value being unity). The maximum correlation coefficient is now 0.78 (between σd and E).

The refined value(s) of σs(horiz) is in good agreement with the nominal source size for the Australian Synchrotron (320 µm). If there was a significant degree of coherence across the spatial extent of the source we would expect the refined value(s) to be somewhat smaller than the design value. The value obtained for σd, which is equivalent to 44.6 (0.6) µm (unit weights case) in terms of FWHM, is somewhat larger than expected considering the specifications of the Gadox phosphor, but not unreasonable. It could be that some factor not accounted for in our model is causing an apparent degradation of the detector resolution, e.g. the presence of the Al filters and the associated scattered radiation (see, for example, Illers et al., 2005[Illers, H., Buhr, E., Günther-Kohfahl, S. & Neitzel, U. (2005). Radiat. Prot. Dosim. 114, 214-219.]). Lastly, the value of E that we obtain is reasonable considering the results shown in Table 1[link] and Fig. 2[link].

4. Discussion and conclusions

The images and tomography results presented here show that the IMBL can produce high-quality data in a very simple `barebone' experimental configuration. Given the plans for the implementation of a large number of imaging modalities in the coming years, we expect this facility to evolve to the forefront of X-ray imaging and tomography research, enabling world-leading contributions in fields ranging from biomedical and pre-clinical imaging to materials science [see, for example, Lewis (2005[Lewis, R. A. (2005). Nucl. Instrum. Methods Phys. Res. A, 548, 23-29.]) and Stevenson et al. (2003[Stevenson, A. W. et al. (2003). Nucl. Instrum. Methods Phys. Res. B, 199, 427-435.]), respectively]. The foremost science drivers are the acquisition of a superconducting multipole wiggler, the installation of custom-designed monochromators for high-resolution imaging and the development of beam focusing devices, in parallel with the commissioning of the long beamline to 136 m.

The preliminary results and analysis for the plastic-edge phantom give parameter values such as the horizontal source size in good agreement with expectations. An important finding is that, for the experimental configuration used, the effective source size (horizontally) deduced from the images is close to the electron-beam size. The simple method used can be extended to encompass various X-ray optics, filters, detectors and other beamline components. Tests can readily be performed for different insertion-device gaps and filter combinations to characterize the source and the beamline.

Whilst the X-ray source size in the horizontal direction is large and the aspect ratio is 20:1, one needs to consider carefully the influence of other experimental parameters to quantify the imaging performance. For example, if σb = 4 µm, σs(horiz) = 400 µm, σs(vert) = 20 µm, σd = 20 µm and M = 1.1, equation (3)[link] will yield σtot(horiz) = 41 µm and σtot(vert) = 19 µm, a ratio of only 2.2:1. However, a high-resolution detector with σd = 2 µm will yield values of 36 µm and 4.8 µm, respectively, a ratio of 7.7:1. In the first case, the detector resolution is the dominant factor vertically and the source size dominates horizontally. In the second case, the detector resolution and source size contributions are almost equal vertically and the source size is even more dominant horizontally. If we use σd = 20 µm, but with M = 1.01, σtot(horiz):σtot(vert) = 20.6:20.2 = 1.02:1. However, to achieve M = 1.01 with R1 = 20 m requires R2 = 20 cm and consequently significantly reduced phase contrast. This can be seen from the above equations as R′ ≃ R2 here and so the numerator in (1)[link] will be significantly reduced. However, if R1 = 140 m (IMBL satellite building), then M = 1.01 can be achieved with R2 = 1.4 m and phase-contrast effects will be enhanced. It should be noted that this increase in R2 (R′) will also increase the diffraction term in (2)[link]. Finally, we should also point out that in these discussions the value of σb has been assumed to be quite small, as is the case for the polyethylene edges. In the case of `real' objects, σb is likely to be significantly larger and it will have an important role to play, e.g. it may reduce detector-resolution requirements.

The present investigations identify useful operating conditions for the prevailing instrument parameters and constraints. Significant phase-contrast effects are observed and used quantitatively to determine and confirm values of these instrument parameters. Access to the satellite building for experiments in the near future will allow imaging with combined high spatial resolution and enhanced phase contrast, as well as a much larger field of view.

Acknowledgements

The authors thank Professor Jill Cornish of University of Auckland, New Zealand, for her support; the mouse tibiae samples were provided through a collaboration between one of us (DEM) and Professor Cornish. The authors are grateful to XRT Ltd for their encouragement of this work. We value discussions with Tim Gureyev and Yakov Nesterets of CSIRO, and Greg LeBlanc and Eugene Tan of the Australian Synchrotron. Many other Australian Synchrotron staff members helped us make these first experiments possible. Special thanks go to Alex Oggenfuss (on leave from the Swiss Light Source), Julian Price, Robert Grubb, Mark Clift, Bryce Karnaghan, Craig Millen, David Wang and Jason Wirthensohn.

References

First citationBoldeman, J. W. & Einfeld, D. (2004). Nucl. Instrum. Methods Phys. Res. A, 521, 306–317.  Web of Science CrossRef CAS Google Scholar
First citationClarke, J. A. (2004). The Science and Technology of Undulators and Wigglers, Oxford Series on Synchrotron Radiation, No. 4. Oxford University Press.  Google Scholar
First citationCornish, J., Callon, K. E., Bava, U., Lin, C., Naot, D., Hill, B. L., Broom, N., Myers, D. E., Nicholson, G. C. & Reid, I. R. (2002). J. Endocrinol. 175, 405–415.  Web of Science CrossRef PubMed CAS Google Scholar
First citationFeldkamp, L. A., Davis, L. C. & Kress, J. W. (1984). J. Opt. Soc. Am. A, 1, 612–619.  CrossRef Web of Science Google Scholar
First citationGureyev, T. E., Nesterets, Ya. I., Stevenson, A. W., Miller, P. R., Pogany, A. & Wilkins, S. W. (2008). Opt. Express, 16, 3223–3241.  Web of Science CrossRef PubMed Google Scholar
First citationHamilton, W. C. (1965). Acta Cryst. 18, 502–510.  CrossRef CAS IUCr Journals Web of Science Google Scholar
First citationIllers, H., Buhr, E., Günther-Kohfahl, S. & Neitzel, U. (2005). Radiat. Prot. Dosim. 114, 214–219.  Web of Science CrossRef Google Scholar
First citationLai, B., Khounsary, A. & Gluskin, E. (1993). Technical Bulletin ANL/APS/TB-11. Argonne National Laboratory/Advanced Photon Source, Argonne, IL, USA.  Google Scholar
First citationLewis, R. A. (2005). Nucl. Instrum. Methods Phys. Res. A, 548, 23–29.  Web of Science CrossRef CAS Google Scholar
First citationMills, D. M., Helliwell, J. R., Kvick, Å., Ohta, T., Robinson, I. A. & Authier, A. (2005). J. Synchrotron Rad. 12, 385.  Web of Science CrossRef IUCr Journals Google Scholar
First citationPaganin, D., Mayo, S. C., Gureyev, T. E., Miller, P. R. & Wilkins, S. W. (2002). J. Microsc. 206, 33–40.  Web of Science CrossRef PubMed CAS Google Scholar
First citationSnigirev, A., Snigireva, I., Kohn, V., Kuznetsov, S. & Schelokov, I. (1995). Rev. Sci. Instrum. 66, 5486–5492.  CrossRef CAS Web of Science Google Scholar
First citationStevenson, A. W. et al. (2003). Nucl. Instrum. Methods Phys. Res. B, 199, 427–435.  Web of Science CrossRef CAS Google Scholar
First citationTanaka, T. & Kitamura, H. (2001). J. Synchrotron Rad. 8, 1221–1228.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationTeague, M. R. (1983). J. Opt. Soc. Am. 73, 1434–1441.  Google Scholar
First citationWilkins, S. W., Gureyev, T. E., Gao, D., Pogany, A. & Stevenson, A. W. (1996). Nature (London), 384, 335–338.  CrossRef CAS Web of Science Google Scholar
First citationZschornack, G. (2007). Handbook of X-ray Data. Springer: Berlin.  Google Scholar

© International Union of Crystallography. Prior permission is not required to reproduce short quotations, tables and figures from this article, provided the original authors and source are cited. For more information, click here.

Journal logoJOURNAL OF
SYNCHROTRON
RADIATION
ISSN: 1600-5775
Follow J. Synchrotron Rad.
Sign up for e-alerts
Follow J. Synchrotron Rad. on Twitter
Follow us on facebook
Sign up for RSS feeds