research papers
One-dimensional small-angle X-ray scattering tomography of dip-coated polyamide 6 monofilaments
aPaul Scherrer Institut, CH-5232 Villigen PSI, Switzerland, bEmpa, Swiss Federal Laboratories for Materials Testing and Research, Laboratory for Advanced Fibers, Lerchenfeldstrasse 5, CH-9014 St Gallen, Switzerland, and cTechnische Universität München, James-Franck-Strasse, 85748 Garching, Germany
*Correspondence e-mail: philipp.kraft@scs.ch
A synchrotron study is presented in which the concept of one-dimensional tomographic reconstruction of small-angle X-ray scattering patterns is applied to investigate polyamide 6 monofilaments, dip-coated with alumina particles. The filaments are scanned with a focused synchrotron beam and the resulting scattering patterns are recorded with a PILATUS 2M detector. The reconstructed sequence of SAXS images reflects the local nanostructure variation along the filament radius. In particular, the influence of coating process parameters on the polyamide 6 is investigated.
Keywords: SAXS; one-dimensional tomography; polyamide 6; Al2O3; dip-coating; PILATUS; Abel transform; BASEX; two-point Abel deconvolution.
1. Introduction
Synchrotron beamlines featuring high et al. (2006), scanning SAXS data from samples with axial symmetry is recorded at the cSAXS beamline of the Swiss Light Source (SLS) at the Paul Scherrer Institut (PSI). The goal is to test the feasibility of one-dimensional tomographic reconstruction by means of Abel inversion. Independently, this concept has been developed and evaluated by Stribeck et al. (2008a,b).
advanced optics and fast detectors are predestined for scanning microscopy experiments such as small-angle X-ray scattering (SAXS) tomography. SAXS tomography reveals information about the local nanostructure without destructive sample preparation. Based on the work presented by SchroerIn the present work the aforementioned method is employed to study thick polymer monofilaments coated with ceramic particles. The two materials exhibit completely different nanostructures of which the characteristic SAXS can be disentangled by tomographic reconstruction such that radial structure variations in the polymer are unveiled. The data are recorded with the novel PILATUS 2M detector (Kraft et al., 2009) featuring single-photon counting and a maximum frame rate of 31.4 Hz for the full detector at 88.2% duty cycle, which improves data quality and data acquisition remarkably compared with standard charged coupled device (CCD) X-ray detectors. Although the beamline is still in the commissioning phase and not fully equipped at the time of the measurements, a focused beam of about 20 µm full width at half-maximum parallel to the filament and 5 µm perpendicular to it is achieved which permits the sample to be scanned in 2 µm steps with minor overlap. Instead of measuring the transmitted beam with an or a positive-intrinsic negative (PIN) diode inside the beamstop (Schroer et al., 2006), the attenuated direct beam is measured with the PILATUS 2M in a second scan, since modulation of the beam intensity at the SLS is relatively small owing to top-up injection.
2. Experimental
2.1. Samples
The studied samples are commercial polyamide 6 (PA6) monofilaments.1 The diameter of filament 72 is 200 µm whereas the diameters of filaments 73 and 74 are twice as much (see Table 1). The density of the PA6 is provided by the manufacturer.
|
The monofilaments are treated in a dip-coating process with commercial alumina (Al2O3) particles2 embedded in a polyester-urethane (PU) binder system (Reifler et al., 2008). The alumina particles are approximately spherical, have an average diameter of 2 µm and a mass percentage with respect to the coating slurry of 39%. The samples are amongst others characterized by means of (SEM) as shown in the inset of Fig. 1. The coating with ceramic microparticles is expected to improve the properties of the filament, such as higher resistance against heat, impact, cut, abrasion, hydrolysis and light while retaining the flexibility and elasticity (Gadow & Von Niessen, 2002), as required in fabric, e.g. for ballistic protective or fireproof clothing.
The coating is accomplished with a custom-built dip-coating machine. After cleaning the filament with acetone it is transited through the slurry and subsequently dried and cured in two separate ovens. In order to find the optimum coating process parameters such that the covering is well adhering and mechanically resistive without losing the general properties of the pure filament, a set of filaments is dip-coated with varying withdrawal speeds and oven temperatures. The coating process parameters of filaments 72 and 73 are identical whereas filament 74 is processed with higher withdrawal speed and higher temperatures of the ovens (see Table 2). The higher withdrawal speed in the case of filament 74 results in a thicker coating. However, the coating thickness of filament 72 is smaller by 23% than the coating thickness of filament 73 although both are processed with identical coating parameters and only differ in diameter.
|
2.2. Set-up
The scanning microbeam SAXS experiments were carried out at the cSAXS beamline of the SLS at PSI, Switzerland. The basic measuring set-up is presented in Fig. 1. The incident beam (wavelength λ = 1 Å) is focused by means of the second monochromator crystal and a dynamically bendable mirror. The achieved footprint of the direct beam at sample level is investigated with a phosphor screen in front of a video camera with microscopic lens. The thereby determined beam widths are 20 µm parallel to the filament axis and 5 µm transversal to it. The distance between the sample mounted on a translation stage and the PILATUS 2M detector is l = 2.160 m. The detector's threshold is set to 2 Å. A flight tube filled with helium and terminated with a mylar window is mounted between sample and detector to reduce parasitic scattering from gas molecules. The filament is moved perpendicular with respect to the filament axis in steps of 2 µm and for each position a 1 s exposure is acquired. Each sample is measured twice. In the first data set the direct beam without beamstop is recorded using a silicon attenuator of thickness 2.4 mm before the sample to investigate beam stability (outside sample) and absorption (inside sample). The second data set contains the SAXS patterns using a beamstop after the sample and no beam attenuator. The parasitic scattering background of the set-up is determined with a few exposures in the SAXS configuration without sample.
2.3. Preprocessing and reconstruction
Owing to the axial symmetry and thermal gradients occurring radially inside a monofilament during its fabrication and the dip-coating process, radial structure variations in the PA6 can be anticipated. This means that regions with the same radial distance from the filament center (sheath) are likely to exhibit an identical local nanostructure and a characteristic scattering pattern reflecting it. If the monofilament is probed with a microbeam, the illuminated volume includes mostly several sheaths. Hence, the resulting scattering contains contributions from different local structures. In order to glean the characteristic scattering for a volume element at a distinct position, a tomographic method is required. The axial symmetry of the sample and the assumptions made regarding the local structure allow for one-dimensional tomographic reconstruction by means of the inverse Abel transformation from a scattering data set of a single transversal microbeam scan (Stribeck et al., 2008a).
Basically the workflow as depicted by Stribeck et al. (2008b) is applied for preprocessing and tomographic reconstruction of the presented data. The beam center and the transmission for each position in the sample is determined from the center of gravity and integration of the direct beam images, respectively. The images are centered and cropped to a square of size −0.3 nm−1 ≤ s12, s3 ≤ 0.3 nm−1, where s12 = x12/(λl), s3 = x3/(λl), x12 denotes the distance on the detector from the beam center transverse to the filament axis and x3 that parallel to the filament axis (see Fig. 1 and Figs. 2a and 2b). From the shift of the direct beam center owing to refraction the exact transverse dimension and the center of the filament is determined. Owing to the filament symmetry, each image apart from the central one has a mirror image at the same radial distance from the center. Thus the two related mirror images are added to increase the counting statistics. The parasitic scattering background weighted by the corresponding transmission is subtracted from each SAXS image. Dead pixels and the gaps between detector modules are filled exploiting the centro-symmetry of the SAXS patterns (see Figs. 2c and 2d). The area of the beamstop is left unaltered since tomographic reconstruction and the subsequent data analysis are unaffected by it.
The one-dimensional tomographic image reconstruction is accomplished by a matrix-vector multiplication. Both recommended algorithms are employed: the two-point Abel deconvolution (Dasch, 1992) and the BASEX Abel transform method (Dribinski et al., 2002).
3. Results and discussion
All recorded SAXS patterns exhibit isotropic scattering around the beamstop. In addition, a four-point pattern appears in the images taken within the PA6. The tomographic reconstruction reveals that the isotropic scattering is generated in the coating whereas the four-point pattern stems from the structure of the PA6 (see Figs. 2e and 2f). Neither below- nor ultra-reconstruction artefacts were noticed in the reconstructed SAXS data (Stribeck et al., 2008a).
3.1. Long period of the PA6
Since the scattering of PA6 is weak, the short exposure time leads to a low signal-to-noise ratio in the reconstructed SAXS patterns. Thus, automated analysis by means of the interface distribution function or the multidimensional chord distribution function (Stribeck, 2001, 2007) is not viable. However, the determination of the long period from the four-point pattern maxima, L12 = |1/s12L| and L3 = |1/s3L|, is found to be feasible. For this purpose the patterns are projected on the s3 axis and the s12 axis, respectively (Stribeck, 2000). The central scattering has to be masked in the case of the s12 projection in order to render the four-point scattering feature clearly visible. For each resulting scattering curve the monotonously decaying background is smoothed outside of the feature and interpolated under it. The feature itself is smoothed and the background subtracted yielding a single peak whereby determination of the long period is possible (Stribeck, 2007). None of the two applied reconstruction algorithms proves in this case to be notably less advantageous regarding the resulting data quality.
The observed four-point SAXS pattern from PA6 can be explained by two different structural models (Fronk & Wilke, 1985; Stribeck et al., 2008a). On the one hand a macro-lattice of block stacks with a longitudinal distance L3 between the hard-domain blocks is considered to yield such a pattern. The block stacks can be regarded as microfibrils consisting of linearly alternating hard and soft domains, with a transverse distance between the microfibrils of L12 (see Fig. 3a). Description of the PA6 fiber structure by this particular model is proposed by Bukošek & Prevoršek (2000). On the other hand a four-point SAXS pattern can be due to a system of tilted lamellar stacks with a long period of L = (s12L2 + s3L2)-2 between lamellae and a tilt angle with respect to the filament axis of α = ±arctan(L3/L12) (see Fig. 3b). This model is also applied by Murthy et al. (1997) to describe the structure of PA6 fibers. None of the aforementioned structural models can be excluded owing to the low signal-to-noise ratio in the present SAXS data. Even a core-sheath structure (Iwata et al., 2006) with two kinds of conformations is conceivable. Hence the structural parameters L(r)12, L(r)3 for the microfibrils model and the accordingly computed parameters L(r), α(r) for the tilted lamellar stacks model are presented in Fig. 4. The diameter of the PA6 strand of filament 72 is only half that of the other two filaments. Thus the scattering, in particular in the reconstructed data, is weaker which is reflected in the stronger modulation of the determined long periods. Considering the microfibrils model (see Fig. 3a), the obtained average long periods (L12 ≃ 17 nm, L3 ≃ 9 nm) are assumed to be of the order of the crystallite size. This is affirmed by measured PA6 crystal heights determined from wide-angle X-ray diffraction (Vasanthan, 2007) which are comparable with the present long periods.
The parameter L(r)12 obtained from measured non-reconstructed data is notably increasing for larger radii, which is observed for all samples (see Fig. 4). This behavior is not reflected in the reconstructed data and thus is not attributable to the PA6 structure. But it can be explained by the increase of isotropic scattering background towards larger radii which is subtracted from the four-point pattern (see §2.3), meaning that systematic errors owing to smoothing and interpolation can bias the in non-reconstructed data notably. The isotropic background becomes so dominant that the automatic algorithm even fails to determine a proper 20 to 30 µm before the position at which the beam leaves the PA6 domain. However, the parameter L(r)3 from non-reconstructed data is not remarkably differing from the same parameter determined from reconstructed data.
The long periods L(r)12 obtained from reconstructed data are very similar for filaments 72 and 73, namely a slight increase in L(r)12 beginning at r = 20 to 30 µm before the transition to the coating (see Fig. 4). This can be explained by the following considerations. Heating-up of the PA6 can lead to additional crystallization and thus to larger crystallites resulting in a remarkable change in structure parameters (Bukošek & Prevoršek, 2000). Heat absorption of the PA6 during the coating process is significantly higher for filaments 72 and 73 since their coating is thinner and the drying and curing times are 3.8 times longer, whereas thermally induced crystallization takes place at temperatures above 423 K (Vasanthan, 2004). Thus, it is likely that sufficient temperatures are reached in the PA6 which permit crystal growth during the curing process. Since the resulting temperature gradient points from the filament center to the surface, crystallites at higher filament radii grow more compared with crystallites closer to the filament center. Alternatively, the observed increase in L(r)12 close to the PA6 surface is explicable with a crystal-to-crystal transition which has been demonstrated to occur at temperatures of 433 K (Vasanthan, 2004).
By contrast, L(r)12 obtained for filament 74 stays constant. The thick coating of filament 74 isolates its PA6 well during the drying and curing process. Furthermore, the withdrawal speed is higher compared with filaments 72 and 73 which results in shorter drying and curing times. This results in minor heat absorption of the PA6. Hence, the nanostructure of filament 74 is expected to remain almost unaltered and thus represents best the state after the filament production, e.g. before the coating process.
During the filament production the heat dissipation at the surface of the PA6 filament results in a temperature gradient from the surface towards the filament center. This leads to a decreasing crystallite size in the opposite direction. Thus, the decrease in L(r)3 towards the PA6 surface of filament 74 can be explained by the decrease in crystallite size owing to production. Assuming that L(r)3 of filaments 72 and 73 originally followed the same trend as L(r)3 of filament 74 after filament production, it could be ascribed to the coating process that the expected decrease in L(r)3 towards the PA6 surface is cancelled out in filaments 72 and 73.
The aforementioned considerations apply analogously for the parameters L(r) and α(r) of the tilted lamellae structure model.
3.2. Linear absorption coefficient
The μPA6 can be calculated from the total formula of PA6 (C6H11NO), the density provided by the manufacturer and the wavelength of the radiation (see Table 1). The transmission measured when the direct beam travels through the center of the filament is given by Tcenter = exp[−(μPA6D + μc2d)] with D being the diameter of the PA6 strand, d the thickness of the coating and μc the of the coating. Therefore, μc is readily obtained (see Table 3) utilizing the values given in Table 1 and the measured absorption Tcenter. This method for determination of μc is found to be the most robust.
|
By means of He pycnometry the volume of a solid can precisely be detected by measuring the amount of displaced helium gas by the solid. Successive weighing of the same specimen permits computation of its density. The densities of the alumina powder and of the coating (with and without alumina powder) have been determined using this technique (Sánchez, 2007). The obtained densities permit the calculation of the density of the alumina proportion in the coating. Thus the for X-ray radiation can be computed for this proportion ( = 17.04 cm−1). Although this value considers only the alumina proportion, it approximates the real value closely owing to the low absorption power of the PU matrix compared with the alumina particles. The exact composition of the commercial PU-based binder system is not known and hence the calculation of its is impossible.
3.3. Dimensionality of aluminium particles
Determination of the average size of the alumina particles by means of the Guinier law requires data in the ultra SAXS regime and is thus impossible with the data at hand. However, the dimensions of the particles can be determined from the asymptotic form of the azimuthally integrated intensity (I) curves by the relation ln(I) ∝ −aln(s), where s = (s12 2+s3 2)1/2 (Roe, 2000). The fitted slope parameter3 a, which is for all samples close to 4 (see Table 3), indicates a spherical shape of the alumina particles.
4. Conclusions
Scanning SAXS data recorded across dip-coated PA6 exhibiting axial symmetry are reconstructed tomographically. The method is successfully applied and the subsequent data analysis reveals parameters indicating the local nanostructure in the PA6. Comparison with results in the literature discloses coincidence with two models which have been applied earlier to describe the structure of PA6. Notable local trends in those parameters indicate an influence of the heat treatment during the dip-coating process on the nanostructure of the filaments. To reliably attribute this observation to the influence of postprocessing, ideally both a processed and an unprocessed sample should be measured for comparison. Furthermore, the dimensionality of the alumina particles in the coating is determined and the
for the coating derived.Therefore, we conclude that the presented method permits investigation of the influence of the dip-coating process on the morphology and structure of a monofilament. The gained results could then be used to optimize the dip-coating process regarding the filament structure. Moreover, this method is suitable for studying and improving the morphology and local structure of filaments in general with respect to composition (e.g. bicomponent fibers), production conditions and processing.
Apparently SAXS images for one-dimensional tomography require better statistics with respect to the presented data in order to ease automated analysis, to extract more information and to achieve better accuracy of the obtained structural parameters. An estimated exposure time of 5 to 10 s per image instead of 1 s should provide sufficient intensity from the weakly scattering PA6. Consequently, the overall measurement period of a sample is increased. On the other hand, an increase in the scanning step size, and acquisition of images from the filament center to one sample edge only, could be considered to shorten the experiment without remarkable loss of information.
Measuring the attenuated direct beam with a position-sensitive detector reveals additional information such as the beam deflection. In particular, the reconstruction of the sample's local phase shift from the beam deflection, as it is obtained from filtered back projection tomography of differential phase contrast images (Pfeiffer et al., 2007), seems a possible extension of the presented technique. The method of deriving the unknown of the coating from the transmission measured in the filament center works well, since the obtained values agree soundly with the calculated approximation. Unfortunately this method is only applicable if one of the two linear absorption coefficients is known.
Footnotes
‡Current address: Supercomputing Systems AG, Technoparkstrasse 1, CH-8005 Zürich, Switzerland.
1Polyamide 6 monofilaments, type A172 by Monosuisse AG, Emmenbrücke, Switzerland (formerly Nexis Fibers AG, Emmenbrücke, Switzerland).
2α-Al2O3 Sumicorundum, high-purity alumina AA-2 [particle size: D50 = 2.05 µm (PIDS)] supplied by Sumitomo Deutschland GmbH, Düsseldorf, Germany.
3The results for SAXS images taken in the coating region are independent of the tomographic reconstruction, since the scattering is in both cases generated by the coating only.
Acknowledgements
We would like to thank F. A. L. Sanchéz for dip-coating the monofilaments. We gratefully acknowledge Dr N. Stribeck for making all of his processing functions available on the web. Provision of the BASEX algorithm in the form of a MATLAB function by Dr H. Reisler and Dr B. Karpichev is acknowledged.
References
Bukošek, V. & Prevoršek, D. C. (2000). J. Polym. Matter, 47, 569–592. Google Scholar
Dasch, C. J. (1992). Appl. Opt. 31, 1146–1152. CrossRef CAS PubMed Web of Science Google Scholar
Dribinski, V., Ossadtchi, A., Mandelshtam, V. A. & Reisler, H. (2002). Rev. Sci. Instrum. 73, 2634–2642. Web of Science CrossRef CAS Google Scholar
Fronk, W. & Wilke, W. (1985). Colloid Polym. Sci. 263, 97–108. CrossRef CAS Web of Science Google Scholar
Gadow, R. & Von Niessen, K. (2002). Ceramic Engineering and Science Proceedings 23, edited by H.-T. Lin and S. Singh, pp. 277–285, Westerville, Ohio, USA. Google Scholar
Iwata, T., Aoyagi, Y., Tanaka, T., Fujita, M., Takeuchi, A., Suzuki, Y. & Uesugi, K. (2006). Macromolecules, 39, 5789–5795. Web of Science CrossRef CAS Google Scholar
Kraft, P., Bergamaschi, A., Broennimann, Ch., Dinapoli, R., Eikenberry, E. F., Henrich, B., Johnson, I., Mozzanica, A., Schlepütz, C. M., Willmott, P. R. & Schmitt, B. (2009). J. Synchrotron Rad. 16, 368–375. Web of Science CrossRef CAS IUCr Journals Google Scholar
Murthy, N. S., Zero, K. & Grubb, D. T. (1997). Polymer, 38, 1021–1028. CrossRef CAS Web of Science Google Scholar
Pfeiffer, F., Bunk, O., Kottler, C. & David, C. (2007). Nucl. Instrum. Methods Phys. Res. A, 580, 925–928. Web of Science CrossRef CAS Google Scholar
Reifler, F., Clemens, F., Sánchez, F. A. L., Varga, K. & Hufenus, R. (2008). Fiber Society Spring Conference, pp. 133–134, Mulhouse, France. Google Scholar
Roe, R. J. (2000). Methods of X-ray and Neutron Scattering in Polymer Science. Oxford University Press. Google Scholar
Sánchez, F. A. L. (2007). Unpublished results. Google Scholar
Schroer, C. G., Kuhlmann, M., Roth, S. V., Gehrke, R., Stribeck, N., Almendarez-Camarillo, A. & Lengeler, B. (2006). Appl. Phys. Lett. 88, 164102. Web of Science CrossRef Google Scholar
Stribeck, N. (2000). ACS Symp. Ser. 739, 41–56. CrossRef CAS Google Scholar
Stribeck, N. (2001). J. Appl. Cryst. 34, 496–503. Web of Science CrossRef CAS IUCr Journals Google Scholar
Stribeck, N. (2007). X-ray Scattering of Soft Matter. Berlin/Heidelberg/New York: Springer. Google Scholar
Stribeck, N., Nöchel, U. & Almendarez-Camarillo, A. (2008b). Macromol. Chem. Phys. 209, 1976–1982. Web of Science CrossRef CAS Google Scholar
Stribeck, N., Nöchel, U., Fakirov, S., Feldkamp, J., Schroer, C., Timman, A. & Kuhlmann, M. (2008a). Macromolecules, 41, 7637–7647. Web of Science CrossRef CAS Google Scholar
Vasanthan, N. (2004). Textile Res. J. 74, 545–550. Web of Science CrossRef CAS Google Scholar
Vasanthan, N. (2007). J. Polym. Sci. B, 45, 349–357. CrossRef CAS Google Scholar
© International Union of Crystallography. Prior permission is not required to reproduce short quotations, tables and figures from this article, provided the original authors and source are cited. For more information, click here.