Figure 4
(Left) Performance measurement of the present differential pumping system. The pressure distribution curves were measured for the atmospheric pressure of air (filled circles) and helium (filled triangles) conditions. The calculated pressure for the optimized design is also indicated for air (empty circles) and helium (empty triangles). (Right) Pumping speed curves of the vacuum pumps used in the present system. |