VUV synchrotron radiation: a new activation technique for tandem mass spectrometry
aSynchrotron SOLEIL, L'Orme des Merisiers, 91192 Gif sur Yvette Cedex, France, bInstitute of Physics, University of Belgrade, Pregrevica 118, 11080 Belgrade, Serbia, and cINRA, U1008 CEPIA, Rue de la Géraudière, F-44316 Nantes, France
*Correspondence e-mail: alexandre.giuliani@synchrotron-soleil.fr
A novel experimental technique for tandem
and ion spectroscopy of electrosprayed ions using vacuum-ultraviolet (VUV) synchrotron radiation is presented. of trapped precursor ions has been performed by coupling a commercial linear quadrupole ion trap (Thermo scientific LTQ XL), equipped with the electrosprayed ions source, to the DESIRS beamline at the SOLEIL synchrotron radiation facility. The obtained results include, for the first time on biopolymers, photodetachment spectroscopy using monochromated synchrotron radiation of multi-charged anions and the single photon ionization of large charge-selected polycations. The high efficiency and signal-to-noise ratio achieved by the present set-up open up possibilities of using synchrotron light as a new controllable activation method in tandem of biopolymers and VUV-photon spectroscopy of large biological ions.Keywords: mass spectrometry; tandem mass spectrometry; ion; spectroscopy; VUV; activation; biopolymer; sequencing.
1. Introduction
m/z), whose distribution produces a Since the introduction of electrospray ionization (ESI) and matrix-assisted laser ionization (MALDI), has been recognized as a powerful tool for the study of the structures of biological molecules (Biemann, 2002). Particular instrumental arrangements offer the possibility of isolating a specific m/z window either in time or in space, leading to so-called tandem capabilities. Tandem also referred to as MS/MS or MS2, can be briefly described as the following sequence of events (Dass, 2001; de Hoffmann, 1996). First, a selected m/z window, corresponding to the targeted is isolated. In a second step, those precursor ions are activated, meaning that their internal energy is increased which eventually leads to the fragmentation of the precursor. Usually, this process, termed (CID; Haddon & McLafferty, 1968; Rozett & Koski, 1968), is carried out by inelastic collisions with a gas, such as helium or nitrogen. The product ions are finally analyzed according to their m/z giving a tandem Note that the previous isolation, activation and fragmentation procedure can be repeated further up to the nth level (MSn). The goal of tandem is to provide detailed structural information on a particular species of interest through successive fragmentation. In particular, tandem has proven extremely useful for determination of the primary structure of biopolymers (Biemann, 2002; de Hoffmann, 2001). However, the crucial point in the application of the MS2 technique to the study of large biopolymers is the need for an efficient activation method, which can produce an intense and selective fragmentation of such species.
(MS) is based upon the analysis of gas-phase ions (The aim of this paper is to show that vacuum-ultraviolet (VUV) synchrotron radiation may be used as an efficient and precise activation technique for tandem
with several important advantages over classical methods.Conventional activation methods in MS2 are based upon the vibrational excitation of the which can be performed via CID or infrared multiphoton dissociation (Little et al., 1994; Price et al., 1996). However, those slow heating activation techniques still impose some important limitations. Firstly, since the energy required for an ion to dissociate increases with its size, these methods may fail to efficiently cleave large biomolecules (Dass, 2001). Furthermore, dissociations by slow heating do not preserve non-covalent associations forming larger assemblies. In recent years, several new techniques allowing improved MS2 have emerged, such as dissociation (ECD) (Zubarev et al., 1998), dissociation (Fung et al., 2009) and femtosecond laser-based ionization/dissociation (Kalcic et al., 2009). However, they still have some limitations (such as low efficiency for dications and the impossibility to apply it on singly charged ions for ECD) or they need to be improved in terms of analytical capabilities. Therefore, there is a clear need in MS2 biological analysis for alternative modes of ion activations, which can improve the efficiency of the internal energy deposition and provide selective fast fragmentation.
A simple and straightforward means of depositing a large amount of internal energy into the ions of interest is absorption of energetic photons by the target, typically in the VUV range. Indeed the absorption of a single photon in this energy domain is enough to activate a given e.g. prolines). In addition, a very important advantage of the present method is the photon tunability, allowing the activation to be controlled by accessing different fragmentation/ionization limits. As a consequence, it might be possible for certain photon energy to localize specific structural properties of biomolecules (e.g. sites of non-covalent interaction). In that sense, tunability is of course mandatory not only for MS2 analytical applications but also for opening a large alley of research on the VUV spectroscopy of gas-phase biopolymer ions.
by a `clean' fragmentation/ionization (or photodetachment in the case of anions) process. The first advantage of such VUV-based activation techniques over previous ones would be the large quantity of analytical information (intensive and clean fragmentation), even for biological species where modern and usually very efficient activation methods such as ECD do not perform well (However, coupling an ion trap to the synchrotron beam is not a trivial task because of difficulties in achieving high enough ion densities and photon fluxes, even though the signal-to-noise ratio benefits from the possibility of performing longer irradiation of the trapped ionic species. The present linear ion trap spectrometer (LTQ XL) can confine large ionic densities with limited space charge using a helium bath. On the other hand, intense laser sources are unable to provide high-energy photons in a large tunable domain, while the conventional UV lamps are not brilliant enough for the purpose. Therefore, the only light source that can deliver sufficient photon fluxes over a wide energy range is synchrotron radiation, especially on third-generation sources, offering both an intrinsic tunability and a high et al., 1991); however, the limited success of the experiment came from their use of the white-beam emission from a bending magnet. The feasibility of coupling a Fourier-transform ion cyclotron resonance ion trap with a soft X-ray synchrotron radiation beamline has been demonstrated very recently by Thissen et al. (2008) and by Hirsch et al. (2009) for the coupling of a linear ion trap for the study of gas-phase atomic species and clusters. The first successful use of non-dispersed undulator radiation for spectroscopy of electrosprayed negative ions stored in a three-dimensional quadrupole ion trap, in the 290–80 nm wavelength range (4.3–15.5 eV), was preliminary reported only in 2009 (Giulliani, 2009). Still, it should be noted that both reported experiments (Thissen et al., 2008; Kravis et al., 1991) suffered from a rather low signal-to-noise ratio. Note that, very recently, Bari et al. (2011) have reported the coupling of a Paul trap with a synchrotron radiation beamline at the BESSY synchrotron facility.
To our knowledge, a first attempt of coupling a Paul trap to synchrotron radiation was reported some years ago (KravisIn the present communication we report on a novel experimental set-up, coupling a linear quadrupole ion trap equipped with an ESI source to a third-generation synchrotron undulator-based VUV beamline. Our set-up shows the possibility of using synchrotron-radiation-induced activation as a new complementary MS2 technique as well as it providing an efficient VUV spectroscopy technique for electrosprayed biopolymer ions isolated in the gas phase. The system involves several important advantages and improvements leading to an excellent signal-to-noise ratio for an ion spectroscopy and activation set-up.
2. Experimental set-up
A schematic drawing of the experimental assembly is given in Fig. 1. It is based upon a commercial linear quadrupole (LTQ XL; Thermo Fisher), equipped with an ESI source. The electrosprayed ions are introduced from the front side of the spectrometer and guided throughout the system into a linear quadrupole ion trap (Schwartz et al., 2002). is performed by ejecting ions out of a slot in two opposite rods using the mass selective instability mode of operation. The synchrotron beam is sent directly into the trap through the back lens of the quadrupole ion trap from the back of the spectrometer. To achieve this, the original back plate of the spectrometer has been replaced by a home-made plate fitted with a flange to connect with the differential pumping stage (turbomolecular pump of 300 l s−1) that accommodates the pressure difference between the spectrometer (10−5 mbar) and the beamline port (10−8 mbar). The pumping stage includes a retractable calibrated photodiode (AXUV100; International Radiation Detectors) for measurements and a photon beam shutter (Milosavljević et al., 2011a). The latter has been especially designed to provide a short rise time (≃ 1 ms) and reproducible chopping operation under high-vacuum conditions.
The main difficulties in photo-excitation of ions are: (i) achieving a high enough ion density for a sufficient long period of time, (ii) sufficient overlap between the ion beam and the synchrotron radiation beam, (iii) sufficient ; Hirsch et al., 2009): S = FIJLσ, where S is the measured ionic signal per second, F is the form factor describing the overlap between the two beams, I is the (photons s−1), J is the ion density (ions cm−3), L is the interaction path length (cm) and σ is the of the process. The ionic trapping cylindrical volume of length 30 mm and diameter 2 mm is formed by four hyperbolic rods. The LTQ XL may confine up to 3 × 104 ions within that trapping volume (106 ions cm−3) without space charge. Hence, for irradiation times of 1 s, processes with cross sections as low as a tenth of a Mbarn may be observed with photon fluxes of 1012 photons s−1. Hence, a high and optimal overlap between the incident photons and trapped ion packet is crucial to achieve a good signal-to-noise ratio. Finally, tunability over a wide wavelength range in the VUV is mandatory for obtaining selectivity in the fragmentation processes. These constraints on the photon source are only met at third-generation synchrotron radiation sources, especially on undulator-based beamlines.
(iv) a long enough interaction path length and (v) enough for ions. The product ion count rate may be estimated by the following formula (West, 2001The experimental set-up has been installed as a non-permanent endstation on the VUV beamline DESIRS (https://www.synchrotron-soleil.fr/portal/page/portal/Recherche/LignesLumiere/DESIRS ; Nahon et al., 2012) at the SOLEIL storage ring in St Aubin (France). The electromagnetic undulator OPHELIE2 (HU640 type) provides tunable radiation in the 5–40 eV energy range (Marcouille et al., 2007). The wavelength is selected by a 6.65 m-long normal-incidence monochromator (Nahon et al., 2001) equipped with a low-dispersion (200 grooves mm−1) grating providing a high of the order of 1012–1013 photons s−1 over the energy range, within a photon bandwidth of typically 12 meV at 10 eV photon energy with typically a 200 µm exit slit. Higher harmonics of the undulator are cut off by a gas filter filled with Ar or Xe (Mercier et al., 2000) or at low energies by MgF2 or SUPRASIL windows.
The experimental procedure consists of the following sequence of events, for a given wavelength: (i) electrosprayed ions are injected, selected and stored in the trap; (ii) when the desired ion capacity is reached, the beam shutter opens, thus starting the irradiation; (iii) after the desired time of irradiation, the shutter intercepts the beam; (iv) the
is recorded; (v) the monochromator and the position of the undulator maximum are set to the next wavelength, in the so-called gap-scan mode, and the procedure is repeated. The irradiation time is regulated directly by the LTQ XL that masters opening/closing of the photon shutter. The automatic energy scanning and mass spectra recording has been achieved by use of a separate personal computer and a home-made program, which synchronizes the operation of both the beamline and the mass spectrometer.3. Results
A model protein, the cytochrome c (104 residues, 12384 Da), was submitted to the experimental procedure previously described. Fig. 2 presents a typical MS2 spectrum of the 4+ precursor of cytochrome c (m/z = 3090), recorded after about 600 ms of irradiation at a photon energy of 14.25 eV. This spectrum reveals a very efficient ionization of polyprotonated cytochrome c protein upon absorption of a VUV photon. Moreover, following the dominant ionized 5+ species at m/z = 2477.6, a series of features corresponding to 6+, 7+ and 8+ ionized precursor can be observed in the spectrum. The multiply ionized molecules are most probably formed by a successive single as their intensity linearly decreases with decreasing irradiation time. A more detailed discussion on the efficiency curves and fragmentation patterns exceeds the scope of the present communication and is partly the subject of a separate publication (Milosavljević et al., 2011b). Nevertheless, it is important to point out that VUV synchrotron radiation appears to be a very efficient activation method for the trapped protein precursor ions. Note that the MS2 spectrum recorded under the same conditions without the photon beam does not exhibit, as expected, any additional features to the precursor peak (inset of Fig. 2). Furthermore, the signal-to-noise ratio is high enough to resolve all features in the even in the monochromated undulator radiation, while a noticeable noise appears only at low m/z values. It should also be noted that the noise further decreases with decreasing irradiation time and photon energy.
Fig. 3 shows a MS2 spectrum of the negative 6− precursor of cytochrome c molecule (m/z = 2063), recorded after about 200 ms of irradiation at a photon energy of 15.0 eV. The dominant channel at m/z = 2475.6 corresponds to the formation of the radical species , which is formed by electron photodetachment from the Additionally, the feature corresponding to double photodetachment (m/z = 3083) is clearly resolved in the spectrum. It should be noted that from trapped peptide and protein polyanions in vacuo by ultraviolet/visible (UV/Vis) laser irradiation has been recently studied (Joly et al., 2007; Brunet et al., 2011). The proposed mechanism involves an electronic excitation of the chromosphere state leading to the via a crossing with an auto-ionizing state of the molecule. As a consequence, photodetachment spectroscopy can be used to probe the electronic structure of a complex biomolecule isolated in the gas phase. The first VUV photodetachment study (above about 4 eV of photon energy) of electrosprayed peptide ions, and only conducted with non-dispersed undulator radiation light, has been preliminary reported recently (Giulliani, 2009). The present results demonstrate that photodetachment spectroscopy of a 12.4 kDa cytochrome c protein precursor can be efficiently performed using a VUV synchrotron radiation source.
4. Conclusion
The present paper reports on a novel experimental system that uses VUV synchrotron radiation as an activation method for quadrupole ion trap tandem
The achievements of the present system are the first photodetachment spectroscopy of electrosprayed biopolymers using monochromated synchrotron radiation, as well as the first ionization spectroscopy study of large polycations. More than satisfactory performances and sensitivity of the present experimental set-up provide the possibility to study fundamental physical and structural properties of large biological ions isolated in the gas phase by use of synchrotron-radiation-based VUV spectroscopy. Furthermore, the efficient formation of radical species upon of trapped ions paves the way for the development of a new complementary tandem technique, which uses VUV synchrotron radiation as a controllable soft-ionization activation method. Besides existing and well established methods such as macromolecular crystallography, small- and wide-angle X-ray scattering or synchrotron radiation our technique appears then as a novel and complementary method for studying the structure and physical properties of biopolymers using synchrotron radiation.Acknowledgements
This work is supported by the Agence Nationale de la Recherche Scientifique, France, under the project #BLAN08-1_348053. We gratefully acknowledge the help of Jocelyn Dupuis from Thermo Scientific, France, and August Specht from Thermo Scientific, USA. ARM acknowledges support for a short-term scientific mission to SOLEIL from the COST Action CM0601, Electron Controlled Chemical Lithography (ECCL), as well as partial support by the Ministry of Education and Science of Republic of Serbia (Project No. 171020). We are grateful to the general SOLEIL staff for running the overall facility, and support under projects #20090295 and #20080023.
References
Bari, S., Gonzalez-Magaña, O., Reitsma, G., Werner, J., Schippers, S., Hoekstra, R. & Schlathölter, T. (2011). J. Chem. Phys. 134, 024314. Web of Science CrossRef PubMed Google Scholar
Biemann, K. (2002). J. Am. Soc. Mass Spectrom. 13, 1254–1272. Web of Science CrossRef PubMed CAS Google Scholar
Brunet, C., Antoine, R., Allouche, A. R., Dugourd, P., Canon, F., Giuliani, A. & Nahon, L. (2011). J. Phys. Chem. A, 115, 8933–8939. Web of Science CrossRef CAS PubMed Google Scholar
Dass, C. (2001). Principles and Practice of Biological Mass Spectrometry. New York: John Wiley and Sons. Google Scholar
Fung, Y. M., Adams, C. M. & Zubarev, R. A. (2009). J. Am. Chem. Soc. 131, 9977–9985. Web of Science CrossRef PubMed CAS Google Scholar
Giulliani, A. et al. (2009). Proceedings of the 57th ASMS Conference on Mass Spectrometry and Allied Topics, Philadelphia, PA, USA, 31 May–5 June 2009. Google Scholar
Haddon, W. F. & McLafferty, F. W. (1968). J. Am. Chem. Soc. 90, 4745–4746. CrossRef CAS Web of Science Google Scholar
Hirsch, K., Lau, J., Klar, P., Langenberg, A., Probst, J., Rittmann, J., Vogel, M., Zamudio-Bayer, V., Möller, T. & Issendorff, B. (2009). J. Phys. B, 42, 154029. Web of Science CrossRef Google Scholar
Hoffmann, E. de (1996). J. Mass Spectrom. 31, 129–137. Google Scholar
Hoffmann, E. de (2001). Mass Spectrometry. Principle and Applications. New York: Wiley. Google Scholar
Joly, L., Antoine, R., Allouche, A. R., Broyer, M., Lemoine, J. & Dugourd, P. (2007). J. Am. Chem. Soc. 129, 8428–8429. Web of Science CrossRef PubMed CAS Google Scholar
Kalcic, C. L., Gunaratne, T. C., Jones, A. D., Dantus, M. & Reid, G. E. (2009). J. Am. Chem. Soc. 131, 940–942. Web of Science CrossRef PubMed CAS Google Scholar
Kravis, S. D., Church, D. A., Johnson, B. M., Meron, M., Jones, K. W., Levin, J., Sellin, I. A., Azuma, Y., Mansour, N. B., Berry, H. G. & Druetta, M. (1991). Phys. Rev. Lett. 66, 2956–2959. CrossRef PubMed CAS Web of Science Google Scholar
Little, D. P., Speir, J. P., Senko, M. W., O'Connor, P. B. & McLafferty, F. W. (1994). Anal. Chem. 66, 2809–2815. CrossRef CAS PubMed Web of Science Google Scholar
Marcouille, O., Brunelle, P., Chubar, O., Marteau, F., Massal, M., Nahon, L., Tavakoli, K., Veteran, J. & Filhol, J. M. (2007). AIP Conf. Proc. 879, 311–314. CrossRef CAS Google Scholar
Mercier, B., Compin, M., Prevost, C., Bellec, G., Thissen, R., Dutuit, O. & Nahon, L. (2000). J. Vac. Sci. Technol. A, 18, 2533–2541. Web of Science CrossRef CAS Google Scholar
Milosavlejić, A. R., Nicolas, C., Gil, J.-F., Canon, F., RéFrégiers, M., Nahon, L. & Giuliani, A. (2011a). Nucl. Instrum. Methods Phys. Res. B, doi:10.1016/j.nimb.2011.10.032. Google Scholar
Milosavlejić, A. R., Nicolas, C., Lemaire, J., Déhon, C., Thissen, R., Bizau, J.-M., RéFrégiers, M., Nahon, L. & Giuliani, A. (2011b). Phys. Chem. Chem. Phys. 13, 15432–15436. Web of Science PubMed Google Scholar
Nahon, L., Alcaraz, C., Marlats, J. L., Lagarde, B., Polack, F., Thissen, R., Lepere, D. & Ito, K. (2001). Rev. Sci. Instrum. 72, 1320–1329. Web of Science CrossRef CAS Google Scholar
Nahon, L., de Oliveira, N., Garcia, G. A., Gil, J.-F., Pilette, B., Marcouillé, O., Lagarde, B. & Polack, F. (2012). J Synchrotron Rad. In preparation. Google Scholar
Price, W. D., Schnier, P. D. & Williams, E. R. (1996). Anal. Chem. 68, 859–866. CrossRef CAS PubMed Web of Science Google Scholar
Rozett, R. W. & Koski, W. S. (1968). J. Chem. Phys. 49, 2691–2695. CrossRef CAS Web of Science Google Scholar
Schwartz, J. C., Senko, M. W. & Syka, J. E. (2002). J. Am. Soc. Mass Spectrom. 13, 659–669. Web of Science CrossRef PubMed CAS Google Scholar
Thissen, R., Bizau, J. M., Blancard, C., Coreno, M., Dehon, C., Franceschi, P., Giuliani, A., Lemaire, J. & Nicolas, C. (2008). Phys. Rev. Lett. 100, 223001. Web of Science CrossRef PubMed Google Scholar
West, J. B. (2001). J. Phys. B, 34, R45. CrossRef Google Scholar
Zubarev, R. A., Kelleher, N. L. & McLafferty, F. W. (1998). J. Am. Chem. Soc. 120, 3265–3266. Web of Science CrossRef CAS Google Scholar
© International Union of Crystallography. Prior permission is not required to reproduce short quotations, tables and figures from this article, provided the original authors and source are cited. For more information, click here.