In vivo physiological saline-infused hepatic vessel imaging using a two-crystal-interferometer-based phase-contrast X-ray technique
aAllied Health Sciences and Graduate School of Medical Science, Kitasato University, 1-15-1 Kitasato, Minami-ku, Sagamihara, Kanagawa 252-0373, Japan, bInstitute of Clinical Medicine, University of Tsukuba, Tsukuba-shi, Ibaraki 305-8575, Japan, cAdvanced Research Laboratory, Hitachi Ltd, Hatoyama, Saitama 350-0395, Japan, dDivision of Clinical Neuroscience, Chiba University Center for Forensic Mental Health, 1-8-1 Inohana, Chiba 260-8670, Japan, eDepartment of Advanced Materials Science, Graduate School of Frontier Sciences, The University of Tokyo, 5-1-5 Kashiwanoha, Kashiwa-shi, Chiba 277-8561, Japan, and fHigh Energy Accelerator Research Organization, Tsukuba-shi, Ibaraki 305-0801, Japan
*Correspondence e-mail: t.takeda@kitasato-u.ac.jp
Using a two-crystal-interferometer-based phase-contrast X-ray imaging system, the portal vein, capillary vessel area and hepatic vein of live rats were revealed sequentially by injecting physiological saline via the portal vein. Vessels greater than 0.06 mm in diameter were clearly shown with low levels of X-rays (552 µGy). This suggests that in vivo vessel imaging of small animals can be performed as conventional angiography without the side effects of the presently used iodine contrast agents.
Keywords: in vivo hepatic vessel imaging; two-crystal interferometer; phase-contrast X-ray technique; physiological saline.
1. Introduction
Vessel imaging plays a vital role in the diagnosis of stenosis, obstruction and collateral formation in coronary arteries and cerebral arteries, vessel structures of various organs and neovascularity in tumors. Conventional X-ray angiography using the absorption-contrast phenomenon is widely used to demonstrate vessel structures. Since the difference in the X-ray absorption of the blood against surrounding tissues is very small, the use of an iodine contrast agent composed of triiodobenzoic acid derivatives is indispensable in vessel imaging. However, an iodine contrast agent with various demerits, such as high viscosity, high et al., 1990).
and allergenicity, causes adverse reactions ranging from minor to severe, sometimes resulting in death (KatayamaIn contrast, the phase-contrast X-ray technique that detects minute changes of density has about 1000-fold greater sensitivity than absorption for low-atomic-number elements such as hydrogen, carbon, nitrogen and oxygen (Momose & Fukuda, 1995; Momose, 1995; Takeda et al., 1995). Several phase-contrast X-ray imaging techniques (Fitzgerald, 2000; Momose, 2005) such as the Bonse–Hart-type crystal X-ray interferometer-based technique (Bonse & Hart, 1965), the refraction-based technique using crystal diffraction (diffraction-enhanced imaging) (Davis et al., 1995; Ingal & Beliaevskaya, 1995; Chapman et al., 1997; Arfelli et al., 2000; Pisano et al., 2000), the propagation-based technique (Snigirev et al., 1995; Wilkins et al., 1996; Yagi et al., 1999; Hwu et al., 1999) and the grating (Talbot) interferometer technique (Momose et al., 2003; Weitkamp et al., 2005; Pfeiffer et al., 2006) have been developed. Among these methods an interferometer-based method using a crystal X-ray interferometer is the most sensitive method for detecting minute differences of density (Fitzgerald, 2000; Yoneyama et al., 2008). Detailed inner structures of soft tissue that could not be revealed by the absorption-contrast X-ray technique were clearly depicted in phase maps of rat cerebellum (Momose & Fukuda, 1995) and human metastatic liver tumor (Takeda et al., 1995), and phase-contrast X-ray CT images of rabbit cancer (Momose et al., 1996), rat brain (Bonse & Busch, 1996; Beckmann et al., 1997) and human cancer (Takeda et al., 2000) using a monolithic X-ray interferometer.
In phase-contrast X-ray imaging, images of vessels can be generated to detect the minute density difference against the surrounding vessel wall, so the image contrast is obtained by the use of only physiological saline. Our previous study demonstrated that ex vivo phase-contrast X-ray imaging with a monolithic interferometer enabled clear depiction of the hepatic vessels using physiological saline as a contrast agent (Takeda et al., 2002). In the monolithic X-ray interferometer-based imaging system, the distance between the object and the half-mirror was less than 2 cm owing to the limited geometrical structure of the monolithic X-ray interferometer (Takeda & Itai, 2001; Takeda et al., 2009). As such, the heat effects from the live object disturbed the recording of phase-contrast X-ray images because the object was set near the half-mirror during the data acquisition, and the half-mirror of the interferometer could not be stabilized at the ångström order by heat.
A two-crystal X-ray interferometer-based imaging system enables reduction of the heat effects of a living object by extending the geometrical distance between the object and the half-mirror to more than 15 cm. Here, using this phase-contrast X-ray imaging system, in vivo hepatic vessel imaging is reported for the first time using physiological saline as a contrast agent.
2. Methods
2.1. Phase-contrast X-ray imaging system
The phase-contrast X-ray imaging system (Yoneyama et al., 2004) consisted of a Si (220) asymmetric-cut crystal for forming a two-dimensional beam, a two-crystal X-ray interferometer, a phase shifter, an object cell and an X-ray CCD camera (Fig. 1). The two-crystal X-ray interferometer was made using a commercially available 10 cm-diameter highly perfect single-crystal silicon ingot. The pixel size of the X-ray CCD camera was 0.018 mm × 0.018 mm with 2048 × 2048 pixels, and a of 14 bits. The experiment was performed at the Photon Factory synchrotron X-ray source in Tsukuba, Japan. The X-ray energy was set at 17.7 keV and the X-ray in front of the sample cell was 184 µGy s−1 at a beam current of 430 mA with 2.5 GeV energy.
2.2. Phase map
The spatial distribution of the phase shift caused by a sample was depicted using the Fourier technique because this technique (Takeda et al., 1982) can acquire the image data five-fold faster than the conventionally used interferometric technique (Momose & Fukuda, 1995). A phase map of the object was acquired by a 3 s exposure. Data transfer from the CCD camera to a computer required 1.8 s, so the time interval of each image became 4.8 s. The rat liver was imaged in a cell filled with physiological saline to reduce the variation of the object thickness. The thickness of the cell was 12 mm and the total X-ray exposure was 552 µGy frame−1.
2.3. Selection of contrast agent
Since, in phase-contrast X-ray imaging, low-atomic-number elements cause a sufficient phase shift, contrast agents not containing heavy elements such as iodine were investigated. Physiological saline had the highest contrast enhancement, and enabled clear imaging of the vessels in the excised liver of rats, so it was selected for use as the contrast agent of the vessels (Takeda et al., 2002).
2.4. Animal preparation
Five Wistar rats were anesthetized using pentobarbital. A 25-gauge catheter was surgically inserted into the portal vein and was ligated by silk thread. Then, a part of the hepatic lobes is bent and protruded from the abdomen in order to take a projection image of the liver. 2 ml of physiological saline was injected using an injector at a rate of 0.1 ml s−1 (M-800C, Nemoto-Kyorindo, Tokyo, Japan). The present experiment was approved by the Medical Committee for the Use of Animals in Research of the University of Tsukuba, and conformed to the guidelines of the American Physiological Society.
3. Results
Hepatic vessels of rat injected with physiological saline via the portal vein clearly showed up as dark grey areas in the phase map (Fig. 2), since physiological saline has a low compared with normal vessel walls and liver tissue. The phase map showed the portal vein, capillary vessel area and hepatic vein sequentially. A hepatic vessel of about 0.06 mm in diameter was clearly shown.
One rat showed obstruction of the distal portal branch, and a perfusion defect was revealed in the region of the obstructed vessel at the capillary phase (Fig. 3). In another rat, a catheter was inserted into a deep portion of the portal vein, and selective angiography of the portal vein, named super-selective catheterization, was carried out. The super-selected portal branch was revealed on the right-hand side of the hepatic lobe; however, the left-hand side of the hepatic lobe was not shown at all owing to the lack of injection of physiological saline (Fig. 4).
In addition, the time density curve of the hepatic vessel shown in Fig. 2 indicated that the peak times for the portal vein, capillary vessel area and hepatic vein were about 4.8 s, 19.2 s and 28.8 s, respectively, and enabled an estimation of the flow speed (Fig. 5).
4. Discussion
In vivo hepatic vessels of rats could be clearly visualized by the two-crystal phase-contrast X-ray imaging system using physiological saline as a contrast agent. Owing to the lower specific gravity of physiological saline, the high difference in between physiological saline and the surrounding soft tissues caused a high contrast in saline-infused vessels. Flow dynamics from the portal vein phase, capillary phase and venous phase were shown sequentially. A hepatic vessel of about 0.06 mm in diameter was shown well. Obstruction of the distal portal branch and a related perfusion defect were also visualized. In addition, the super-selective angiographic procedure also allowed selected vessel imaging as with typical angiography using a conventional iodine contrast agent. Furthermore, quantitative analysis enabled measurement of the density changes of the portal vein, capillary vessel area and hepatic vein. Thus, the phase-contrast X-ray imaging technique allows visualization of the in vivo vessel as with conventional angiography, but using only physiological saline.
To resolve the small difference in the X-ray absorption of the blood against that of the vessel wall and surrounding tissue, the vessel structure must be imaged by filling the vessel with a contrast agent containing a heavy element such as iodine. However, in such angiographic techniques, unpleasant side effects occur when using iodine contrast agent, such as those owing to high et al., 1990). Physiological saline composed of sodium chloride without iodine has excellent merits for vessel imaging because the and viscosity are reduced by one-half to one-third and one-thirteenth to one-quarter, respectively, compared with iodine contrast agent.
high viscosity and iodine allergenicity (KatayamaIn this phase-contrast technique a vessel of about 0.06 mm in diameter was visualized using an X-ray dose of 552 µGy frame−1. However, in an absorption-contrast method using the iodine contrast agent, the X-ray dose is estimated to be approximately 41-fold greater for depiction of the same vessel diameter. A small amount of radiation exposure to the object will reduce the risk of harm and has excellent benefits for examination.
Thus, the in vivo phase-contrast X-ray imaging technique enables the imaging of vessels without the use of iodine contrast agent and with low X-ray dose, and might be applied to assess various vascular diseases, neovascularity of tumors and angiogenesis in the near future.
4.1. Present limitations
In this experiment the time interval for each image must be set to 4.8 s owing to the low sensitivity of the lens-coupled X-ray CCD camera and the long transfer time of the analog-to-digital converter (ADC). Therefore, the data acquisition speed is insufficient for imaging of the vessel and analysis of the detailed flow dynamics. In addition, the visualized vessel diameter was limited to about 0.06 mm. If a fibre-coupled-type X-ray CCD camera and high-speed ADC could be used, the light in front of the CCD camera could be increased by about five-fold, and the data transfer time could be shortened to less than 0.2 s; then one image would be obtained in less than 1 s, and clear in vivo images with less blurring would be acquired. Furthermore, an increment of the X-ray could enable a shortening of the image acquisition time and the blood flow might be revealed at a rate sufficient for video-imaging.
The present system with small interferometer size has been limited to use in clinical applications, and is applicable for small-animal experiments. However, further developments to widen the view size of the interferometer system may allow clinical applications in the future.
Acknowledgements
This research was supported by Special Coordination Funds from the Ministry of Education, Culture, Sports, Science and Technology, and was performed under proposal numbers 2002S2-001 and 2005S2-001, approved by the High Energy Accelerator Research Organization. We are grateful to Mr Kouzou Kobayashi for preparation of the experimental apparatus.
References
Arfelli, F. et al. (2000). Radiology, 215, 286–293. Web of Science CrossRef PubMed CAS Google Scholar
Beckmann, F., Bonse, U., Busch, F. & Günnewig, O. (1997). J. Comput. Assist. Tomogr. 21, 539–553. CrossRef CAS PubMed Web of Science Google Scholar
Bonse, U. & Busch, F. (1996). Prog. Biophys. Mol. Biol. 65, 133–169. CrossRef CAS PubMed Web of Science Google Scholar
Bonse, U. & Hart, M. (1965). Appl. Phys. Lett. 6, 155–156. CrossRef Web of Science Google Scholar
Chapman, D., Thomlinson, W., Johnston, R. E., Washburn, D., Pisano, E., Gmür, N., Zhong, Z., Menk, R., Arfelli, F. & Sayers, D. (1997). Phys. Med. Biol. 42, 2015–2025. CrossRef CAS PubMed Web of Science Google Scholar
Davis, T. J., Gao, D., Gureyev, T. E., Stevenson, A. W. & Wilkins, S. W. (1995). Nature (London), 373, 595–598. CrossRef CAS Web of Science Google Scholar
Fitzgerald, R. (2000). Phys. Today, 53, 23–28. Web of Science CrossRef Google Scholar
Hwu, Y., Hsieh, H., Lu, M. J., Tsai, W. L., Lin, H. M., Goh, W. C., Lai, B., Je, J. H., Kim, C. K., Noh, D. Y., Youn, H. S., Tromba, G. & Margaritondo, G. (1999). J. Appl. Phys. 86, 4613–4618. Web of Science CrossRef CAS Google Scholar
Ingal, V. N. & Beliaevskaya, E. A. (1995). J. Phys. D, 28, 2314–2317. CrossRef CAS Web of Science Google Scholar
Katayama, H., Yamaguchi, K., Kozuka, T., Takashima, T., Seez, P. & Matsuura, K. (1990). Radiology, 175, 621–628. CrossRef CAS PubMed Web of Science Google Scholar
Momose, A. (1995). Nucl. Instrum. Methods Phys. Res. A, 352, 622–628. CrossRef CAS Web of Science Google Scholar
Momose, A. (2005). Jpn. J. Appl. Phys. 44, 6355–6367. Web of Science CrossRef CAS Google Scholar
Momose, A. & Fukuda, J. (1995). Med. Phys. 22, 375–379. CrossRef CAS PubMed Web of Science Google Scholar
Momose, A., Kawamoto, S., Koyama, I., Hamaishi, Y., Takai, K. & Suzuki, Y. (2003). Jpn. J. Appl. Phys. 42, L866–L868. Web of Science CrossRef CAS Google Scholar
Momose, A., Takeda, T., Itai, Y. & Hirano, K. (1996). Nat. Med. 2, 473–475. CrossRef CAS PubMed Web of Science Google Scholar
Pfeiffer, F., Weitkamp, T., Bunk, O. & David, C. (2006). Nat. Phys. 2, 258–261. Web of Science CrossRef CAS Google Scholar
Pisano, E. D., Johnston, R. E., Chapman, D., Geradts, J., Iacocca, M. V., Livasy, C. A., Washburn, D. B., Sayers, D. E., Zhong, Z., Kiss, M. Z. & Thomlinson, W. C. (2000). Radiology, 214, 895–901. Web of Science CrossRef PubMed CAS Google Scholar
Snigirev, A., Snigireva, I., Kohn, V., Kuznetsov, S. & Schelokov, I. (1995). Rev. Sci. Instrum. 66, 5486–5492. CrossRef CAS Web of Science Google Scholar
Takeda, M., Ina, H. & Kobayashi, S. (1982). J. Opt. Soc. Am. 72, 156–160. CrossRef Web of Science Google Scholar
Takeda, T. & Itai, Y. (2001). Progress in Pathology, edited by N. Kirkham and N. R. Lemoine, pp. 81–102. London: Greenwich Medical Media. Google Scholar
Takeda, T., Momose, A., Hirano, K., Haraoka, S., Watanabe, T. & Itai, Y. (2000). Radiology, 214, 298–301. CrossRef PubMed CAS Google Scholar
Takeda, T., Momose, A., Itai, Y., Wu, J. & Hirano, K. (1995). Acad. Radiol. 2, 799–803. CrossRef CAS PubMed Web of Science Google Scholar
Takeda, T., Momose, A., Wu, J., Yu, Q., Zeniya, T., Thet-Thet-Lwin, Yoneyama, A. & Itai, Y. (2002). Circulation, 105, 1708–1712. Web of Science CrossRef PubMed Google Scholar
Takeda, T., Yoneyama, A., Takeya, S., Thet-Thet-Lwin & Wu, J. (2009). Handbook of Interferometers, edited by D. Halsey and W. Raynor, pp. 288–316. New York: Nova Science. Google Scholar
Weitkamp, T., Diaz, A., David, C., Pfeiffer, F., Stampanoni, M., Cloetens, P. & Ziegler, E. (2005). Opt. Express, 13, 6296–6304. Web of Science CrossRef PubMed Google Scholar
Wilkins, S. W., Gureyev, T. E., Gao, D., Pogany, A. & Stevenson, A. W. (1996). Nature (London), 384, 335–338. CrossRef CAS Web of Science Google Scholar
Yagi, N., Suzuki, Y., Umetani, K., Kohmura, Y. & Yamasaki, K. (1999). Med. Phys. 26, 2190–2193. Web of Science CrossRef PubMed CAS Google Scholar
Yoneyama, A., Takeda, T., Tsuchiya, Y., Wu, J., Thet-Thet-Lwin, Koizumi, A., Hyodo, K. & Itai, Y. (2004). Nucl. Instrum. Methods Phys. Res. A, 523, 217–222. Web of Science CrossRef CAS Google Scholar
Yoneyama, A., Wu, J., Hyodo, K. & Takeda, T. (2008). Med. Phys. 35, 4724–4734. Web of Science PubMed Google Scholar
© International Union of Crystallography. Prior permission is not required to reproduce short quotations, tables and figures from this article, provided the original authors and source are cited. For more information, click here.