research papers
A variable ultra-short-pathlength solution cell for XAFS transmission spectroscopy of light elements
aChemical and Materials Sciences Division, Pacific Northwest National Laboratory, Richland, WA 99354, USA, bX-ray Science Division, Argonne National Laboratory, Argonne, IL 60439, USA, and cEnergy Processes and Materials Division, Pacific Northwest National Laboratory, Richland, WA 99354, USA
*Correspondence e-mail: john.fulton@pnnl.gov
An X-ray absorption fine-structure spectroscopy (XAFS) cell that is suitable for solution-phase studies of the light elements in the series from Na+ and Ca2+ is described. This cell has an ultra-short pathlength that can be remotely adjusted using a miniature stepper-motor drive and thereby readily provides transmission pathlengths in the range from submicrometer to several hundred micrometers. The flexibility to vary the pathlength enables acquisition of high-quality spectra and also allows one to check for potential distortions in the spectra from thickness effects. The primary components are mostly commercially available optical parts. The performance of this device is demonstrated at the Cl K-edge (2.8 keV) for several different aqueous Cl− solutions.
Keywords: X-ray absorption fine structure; XAFS; low-Z cell; short pathlength.
1. Introduction
The light elements in the third-row and early fourth-row series (including Na, Mg, Al, Si, P, S, Cl, K and Ca) contain six of the most prevalent elements in the earth's crust. In their various chemical states they play central roles in a wide variety of geochemical and biochemical processes. Some of these elements (e.g. P and S) also play important roles in catalysis and energy storage. The chemical state of these elements can be probed using the method of X-ray absorption fine structure (XAFS). provides detailed molecular structure information on the coordination state including bond distances, bond disorder, coordination symmetry, and the chemical identities of coordination species and even in disordered systems. In earlier studies of these light elements, spectroscopy has been used to determine the hydration structure about aqueous ions (Dang et al., 2006) (Cl−, K+ and Ca2+) both under ambient conditions and for temperatures up to approximately 700 K in water (Fulton et al., 2004, 2006). More recently the structure between Cl− and hydronium (H3O+) has been elucidated (Fulton & Balasubramanian, 2010). Further, studies have provided important information on protonation of aqueous phosphate species (Rouff et al., 2009) and for ion chemical states in rechargeable lithium/sulfur batteries (Gao et al., 2011). However, many of these studies were limited by the difficulty of preparing samples with suitably short pathlengths. In this report we describe an adjustable short-pathlength transmission cell that is specifically designed to address liquid-phase solutions for this series of light elements. The K-edge absorption energies for these elements span the range from 1 keV to 4.5 keV. This energy range also coincides with the L-edge energies for the series of elements above Zn including almost all of the second-row transition metals up to Cd that are of primary interest for catalysis. This energy range also covers the M4,5-edges for the lanthanides and actinides. Thus this ultra-short-pathlength cell is also useful for L-edge studies of the second-row transition metals and M-edge studies of the actinides and lanthanides.
The cell is optimized for transmission measurements of solutes at moderate to high concentrations. In this case transmission measurements provide superior signal-to-noise ratios since the fluorescence yields of the light elements are rather low. However, the challenge is that the X-ray absorbance lengths at these low energies (1 keV to 4.5 keV) are rather short and therefore both the windows and the sample pathlengths need to be short. In order to maximize the transmission of the X-ray beam through the sample, the use of thin membrane windows (100–500 nm thick) having relatively little beam attenuation is preferred. Thin Si3N4 membrane windows were first used by Yang for an transmission study of water (Yang & Kirz, 1987). Recently, Wernet et al. (Schreck et al., 2011) and Nagasaka et al. (2010) have reported a significant improvement on the concept for very low energies (300–1000 eV) including the study of the O K-edge of liquid water. Huthwelker et al. (2010) developed a cell for O K-edge studies of individual aerosol particles under controlled atmospheres. Variants of these cells have been described for both transmission (Näslund et al., 2005) and fluorescence measurements (Aziz, 2010; Aziz et al., 2006; Guo & Luo, 2010; Edwards & Myneni, 2006; Freiwald et al., 2004). In this report we describe a cell that is optimized for both X-ray absorption near-edge spectroscopy (XANES) and extended X-ray absorption fine-structure spectroscopy (EXAFS) at somewhat higher energies. This cell is optimal for those elements having edge energies in the range from 1 keV to 4.5 keV. These require relatively longer pathlengths when compared with cells used for soft X-rays (<1 keV). This cell also uses membrane windows; however, one of the windows is connected to a stepper-motor drive. Using this approach a broad range of pathlengths (1–500 µm) can be selected by remotely adjusting the stepper-motor position. The resolution of the stepper-motor drive is 50 nm, thereby allowing achievement of pathlengths well below 1 µm. In this design the parts having wetted surfaces are machined from a chemically resistant polymeric material whereas the primary mechanical components are mostly commercially available. This allows for more ready implementation of the method at a multitude of beamlines.
Finally, the cell is suitable for transmission measurements for a variety of different aqueous, organic, ionic liquid and nanoparticle-containing phases. Although designed specifically for transmission measurement there are instances where the device could be potentially beneficial for fluorescence measurements at 45°, since the adjustable pathlength feature of the device would allow for optimizing the effective sample thickness to minimize self-absorption effects from concentrated samples.
2. Cell design
Schematics of the primary components of the cell are shown in Fig. 1. Images of the entire cell are shown in Fig. 2. The sample liquid is contained between two polyether ether ketone (PEEK) discs. PEEK provides very high chemical resistance for a broad range of compounds. One of the 3.8 cm-diameter discs is essentially rigid and contains the connections for introducing the sample through 1.6 mm outer-diameter PEEK tubing (Upchurch Scientific) into the cell. The other disc is flat, thin and flexible and is machined to a relatively high thickness uniformity (±0.025 mm). These two discs are sealed together using an elastomeric O-ring to create the disc-shaped liquid-filled cavity between the two PEEK plates. The thin PEEK disc is deflected by applying a uniform circular load near the center of the disc in the region where the X-ray beam passes through the cell. The load is applied using a miniature stepper-motor translation stage (Micos, VT-21 S) having 50 nm resolution. In this way the X-ray pathlength through the liquid can be remotely adjusted in the range from approximately 1 to 500 µm. The deflection of this plate under load can be calculated using an analytical formula for a flat circular plate having a constant thickness, with the outer edge fixed, and with a line load on the inner circumference (Roark & Young, 1982). Fig. 3 shows typical deflections for a 3.8 cm-diameter PEEK disc of various thicknesses.
Portions of the device use standard optical components routinely found at synchrotron beamlines. The two PEEK discs are held together using a standard lens mount (Thorlabs, LMR1.5). The actuator shown in Fig. 1(a) is driven by the stepper-motor drive (Micos, VT-21 S). To be sure that the actuator applies the load uniformly, it is attached to a gimbal (Thorlabs, #GMB1) that is mounted directly onto the miniature stepper-motor drive. The gimbal mount assures that the circular load is evenly distributed and that the load remains perpendicular to the surface of the deflection disc. The X-ray windows are two silicon nitride (Si3N4) windows (Norcada) that are affixed to the inner surfaces of the PEEK discs. These 2 mm × 2 mm by 200 nm-thick Si3N4 membranes are part of a larger 7.5 mm × 7.5 mm by 500 µm-thick Si frame (Norcada). These frames are then attached to the PEEK discs using a silicone-resin-based high-vacuum adhesive (Vacseal, SPI Supplies). The two 200 nm-thick Si3N4 membranes provide >75% transmission in the range from 1 to 4.5 keV. The maximum pressure differential across these membrane windows is approximately 1 bar. Before filling the sample cell with liquid, the stepper motor is fully retracted to open the gap between the Si3N4 windows. This allows the solute solution to fully replenish the region between the windows. The gap is then reduced for the measurement. The maximum velocity of the translator is limited to allow the fluid between the windows to be expelled, without build up of pressure, as the gap is being reduced. The configuration shown in Fig. 2 is intended for use under ambient temperatures although the thicker of the two sample containment discs (`cell body' in Fig. 1) could be constructed from a metal alloy and thereby electrically heated to provide temperature control up to approximately 373 K.
Even though the Si3N4 windows are quite thin they can introduce artifacts from the Si edge (1839 eV) and the Si structure (from the Si3N4) into the sample spectra in this energy region. An alternate window material for this spectral region is 800 nm-thick CVD diamond membranes that are also provided mounted on Si frames (Applied Diamond). Two 800 nm-thick CVD diamond membranes provide >70% transmission in the region above 1.5 keV and provide a clean background in this region.
As shown in Figs. 2(b) and 4, the solution is introduced into the cell using a syringe. A second, open-top, vessel is used as a receiver to capture excess fluid during the filling process. For the ultra-thin X-ray windows used in this study, a small amount of hydrostatic pressure is sufficient to cause a small distension of the Si3N4 membrane windows. For this reason the elevation of the meniscus in the receiver vessel should approximately coincide with the elevation of the X-ray window. Small changes in the elevation of the receiver can be used to fine-tune the gap between the two X-ray windows. The receiver could also be attached to a vertical translator and stepper motor to remotely fine-tune the gap.
3. results
Fig. 5 shows a series of Cl K-edge spectra of hydrated Cl− in aqueous 6 M NaCl. Fig. 6 shows the XANES spectra for a series of polyoxoanions of chlorine. All of these chlorine K-edge (2822 eV) spectra were acquired at the bending-magnet beamline at the Advanced Photon Source, Argonne National Laboratory (XSD/PNC, Sector 20). A single 10 min scan is sufficient to obtain high-quality transmission spectra at this beamline. The incident beam was focused to a small spot (650 µm × 350 µm) using an Al2O3/Pt-coated toroidal mirror. The Al2O3 coating suppresses the influence of the Pt M-edges (Heald, 2011). Harmonic contamination was minimized by setting a Ni-coated X-ray mirror to ∼10.5 mrad, inserted in front of the beam-defining slits near the I0 detector. In combination with ∼25% crystal detuning, a high degree of harmonic rejection is attained. Energy calibration was accomplished using the first inflection point of the Ar K-edge (3205.9 eV). The estimated incident was ∼1 × 1010 photon s−1.
We used two standard gas-ionization detectors for transmission measurements at the Cl K-edge. The 18 cm-long upstream ionization detector, I0, was filled with a 4% N2/96% He mixture whereas the 30 cm-long downstream detector, IT, was filled with a 35% N2/65% He mixture. As shown in Fig. 2(b), the ends of these two detectors were extended to within 1–2 mm of the Si3N4 cell windows so that the air gap was reduced to minimize beam absorption by air. Each of these extenders was terminated with 4 µm-thick Prolene films. All of the beam path (e.g. slits and harmonic rejection mirror) upstream from these detectors was helium purged. As an alternate configuration the entire cell assembly could be placed inside a helium-purged tank or enclosure that is a common component of many low-energy beamlines.
In Fig. 5, three different sets of μ(E) spectra for a 6 M NaCl solution are presented. The spectra were acquired using three different pathlength settings, 60, 70 and 85 µm, resulting in absorption-edge heights of 1.42, 1.75 and 2.37, respectively. In the case of the 85 µm pathlength the beam transmission through the solution is only about 5%. The fact that the three spectra are almost identical, especially in the strongly absorbing white-line region about 2825 eV, indicates that the windows are parallel as required to prevent any leakage effects (Heald, 1988). The resulting k2χ(k) plots are shown in Fig. 5(b) after background subtraction. Notably, the spectra are almost identical, again supporting the equivalency of the different pathlengths and the parallelism of the windows. The Cl—O bond for the hydrated Cl−·H2On is quite disordered (Fulton & Balasubramanian, 2010) with a Debye–Waller factor of approximately 0.024 Å2. This means that the primary oscillations from the Cl—H2O scattering are strongly damped at higher k. Even so, with a single 10 min scan, the signal-to-noise ratio is sufficiently high to capture five or six oscillations from the backscattering of water. There is a well known multielectron excitation (KLII,III) at k = 8 Å−1 that interferes with the oscillation at higher k (Fulton & Balasubramanian, 2010). Finally, Fig. 6 shows the XANES region for aqueous Cl− and three different polyoxyanions of chlorine. The obtained spectra show the expected sensitivity to Cl speciation and validate the ability to obtain distortion-free spectra with high signal-to-noise ratios. The cell is easily loaded for each sample and the optimization of the pathlength while the cell is in the beam provides for efficient evaluation of multiple samples of this type.
Acknowledgements
This work was supported by the US Department of Energy's (DOE) Office of Basic Energy Sciences, Division of Chemical Sciences, Geosciences and Biosciences. PNNL is operated for the Department of Energy by Battelle. XSD/PNC facilities at the Advanced Photon Source, and research at these facilities, are supported by the US Department of Energy, Basic Energy Sciences, a major facilities access grant from NSERC, the University of Washington, Simon Fraser University, the Pacific Northwest National Laboratory and the Advanced Photon Source. Use of the Advanced Photon Source is also supported by the US Department of Energy, Office of Science, Office of Basic Energy Sciences, under Contract DE-AC02-06CH11357.
References
Aziz, E. F. (2010). J. Electron Spectrosc. Relat. Phenom. 177, 168–180. Web of Science CrossRef CAS Google Scholar
Aziz, E. F., Zimina, A., Freiwald, M., Eisebitt, S. & Eberhardt, W. (2006). J. Chem. Phys. 124, 114502. Web of Science CrossRef PubMed Google Scholar
Dang, L. X., Schenter, G. K., Glezakou, V. A. & Fulton, J. L. (2006). J. Phys. Chem. B, 110, 23644–23654. Web of Science CrossRef PubMed CAS Google Scholar
Edwards, D. C. & Myneni, S. C. (2006). J. Phys. Chem. A, 110, 11809–11818. Web of Science CrossRef PubMed CAS Google Scholar
Freiwald, M., Cramm, S., Eberhardt, W. & Eisebitt, S. (2004). J. Electron Spectrosc. Relat. Phenom. 137, 413–416. Web of Science CrossRef Google Scholar
Fulton, J. L. & Balasubramanian, M. (2010). J. Am. Chem. Soc. 132, 12597–12604. Web of Science CrossRef CAS PubMed Google Scholar
Fulton, J. L., Chen, Y., Heald, S. M. & Balasubramanian, M. (2004). Rev. Sci. Instrum. 75, 5228–5231. Web of Science CrossRef CAS Google Scholar
Fulton, J. L., Chen, Y., Heald, S. M. & Balasubramanian, M. (2006). J. Chem. Phys. 125, 094507. Web of Science CrossRef PubMed Google Scholar
Gao, J., Lowe, M. A., Kiya, Y. & Abruna, H. D. (2011). J. Phys. Chem. C, 115, 25132–25137. Web of Science CrossRef CAS Google Scholar
Guo, J. H. & Luo, Y. (2010). J. Electron Spectrosc. Relat. Phenom. 177, 181–191. Web of Science CrossRef CAS Google Scholar
Heald, S. M. (1988). X-ray Absorption: Principles, Applications, Techniques of EXAFS, SEXAFS and XANES, edited by D. C. Koningsberger and R. Prins, pp. 87–118. New York: John Wiley and Son. Google Scholar
Heald, S. M. (2011). Nucl. Instrum. Methods Phys. Res. A, 649, 128–130. Web of Science CrossRef CAS Google Scholar
Huthwelker, T., Zelenay, V., Birrer, M., Krepelova, A., Raabe, J., Tzvetkov, G., Vernooij, M. G. & Ammann, M. (2010). Rev. Sci. Instrum. 81, 113706. Web of Science CrossRef PubMed Google Scholar
Nagasaka, M., Hatsui, T., Horigome, T., Hamamura, Y. & Kosugi, N. (2010). J. Electron Spectrosc. Relat. Phenom. 177, 130–134. Web of Science CrossRef CAS Google Scholar
Näslund, L. A., Lüning, J., Ufuktepe, Y., Ogasawara, H., Wernet, P., Bergmann, U., Pettersson, L. G. & Nilsson, A. (2005). J. Phys. Chem. B, 109, 13835–13839. Web of Science PubMed Google Scholar
Roark, R. J. & Young, W. C. (1982). Formulas for Stress and Strain. New York: McGraw-Hill. Google Scholar
Rouff, A. A., Rabe, S., Nachtegaal, M. & Vogel, F. (2009). J. Phys. Chem. A, 113, 6895–6903. Web of Science CrossRef PubMed CAS Google Scholar
Schreck, S., Gavrila, G., Weniger, C. & Wernet, P. (2011). Rev. Sci. Instrum. 82, 103101. Web of Science CrossRef PubMed Google Scholar
Yang, B. X. & Kirz, J. (1987). Phys. Rev. B, 36, 1361–1364. CrossRef CAS Web of Science Google Scholar
© International Union of Crystallography. Prior permission is not required to reproduce short quotations, tables and figures from this article, provided the original authors and source are cited. For more information, click here.