research papers
A 30 nm-resolution hard X-ray microscope with
mapping capability at BSRFaBeijing Synchrotron Radiation Facility, Institute of High Energy Physics, Chinese Academy of Sciences, Beijing 100049, People's Republic of China, bNational Synchrotron Radiation Laboratory, University of Science and Technology of China, Hefei, Anhui 230026, People's Republic of China, and cXradia Inc., 4385 Hopyard Road, Suite 100, Pleasanton, CA 94588, USA
*Correspondence e-mail: wuzy@ihep.ac.cn
A full-field transmission X-ray microscope (TXM) operating continuously from 5 keV to 12 keV with fluorescence mapping capability has been designed and constructed at the Beijing Synchrotron Radiation Facility, a first-generation synchrotron radiation facility operating at 2.5 GeV. Spatial resolution better than 30 nm has been demonstrated using a Siemens star pattern in both absorption mode and Zernike phase-contrast mode. A scanning-probe mode fluorescence mapping capability integrated with the TXM has been shown to provide 50 p.p.m. sensitivity for trace elements with a spatial resolution (limited by probing beam spot size) of 20 µm. The optics design, testing of spatial resolution and fluorescence sensitivity are presented here, including performance measurement results.
Keywords: hard X-ray; transmission X-ray microscope; fluorescence mapping capability; trace-element mapping; zone plate microscopy.
1. Introduction
In the recent past, full-field transmission X-ray microscopy (TXM) operating in the hard X-ray regime has undergone rapid improvements based on the development of highly efficient X-ray optics for that regime (Zeng et al., 2008; Feng et al., 2007; Tkachuk et al., 2007). With outstanding high-throughput non-destructive three-dimensional imaging capabilities, TXM has attracted significant interest across many different disciplines. Many synchrotron radiation facilities have built this kind of hard X-ray microscope (Yin et al., 2006; Andrews et al., 2008; Chu et al., 2008; Tian et al., 2008) and made significant progress in this field (Chen et al., 2011a,b; Yi et al., 2011), along with a wide range of research work in biological (Andrews et al., 2010a,b, 2011; Liu et al., 2011a; Chen et al., 2011c) and materials science fields (Chen et al., 2008; Hsu et al., 2010; Meirer et al., 2011; Nelson et al., 2011; Sun & Wang, 2011).
The TXM program at the Beijing Synchrotron Radiation Facility (BSRF) is driven by the needs of a diverse science community preferring a non-destructive high-resolution probe. Examples are the analysis of failure mechanisms in microelectronic devices due to electro-migration, thermal breakdown or inhomogeneity in the industry field, and the characterization of porous materials such as soils and rock as well as the investigation of the transportation behavior in these porous structures. In the field of biological applications, TXMs working in the soft X-ray region around 500 eV (`water window' region) provide strong contrast between organic matter and water and have shown outstanding results recently (Hertz et al., 2012; Chichon et al., 2012). However, water-window TXM is limited to small micrometre-size cells because of the limited penetration and depth of focus at these photon energies. Hard X-ray TXM provides more flexibility to image larger cells, tissue sections and bone, as well as non-biological materials as mentioned above. However, for soft materials, the use of phase contrast is required. All these potential applications of TXM promoted the design and construction process in 2007. In October 2010, the BSRF TXM was commissioned in the 5–12 keV X-ray energy range, demonstrating spatial resolution better than 30 nm. In addition to high-resolution two-dimensional and three-dimensional full-field imaging, a fluorescence mapping mode is implemented in this instrument. BSRF is the only first-generation parasitical synchrotron radiation facility known to the authors at the time of writing where correlative 30 nm spatial resolution and fluorescence mapping have been achieved with TXM. The solution has been developed in the context of the BSRF instrumentation development program for a wide range of applications in materials science and biological research.
2. Instrument design
2.1. General description of the instrument
The TXM instrument, shown in Fig. 1, has an optical layout similar to a visible-light microscope. The X-ray beam is focused onto the sample by a condenser consisting of an elliptically shaped capillary (Zeng et al., 2008). Hereafter, the objective zone plate (Feng et al., 2007) produces a magnified image of the sample on a scintillator crystal. The resulting visual image is then further enlarged with a microscope objective lens and captured by a 16-bit 1024 × 1024 CCD camera for imaging with resolution down to 30 nm. In addition, the TXM is equipped with a set of phase rings to enable Zernike phase-contrast imaging at the same resolution level at fixed X-ray energies of 5 keV, 8 keV and 11 keV. The choice of discrete energies for Zernike phase contrast is due to the required matching of ring illumination from the (achromatic) condenser and the phase ring in the back focal plane of the (chromatic) objective lens. Since phase contrast is typically not used for spectroscopic studies, a set of discrete energies is adequate for most experiments of interest. Note that Zernike phase contrast is not a quantitative phase-contrast imaging method, and a few methods have been presented to achieve quantitative phase recovery (Yin et al., 2007; Liu et al., 2011b).
The TXM is also designed with a mechanism that enables switching from the synchrotron X-ray source to a rotating-anode laboratory source. This developed dual-source concept extends the usable operating time of the microscope since BSRF operates in a parasitic mode and only provides synchrotron radiation for a short part of the year. The rotating-anode X-ray generator has switchable copper and chromium targets with operating energies of 8.0 keV (Cu Kα) and 5.4 keV (Cr Kα), respectively. The sample and most optical components (except for the scintillator, visual magnification system and camera) are located in a vacuum chamber to reduce air absorption at the lower end of the energy range and to allow for experiments requiring specific atmospheric conditions. A total of five separate capillary condenser optics with optimized parameters are available in the instrument to cover all imaging modes available with the rotating-anode laboratory source and synchrotron source. Three imaging zone plates are installed to cover the energy range 5–12 keV and different imaging modes of the TXM. All optics are motorized and configuration changes between imaging modes are executed automatically.
An ion chamber is installed before the condenser to normalize exposure time to beam intensity, which has been measured to decay over the time scale of a few hours. To minimize exposure of the sample to radiation when not imaging, a shutter is installed just after the ion chamber. The shutter can be automatically controlled by the software.
The support table is equipped with passive dampening to isolate floor vibrations. According to the survey, vibrations are not a concern as proven by the resolution achieved (<30 nm).
2.2. Choice of beamline parameters
At the BSRF TXM beamline, polychromatic radiation is delivered by the single-period wiggler (Jiang et al., 1993) (wavelength shifter) 4W1 with a of 1.8 T and of 7.5 keV. A focusing beamline scheme is chosen to maximize the efficiency of the coupling to the TXM capillary condenser.
A general requirement for a TXM beamline and condenser system is that the illumination fills the beam acceptance of the microscope, defined as the product of the field of view and the acceptance angle of the objective lens (equal to two times the numerical aperture). For beamlines at third-generation synchrotron sources, the beam emittance, defined as the product of beam size and beam divergence, is typically smaller than the TXM phase-space acceptance at least in the vertical plane and sometimes also in the horizontal plane. The common practice is to choose the parameters of the beamline and TXM condenser optics to match the illumination angle of the sample to the numerical aperture (NA) of the objective zone plate (critical illumination) and to homogeneously distribute the focused beam by raster-scanning the condenser optic, with a corresponding loss in average
As shown below, the provided phase space of the BSRF source overfills the phase-space requirement of the microscope, avoiding the need for raster-scanning of the illumination. Note, however, that the total delivered into the microscope field of view is still significantly lower at a first-generation synchrotron than a similarly designed third-generation synchrotron beamline, emphasizing the need for a thorough design process for all optics.In the case of the BSRF TXM, the beamline source size is 2.46 mm horizontally (H) × 0.84 mm vertically (V) (FWHM), and the beam divergence is 0.38 mrad (H) and 0.26 mrad (V) FWHM. To illustrate the design considerations, we first calculate the matching of the available beamline phase space to the required phase space of the microscope for the case of 8 keV X-rays. The optical layout of beamline and TXM is shown in Fig. 2.
As shown in Fig. 2(a), the synchrotron X-ray beam passing through slit 1 is collimated vertically by mirror 1 and then focused by the cylindrical bending mirror 2. A double-crystal monochromator with a pair of Si(111) crystals is set between the two mirrors and tunes the X-ray energy in the range 5–12 keV. As the secondary source, slit 2 is set at the focus point of mirror 2. The Fresnel zone plate, installed after the sample position, serves as the objective imaging lens to produce a magnified image of the sample on the detector. When using the laboratory source, the laboratory source condenser illuminates the sample with a magnification factor of ∼1.
For zone-plate imaging, when illuminated by a parallel beam, the Rayleigh resolution is
According to Fig. 3, if the illumination is an inclined beam or hollow cone beam, the resolution δ on the sample obeys the Rayleigh criterion according to the following equation,
where f is the zone-plate focal length, D is the zone-plate diameter, λ is the X-ray wavelength, and 2θ is the angle deflected from the incident direction to the far side of the objective zone plate (two times the zone-plate numerical aperture, θ = λ/Δ2Rn). Then we have
According to (3), for an objective zone plate with 40 nm outermost zone width, a spatial resolution of 24–25 nm can be expected.
Based on (2), if we want a spatial resolution of 30 nm we have to have 2θ = 3.05 mrad with 8 keV X-rays (λ = 0.15 nm). Therefore, for a 20 µm field of view the required phase space provided by the condenser at the sample position should be equivalent to 20 µm × 3 mrad or 60 µm mrad.
The demagnification factor of the condenser optics has been chosen to be 10 for the 8 keV condenser in synchrotron operation, which translates into a required phase space of 200 µm × 0.3 mrad for the secondary source (point of the optical axis at the position of slit 2). According to Fig. 4, it is necessary for the condenser to obtain the same phase-space illumination from every point of the secondary source away from the central optical axis. Therefore, the divergence requirement should be enlarged by 200 µm/1820 mm = 0.11 mrad (200 µm is the size of slit 2). This results in the total phase space needed at the secondary source of 200 µm × 0.41 mrad, which will be shown below to be satisfied by the synchrotron source properties and the choice of beamline optics. The slope errors of the reflective monocapillary condensers are typically below 100 µrad FWHM, which for a working distance around 200 mm translates to a blur in the imaging plane below 40 µm.
From Figs. 2(b) and 2(c) the demagnification factor of the beamline can be seen to be 31/11.8 horizontally and 22/11.8 vertically. Thus, we calculate the necessary phase space at the source to be 525 µm × 0.156 mrad (H) and 373 µm × 0.22 mrad (V). The phase space provided by the BSRF-4W1 source is 2460 µm × 0.38 mrad horizontally and 840 µm × 0.26 mrad vertically, which easily overfills the phase-space requirement of the TXM provided the mirrors do not clip the beam divergence.
The incidence angle of the X-ray beam on both mirrors has been chosen to be 5 mrad to cover the planned operating energy range of 5–12 keV. Both mirrors are Rh-coated to push the critical angle to 5.6 mrad at 12 keV. The length of both mirrors has been chosen to be 1000 mm with a usable length of 900 mm, which limits the vertical angular acceptance of the first mirror to 0.20 mrad, just meeting the requirement of the phase space intercepted from the source. Since the beam is parallel between mirrors 1 and 2, the second mirror intercepts the whole vertical beam as well. In the horizontal plane the mirror width of 40 mm exceeds the necessary calculated width of 4.84 mm on the second mirror by a wide margin.
According to the above calculations, slit 1 should be set to restrict the angular divergence of the beam to 0.2 mrad vertically and 0.156 mrad horizontally to eliminate unnecessary beam on the optical elements.
The above design is based on ideal conditions of each optical element. The imperfections of optical elements (slope errors) will increase the phase space of the X-ray beam during its propagating process through beamline and TXM, which increases the uniformity of the illumination in the sample plane at the expense of
density.2.3. Design for using both synchrotron and laboratory source
As a parasitic facility, the beam time for BSRF for user operation is limited currently to three to four months per year. In order to fully utilize the instrument, a laboratory rotating-anode X-ray source with swappable copper and chromium target (Tkachuk et al., 2007) with 70 µm spot size was implemented in the BSRF-TXM. This source is on a manually operated stage and can be inserted into the beam path when the synchrotron does not supply X-ray beam (Wu et al., 2010).
In order to properly focus the beam at the same sample position, two different types of condensers were designed, one set for the synchrotron beam and the other set for the laboratory source, as shown in Fig. 5. The focusing conjugates of these condensers are significantly different. In the case of the synchrotron beam, the distance between the secondary source (slit 2) and the capillary condenser is 2 m and the condenser demagnifies the source by ten times. In the case of the laboratory source, the condenser is a 1:1 focusing lens with a conjugate distance between source and sample of 0.5 m only. This is chosen to most effectively utilize each X-ray source, appropriately filling a large field of view on the sample plane while simultaneously providing high spatial resolution.
2.4. Design for fluorescence capability
The ability to collect an
signal from the sample has been integrated into the design of the instrument. Based on the described design, the TXM condenser illuminates a field of view of approximately 20 µm × 20 µm at the sample position. Although this size is not comparable with micrometre or even submicrometre spots achieved by dedicated state-of-the-art fluorescence instruments existing in many synchrotron facilities, it is still very useful for users to measure the elemental composition within the field of view in addition to taking high-resolution images. Additionally, a raster scan of the sample can provide a coarse-resolution map of an extended sample area. For fluorescence operation, a Si(Li) detector (model XR-100CR, Amptek) is placed perpendicular to the TXM optical axis as EDS detector to minimize the contribution to the signal by The software allows data acquisition for to be run concurrently with the TXM control software.3. Instrument performance
3.1. Photon flux
The 10 photons s−1 at 8 keV. A measurement after the slit with the slit setting typically used for imaging is not available at the time of writing.
of the beamline just before slit 2 has been measured using an ion chamber to be about 3 × 10For a typical flat-field image (no sample) with 15 nm pixel size, we measure about 200 camera counts per second per pixel without pixel binning. Using a calibration of 16 camera counts per 8 keV X-ray photon (as measured previously for an equivalent scintillator/optics/camera set-up using a large-area photon-counting detector) and 12% measured efficiency for the first diffraction order of the 1400 nm-thick double-stacked zone plate (Feng et al., 2007; Chen et al., 2008), this results in a total of about 100 photons s−1 pixel−1 in the sample plane. Comparing this with the of 3 × 1010 photons s−1 before slit 2 and considering the image size of 1024 × 1024 pixels, the effective transmission of the slit/condenser system to the image plane is about 3.5 × 10−3. Part of this loss of is due to intentional blockage by the slits and central stop on the condenser (which is needed to provide hollow cone illumination, see Fig. 2, and for the condenser used covers about 80% of the condenser area), as well as the known condenser reflectivity of about 80%. The remaining losses are possibly due to illumination spread outside the condenser aperture or outside the imaged area in the sample plane.
Typical exposure times without pixel binning are about 90–120 s per two-dimensional
using 8 keV synchrotron radiation X-rays. Using 181 radiographs at 90 s per view will result in a total exposure time of about 5.5 h in synchrotron operation mode (scaling exposure time to match ion chamber current) for a full tomogram. Depending on the experimental requirements, the exposure time can be reduced by using pixel binning, with a corresponding loss of pixel resolution.Comparing the camera counts in flat-field images, the laboratory source with its dedicated condenser provides about 10% of the
of the synchrotron in the sample plane.3.2. TXM spatial resolution
The spatial resolution was tested by imaging an electroplated gold star pattern with the laboratory X-ray source and with the synchrotron beam.
With the laboratory source using the copper target (8.0 keV), a Siemens star with 50 nm finest features was imaged in two-dimensions as well as three-dimensions using a zone plate with 40 nm outermost zone width. A total of 121 sequential radiographs were acquired over an angular rotation range of 120° with an exposure time of 900 s for each indicate that both two-dimensional and three-dimensional spatial resolution of 50 nm have been achieved for the laboratory X-ray source, limited by the large spectral bandwidth of the Cu Kα emission, which introduces chromatic aberration.
and a pixel size of 15 nm. Tomographic reconstruction was performed using a filtered backprojection algorithm. The results shown in Fig. 6For the synchrotron beam, TXM spatial resolution was tested both in absorption and Zernike phase-contrast mode. Fig. 7(a) shows an absorption image taken at 8 keV of a star pattern with features tapering to a 30 nm minimum feature size at the centre of the pattern. Fig. 7(b) is a Zernike phase-contrast image taken at 8 keV of the same star pattern. Fig. 7(c) shows an absorption image taken at 8 keV of another star pattern with 24 nm minimum feature size at the centre. A more accurate analysis of the whole area of Fig. 7(c) using the accepted technique (Wang et al., 2012; Yi et al., 2011; Chen et al., 2011a,b) of radial power spectrum density (RPSD) is shown in Fig. 7(d). It can be seen that the power spectrum of the sample cuts off at a spatial frequency of 19 µm−1, corresponding to a period of about 52 nm. Thus, the TXM spatial resolution, which is usually represented by half the spatial period (line width), can be concluded to be 26 nm.
3.3. Fluorescence capability
), which allows the role of trace elements to be investigated in the function of biological systems. According to the blade scan results shown in Fig. 8, with slit 2 set at 200 µm × 200 µm, the condenser focal spot size at the sample plane is about 18.5 µm × 17.5 µm (FWHM), which can be used to carry out fluorescent mapping experiments while acquiring images.
offers elemental analysis with high sensitivity (Gordon, 1982To assess the sensitivity and K or L series, as shown in Fig. 9, were fitted with PyMCA (Solé et al., 2007) freely provided by the software group of ESRF for interactive data processing. Excitation lines with energies lower than 1 keV cannot be detected with good signal-to-noise ratio owing to absorption of the air path of about 5 cm. Note that a 5 cm air path before the fluorescence detector only existed at the time of commissioning when the fluorescence sensitivity was tested. Right now, the fluorescence detector can be put in the whole The fitting results match with elements given by NIST (including the Ar K peak 2.9 keV, which exists in the air). Thus, we conclude that the fluorescence sensitivity of the BSRF TXM is substantially better than 50 p.p.m, since the peaks are very nicely resolved indicating that lower concentrations of the elements of interest would still be readily detectable.
of the mode, a trace-element analysis experiment was performed using the number 612 standard sample from US National Institute of Standards and Technology (NIST), whose nominal trace-element concentration is 50 p.p.m. The `fluorescence spectra with elements' emission line series, including the4. Initial applications examples
The first sample imaged with BSRF-TXM is shale as shown in Fig. 10. Shale is typically composed of variable amounts of clay minerals and quartz grains. The oil and gas industry is now looking toward shale as a source of stored hydrocarbons, which may be extracted and converted into fossil fuel. It is believed that the stored energy may exist within the pore space, but the pores typically range in size from ∼1 µm to ∼10 nm, so conventional micro-computed tomography does not allow for visualization of these fine features. Fig. 10(a) presents a virtual slice image of the shale sample, in which the grey value of each pixel corresponds to the extent of the shift of X-ray phase and amplitude. The image, collected at 8 keV utilizing absorption contrast, demonstrates sufficient contrast and spatial resolution to differentiate individual pore segments (dark grey), quartz matrix (medium grey) and some inclusions (bright grey/white). The three-dimensional volume rendering of this result is also displayed in Fig. 10(b). Figs. 10(c) and 10(d) present the two-dimensional and three-dimensional image segmentation, respectively, of this sample, produced using the Avizo Fire software (Visualization Sciences Group, Bordeaux, France). According to the results of image segmentation, the quantitative volume for the pores, quartz matrix and inclusions are 1.58 × 1012 nm3, 4.32 × 1012 nm3 and 4.49 × 1012 nm3, respectively, corresponding to a porosity of 3.20%.
Another sample is an assembled particle made of zinc silicate (Qu et al., 2012), roughly spherical with a diameter of about 20 µm. This kind of hierarchical structured material is used as adsorbents in water treatment and catalyst applications and its internal structure influences the growth mechanism and corresponding chemical performance. However, because the sample is too thick for the transmission of the TEM (transmission electron microscope) electron beam, it is difficult to observe the internal structure by TEM (Qu et al., 2012). Using TXM with Zernike phase contrast at 8 keV, the true internal structure of the particle was successfully imaged, as shown in Fig. 11. From the virtual slice shown in Fig. 11(a), the interior of the sample can be clearly observed, while Fig. 11(b) shows some assembled defects, indicated with elliptical annotations. These show that some micro-rods are misaligned inside the regular micro-rod array, possibly reflecting improper processing parameters. The structures revealed by TXM may play a great role in the research and development of assembly conditions.
5. Conclusions
A high-resolution full-field transmission hard X-ray microscope with fluorescence mapping capability was constructed at the first-generation synchrotron facility BSRF. The beamline design criteria for the instrument are described in detail based on the transfer of phase space. Testing results give a spatial resolution of better than 30 nm and fluorescence sensitivity better than 50 p.p.m. The instrument is now entering user operation and will be put to use as a new tool in the fields of biology and materials science.
Acknowledgements
This work was supported by the States Key Project for Fundamental Research (2012CB825800, 2009CB930804), the Key Important Project of the National Natural Science Foundation of China (10734070), the Knowledge Innovation Program of the Chinese Academy of Sciences (KJCX2-YW-N42) and the National Natural Science Foundation of China (10979055). The authors also wish to acknowledge the assistance of Mr George Hsu, Mr Jeff Irwin and Dr Christian Holzner for their assistance on this project.
References
Andrews, J. C., Almeida, E., Meulen, M. C. H., Alwood, J. S., Lee, C., Liu, Y., Chen, J., Meirer, F., Feser, M., Gelb, J., Rudati, J., Tkachuk, A., Yun, W. & Pianetta, P. (2010a). Microsc. Microanal. 16, 327–336. Web of Science CrossRef CAS PubMed Google Scholar
Andrews, J. C., Brennan, S., Patty, C., Luening, K., Pianetta, P., Almeida, E., Meulen, M. C. H., Feser, M., Gelb, J., Rudati, J., Tkachuk, A. & Yun, W. (2008). Synchrotron Rad. News, 21, 17–26. CrossRef Google Scholar
Andrews, J. C., Meirer, F., Liu, Y., Mester, Z. & Pianetta, P. (2011). Microsc. Res. Tech. 74, 671–681. Web of Science CrossRef CAS PubMed Google Scholar
Andrews, J. C., Pianetta, P., Meirer, F., Chen, J., Almeida, E., van der Meulen, M. C., Alwood, J. S., Lee, C., Zhu, J. & Cui, Y. (2010b). AIP Conf. Proc. 1234, 79–82. CrossRef PubMed Google Scholar
Chen, T. Y., Chen, Y. T., Wang, C. L., Kempson, I. M., Lee, W. K., Chu, Y. S., Hwu, Y. & Margaritondo, G. (2011a). Opt. Express, 19, 19919–19924. Web of Science CrossRef PubMed Google Scholar
Chen, Y. T., Chen, T. Y., Yi, J., Chu, Y. S., Lee, W. K., Wang, C. L., Kempson, I. M., Hwu, Y., Gajdosik, V. & Margaritondo, G. (2011b). Opt. Lett. 36, 1269–1271. CrossRef PubMed Google Scholar
Chen, H. H., Chien, C. C., Petibois, C., Wang, C. L., Chu, Y. S., Lai, S. F., Hua, T. E., Chen, Y. Y., Cai, X., Kempson, I. M., Hwu, Y. & Margaritondo, G. (2011c). J. Nanobiotechnol. 9, 14. Web of Science CrossRef Google Scholar
Chen, J., Wu, C., Tian, J., Li, W., Yu, S. & Tian, Y. (2008). Appl. Phys. Lett. 92, 233104. Web of Science CrossRef Google Scholar
Chichon, F. J., Rodriguez, M. J., Pereiro, E., Chiappi, M., Perdiguero, B., Guttmann, P., Werner, S., Rehbein, S., Schneider, G., Esteban, M. & Carrascosa, J. L. (2012). J. Struct. Biol. 177, 202–211. Web of Science PubMed Google Scholar
Chu, Y. S., Yi, J. M., De Carlo, F., Shen, Q., Lee, Wah-Keat, Wu, H. J., Wang, C. L., Wang, J. Y., Liu, C. J., Wang, C. H., Wu, S. R., Chien, C. C., Hwu, Y., Tkachuk, A., Yun, W., Feser, M., Liang, K. S., Yang, C. S., Je, J. H. & Margaritondo, G. (2008). Appl. Phys. Lett. 92, 103119. Web of Science CrossRef Google Scholar
Feng, Y., Feser, M., Lyon, A., Rishton, S., Zeng, X., Chen, S., Sassolini, S. & Yun, W. (2007). J. Vac. Sci. Technol. B, 25, 2004–2007. Web of Science CrossRef CAS Google Scholar
Gordon, B. M. (1982). Nucl. Instrum. Methods, 204, 223–229. CrossRef CAS Google Scholar
Hertz, H. M., Hofsten, O., Bertilson, M., Vogt, U., Holmberg, A., Reinspach, J., Martz, D., Selin, M., Christakou, A. E., Jerlstrom-Hultqvist, J. & Svard, S. (2012). J. Struct. Biol. 177, 267–272. Web of Science CrossRef CAS PubMed Google Scholar
Hsu, P. C., Chu, Y., Yi, J., Wang, C. L., Wu, S. R., Hwu, Y. & Margaritondo, G. (2010). Appl. Phys. Lett. 97, 033101. Web of Science CrossRef Google Scholar
Jiang, J., Zhao, J., Tian, Y., Han, Y., Chao, Z., Jiang, X. & Xian, D. (1993). Nucl. Instrum. Methods Phys. Res. A, 336, 354–360. CrossRef CAS Web of Science Google Scholar
Liu, Y., Andrews, J. C., Meirer, F., Mehta, A., Gil, S. C., Sciau, P., Mester, Z. & Pianetta, P. (2011a). AIP Conf. Proc. 1365, 357–360. CrossRef Google Scholar
Liu, Y., Andrews, J. C., Wang, J., Meirer, F., Zhu, P., Wu, Z. & Pianetta, P. (2011b). Opt. Express, 19, 540–545. Web of Science CrossRef PubMed Google Scholar
Meirer, F., Cabana, J., Liu, Y., Mehta, A., Andrews, J. C. & Pianetta, P. (2011). J. Synchrotron Rad. 18, 773–781. Web of Science CrossRef CAS IUCr Journals Google Scholar
Nelson, G. J., Harris, W. M., Izzo, J. R. Jr, Grew, K. N., Chiu, W. K. S., Chu, Y. S., Yi, J., Andrews, J. C., Liu, Y. & Pianetta, P. (2011). Appl. Phys. Lett. 98, 173109. Web of Science CrossRef Google Scholar
Qu, J., Cao, C., Hong, Y., Chen, C., Zhu, P., Song, W. & Wu, Z. (2012). J. Mater. Chem. 22, 3562–3567. Web of Science CrossRef CAS Google Scholar
Solé, V. A., Papillon, E., Cotte, M., Walter, P. & Susini, J. (2007). Spectrochim. Acta B, 62, 63–68. Google Scholar
Sun, Y. & Wang, Y. (2011). Nano Lett. 11, 4386–4392. Web of Science CrossRef CAS PubMed Google Scholar
Tian, Y., Li, W., Chen, J., Liu, L., Liu, G., Tkachuk, A., Tian, J., Xiong, Y., Gelb, J., Hsu, G. & Yun, W. (2008). Rev. Sci. Instrum. 79, 103708. Web of Science CrossRef PubMed Google Scholar
Tkachuk, A., Duewer, F., Cui, H., Feser, M., Wang, S. & Yun, W. (2007). Z. Kristallogr. 222, 650–655. Web of Science CrossRef CAS Google Scholar
Wang, J., Chen, Y. K., Yuan, Q., Tkachuk, A., Erdonmez, C., Hornberger, B. & Feser, M. (2012). Appl. Phys. Lett. 100, 143107. Web of Science CrossRef Google Scholar
Wu, Z., Yun, W., Zhu, P., Wang, Y., Yuan, Q., Tkachuk, A., Huang, W. & Fester, M. (2010). US Patent 7 796 725 B1. Google Scholar
Yi, J., Chu, Y. S., Chen, Y. T., Chen, T. Y., Hwu, Y. & Margaritondo, G. (2011). J. Phys. D, 44, 232001. Web of Science CrossRef Google Scholar
Yin, G.-C., Chen, F.-R., Hwu, Y., Shieh, H.-P. & Liang, K. (2007). Appl. Phys. Lett. 90, 181118. Web of Science CrossRef Google Scholar
Yin, G.-C., Tang, M.-T., Song, Y.-F., Chen, F.-R., Liang, K. S., Duewer, F. W., Yun, W., Ko, C.-H. & Shieh, H.-P. D. (2006). Appl. Phys. Lett. 88, 241115. Web of Science CrossRef Google Scholar
Zeng, X., Duewer, F., Feser, M., Huang, C., Lyon, A., Tkachuk, A. & Yun, W. (2008). Appl. Opt. 41, 2376–2381. Web of Science CrossRef Google Scholar
© International Union of Crystallography. Prior permission is not required to reproduce short quotations, tables and figures from this article, provided the original authors and source are cited. For more information, click here.