research papers
High-resolution high-efficiency X-ray imaging system based on the in-line Bragg magnifier and the Medipix detector
aANKA Light Source, Karlsruhe Institute of Technology, Karlsruhe, Germany, bInstitute for Electrical Engineering, Slovak Academy of Sciences, Bratislava, Slovakia, cFMF, Freiburg, Germany, dFNSPE CTU in Prague, Czech Republic, eSchool of Physics, Monash University, Victoria 3800, Australia, fESRF, Grenoble, France, gAcademy of Sciences of the Czech Republic, Prague, Czech Republic, hRigaku Innovative Technologies Europe, Prague, Czech Republic, iDiamond Light Source, UK, and jSwiss Light Source, PSI, Switzerland
*Correspondence e-mail: patrik@tagen.tohoku.ac.jp
The performance of a recently developed full-field X-ray micro-imaging system based on an in-line Bragg magnifier is reported. The system is composed of quasi-channel-cut crystals in combination with a Medipix single-photon-counting detector. A theoretical and experimental study of the imaging performance of the crystals–detector combination and a comparison with a standard indirect detector typically used in high-resolution X-ray imaging schemes are reported. The spatial resolution attained by our system is about 0.75 µm, limited only by the current magnification. Compared with an indirect detector system, this system features a better efficiency, signal-to-noise ratio and spatial resolution. The optimal working resolution range of this system is between ∼0.4 µm and 1 µm, filling the gap between transmission X-ray microscopes and indirect detectors. Applications for coherent full-field imaging of weakly absorbing samples are shown and discussed.
Keywords: Bragg magnifier; germanium; holography; high resolution; in-line; X-ray imaging.
1. Introduction
Hard X-ray methods with sub-micrometer spatial resolution are currently utilized in modern synchrotron facilities and, in selected cases, with laboratory sources. Typically one can use a configuration with parallel X-ray beam and high-resolution detector or X-ray magnification in projection geometry. The former is commonly a synchrotron technique while the latter, mainly implemented in laboratory set-ups, is generally limited by the size of the source (typically a few micrometers). On the other hand, established high-resolution X-ray detecting systems are based on the combination of scintillating screens, diffraction-limited optical microscopes and CCD or CMOS cameras, called indirect detectors (Bonse & Bush, 1996). Typical resolution is slightly above 1 µm, obtainable by magnifying the converted visible light. The need for resolution implies that the scintillating screen must be thin enough to avoid spreading the (incoherent) scintillating process over a large volume. Therefore such conversion is particularly inefficient for hard X-rays, where more than 60% of the usable signal is lost (Cecilia et al., 2011).
Detector limitations can be easily overcome by magnifying the X-ray beam prior to a detector. In this way both resolution and sensitivity of X-ray imaging systems can be improved, by employing magnifying optics such as Fresnel zone plates (Baez, 1960), refractive X-ray lenses (Suehiro et al., 1991), curved mirrors in crossed configuration (Kirkpatrick–Baez, KB) (Kirkpatrick & Baez, 1948) or X-ray waveguides (Pelliccia et al., 2010). Where the objective lens is placed after the sample, high resolution has been demonstrated in a full-field configuration. Another approach, which does not need a high-resolution objective lens, consists of the use of a Bragg magnifying optics (Boetinger et al., 1979; Kagoshima et al., 2000; Kobayashi et al., 2001; Spal, 2001; Stampanoni et al., 2002; Schäfer & Köhler, 2003; Vagovič et al., 2011). The advantage of such an approach is to keep the X-ray beam divergence very low, so the effective source size does not play a significant role, as in cone-beam laboratory imaging set-ups. The Bragg magnifier can be placed in front of or behind the sample. In the former case, the divergence decrease provides, after the magnifier, a quasi-parallel larger beam illumination to the sample. The latter case, which is the subject of the present study, corresponds to a magnification of the sample projected image. The image is then recorded by a sensitive moderate-resolution X-ray detector such as the single-photon-counting detector Medipix (Llopart et al., 2002).
The combination of a single-photon-counting detector (PILATUS) with silicon-based Bragg magnifier was first demonstrated by Stampanoni et al. (2006). In this work, besides the novelty of the approach, the obtainable image quality was far from being of practical use because of strong geometrical deformations and the very small field of view. This was due to very high magnification obtained with a single reflection, in conjunction with a low-spatial-resolution detector such as PILATUS (172 µm pixel size).
In this work we demonstrate the performance of a full-field micro-imaging system based on an in-line germanium Bragg magnifier composed of two identical quasi-channel-cut crystals in combination with the single-photon-counting detector Medipix (BMM). With this approach it is not necessary to produce extremely high magnification with a single crystal; therefore, the requirements for the crystal surface quality are less stringent, yet the total magnification can be extremely high. Furthermore, the combination of two crystals in the (+n, −n) configuration leads to an in-line set-up which is technically more convenient compared with Bragg magnifiers composed only of a single crystal (magnifying in one direction). We present the description of the device, design parameters and the experimental results obtained on beamline B16 (Sawhney et al., 2010) at Diamond Light Source (Oxford, UK), the Optics beamline at the Swiss Light Source (Villigen, Switzerland) and ANKA light source (Karlsruhe, Germany). We confront, both qualitatively and quantitatively, the imaging performances of the BMM with an indirect detector with comparable resolution. Finally we address the capability of our set-up for coherent full-field imaging schemes.
2. Description of the device
The Bragg magnifier scheme introduced by Vagovič et al. (2011) demonstrated the possibility of performing high-resolution imaging with a pair of crossed monolithic channel-cut crystals. This configuration is clearly favorable in terms of easy alignment procedures. Nevertheless, a major limitation has been identified. In order to build a two-dimensional system, two channel-cut crystals have to be employed. It is extremely difficult to have exactly the same magnification (i.e. the same asymmetry angles) for the two monolithic pieces and therefore the magnification factor is usually different for the two directions. If we consider in addition that the surface quality of a channel-cut crystal is worse than the surface quality of a flat crystal, we understand that such a set-up is prone to distortions and aberrations when high magnification is to be achieved. Therefore such a configuration was not able to reach an aberration-free sub-micrometer resolution with a Medipix detector. Based on this experience we designed and developed an improved in-line Bragg magnifier based on quasi-channel-cut germanium crystals. The improved set-up consists of four independent crystals, each using the (220) diffraction plane, arranged in the σ(+n, −n) − π(+n, −n) configuration. This choice has several advantages over the use of a channel-cut crystal. (i) The crystal size can be easily made larger, leading to an increased field of view of the optical system. (ii) A flat crystal can be polished to a higher degree of accuracy with respect to a channel-cut crystal, which is absolutely crucial to minimize the effect on the coherence properties of the X-ray beam. (iii) Each crystal can be designed and aligned to impart the required magnification factor, to realise a uniform two-dimensional magnification. The full-field microscope is built with the crystals arranged in crossed configuration. A three-dimensional model of the optical set-up is shown in Fig. 1. The optical design allowed a 250-fold magnification to be obtained. The crystals have been prepared in the optical workshop of the European Synchrotron Radiation Facility (Grenoble, France) using a planar polishing technology (Barrett et al., 2010). On the other hand, the use of independent crystals imposes more stringent requirements on the mechanical housing and the use of a more accurate alignment system. The adjustment of the diffraction planes in each quasi-channel-cut crystal is made with three piezo-driven motors.
The above-mentioned modifications make the set-up suitable for real-time X-ray imaging with high spatial resolution, estimated to be below 0.5 µm with the Medipix detector having 55 µm pixel size. The spatial resolution is limited by the numerical aperture of the Bragg magnifier (input angular acceptances) and the used experimental geometry (source size, source-to-sample and sample-to-magnifier distances). The blur caused by the penetration of X-rays into the crystals, as shown by Vagovič et al. (2011), in the case of extreme asymmetric cases is very small, below the resolution limit of the Bragg magnifier given by its numerical aperture, and therefore it does not limit the spatial resolution.
2.1. Device parameters
The asymmetry angles α1 and α2 of the first and second crystal forming the quasi-channel-cut have been specified to reach a 250-fold magnification at an X-ray energy of 11 keV. Such a value is the energy for which the angular acceptance of the system is maximized. In this sense we speak about the optimal energy. As discussed in the preceding sections, such magnifications cannot be reached with a single crystal. However, using two crystals, the magnification required for the second crystal is smaller (lower asymmetry angle) than the first, the total magnification being the product of the magnification on the first and second crystal. A plot of the total magnification as a function of the incident X-ray energy is shown in Fig. 2. The main parameters are tabulated in Table 1.
|
The angular acceptances and the spectral bandwidth were calculated from the spectral-angular function (SAF) which is the distribution of the reflection coefficients as a function of three coordinates: input angle of the X-ray wavevector with respect to the central beam in the horizontal plane θ, in the vertical plane φ and wavelength λ (or energy E) (Vagovič et al., 2007). The BMM can be modeled as a σ(+n, −n) − π(+n, −n) successive crystal configuration. The SAF of this system can be calculated using the two-beam of X-ray diffraction (Authier, 2004) employing universal computation algorithms (Huang & Dudley, 2003). The SAF is first calculated for each energy point, yielding a function of three variables R(θ, φ, E); then it is integrated over any two variables and the FWHM of the one-dimensional resulting function is estimated. We call these values Δθ(E), Δφ(E) and ΔE(E). In our case, the system being symmetric in the horizontal and vertical directions, the two angular variables are equivalent. A plot of angular acceptances and bandwidth is shown in Fig. 3. The position of the maxima of the angular acceptances and bandwidth are slightly different. This means that the optimal efficiency (larger accepted bandwidth) and the optimal angular acceptance are not reached for the same energy, and a compromise is required when choosing the working parameters. We opted to design the system for optimal angular acceptance, in order to maximize the attainable spatial resolution. For illustration, the angular acceptance window of the Bragg magnifier, i.e. the SAF integrated over an energy range of 2 eV, is shown in Fig. 4. It is worth noting that the spectral dispersion of each quasi-channel-cut crystal is considerably large in the diffraction plane while it is rather small in the perpendicular plane. Therefore, in the crossed σ–π configuration (the diffraction planes being perpendicular), the spectral dispersion of either channel-cut crystal in its diffraction plane is limited by the small spectral dispersion of the other one. Thus the magnifier is selecting a narrow bandwidth of about 1 eV out of a polychromatic illumination.
2.2. Efficiency of the Bragg magnifier and Medipix detector
The efficiency of the device is a crucial parameter, especially in view of possible applications in real-time imaging with a photon-counting detector. Because the Bragg magnifier consists of perfect dislocation-free germanium crystals, one can estimate the efficiency from its SAF. We call this parameter the diffraction efficiency of the Bragg magnifier, ηD. The radiation which is transmitted through the (whose reflectivity is governed by the reflection coefficient) is then recorded by the detector. We call ηC the of the detector. Finally, the efficiency of the system is defined as
In order to estimate the diffraction efficiency of the BMM one has to integrate the SAF over the three variables and normalize it by the integration volume as was suggested by Vagovič et al. (2011),
In (2), Ω is the integrating volume and R(θ, φ, λ) is the reflection coefficient of the crystal configuration given by
where n = 1, 2, 3, 4 is an index representing at which consecutive reflection the SAF has to be calculated. In Fig. 5 the plot of the BMM efficiency versus the magnification is shown, calculated for an 11 keV central photon energy. As we can see the BMM efficiency is, at first, growing with the magnification. The reason for this is the increase of the angular and spectral acceptances of the device. Finally, when the angle of input radiation is approaching the critical angle, the efficiency together with the angular and spectral acceptances are decreasing. The optimal efficiency of 4.85% is reached for a total magnification of 175. As previously explained, we designed the BMM to have 250-fold magnification at 11 keV. Therefore the diffraction efficiency is reduced to 4.10%.
3. Experimental tests
The BMM was experimentally tested on the TopoTomo beamline at ANKA, beamline B16 at Diamond Light Source (Sawhney et al., 2010) and on the X05DA beamline at SLS (Flechsig et al., 2009). A photograph of the experimental arrangement at B16 is shown in Fig. 6. The detector used was a silicon QUAD Medipix2 detector that was assembled at the National Institute for Subatomic Physics (NIKHEF, Amsterdam, The Netherlands). The sensor material thickness was equal to 300 µm with an active detector size of 28 mm × 28 mm (four Medipix readout chips in a 2 × 2 matrix) and a pixel dimension of 55 µm × 55 µm. The QUAD detector was controlled by using the Pixelman software package (Turecek et al., 2011) and the USB interface for Medipix2 (Vykydal et al., 2006).
3.1. Imaging performance comparison between the indirect detector and Bragg magnifier with Medipix
A comparison of the performances of the BMM with an imaging system based on a scintillator coupled to a CCD is not a straightforward task. The Medipix detector is working in single-photon-counting mode and by setting an ) which is comparable with the bandwidth of the DCM. Thus both the Medipix and the indirect detector system are illuminated with radiation of similar bandwidth. The indirect detector used was composed of (i) LAG 5 µm-thick scintillator, (ii) Optique Peter diffraction-limited microscope with available magnifications of 4, 10 and 20-fold, and (iii) CCD camera PCO4000 with a pixel size of 9 µm × 9 µm.
with essentially no The indirect detector with CCD is integrating charges and a is present. To fairly compare these two systems we should proceed on a qualitative and quantitative basis. We can compare the achieved contrast (or visibility) for a given acquisition time, the signal-to-noise ratio (SNR) as a function of the acquisition time and the detected passing per unit of area through the sample per unit of time. For the illumination of the Bragg magnifier we used the radiation pre-monochromated by a double-multilayer monochromator (DMM). Irradiation of the indirect detector system was performed using radiation delivered by a double-crystal monochromator (DCM). We consider the illumination conditions for the comparison to be correct because the Bragg magnifier itself acts as a crystal monochromator with a bandwidth of about 1 eV (see Fig. 33.1.1. Qualitative comparison
For the first test, suitable for an easy comparison, we recorded the image of an X-radia X500-300-20 test pattern within 1 s using both the BMM and the indirect detector systems. The exposure time was 1 s and the X-ray energy was set to 10.6 keV. In the case of the BMM the effective pixel size was 0.39 µm and in the case of the indirect detector it was 0.45 µm, thus very similar values. The result is shown in Fig. 7. The quality of the image recorded by the BMM is significantly better while the poor SNR is the single most limiting factor for the indirect detector.
Next we analysed the image contrast at selected spatial frequencies of increasing the horizontal half-pitch pattern of X500-300-20: the values are gathered in Table 2. The visibility was calculated according to V = (Imax − Imin)/(Imax + Imin), where Imax and Imin are the maximum and minimum intensities at the given spatial frequency. For the analysis of the contrast we used an exposure time of 1 s in the case of the BMM and 10 s for the indirect detector. This choice was unavoidable in order to reach sufficient contrast to perform a meaningful visibility estimation in the latter case. As can be clearly seen, the contrast in the case of the BMM is significantly better even with a ten times shorter acquisition time.
|
3.1.2. Quantitative comparison
To compare both systems in a quantitative way we can, for example, analyse the SNR of recorded images or we can compare the
of X-rays passing through a unit of area at the sample which is detected by the system. The second comparison will give us the information about the efficiency of both systems.For the signal-to-noise analysis we used flat-field images recorded at an X-ray energy of 10.6 keV. The BMM was irradiated by the radiation delivered by the DMM and the indirect detector with the radiation delivered by the DCM. We calculated the signal-to-noise ratio according to following equation,
where S is the measured signal, σD is the dark noise and σP is the photonic noise. In the case of the Medipix detector the dark noise is absent. The SNR has been estimated for different exposure times and the profiles are plotted in Fig. 8 for both BMM and indirect detector. The absence of greatly contributes to the better BMM performances.
To compare the detected . In the case of the BMM we can see that the increase of the detected per unit of area towards the smaller effective pixel sizes (larger magnifications) is caused by an increase of the BMM efficiency. This is in very good agreement with the theoretical simulation, as shown in Fig. 5. In the case of the indirect detector the trend is opposite, influenced by the decrease of the SNR towards the smaller effective pixel sizes. The detected in the case of the indirect detector was calculated using the conversion efficiency equation given in Appendix A.
passing through a unit area at the sample we used several magnifications for both systems. The results are shown in Fig. 93.2. Spatial resolution estimation
Because the images are recorded in a holographic regime, the spatial resolution cannot be estimated by straightforward analysis. However, one can estimate the maximum spatial frequency for which the Fourier transform of the recorded image still displays non-negligible amplitude. For this analysis we used images recorded at 10.7 keV. For practical reasons we stayed at 10.7 keV because for higher energies the field of view was very small (<200 µm). The spatial resolution at 10.7 keV estimated by the Fourier analysis was 0.75 µm (Fig. 10).
4. Imaging examples
As mentioned above, we address the capability of our set-up for coherent full-field imaging schemes. We report the image of weakly absorbing PMMA spheres on SiN membrane, recorded in the holographic regime (Fig. 11a). Being able to record holograms enables the phase shift introduced by the sample to be retrieved (if the field of view is coherently illuminated). This can be done, for example, by adopting iterative phase-retrieval algorithms from coherent diffraction imaging (Miao et al., 1999). To demonstrate this approach we modified the hybrid input–output algorithm (Fienup, 1982) for the Bragg magnifier and applied it to the phase reconstruction of the image of PMMA spheres. The image (Fig. 11b) shows the preliminary result (still containing some reconstruction artefacts) of the retrieved phase map. The development of the algorithm is in progress and its description will be given in a future work.
5. Conclusions
We have shown the performance of a novel X-ray imaging system based on a Bragg magnifier and Medipix. The results demonstrate that this system can be used for high-resolution hard X-ray holography. The performances of the system have been compared with those of a high-resolution indirect detection system. The achieved results show a clear advantage in using an X-ray magnifying system combined with single-photon-counting detectors. Compared with an indirect detector, our system has better spatial resolution, significantly higher signal-to-noise ratio and sensitivity. The optimal working resolution range of this system is between ∼0.4 µm and 1 µm, filling the gap between transmission X-ray microscopes and indirect detectors. The application of this system is suggested in the quantitative three-dimensional (spatial) and four-dimensional (space and time) imaging of biological and other weakly absorbing specimens with typical sizes in the range 100–500 µm.
APPENDIX A
Conversion efficiency of the indirect detector
An indirect detector is based on a luminescent screen (scintillator) which converts the X-ray image into a visible-light image. Then, the visible-light image is projected onto a CCD detector through a diffraction-limited magnifying optics. Taking into account the different steps of the image formation, the signal detected in a CCD pixel is given by the following expression,
where Φ is the X-ray per unit of time and area, E is the energy of X-rays (keV), ηabs is the X-ray absorption efficiency of the scintillator, LY is the scintillator light yield (photons keV−1), ηColl is the of the magnifying optics given by (NA/n)2/4, where NA is the numerical aperture and n is the index of refraction, QE is the of the CCD detector and ADU is the analog-to-digital unit of the CCD detector.
Acknowledgements
We would like to acknowledge Andrew Malandain for his great technical support at the Diamond B16 beamline, Daniel Gauder and Julian Schäfer for production of mounting adapters, and David Haas for his outstanding IT support. DK acknowledges the support of the Scientific Grant Agency of the Ministry of Education of Slovak Republic and the Slovak Academy of Sciences No. VEGA-2/0153/10 and by the Grant of Science and Technology Assistance Agency, project No. APVV-0308-11. DP gratefully acknowledges funding from the Australian Research Council and PO acknowledges MPO CR (FR-TI1/421) for financial support.
References
Authier, A. (2004). Dynamical Theory of X-ray Diffraction. Oxford University Press. Google Scholar
Baez, A. V. (1960). Nature (London), 186, 958. CrossRef Web of Science Google Scholar
Barrett, R., Härtwig, J., Morawe, C., Rommeveaux, A. & Snigirev, A. (2010). Synchrotron Radiat. News, 23, 36–42. CrossRef Google Scholar
Boettinger, W. J., Burdette, H. E. & Kuriyama, M. (1979). Rev. Sci. Instrum. 50, 26. CrossRef PubMed Web of Science Google Scholar
Bonse, U. & Busch, F. (1996). Prog. Biophys. Mol. Biol. 65, 133–169. CrossRef CAS PubMed Web of Science Google Scholar
Cecilia, A., Rack, A., Douissard, P., Martin, T., dos Santos Rolo, T., Vagovič, P., Hamann, E., van de Kamp, T., Riedel, A., Fiederle, M. & Baumbach, T. (2011). Nucl. Instrum. Methods Phys. Res. A, 648, S321–S323. Web of Science CrossRef CAS Google Scholar
Fienup, J. R. (1982). Appl. Opt. 21, 2758–2769. CrossRef CAS PubMed Web of Science Google Scholar
Flechsig, U., Jaggi, A., Spielmann, S., Padmore, H. A. & MacDowell, A. A. (2009). Nucl. Instrum. Methods Phys. Res. A, 609, 281–285. Web of Science CrossRef CAS Google Scholar
Huang, X. & Dudley, M. (2003). Acta Cryst. A59, 163–167. Web of Science CrossRef CAS IUCr Journals Google Scholar
Kagoshima, Y., Tsusaka, Y., Yokoyama, K., Takai, K., Takeda, S., Kobayashi, K., Kimura, H., Kimura, S. & Izumi, K. (2000). AIP Conf. Proc. 507, 41–44. CrossRef CAS Google Scholar
Kirkpatrick, P. & Baez, A. V. (1948). J. Opt. Soc. Am. 38, 766–774. CrossRef PubMed CAS Web of Science Google Scholar
Kobayashi, K., Izumi, K., Kimura, H., Kimura, S., Ibuki, T., Yokoyama, Y., Tsusaka, Y., Kagoshima, Y. & Matsui, J. (2001). Appl. Phys. Lett. 78, 132–134. Web of Science CrossRef CAS Google Scholar
Llopart, X., Campbell, M., Dinapoli, R., San Segundo, D. & Pernigotti, E. (2002). IEEE Trans. Nucl. Sci. 49, 2279–2283. Web of Science CrossRef Google Scholar
Miao, J., Charalambous, P., Kirz, J. & Sayre, D. (1999). Nature (London), 400, 342–344. Web of Science CrossRef CAS Google Scholar
Pelliccia, D., Sorrentino, A., Bukreeva, I., Cedola, A., Scarinci, F., Ilie, M., Gerardino, A. M., Fratini, M. & Lagomarsino, S. (2010). Opt. Express, 18, 15998–16004. Web of Science CrossRef PubMed Google Scholar
Sawhney, K., Dolbnya, I., Tiwari, M., Alianelli, L., Scott, S., Preece, G., Pedersen, U. & Walton, R. (2010). AIP Conf. Proc. 1234, 387. Google Scholar
Schäfer, P. & Köhler, R. (2003). J. Phys. D, 36, A113–A117. Google Scholar
Spal, R. D. (2001). Phys. Rev. Lett. 14, 3044–3047. Web of Science CrossRef Google Scholar
Stampanoni, M., Borchert, G. & Abela, R. (2006). Radiat. Phys. Chem. 75, 1956–1961. Web of Science CrossRef CAS Google Scholar
Stampanoni, M., Borchert, G., Abela, R. & Rüegsegger, P. (2002). J. Appl. Phys. 92, 7630–7635. Web of Science CrossRef CAS Google Scholar
Suehiro, S., Miyaji, H. & Hayashi, H. (1991). Nature (London), 352, 385–386. CrossRef PubMed Web of Science Google Scholar
Turecek, D., Holy, T., Jakubek, J., Pospisil, S. & Vykydal, Z. (2011). J. Instrum. 6, C01046. Google Scholar
Vagovič, P., Korytár, D., Mikulík, P., Cecilia, A., Ferrari, C., Yang, Y., Hänschke, D., Hamann, E., Pelliccia, D., Lafford, T. A., Fiederle, M. & Baumbach, T. (2011). J. Synchrotron Rad. 18, 753–760. Web of Science CrossRef IUCr Journals Google Scholar
Vagovič, P., Korytár, D., Mikulík, P. & Ferrari, C. (2007). Nucl. Instrum. Methods Phys. Res. B, 265, 599–604. Google Scholar
Vykydal, Z., Jakubek, J. & Pospisil, S. (2006). Nucl. Instrum. Methods Phys. Res. A, 563, 112–115. Web of Science CrossRef CAS Google Scholar
© International Union of Crystallography. Prior permission is not required to reproduce short quotations, tables and figures from this article, provided the original authors and source are cited. For more information, click here.