research papers
Electron yield soft X-ray photoabsorption spectroscopy under normal ambient-pressure conditions
aJapan Synchrotron Radiation Research Institute/SPring-8, 1-1-1 Kouto, Sayo, Hyogo 679-5198, Japan
*Correspondence e-mail: tamenori@spring8.or.jp
Ambient-pressure soft X-ray photoabsorption spectroscopy (XAS) was demonstrated to be applicable to the chemical analysis of hydrated transition-metal compounds. For this purpose, even under ambient-pressure conditions, electron yield detection 2) and to hydrated (CoCl2·6H2O) cobalt chloride at the Co L23-edge. The present measurements demonstrated the ability to unambiguously distinguish the different chemical states of cobalt ions, relying upon spectral differences that indicate octahedral/quasi-octahedral structural changes as a result of hydration/dehydration reactions.
(EY-XAS), based on a simple drain-current set-up, was used to overcome a weakness in detection (FY-XAS), which does not give a pure soft The feasibility of EY-XAS was investigated and it was clarified that the EY-XAS under ambient-pressure conditions corresponds to the mixed data of the total EY and conversion EY spectra. Normal ambient-pressure EY-XAS analysis was applied to anhydrous (CoClKeywords: soft X-ray photoabsorption spectroscopy; normal ambient pressure; electron yield detection; conversion electron yield detection; hydration.
1. Introduction
X-ray photoabsorption spectroscopy (XAS), which is often called X-ray absorption fine-structure (XAFS) spectroscopy, is one of the best techniques for the determination of the electronic structures of materials (Stöhr, 1992). One of the advantage of analysis is that samples can be analyzed regardless of their physical state (solid, gas or liquid), and it is not restricted to the analysis of certain elements, as are Mössbauer and nuclear magnetic resonance (NMR) analysis. Furthermore, because the core electrons excited by X-ray absorption are localized to the atoms from which they originate, X-ray core-electron transitions allow investigation of the chemical properties of specific elements in complex materials. In particular, analyses using soft X-rays are uniquely appropriate not only for studying light elements but also for studying transition metals (Zaanen et al., 1985; de Groot & Kotani, 2008). Since is governed by dipole selection rules, the d-shell properties of first-row transition metals are best probed by the L23-edge absorption (2p–3d transition) in the soft X-ray region.
However, so far the application of soft X-ray analysis has been highly restricted. Since soft X-rays below 2.0 keV are largely absorbed by air under ambient pressure, it is standard procedure to perform soft X-ray analysis under high vacuum. Although the high-vacuum chamber produces an ideally clean environment for the sample, chemically significant phenomena take place under ambient pressure, such as the coexistence of water and oxygen gas. One of the active areas of modern research in soft X-ray analysis is to improve the versatility of the sample environment. For example, vacuum-compatible sample holders containing a liquid cell have been developed for the analysis of liquid materials (Guo et al., 2004; Liu et al., 2007; Schreck et al., 2011) and for the observation of chemical reactions (de Smit et al., 2008; de Groot et al., 2010; Huse et al., 2010; Kolmakov et al., 2012).
Ideally, it would be convenient for researchers to be able to apply soft X-ray analytical techniques to their samples without using high-vacuum chambers and special vacuum-compatible reaction cells. One solution is to use a helium path because helium gas has a higher transmission than air in the soft X-ray region as a result of its smaller et al. in 1992, several investigations have been reported to obtain reliable spectra under atmospheric pressure conditions using soft X-rays as a probe (Roper et al., 1992; Yagi et al., 2004; Nakanishi et al., 2010). In this scenario, the high vacuum of the light source or synchrotron radiation beam is protected by isolating the analysis chamber using a vacuum window. For example, Nakanishi et al. (2010) have reported their recent successful analysis of hydrated compounds at the Mg K-edge region (1.3 keV) using a thin Be foil (15 µm) as a vacuum window. However, soft X-ray analysis under atmospheric pressure is still limited to energies higher than 1 keV, since photoabsorption by the window reduces the arriving at the sample. To overcome the current limitations, we developed a new approach to analysis using a differential pump instead of a vacuum window (Tamenori, 2010). We specifically focused on the application of lower-energy soft X-rays (below 1 keV) under normal ambient-pressure conditions, the range in which the K-edges of carbon and oxygen and the L23-edges of 3d transition metals are found.
Since the pioneering work by RoperAnother issue to be solved is the development of an et al., 1977; Zschech et al., 1992; Meitzner & Fischer, 2002). Another issue is that FY measurements do not produce pure when multiplet effects are important, because the fluorescence decay probability strongly depends on the final state (de Groot et al., 1994). In contrast to fluorescence decay, the Auger decay probability is approximately constant. Therefore, even under normal ambient-pressure conditions, the application of electron yield (EY) measurements is indispensable to obtaining reliable in the soft X-ray region.
signal detection technique instead of fluorescence detection. (FY) detection, which is the more common signal detection method under high pressure, has two drawbacks. One is the spectral line shape distortion, which is well known as the `self-absorption effect' (JaklevicIn the present research, we demonstrate the capabilities of normal ambient-pressure soft via hydration/dehydration using L23-edge analysis. Such hydrated compounds adopt different crystal structures from those of their dehydrated counterparts and, consequently, generate distinct This technique provides information about the symmetry and of the analyte and allows for the identification of unknown species, establishing it as a powerful tool for of transition metals under normal ambient-pressure conditions.
using EY detection. This technique was used for measurements of the coordination states in hydrated transition-metal compounds. We observed a change in the coordination states of cobalt ions in cobalt(II) chloride2. Experimental
2.1. Soft X-ray source
The normal ambient-pressure soft et al., 1998; Ohashi et al., 2001). were obtained by scanning the undulator gap as well as the monochromator to maintain maximum intensity of the incident soft X-rays, and by scanning the width of the entrance and exit slits to maintain constant The photon-energy-resolving power was set to E/ΔE = 5000. The X-ray beam diameter was focused to a 200 µm spot on the sample. During analysis, the intensity of the incoming photon beam was monitored by measuring the drain current on the surface of a post-focusing mirror.
set-up was installed at the c-branch of the soft X-ray beamline (BL27SU) of SPring-8. Radiation from a figure-8 undulator was dispersed by a soft X-ray monochromator with varied-line-spacing plane gratings, and then introduced into the instrument (Tanaka2.2. Ambient-pressure soft set-up
A schematic diagram of the experimental set-up is shown in Fig. 1. The chamber was connected to the beamline via a four-stage differential pump (Tamenori, 2010). When the sample chamber was filled with helium (1 atm), the pressure at the uppermost stage of the differential pump was maintained at 1.0 × 10−4 Pa. The system provided enough isolation between the sample chamber atmospheric pressure and the high vacuum of the beamline across a short distance, eliminating the need for a vacuum window for the soft X-ray beamline. The loss of in the differential pump section was less than 5% at 800 eV, roughly one-sixth of the loss observed when using a 200 nm Si3N4 vacuum window. The total transmission at 800 eV (at the cobalt L23-edge) measured 50 mm downstream from the last aperture (diameter 0.5 mm) was about 90%. Additionally, there was no steep reduction in photon intensity at the of the elements contained in the vacuum window, providing ideal conditions for analysis. The pressure of the analysis region could be reduced to 1.0 × 100 Pa via evacuation through the aperture of the differential pump, and pure helium gas was installed to produce the normal ambient-pressure conditions.
Evaporation of water from the sample was significantly reduced under normal ambient pressure compared with in a vacuum. However, even under conditions of 1 atm of helium, drying of the sample could not be prevented completely. Therefore, in order to keep the sample hydrated, a helium–water mixed gas was introduced into the 2 sample. Because the hydration/dehydration reaction is fast and is accompanied by a change in color, cobalt chloride is used as a moisture indicator in drying agents (Solomon, 1945). CoCl2·6H2O is deep purple in color, whereas the anhydrous form is sky blue (Grime & Santos, 1934).
sample chamber. During analysis of the hydrated samples, helium gas was passed through a reservoir containing purified water at room temperature, and the mixed gas was introduced through a nozzle with an inner diameter of 6.25 mm. The moisture level of the sample chamber was monitored by the color of the CoCl2.3. measurements
). EY-XAS and FY-XAS were recorded simultaneously for a sequence of incident X-ray energies in the usual fashion used for measurements. Monochromatic light was irradiated at an angle of ∼45° relative to the sample normal. Data analysis to remove the background and qualitatively analyze the was carried out manually. The data were normalized for variations in the primary A linear pre-edge was removed for each spectrum and the data were normalized by the height of the edge jump.
were obtained using both EY and FY detection. EY-XAS were recorded using the same experimental set-up as that for conventional sample drain-current measurements. The aluminium sample holder was biased (−18 V) by two dry batteries (9 V) with respect to the analysis chamber, and the drain current was monitored by an ammeter. The FY-XAS were recorded using a Si photodiode detector (AXUV-100; IRD), which was installed in front of the sample (Fig. 12.4. Sample preparation
In the present research, we chose a colored silica gel ball as a sample, since it is easier to check the color of the sample and to control hydrating/dehydrating conditions than it is using pure CoCl2 powder. Anhydrous colored silica gel (CoCl2/SiO2) was purchased from Hakuyo. The concentration of CoCl2 in the silica gel was 4%. The hydrated sample was prepared by immersing the anhydrous sample in purified water for 1 h. Purified water was obtained using a water purification system (Barinsted EASYpure RF).
As a standard material, high-purity cobalt(II) chloride hexahydrate (CoCl2·6H2O, 99.3% purity) was purchased from MP Biomedicals, and anhydrous cobalt(II) chloride (CoCl2, 99.4% purity) was purchased from Wako Pure Chemical Industries. The standard samples of CoO and LiCoO2 were purchased from Wako Pure Chemical Industries and Kishida Chemical Company, respectively. All of the above materials were used without further purification. Samples were fixed with conductive double-sided carbon tape onto an aluminium sample holder and installed in the chamber.
2.5. Atomic multiplet calculations
Atomic multiple simulations for Co(II) were performed using the CTM4XAS 5.0 program, including full spin–orbit coupling and crystal-field effects (Cowan, 1981; Stavitski & Groot, 2010). The simulation default values were determined by referring to the literature concerning cobalt ions of the same and symmetry, and optimized by repeated simulations (de Groot et al., 1993; Papaefthimiou et al., 2011).
3. Results and discussion
3.1. EY-XAS measurements under normal ambient-pressure conditions
The feasibility of EY detection under normal ambient-pressure conditions was confirmed. Fig. 2 shows the of 2 mm-thick α-Al2O3 obtained using the EY-XAS (solid line) and FY-XAS (dashed line) techniques in the Al K-edge region. The indicated spectra were obtained by background subtraction and then normalized by the edge-jump height. One significant feature of these measurements is that the can be obtained by the EY method, even though α-Al2O3 is an insulating material. FY detection is insensitive to charging of the sample, making it applicable for measuring the of insulating materials, but measurements using a conventional drain-current method under vacuum conditions are inhibited by sample charging.
Here, the question of how the EY is detected by this experimental set-up arises. The EY-XAS in Fig. 2 was obtained with the same set-up as that used in the conventional sample drain-current method. Since a thick full insulator is used as the sample, direct detection of the sample drain-current can be ruled out. Another possibility is that the positive ions are collected by the sample holder. Because the probabilities of ion by soft X-ray absorption are negligible, the production of secondary ions is considered to be an important process. This is similar to well known processes such as conversion EY (CEY) detection in the hard X-ray region (Kordesch & Hoffman, 1984; Tourillon et al., 1987). A conceivable EY detection scheme is as follows. Auger electrons and which are eliminated in the core-hole decay processes, ionize the surrounding helium gas, and secondary ions can be collected by the negatively biased sample holder. The collected ions are neutralized on the surface of the sample holder, and the number of lost electrons is measured as the drain current. In order to confirm the proposed model, the effects on the sample current of positive and negative bias voltages were investigated by changing the polarities of the dry batteries. Reversing the bias voltage reversed the polarity of the signal. This result can be interpreted as the positively biased sample holder collecting eliminated and secondary electrons.
The EY-XAS measurements of insulating samples without sample charging can also be interpreted by considering the CEY detection scheme. Soft X-ray irradiation eliminates electrons from the sample surface, and positive charges will be distributed on the sample surface. Under vacuum conditions, distortion of the spectral shape should occur as a result of the accumulation of positive charges on the sample surface. On the other hand, under low-vacuum or ambient-pressure conditions, the photoelectrons and Auger electrons emitted from the sample ionize the surrounding helium gas, and this produces several et al., 1978).
These electrons are flooded around the sample surface and neutralize the positive surface charges. Similar phenomena are well known in secondary (SEM) measurements as low-vacuum SEM (MoncrieffThe et al., 1977; Zschech et al., 1992; Meitzner & Fischer, 2002). Under ambient-pressure conditions, EY detection produces more reliable than does FY detection. On the other hand, Nakanishi & Ohta (2008) determined that the peak height at the white line is 5.4 by total EY (TEY) measurements using a microchannel-plate detector. The obtained peak height of the EY-XAS (3.08) is lower than those reported in earlier papers. This result suggests that the EY-XAS measurement was also affected by the saturation effect.
obtained using two different detection modes show different spectral profiles. The peak height at the white line (1569.0 eV) is 1.45 in the FY-XAS, whereas in the EY-XAS it is 3.08. The clear peak intensity reduction in the FY-XAS is known to be caused by the `self-absorption effect' (JaklevicZheng et al. (1997) indicated that the probing depth of the CEY method is deeper than that of the TEY method. This phenomenon, caused by the escape depth of electrons in solid materials, strongly depends on the kinetic energies of these electrons. The TEY signal is dominated by low-energy electrons with kinetic energies below about 20 eV, and the mean free paths of these electrons are shorter than 5 nm (Stöhr, 1992). However, these low-energy electrons cannot contribute to the production of helium ions, since the kinetic energies of most are not enough to ionize helium atoms [24.58 eV (Siegbahn et al., 1969)]. In CEY detection, higher-energy electrons such as Auger electrons can contribute to the production of secondary ions, and these electrons are eliminated from much deeper inside the sample. As a result, the obtained by the CEY technique is more sensitive to the than is that obtained by the TEY technique.
3.2. EY-XAS measurement of dehydrated CoCl2 under various pressure conditions
From the EY-XAS measurement of an insulating material, the contribution of the CEY signal to the EY-XAS measurement was confirmed. On the other hand, a conductive sample, which can measure the sample drain-current, should produce identical data to those obtained by TEY detection, even under ambient-pressure conditions. This assumption will be experimentally confirmed in this section.
Figs. 3(a)–3(d) show of dehydrated CoCl2 samples obtained at the Co L23-edge under various pressure conditions. To understand the spectral profiles, the of a standard material (CoO) is also shown in Fig. 3(e). Figs. 3(a) and 3(b) show the EY-XAS for pure CoCl2 powder under a high vacuum (1 × 10−5 Pa) and normal ambient atmospheric pressure conditions, respectively. The L23-edge contain two regions (L2 and L3) arising from spin–orbit splitting of the 2p3/2-1 and 2p1/2-1 core holes. These spectra are almost identical, and the identical obtained under different pressure conditions suggests two facts. The first is that the high pressure caused by the surrounding helium gas does not affect the chemical and electronic properties of the sample molecules. The second is that TEY detection is achievable for conductive samples, even under ambient-pressure conditions, as described above. As a result, the present measurements clarified that the EY-XAS under ambient-pressure conditions corresponds to the mixed data of the TEY and CEY spectra. For conductive materials, the EY-XAS measurements mainly correspond to the data obtained by TEY detection, whereas the contribution of CEY detection becomes significant for insulating materials, and is affected by the saturation effect. However, even for measurements of insulating materials, the obtained EY-XAS was more reliable than the FY-XAS measurements. Therefore, EY-XAS measurements are applicable in measurements under ambient-pressure conditions.
Fig. 3(c) shows the EY-XAS for a colored desiccant under atmospheric pressure conditions. This spectrum is identical to the top two spectra, and suggests that CoCl2 contained in the silica gel ball is identical to the pure CoCl2 reagent. This result indicates that, although silica gel is an insulating material, the saturation effect is negligible in the present EY-XAS because of the low concentration of silica gel. Fig. 3(d) shows the FY-XAS of a colored desiccant under atmospheric pressure conditions, as a typical spectrum obtained by FY detection. The FY-XAS obtained under the top two experimental conditions were also almost identical to the FY-XAS of the colored desiccant (data not shown). The spectral profile obtained by FY detection is clearly different from that obtained by EY detection. The intensity of the L3 band (in the 776–778 eV range) decreased in the FY-XAS. The area intensity ratio of the L2:L3 band in the EY-XAS is 1:3.1, whereas that of FY-XAS is 1:1.8. A similar phenomenon was indicated in earlier reports, and is interpreted as suppression of the L3 band intensity by the saturation effect (Liu et al., 2007). Furthermore, depletion of some specific peaks can be confirmed in the multiplet structures. For example, the first and second peaks at 776.9 and 778.1 eV are clearly suppressed in the FY-XAS. This phenomenon seems to arise because the fluorescence decay probabilities strongly depend on the final states (de Groot et al., 1994). The elimination of fluorescence X-rays is expected to be suppressed on excitation of these final states. The present results also demonstrate that EY-XAS measurements are applicable even under normal ambient-pressure conditions, and produce more reliable data than do FY-XAS measurements.
Fig. 3(e) shows the EY-XAS of CoO as a reference material corresponding to dehydrous CoCl2. The of anhydrous CoCl2 was found to be similar to that of CoO, indicating that the oxidation states of the cobalt ions as well as the symmetries of the CoCl2 crystals are analogous to those of CoO. In CoO, the Co(II) ions are coordinated to oxygen anions with octahedral geometry and Oh symmetry (de Groot et al., 1993). This interpretation is consistent with previous X-ray diffraction investigations of CoCl2 crystals (Grime & Santos, 1934). The chlorine atoms adopt a cubic close-packed arrangement with the cobalt atoms located at the centers of octahedral groups of chlorine atoms.
3.3. Comparison of EY-XAS of unhydrated and hydrated CoCl2
Figs. 4(a) and 4(b) show the EY-XAS for a colored desiccant under normal ambient-pressure conditions. The former is the spectrum obtained for dehydrated CoCl2, which is sky blue, recorded under 1 atm of pure helium. The latter is the spectrum of CoCl2 hydrate, which is deep purple, recorded under 1 atm of a mixed helium/water gas. In order to confirm the sample conditions, photographs of the samples are shown as insets. Of particular interest are the spectral profiles of the hydrated and dehydrated CoCl2 obtained using EY-XAS, which differ substantially. The intensities of the multiplet peaks in the range 777–779 eV decreased in the of CoCl2·6H2O, whereas the peak at 780.5 eV was enhanced. Since both measurements were performed under high-pressure conditions, the possibility that the high pressure changes the electronic structure of the sample can be ruled out. Another interpretation of the spectral change is a change in the valence or molecular structure as a result of hydration. The pronounced differences in the line shapes suggest that the local structures and/or the oxidation states of the cobalt ions are significantly different in the two samples.
The 2·6H2O and standard trivalent cobalt compounds such as LiCoO2 are very similar (Fig. 4c). At first glance, this agreement suggests that the changed from divalent to trivalent. However, this is unlikely because of the instability of trivalent cobalt ions. Although cobalt(III) is stable in the presence of a complexing agent such as ammonia, in an aqueous solution containing no complexing agent cobalt(III) is easily reduced to cobalt(II) (Rajbir, 2002). The impurities contained in the silica gel could act as complexing agents and could affect the of the cobalt during the hydration/dehydration processes. However, this scenario can be ruled out because the change in the spectral profile was replicated by analysis of pure CoCl2 (data not shown). Therefore, we conclude that it is unlikely that water molecules added by hydration oxidized the cobalt ions in CoCl2·6H2O.
of CoClAnother interpretation is that the crystal symmetry is changed by hydration. X-ray diffraction studies suggested the following 2·6H2O (Mizuno et al., 1959; Mizuno, 1960; El Saffar, 1960; Waizumi et al., 1990; Souissi & Kammoun, 2011): two Cl− ions and four water molecules are coordinated to a Co2+ ion in an octahedral geometry to form a CoCl2·4H2O group, whereas the other two water molecules of the formula unit are relatively free, and the groups are joined to one another by hydrogen bonds (O⋯H—O—H⋯O and O—H⋯Cl). The CoCl2·6H2O crystal has a quasi-octahedral geometry, and the complex unit belongs to the point-group D4h symmetry. Based on these X-ray diffraction reports, we can assume that the symmetry of the CoCl2 crystal changed from Oh to D4h as a result of hydration.
for CoClThis hypothesis has been confirmed by ligand-field multiplet simulation (de Groot et al., 1993; Papaefthimiou et al., 2011). Fig. 4(d) shows the simulated spectra, assuming a D4h symmetry with a ligand-field strength of 10Dp/dt/ds = 1.2/−0.5/0.1 eV for the divalent cobalt ion. The simulations faithfully reproduce the main spectral line profiles, whereas the spectrum from the simulation does not reproduce the lower-energy sides of the L2 and L3 bands. The discrepancy can be interpreted as contamination of the dehydrated CoCl2 signal. For example, the small peaks observed in the range 777–779 eV could not be reproduced in the simulated spectrum. However, these peaks correspond to the first three peaks observed in the spectrum of dehydrated CoCl2 (Fig. 4a). Therefore, the discrepancy can be interpreted as partly dried CoCl2 being contained in the colored silica gel.
Furthermore, our explanation is also strengthened by the repeatable nature of the 2 underwent dehydration. Dehydration of the sample can be unambiguously confirmed by the change in the sample color (a color change from purple to blue). When we performed measurements on this dehydrated sample, the spectrum indicated that the sample possessed Oh symmetry. Once the sample was rehydrated by immersion in water (undergoing a color change from blue to purple), subsequent analysis revealed the D4h nature of the analyte. The repeatable change in the is achieved through the hydration and dehydration reactions alone, without adding a special oxidizing or reducing agent. Therefore, although the of CoCl2·6H2O mimics the of standard trivalent materials, we can eliminate the possibility that the spectral changes originate from a change in the of the cobalt ion. We can conclude that the valence of the cobalt ion in CoCl2·6H2O is divalent and the crystal has D4h symmetry.
analysis. After measurement of a hydrated sample, we stopped the helium/water mixed gas flow. This procedure evacuated the sample region and the CoCl4. Conclusions
In the present study, it was demonstrated that EY soft 2O3. EY can be understood to have the mixed data of the total EY and conversion EY spectra. The drawback of EY-XAS under ambient conditions is that the obtained spectrum contains CEY signals. The CEY signals are enhanced on the measurement of insulating materials, and this causes spectral suppression as a result of the Despite this drawback, the obtained spectrum was more reliable than that using FY-XAS measurements, and is applicable in measurements under ambient-pressure conditions.
under normal ambient-pressure conditions can be applied by using the same layout as that with a normal sample drain-current used in conventional soft X-ray analysis. The feasibility of this method was confirmed by measurements of AlEY-XAS analysis was applied to the analysis of cobalt ions in anhydrous CoCl2 crystals and to hydrated CoCl2 crystals under normal ambient-pressure conditions. The EY-XAS results unambiguously distinguished the chemical states of the cobalt ions in each sample. simulation was used to show that the CoCl2 crystal underwent an octahedral/quasi-octahedral structural change by a hydration/dehydration process. We have successfully demonstrated that EY-XAS analysis under ambient pressure, using a helium path, is applicable to chemical state analysis, even for hydrated samples. This represents a significant milestone in the development of analysis using soft X-rays under ambient conditions.
Acknowledgements
This work was supported by the GIGNO project at SPring-8. The author gratefully thanks the members of the GIGNO committee for useful advice concerning this study. This work was financially supported by a Grant-in-Aid for challenging Exploratory Research (22655023). This study was conducted with the approval of the SPring-8 Proposal Review Committee (Proposals 2010A1104, 2010B1198, 2011B1993 and 2012A1847).
References
Cowan, R. D. (1981). The Theory of Atomic Structure and Spectra. Berkeley: University of California Press. Google Scholar
El Saffar, Z. M. (1960). J. Phys. Soc. Jpn, 17, 1334–1337. Google Scholar
Grime, H. & Santos, J. A. (1934). Z. Kristallogr. 88, 136–141. CAS Google Scholar
Groot, F. de, Arrio, M., Sainctavit, P., Cartier, C. & Chen, C. (1994). Solid State Commun. 92, 991–995. Google Scholar
Groot, F. M. de, de Smit, E., van Schooneveld, M. M., Aramburo, L. R. & Weckhuysen, B. M. (2010). ChemPhysChem, 11, 951–962. Web of Science PubMed Google Scholar
Groot, F. M. F. de, Abbate, M., Elp, J. v., Sawatzky, G. A., Ma, Y. J., Chen, C. T. & Sette, F. (1993). J. Phys. Condens. Matter, 5, 2277–2288. CrossRef Google Scholar
Groot, F. M. F. de & Kotani, A. (2008). Core Level Spectroscopy of Solids. Boca Raton: CRC Press. Google Scholar
Guo, J.-H., Augustsson, A., Englund, C.-J. & Nordgren, J. (2004). AIP Conf. Proc. 705, 1066–1069. CrossRef CAS Google Scholar
Huse, N., Kim, T. K., Jamula, L., McCusker, J. K., de Groot, F. M. & Schoenlein, R. W. (2010). J. Am. Chem. Soc. 132, 6809–6816. Web of Science CrossRef CAS PubMed Google Scholar
Jaklevic, J., Kirby, J., Klein, M., Robertson, A., Brown, G. & Eisenberger, P. (1977). Solid State Commun. 23, 679–682. CrossRef CAS Web of Science Google Scholar
Kolmakov, A., Dikin, D. A., Cote, L. J., Huang, J., Abyaneh, M. K., Amati, M., Gregoratti, L., Günther, S. & Kiskinova, M. (2012). Nat. Nanotechnol. 6, 651–657. Web of Science CrossRef Google Scholar
Kordesch, M. & Hoffman, R. (1984). Phys. Rev. B, 29, 491–492. CrossRef CAS Web of Science Google Scholar
Liu, H., Guo, J., Yin, Y., Augustsson, A., Dong, C., Nordgren, J., Chang, C., Alivisatos, P., Thornton, G., Ogletree, D. F., Requejo, F. G., de Groot, F. & Salmeron, M. (2007). Nano Lett. 7, 1919–1922. Web of Science CrossRef CAS Google Scholar
Meitzner, G. & Fischer, D. (2002). Microchem. J. 71, 281–286. Web of Science CrossRef CAS Google Scholar
Mizuno, J. (1960). J. Phys. Soc. Jpn, 15, 1412–1420. CrossRef CAS Web of Science Google Scholar
Mizuno, J., Ukei, K. & Sugawara, T. (1959). J. Phys. Soc. Jpn, 14, 383. CrossRef Web of Science Google Scholar
Moncrieff, D. A., Robinson, V. N. E. & Harris, L. B. (1978). J. Phys. D, 11, 2315–2325. CrossRef CAS Web of Science Google Scholar
Nakanishi, K. & Ohta, T. (2008). Mem. SR Cent. Ritsumeikan Univ. 10, 105. Google Scholar
Nakanishi, K., Yagi, S. & Ohta, T. (2010). AIP Conf. Proc. 1234, 931–934. CrossRef CAS Google Scholar
Ohashi, H., Ishiguro, E., Tamenori, Y., Okumura, H., Hiraya, A., Yoshida, H., Senba, Y., Okada, K., Saito, N., Suzuki, I., Ueda, K., Ibuki, T., Nagaoka, S., Koyano, I. & Ishikawa, T. (2001). Nucl. Instrum. Methods Phys. Res. A, 467, 533–536. Web of Science CrossRef Google Scholar
Papaefthimiou, V., Dintzer, T., Dupuis, V., Tamion, A., Tournus, F., Hillion, A., Teschner, D., Hävecker, M., Knop-Gericke, A., Schlögl, R. & Zafeiratos, S. (2011). ACS Nano, 5, 2182–2190. Web of Science CrossRef CAS PubMed Google Scholar
Rajbir, S. (2002). Inorganic Chemistry. Transition Metals. New Delhi: Mittal. Google Scholar
Roper, M. D., van der Laan, G., Flaherty, J. V. & Padmore, H. A. (1992). Rev. Sci. Instrum. 63, 1482. CrossRef Web of Science Google Scholar
Schreck, S., Gavrila, G., Weniger, C. & Wernet, P. (2011). Rev. Sci. Instrum. 82, 103101. Web of Science CrossRef PubMed Google Scholar
Siegbahn, K., Nordling, C., Johansson, G., Hedman, J., Heden, P. F., Hamrin, K., Gelius, U., Bergmark, T., Werme, L. O., Manne, R. & Baer, Y. (1969). ESCA Applied to Free Molecules. New York: Elsevier. Google Scholar
Smit, E. de, Swart, I., Creemer, J. F., Hoveling, G. H., Gilles, M. K., Tyliszczak, T., Kooyman, P. J., Zandbergen, H. W., Morin, C., Weckhuysen, B. M. & de Groot, F. M. (2008). Nature (London), 456, 222–225. Web of Science PubMed Google Scholar
Solomon, M. E. (1945). Ann. Appl. Biol. 32, 75–85. CrossRef CAS Web of Science Google Scholar
Souissi, H. & Kammoun, S. (2011). Mater. Sci. Appl. 2, 1121–1126. CAS Google Scholar
Stavitski, E. & de Groot, F. M. (2010). Micron, 41, 687–694. Web of Science CrossRef CAS PubMed Google Scholar
Stöhr, J. (1992). NEXAFS Spectroscopy. Berlin: Springer. Google Scholar
Tamenori, Y. (2010). J. Synchrotron Rad. 17, 243–249. Web of Science CrossRef CAS IUCr Journals Google Scholar
Tanaka, T., Maréchal, X.-M., Hara, T., Tanabe, T. & Kitamura, H. (1998). J. Synchrotron Rad. 5, 459–461. Web of Science CrossRef CAS IUCr Journals Google Scholar
Tourillon, G., Dartyge, E., Fontaine, A., Lemonnier, M. & Bartol, F. (1987). Phys. Lett. A, 121, 251–257. CrossRef CAS Web of Science Google Scholar
Waizumi, K., Masuda, H., Ohtaki, H., Tsukamoto, K. & Sunagawa, I. (1990). Bull. Chem. Soc. Jpn, 63, 3426–3433. CrossRef CAS Web of Science Google Scholar
Yagi, S., Matsumura, Y., Soda, K., Hashimoto, E. & Taniguchi, M. (2004). Surf. Interface Anal. 36, 4–1066. Web of Science CrossRef Google Scholar
Zaanen, J., Sawatzky, G., Fink, J., Speier, W. & Fuggle, J. (1985). Phys. Rev. B, 32, 4905–4913. CrossRef CAS Web of Science Google Scholar
Zheng, S., Hayakawa, S. & Gohshi, Y. (1997). J. Electron Spectrosc. Relat. Phenom. 87, 81–89. CrossRef CAS Web of Science Google Scholar
Zschech, E., Tröger, L., Arvanitis, D., Michaelis, H., Grimm, U. & Baberschke, K. (1992). Solid State Commun. 82, 1–5. CrossRef CAS Web of Science Google Scholar
This is an open-access article distributed under the terms of the Creative Commons Attribution (CC-BY) Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.