research papers
Measurement of the X-ray mass attenuation coefficients of silver in the 5–20 keV range
aSchool of Physics, University of Melbourne, Australia, bSchool of Chemistry, University of Melbourne, Australia, and cSchool of Physics, La Trobe University, Australia
*Correspondence e-mail: chantler@unimelb.edu.au
The X-ray mass attenuation coefficients of silver were measured in the energy range 5–20 keV with an accuracy of 0.01–0.2% on a relative scale down to 5.3 keV, and of 0.09–1.22% on an absolute scale to 5.0 keV. This analysis confirms that with careful choice of foil thickness and careful correction for systematics, especially including harmonic contents at lower energies, the X-ray attenuation of high-Z elements can be measured with high accuracy even at low X-ray energies (<6 keV). This is the first high-accuracy measurement of X-ray mass attenuation coefficients of silver in the low energy range, indicating the possibility of obtaining high-accuracy X-ray absorption fine structure down to the L1 edge (3.8 keV) of silver. Comparison of results reported here with an earlier data set optimized for higher energies confirms accuracy to within one standard error of each data set collected and analysed using the principles of the X-ray extended-range technique (XERT). Comparison with theory shows a slow divergence towards lower energies in this region away from absorption edges. The methodology developed can be used for the analysis of compounds and solutions to investigate structural features, bonding and coordination chemistry.
Keywords: high accuracy; XERT; X-ray absorption fine structure; silver.
1. Introduction
The development of theoretical and experimental investigations of the interaction of X-rays with atoms has been a continuing effort (Rehr & Albers, 2000; Joly, 2001; Chantler, 1995, 2009; Pettifer et al., 1999; Machali et al., 1987; Mica et al., 1985; Lytle, 2007). X-ray mass attenuation coefficients [μ/ρ] and form factors describe the interaction of X-rays with atoms and are widely used. Absolute measurement of [μ/ρ] is of great interest for testing of atomic theories and for comparison with existing tabulations of [μ/ρ] in the literature (e.g. FFAST and XCOM tabulations). X-ray mass attenuation coefficients are used in medical science to generate X-ray images and computed tomography (Lee et al., 1991). Measurements of X-ray mass attenuation coefficients are also important for the study of bonding and local atomic structure of materials and molecules (Han et al., 2002; Ignatov et al., 2001), of the density of electronic states (Hossain et al., 2005) and to test predictions of the photoabsorption (Gerward et al., 1979; Gerward, 1989; Karabulut et al., 2005) and (Hopersky et al., 2004; Khoperskii et al., 2005) using bound state electron wavefunctions. Recently, nano-roughness measurement of thin foils has also become possible using X-ray mass attenuation data (Glover et al., 2009). Measurements of X-ray mass attenuation coefficients have, therefore, been a subject of great interest of several research groups around the world.
This work measures [μ/ρ] in the region between the K and L1 absorption edges of silver and tests the XERT (X-ray extended-range technique) for its suitability for high-accuracy measurements of such a high-Z element. We also extend measurements of the absorption of silver to lower energies, down to 5 keV, close to its L edge at 3.8 keV. To do this, we tested a number of foils between 5 µm and 100 µm thick for their suitability for accurate measurements at these lower energies. The thinnest 5 µm foil is well suited for accurate measurements at these low energies and will enable us to extend measurements to the K and L edges using silver foils and silver compounds in solution.
To confirm the high-accuracy in the present measurements, error analysis was performed for experimental systematics relating to dark-current, air-path attenuation, harmonic contamination and energy calibration; statistical reliability is a key question central to the application of et al., 2012).
for structural analysis of elements and compounds (Chantler2. Experimental
The XERT (Chantler, 2009; Chantler et al., 2001; de Jonge et al., 2006) has been used for the high-accuracy measurement of mass attenuation coefficients of various elements, of form factors and X-ray absorption fine structure (XAFS), and has produced some of the most accurate measurements of mass attenuation coefficients (de Jonge et al., 2005; Glover et al., 2008; Rae et al., 2010; Islam et al., 2010). XERT measurements are performed over an extended range of experimental parameters allowing the determination of sources of systematic errors such as scattering and fluorescence (Tran et al., 2004), harmonic content (Barnea & Mohyla, 1974), energy drift (Glover et al., 2010), bandwidth (de Jonge et al., 2004a), and the attenuation by the detectors and by the air path.
Fig. 1 represents a schematic of the experimental set-up at the ANBF, Tsukuba, Japan. In this work, the mass attenuation of silver was measured using an X-ray beam produced by a bending magnet at beamline 20B of the Photon Factory in Tsukuba, Japan. The X-ray beam was monochromated by a double-crystal monochromator using a monolithic silicon 111 crystal located in the first optical enclosure. The monochromated X-rays were passed through a pair of orthogonal adjustable slits for defining the beam (approximately 2 mm × 1.5 mm).
Matched ion chambers were located upstream and downstream from the attenuating silver foils and were run using a flow of nitrogen gas. The beam intensity was continuously monitored. To achieve high precision, each of the measurements was repeated at least ten times. The monochromator was set to produce photons of energy well above 20 keV and then gradually stepped down to lower energies during the attenuation measurements to avoid effects of backlash.
The sample stage was configured so that it could hold three silver samples and translate and rotate them about horizontal and vertical axes perpendicular to the beam. Two daisy wheels containing 15 aluminium foils each around its perimeter were mounted between the sample stage and the two ion chambers. The aluminium foils had thicknesses varying over several orders of magnitude and allowed us to record additional information about thickness dependence of systematic errors.
2.1. The silver foil samples
A total of seven silver foils with nominal thicknesses ranging from 5 µm to 275 µm were used to measure the X-ray mass attenuation coefficients of the silver foils. The foils were supplied by Goodfellow with claimed purity of 99.99%. The masses and areas of the foils were determined from repeated measurements using a microgram balance and an optical comparator, respectively. The average for each foil was used with a respective standard error determined from the variance of the repeated measurements. Of the two 100 µm foils, the foil in the 15–20 keV energy range (Fig. 4) was used as the reference with a determined mass M = 0.6911 g ± 0.000001 (0.0014%) and area A = 6.493 cm2 ± 0.0019 (0.03%). The average integrated column density [ρt]av of the reference foil was found to be [ρt]av = 0.10644 g cm−2 ± 0.00003 (0.03%) from the ratio of the mass M to the area A of the foil. We identify the foils by their nominal thicknesses, but of course only measured values were used in the analysis.
3. Analysis details
3.1. Dark-current analysis
Dark currents were recorded by the ion chambers at each of the energies to correct for its effect on sample attenuations. The ). For the first time, a stepwise fitting procedure was followed to fit the recorded dark currents at two successive energies with respect to the time difference between the two measurements using a covariance matrix. The fitted dark currents were then subtracted from measurements recorded with and without the foil in both upstream and downstream ion chambers. This new approach to dark-current fitting produced greater consistency and lower χ2 values, thereby allowing reduction of uncertainty contributions from this particular source of systematic error.
is a function of time and has been shown to have significant effect on the attenuation measurements (Glover & Chantler, 20073.2. Determination of integrated column density
The accurate determination of the integrated column density [ρt]c along the beam path is crucial for the determination of the X-ray mass attenuation coefficients [μ/ρ]. The present analysis used two different methods, i.e. full-foil mapping (de Jonge et al., 2004b) and `central-region mapping' (Tran et al., 2003b,c; Rae et al., 2010) of the foil, to determine the integrated column density [ρt]c of the reference foil of nominal 100 µm thickness at the (central) point through which the 2 mm × 1.5 mm X-ray beam passed. The integrated column densities determined by the two methods were consistent. Table 1 and the rest of the paper uses the results of the full-foil mapping technique because of the lower uncertainty in the determined [ρt]c.
The attenuations across the area of the foil were measured using an X-ray raster scan with the 2 mm × 1.5 mm beam size. The average [μ/ρ][ρt]av of the attenuations [μ/ρ][ρt]i across the area and the attenuation at the central point used for all attenuation measurements were determined from the obtained attenuation map of the foil. This followed the modelling of the foil and a fitting procedure applied to the experimental data. The mass m and area A of the foil were used to determine the average integrated column density [ρt]av. Finally, the information was used in equation (1) to determine the integrated column density [ρt]c and the corresponding uncertainty of the reference foil,
3.3. Modelling of the foil and removing the effect of the attenuation of the sample holder
To mount each of the silver foils, a Perspex sample holder was used which also attenuated the X-ray beam (Fig. 2). The average attenuation of the foil was obtained following the subtraction of the attenuation of the Perspex holder used for mounting the foil. In order to remove the effect of the attenuation of the holder, a model of the attenuation of the foil and holder was constructed and fitted to the experimental data (Fig. 3). A square-shaped foil was modelled with two parameters (x0, y0) as the central position where x and y describe its axes. The model was then fitted to the experimental data of X-ray attenuations across the entire area of the reference foil. The raster scan mapped the foil completely in both x and y directions. The thickness profile of the modelled foil was considered to have the attenuation of a truncated wedge-shaped foil described by Glover et al. (2010),
where [μ/ρ][ρt]0 represents the attenuation at the central point (x0, y0) of the foil. The gradients of the wedge in the x and y directions are represented by the parameters mx and my.
3.4. Determining the average and central-point attenuations, and uncertainties
The attenuation of the well defined holder was then subtracted from the fitted data to obtain the attenuation of the foil (Fig. 4). For the accurate determination of the average attenuation of the foil, the average of all individual attenuations across the scanned portion of the foil was determined. Details of the procedure are described by Glover et al. (2010). The standard error determined from the measurements across the foil was taken as the uncertainty of the average attenuation [μ/ρ][ρt]av.
3.4.1. The central point attenuation and uncertainty
The area scan of the reference foil (100 µm) was performed with single-point measurements at 17 keV.
The central point attenuation of the reference foil incorporated information of repeated measurements of the central point together with the single-point measurements over the full-foil map,
The central-point attenuation was obtained to be 3.039249 ± 0.000278. The average attenuation [μ/ρ][ρt]av was determined from the average of all measurements across the area map shown in Fig. 4, and found to be 2.993479 ± 0.002679. The obtained average and central-point attenuations were then used to determine the absolute integrated column density [ρt]c of the reference foil along the beam.
3.5. Determining the integrated column densities [ρt]c of the foils relative to the reference foil
A total of seven silver foils were used including the reference foil for the measurements. The full-foil map of the reference foil was used to determine the column densities [ρt] of the other foils using a fitting procedure.
A hysteretic effect occurred due to a non-monotonic energy cycling. Therefore, we fitted the foils before and after the hysteretic loop. Hence a total of 11 foils were fitted. For instance, the 50 µm foil was fitted over two independent ranges applied at energies before and after 16.5 keV. Similarly, the 10 µm and 12 µm foils were divided into multiple thicknesses to correct for the hysteresis effect across repeated energies 16.5 keV and 16 keV. This approach improved consistency between the μ/ρ]wm and the individual measurements [μ/ρ]i of the foils using two aperture sizes.
[The local consistency between the [μ/ρ] of different foils and the result with the full-foil mapping is maintained to minimize χ2 following
where is the Si at energy Ei with the associated uncertainty , and [μ/ρ]wm is the of of all foils and aperture-dependent measurements at that energy.
of the sample3.6. Determining X-ray attenuation
X-ray intensities were recorded using an upstream (u) and a downstream (d) ion chamber with a sample placed in the path of the X-ray beam, without a sample, and shuttering the X-ray beam for dark-current measurements. Intensities without a sample in the path of the X-ray beam (blank measurements) were also recorded to characterize the air-path attenuation for the X-ray beam travelling the path between the ion chambers. X-ray intensities recorded with the sample, without the sample and the
were then used to determine the accurate X-ray attenuation of the sample.Recorded intensities with the sample and without sample were normalized by subtracting the recorded . The Beer–Lambert law was used for measuring the foil attenuation,
at each of the energies and used for determining the X-ray beam intensity for both samples and blanks (as a function of energy). The procedure of dark-current fitting and subtraction is discussed in §3.1where the subscripts s and b refer to the intensities measured with a sample in the path of the beam and without any sample in the path, respectively. The foil attenuations as a function of energy are shown in Fig. 5.
The uncertainty of attenuation measurements was determined from the uncertainty contributions from dark-current measurements and from the intensities with and without a sample. A total of ten repeated measurements were made for each type of measurement (i.e. blank and sample measurements). The standard errors of the repeated measurements were used to determine the final uncertainty of the attenuation. Thus, the total uncertainty of foil attenuations was determined as
where Idc represents the (i.e. recorded intensity when there is no X-ray beam), and the subscripts s and b denote the measurements with a sample in the beam and without a sample in the beam, respectively. The corresponding uncertainties of the foil attenuation are shown in Fig. 6 using the same symbols as in Fig. 5 to represent attenuations for foils of different thicknesses.
3.7. Energy calibration
A key experimental systematic is the energy offset functional caused by the overall hysteresis of the monochromator motor control, which contributes error to the recorded encoder angle of the monochromator (Tantau et al., 2013). This is generically true for all motor or monochromator systems and should be characterized regularly. We used a powder diffractometer to measure directly the energies at which the mass attenuation coefficients of silver were measured, which is much more reliable than encoder settings and currently much more reliable than using an uncalibrated standard foil edge of indeterminate thickness. Two NIST powder standards, Si(640b) and LaB6(660), were used, the most accurate standards described in the literature. Energy measurements were carried out at every 1–2 keV covering the range of the energies where the attenuation measurements were performed. Detailed energy measurements have been reported by Rae et al. (2006). Figs. 7 and 8 show the difference between nominal and calibrated energies and the associated uncertainties of interpolated energies.
A least-squares procedure was performed to fit the differences between the nominal and calibrated energies. For determining the calibrated energy of every attenuation measurement, the best-fit parameters were used. The uncertainties of the interpolated energies were less than 0.62 eV.
4. Effect of harmonic contamination, roughness and their corrections
A 0.01–30% harmonic contamination was observed at lower energies from 8 keV down to 5 keV, which is the highest harmonic contamination observed by our group. This large higher-order harmonic contamination required special treatment for its correction. We investigated two methodologies to determine the contribution of harmonic contamination as harmonic fraction. The `sample method' used the attenuations of the three main silver foils in the regions of harmonic contamination (10 µm, 12 µm and 5 µm), while the `daisy-wheel method' used 14 aluminium foils with different thicknesses as mounted on the daisy wheel for harmonic and scattering signatures. Both methods found similar levels of contamination across the range of energies to within a few standard deviations. However, the `sample method' underestimated the correction, in part because of the sparse information with only three data points per energy. The daisy-wheel method, however, slightly overestimated the harmonic correction, despite a large number of independent measurements (14), in part due to the mismatch between sample and foil attenuation and the different location. Therefore, we investigated the pattern of discrepancies across the range between these estimates in order to minimize the variance. Several patterns were investigated; the optimum fit to the full data set was found using 0.808 times the estimated harmonic fraction from the daisy-wheel method. This approach achieved 0.09–1.22% accuracy in [μ/ρ] across the full range of energies and samples, and is reported in this paper. We note that there is obviously a residual systematic variance associated with this harmonic, which is dominated at the lowest three measured energies where the correction and its uncertainty are greatest; as such, the residual variance correctly estimates the uncertainty of the corrections in those regions.
4.1. Harmonic contamination and correction
When a monochromator is set to select a particular energy of X-rays, higher-order reflections may also contribute harmonic components to the incident X-ray beam (Tran et al., 2003a; Barnea & Mohyla, 1974). These harmonic components were characterized at each of the energies where sample attenuations were measured using a set of aluminium filters. An upstream daisy wheel was used in the experimental set-up to measure the attenuations of 15 aluminium foils mounted around the perimeter of the daisy wheel with thicknesses ranging from 10 µm to 4000 µm. Harmonic components were detected at energies lower than 6.5 keV (Fig. 9).
We used the FFAST tabulated X-ray mass attenuation coefficients of aluminium to characterize and correct for the effect of harmonic components in the X-ray beam. In the presence of harmonic components of energy En, the total transmission probability I/I0 equals the sum of the transmission probabilities for each of the energy components weighted by the relative intensity of each component in the incident beam and the relative for each energy component. If a fundamental energy E1 is contaminated by only one harmonic component of energy En, the measured attenuation is given by
where fn is the fraction of the harmonic component of energy En present in the fundamental energy.
4.1.1. Fitting procedure and correction
With the use of this model, we fitted five parameters (harmonic fraction fn; attenuation for fundamental energy E1; attenuation for the third-order energy E3; a correction; and an offset) to obtain the harmonic effect on the measurements. To obtain the best fit with the lowest , the parameter was constrained to the FFAST tabulated values at higher energies and was free at lower energies, while the critical output was free for all energies. The dark-current correction and offset were consistent with zero. Following this procedure, harmonic contamination was found to be smooth and monotonic at lower energies <7 keV whereas at higher energies no harmonic contamination was detected. At 5 keV, the obtained harmonic fraction was fn = 0.0029 ± 0.00011; the for the fundamental energy was and was found to be 0.0512 at 5 keV. The highest contribution was found at 5 keV (0.29%). The harmonic effect on the attenuations of silver foil was then corrected for making use of the fractional harmonic contribution at each of the energies by reversing equation (6). This subtracts the attenuation for the higher-order Bragg peaks from the attenuation for the fundamental energy, thereby correcting for the harmonic contamination to the fundamental measurement. Harmonic effects up to 15% at energies lower than 7 keV were observed with two foils (10 µm and 12 µm).
This is the highest level of harmonic correction thus far achieved through applying this method.
4.2. Roughness correction
Attenuation of the 10 µm foil was affected by roughness, which was clear from the residual signature of the foil.
The roughness r.m.s. of the 10 µm foil was found to be 1.1 µm for the energy range 7–17 keV, for which the et al., 2008).
of the foil was corrected (Glover4.3. Effects of fluorescence
The interaction between X-rays with energies higher than that of the silver L1 edge and the silver samples can produce fluorescent photons. The present measurements were made in the 5–20 keV energy range which does not contain any but the energies fall in between the L1 and K edges which can produce fluorescent photons due to the X-rays with energies higher than the L1 edge. The fraction of fluorescent photons scattered can be determined from aperture-dependent measurements (Glover et al., 2008; de Jonge et al., 2005). To observe the effect of scattering, each of the scattered intensities was recorded using two different-sized apertures located on the daisy wheels between the samples and ion chambers (Fig. 10).
As expected, away from absorption edges the fluorescence correction was almost negligible (0.025%) (Fig. 10).
4.4. Aperture- and thickness-dependent consistency of the results
At each of the measured energies, multiple values of the X-ray μ/ρ] of silver were obtained by the use of multiple thicknesses, and two different-sized apertures on the daisy wheels allowed aperture-dependent measurements. We tested the consistency of the obtained X-ray mass attenuation coefficients of silver using those multiple measurements at each of the energies. For this consistency test, the percentage variation between the and individual values was investigated (Fig. 11).
[Excellent consistency was achieved within 0.1% from 6 keV to 17 keV, with somewhat poorer consistency found below 6.5 keV and above 17 keV. The measurements below 6.5 keV contributed slightly larger uncertainties with the 10 µm and 12 µm foils, possibly due to incomplete subtraction of harmonic contributions. At the lowest two to three energies measured the influence of several systematics increases the variance and uncertainty.
4.5. Comparison of this work with theories and other experiments
The US National Institute for Standards and Technology (NIST) recommends two theoretical tabulations, FFAST (Chantler, 1995, 2000) and XCOM (Berger et al., 1998; Saloman et al., 1988), for the photoelectric absorption, and Compton and of X-rays are commonly also used in the calculation of the X-ray The measured X-ray mass attenuation coefficients of silver of this work are compared with the FFAST and XCOM tabulated values and found to be within 2% at the higher energies, with discrepancies increasing towards lower energies. The results were also compared with the few other experimental measurements in the literature in this energy range (5–20 keV).
Our results (Figs. 12 and Fig. 13) show excellent agreement with the careful work of Tran et al. (2005) collected under independent conditions, with agreement to within one standard error across the range 16–20 keV, with a standard error of about 0.1%. The results also confirm a discrepancy with the FFAST tabulation of about 2% at the higher energies, which appears to increase towards lower energies. Comparison of the data with XCOM theory is smoother but also reaches 2%. This is not surprising since the theoretical predictions have an accuracy of order 1%.
This investigation calls for further investigations of the L1 and K edges of silver using the XERT which is able to obtain high accuracy in the energy ranges across the L1 and K absorption edges.
5. Results and discussion
5.1. The mass attenuation coefficients of silver
The mass attenuation coefficients of silver metal foils were determined by the XERT approach and presented in Table 2. Six silver foils with different thicknesses were used in the energy range between 5 keV and 20 keV. To obtain the final at a given energy, the of obtained from the different foil thicknesses and aperture-dependent measurements at that energy was used,
where are the measured mass attenuation coefficients obtained from the different thicknesses and are the corresponding errors of the measurements,
where is the weighted average of the , are the corresponding statistical errors in , and N is the number of foils with different thicknesses measured at a given energy.
|
The obtained mass attenuation coefficients using the integrated column densities [ρt]c by central-foil mapping have been found to be consistent with the full-foil mapping to within 0.15%. A summary of uncertainty and error contributions is presented in Table 3.
|
5.2. Derivation of photoelectric absorption, form factor and scattering components
Theoretically tabulated (FFAST and XCOM) scattering cross sections were derived (by interpolation) for the measured experimental energies at which X-ray mass attenuation coefficients were determined. The photoelectric mass et al., 2004). In this analysis, we used averaging the XCOM (Berger et al., 1998; Saloman et al., 1988) and FFAST (Chantler, 1995, 2000) tabulated values. The corresponding uncertainty of was determined from the half of the variation between the two tabulations. The imaginary component of the atomic form factor is directly related to the photoelectric absorption,
was then determined by subtracting the theoretically tabulated scattering cross sections from the measured total mass attenuation coefficients (Tranwhere E is the photon energy in eV, u is the A is the of silver, h is the Planck constant, c is the speed of light and re is the classical electron radius. The corresponding uncertainty in was evaluated using
5.3. Independent verification of potential accuracy of XERT
Comparison of results reported here with an earlier data set optimized for higher energies (Tran et al., 2005) confirms agreement to within one standard error of each data set collected and analysed using the principles of the XERT, in the region over which they overlap. These results (Figs. 12 and 13) show excellent agreement with the careful earlier work (Tran et al., 2005) collected under independent conditions, with agreement to within one standard error across the range 16 keV to 20 keV, with a standard error of about 0.1%.
The earlier work reported accuracies between 0.27% and 0.4% away from the K-edge from measurements at the Advanced Photon Source, Chicago, 1BM, using a silicon 400 monochromating diffracting crystal, 2 mm × 2 mm aperture and N2 in ion chambers. The current data were collected at beamline 20B of the Photon Factory in Tsukuba, Japan. The X-ray beam was monochromated by a double-crystal monochromator using a monolithic silicon 111 crystal located in the first optical enclosure. The monochromated X-rays were passed through a pair of orthogonal adjustable slits for defining the beam (approximately 2 mm × 1.5 mm). Harmonics, collimation, monochromatic and energy range were different in the two experiments. The sample stage location was also different, implying that systematic corrections including scattering were also of a different magnitude. While some of the foils used as samples were the same, the principle foil thicknesses used in the experiment were different. Even when the same foils were used, the points on the foil through which the X-ray beam passed were not the same.
Interestingly, both experiments used powder diffraction to calibrate the energies of the measured attenuations; but whereas the earlier experiment used a single Si 640b standard powder on a scanning six-circle goniometer, the current experiment used the powder in an image-plate environment using BigDiff, the large Australian powder diffractometer. The systematics for each of these differences are independent and the agreement from such different beamlines and arrangements is therefore quite remarkable.
The key marker for the self-consistency of the data after removal of systematic contributions is Fig. 10 of the earlier publication compared with Fig. 11 of this current paper. The consistencies are very similar, and it is not clear which results are superior on that basis. In the tables, the earlier experiment separated the statistical precision from the absolute measurement, and noted that the dominant systematic was the absolute calibration of the foils, with the second most dominant being the statistical precision, especially at the high energies involved.
Conversely, in the current experiment, the precision is generally improved, and the relative accuracy is generally improved; a significant uncertainty arises from the absolute calibration of the primary foil, and a significant uncertainty arises from the variance between measurements of different foils, including the accuracy of the transfer of thickness determination from primary foils to secondary foils (at higher energies). The fact that so many variables were so different and yet the overall technique generated accurate and consistent results in the region of overlap, at the extremes of energy of both data sets, is perhaps the greatest achievement of the X-ray extended range technique.
6. Conclusion
The X-ray mass attenuation coefficients of silver were determined with an accuracy of 0.01–0.2% on a relative scale down to 5.3 keV, and of 0.09–1.22% on an absolute scale to 5.0 keV. This analysis provides the most accurate measured X-ray mass attenuation coefficients of silver in the 5–20 keV energy range. The methodology developed in this analysis will be important for the L and K edges of silver.
investigation of theOne of the key achievements from this analysis is the attainment of high accuracy at lower energies (lower than 10 keV) for such a high-Z element. This analysis indicates that high accuracy is obtainable at lower energies by the use of comparatively thin foils (5 µm) and by the use of dilute solutions of silver compounds if accurate transfer is possible. This work thus indicates that measurements of L-edge of silver (theoretically at 3.8 keV) are possible using XERT by making use of thinner silver foils at lower energies (3–6 keV).
Perhaps just as significant is the independent verification of the accuracy of the earlier work of Tran et al. (2005) to within one standard error, which confirms the potential accuracy of this technique and the portability with different diffracting monochromator crystals, energy ranges and foils.
Acknowledgements
The Australian Research Council (ARC) and the science faculty of the University of Melbourne are acknowledged for funding this work. The authors would like to thank the staff of the Australian National Beamline Facility (ANBF), Tsukuba, Japan, where the experiment was performed for their assistance in this work. As the ANBF is now closed, we dedicate this work to the efforts of the Australian and Japanese scientists who have worked together to make the beamline and collaboration such a success.
References
Barnea, Z. & Mohyla, J. (1974). J. Appl. Cryst. 7, 298–299. CrossRef IUCr Journals Web of Science Google Scholar
Berger, M. J., Hubbell, J. H., Seltzer, S. M., Chang, J., Coursey, J. S., Sukumar, R. & Zucker, D. S. (1998). NIST Stand. Ref. Database, 8, 87–3597. Google Scholar
Chantler, C. T. (1995). J. Phys. Chem. Ref. Data, 24, 71–82. CrossRef CAS Google Scholar
Chantler, C. T. (2000). J. Phys. Chem. Ref. Data, 29, 597–1048. Web of Science CrossRef CAS Google Scholar
Chantler, C. T. (2009). Eur. Phys. J. 169, 147–153. Google Scholar
Chantler, C. T., Rae, N. A., Islam, M. T., Best, S. P., Yeo, J., Smale, L. F., Hester, J., Mohammadi, N. & Wang, F. (2012). J. Synchrotron Rad. 19, 145–158. Web of Science CrossRef CAS IUCr Journals Google Scholar
Chantler, C. T., Tran, C. Q., Paterson, D. & Barnea, Z. (2001). Phys. Lett. A, 286, 338–346. Web of Science CrossRef CAS Google Scholar
Gerward, L. (1989). Acta Cryst. A45, 1–3. CrossRef CAS Web of Science IUCr Journals Google Scholar
Gerward, L., Thuesen, G., Stibius Jensen, M. & Alstrup, I. (1979). Acta Cryst. A35, 852–857. CrossRef CAS IUCr Journals Web of Science Google Scholar
Glover, J. L. & Chantler, C. T. (2007). Meas. Sci. Technol. 18, 2916–2920. Web of Science CrossRef CAS Google Scholar
Glover, J. L., Chantler, C. T., Barnea, Z., Rae, N. A., Creagh, D. C., Paterson, D. & Dhal, B. B. (2008). Phys. Rev. A, 78, 052902. Web of Science CrossRef Google Scholar
Glover, J. L., Chantler, C. T., Barnea, Z., Rae, N. A. & Tran, C. Q. (2010). J. Phys. B, 43, 085001. Web of Science CrossRef Google Scholar
Glover, J. L., Chantler, C. T. & de Jonge, M. D. (2009). Phys. Lett. A, 373, 1177–1180. Web of Science CrossRef CAS Google Scholar
Han, S. W., Stern, E. A., Haskel, D. & Moodenbaugh, A. R. (2002). Phys. Rev. B, 66, 094101. Web of Science CrossRef Google Scholar
Hopersky, A. N., Petrov, I. D., Nadolinsky, A. M., Yavna, V. A. & Koneev, R. V. (2004). J. Phys. B, 37, 3313–3320. Web of Science CrossRef CAS Google Scholar
Hossain, F., Riley, D. & Murch, G. (2005). Phys. Rev. B, 72, 235101. Web of Science CrossRef Google Scholar
Ignatov, A., Yu, Ali, N. & Khalid, S. (2001). Phys. Rev. B, 64, 014413. Web of Science CrossRef Google Scholar
Islam, M. T., Rae, N. A., Glover, J. L., Barnea, Z., de Jonge, M. D., Tran, C. Q., Wang, J. & Chantler, C. T. (2010). Phys. Rev. A, 81, 022903. Web of Science CrossRef Google Scholar
Joly, Y. (2001). J. Phys. B, 63, 125120. Google Scholar
Jonge, M. D. de, Barnea, Z., Tran, C. Q. & Chantler, C. T. (2004a). Phys. Rev. A, 69, 022717. Google Scholar
Jonge, M. D. de, Tran, C. Q., Chantler, C. T. & Barnea, Z. (2006). Opt. Eng. 45, 046501. Google Scholar
Jonge, M. D. de, Tran, C. Q., Chantler, C. T., Barnea, Z., Dhal, B. B., Cookson, D. J., Lee, W. K. & Mashayekhi, A. (2004b). Meas. Sci. Technol. 15, 1811–1822. Web of Science CrossRef Google Scholar
Jonge, M. D. de, Tran, C. Q., Chantler, C. T., Barnea, Z., Dhal, B. B., Cookson, D. J., Lee, W. K. & Mashayekhi, A. (2005). Phys. Rev. A, 71, 032702. Google Scholar
Karabulut, A., Gurol, A., Budak, G. & Ertugrul, M. (2005). Nucl. Instrum. Methods Phys. Res. B, 227, 485–489. Web of Science CrossRef CAS Google Scholar
Khoperskii, A., Nadolinskii, A., Yavna, V. & Koneev, R. (2005). Opt. Spectrosc. 98, 161–165. Web of Science CrossRef CAS Google Scholar
Lee, M. J., Hahn, P. F., Papanicolaou, N., Egglin, T. K., Saini, P. R., Mueller, S. & Simeone, J. F. (1991). Radiology, 179, 415–418. CrossRef CAS PubMed Web of Science Google Scholar
Lytle, F. (2007). Synchrotron Radiat. News, 20, 9. CrossRef Google Scholar
Machali, F., Al-Barakati, G. G., El-Sayed, A. A. & Altaf, W. J. (1987). J. Phys. F, 17, 1279–1284. CrossRef CAS Web of Science Google Scholar
Mica, J. F., Martin, L. J. & Barnea, Z. (1985). J. Phys. C, 18, 5215–5223. Google Scholar
Pettifer, R. F., Borowski, M. & Loeffen, P. W. (1999). J. Synchrotron Rad. 6, 217–219. Web of Science CrossRef CAS IUCr Journals Google Scholar
Rae, N. A., Chantler, C. T., Barnea, Z., de Jonge, M. D., Tran, C. Q. & Hester, J. R. (2010). Phys. Rev. A, 81, 022904. Web of Science CrossRef Google Scholar
Rae, N. A., Chantler, C. T., Tran, C. Q. & Barnea, Z. (2006). Radiat. Phys. Chem. 75, 2063–2066. Web of Science CrossRef CAS Google Scholar
Rehr, J. & Albers, R. (2000). Rev. Mod. Phys. 72, 621–654. Web of Science CrossRef CAS Google Scholar
Saloman, E. B., Hubbell, J. H. & Scofield, J. H. (1988). At. Data Nucl. Data Tables, 38, 1. CrossRef Web of Science Google Scholar
Sandiago, T. K. U. & Gowa, R. (1997). Pramana, 48, 1077. CrossRef Web of Science Google Scholar
Tajuddin, A. A., Chong, C., Shukri, A., Bandyopadhyay, T. & Bradley, A. (1995). Appl. Radiat. Isot. 46, 113–115. CrossRef CAS Web of Science Google Scholar
Tantau, L. J., Islam, M. T., Payne, A. T., Tran, C. Q., Cheah, M. H., Best, S. P. & Chantler, C. T. (2013). Radiat. Phys. Chem. In the press. Google Scholar
Tran, C. Q., Barnea, Z., de Jonge, M. D., Dhal, B. B., Paterson, D., Cookson, D. J. & Chantler, C. T. (2003a). X-ray Spectrom. 32, 69–74. Web of Science CrossRef CAS Google Scholar
Tran, C. Q., Chantler, C. T. & Barnea, Z. (2003b). Phys. Rev. Lett. 90, 257401. Web of Science CrossRef PubMed Google Scholar
Tran, C. Q., Chantler, C. T., Barnea, Z., de Jonge, M. D., Dhal, B. B., Chung, C. T. Y., Paterson, D. & Wang, J. (2005). J. Phys. B, 38, 89–107. Web of Science CrossRef CAS Google Scholar
Tran, C. Q., Chantler, C. T., Barnea, Z., Paterson, D., Cookson, D. J. & Balaic, D. X. (2003c). Phys. Rev. A, 67, 042716. Web of Science CrossRef Google Scholar
Tran, C. Q., de Jonge, M. D., Barnea, Z. & Chantler, C. T. (2004). J. Phys. B, 37, 3163–3176. Web of Science CrossRef CAS Google Scholar
© International Union of Crystallography. Prior permission is not required to reproduce short quotations, tables and figures from this article, provided the original authors and source are cited. For more information, click here.