view article

Figure 2
(a) Sketch of the vacuum-compatible version of the PILATUS detector. The vacuum side (on the right) consists of the ten detector modules (grey) mounted on the downsized module carrier plate (blue). This is attached to the feed-through flange plate that contains the 575 electric connections and the water supply lines. The vacuum chamber (semi-transparent structure) is directly connected to the beamline. The air side (on the left) consists of the standard electronic units of the PILATUS and the water-cooling supply. (b) Sketch of the feed-through flange plate that separates the vacuum and the air side and facilitates the electrical connection of detector modules.