Figure 5
(Left column) Blow-up of the 100k modules in column 1, rows 2 through 5, featuring the typical correlation between one-photon mean and one-photon standard deviation, and the anti-correlation with point spread. (Right column) Blow-up of a small section of pixels which slightly deviates from the trends depicted in the first column where a smaller voltage per photon resulted in a larger variance and point spread, from the 100k module in row 4 and column 3. The high sensitivity of the mathematical calibration was able to detect this deviance, despite all pixels remaining within 2% of the average detector parameters.