Figure 8
Count rate characteristic (top) and deviation from simulation (bottom) for mid gain settings, an energy threshold of half the X-ray energy, and disabled retrigger mode. In the upper plot the points show the experimental data while the lines are the results of the Monte Carlo simulation. |