research papers
Concept of a multichannel spin-resolving electron analyzer based on Mott scattering
aSwiss Light Source, Paul Scherrer Institute, CH-5232 Villigen-PSI, Switzerland, bSt Petersburg Polytechnical University, Polytechnicheskaya Str. 29, St Petersburg RU-195251, Russian Federation, and cInstitut de Physique de la Matière Condensée, École Polytechnique Fédérale de Lausanne, CH-1015 Lausanne, Switzerland
*Correspondence e-mail: vladimir.strocov@psi.ch
The concept of a multichannel electron spin detector based on optical imaging principles and Mott scattering (iMott) is presented. A multichannel electron image produced by a standard angle-resolving (photo) electron analyzer or microscope is re-imaged by an electrostatic lens at an accelerating voltage of 40 kV onto the Au target. Quasi-elastic electrons bearing spin asymmetry of the Mott scattering are imaged by magnetic lenses onto position-sensitive electron CCDs whose differential signals yield the multichannel spin asymmetry image. Fundamental advantages of this concept include acceptance of inherently divergent electron sources from the electron analyzer or microscope focal plane as well as small aberrations achieved by virtue of high accelerating voltages, as demonstrated by extensive ray-tracing analysis. The efficiency gain compared with the single-channel Mott detector can be a factor of more than 104 which opens new prospects of spin-resolved spectroscopies in application not only to standard bulk and surface systems (Rashba effect, topological insulators, etc.) but also to buried heterostructures. The simultaneous spin detection combined with fast CCD readout enables efficient use of the iMott detectors at X-ray free-electron laser facilities.
Keywords: electron spin; Mott scattering; electron optics; angle-resolved photoelectron spectroscopy and microscopy; X-ray free-electron lasers.
1. Introduction
The spin of the electron plays a crucial role in many physical phenomena, ranging from the obvious example of magnetism, via novel materials for spintronics applications, to high-temperature superconductivity. The direct detection of the spin has therefore played an important role in the understanding of processes such as giant magneto-resistance (GMR) (Baibich et al., 1988; Bonell et al., 2012), the formation of ferromagnetic domains (Kirschner et al., 1984), the Rashba effect (Slomski et al., 2013; Landolt et al., 2012) and topological insulators (Hasan & Kane, 2010; Landolt et al., 2014). Starting from the classical experiments of Schönhense (1980), it was realised that any angle-resolved (ARPES) experiment yields in fact spin-polarized intensities. The most direct access to the electronic structure of crystalline solids resolved in electron spin and momentum is therefore delivered by spin-resolved ARPES (SARPES); for a recent review of the field see Kirschner (1985), Dil (2009) and Okuda & Kimura (2013). In the aforementioned examples the focus was on the detection of the spin of the valence and core level states, but it has also been realised that the spin polarization as induced by the excitation process can yield a rich variety of information (Heinzmann & Dil, 2012). Such measurements typically require the possibility to vary the and/or energy, but then provide a rich variety of information such as above the Curie temperature (Sinkovic et al., 1997; Müller et al., 2001), the distinction between singlet and triplet spin states (Tjeng et al., 1997), and the study of unconventional superconductors (Veenstra et al., 2014). The appeal of these possibilities has led to a revival in the interest of spin detection in combination with spectroscopic techniques and the construction of new experimental setups primarily using synchrotron radiation sources with their high and tunable photon energy hν.
On the technical side, the general evolution trend of the electron as well as X-ray spectroscopic instrumentation can be characterized by moving from the use of slits for selecting a certain energy channel to the use of dispersion and imaging principles for multichannel acquisition in an extended region of phase space. For ARPES, in particular, the first major hallmark of this development vector has been the use of the dispersing and focusing properties of the hemispherical capacitor to produce a multichannel one-dimensional image of electron intensity IARPES over a certain region of kinetic energies Ek. The second hallmark has been the invention of electrostatic lenses which could image the emission angles θ onto the entrance slit of the hemispherical capacitor to produce in its exit plane a multichannel two-dimensional image of IARPES(Ek, θ) over a certain region of Ek and θ (Mårtensson et al., 1994; Wannberg, 2009). Finally, the angle-resolved time-of-flight (ARTOF) analyzers (Öhrwall et al., 2011) allow multichannel acquisition of three-dimensional images of IARPES(Ek, θx, θy) as a function of Ek and two orthogonal emission angles θx and θy. Similar development in resonant inelastic X-ray scattering (RIXS) has been marked by the spectrometers based on the Rowland-circle (Nordgren et al., 1989) and variable-line-spacing (Ghiringhelli et al., 2006; Strocov et al., 2011) spherical gratings whose combined dispersion and focusing actions produced a multichannel one-dimensional image of scattered IRIXS over a certain region of scattered energies hνout. Finally, the concept of an hν2-spectrometer (Strocov, 2010; Warwick et al., 2014) has been advanced where simultaneous imaging action in the orthogonal plane produces a full multichannel two-dimensional image IRIXS(hνin, hνout) as a function of incoming hνin and scattered hνout energies. The overall efficiency gain due to the multichannel detection in the two-dimensional case compared with the single-channel detection can measure for various instruments up to four or more orders of magnitude.
Although new exciting schemes of more efficient spin detection have recently been established (Jozwiak et al., 2010) the main technical difficulty of SARPES remains a dramatic intensity loss. This is characterized by the figure of merit (FOM) F = S2(I/I0) where I/I0 is the scattered-to-incident intensity ratio, and S is the asymmetry (also called Sherman) function defined as the measured intensity asymmetry A = I+-I-/I++I- between two opposite polarization directions divided by the electron polarization P. With small FOM values of less than 10−2 for one spin projection, high-quality spin-resolved measurements are primarily possible only at synchrotron radiation facilities or using high-power lasers, and then still only within a limited range of experimental parameters and/or using measurement times of many hours or even days. This has inhibited the expansion of spin detection to other techniques which may be inefficient in itself because of a lower or because they rely on detection of the time structure, or because they sample only a small amount of matter. Obviously, SARPES and other spin-resolving techniques appear in most severe need of the multichannel detection. Surprisingly, only most recently have such instruments come into play, starting from photoemission microscopes with their natural multichannel detection. The spin resolution was achieved here using spin-polarized reflectivity of a collimated low-energy (below 100 eV) electron beam from W(100) or (Au-passivated) Ir(100) crystals working as imaging spin filters (Tusche et al., 2011; Kutnyakhov et al., 2013; Vasilyev et al., 2015; Kutnyakhov et al., 2015). Furthermore, installation of such a spin filter behind the standard hemispherical analyzer (HSA) has for the first time enabled spin-resolved energy- and angle-multichannel ARPES measurements (Kolbe et al., 2011) although suffering from the inherent divergence of electron trajectories in HSA (see the discussion later).
The Mott spin detectors (for entries see Petrov et al., 1997, 2007; Hoesch et al., 2002) are based on quasielastic scattering of high-energy electrons (about 40 keV) from a target of a high-Z material such as Au. As a spin selective process, the Mott scattering is characterized by a FOM value of about 6 × 10−4 per one spin projection (Petrov et al., 1997, 2007) measured by two opposite electron detectors working simultaneously. Other spin selective processes such as the low-energy electron reflectivity from W(100) or Ir(100) (Tusche et al., 2011; Kutnyakhov et al., 2013; Vasilyev et al., 2015; Kolbe et al., 2011) or exchange scattering from the Fe2O3 surface (Okuda et al., 2008, 2011) may certainly have larger FOM values, in the latter case reaching 9.5 × 10−3 per one spin projection (Okuda et al., 2008) acquired in two successive measurements under re-magnetization of the target or sample. However, the Mott detectors, brought to equal efficiency in all channels by proper calibration (Petrov et al., 1997, 2007), allow simultaneous measurements of two spin projections, which doubles their effective FOM to an already comparable value of 1.2 × 10−3.
From a practical point of view, the FOM analysis alone ignores other aspects of the spin-resolved experiment. First, the simultaneous differential measurements render the Mott detectors immune to the statistical fluctuations brought, for example, by those of the excitation source such as the synchrotron beam or ripple of voltages on the electron optics (Petrov et al., 2001). Furthermore, the long-term stability and reliability of the Mott detector are just as important for obtaining reliable data because of potential degradation of the sample during the acquisition time and drifts in the electron optics. This can be illustrated, for example, by recent studies of the topological Kondo insulator SmB6 where the spin-polarized have a low spectral intensity and are located in a 50 meV band gap. Although attempts were made using a detector with a higher FOM (Suga et al., 2014) only the Mott detectors provided enough long-term stability to explore this spin texture (Xu et al., 2014). The reliability of the Mott detector is expressed, in particular, in the fact that the high energies make it insensitive to the surface quality of the target, and the detector will function in the same way every time it is switched on and for a long period of time. This is critical at large-scale research facilities such as synchrotrons. A further major advantage of the Mott detectors lies in the fact that the spin contrast is obtained without repeating the measurement under different conditions such as flipping the spin of the incident electron beam in electron excitation experiments or re-magnetization of the target or sample. This simplifies studies of non-magnetic samples and, most important, enables experiments under conditions where an exact repetition is fundamentally impossible, such as studies on quickly degrading samples or using free-electron laser (FEL) pulses. We note that the above practical advantages are relevant only to the classical-type Mott detector where the electrons scattered off the target move in field-free space. In the retarding- or Rice-type Mott detectors the scattered electrons move in a retarding potential (for comparative tests of various Mott detectors, see Petrov et al., 2001). Such detectors are actually highly sensitive to parameters of the incident beam, which results in unstable measurements of the polarization asymmetries.
The most important advantage of the Mott detectors appears, however, in the multichannel detection perspective. By virtue of high electron energies, the Liouville–Helmholtz theorem works in these detectors completely in advantage of the electron optics, which delivers then minimal aberrations at large angle and energy acceptance, as will be explained below.
Here, we present the concept of an angle- and energy-multiplexing spin-resolving electron analyzer which combines electron optics based on the imaging principles with the detector based on the Mott scattering. This `imaging Mott' detector will be nicknamed `iMott'.
2. Operation principle
We will sketch the iMott concept as tailored to the standard angle-resolving HSA. Fig. 1(a) shows the iMott detector attached to the HSA. The latter creates in its exit (focal) plane an image of electrons dispersed in the X and Y coordinates corresponding to Ek and θ. Fig. 1(b) shows a blow-up of the iMott itself. The electrostatic lens (EL) operating at high voltage accelerates the electrons from the HSA focal plane and images them onto the polycrystalline Au target to create there a demagnified image again stretching along the Ek and θ directions. The electrons quasi-elastically scattered from the Au target and bearing spin asymmetry of the Mott scattering are then imaged by four magnetic lenses (MLs) (chosen instead of electrostatic ones because of smaller aberrations with extended sources) onto the energy-selective and position-sensitive detectors [in our case implemented as electron-sensitive CCD (eCCD) detectors] to create there images of scattered intensity stretching along the Ek and θ directions (distortions of these images by the magnetic field are corrected on the post-processing stage). The differential images between the opposite eCCDs immediately yield a multichannel image of the A(E, θ) spin asymmetry in the E and θ coordinates. In addition, another two MLs and eCCDs installed in the orthogonal will simultaneously measure the orthogonal spin component. As we will see below, the overall efficiency gain granted by our multichannel concept can be well above 104.
3. Ray-tracing analysis
We will now illustrate the feasibility of the iMott concept with ray-tracing simulations. We restricted ourselves to a compact design of the instrument with a radius of the Mott hemisphere RM = 100 mm, allowing its seamless use with virtually any experimental setup. Without restrictions on generality, we will tailor our analysis to the spatial and angular characteristics of the electron trajectories in the commercial analyzer PHOIBOS-150 having a hemisphere radius of 150 mm, which operates in the medium angle acceptance mode with an entrance slit of 200 µm and pass energy Ep of 40 eV.1 The energy resolution ΔE and angle resolution Δθ are in this case 36 meV and 0.07° FWHM, respectively, with the latter achieved with the source size below 100 µm typical of the synchrotron radiation sources. For the simulations we select a beam of incident electrons comprising five values of Ek stretching within ±2 eV around 400 eV and five values of θ stretching within ±9.5°. The electron image created in this case by the HSA in its exit (focal) plane (ignoring for brevity the angular distortions caused by the hemisphere and aberrations in the HSA lens (Mårtensson et al., 1994; Wannberg, 2009) is shown in Fig. 2(a). Note that for better visibility the broadenings in this image are artificially inflated. The imaging properties of the HSA ensure that the X coordinate in this image is a linear function of E, and Y a linear function of θ. It is important to take into account that the electron trajectories come into each point of the HSA exit plane focused with an inherent angular divergence in both E and θ directions, in our case 1.32° and 0.93° FWHM, respectively. These divergent rays form the source for the subsequent iMott optics.
The electrostatic lens EL, following the HSA, is shown in Figs. 3(a) and 3(b). Operating at a high accelerating voltage of 40 kV, the lens adopts a cascade design with increasing diameter of the cylindric electrodes. We note the absence of any apertures in the lens. The last electrode is connected to the Mott hemisphere forming a field-free region around the target. In our ray-tracing analysis, first we investigated the imaging properties of this lens using an ideal 25-point source simulating the image produced by the HSA (Fig. 2a). Here the spots were replaced by points and the angular divergence, as the most critical test, was increased to 4° FWHM along both directions. Figs. 3(a) and 3(b) show the electron trajectories originating from the central and two ±2 eV and ±9.5° end points of the source in the EL along the two axial cross-sections in the E and θ directions, respectively. We note that in each point of the HSA focal plane the central trajectories are normal to this plane in the E direction, but inclined in the θ direction with the angle relative to the normal progressively increasing with θ, in our case to ±7° at the limits of the θ range. The image transferred from the above point source to the Au target is shown in Fig. 3(c) (note the artificially inflated broadening). The lens demagnifies the image by a factor of about three in order to reduce aberrations in the subsequent MLs. The focusing voltages at the lens are optimized for the points slightly away from the center. The image shows practically no distortions and is characterized by average ΔE and Δθ broadenings of 3.8 meV and 0.036° FWHM, respectively, which can be considered negligible compared with the HSA resolution figures even in our most critical case. Indeed, Fig. 3(c) shows the image transferred from the real broadened source at the HSA focal plane, Fig. 2(a). Evidently, the EL introduces practically no broadening or aberrations. As we discuss below, such extraordinary imaging properties are achieved by virtue of the high accelerating voltage. Finally, our electron optics allows efficient spin-integrated measurements by mechanical in situ exchange of the Au target with a direct-view eCCD. As discussed below, this electron registration method much supersedes the MCP/phosphor screen/CCD stack conventionally used in ARPES analyzers (Mårtensson et al., 1994).
The image produced by electrons quasi-elastically scattered from the Au target is transferred to the eCCDs by the magnetic lens ML1 in one E direction and ML2 in the orthogonal along the θ direction, Fig. 1(b). The lenses are installed at an angle of 60° relative to the incident electron beam. The use of magnetic lenses delivered in our case better performance compared with their electrostatic counterparts because their focal source and image planes are flatter. This allowed better matching of the flat extended source at the Au target to the images at the eCCDs, dramatically reducing aberrations away from the central ray. Technically, the smaller size of magnetic lenses compared with electrostatic ones operating at large accelerating voltages allows compact design of our MLs restricted by the Mott hemisphere, Fig. 3. The latter is made of µ-metal, which protects the HSA from leakage of magnetic fields from the lenses. The MLs and eCCDs are biased with 40 kV, equating their potential to that of the Au target. Power to each ML coil is provided by a separate `floating point' power supply. Data transmission from the eCCDs is organized with opto-cables. The MLs are electromagnetic which allows tuning their focusing properties. Their solenoids are embraced in UHV-compatible jackets allowing their air cooling through insulating plastic hoses. An aperture with a diameter of 12 mm in front of the lenses restricts the acceptance area on the target, and an iris with a diameter of 4.5 mm in the middle of the lenses restricts their angular acceptance from each point on the target to ±5°. We note that the latter, with almost isotropic distribution of the quasi-elastically scattered electrons, reduces the geometrical lens acceptance by a factor of 6.25 in comparison with the one-channel Mott detector whose acceptance may reach ±15° (Petrov et al., 2007). Taking into account the angular dependence of the Sherman function having a lobe near the 120° scattering angle (Holzwarth & Meister, 1964), we obtain an effective reduction of the electron optics transmission by a factor of about 5.5.
along theAgain, we started our ray-tracing simulations with the imaging properties of the MLs using the ideal 25-point source, where the points resembled the image produced at the Au target, Fig. 2(b). The electron trajectories in ML1 originating from the central point of the source are shown in Fig. 4 for the axial of the lens along the E direction (a) and along the θ direction (b). For clarity, the angular spread of electrons in the simulations was restricted by the E direction. Importantly, although the angular spread of electrons was one-dimensional in our simulations, the resulting trajectories become three-dimensional. This is caused by the Lorentz force in the magnetic field which curls the electron trajectories. Furthermore, for the axial of ML1 along the E direction, the trajectories originating from the ±2 eV end points are shown in Fig. 4(c). The best matching to the lens focal plane is achieved by inclination of the eCCDs by ∼45°. The trajectories in ML2 in the orthogonal are identical, except that the E and θ directions are swapped. Images of the above 25-point point source transferred by ML1 onto eCCD1 and by ML2 onto eCCD2 are shown in Figs. 4(d) and 4(e), respectively. As the magnetic field curls the electron trajectories, the rectangular pattern of the source is rotated by ∼30°. The square-like versus stripe-like appearance of the eCCD1 and eCCD2 images expresses the fact that the view angles of ML1 and ML2 onto the Au target are inclined in two orthogonal planes along the E and θ directions, respectively. We note that the aberrations almost vanish near the central ray but significantly scale up away from the center, which is typical of electron optics working with extended sources. The rotation and distortion of the image breaks the linear and independent relations of the X-coordinate to E and the Y-coordinate to θ. Their relation can nevertheless be described using an n-order polynomial morphing transformation defined by the equations E = and θ = , with the coefficients Aij and Bij determined by linear least-squares fitting. The images in Figs. 4(d) and 4(e) corrected using the n = 3 transformation are shown in Figs. 4(f) and 4(g), respectively.
Finally, we have performed ray-tracing simulations with the real broadened source, Fig. 2(b). The images analogous to those in Figs. 4(c) and 4(d), rotated and distorted in the spatial coordinates, were transformed into the physical E and θ coordinates using the above morphing transformation. The resulting images are shown in Figs. 2(c) and 2(d) for eCCD1 and eCCD2, respectively. The transformation has fully recovered the rectangular pattern of the source in the E and θ coordinates. With negligible contribution of the EL, the MLs introduce into these images aberrations which scale up away from the central ray and can be characterized by average 〈ΔE〉 and 〈Δθ〉 broadenings of 50 meV and 0.2° FWHM for eCCD1, and 30 meV and 0.63° for eCCD2. These figures are certainly significant compared with the resolution figures of the HSA itself, but can be considered acceptable in view of the efficiency gain delivered by the multichannel detection. Moreover, our ray-tracing analysis suggests that the iMott resolutions can be improved by reduction of the image size on the Au foil by increasing the EL demagnification. The optimal demagnification, which will balance the aberrations of all lenses and pixel size of the eCCDs, has yet to be determined with the real instrument. At the time of writing, the iMott detector described above is under construction for the soft-X-ray ARPES facility (Strocov et al., 2014a) at the ADRESS beamline of the Swiss Light Source.
4. Properties and advantages of the iMott concept
We start the discussion of the iMott properties with the exact definition of efficiency gain delivered by the multichannel detection. For the energy- and angle-resolving analyzer (with obvious generalization for the microscope) this figure should be defined as G = , where the first term is the ratio of the intercepted area SE,θ in the (E, θ) coordinates to the average resolutions product 〈ΔE〉〈Δθ〉 representing the single-channel detection, and the second term is the ratio of the (aperture-limited) multichannel electron optics transmission T to the single-channel one T0. In our case of the compact iMott, the above resolutions figures and aperture-limited angular acceptance of the MLs yield G = 8.4 × 102 which is almost three orders of magnitude.
Our ray-tracing analysis shows that the aberrations dramatically increase away from the central ray, an inherent property of electron optics working with an extended source. An obvious way to further improve ΔE and Δθ will be to scale up the dimensions of the magnetic lenses, which will flatten the source and image focal planes and therefore reduce the aberrations. As we mentioned above, the present implementation of the iMott concept was restricted by RM = 100 mm fitted to the PHOIBOS-150 analyzer. Larger analyzers such as PHOIBOS-225 or microscopes allow scaling up of the iMott size. Our preliminary ray-tracing analysis indicates that in this case, most conservatively, the ΔE and Δθ resolutions improve and the ML aperture increase linearly, the latter meaning a quadratic increase of the ML solid angle acceptance. Correspondingly, only doubling of the iMott dimensions increases the multichannel efficiency to a colossal value of G = 1.4 × 104.
The iMott scheme has several conceptual advantages compared with the previous multichannel spin detection schemes (Tusche et al., 2011; Kolbe et al., 2011):
(1) The imaging principles of the iMott electron optics imply the transfer from one image plane to another of inherently divergent electron beams, in contrast to the spin-filter ARPES analyzer (Kolbe et al., 2011) which relies on a trade-off between collimation and focusing of the electron beams. This relieves the iMott from any need to reduce the normal beam divergence from the HSA [in particular, the divergence in the E direction resulting in the so-called α-factor in the HSA energy resolution (Wannberg, 2009) concomitantly reducing the electron optics transmission, or introduce any apertures to restrict the (E, θ) area delivered by the HSA]. The resulting high transmission of the electron optics compensates the Mott scattering FOM being much smaller than that of the low-energy spin filters (Tusche et al., 2011; Kolbe et al., 2011).
(2) The iMott concept positively utilizes the Liouville–Helmholtz theorem which states that the product is constant, where Δx and Δα are the spatial and angular broadening of the electron beam, respectively, and V is its energy. Simultaneously reducing Δx and Δα, the high accelerating voltages improve the focusing and thus ΔEk and Δθ without compromising the electron optics transmission.
(3) The intercepted Ek bandpass is limited only by the HSA rather than by the working regions of the W, Ir or Fe2O3 spin filters with an energy width varying from several eV to less than 1 eV as determined by peaks of their FOM energy dependence (Kutnyakhov et al., 2013). Furthermore, the iMott's Sherman function is practically identical for all the energies and angles which are simultaneously detected.
There are also advantages of the iMott concept on the technical and practical sides:
(1) The measurements do not require re-magnetization of the sample or detector, or changing the energy window to measure the A(Ek, θ) asymmetry inherently utilized in the spin-filter ARPES instruments. Instead, the asymmetry is derived from simultaneous measurements on two eCCDs. Furthermore, the pair of eCCDs installed in the orthogonal simultaneously deliver the second spin component.
(2) Scattering of high-energy electrons is almost insensitive to the surface conditions of the Au target. This relieves laborious surface preparation procedures and surface degradation problems typical of the LEED or spin-filter based detectors.
5. Position-sensitive detectors
Finally, we comment on the choice of position-sensitive detectors for iMott. One of the most important technical advantages of the iMott concept is that the high electron energies enable implementation of the position-sensitive detectors as directly irradiated eCCDs. Their 2 which feature an active area of 12 mm × 12 mm with a matrix of 512 × 512 effective pixels having a size of 24 µm × 24 µm. This delivers well sufficient spatial resolution compared with the focused spot size delivered by the iMott optics.
with electron energies above 20 keV is nearly 100%. These devices have numerous advantages compared with the MCP/phosphor screen/CCD stacks conventionally used in ARPES for multichannel detection. Essential for the iMott concept, the eCCDs are energy selective with a bandwidth of about 200 eV which is achieved with an amplitude discriminator at their output. This allows selection of the quasi-elastically scattered electrons and rejection of the inelastically scattered ones, which carry less spin information and are focused by MLs away from the nominal imaging plane. Furthermore, the eCCDs benefit from their simplicity, larger and possibility to measure absolute electron counts. The iMott detector can use standard, not even high-end, eCCDs such as back-thinned ones from HamamatsuThe maximal readout frequency of high-end eCCDs can nowadays top up 10 MHz. This ensures the absence of any charge saturation and, moreover, enables measurements in single-pulse-counting mode. In this case fast online data processing to calculate the center of gravity of each event allows an increase of the effective spatial resolution.
6. Potential applications
In the above, we have primarily focused on the use of the iMott combined with the angle- and energy-resolving HSA. This configuration will be especially useful for SARPES which has so far been one of the main applications of the single-channel Mott detectors. The advantage compared with the existing experimental setups is that a spin-resolved band map will be directly obtained and, depending on the requirements, the data can be binned afterwards to enhance statistics for small signals. The colossal achieved efficiency gain will also allow for SARPES in the soft X-ray energy range which will extend from spin phenomena in the bulk such as the bulk Rashba states (Landolt et al., 2012) to previously unthinkable applications to buried heterostructures, interfaces (Cancellieri et al., 2013) and impurities (Kobayashi et al., 2014) [see Strocov et al. (2014b) for a recent review]. Furthermore, SARPES can now be extended to less intense VUV and X-ray sources including bending-magnet beamlines and laboratory sources.
Applications of the iMott spin detector are certainly not limited to its combination with the HSA. Its unique capabilities can be used in combination with any instrument which produces a two-dimensional image of electrons at its exit plane, with energies up to several keV. As described above, the high energies used in the iMott create ideal focusing conditions, and the angular divergence of the electrons delivered by the instrument it is combined with is not crucial. At most, only the first element of the electrostatic lens would have to be adapted. Based on these characteristics, we can envision the following (and not limited to) applications of the iMott. First, it can be placed behind time-of-flight analyzers such as ARTOF (Öhrwall et al., 2011) to allow for direct spin-resolved Fermi surface mapping. Although the angular variance of electron trajectories in the MLs introduces some time-of-flight uncertainties, high electron energies render their contribution negligible. However, for this type of analyzer the eCCDs should be replaced by faster position-sensitive detectors (for example, delay-line detectors) having a readout speed better than 100 MHz to ensure sufficient energy resolution. Second, the iMott can be combined with a photoemission electron microscope (PEEM) for imaging of magnetic domains even with unpolarized light. This could open the possibility of laboratory-based measurements to complement the current synchrotron-based experiments. Furthermore, one could use the PEEM in combination with circular polarized light for spatial resolved chiral imaging of molecules (Göhler et al., 2011). Third, combined with a low-energy electron microscope (LEEM) the iMott can be used for magnetic imaging, or for the direct visualization of spin transfer torque through a thin layer when combined with a polarized source. Even more ambitious would be the first direct visualization of spatial entanglement in a solid.
In connection with the completion and planned construction worldwide of several X-ray FEL facilities, delivering short and very intense X-ray pulses, we note that in contrast to other spin detection schemes the iMott is perfectly suited for time-resolved studies, especially under the influence of a so-called jitter of the FEL pulses. This virtue comes because the spin-asymmetry is obtained in a simultaneous measurement which does not have to be repeated with different FEL pulses and different sample conditions. For each pulse, the complete data acquired at every eCCD can be stored and later analysed accordingly, as is now common practice for spin-integrated measurements at such facilities (Hellmann et al., 2012), in order to recover the whole time evolution picture. The readout frequency of the eCCDs up to 10 MHz enables seamless handling of each pulse even at the highest repetition rates, which is presently 27 kHz at the European XFEL. We will now comment on possible saturation of the eCCDs with the very intense FEL pulses. Recent spin-polarized photoemission experiments at FLASH (Fognini et al., 2014) with a single-channel Mott detector have found that more than one electron per pulse reaches the detector of passivated implanted planar silicon (PIPS) which was prohibitive for single-pulse counting. However, these experiments were essentially angle- and energy-integrated, bringing to one detector the whole integral photoemission intensity. On the other hand, the most recent time-resolved HAXPES experiments at SACLA (Oloff et al., 2014) have shown that every pulse produces about 3 × 106 photoelectrons. For the actual iMott design, with scattering at the Au target characterized by an efficiency of the order of 10−1 of the total (inelastic and elastic) reflectivity and with the actual iris of the MLs, only about 5 × 104 electrons will pass to the eCCDs. They will distribute over more than 2.5 × 105 channels of our eCCD, leaving about one-fifth of an electron per channel. This number of events remains sufficiently low for single-pulse counting and thus energy resolution of the scattered electrons essential for the iMott operation. However, given the very approximate character of this estimate, we cannot rule out that certain pulse energy attenuation may still be necessary. In any case, the problem of single-pulse counting in iMott is less restrictive compared with that of space charge (Fognini et al., 2014), the main encumbrance of at FEL sources, which is relieved by an increase of the pulse repetition rate with simultaneous reduction of intensity of each pulse.
7. Conclusion
We have presented the concept of the multichannel electron spin detector iMott which combines imaging electron optics principles to achieve multichannel detection with Mott scattering as the spin selective process. The detector can be fitted to a standard (photo) electron analyzer to yield a spin-resolved image in the energy and angle coordinates, or to a microscope to yield an image in the spatial coordinates.
The iMott electron optics uses consecutive imaging principles: (i) The multichannel electron image from the analyzer or microscope focal plane is re-imaged by the electrostatic lens at an accelerating voltage of 40 kV onto the Au target. (ii) Quasi-elastic electrons bearing spin asymmetry of the Mott scattering are imaged by four magnetic lenses in two orthogonal planes onto energy-selective position-sensitive eCCDs to yield the multichannel spin asymmetry image. Ray-tracing calculations for a compact detector with RM = 100 mm fitted to a standard HSA have demonstrated an efficiency gain of 8.4 × 102 compared with the single-channel detector. Scaling up of the iMott dimensions dramatically improves the resolutions and electron optics transmission, pushing the gain to a colossal factor above 104 which becomes 1.4 × 104 already for RM = 200 mm.
The iMott concept has a few fundamental advantages: (i) The imaging electron optics accepts divergent electron sources typical of electron analyzers or microscopes. (ii) High accelerating voltage ensures almost ideal imaging properties of the EL stage, with replacement of the Au target by an eCCD allowing efficient spin-integrated measurements. (iii) The use of directly irradiated eCCDs with their simplicity, large etc.) to previously unthinkable cases of buried heterostructures actual for current device applications. Furthermore, the simultaneous spin detection and fast eCCDs readout enable efficient exploitation of the iMott detectors not only at synchrotron but also at X-ray FEL facilities.
and fast readout. (iv) Stability and simultaneous measurements of two spin components inherited from the Mott detectors. By virtue of the colossal efficiency gain, the iMott concept enables expansion of the spin-resolved spectroscopies from standard bulk and surface systems (Rashba effect, topological insulators, magnetic pairing in unconventional superconductors,Footnotes
1Detailed information on the PHOIBOS-150 analyzer is available at the SPECS GmbH website https://www.specs.de. Parameters of the electron trajectories were supplied by S. Maehl.
2Product information is available at https://www.hamamatsu.com/resources/pdf/ssd/s7170-0909_etc_kmpd1028e.pdf.
Acknowledgements
The authors thank J. Friso van der Veen for attracting their attention to the challenge of multichannel spin detection and for promoting discussions. We thank G. Schönhense for encouragement and fruitful scientific exchange. Logistic support of the iMott project by F. Nolting and T. Schmitt is gratefully acknowledged. We are obliged to S. Maehl for providing ray-tracing data of the PHOIBOS-150 analyzer, and S. Muff for the CAD drawings. The project is partially supported by the Swiss National Science Foundation.
References
Baibich, M. N., Broto, J. M., Fert, A., Nguyen Van Dau, F., Petroff, F., Etienne, P., Creuzet, G., Friederich, A. & Chazelas, J. (1988). Phys. Rev. Lett. 61, 2472–2475. CrossRef PubMed CAS Web of Science Google Scholar
Bonell, F., Hauet, T., Andrieu, S., Bertran, F., Le Fèvre, P., Calmels, L., Tejeda, A., Montaigne, F., Warot-Fonrose, B., Belhadji, B., Nicolaou, A. & Taleb-Ibrahimi, A. (2012). Phys. Rev. Lett. 108, 176602. Web of Science CrossRef PubMed Google Scholar
Cancellieri, C., Reinle-Schmitt, M. L., Kobayashi, M., Strocov, M., Schmitt, T., Willmott, P. R., Gariglio, S. & Triscone, J. M. (2013). Phys. Rev. Lett. 110, 137601. Web of Science CrossRef PubMed Google Scholar
Dil, J. H. (2009) J. Phys. Condens. Matter, 21, 403001. Web of Science CrossRef PubMed Google Scholar
Fognini, A., Salvatella, G., Michlmayr, G., Wetli, C., Ramsperger, U., Bähler, T., Sorgenfrei, F., Beye, M., Eschenlohr, A., Pontius, N., Stamm, C., Hieke, F., Dell'Angela, M., de Jong, M., Kukreja, S., Gerasimova, N., Rybnikov, V., Redlin, H., Raabe, J., Föhlisch, A., Dürr, A., Wurth, W., Pescia, D., Vaterlaus, A. & Acremann, Y. (2014). New J. Phys. 16, 043031. Web of Science CrossRef Google Scholar
Ghiringhelli, G., Piazzalunga, A., Dallera, C., Trezzi, G., Braicovich, L., Schmitt, T., Strocov, T., Betemps, R., Patthey, L., Wang, X. & Grioni, M. (2006). Rev. Sci. Instrum. 77, 113108. Web of Science CrossRef Google Scholar
Göhler, B., Hamelbeck, V., Markus, V., Kettner, M., Hanne, M., Vager, Z., Naaman, R. & Zacharias, H. (2011). Science, 331, 894–897. Web of Science PubMed Google Scholar
Hasan, M. Z. & Kane, C. L. (2010). Rev. Mod. Phys. 82, 3045–3067. Web of Science CrossRef CAS Google Scholar
Heinzmann, U. & Dil, J. H. (2012). J. Phys. Condens. Matter, 24, 173001. Web of Science CrossRef PubMed Google Scholar
Hellmann, S., Sohrt, C., Beye, M., Rohwer, T., Sorgenfrei, F., Marczynski-Bühlow, M., Kalläne, M., Redlin, H., Hennies, F., Bauer, M., Föhlisch, A., Kipp, L., Wurth, W. & Rossnagel, K. (2012). New J. Phys. 14, 013062. Web of Science CrossRef Google Scholar
Hoesch, M., Greber, T., Petrov, T., Muntwiler, M., Hengsberger, M., Auwärter, W. & Osterwalder, J. (2002). J. Electron Spectrosc. Relat. Phenom. 124, 263–279. Web of Science CrossRef CAS Google Scholar
Holzwarth, G. & Meister, H. J. (1964). Nucl. Phys. 59, 56–64. CrossRef CAS Web of Science Google Scholar
Jozwiak, C., Graf, J., Lebedev, G., Andresen, N., Schmid, N., Fedorov, A. V., El Gabaly, F. E., Wan, F., Lanzara, A. & Hussain, Z. (2010). Rev. Sci. Instrum. 81, 053904. Web of Science CrossRef PubMed Google Scholar
Kirschner, J. (1985). Polarized Electrons At Surfaces. Berlin: Springer. Google Scholar
Kirschner, J., Rebenstorff, D. & Ibach, H. (1984). Phys. Rev. Lett. 53, 698–701. CrossRef CAS Web of Science Google Scholar
Kobayashi, M., Muneta, I., Takeda, Y., Harada, Y., Fujimori, A., Krempaský, J., Schmitt, T., Ohya, S., Tanaka, M., Oshima, M. & Strocov, V. N. (2014). Phys. Rev. B, 89, 205204. Web of Science CrossRef Google Scholar
Kolbe, M., Lushchyk, P., Petereit, B., Elmers, B., Schönhense, G., Oelsner, A., Tusche, C. & Kirschner, J. (2011). Phys. Rev. Lett. 107, 207601. Web of Science CrossRef PubMed Google Scholar
Kutnyakhov, D., Elmers, H. J., Schönhense, G., Tusche, C., Borek, S., Braun, J., Minár, J. & Ebert, H. (2015). Phys. Rev. B, 91, 014416. Web of Science CrossRef Google Scholar
Kutnyakhov, D., Lushchyk, P., Fognini, A., Perriard, D., Kolbe, M., Medjanik, K., Fedchenko, E., Nepijko, S. A., Elmers, H. J., Salvatella, G., Stieger, C., Gort, R., Bähler, T., Michlmayer, T., Acremann, Y., Vaterlaus, A., Giebels, F., Gollisch, H., Feder, R., Tusche, C., Krasyuk, A., Kirschner, J. & Schönhense, G. (2013). Ultramicroscopym 130, 63. Web of Science CrossRef Google Scholar
Landolt, G., Eremeev, S. V., Koroteev, Y. M., Slomski, B., Muff, S., Neupert, T., Kobayashi, M., Strocov, M., Schmitt, T., Aliev, T., Babanly, M. B., Amiraslanov, I. R., Chulkov, E. V., Osterwalder, J. & Dil, J. H. (2012). Phys. Rev. Lett. 109, 116403. Web of Science CrossRef PubMed Google Scholar
Landolt, G., Schreyeck, S., Eremeev, S. V., Slomski, B., Muff, S., Osterwalder, J., Chulkov, J., Gould, C., Karczewski, G., Brunner, K., Buhmann, H., Molenkamp, H. & Dil, J. H. (2014). Phys. Rev. Lett. 112, 057601. Web of Science CrossRef PubMed Google Scholar
Mårtensson, N., Baltzer, P., Brühwiler, P. A., Forsell, J.-O., Nilsson, A., Stenborg, A. & Wannberg, B. (1994). J. Electron Spectrosc. Relat. Phenom. 70, 117–128. Google Scholar
Müller, N., Lischke, T., Weiss, T. & Heinzmann, U. (2001). J. Electron Spectrosc. Relat. Phenom. 114–116, 777–782. Google Scholar
Nordgren, J., Bray, J., Cramm, S., Nyholm, S., Rubensson, R. & Wassdahl, N. (1989). Rev. Sci. Instrum. 60, 1690. CrossRef Web of Science Google Scholar
Öhrwall, G., Karlsson, P., Wirde, M., Lundqvist, M., Andersson, P., Ceolin, D., Wannberg, B., Kachel, T., Dürr, H., Eberhardt, W. & Svensson, S. (2011). J. Electron Spectrosc. Relat. Phenom. 183, 125–131. Google Scholar
Okuda, T. & Kimura, A. (2013). J. Phys. Soc. Jpn, 82, 021002. Web of Science CrossRef Google Scholar
Okuda, T., Miyamaoto, K., Miyahara, H., Kuroda, K., Kimura, A., Namatame, H. & Taniguchi, M. (2011). Rev. Sci. Instrum. 82, 103302. Web of Science CrossRef PubMed Google Scholar
Okuda, T., Takeichi, Y., Maeda, Y., Harasawa, A., Matsuda, I., Kinoshita, T. & Kakizaki, A. (2008). Rev. Sci. Instrum. 79, 123117. Web of Science CrossRef PubMed Google Scholar
Oloff, L.-P., Oura, M., Rossnagel, K., Chainani, K., Matsunami, M., Eguchi, R., Kiss, T., Nakatani, Y., Yamaguchi, T., Miyawaki, J., Taguchi, M., Yamagami, K., Togashi, T., Katayama, T., Ogawa, K., Yabashi, M. & Ishikawa, T. (2014). New J. Phys. 16, 123045. Web of Science CrossRef Google Scholar
Petrov, V. N., Galaktionov, M. S. & Kamochkin, A. S. (2001). Rev. Sci. Instrum. 72, 3728. Web of Science CrossRef Google Scholar
Petrov, V. N., Grebenshikov, V. V., Andronov, A. N., Gabdullin, P. G. & Maslevtcov, A. V. (2007). Rev. Sci. Instrum. 78, 025102. Web of Science CrossRef PubMed Google Scholar
Petrov, V. N., Landolt, M., Galaktionov, M. S. & Yushenkov, B. V. (1997). Rev. Sci. Instrum. 68, 4385. Web of Science CrossRef Google Scholar
Schönhense, G. (1980). Phys. Rev. Lett. 44, 640–643. Google Scholar
Sinkovic, B., Tjeng, L. H., Brookes, L., Goedkoop, N., Hesper, R., Pellegrin, E., de Groot, E., Altieri, F., Hulbert, S. L., Shekel, E. & Sawatzky, G. A. (1997). Phys. Rev. Lett. 79, 3510–3513. CrossRef CAS Web of Science Google Scholar
Slomski, B., Landolt, G., Bihlmayer, G., Osterwalder, J. & Dil, J. H. (2013). Sci. Rep. 3, 1963. Web of Science PubMed Google Scholar
Strocov, V. N. (2010). J. Synchrotron Rad. 17, 103–106. Web of Science CrossRef CAS IUCr Journals Google Scholar
Strocov, V. N., Kobayashi, M., Wang, X., Lev, L. L., Krempasky, J., Rogalev, V. A., Schmitt, T., Cancellieri, C. & Reinle-Schmitt, M. L. (2014b). Synchrotron Radiat. News, 27(2), 31. CrossRef Google Scholar
Strocov, V. N., Schmitt, T., Flechsig, U., Patthey, L. & Chiuzbăian, G. S. (2011). J. Synchrotron Rad. 18, 134–142. Web of Science CrossRef CAS IUCr Journals Google Scholar
Strocov, V. N., Wang, X., Shi, M., Kobayashi, M., Krempasky, J., Hess, C., Schmitt, T. & Patthey, L. (2014a). J. Synchrotron Rad. 21, 32–44. Web of Science CrossRef CAS IUCr Journals Google Scholar
Suga, S., Sakamoto, K., Okuda, T., Miyamoto, K., Kuroda, K., Sekiyama, A., Yamaguchi, J., Fujiwara, H., Irizawa, A., Ito, T., Kimura, S., Balashov, T., Wulfhekel, W., Yeo, S., Iga, F. & Imada, S. (2014). J. Phys. Soc. Jpn, 83, 014705. Web of Science CrossRef Google Scholar
Tjeng, L. H., Sinkovic, B., Brookes, B., Goedkoop, N., Hesper, R., Pellegrin, E., de Groot, E., Altieri, F., Hulbert, S. L., Shekel, E. & Sawatzky, G. A. (1997). Phys. Rev. Lett. 78, 1126–1129. CrossRef CAS Web of Science Google Scholar
Tusche, C., Ellguth, M., Ünal, M., Chiang, C.-T., Winkelmann, A., Krasyuk, A., Hahn, M., Schönhense, G. & Kirschner, J. (2011). Appl. Phys. Lett. 99, 032505. Web of Science CrossRef Google Scholar
Vasilyev, D., Tusche, C., Giebels, F., Gollisch, H., Feder, R. & Kirschner, J. (2015). J. Electron Spectrosc. Relat. Phenom. 199, 10–18. Web of Science CrossRef CAS Google Scholar
Veenstra, C. N., Zhu, Z.-H., Raichle, M., Ludbrook, M., Nicolaou, A., Slomski, B., Landolt, G., Kittaka, S., Maeno, Y., Dil, Y., Elfimov, I. S., Haverkort, M. W. & Damascelli, A. (2014). Phys. Rev. Lett. 112, 127002. Web of Science CrossRef PubMed Google Scholar
Wannberg, B. (2009). Nucl. Instrum. Methods Phys. Res. A, 601, 182–194. Web of Science CrossRef CAS Google Scholar
Warwick, T., Chuang, Y.-D., Voronov, D. L. & Padmore, H. A. (2014). J. Synchrotron Rad. 21, 736–743. Web of Science CrossRef CAS IUCr Journals Google Scholar
Xu, N., Biswas, P. K., Dil, J. H., Dhaka, R. S., Landolt, G., Muff, S., Matt, S., Shi, X., Plumb, X., Radović, M., Pomjakushina, E., Conder, K., Amato, A., Borisenko, A., Yu, R., Weng, R., Fang, Z., Dai, X., Mesot, J., Ding, H. & Shi, M. (2014). Nat. Commun. 5, 4566. Web of Science PubMed Google Scholar