PRIGo: a new multi-axis goniometer for macromolecular crystallography
aSwiss Light Source, Paul Scherrer Institute, Villigen PSI, Switzerland, bCentre Suisse d'Electronique et Microtechnique SA, Neuchâtel 2002, Switzerland, and cDectris Ltd, Baden 5400, Switzerland
*Correspondence e-mail: vincent.olieric@psi.ch, meitian.wang@psi.ch
The Parallel Robotics Inspired Goniometer (PRIGo) is a novel compact and high-precision goniometer providing an alternative to (mini-)kappa, traditional three-circle goniometers and Eulerian cradles used for sample reorientation in macromolecular crystallography. Based on a combination of serial and parallel kinematics, PRIGo emulates an arc. It is mounted on an air-bearing stage for rotation around ω and consists of four linear positioners working synchronously to achieve x, y, z translations and χ rotation (0–90°), followed by a φ stage (0–360°) for rotation around the sample holder axis. Owing to the use of piezo linear positioners and active correction, PRIGo features spheres of confusion of <1 µm, <7 µm and <10 µm for ω, χ and φ, respectively, and is therefore very well suited for micro-crystallography. PRIGo enables optimal strategies for both native and experimental phasing crystallographic data collection. Herein, PRIGo hardware and software, its calibration, as well as applications in macromolecular crystallography are described.
Keywords: multi-axis goniometry; macromolecular crystallography; diffractometer; diffraction data; collection strategy; beamline endstation.
1. Introduction
The success of a macromolecular crystallography (MX) diffraction experiment depends on the accurate measurement of intensities and the completeness of the data. The main causes of failure are overlapping reflections (Dauter, 1999), radiation damage (Blake & Phillips, 1962; Ravelli & Garman, 2006; Garman & Weik, 2011) and noise originating from various sources (Holton & Frankel, 2010). Consideration and optimization of these parameters are crucial for obtaining the best outcome (Dauter, 2005, 2010). To this end, crystallographers have always employed data collection protocols involving crystal alignment using multi-axis goniometers. They are still routinely used on X-ray laboratory sources to obtain complete data sets at high resolution from small molecule crystals (Helliwell, 1992). At synchrotron MX beamlines, however, multi-axis goniometers were almost abandoned after the arrival of two-dimensional detectors, which restricted motions around the sample, due to the risk of collision. In addition, these goniometers never met the specifications imposed by the trend towards collecting on ever smaller crystals with ever smaller X-ray beam sizes. Their presence at third-generation synchrotrons was thus limited to few beamlines (Rosenbaum et al., 2006; Shi et al., 2006; Ascone et al., 2007) until collision issues were addressed with the development of an inverse mini-kappa goniometer head with dedicated control software at ESRF/EMBL Grenoble (Brockhauser et al., 2011). Several such systems have been installed at synchrotron beamlines (Brockhauser et al., 2013) but the vast majority of data sets remain collected from a single axis, primarily around ω.
Developed at the Swiss Light Source (SLS) at Paul Scherrer Institute, the Parallel Robotics Inspired Goniometer (PRIGo) (Glettig et al., 2009, 2011) is a new type of multi-axis goniometer with high-end features such as micrometer precision around the main rotation axis ω, a large collision-free angular range, reduced self-shadowing and low-maintenance device mechanics. Fully integrated into the beamline software architecture, it enables the realisation of advanced data collection protocols for both native and experimental phasing experiments in a user-friendly manner. Herein we present the geometric model, calibration and accuracy of the PRIGo goniometer as well as its integration at beamline X06DA (PXIII)1 at SLS. New opportunities for anomalous diffraction data collection are also presented.
2. Hardware and software description of PRIGo
Based on a combination of serial and parallel kinematics, PRIGo is a compact and precise multi-axis goniometer, which emulates the movements of an arc (Fig. 1). PRIGo utilizes linear and rotary piezo positioners (SLC-P-2490, SLC-P-24120 and SR-P-2013-S, SmarAct GmbH) driven by a combined stick slip and direct current principle, with built-in linear encoders operating in closed loop and enabling nanometer resolution and positional reproducibility <10 nm. Three of the four linear positioners are parallel to the ω rotation axis and connected by spherical joints to a tetrahedron-shaped platform, called the `Orion table' (Henein, 2000) (Fig. 2). The fourth positioner is inclined by an angle of 18° and connects a swing head supporting the φ rotational stage. PRIGo is mounted on an air-bearing rotary stage (Aerotech ABRS-250) used for the ω rotation with a 36-channel slip-ring in the center enabling unlimited rotary motion. Translations at the sample position, as well as the χ rotation, are achieved with synchronous movements of the four linear positioners.
PRIGo's control system was implemented in the Orchestra Control Engine2, an open-source real-time control framework running on RTAI/Linux, which provides powerful and stable performance in a low-cost package. The various software modules of PRIGo (Fig. 3) communicate with a simple and efficient set of commands, keeping the network overhead to a minimum. PRIGo's movements are described in the beamline coordinate system (see insert `measurement setup' on Fig. 4), which does not correspond to linear movements of the individual motors. The conversion between the beamline coordinates (x, y, z, ω, χ, φ) and the motor positions (four linear positioners, ω and φ rotation stages) is achieved by a forward and inverse kinematics calculation. Performed by the control system at every cycle of the control loop (every 1 ms), this calculation ensures synchronized and smooth movements of the linear positioners.
3. Calibration of PRIGo
The geometric model of PRIGo has 21 independents parameters, which correspond to PRIGo's real dimensions (Fig. 2). Most could be measured using a three-dimensional coordinate-measuring machine (Wenzel LKS) with a precision of ∼10 µm, yielding the relative position of the four linear positioners (positions B1 to B5 on Fig. 2). The parameters linked to the fourth linear positioner (slider 4 on Fig. 2), which have the largest effect on the geometric model, were then refined in an iterative manner by minimizing the sphere of confusion (SoC) for the χ rotation axis (0–90°). Such SoC measurements were performed using a precise stainless steel reference sphere (Ø = 12.7 mm) with form tolerance below 0.25 µm, representing the sample mounted on the PRIGo sample holder and three capacitive sensors (Probe Type: C23/B, Driver System: Elite Series CPL190, Lion Precision/IBS Precision Engineering) aligned with the beamline coordinate system (Fuchs et al., 2014) (Fig. 4). The same SoC measurements were also performed for the ω (0–360°) and φ (0–360°) rotation axes. We obtained a `geometric model' peak-to-peak SoC of <5 µm, 25 µm and 10 µm for ω, χ and φ, respectively (Fig. 4, left side). To further improve these results, we implemented an active correction of the four motor positioners for the rotations around the ω (when χ = 0°) and χ axes (no correction was implemented for the φ axis yet). The active correction consists of parameterizing the measured deviation for a rotation (here, ω and χ), i.e. fitting a function to the SoC measurements. The deviations for the ω and χ rotations are described by a sum of two sinusoids (six parameters) and by a quadratic polynomial (five parameters), respectively. Implemented in the PRIGo control system, the active correction around the ω axis is applied up to a speed of 180° s−1. To achieve this, it is necessary to interpolate the position of the air-bearing rotary stage (ABR) ω position, which is obtained via EPICS (Fig. 3) at a rate of 10 Hz only. This is done within the calibration module of the PRIGo control, knowing the last ABR ω position and the rotation speed. The active correction is applied around the χ axis up to a speed of 5° s−1.
Owing to the high resolution and repeatability of the SmarAct positioners as well as the reproducibility of PRIGo's mechanics, the SoC improved to <1 µm for ω and <7 µm for χ (Fig. 4, right side). This accuracy is valid within ±2.5 mm deviation along x, y and z from the ideal sample position. The calibration, together with the active correction, have proven to be extremely stable over time and were repeated only once after 15 months of operation, following the maintenance of two linear positioners.
4. Integration of PRIGo at beamline X06DA (PXIII) at SLS
PRIGo was installed at beamline X06DA (PXIII) at the Swiss Light Source in January 2012 and is now fully integrated into the new data acquisition user interface DA+ (E. Panepucci et al., unpublished). Its compact design and small space envelope at large χ angles (Maag et al., 2013) (Fig. 5) maximize the achievable angular range but self-shadowing on the detector and potential collisions with beamline devices such as the collimator or the beam stop still impose angular restrictions. Currently, PRIGo offers collision-free ω rotation with a χ angle up to 40°, more than 210° ω rotation with a χ angle between 40° and 60°, and more than 120° ω rotation with a χ angle of 90° (Fig. 6). The cryogenic nitrogen gas blowing directly on parts of PRIGo (mainly the head) can result in deflection of the device and movement at the sample position (∼10 µm). This is not of practical concern at beamline X06DA (PXIII) because the cryojet is oriented such that the affected positions are limited (light blue on Fig. 6), and in the cryo-stream for only a short time (∼20 s) in a typical 360° data collection at a speed of 2° s−1.
In addition, the PRIGo φ axis can also serve as a data collection axis (Table 1) thanks to continuous shutterless data acquisition capabilities offered by fast readout detectors such as the PILATUS (Broennimann et al., 2006; Henrich et al., 2009). Finally, PRIGo is compatible with both sample changer usage [an IRELEC CATS system with vial mounting is in place at X06DA (PXIII)] and manual transfer where a χ offset can facilitate both crystal mounting and recovery. Possible bump or collision on PRIGo upon sample exchange (from either robot or manual mounting) are recovered with an initialization procedure, which brings back all linear positioners to their reference positions. This procedure is completed in about 10 s.
|
5. Applications
Multi-axis goniometers like PRIGo are beneficial in a wide variety of cases including a simple reorientation to rotate around the crystal's best spot or data collection while avoiding contamination from satellite crystals. Other advantages are the unambiguous et al., 2013). For crystals diffracting to atomic resolution, data measured at a different χ angle to collect the `missing cusp' yield complete data in the highest-resolution shell after merging.
determination, alignment of a long crystallographic axis along the spindle in order to avoid overlapping reflections or minimization of the total oscillation range needed for a complete data set on a radiation-sensitive crystal (BrockhauserMulti-axis goniometers also enable optimal strategies for the measurements of anomalous diffraction data. In the case of single-wavelength et al., 1999), they provide an elegant way to acquire true redundant data (Debreczeni et al., 2003) with improved accuracy by changing the crystal orientation (here, multiple low-dose 360° data sets at various χ and φ angles). We successfully implemented such a strategy at beamline X06DA (PXIII) at SLS (Weinert et al., 2015) and solved, using only the anomalous signal of low-Z elements and a single crystal entity, a 266 kDa multiprotein–ligand tubulin complex (Prota et al., 2013) (T2R-TTL) (Fig. 7). With 2317 residues, this is the largest structure solved by native-SAD phasing to date.
(SAD) requiring high multiplicity (DauterIn the case of multiple-wavelength Bijvoet mates on the same image, i.e. with essentially similar X-ray dose and absorption effects (Kahn et al., 1985; Hendrickson et al., 1988). For this, we use XOalign (Legrand, 2009) with an orientation matrix from indexed diffraction patterns to calculate possible goniometer angles compatible with the beamline collision map (Fig. 6). Phasing experiments can also be designed to take advantage of the anisotropy of the anomalous signal at an (Bricogne et al., 2005; Schiltz & Bricogne, 2008, 2010). Here, the possibility offered by PRIGo to collect data around the φ axis and to orient this axis at a chosen angle with respect to the X-ray beam polarization is potentially a great advantage.
(MAD) experiments, PRIGo can be used to align an even-fold symmetry axis parallel to the spindle in order to collectIn conclusion, with a large angular range and micrometer precision, we have shown that the multi-axis goniometer PRIGo offers new possibilities for optimizing MX data collection protocols. By facilitating multi-axis data acquisitions, we aim at encouraging crystallographers to measure data with the highest accuracy, especially when attempting to obtain phases experimentally.
Footnotes
‡These authors contributed equally to this work.
§Present address: NSLS-II, Brookhaven National Laboratory, Upton, NY 11973, USA.
1X06DA (PXIII), https://www.psi.ch/sls/pxiii/.
2Orchestra, https://en.wikipedia.org/wiki/Orchestra_Control_Engine.
Acknowledgements
The authors would like to thank Markus Vitins and Stephan Maag for their help in manufacturing PRIGo, and Katja Bargsten, Natacha Oliéric, Andrea Prota and Michel Steinmetz for providing the T2R-TTL crystals.
References
Ascone, I., Girard, E., Gourhant, P., Legrand, P., Roudenko, O., Roussier, L. & Thompson, A. W. (2007). AIP Conf. Proc. 872, 874. Google Scholar
Blake, C. C. F. & Phillips, D. C. (1962). Vienna Int. At. Energ. Agency. pp. 183–191. Google Scholar
Bricogne, G., Capelli, S. C., Evans, G., Mitschler, A., Pattison, P., Roversi, P. & Schiltz, M. (2005). J. Appl. Cryst. 38, 168–182. Web of Science CrossRef IUCr Journals Google Scholar
Brockhauser, S., Ravelli, R. B. G. & McCarthy, A. A. (2013). Acta Cryst. D69, 1241–1251. Web of Science CrossRef IUCr Journals Google Scholar
Brockhauser, S., White, K. I., McCarthy, A. A. & Ravelli, R. B. G. (2011). Acta Cryst. A67, 219–228. Web of Science CrossRef CAS IUCr Journals Google Scholar
Broennimann, Ch., Eikenberry, E. F., Henrich, B., Horisberger, R., Huelsen, G., Pohl, E., Schmitt, B., Schulze-Briese, C., Suzuki, M., Tomizaki, T., Toyokawa, H. & Wagner, A. (2006). J. Synchrotron Rad. 13, 120–130. Web of Science CrossRef CAS IUCr Journals Google Scholar
Dauter, Z. (1999). Acta Cryst. D55, 1703–1717. Web of Science CrossRef CAS IUCr Journals Google Scholar
Dauter, Z. (2005). Prog. Biophys. Mol. Biol. 89, 153–172. Web of Science CrossRef PubMed CAS Google Scholar
Dauter, Z. (2010). Acta Cryst. D66, 389–392. Web of Science CrossRef CAS IUCr Journals Google Scholar
Dauter, Z., Dauter, M., de La Fortelle, E., Bricogne, G. & Sheldrick, G. M. (1999). J. Mol. Biol. 289, 83–92. Web of Science CrossRef PubMed CAS Google Scholar
Debreczeni, J. É., Bunkóczi, G., Ma, Q., Blaser, H. & Sheldrick, G. M. (2003). Acta Cryst. D59, 688–696. Web of Science CrossRef CAS IUCr Journals Google Scholar
Fuchs, M. R., Pradervand, C., Thominet, V., Schneider, R., Panepucci, E., Grunder, M., Gabadinho, J., Dworkowski, F. S. N., Tomizaki, T., Schneider, J., Mayer, A., Curtin, A., Olieric, V., Frommherz, U., Kotrle, G., Welte, J., Wang, X., Maag, S., Schulze-Briese, C. & Wang, M. (2014). J. Synchrotron Rad. 21, 340–351. Web of Science CrossRef CAS IUCr Journals Google Scholar
Garman, E. F. & Weik, M. (2011). J. Synchrotron Rad. 18, 313–317. Web of Science CrossRef IUCr Journals Google Scholar
Glettig, W., Henein, S., Breguet, J.-M. & Schulze-Briese, C. (2009). Proceedings of the 9th Euspen International Conference, pp. 192–195, San Sebastian, Spain. Google Scholar
Glettig, W., Vitins, M., Schwarb, A., Maag, S. & Schulze-Briese, C. (2011). Proceedings of the 11th Euspen International Conference, pp. 31–34, Como, Italy. Google Scholar
Helliwell, J. (1992). Macromolecular Crystallography with Synchrotron Radiation. Cambridge University Press. Google Scholar
Hendrickson, W. A., Smith, J. L., Phizackerley, R. P. & Merritt, E. A. (1988). Proteins, 4, 77–88. CrossRef CAS PubMed Web of Science Google Scholar
Henein, S. (2000). Thesis. EPFL, Switzerland. Google Scholar
Henrich, B., Bergamaschi, A., Broennimann, C., Dinapoli, R., Eikenberry, E. F., Johnson, I., Kobas, M., Kraft, P., Mozzanica, A. & Schmitt, B. (2009). Nucl. Instrum. Methods Phys. Res. A, 607, 247–249. Web of Science CrossRef CAS Google Scholar
Holton, J. M. & Frankel, K. A. (2010). Acta Cryst. D66, 393–408. Web of Science CrossRef CAS IUCr Journals Google Scholar
Kahn, R., Fourme, R., Bosshard, R., Chiadmi, M., Risler, J. L., Dideberg, O. & Wery, J. P. (1985). FEBS Lett. 179, 133–137. CrossRef CAS PubMed Web of Science Google Scholar
Legrand, P. (2009). xdsme, https://code.google.com/p/xdsme/. Google Scholar
Maag, S., Frommherz, U., Kotrle, G., Thominet, V., Welte, J., Pradervand, C., Thermer, R., Schnorf, D., Ellenberger, U. & Fuchs, M. R. (2013). J. Phys. Conf. Ser. 425, 012009. CrossRef Google Scholar
Prota, A. E., Bargsten, K., Zurwerra, D., Field, J. J., Díaz, J. F., Altmann, K.-H. & Steinmetz, M. O. (2013). Science, 339, 587–590. Web of Science CrossRef CAS PubMed Google Scholar
Ravelli, R. B. G. & Garman, E. F. (2006). Curr. Opin. Struct. Biol. 16, 624–629. Web of Science CrossRef PubMed CAS Google Scholar
Rosenbaum, G., Alkire, R. W., Evans, G., Rotella, F. J., Lazarski, K., Zhang, R.-G., Ginell, S. L., Duke, N., Naday, I., Lazarz, J., Molitsky, M. J., Keefe, L., Gonczy, J., Rock, L., Sanishvili, R., Walsh, M. A., Westbrook, E. & Joachimiak, A. (2006). J. Synchrotron Rad. 13, 30–45. Web of Science CrossRef CAS IUCr Journals Google Scholar
Schiltz, M. & Bricogne, G. (2008). Acta Cryst. D64, 711–729. Web of Science CrossRef CAS IUCr Journals Google Scholar
Schiltz, M. & Bricogne, G. (2010). Acta Cryst. D66, 447–457. Web of Science CrossRef CAS IUCr Journals Google Scholar
Shi, W., Robinson, H., Sullivan, M., Abel, D., Toomey, J., Berman, L. E., Lynch, D., Rosenbaum, G., Rakowsky, G., Rock, L., Nolan, B., Shea-McCarthy, G., Schneider, D., Johnson, E., Sweet, R. M. & Chance, M. R. (2006). J. Synchrotron Rad. 13, 365–372. Web of Science CrossRef CAS IUCr Journals Google Scholar
Weinert, T. et al. (2015). Nat. Methods, 12, 131–133. Web of Science CrossRef CAS PubMed Google Scholar
This is an open-access article distributed under the terms of the Creative Commons Attribution (CC-BY) Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.