short communications
A plug-and-measure diagnostic tool for
resolution and beam size of a soft X-ray beamlineaDepartment of Material Science and Engineering, National Chiao Tung University, Hsinchu 30013, Taiwan, and bNational Synchrotron Radiation Research Center, 101 Hsin-Ann Road, Hsinchu Science Park, Hsinchu 30076, Taiwan
*Correspondence e-mail: linhc@mail.nctu.edu.tw, jyyuh@nsrrc.org.tw
A compact multi-functional diagnostic tool has been installed for commissioning beamlines at the Taiwan Light Source. For a photon beam, the instrument can measure the −1 and dispersing photons of energies from 80 to 1200 eV were calibrated with a of 1011 photon s−1 at slit openings of 50 µm × 50 µm; a maximum of greater than 10000 was certified with an absorption spectra of gaseous samples. Pressure differences over four orders of magnitude were achieved between the ion chamber and the measurement chamber with a single capillary differential pumping stage. A sequence of beam profiles was measured by moving continuously in the vicinity of the nominal focal positions. This tool is useful in commissioning or trouble-shooting at most beamlines in a synchrotron facility.
energy resolution and beam size, consecutively with a photodiode or gold mesh, a windowless gas cell and a movable fluorescence screen viewport. Two gratings with ruling densities of 350 and 1000 lines mmKeywords: windowless ion chamber; soft X-ray beamline; gas phase absorption; dynamic beam size monitor.
1. Introduction
The Taiwan Light Source is a 3 GeV storage ring operating at 360 mA in a top-up mode. The average up time of the machine is over 98% annually (Luo et al., 2007); after prolonged operation during routinely scheduled beam time, carbon contamination (Boller et al., 1983; Chauvet et al., 2011) or misalignment of optical elements (mirrors, gratings) is unavoidable, so a quick test of the beamline performance is regularly necessary. The most important parameters such as resolution, alignment and size of the beam spot (focal position) must be fine-tuned using devices such as a photodiode or gold mesh, an ion chamber, a laser and a fluorescence screen, which are generally installed separately at various sections of the beamline. Some of these tools dedicated to beamline diagnostics have recently been developed to test the beam quality and sample alignment (Muro et al., 2009; Yuh, 2014), but none of those devices were integrated to measure the beamline parameters at the same time – above all, to dynamically monitor the beam quality (shape, focal position) along the beam path. We have implemented such a compact yet easily installed beamline diagnostic tool equipped with a detector, a windowless gas cell and a movable fluorescence screen viewport to characterize the beam quality simultaneously. The tool was applied to a newly built monochromator at beamline 08B that contains two gratings, with ruling densities of 350 and 1000 lines mm−1, to disperse photons of energy from 80 to 1200 eV with a measuring of 1 × 1011 photon s−1 and a greater than 10000. The variation in the size of the beam spots could be observed and was recorded in real time with a digital microscope. The focal point and the size of the beam spot by design can thus be verified.
2. Apparatus and measurements
The design (see Fig. 1a) comprises three major parts: a chamber, an ion chamber and an imaging chamber. The photon beam from the right side enters a multiport flux-measuring chamber [70 mm OD, (a)]; inside the chamber the sample holder (h) (ceramic or Teflon) is connected to a linear-motion feedthrough. On adjusting the feedthrough up and down, two measuring devices are selected: a photodiode (IRD XUV100) at the upper position and a gold mesh at the lower position. The beam passes through the capillary (b) into the ion chamber; absorption by a gas generates the currents measured and recorded with a current meter. The photon beam is eventually dumped onto a fluorescence screen viewport (f) at the end of the retractable welded bellows (d), which generated a bright spot on the screen. On adjusting the length of the bellows, the beam profile at various positions is sectioned and a sequence of images is recorded with a digital microscope. A photograph of the system is displayed for comparison in Fig. 1(b); the system is installed on a pumping cart. For our tests, using a turbomolecular pump (450 l s−1) a pressure of 10−7 torr within 2–3 h was attained. Baking was optional, but cycles of flushing the system with dry nitrogen were essential. The installation was completed within 1 h after the experimental endstation of the user was removed. A transmission mode (without removing the endstation) is practical under the following conditions: the length of the working space for the system should exceed 600 mm with the fluorescence screen removed and the can be measured only with a highly transmitting gold mesh.
An ion chamber is widely used to determine the absolute intensity of extreme ultraviolet light (Samson, 1964) and in photon absorption spectra experiments (Ho, 1998), in particular, to verify the total instrumental resolution of the beamline (Domke et al., 1992; Chung et al., 1995; Song et al., 2006). Fig. 1(a) shows the structure of a modified ion chamber (c) decreased in size; it consists of a standard nipple (70 mm OD, length 100 cm) and three flange ports (34 mm OD). In its interior are two curved electrode plates (g) to collect the currents with a potential difference supplied at 40 V. The back-fill pressure in the ion chamber is 7–18 mtorr. A pressure difference between the ion chamber (10−3 torr) and the measurement chamber (10−7 torr) was achieved by isolating with a capillary (diameter 1 mm, length 80 mm), which was optimized to diminish the conductance of the gas flow. To visualize the variation of the spot size in the x–y plane, a fluorescence screen moved continuously in the z direction with a retractable stage (102 mm) made of a welded bellows. The fluorescence screen viewport (70 mm OD) was uniformly coated with CdSe (material with a small work function), which, when struck with the intense beam of light from the synchrotron, clearly displayed the beam profile on the screen which was captured with a digital microscope from the atmosphere side. The spatial resolution at the image plane was defined with a CMOS detector (2 Mpixel, 2.55 mm × 1.9 mm); the pixel size is estimated to be less than 10 µm, which is sufficient for sampling a beam of size on the millimeter scale. The translational resolution along the beamline axis (z) is within 1 mm through manual adjustment, but can be improved with a stepping motor.
3. Results
The photocurrents of the photodiode were converted to a shows the of photon energy measured with two gratings; the maximum densities at 170 and 700 eV reached 6 × 1011 and 8 × 1010 photon s−1, respectively. Two pronounced absorption dips near 270 and 540 eV arose from the strong absorption of carbon and oxide contamination of the optical surfaces. The range covering from 80 to 420 eV and from 250 to 1200 eV was tested with four gases, Kr, CO, N2 and Ne, for an analysis of the line width and energy calibration. Here, only the of the nitrogen K-edge is presented. The spectrum in Fig. 3 was fitted with seven Voigt functions for the line of best fit after subtraction of a linear background. The Lorentzian and Gaussian widths represent the core lifetime broadening and the total instrumental resolution (slit opening, slope error, aberration and misalignment). The best-fitted widths of the first vibrational state are 127.6 meV and 38 meV for Lorentzian and Gaussian shapes, respectively. By definition, the (E/ΔE) of the monochromator is greater than 10000, which is comparable with the results reported by Domke et al. (1992) and Prince et al. (1999). To verify the focal position, we sectioned the beam spot in the x–y plane with a continuously moving screen along the z axis. Fig. 4(a) shows nine snapshots, captured with a digital microscope, of the beam profile at 400 eV in the vicinity of the nominal focal point. An analysis of their detailed beam profiles (grey scale of each pixel) with a Gaussian fit in the vertical direction (y direction) is shown in the inset of Fig. 4(b). The beam spot at the focal point has a vertical dimension of 0.55 mm and a horizontal dimension of 2.0 mm with both slits set at 50 µm.
on normalization of the storage ring current (360 mA) and the curve of of the photodiode. Fig. 24. Discussion
For maintaining a beamline, this tool is useful both for commissioning a new beamline and after restoration of the optical components (grating, mirror). There is no point yet in installing this tool permanently in each beamline; its highly mobile design and ease of operation should be the major benefits. Compared with our early designs (Chung et al., 1995; Song et al., 2006), the mass of the new ion chamber is greatly decreased from 10 kg to 0.5 kg and the length from 28 cm to 13 cm; these features facilitate the easy installation. The most challenging task was to pass the synchrotron beam through the tiny holes of the capillary; this problem was solved with the aid of a fluorescent material coated on the wedge surfaces of the capillary. In the analysis of the beam profile in Fig. 4(b), there are sudden decreases of intensity (indicated with arrows) that were caused by the blurred (horizontal) lines [Fig. 4(a)]. These millimeter-scale rulings attached from the atmosphere side caused a shadowing effect on the digital microscope. In contrast, the intensity truncation in Fig. 4(b) is due to the saturation of the CMOS sensor struck by the intense beam from the synchrotron (the detector was blinded). The gradually decreasing beam profile is attributed to the slow converging power of the refocusing mirror; there is an uneven number of data points at the right side of the inset in Fig. 4(b) as the presumed focal position is not in the middle of the translation range of the bellows. The position of the focal point is, nevertheless, indicated by the minimum of the fitting curve. When this tool is applied for a smaller beam such as from an undulator beamline, the magnification of the microscope should be increased.
In conclusion, we have developed a compact beamline diagnostic tool that combines the abilities to measure the s to π* indicates that our system is capable of enabling gas-phase experiments for beamline diagnostic purposes. The installation is easy and saves space; it might also be beneficial for applications outside the synchrotron community with having a lamp in vacuum and a small divergence.
and spot size in real time. The small width of the lines in the vibrational structure in the of the nitrogen transition 1References
Boller, K., Haelbich, R.-P., Hogrefe, H., Jark, W. & Kunz, C. (1983). Nucl. Instrum. Methods Phys. Res. 208, 273–279. CrossRef CAS Web of Science Google Scholar
Chauvet, C., Polack, F., Silly, M. G., Lagarde, B., Thomasset, M., Kubsky, S., Duval, J. P., Risterucci, P., Pilette, B., Yao, I., Bergeard, N. & Sirotti, F. (2011). J. Synchrotron Rad. 18, 761–764. Web of Science CrossRef CAS IUCr Journals Google Scholar
Chung, S.-C., Chen, C.-I., Tseng, P.-C., Lin, H.-F., Dann, T.-E., Song, Y.-F., Huang, L.-R., Chen, C.-C., Chuang, J.-M., Tsang, K.-L. & Chang, C.-N. (1995). Rev. Sci. Instrum. 66, 1655–1657. CrossRef CAS Web of Science Google Scholar
Domke, M., Mandel, T., Puschmann, A., Xue, C., Shirley, D. A., Kaindl, G., Petersen, H. & Kuske, P. (1992). Rev. Sci. Instrum. 63, 80–89. CrossRef CAS Web of Science Google Scholar
Ho, G. H. (1998). Chem. Phys. 226, 101–111. Web of Science CrossRef CAS Google Scholar
Luo, G. H. et al. (2007). AIP Conf. Proc. 879, 13–16. CrossRef CAS Google Scholar
Muro, T., Kato, Y., Kinoshita, T. & Watanabe, Y. (2009). J. Synchrotron Rad. 16, 595–596. Web of Science CrossRef IUCr Journals Google Scholar
Prince, K. C., Vondráček, M., Karvonen, J., Coreno, M., Camilloni, R., Avaldi, L. & de Simone, M. (1999). J. Electron Spectrosc. Relat. Phenom. 101–103, 141–147. Web of Science CrossRef CAS Google Scholar
Samson, J. A. R. (1964). J. Opt. Soc. Am. 54, 6–15. CrossRef CAS Web of Science Google Scholar
Song, Y. F., Yuh, J.-Y., Lee, Y.-Y., Chung, S. C., Huang, L. R., Tsuei, K.-D., Perng, S. Y., Lin, T. F., Fung, H. S., Ma, C.-I., Chen, C. T. & Tsang, K. L. (2006). Rev. Sci. Instrum. 77, 085102. Web of Science CrossRef Google Scholar
Yuh, J.-Y. (2014). J. Synchrotron Rad. 21, 1213–1214. Web of Science CrossRef IUCr Journals Google Scholar
© International Union of Crystallography. Prior permission is not required to reproduce short quotations, tables and figures from this article, provided the original authors and source are cited. For more information, click here.